δ 13 C of DIC and carbonate samples: comparison of traditional mass spectrometry methods with an infrared spectrometry method

Similar documents
Determination of Carbon d13c isotopic composition from DIC samples

Thermo Scientific GasBench II

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

of mass spectrometry

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Elemental Analysis: NCS characterization of paper

Advanced ICP-MS techniques for overcoming interferences in LA-ICP-MS bioimaging

Expanded capabilities in ammonium and amine detection

Defining quality standards for the analysis of solid samples

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer

EA-IRMS: Fast and Precise Isotope Analysis of Liquids on a Delta V Isotope Ratio MS with a High Temperature Conversion Elemental Analyzer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

ProPac WCX-10 Columns

Introduction to Fourier Transform Infrared Spectroscopy

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Thermo Scientific 253 Plus 10 kv Isotope Ratio MS. Utilize the Gold Standard. in isotope ratio MS. Robustness Sensitivity Precision

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Fast methods for the determination of ibuprofen in drug products

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Introduction to Fourier Transform Infrared Spectroscopy

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Speciation and δ 34 S analysis of volatile organic compounds in crude oil by GC-MC-ICP-MS using the Thermo Scientific GCI 300 Interface

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Speed and flexibility

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Determination of trace anions in concentrated hydrofluoric acid

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Fusion UV/Persulfate TOC Analyzer

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Calibration strategies for FTIR and other IRIS instruments for accurate δ 13 C and δ 18 O measurements of CO 2 in air

LCTech GmbH Chromatographie

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

*Correspondence to:

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Method for Nitrite determination on Low-Range Samples

Fusion UV/Persulfate TOC Analyzer

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of Volatile Substances Proof of Food Adulteration

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Determination of urea in ultrapure water by IC-MS/MS

Dionex IonPac AS28-Fast-4µm column

Elemental Analysis: CHNS/O characterization of polymers and plastics

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

for XPS surface analysis

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Advances in Sample Preparation for Accelerated Solvent Extraction

IRMS. perspective. Isotope Ratio Mass Spectrometry.

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

HELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer

METHOD VALIDATION REPORT

DI 3500 Discrete Aoalvzer

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients

ION CHROMATOGRAPHY SYSTEM S 150

Thermo Scientific DELTA V Advantage DELTA V Plus

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

AcroSeal Packaging. Your Solution for Air- and Moisture-Sensitive Organometallic Reagents

Application Note PB 401. Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process

Monitoring Protein PEGylation with Ion Exchange Chromatography

Ideal and powerful solution

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

HPLC Praktikum Skript

AppNote 5/2008. Automation of Sample Preparation Steps using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Transcription:

APPLICATION NOTE 30486 δ 13 C of DIC and carbonate samples: comparison of traditional mass spectrometry methods with an infrared spectrometry method Authors Danijela Smajgl, 1 Magda Mandic, 1 Dirk Nürnberg, 2 Sebastian Fessler, 2 Albrecht Leis 3 1 Thermo Fisher Scientific, Bremen, Germany 2 GEOMAR Helmholtz Centre for Ocean Research Kiel, Kiel, Germany 3 JR-AquaConSol GmbH, Graz, Austria Keywords Delta Ray Connect, IRIS, Carbonates, Dissolved Inorganic Carbon (DIC), IRMS Goal To compare measurements from an infrared spectrometry method for δ 13 C conducted with a Thermo Scientific Delta Ray IRIS with URI Connect with those from traditional mass spectrometry methods. Introduction Isotope ratio mass spectrometry (IRMS) is a widely used technique for measurements of stable isotopic compositions. It is employed in different applications for environmental and geochemical studies. One example is paleo-environmental reconstruction based on carbonate measurements. In the past, high throughput and high precision of the measurements could only be achieved with a Thermo Scientific Kiel Carbonate Device connected to the IRMS via the dual viscous flow inlet system or as a second option with the Thermo Scientific GasBench II with the Carbonate Option. Now, in addition to IRMS systems, Thermo Fisher Scientific offers a laser-based and field deployable instrument: the Thermo Scientific Delta Ray Isotope

Ratio Infrared Spectrometer (IRIS) with the Universal Reference Interface (URI) Connect (Figure 1). Because of its size and robustness, the Delta Ray IRIS with URI Connect can be moved and installed much easier than mass spectrometers and it offers a possibility of direct measurements in the field. Here we present a sample preparation and sample processing method for determination of the isotopic composition of carbon (δ 13 C) in dissolved inorganic carbon (DIC) and carbonate material with the Delta Ray IRIS with URI Connect and provide a comparison with traditional mass spectrometry methods. Figure 1. Delta Ray IRIS with URI Connect and Teledyne Cetac ASX-7100 Autosampler with heated sample rack, and in the background, Kiel IV Carbonate Device with Thermo Scientific MAT 253 Mass Spectrometer Mid-infrared isotope ratio spectroscopy The Delta Ray IRIS with URI Connect is based on direct absorption spectroscopy and uses a mid-infrared laser that operates at 4.3 µm. The laser light is generated by two tunable near-infrared diode lasers that are combined in nonlinear crystal. The laser scans over the spectral region containing four CO 2 absorption lines, and isotope ratios are calculated from the spectrum fit. Calculation of different CO 2 isotopologues and determination of stable isotope ratios from the spectrum is possible due to absorption lines that are shifted relative to each other (Figure 2). Detector Voltage (V) Wavelength (µm) Figure 2. Mid-infrared spectral region at a wavelength of 4.3 μm and characteristic absorption lines for CO 2 isotopologues URI (Universal Reference Interface) Connect With an option of continuous air measurements, URI Connect offers the possibility to measure discrete samples. It is composed of three functional units: MITCH, Variable Volume (sample inlet), and Dryer (Figure 3). The Variable Volume (VV) collects the sample gas, which comes from a sample port. The system automatically determines the concentration of the sample in the VV and adjusts the dilution during measurement. Dilution is done by the reference section with the MITCH valve block (gas MIxing and switching device). This enables the instrument to perform the measurements at constant concentration. The referencing for isotope values is done by introduction of pure CO 2 gases diluted using a carrier gas. The change of concentration allows the determination of linearity (change of delta value with concentration), which can be used for subsequent correction of data. The concentration of the reference gas is balanced against that of the sample to improve accuracy. Figure 3. Schematic view of the URI Connect 2

Experimental δ 13 C of carbonate samples method Delta Ray IRIS with URI Connect Borosilicate vials were washed and dried overnight at 70 C. After adding 400 500 μg of sample, vials were sealed with new septa and residual air was removed from the vials by an automated autosampler-assisted flushing procedure. The flushing procedure is carried out at a flow of 30 ml/min for 3 min. Further, samples were manually acidified using three droplets of 104% phosphoric acid (anhydrous phosphoric acid; 104% of H 3 PO 4 = 0.25 mol H 2 O/L H 3 PO 4 ) and placed into a heated tray at a constant temperature of 80 C. After one hour at 80 C, the measurement was started. The Principle of Identical Treatment was applied in sample and standard preparation, in the measurement procedure, as well as in the evaluation of the results. The international reference materials NBS 18 and IAEA-603 were used to perform a two-point calibration. The normalized carbon isotope ratios of the samples were reported in per mill ( ) relative to VPDB. Kiel IV Carbonate Device coupled to a MAT 253 Mass Spectrometer (Figure 4) Phosphoric acid was added dropwise at full reaction temperature into the evacuated septum-free vial. The samples reacted at 70 C with the oversaturated phosphoric acid (>1.92 kg/l phosphoric acid) to produce CO 2 sample gas. The developed CO 2 gas was subsequently separated from moisture and other gases remaining in the samples by leading the gas over two cooling traps (fluid nitrogen). Subsequently, the CO 2 gas was injected under vacuum into the mass spectrometer. δ 13 C of DIC (water samples) method Delta Ray IRIS with URI Connect Vials were preloaded with 1 droplet of 104% phosphoric acid, capped and flushed with zero air (synthetic air without CO 2 ). Then, 0.8 ml per sample was injected into the vials using a disposable needle syringe. Reaction time was approximately 15 hours (overnight) at a constant room temperature. The Principle of Identical Treatment was applied in the sample and standard preparation, in the measurement procedure, as well as in the evaluation of the results. Three international standards of carbonate materials (NBS 18, IAEA-603 and LSVEC) were analyzed along with laboratory standards NaHCO 3 (s), NaHCO 3 (aq), Na 2 CO 3 (s), and Na 2 CO 3 (aq). NBS 18, IAEA-603, and LSVEC were used as standards for calibration, while laboratory standards were used for quality control. NBS 18, IAEA-603, and LSVEC were used to perform a three-point calibration. Assigned δ 13 C values of the samples were reported in per mill ( ) vs. VPDB scale. GasBench II System coupled to a Thermo Scientific DeltaXP Mass Spectrometer (Figure 5) Vials were preloaded with five droplets of ~ 90% phosphoric acid, capped and the headspace was flushed with He. Further on, 1 ml of water sample was injected into the vials using a disposable needle syringe. The reaction of sample and acid started immediately. The tray holder of the CTC Combi PAL Autosampler was routinely loaded with DIC samples and with carbonate standards that are used because international standards for δ 13 C of DIC are not available. Figure 4. Kiel IV Carbonate Device coupled to MAT 253 Mass Spectrometer Figure 5. The GasBench II System in conjunction with a CTC Combi PAL Autosampler, coupled to Thermo Scientific Delta V Isotope Ratio Mass Spectrometer 3

Results and discussion δ 13 C of carbonate samples Nineteen carbonate samples were measured with the Delta Ray IRIS with URI Connect and the Kiel IV Carbonate Device coupled to a MAT 253 Mass Spectrometer. Samples present different calcifying organisms collected from coral culturing tank at GEOMAR, Kiel (snail shells, red algae, corals, etc.) (Figure 6). Two batches with three replicates of each sample were measured using the Delta Ray IRIS with URI Connect. Repeatability (three aliquots of one sample) was better than 0.05 (MIN = 0.008 ; MAX = 0.165 ), while reproducibility of data (comparison of Batch 1 and Batch 2) was better than 0.06. When comparing the IRMS and IRIS methods, the average of absolute difference is 0.1 with minimal difference of 0.01 and maximal difference of 0.39 (Figure 7). Figure 6. Example of corals used for isotopic analysis. Seriatopora caliendrum (left) and Montipora digitate (right). Figure 7. δ 13 C values of 19 carbonates samples measured using a Delta Ray IRIS with URI Connect (marked in red) and Kiel IV Carbonate Device coupled to a MAT 253 Mass Spectrometer (marked in black) 4

δ 13 C of DIC - water samples Ten seawater samples were measured with the Delta Ray IRIS with URI Connect and with the GasBench II System coupled to the DeltaXP Mass Spectrometer. Two batches with three replicates of each sample were measured using the Delta Ray IRIS with URI Connect. Repeatability (three aliquots of one sample) was better than 0.05 (MIN = 0.01 ; MAX = 0.13 ). When comparing the IRMS and IRIS methods, the average of absolute difference is 0.2 with minimal difference of 0.03 and maximal difference of 0.73 (Figure 8). Conclusions We presented preliminary results of comparing traditional IRMS and newly developed IRIS methods for δ 13 C measurements. Both methods show high accuracy and precision when measuring δ 13 C of carbonates and DIC of seawater samples. In addition to traditional IRMS methods for laboratory analysis of stable isotopes of carbonates and DIC, IRIS offers a new option of performing high precision isotopic measurements in the laboratory as well as in the field. References 1. Geldern, R. V.; Nowak, M. E.; Zimmer, M.; Szizybalski, A.; Myrttinen, A.; Barth, J. A. C.; and Jost, H.-J. Field-based stable isotope analysis of carbon dioxide by mid-infrared laser spectroscopy for carbon capture and storage monitoring, Anal. Chem., 2014, 86, 2191 12198, https://doi.org/10.1021/ac5031732, 2014. 2. Spötl, C. A robust and fast method of sampling and analysis of δ 13 C of dissolved inorganic carbon in ground waters. Isotopes in Environmental and Health Studies, 2005, 41(3), 217 221. Figure 8. δ 13 C DIC values of water samples measured using the Delta Ray IRIS with URI Connect (marked in red) and the GasBench II System (marked in black) Find out more at thermofisher.com/deltaray 2018 Thermo Fisher Scientific Inc. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. CTC and Combi PAL are trademarks of or are used in trade by CTC Analytics AG. Teledyne is a registered trademark of Teledyne Technologies Inc. CETAC is a registered trademark of Teledyne Instruments Inc. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN30486-EN 0318S