A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Similar documents
Orbitrap GC-MS: An Opportunity to Help Address the Challenges of Chlorinated Paraffins Analysis

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

APPLICATION NOTE Authors Jane Cooper, 1 Jordi Parera, 2 Esteban Abad, 2 Richard Law,¹ and Cristian Cojocariu 1

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Multi-residue analysis of pesticides by GC-HRMS

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Automated and accurate component detection using reference mass spectra

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT

ApplicationNOTE THE ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBS) BY GC-HIGH RESOLUTION MASS SPECTROMETRY USING THE MICROMASS AUTOSPEC ULTIMA NT

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

A New Approach to the Analysis of Chlorinated Paraffins by Gas Chromatography Quadrupole Time of-flight Mass Spectrometry

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

U.S. EPA Method 8270 for multicomponent analyte determination

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

GC Analysis of Polybrominated Flame Retardants Application

Determination of urea in ultrapure water by IC-MS/MS

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Two Years On: The Impact of Orbitrap GC-MS in Food Analysis

GC-CI-MS analysis of TMS derivatives

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

GAS CHROMATOGRAPHY (GC)

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Kendrick mass = IUPAC mass x (14/ ) (1) Kendrick mass defect = nominal Kendrick mass exact Kendrick mass (2)

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Analysis of alkylphenols using GC-MS/MS and automated SRM development

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

GC/MS/MS Dioxin and Dioxin-like PCBs in Feed and Food. Jessica Westland, LSAG Applications Chemist

Elemental Analysis: CHNS/O characterization of polymers and plastics

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

Highly Sensitive and Rugged GC/MS/MS Tool

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

Dionex IonPac AS28-Fast-4µm column

Defining quality standards for the analysis of solid samples

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Discovery of Emerging Disinfection By-Products in Water Using Gas Chromatography Coupled with Orbitrap-based Mass Spectrometry

Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

The key to your laboratory s success

Elemental Analysis: NCS characterization of paper

PosterReprint INTRODUCTION

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

ENVIRONMENTAL analysis

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Determination of trace anions in concentrated hydrofluoric acid

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Expanded capabilities in ammonium and amine detection

Organochlorine Pesticides by GCxGC-ECD

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Transcription:

TECHNICAL NOTE 39 A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins Author Cristian Cojocariu Thermo Fisher Scientific, Runcorn, UK Keywords Exactive GC, Persistent Organic Pollutants (POPs), Short Chained Chlorinated Paraffins (SCCPs), environment, emerging contaminants, sensitivity, mass accuracy, high resolution, Orbitrap technology, gas chromatography Introduction Short Chained Chlorinated Paraffins (SCCPs) are emerging contaminants that, once released, will remain in the environment for long periods of time with the potential to bioaccumulate in living organisms. SCCPs are intentionally manufactured and used as lubricants and coolants in the metal processing industry or as plasticizers and flame retardants in plastic products. Chronic exposure to SCCPs is believed to have harmful and irreversible effects for humans and the environment. 1 As a consequence, SCCPs are listed in the Stockholm convention as chemicals with potential adverse effects and their production and use in Europe is restricted and regulated. 2 Detection and quantification of SCCPs poses analytical challenges due to the fact that these compounds are present in the environment at low levels, as very complex isomeric mixtures and are difficult to separate chromatographically. Although there is no consensus for the use of a validated analytical procedure for the routine monitoring of SCCPs in environmental samples, there are several analytical methods that are used to detect and quantify SCCPs. Details of these methods and their limitations are listed in Table 1.

Table 1. Current analytical methodology used for the analysis of SCCPs. Carbon skeleton analysis by GC-FID or GC-MS Details: Uses Pd catalyst held in the gas chromatograph injector to simultaneously dechlorinate the CPs and separate the resulting alkanes Disadvantages: Lack in sensitivity and selectivity No information on the degree of chlorination of the SCCPs can be achieved GC-ECD Details: GC coupled to electron capture detector, sensitive for halogenated compounds Disadvantages: Relatively non-specific Interferences from other halogenated compounds GC-NCI-MS Details: Uses soft ionisation (negative chemical ionisation) with methane and/or dichloromethane Disadvantages: Low resolution GC-MS nominal mass interferences from higher chlorinated PCBs, toxaphenes and chlordanerelated compounds, have similar molecular masses In this study, the performance of a novel bench top, high resolution accurate mass Orbitrap-based GC-MS was tested for the analysis of SCCPs. System performance was tested using full-scan acquisition and simple instrumental setup. The experiments performed focused on assessing the sensitivity, linear dynamic range, selectivity and analytical precision for the analysis of two SCCPs technical mixtures. Both electron ionization (EI) and negative chemical ionization (NCI) were used and the results compared and discussed. Experimental In the experiments described here, a Thermo Scientific Exactive GC Orbitrap mass spectrometer was coupled to a Thermo Scientific TRACE 13 Gas Chromatograph for gas-phase separation of target compounds, achieved on a Thermo Scientific TraceGOLD TG-SilMS m x. mm x. µm column (P/N 296-). Injection of liquid samples was performed automatically using a Thermo Scientific TriPlus RSH autosampler. The Exactive GC was tuned and calibrated in under one minute using PFTBA to achieve the best ion transmission and sub-ppm mass accuracy. The mass spectrometer was operated in fullscan using k mass resolution (measured as FWHM at ). Lockmass corrected data was processed with Thermo Scientific TraceFinder software. Additional details regarding the GC and MS conditions are given in Table 2. 2

Table 2. Gas chromatography and mass spectrometers analytical parameters. TRACE 13 GC System Parameters Injection Volume (µl): 2. Liner: Inlet ( C): 2 Inlet Module and Mode: Splitless Carrier Gas, (ml/min): He, 1.2 Oven Temperature Program Temperature 1 ( C): Hold Time (min): 2. Temperature 2 ( C): 3 Rate ( C/min): Hold Time (min): 4. LinerGOLD, single taper (P/N:3A344-UI) Electron Ionization MS Parameters Transfer line ( C): 2 Ionization type: EI Ion source ( C): 2 Electron energy (ev): Acquisition Mode: Full-scan Mass range (Da): -6 C-Trap voltage (V):. Mass resolution (FWHM at ): k, k and k Lockmass (): 7.32 281.114 3.6993 Negative Chemical Ionization MS Parameters Transfer line ( C): 2 Ionization type: NCI Ion source ( C): Reagent gas and flow (ml/min): Methane, 1.2 Electron energy (ev): Electron Lens Voltage (V): Emission current (µa): 1 C-Trap voltage: 2. Acquisition Mode: Full-scan Mass range (Da): - Mass resolution (FWHM at ): k Lockmass (): 234.944 3

Complete separation of SCCPs is very difficult due to the very high numbers of isomers and homologues with similar physiochemical properties (Figure 1). Due to the fragmentation obtained in EI, it is often difficult to find homologue specific ions with sufficient intensity for use as quantification masses, highly efficient electron ionization can be used to detect and quantify SCCPs. To provide even higher selectivity for homologue groups, GC-MS with negative chemical ionization is the method preferred by many laboratories for the detection and quantification of SCCPs. NCI allows for sensitive and selective detection of SCCPs by using ions characteristic for various homologue groups in a mixture (in particular [M-Cl]-, [M-HCl]-). Using NCI, the SCCP congeners in the two technical mixtures were easily separated based on the number of chlorine substitutes for a certain carbon chain length and according to the number of carbon atoms and chlorine atoms for various carbon chain lengths. Examples of congener specific extracted ion chromatograms are given in Figure 2. SCCP 63% chlorine SCCP.% chlorine RT:. - 17. RT: 9.68 RT:. - 17. RT: 7.89 8.43 EIC 2.28 (C H 7 CI) EIC 2.28 (C H 7 CI) 2 4 6 8 12 14 16 Time (min) 2 4 6 8 12 14 16 Time (min) Figure 1. Extracted ion chromatogram of the fragment ion 2.28 corresponding to C H 7 Cl 7, ±ppm extraction window) showing the chromatographic complexity of two SCCP technical mixtures (63% and.% chlorine). Data acquired in EI, full-scan, using k resolution. 4 RT: 6.48-14.43 C H Cl 6 C 11 H 17 Cl 6 C 11 H 16 Cl 7 C 12 H 18 Cl 7 9.72 TIC.22..69.73 11.21 11. 6. 7. 7. 8. 8. 9. 9... 11. 11. 12. 12. 13. 13. 14. Time (min) Figure 2. High resolution accurate mass selectivity demonstrated for the C -C 13 63% technical mix acquired in NCI at k resolution. Examples of extracted ion chromatograms for individual homologues with various chlorination degrees are shown. 11. 11.73 C 12 H 17 Cl 8 C 13 H 19 Cl 8 C 13 H 18 Cl 9 2.E8 TIC MS 2.E7 346.922 3.E7 3.948 3.E7 394.3 2.E7 8.91872 2.E7 444.897 2.E7 8.897 2.E7 492.283

Sensitivity The sensitivity of the Exactive GC was tested by injecting low concentration solvent standards (in cyclohexane) prepared by a serial dilution of the two SCCP technical mixtures (Figure 3). The limit of detection varies depending on the relative concentration of a particular congener in a technical mixture. From the EI data, the instrumental detection limits (IDL) were ~ pg/µl (calculated as total homologues response for each of the two SCCPs technical mixtures). In addition, the NCI data demonstrates that IDLs as low as 3 pg/µl can be obtained for individual homologue groups (Table 3). Table 3. Peak area repeatability calculated as %RSD from n= repeat injections at pg/µl level for two SCCP congeners acquired using NCI. IDL calculated taking into account the Student s-t critical values for the corresponding degrees of freedom (99% confidence). inj. no 492.46 (C 13 H 18 Cl 9 ) 8.8936 (C 13 H 19 Cl 8 ) 1 7881 18232 2 8221 142 3 7941 13613 4 781911 1363928 7797 13218 6 731874 128 7 7611 483 8 749797 1284342 9 737987 17718 7772 1286412 mean 7661 1316822 StDev 27217 %RSD 3. 4.1 IDL 2. 2.9 RT:. - 17. pg pg pg pg 2 4 6 8 12 14 16 Time (min).e4 = 2.27-2.29 MS ppbei.e4 = 2.27-2.29 MS ppbei.e4 = 2.27-2.29 MS ppbei.e4 = 2.27-2.29 MS ppbei Figure 3. Extracted ion chromatograms ( 2.28, mass window ±ppm) representing the total SCCP C -C 13.% chlorine homologues. Peak area response at,, and pg on column concentrations) are shown. Linearity and dynamic range SCCPs linearity and dynamic range was assessed for each SCCP technical mix (63% and.%) using the following dilution series: 1,,,,, 2,,, and pg/µl (in cyclohexane). This test was performed using both EI and NCI and examples are given in Figure 4. The coefficient of determination was >.99 indicating excellent linearity across this concentration range.

EI NCI 2.28 (±ppm) SCPP.% chlorine homologues 346.92699 (±ppm) C H C I6 [M-CI] - 3.E-8 3E-8 y =.3x +.8 R =.99879 1.E+8 1.E+8 y = 94x + 217 R =.99943 2.E-8 8.E+7 2E-8 6.E+7 1.E-8 1E-8 4.E+7 E-9 2.E+7 Concentration [pg/µl].e+ Concentration [pg/µl] Figure 4. Example of linearity obtained for the C -C 13.% chlorine homologues across pg/µl in EI and NCI at k resolution. Moreover, selected individual homologue masses showed excellent linearity when acquired in NCI with R 2 values >.999 (Figure ). 6.E+7 y = 163x - 293814 R² =.9998.E+7 4.E+7 y = 121x - 3111 R² =.9997 peak area counts 3.E+7 312.96 346.92 3.8886 2.E+7 y = 2948.3x - 11 R² =.9996 416.8466 1.E+7 y = 16.6x - 122811 R² =.9998.E+ Concentration [pg/µl] Figure. Example of linearity obtained individual SCCP congeners in the C -C 13 63% technical mixture across 1 pg/µl concentration range. Data acquired in full-scan, NCI at k resolution. 6

Selectivity through high mass resolution High resolution and high mass accuracy enable excellent selectivity and specificity. This is particularly important for full-scan data when background masses such as matrix ions, other organochloro contaminants or ions from other SCCP homologous group interfere with the target masses, leading to erroneous quantification. As demonstrated in Figure 6 for a SCCPs % chlorine standard spiked with polychlorinated biphenyls, a resolving power of k is not sufficient to differentiate between an SCCP ion ( 3.3121, C 11 H 16 Cl 3 ) and a PCB interference ( 3.1733, C 14 H 11 Cl 2 ). Instead, at k resolution a single ion is detected, which in turn will significantly affect the peak area determination and precise estimation of SCCP concentration. At low resolving power, the extracted peak area of the target SCCP ion 3.3121 is significantly lower than those obtained at k or k resolution due to higher errors in mass measurements (ppm). To achieve the sub-ppm mass accuracy and the selectivity required for consistent separation and quantification of target compounds, resolving powers of >k are needed. RT: 7.48 -.43 RT: 9.2 MA: 112874 7. 8. 8. 9. 9.. 3.32 C 11 H 16 Cl 3 = 3.3121 6.68 ppm 2.99 3. 3.1 3.2 3.3 3.4 3. RT: 9.2 MA: 16413 7. 8. 8. 9. 9... 3.3128 C 11 H 16 Cl 3 = 3.3121.26 ppm SCCP mass PCB mass 3.1669 C 14 H 11 Cl 2 = 3.1733-2. ppm k k k 2.99 3. 3.1 3.2 3.3 3.4 3. RT: 9.3 MA: 6876 7. 8. 8. 9. 9... 3.31 C 11 H 16 Cl 3 = 3.3121.17 ppm PCB mass 3.19 C 14 H 11 Cl 2 = 3.1733 -.96 ppm SCCP mass 2.99 3. 3.1 3.2 3.3 3.4 3. Figure 6. Selectivity enhancement by using narrow mass tolerance windows is possible at high resolving power. The effect of increasing resolving powers of the mass accuracy and peak area of a SCCP ion 2.3121 is demonstrated for a % SCCPs sample spiked with polychlorinated biphenyls. Data acquired in full-scan, EI. 7

ppb level (63% SCCP, EI), ppb level (63% SCCP, EI) ppb63ei RT: 6.27-9.4 AV: 1419 FTMS + p EI Full ms [.-6.] 2.3122 R=922 2.2316 R=89412 C H 7 Cl = 2.28.893 ppm 2.67 R=92394 2.4183 R=921 2.6313 R=339 2.1 2.2 2.3 2.4 2. 2.6 2.7 ppm63ei RT: 6.27-9.4 AV: 1429 FTMS + p EI Full ms [.-6.] 2.2313 R=91363 C H 7 Cl = 2.28.486 ppm 2.1877 R=88428 2.4643 R=7 C 8 H 6 = 2.46.248 ppm 2.1 2.2 2.3 2.4 2. 2.6 2.7 Figure 7. Mass accuracy remained at <1 ppm level irrespective of the by compound concentration as demonstrated for SCCPs fragment ion 2.28 (C H 7 Cl) acquired in full-scan, EI at, mass resolution. The exact mass resolution, as well as the mass precision (ppm), used are annotated to each measured ion. Maintaining sub-ppm mass accuracy at low and high concentrations Outstanding mass accuracy (<1 ppm) was maintained across all compound concentrations (Figure 7). This is essential, as any compromise in accuracy of mass measurements can result in false identification, erroneous quantification and interferences from matrix ions. In addition, with increased selectivity, the NCI data demonstrates that LOQs as low as 3 pg/µl can be reliably obtained for individual homologue groups. Excellent linearity was obtained across a total SCCP mixture concentration range of 1, pg/µl, making the Exactive GC an ideal quantification tool. Conclusions These preliminary results demonstrate that the Exactive GC is a potential solution to address the difficult challenges related to the detection and quantification of SCCPs due to the excellent sensitivity, linearity and selectivity and in combination with an uncomplicated instrumental setup. From the EI data, low instrumental LODs of ~ pg/µl, calculated as total homologues response for each of the two SCCPs technical mixtures, can be easily obtained. Using NCI it is possible to selectively separate C alkanes chains with various chlorination degrees making quantification of homologues with similar Cl content achievable. Find out more at thermofisher.com/exactivegc The high resolving power of the Exactive GC facilitates sub-ppm mass accuracy at low and high concentrations, essential for achieving enough selectivity to confidently separate SCCP specific low mass ions from the interfering background ions (in EI), or higher masses (in NCI), for various SCCP homologue groups. References 1. Geng N, Zhang H, Zhang B, Wu P, Wang F, Yu Z, Chen J. Effects of short-chain chlorinated paraffins exposure on the viability and metabolism of human hepatoma HepG2 cells. Environ Sci Technol, Mar 3;49():76-83. doi:.21/ es2x. 2. European Commission, Commission Regulation (EU) No 19, Off. J. Eur. Union, L 9 1-4, 12. 17 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. TN39-EN 17S