Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Similar documents
Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Methanol Extraction of high level soil samples by USEPA Method 8260C

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Solid Phase Micro Extraction of Flavor Compounds in Beer

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Maximizing Sample Throughput In Purge And Trap Analysis

Optimizing. Abstract: is standardd. procedures. altered to

Analytical Trap Comparison for USEPA Method 8260C

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Ed George and Anaïs Viven Varian, Inc.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Solid Phase Micro Extraction of Flavor Compounds in Beer

Maximizing Production While Minimizing Costs

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Optimizing Standard Preparation

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Determination of Volatile Organic Compounds in Air

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

2017 EPA Method Update Rule and EPA Method 624.1

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

EPA TO-17 Volatile Organic Compounds

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Target Compound Results Summary

Volatile Organic Compounds in Every Day Food

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

CALA Directory of Laboratories

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

STANDARD OPERATING PROCEDURES

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Thermal Desorption Technical Support

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Associates. Project No.: October 25, 2017

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Headspace Technology for GC and GC/MS: Features, benefits & applications

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

Volatile Organic Compounds in Water PBM

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Study of Residual Solvents in Various Matrices by Static Headspace

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Copies: Dave Favero, RACER File

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

METHOD 5030C PURGE-AND-TRAP FOR AQUEOUS SAMPLES

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

In-Tube Extraction Sample Preparation for Gas Chromatography

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Table 1 Soil and Groundwater Sampling Summary Table

January 19, Dear Mr. Nightingale:

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

Author. Abstract. Introduction

CIA Advantage-xr. Cryogen-free automated canister and whole-air sampling system

Transcription:

Validation of USEPA Method 524.2 Using a Stratum PTC and the New AQUATek 100 Autosampler Application Note Abstract Automation is the key to increasing laboratory productivity and minimizing costs. It is equally important to prove that the changes in technology to achieve these results may be done without sacrificing sample integrity or data quality. The AQUATek 100 is a new vial autosampler which features many new benefits to help increase productivity, minimize carryover, and provide more flexibility to the end user in terms of maintaining sample integrity and servicing the unit. The data presented in this application note was provided by a large drinking water facility and validates USEPA Method 524.2 1 with regard to reproducibility, Surrogate percent recovery, and Method Detection Limits (MDLs) using a Stratum Purge and Trap Concentrator (PTC) and the new AQUATek 100 Vial Autosampler interfaced to an Agilent 7890A GC and a 5975C Triple Axis Detector. Introduction Figure 1: AQUATek 100 The AQUATek 100 is a liquid autosampler that interfaces to a purge and trap concentrator for measuring Volatile Organic Compounds (VOCs) in drinking water, groundwater and surface water. It includes a 100-sample position carousel for standard 40mL vials and is removable from the drive assembly for easy vial loading. It also includes a water reservoir to provide auto-blanking which frees up vial space in the carousel increasing sample throughput. It comes standard with a vial chiller tray which allows for sample cooling to 10 C which is now a requirement for the new USEPA Method 524.3 2 for preserving sample integrity. In today s environment, many analysts are being required to troubleshoot and service their own instruments. The system has a Plumbing Access Compartment (PAC) which is a pullout drawer for easy access to the sample loop, plumbing, tubing, valves, electronic boards, and other hardware components. In this study, a linear calibration was performed and the percent Relative Standard Deviation (%RSD) and Method Detection Limits (MDLs) were determined for a target list of analytes. A 25mL purge volume was used and the requirements, conditions and specifications according to EPA Method 524.2 1 were all met. Experimental-Instrument Conditions The Stratum PTC and AQUATek 100 were configured with an Agilent 7890A GC and a 5975C Triple Axis Detector for this study. The column used was an HP-VOC 30m x 0.20mm x 1.12µm (Agilent Technologies P/N 19091R-303). The columns were configured with an ultra large (6mm) inert draw-out lens (Agilent Technologies P/N G2589-20045). A larger draw-out plate was chosen due to better performance for use with Purge and Trap as found in a previous application note. A 1.0mm ID Restek 1mm split 20973-214.5 inlet liner was also employed for this study. The analytical trap used was a Vocarb 3000 adsorbent trap. The column flow was 0.47mL/min and all analyses were performed using a 30:1 split ratio and a septum purge of 3mL/min. Tables 1 and 2 display the GC/MS parameters while Table 3 displays the Stratum PTC and AQUATek 100 conditions.

GC Parameters MSD Parameters GC: Agilent 7890A MSD: 5975C Triple Axis Detector Column: HP-VOC 30m x 0.200mm x1.12um Source: 230 C Oven Program: 35 C for 4 min,8 C/min to 220 C for 0 min, 27.125 min runtime Quad: 150 C Inlet: 220 C Solvent Delay: 2.34 min Column Flow 0.47054mL/min Scan Range: m/z 35-260 Gas: Helium Scans: 5.98 scans/sec Split 30:1 Threshold: 500 Pressure: 8.75 psi MS Transfer Line Temp.: Inlet: Split/Splitless 220 C Tables 1 & 2: GC and MSD Parameters for the DB-624 Column GC Stratum PTC and AQUATek 100 Parameters Water Parameters Variable Value Variable Value Pressurize Time 0.95 min Preheat Time 1.00 min Sample Transfer Time 1.25 min Preheat Temp 40 C Rinse Loop Time 0.85 min Purge Time 11.00 Sweep Needle Time 0.35 min Purge Temp 0 C Bake Rinse On Purge Flow 40mL/min Number of Bake Rinses 1 Dry Purge Time 2.00 min Bake Rinse Drain Time 0.60 min Dry Purge Temp 30 C Presweep Time 0.35 min Dry Purge Flow 100mL/min Valve Oven Temp 150 C GC Start Start of Desorb Transfer Line Temp 150 C Desorb Preheat Temp 245 C Sample Mount Temp 45 C Desorb Time 2.00 min Purge ready Temp 40 C Desorb Temp 250 C Condenser Ready Temp 35 C Drain Flow 200mL/min Condenser Purge Temp 35 C Bake Time 7.00 min Standby Flow 10mL/min Bake Temp 270 C Sparge Vessel Heater OFF Condenser Bake Temp 200 C Pre-Purge Time 0.50 min Bake Flow 250mL/min Pre-Purge Flow 40mL/min Table 3: Stratum PTC and AQUATek 100 Parameters Stratum PTC Parameters are in Blue

Calibration To evaluate the system s linearity, a 5-point calibration curve was performed at 0.5ppb, 1.0ppb, 2.0ppb, 5.0ppb, and 10ppb. A target list of analytes from EPA Method 524.2 in addition to the oxygenate gasoline additives were evaluated. Due to the poor purge efficiency of 4 specific compounds, a higher calibration range was used and is listed in Table 4. The AQUATek 100 stores the Internal Standards and Surrogates in two- 15mL amber glass vessels to prevent transmission of UV radiation. The vessels are sealed with a PEEK TM cap to prevent adsorption and contamination of the standard solution. Each standard vessel can deliver volumes of 1, 2, 5, 10, or 20µL to each sample and delivers the exact amount minimizing waste and reducing overall costs for standards. The vessels are under constant pressure which may extend the life of the standard for 2-3 months. For this study, the required Internal Standard (IS) and Surrogates (SS) were prepared in methanol at a concentration of 62.5ppm. This was then transferred to the standard vessel on the AQUATek 100 and a 2µL aliquot was added to each sample for a final concentration of 2.5ppb. Tert-Butyl Alcohol Methyl Ethyl Ketone Tetrahydrofuran 4-Methyl-2-Pentanone All Other Compounds Compounds Calibration Range 2.5-50ppb 5.0-100ppb 5.0-100ppb 5.0-100ppb 0.5-10ppb Table 4: Calibration Range for Target Analytes The calibration data was processed using Agilent Chemstation software. The relative response factors were evaluated for linearity and %RSD, and the calibration results for all compounds are listed in Table 5. Three calibration curves were run over several weeks and the average %RSD for each curve was 6.48%, 7.06%, and 6.82% respectively. Over 600 samples were run and after cleaning the source, another calibration curve was processed and all analytes were <10%RSD. A chromatogram of the 5ppb calibration standard is shown in Figure 1.

Compound %RSD Compound %RSD fluorobenzene (IS) N/A toluene 4.24 chlorodifluoromethane 6.00 trans-1,3-dichloropropene 12.52 dichlorodifluoromethane 4.57 1,1,2-trichloroethane 2.16 chloromethane 5.06 1,3-dichloropropane 5.12 vinyl chloride 15.08 dibromochloromethane 9.28 bromomethane 6.09 tetrachloroethene 3.75 chloroethane 4.91 chlorobenzene 4.96 trichlorofluoromethane 3.24 1,1,1,2-tetrachloroethane 6.80 1,1-dichloroethene 4.70 ethyl benzene 8.61 1,1,2-Trichlorotrifluoromethane 3.64 m,p-xylene 8.40 tert-butyl-alcohol (TBA) 15.33 styrene 13.74 methylene chloride 3.27 o-xylene 7.29 methyl-tert-butyl-ether (MTBE) 3.30 bromoform 9.17 trans-1,2-dichloroethene 2.60 1,1,2,2-tetrachloroethane 7.08 1,1-dichloroethane 2.37 4-bromofluorobenzene (surr) 3.07 di-isopropyl ether (DIPE) 4.82 isopropylbenzene 9.43 methylethylketone (MEK) 6.85 1,2,3-trichloropropane 6.29 cis-1,2-dichloroethene 2.64 bromobenzene 5.55 2,2-dichloropropane 6.91 n-propylbenzene 8.22 ethyl-tert-butyl-ether (ETBE) 5.32 2-chlorotoluene 4.46 chloroform 2.48 4-chlorotoluene 7.87 bromochloromethane 5.65 1,3,5-trimethylbenzene 12.32 tetrahydrofuran (THF) 5.28 tert-butylbenzene 10.30 1,1,1-trichloroethane 4.18 1,2,4-trimethylbenzene 14.31 1,2-dichloroethane 2.42 sec-butylbenzene 11.27 1,1-dichloropropene 5.29 1,3-dichlorobenzene 7.66 carbon tetrachloride 5.48 4-isopropyltoluene 14.08 benzene 3.08 1,4-dichlorobenzene 8.03 tert-amyl-methyl-ether (TAME) 4.86 1,2-dichlorobenzene-d4 (surr) 3.64 trichloroethene 3.78 1,2-dichlorobenzene 6.75 1,2-dichloropropane 3.65 n-butylbenzene 12.22 dibromomethane 2.53 1,2,4-trichlorobenzene 9.71 bromodichloromethane 5.27 naphthalene 16.09 4-methyl-2-pentanone 7.82 hexachlorobutadiene 15.56 cis-1,3-dichloropropene 9.37 1,2,3-trichlorobenzene 9.72 Table 5: %RSDs from Calibration Curve

Figure 1: Total Ion Chromatogram of a USEPA Method 524.2 5ppb Calibration Standard Minimum Detection Limits MDLs and %recovery were evaluated over a period of 8 days by running seven replicates at the lowest concentration level for individual compounds (Table 6). This data was calculated using a 99% confidence level and is shown in Table 7. Compounds MDL True Concentration Tert-Butyl Alcohol Methyl Ethyl Ketone Tetrahydrofuran 4-Methyl-2-Pentanone Benzenes/Toluene All Other Compounds 2.5ppb 5.0ppb 5.0ppb 5.0ppb 0.25ppb 0.5-10ppb Table 6: MDL Concentration for Target Analytes

Analyte True Conc. Mean Conc. % Recovery Standard Deviation SD MDL chlorodifluoromethane 0.50 0.49 97 0.034 0.11 dichlorodifluoromethane 0.50 0.50 99 5 0.14 chloromethane 0.50 0.48 96 1 0.13 vinyl chloride 0.50 0.46 92 0.037 0.12 bromomethane 0.50 0.49 98 0.035 0.11 Chloroethane 0.50 0.51 102 0.032 0.10 trichlorofluoromethane 0.50 0.53 106 1 0.13 1,1-dichloroethene 0.50 0.52 104 0.029 0.09 1,1,2-Trichlorotrifluoromethane 0.50 0.54 109 6 0.14 tert-butyl-alcohol (TBA) 2.50 2.78 111 0.224 0.70 methylene chloride 0.50 0.55 109 0 0.12 methyl-tert-butyl-ether (MTBE) 0.50 0.49 99 0 0.13 trans-1,2-dichloroethene 0.50 0.52 103 3 0.13 1,1-dichloroethane 0.50 0.52 105 0.035 0.11 di-isopropyl ether (DIPE) 0.50 0.50 99 3 0.13 methylethylketone (MEK) 5.00 4.93 99 0.339 1.06 cis-1,2-dichloroethene 0.50 0.52 105 9 0.15 2,2-dichloropropane 0.50 0.52 103 0.036 0.11 ethyl-tert-butyl-ether (ETBE) 0.50 0.49 99 0.034 0.11 chloroform 0.50 0.53 106 3 0.13 bromochloromethane 0.50 0.53 106 4 0.14 tetrahydrofuran (THF) 0.50 4.79 96 0.301 0.95 1,1,1-trichloroethane 0.50 0.53 105 0.036 0.11 1,2-dichloroethane 0.50 0.54 108 0.037 0.12 1,1-dichloropropene 0.50 0.50 100 0.034 0.11 carbon tetrachloride 0.50 0.52 103 0.036 0.11 benzene 0.25 0.26 105 0.021 0.07 tert-amyl-methyl-ether (TAME) 0.50 0.50 99 2 0.13 trichloroethene 0.50 0.52 105 0 0.12 1,2-dichloropropane 0.50 0.52 104 3 0.14 dibromomethane 0.50 0.53 105 4 0.14 bromodichloromethane 0.50 0.52 104 0 0.13 4-methyl-2-pentanone 5.00 4.71 94 0.356 1.12 cis-1,3-dichloropropene 0.50 0.49 99 0.032 0.10 toluene 0.25 0.25 100 0.016 0.05 trans-1,3-dichloropropene 0.50 0.48 95 0.035 0.11 1,1,2-trichloroethane 0.50 0.53 106 0.053 0.17 Table 7: MDLs and % Recovery for Target Analytes

Analyte 1,3-dichloropropane dibromochloromethane tetrachloroethene chlorobenzene 1,1,1,2-tetrachloroethane ethyl benzene m,p-xylene styrene o-xylene bromoform 1,1,2,2-tetrachloroethane isopropylbenzene 1,2,3-trichloropropane bromobenzene n-propylbenzene 2-chlorotoluene 4-chlorotoluene 1,3,5-trimethylbenzene tert-butylbenzene 1,2,4-trimethylbenzene sec-butylbenzene 1,3-dichlorobenzene 4-isopropyltoluene 1,4-dichlorobenzene 1,2-dichlorobenzene n-butylbenzene 1,2,4-trichlorobenzene naphthalene hexachlorobutadiene 1,2,3-trichlorobenzene True Conc. Mean Conc. % Recovery Standard Deviation SD MDL 0.50 0.51 101 2 0.13 0.50 0.51 101 0.029 0.09 0.50 0.53 107 0.035 0.11 0.25 0.25 102 0.020 0.06 0.50 0.51 101 4 0.14 0.25 0.23 93 0.011 0.03 0.50 0.44 87 0.025 0.25 0.22 87 0.022 0.07 0.25 0.22 89 0.014 0.50 0.50 101 6 0.14 0.50 0.53 107 5 0.14 0.25 0.22 89 0.015 0.05 0.50 0.50 100 0.025 0.25 0.25 101 0.017 0.05 0.25 0.22 90 0.014 0.25 0.23 93 0.016 0.05 0.50 0.46 91 0.026 0.25 0.22 90 0.011 0.25 0.22 88 0.008 0.03 0.25 0.21 85 0.014 0.25 0.22 87 0.011 0.03 0.25 0.24 95 0.013 0.50 0.43 86 0.028 0.09 0.50 0.47 95 0.027 0.50 0.49 99 0.035 0.11 0.25 0.23 91 0.012 0.25 0.23 91 0.021 0.07 0.50 0.42 85 8 0.15 0.50 0.47 93 0.027 0.50 0.47 93 4 0.14 Table 7: MDLs and % Recovery for Target Analytes (cont.) Surrogate Percent Recovery The %recovery for all surrogates must be maintained at ±30% over a period of time according to Section 9.4 of EPA Method 524.2. The 4-bromofluorobenzene and 1,4-dichlorobenzene-d4 were evaluated over a period of 10 days and 224 samples showing an average consistent recovery of 93.25% and 92.97% respectively. This indicates that the AQUATek 100 is very accurate in the amount of standard delivered to each sample and reproducible over time.

Optional ph Module A ph sensing probe option will enable the purge and trap system to measure and record ph values for all samples in the schedule (Figure 2). Figure 3: ph Probe Figure 2: Schedule with ph Values Shown Conclusions The AQUATek 100 provides many enhancements to improving laboratory productivity while maintaining sample integrity and high quality data. Analytically, the AQUATek 100 meets or exceeds the cleanliness, reproducibility, and accuracy specifications required in a vial autosampler. References 1. USEPA Method 524.2, Measurement of Purgeable Organic Compounds in Water By Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS), Revision 4.1, 1995. 2. USEPA Method 524.3, Measurement of Purgeable Organic Compounds in Water By Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS), Version 1.0, June 2009.