The Physics of Doppler Ultrasound. HET408 Medical Imaging

Similar documents
Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Structure of Biological Materials

Pixel-based Beamforming for Ultrasound Imaging

Ultrasonic Measurement of Minute Displacement of Object Cyclically Actuated by Acoustic Radiation Force

SCATTERING OF ULTRASONIC WAVE ON A MODEL OF THE ARTERY J. WÓJCIK, T. POWAŁOWSKI, R. TYMKIEWICZ A. LAMERS, Z. TRAWIŃSKI

Sound wave bends as it hits an interface at an oblique angle. 4. Reflection. Sound wave bounces back to probe

Technical University of Denmark

Basic System. Basic System. Sonosite 180. Acuson Sequoia. Echo occurs at t=2z/c where c is approximately 1500 m/s or 1.5 mm/µs

Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2006 Ultrasound Lecture 1

Today s menu. Last lecture. Ultrasonic measurement systems. What is Ultrasound (cont d...)? What is ultrasound?

ISUOG Basic Training The Principles of Doppler Ultrasound. Juriy Wladimiroff

Technical University of Denmark

31545 Medical Imaging systems

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

Simulation of Contrast Agent Enhanced Ultrasound Imaging based on Field II

Physical principles of Harmonic Imaging Min Joo Choi, PhD

Mandatory Assignment 2013 INF-GEO4310

ECMUS The Safety Committee of EFSUMB : Tutorial

EE 5345 Biomedical Instrumentation Lecture 6: slides

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

Chapter 2. Interaction with Soft Tissue

Vector blood velocity estimation in medical ultrasound

DOPPLER-BASED ULTRASONIC BLOOD VELOCITY ESTIMATION

IEEE Ultrasonic symposium 2002

Medical Imaging Physics Spring Quarter Week 3-2

Chapter 31 Maxwell s Equations and Electromagnetic Waves. Copyright 2009 Pearson Education, Inc.

Friis Transmission Equation and Radar Range Equation 8.1 Friis Transmission Equation

Today s menu. Last lecture. Measurement of volume flow rate. Measurement of volume flow rate (cont d...) Differential pressure flow meters

LECTURE 8 DOPPLER EFFECT AND SHOCK WAVES

Shear waves in solid-state materials

PD233: Design of Biomedical Devices and Systems

Chapter 31 Maxwell s Equations and Electromagnetic Waves. Copyright 2009 Pearson Education, Inc.

Electrical Engineering 3BA3: Structure of Biological Materials

Spectral Velocity Estimation in the Transverse Direction

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

N db compared with that in the single pulse harmonic imaging mode, whereas

Microwave-induced thermoacoustic tomography using multi-sector scanning

IMPROVEMENT OF TIME REVERSAL PROCESSING IN TITANIUM INSPECTIONS

Introduction to Medical Imaging. Medical Imaging

Pulse-echo method can't measure wave attenuation accurately

Outline. Fundamentals of ultrasound Focusing in ultrasound Ultrasonic blood flow estimation Nonlinear ultrasonics

Lecture 20. Wind Lidar (2) Vector Wind Determination

University of Lübeck, Medical Laser Center Lübeck GmbH Optical Coherence Tomography

BROADBAND MIMO SONAR SYSTEM: A THEORETICAL AND EXPERIMENTAL APPROACH

Caution! Pay close attention to the special operation and safety instructions in the manual of the ultrasonic echoscope.

Physics and Knobology

31545 Medical Imaging systems

ULTRASONIC INSPECTION, MATERIAL NOISE AND. Mehmet Bilgen and James H. Center for NDE Iowa State University Ames, IA 50011

EXEMPLARY PROBLEMS APPENDIX B CHAPTER 1

Phononic Crystals. J.H. Page

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

Outline of today Medical Imaging systems. Wave types. 1. Discussion assignment on B-mode imaging

ULTRASONIC WAVE PROPAGATION IN DISSIMILAR METAL WELDS APPLICATION OF A RAY-BASED MODEL AND COMPARISON WITH EXPERIMENTAL RESULTS

A longitudinal wave travels through a medium from left to right.

Electromagnetic fields and waves

Signal Loss. A1 A L[Neper] = ln or L[dB] = 20log 1. Proportional loss of signal amplitude with increasing propagation distance: = α d

Mobile Radio Communications

Donald School Journal of Ultrasound Physical Bases in Obstetrics of Medical and Ultrasound Gynecology, April-June 2009;3(2):1-9

UNIVERSITY OF SOUTHAMPTON

EFIT SIMULATIONS FOR ULTRASONIC NDE

CHAPTER 14. Based on the info about the scattering function we know that the multipath spread is T m =1ms, and the Doppler spread is B d =0.2 Hz.

Applied Nuclear Physics (Fall 2006) Lecture 19 (11/22/06) Gamma Interactions: Compton Scattering

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm.

(INCLUDING THIS FRONT PAGE)

Aberration and harmonic imaging

APPLICATION-DIRECTED MODELING OF RADIATION AND PROPAGATION OF ELASTIC WAVES IN ANISOTROPIC MEDIA: GPSS AND OPOSSM

42 TRAVELING WAVES (A) (B) (C) (D) (E) (F) (G)

General Physics (PHY 2130)

MODELLING AND SIMULATION OF ULTRASOUND NON LINEARITIES MEASUREMENT FOR BIOLOGICAL MEDIUMS

BNG/ECE 487 FINAL (W16)

Ultrasonic Flow Metering Errors due to Pulsating Flow

A NEW BLOOD MIMICKING FLUID USING PROPYLENE GLYCOL AND THEIR PROPERTIES FOR A FLOW PHANTOM TEST OF MEDICAL DOPPLER ULTRASOUND

General Physics (PHY 2130)

The physics US and MRI. Prof. Peter Bogner

Fast simulation of nonlinear radio frequency ultrasound images in inhomogeneous nonlinear media: CREANUIS

Take-Home Final Exam Last Possible Due Date: Dec. 21, 2004, 5 p.m.

Copyright license. Exchanging Information with the Stars. The goal. Some challenges

A new estimator for vector velocity estimation [medical ultrasonics]

Negative refraction of ultrasonic waves in 2D phononic crystals.

Angular Spectrum Decomposition Analysis of Second Harmonic Ultrasound Propagation and its Relation to Tissue Harmonic Imaging

Analog Electronics 2 ICS905

Supplement (videos)

ECE6604 PERSONAL & MOBILE COMMUNICATIONS. Week 3. Flat Fading Channels Envelope Distribution Autocorrelation of a Random Process

MARITIME UNIVERSITY IN SZCZECIN ORGANIZATIONAL UNIT: FACULTY OF NAVIGATION - DEPARTMENT OF NAVIGATION DEVICES. Instruction

CHAPTER 1 INTRODUCTION TO ACOUSTIC SATURATION

High Intensity Focused Ultrasound Test & Measurement Techniques

LECTURE NO. 4-5 INTRODUCTION ULTRASONIC * PULSE VELOCITY METHODS

Phased Array Inspection at Elevated Temperatures

CS6956: Wireless and Mobile Networks Lecture Notes: 2/4/2015

Mathematical analysis of ultrafast ultrasound imaging

ANALYSIS OF FREQUENCY CHARACTERISTICS ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT FOR METAL PIPES

Medical Imaging Injecting Mathematics into the Problem of Bubbly Blood. Sarah McBurnie Prof Jon Chapman OCIAM, University of Oxford

Doppler shift and aberration for spherical electromagnetic waves

MEDT8007 Simulation methods in ultrasound imaging

I have nothing to disclose

Workshop 2: Acoustic Output Measurements

Cambridge International Examinations Cambridge International Advanced Level

The structure of laser pulses

A Brief Introduction to Medical Imaging. Outline

Transcription:

The Physics of Doppler Ultrasound HET408 Medical Imaging

1 The Doppler Principle The basis of Doppler ultrasonography is the fact that reflected/scattered ultrasonic waves from a moving interface will undergo a frequency shift. In general the magnitude and the direction of this shift will provide information regarding the motion of this interface. To appreciate this very general fact we need to consider the relationship between the frequency, f S, of waves produced by a moving source and the frequency, f R, of the waves received by a moving receiver. For simplicity we shall assume that the source and receiver are moving along the same line. The argument that follows will generalise to three dimensions if wave speed is isotropic and the source produces spherical waves. At t = 0, let the source, S, and receiver, R, be separated by a distance d d S R v S v R At t = 0 let S emit a wave that reaches R at a time t later ct d v t R S R R In this time t the receiver will have moved a distance v R t and the wave, propagating with velocity c will have traveled a distance c t. Thus ct = d + v R t or t = d c v R Now at time, the source will have moved a distance v S. Let the wave emitted at that instant be received at time t 0 by R. In this time R would have traveled a total distance of v R t 0, and thus

τ c(t - ) d v t R v τ S S S R R c(t 0 ) = (d v s ) + v R t 0 or t 0 = d + (c v s) c v R Thus for the receiver the interval between the waves has been 0 = t 0 t = c v S c v R whereas for the source the interval between the waves has been. Now the number of waves emitted in by the source must be equal to the number of waves received by the receiver in 0 i.e f R 0 = f S Thus f R = c v R f S (1) c v S For v S, v R c 1 we can develop the following approximation, f R = 1 v R/c 1 v S /c f S v R = 1 1 c v S c 1 1 this is reasonable as c ' 1500m s 1 and in Doppler measurements, e.g in blood flow, v apple 2m s 1

The last term on the right-hand side can be expanded using the binomial expansion (1 + x) n = 1 + nx + n(n 1) x 2 + 2! thus for x 1, higher order terms become unimportant and thus f R ' = 1 1 v R 1 + v S c c v RS f S c where v RS = v R v S is the velocity of the receiver relative to the source. Thus the Doppler shift is f R f S = f D = v RS f S (2) c thus the frequency measured by a receiver moving away from a source will be less than the frequency measured at the source, whereas the frequency measured by a receiver moving towards a source will be greater than the frequency measured at the source. 2 Continuous Wave Doppler Ultrasound In the field of continuous wave Doppler ultrasound the source and receiver are stationary. In addition the transmitting and receiving transducers may not be in line 2 Let t be the angle of the transmitter to the direction of motion and let r be the angle of the receiver to the direction of motion. Then the velocity of the scatterer relative to the transmitter will be v cos( t ) and the velocity of the scatter relative to the receiver will be v cos( r ) 2 Modern pulsed Doppler however uses the same transducer to transmit and receive.

transmitter θ t receiver θ r v scatterer The Doppler shift arising from a moving reflector/scatterer can be calculated by assuming that the reflector/scatterer is a receiver moving away from the source with velocity v cos( t ). that the receiver is moving away form the reflector/scatterer (source) with velocity v cos( r ) This is equivalent to the receiver moving away form the source with velocity v cos( t ) + v cos( r ) even though both the receiver and transducer are stationary. Thus from equation 2 we have f D = = f S v c {cos( t) + cos( r )}! 2f S v r + t cos cos c 2 r 2! t for t r we have f D = 2f Sv c cos( ) (3) where v cos( ) is the velocity of the reflector/scatter relative to the receiver/transmitter. This equation shows f D / f S. When considerations of increased ultrasound attenuation with frequency

increased back-scatter signal power with increasing frequency desired beam width are taken into account f S is chosen to be 2 20 MHz. f D / v. f D / 1/c. f D is dependent on the angles the transmitter and receiver beams make with the velocity vector. In particular if the receiver and transmitter beams are perpendicular to blood flow f D = 0. if v > 0 then f D < 0 and if v < 0 then f D > 0. 3 The Continuous Wave Doppler Instrument 3.1 Basis of operation This section outlines the basis of the instrumentation required to detect Doppler shifts in received ultrasound. Let the transmitted signal be of the form x t (t) = t cos(! S t) and the corresponding signal received from one scatter be given as x r (t) = r cos([! S +! D ]t + 1 ) where! S! R = 2 f S = 2 f R 1 is a phase term dependent on the distance of the scatterer from the transducer and phase shifts produced within the receiver. These two signals can be electronically multiplied together to give

x t (t)x r (t) = t r cos(! S t) cos([! S +! D ]t + 1 ) = t r 2 {cos(! Dt + 1 ) + cos([2! S +! D ]t + 1 )} (4) This resulting signal is then low-pass filtered to remove the 2f S source frequency leaving the Doppler signal x D (t) = t r 2 cos(! Dt + 1 ) (5) However further analogue signal processing may be required because the received ultrasound signal also consists of reflected ultrasound of much greater amplitude ( 40 50dB than backscattered signal from moving scatterer eg blood). Such reflected ultrasound exhibits a low frequency Doppler shift because of movement of the reflecting tissues eg. pulsating arteries and movement of the probe (if hand-held). For this reason some form of high-pass filtering may be required to remove this artefact. Such a procedure will unavoidably lose low frequency Doppler signals from slowly moving blood which may be of clinical significance. 3.2 Discrimination of the direction of flow The Doppler instrument described so far is unable to provide us with any information regarding the direction of motion. In instances where Doppler ultrasound is used to assess blood flow the direction of blood flow may have diagnostic significance. For example, in veins with incompetent valves or A-V malformations/fistulas. The directional information can be preserved in a number of ways side-band filtering o set carrier demodulation in-phase/quadrature demodulation We will consider each of these techniques in turn. In the descriptions that follow it must be remembered that! D > 0 implies that velocity vector components along the beam are directed towards the probe.! D < 0 implies that velocity vector components along the beam are directed away from the probe.

oscillator RF amplifier mixer doppler probe frequency meter audio amplifier blood vessel chart recorder loudspeaker headphones Generic continuous wave Doppler instrumentation 3.2.1 Side-band filtering This method is probably the simplest. The received rf signal is passed to two filters, one passing frequencies over the range! S <! <! S +! m and the other passing frequencies over the range! S! m <! <! S. The output of each filter passes to a multiplier and as before. 3.2.2 O set carrier demodulation In this method ofdetermining the direction of flow the receivedsignal is multiplied by a reference signal! 1 +! S. Thus as before the received signal is given by x r (t) = r cos([! S +! D ]t + 1 )

ω s (ω,ω + ω ) s s m audio +ve shifted doppler rf in (ω ω, ω ) s m s audio -ve shifted doppler Side-band filtering the reference signal is given by x 1 (t) = 1 cos([! S +! 1 ]t) Multiplying these two signals together gives x 1 (t)x r (t) = 1 r 2 {cos([! 1 +! D ]t + 1 ) + cos([2! S +! 1 +! D ]t + 1 )} where! 1 is chosen such that! 1! D,max. As before this multiplied signal is low pass filtered to remove the > 2! S component. Thus! 1 +! D >! 1 +ve shifted doppler! 1 +! D <! 1 -ve shifted doppler Note our tissue movement rejection filter is now a band-stop filter with a centre frequency of! 1. 3.2.3 In-phase/quadrature demodulation The received signal is passed to two separate multipliers, one, the in phase reference, multiplies the signal by

reference signal (ω s+ ω m) rf in amplifier,, notch filter doppler shifted signal Offset carrier demodualtion x ip (t) = t cos(! S t) whereas the second, a + /2 phase-shifted reference, multiplies the signal by x ps (t) = t cos(! S t + /2) = t sin(! S t) The in-phase signal, i(t), is given as before, as i(t) = x r (t)x ip (t) = r t 2 {cos(! Dt + 1 ) + cos([2! S +! D ]t + 1 )} and the quadrature phase signal, q(t), is given by q(t) = x r (t)x ps (t) = r t sin(! S t) cos([! S +! D ]t + 1 ) = r t 2 {sin(! Dt + 1 ) sin([2! S +! D ]t + 1 )} Both i(t) and q(t) are band-pass filtered and amplified as before to give

i f (t) = cos(! D t + 1 ) q f (t) = sin(! D t + 1 ) The direction of the Doppler shift, and hence the direction of flow, is determined by noting the phase relationship between i f (t) and q f (t), i.e! D > 0 then q f (t) is /2 phase retarded with respect to i f (t).! D < 0 then q f (t) is /2 phase advanced with respect to i f (t). reference signal (ω ) s rf in audio amplifier, i (t) f (in phase) +π/2 doppler signals audio amplifier, q (t) f (quadrature) In-phase/quadrature demodulation 4 Pulsed Doppler Flow Detectors There are problems associated with conventional continuous wave (CW) Doppler instrumentation, particularly when used as a flow detector. The most important one being that CW is unable to provide range resolution. In other words CW is not able to separate Doppler signals arising from di erent points along the transmitted ultrasound beam. Thus if two blood vessels intersect the ultrasonic beam it is not possible to separate velocities at the di erent points along the beam. The use of Pulsed Doppler is able to overcome this problem. The significant di erences between CW Doppler and Pulsed Doppler are

a single transducer is used for transmission and reception as transmission and reception are separated in time. pulsed Doppler is often incorporated as an additional signal processing step in conventional pulse echo ultrasound (often known as duplex scanning). periodic bursts of ultrasound (e.g a few cycles) are used. pulsed Doppler in general is only sensitive to flow within a region termed the sampling volume. Range resolution in pulsed Doppler is achieved by transmitting a short burst of ultrasound. Following the burst the received signal is mixed with a delayed version of the transmitted burst as a reference signal. The transit time of the transmitted pulse to the region of interest and back again is equal to this delay. Thus the sampling volume can be moved to di erent positions along the beam by altering this delay. The implications of this are clear: flow at di erent depths or at di erent points within a vessel can be selectively monitored. The width of the sampling volume will be proportional to the width of the transmitted ultrasound beam, whereas the length of this sampling volume will be proportional to the duration of the transmitted burst of ultrasound. 5 References Evans D H. Doppler Ultrasound: Physics, Instrumentation and Clinical Applications. Wiley, Chichester. 1989. Hill C R (ed). Physical Principles of Medical Ultrasonics. Ellis Horwood, Chichester. 1986. Webb S (ed). The Physics of Medical Imaging. IOP Publishing, Bristol. 1992. pp 319-388.