Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Similar documents
GC/MS Application Note

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

PAL SPME Arrow The Better SPME

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

PAL SPME Arrow The Better SPME

Analysis of BTEX in Natural Water with SPME

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

ENVIRONMENTAL analysis

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application. Gas Chromatography February Introduction

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

A New Way to Extract Off-Flavor Compounds in the Aquatic Environment. Stir Bar Sorptive Extraction (SBSE)

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Determination of Volatile Substances Proof of Food Adulteration

Author. Abstract. Introduction

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Detection of Chemical Warfare Agents by Transportable GC/MS

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Highly Sensitive and Rugged GC/MS/MS Tool

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

GC Application Note. Determination of C2-C12 aldehydes in water by SPME on-fiber derivatization.

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

U.S. EPA Method 8270 for multicomponent analyte determination

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Signal, Noise, and Detection Limits in Mass Spectrometry

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Authors. Environmental

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Application Note 032

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Analytical Instrumentation

GC Application Note. Dilution/standard addition: Efficient and traceable.

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

AppNote 1/2011 KEYWORDS ABSTRACT

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Transcription:

Application Note Environmental Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry Authors Ye Kong and Zhe Cao Agilent Technologies, Inc. Beijing, China Abstract In this Application Note, the determination of the typical off-odor substances in drinking water, -methyl isobutyl alcohol (-MIB) and geosmin, using an SPME device with gas chromatography/mass spectrometry (GC/MS) detection is described, and applied to the control of actual water samples. The method is robust, has high sensitivity compared to conventional detection methods, and is fully automated. The detection limit was decreased by one order of magnitude with high accuracy. The precision and reproducibility were good with RSD values of less than.7%, and the interday reproducibility at trace level was less than 8.7% RSD. The performance of this method satisfied the standard test requirements of -MIB and geosmin in drinking water. The test results for actual drinking water samples were excellent.

Introduction In recent years, problems with off-odors in domestic and foreign drinking water were frequently reported. This contamination increasingly affects the quality of drinking water. The two most common off-odor substances monitored are geosmin and -methylisoborneol (-MIB). Their chemical structures shown in Figures and. Geosmin and -MIB are produced by cyanobacteria and are responsible for an unpleasant muddy or earthy off-flavor in drinking water, which is potentially also known from seafood. The olfactory thresholds are extremely low in the ppt level with 5 ng/l for geosmin and even ng/l for -MIB. The regulated maximum limit value is ng/l in the China hygienic standard for domestic drinking water (GB579-6) for both off-flavor components. In 6, the China national standard test method GB/T 7-6 for geosmin and -MIB testing in drinking water was established. Therefore, the requirement for a rapid, high sensitivity and reliable analysis method of trace off-odor levels is of special significance to the efficient control regional drinking water, and provides an early warning in case of an unexpected incident of peculiar water quality. At present, the developments on methods for off-odor analysis and determination are mainly focused on optimization of the instrumental analysis. For water samples, the qualitative and quantitative detection is typically performed by gas chromatography/mass spectrometry (GC/MS). The commonly used sample preparation methods include a wide variety of analytical solutions such as closedloop stripping analysis (CLSA), liquid-liquid extraction (LLE), solid phase extraction (SPE), solid phase microextraction (SPME), liquid phase microextraction (LPME), ultrasonic assisted emulsification microextraction (USAEME), or stir-bar sorptive extraction (SBSE). Many of these methods have shortcomings such as complex technical realization, lack of automation, the need for larger amounts of organic solvents, or low sensitivity. SPME in general is a green solventless extraction technique for small sample volumes and automated high-sensitivity analyses. The SPME Arrow is a new patented SPME device with innovative improvements for ruggedness, high sample throughput, speed of extraction, and high sensitivity for low limits of detection (LODs). The SPME Arrow device comprises an arrow-shaped tip, which penetrates vial and injection port septa more easily and with less force or damage. The sorbent material coating provides a larger surface area and higher sorbent volume than fibers, increasing the extraction capacity and thereby greatly improving the detection sensitivity. This method has the advantages of time saving, high extraction efficiency, zero solvent use, less matrix interference, and ease of automation for large sample series. In this study, an efficient and rugged SPME-GC/MS analysis method of the typical off-odors geosmin and -MIB in water was established and successfully applied in the control of drinking water. Figure. Chemical structure of geosmin Figure. Chemical structure of -methyl-iso-borneol (-MIB) Figure. A: SPME Arrow device (top) compared to SPME Fiber device (bottom). B. Color coded SPME Arrow devices

Experimental Instrumentation The following instrumentation was used to perform the experiments. Sample pretreatment platform SPME Arrow GC/MS platform Reagents Agilent PAL RSI or RTC sample handling system with SPME Arrow accessory, including Arrow tool, Heatex stirrer, injection port assembly, and liner. Optional conditioning module and agitator. mm od, DVB/CAR/PDMS sorbent phase, p/n 59-586. Before first use, the SPME Arrow was conditioned at 7 C for h Agilent 789B GC and Agilent 5977B single quadrupole GC/MS system Standard solution of targets -MIB and geosmin, each mg/l in methanol (Accustandard), as certified standard solution Internal standard of -isobutyl--methoxypyrazine, mg/l in methanol solution (Accustandard), as certified standard solution Sodium chloride, analytical grade (Group Chemical Reagent Co., Ltd.), baked at 5 C for h Methanol, chromatographic purity (Merck) Water, ultrapure/hplc quality SPME Arrow extraction procedure.5 g of sodium chloride were placed in an empty ml screw cap vial. 5 ml of a water sample were added followed by 5 µl of internal standard solution (-isobutyl-- methoxypyrazine, ng/ml). The cap was tightened, and the vial placed in the PAL system sample rack for analysis. Before SPME extraction, the sample was transferred by the PAL system to the Heatex stirrer and incubated at 6 C for min. The SPME Arrow penetrated the vial septum and its sorbent phase was extended into the air phase above the water sample, while maintaining the vial temperature at 6 C for a min headspace extraction. For analysis, the SPME Arrow was moved to the injection port and thermally desorbed at 5 C for 5 min. During the desorption time, the injector split was closed. GC/MS analysis conditions The analysis parameters of the GC/MS instrument are shown in Table. Table. GC/MS parameter settings GC column and DB-5MS UI, m mm µm, p/n -55-UI temperature program 6 C ( min) at C/min to 7 C ( min) Detector type and operating parameter Acquisition mode and masses Injection parameter Carrier gas and flow rate Results and Discussion MS detector Transfer line temperature 8 C Ion source temperature 5 C Selective ion monitoring mode (SIM): -MIB: m/z 95, 7, 8 Geosmin: m/z,, 5 ISTD: m/z 9, Temperature 5 C, splitless min He, 5. quality, constant flow mode, ml/min Chromatogram and linearity of standard samples The standard stock solution was prepared with ultrapure water at a concentration of ng/l. The SPME Arrow extraction of the stock solution provided the chromatogram shown in Figure. 5..... +EI TIC SIM std-ppt.d -MIB -IMP 8. 8.6 9 9. 9.8...8..6. Geosmin Figure. GC/MS chromatogram of -MIB and geosmin aqueous solution ( ng/l) by SPME Arrow extraction A series of standard dilutions with concentration of,,, 5, and ng/l were prepared from the stock solution. The calibration curves were generated by calculating the relative response relative to the internal standard. The calibration curves of -MIB and geosmin were obtained with good precision and linearity as shown in Figure 5. The calibration precision was excellent with the linear correlation coefficient better than 99 for both compounds.

Relative Responses.... -MIB - 5 Levels, 5 Levels Used, 5 Points, 5 Points Used. QCs y = 79 * x -.997 R^ = 999 Type: Linear, Origin: Ignore, Weight: None Reproducibility The standard solution for the ng/l level of -MIB and geosmin was measured five consecutive times to obtain the intraday reproducibility data of this method. The GC/MS chromatograms are shown in Figure 6. The area precision of -MIB and geosmin repeat measurements were.65 and.7% RSD (n = 5) respectively. Figure 7 shows the overlaid chromatograms of these measurements, the reproducibility of the SPME Arrow method is good, as demonstrated by the graphs and data. -.5.5.5.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 Relative Concentration 5.5.5.75.5.5.75.5.5.75.5.5 5 - Relative Responses Geosmin - 5 Levels, 5 Levels Used, 5 Points, 5 Points Used, QCs y =.589 * x -.579 R^ = 9976 Type: Linear, Origin: Ignore, Weight: None -.5.5.5.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 Relative Concentration Figure 5. Calibration curves of -MIB (top) and geosmin (bottom), both in the range of to ng/l with precision better than R = 99 +EI TIC SIM std-ppt.d -MIB +EI TIC SIM std-ppt.d. 9. 9. Geosmin..6..... 8.77. 8.69 9.76.. 8.7 8.8 8.9 9 9. 9. 9. 9. 9.5 9.6 9.7 9.8.5.5.55.6.65.7.75.8.85.9.95 Figure 6. GC/MS chromatograms of -MIB and geosmin at the regulated decision level ng/l

+EI EIC(.) SIM std-ppt. D 6 5.5 5.5.5.5.5 +EI EIC(95.) SIM std-ppt. D 8.5 8 7.5 7 6.5 6 5.5 5.5.5.5.5..58.6.6.6.66.68.7.7 9. 9. 9. 9.6 9.8 9. 9. 9. 9.6 9.8 9.5 9.5 9.5 9.56 9.58 Figure 7. Intraday reproducibility data for geosmin and -MIB ( ng/l, n = 5) Sensitivity of the SPME Arrow method The method uses the newest SPME Arrow patented technology, which delivers greatly improved detection sensitivity. The method detection limit (MDL) of -MIB and geosmin achieved with this method were.7 ng/l for -MIB and. ng/l for geosmin, calculated from eight consecutive runs of a ng/l dilution. The chromatograms obtained from the sample extraction at the lowest level of ng/l are shown in Figure 9. The peak shape is symmetrical with little interference for -MIB. At this low concentration level, a signal-to-noise ratio (S/N) of.6 was achieved for -MIB and an S/N of 6 for geosmin. 7 6.5 6 5.5 5.5.5.5.5 +EI EIC(.) SIM std-5ppt.d Geosmin.58.6.6.6.66.68.7.7 The interday reproducibility of this method was also evaluated. The ng/l aqueous solution was measured by this method during five consecutive days. The reproducibility for -MIB and geosmin was obtained with a precision of the peak area of 5.% and 8.57% RSD respectively. The results of the measurements are shown in Figure 8. It can be seen from these results that this method also has good day-today stability and satisfies completely the requirements of reproducible detection at trace levels. 5..5. 5.5 5 5 5 5 5 +EI EIC(.) SIM std-5ppt.d -MIB 9. 9.6 9.8 9. 9. 9. 9.6 9.8 9.5 9.5 9.5 9.56 9.58 9.6 5 Figure 9. Chromatograms of the lowest concentration level measured ( ng/l) 5 5 5 -MIB Geosmin Figure 8. Day-to-day reproducibility of -MIB and geosmin SPME Arrow analysis ( ng/l) 5

Real-life water analyses Regular tap water was tested. The spiked standard was ng/l, with repeated threefold measurements. The recovery rates of -MIB and geosmin were measured at 6% and 98.%. The precision for -MIB and geosmin in the repeat measurements of the real-life samples was.8% and.8% RSD. The total ion chromatogram obtained from the actual sample determination is shown in Figure. The real-life sample background does not interfere with the determination of the target compounds at retention times of 9. min for -MIB and.6 min for geosmin..75.5.5 +EI TIC SIM spikedppt.d 8. 9. 8. 9. 8.77.6 5 8.9 9.68..6..95.5 8.5 9 9.5.5.5.75 +EI TIC SIM std-ppt.d.5.5 9. 9..6 5 8..5 8.77. 8.6.6 9.76 9..9 8.5 9 9.5.5.5 Figure. Water sample chromatograms, spiked at ng/l (above, the spiked drinking water sample, below a spiked ultrapure water preparation) 6

Conclusions An analytical SPME method for the determination of -MIB and geosmin in water with GC/MS detection was established in this study. The method demonstrated the following advantages: Robust SPME Arrow technology provides high sensitivity, and rugged and highly reproducible quantitative determinations Green analytical method eliminates the use organic solvents, reduces environmental pollution, and protects the safety of operators Method requires small sample volumes, exhibits less interference by the matrix, and is easy to deploy for online monitoring The used SPME Arrow device is robust, unbreakable, has a long life, and lasts for several hundreds of analyses and GC/MS injections The described SPME Arrow GC/MS method for off-flavors is not only simple in operation and automated, but also fast and accurate, and satisfies completely the hygienic test requirements of drinking water and water source control in the China national standard test method for drinking water (GB579-6). 7

www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 8 Printed in the USA November, 8 599-66EN