UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D

Similar documents
Abstract Submitted for the DPP99 Meeting of The American Physical Society

GA A22722 CENTRAL THOMSON SCATTERING UPGRADE ON DIII D

GA A23198 HORIZONTAL THOMSON SCATTERING SYSTEMS FOR DIII D AND SSPX

Plasma Response Control Using Advanced Feedback Techniques

GA A23736 EFFECTS OF CROSS-SECTION SHAPE ON L MODE AND H MODE ENERGY TRANSPORT

GA A22337 INITIAL OPERATION OF THE DIVERTOR THOMSON SCATTERING DIAGNOSTIC ON DIII D

Scaling of divertor heat flux profile widths in DIII-D

CQNl_" RESPONSE TO 100% INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES TO LOW ENERGY ELECTRONS AND IONS

TARGET PLATE CONDITIONS DURING STOCHASTIC BOUNDARY OPERATION ON DIII D

Simulation of Double-Null Divertor Plasmas with the UEDGE Code

RWM FEEDBACK STABILIZATION IN DIII D: EXPERIMENT-THEORY COMPARISONS AND IMPLICATIONS FOR ITER

GA A23713 RECENT ECCD EXPERIMENTAL STUDIES ON DIII D

GA A26474 SYNERGY IN TWO-FREQUENCY FAST WAVE CYCLOTRON HARMONIC ABSORPTION IN DIII-D

PLASMA MASS DENSITY, SPECIES MIX AND FLUCTUATION DIAGNOSTICS USING FAST ALFVEN WAVE

GA A26686 FAST ION EFFECTS DURING TEST BLANKET MODULE SIMULATION EXPERIMENTS IN DIII-D

PREDICTIVE MODELING OF PLASMA HALO EVOLUTION IN POST-THERMAL QUENCH DISRUPTING PLASMAS

GA A25658 PRECISION X-RAY OPTICAL DEPTH MEASUREMENTS IN ICF SHELLS

GA A23411 COMPARISON OF LANGMUIR PROBE AND THOMSON SCATTERING MEASUREMENTS IN DIII D

PROJECT PROGRESS REPORT (03/lfi?lfibr-~/15/1998):

GA A27806 TURBULENCE BEHAVIOR AND TRANSPORT RESPONSE APPROACHING BURNING PLASMA RELEVANT PARAMETERS

GA A22689 SLOW LINER FUSION

Modeling Laser and e-beam Generated Plasma-Plume Experiments Using LASNEX

Analysis of Shane Telescope Aberration and After Collimation

GA A22677 THERMAL ANALYSIS AND TESTING FOR DIII D OHMIC HEATING COIL

GA A27857 IMPACT OF PLASMA RESPONSE ON RMP ELM SUPPRESSION IN DIII-D

UNDERSTANDING TRANSPORT THROUGH DIMENSIONLESS PARAMETER SCALING EXPERIMENTS

GA A27235 EULERIAN SIMULATIONS OF NEOCLASSICAL FLOWS AND TRANSPORT IN THE TOKAMAK PLASMA EDGE AND OUTER CORE

Recent improvement of the LHD Thomson scattering system

Using the X-FEL to understand X-ray Thomson scattering for partially ionized plasmas

SciDAC CENTER FOR SIMULATION OF WAVE-PLASMA INTERACTIONS

12/16/95-3/15/96 PERIOD MULTI-PARAMETER ON-LINE COAL BULK ANALYSIS. 2, 1. Thermal Neutron Flux in Coal: New Coal Container Geometry

in the pinch. This paper describes the computer modeling behind the shielding design of a

GA A26741 SCINTILLATOR-BASED DIAGNOSTIC FOR FAST ION LOSS MEASUREMENTS ON DIII-D

GA A26123 PARTICLE, HEAT, AND SHEATH POWER TRANSMISSION FACTOR PROFILES DURING ELM SUPPRESSION EXPERIMENTS ON DIII-D

GA A23797 UTILIZATION OF LIBEAM POLARIMETRY FOR EDGE CURRENT DETERMINATION ON DIII-D

Development of a High Intensity EBIT for Basic and Applied Science

Optimization of NSLS-II Blade X-ray Beam Position Monitors: from Photoemission type to Diamond Detector. P. Ilinski

CONTRIBUTIONS FROM THE NEUTRAL BEAMLINE IRON TO PLASMA NON-AXISYMMETRIC FIELDS

GA A22863 PLASMA PRESSURE AND FLOWS DURING DIVERTOR DETACHMENT

ONION-SKIN METHOD (OSM) ANALYSIS OF DIII D EDGE MEASUREMENTS

(4) How do you develop an optimal signal detection technique from the knowledge of

Applications of Pulse Shape Analysis to HPGe Gamma-Ray Detectors

Bulk Modulus Capacitor Load Cells

GA A27805 EXPANDING THE PHYSICS BASIS OF THE BASELINE Q=10 SCENRAIO TOWARD ITER CONDITIONS

Controlling Backstreaming Ions from X-ray Converter Targets with Time Varying Final Focusing Solenoidal Lens and Beam Energy Variation

BWXT Y-12 Y-12. A BWXT/Bechtel Enterprise SMALL, PORTABLE, LIGHTWEIGHT DT NEUTRON GENERATOR FOR USE WITH NMIS

Multicusp Sources for Ion Beam Lithography Applications

GA A25853 FAST ION REDISTRIBUTION AND IMPLICATIONS FOR THE HYBRID REGIME

GA A26057 DEMONSTRATION OF ITER OPERATIONAL SCENARIOS ON DIII-D

GA A THERMAL ION ORBIT LOSS AND RADIAL ELECTRIC FIELD IN DIII-D by J.S. degrassie, J.A. BOEDO, B.A. GRIERSON, and R.J.

GA A26785 GIANT SAWTEETH IN DIII-D AND THE QUASI-INTERCHANGE MODE

IMPACT OF EDGE CURRENT DENSITY AND PRESSURE GRADIENT ON THE STABILITY OF DIII-D HIGH PERFORMANCE DISCHARGES

GA A27290 CALCULATION OF IMPURITY POLOIDAL ROTATION FROM MEASURED POLOIDAL ASYMMETRIES IN THE TOROIDAL ROTATION OF A TOKAMAK PLASMA

Initial Results from the Lick Observatory Laser Guide Star Adaptive Optics System

GA A24016 PHYSICS OF OFF-AXIS ELECTRON CYCLOTRON CURRENT DRIVE

C S N f - 7 7Ob / 5'+RADIATION TEMPERATURE MEASUREMENTS

GA A23114 DEPENDENCE OF HEAT AND PARTICLE TRANSPORT ON THE RATIO OF THE ION AND ELECTRON TEMPERATURES

STABILIZATION OF m=2/n=1 TEARING MODES BY ELECTRON CYCLOTRON CURRENT DRIVE IN THE DIII D TOKAMAK

Tell uric prof i 1 es across the Darrough Known Geothermal Resource Area, Nevada. Harold Kaufniann. Open-file Report No.

GA A27437 SCALING OF THE DIVERTOR HEAT FLUX WIDTH IN THE DIII-D TOKAMAK

RESONANCE INTERFERENCE AND ABSOLUTE CROSS SECTIONS IN NEAR-THRESHOLD ELECTRON-IMPACT EXCITATION OF MULTICHARGED IONS

Colliding Crystalline Beams

Design study of the divertor Thomson scattering system for JT-60SA

GA A24166 SUPER-INTENSE QUASI-NEUTRAL PROTON BEAMS INTERACTING WITH PLASMA: A NUMERICAL INVESTIGATION

Three-Dimensional Silicon Photonic Crystals

sample-specific X-ray speckle contrast variation at absorption edges $ & ~ 0

A NEW TARGET CONCEPT FOR PROTON ACCELERATOR DRIVEN BORON NEUTRON CAPTURE THERAPY APPLICATIONS* Brookhaven National Laboratory. P.O.

channel system which covers both sides of the laser line. Coverage on both sides of the spectrum allows for

Impurity Transport Studies of Intrinsic MO and Injected Ge in High Temperature ECRH Heated FTU Tokamak Plasmas

Magnetic Measurements of the Elliptical Multipole Wiggler Prototype

To be published in the Proceedings of ICEC-22, Seoul Korea, July 2008 MICE Note 232 1

Nuclear Science and Security Consortium. September Workshop and Advisory Board Meeting. ACED: Argon Capture Experiment at DANCE

Data Comparisons Y-12 West Tower Data

N. Tsoupas, E. Rodger, J. Claus, H.W. Foelsche, and P. Wanderer Brookhaven National Laboratory Associated Universities, Inc. Upton, New York 11973

INTERMOLECULAR POTENTIAL FUNCTIONS AND HIGH RESOLUTION MOLECULAR SPECTROSCOPY OF WEAKLY BOUND COMPLEXES. Final Progress Report

Parametric Instabilities in Laser/Matter Interaction: From Noise Levels to Relativistic Regimes

Incoherent wind velocity sensing with a two-beam interferometer

FAR SCRAPE-OFF LAYER AND NEAR WALL PLASMA STUDIES IN DIII D

Development of Polarization Interferometer Based on Fourier Transform Spectroscopy for Thomson Scattering Diagnostics

PROJECT PROGRESS REPORT (06/16/1998-9/15/1998):

DML and Foil Measurements of ETA Beam Radius

GA A22571 REDUCTION OF TOROIDAL ROTATION BY FAST WAVE POWER IN DIII D

Abstract. results that address this question for the main linacs of the NLC. We will show that the effects of alignment drifts can indeed be handled.

GA A23977 DETAILED MEASUREMENTS OF ECCD EFFICIENCY ON DIII D FOR COMPARISON WITH THEORY

~ _- IDOCLRXKT DMSP SATELLITE DETECTIONS OF GAMMA-RAY BURSTS. J. Terrell, NIS-2 P. Lee, NIS-2 R W.

DIII-D TOKAMAK MODELING OF THE RECYCLING PARTICLE FLUX AND ELECTRON PARTICLE TRANSPORT IN THE

J. C. Batchelder', K. S. Toth2, D. M. Moltz3, E. F. Zganjarl, T. J. Ognibene3, M. W. Rowe3, C. R. Binghan12.~,J. Powell3, and B. E.

Hydra Fault Current Limiting HTS Cable to be Installed in the Consolidated Edison Grid

Rotation and Particle Loss in Tore Supra. R. B. Whitel, F. W. Perkinsz, X. Garbet3j C. Bourdelle3, V. Basiuk3, L. G. Eriksson

FUSION TECHNOLOGY INSTITUTE

The temperature dependence of the C1+ propylene rate coefficient and the Arrhenius fit are shown in Figure 1.

+.V) eo(o) -2 so - sx. hngc)90w3-- Beam-beam collisions and crossing angles in RHIC*

A Distributed Radiator, Heavy Ion Driven Inertial Confinement Fusion Target with Realistic, Multibeam Illumination Geometry

OF A HELIUM-COOLED MODULE

August 3,1999. Stiffness and Strength Properties for Basic Sandwich Material Core Types UCRL-JC B. Kim, R.M. Christensen.

9 7og$y4- International Conference On Neutron Scattering, Toronto August Spin Dynamics of the reentrant spin glass Fe0.7A10.3.

Start-up Noise in 3-D Self-AmpMed

Partial Discharge Characterization of High-Voltage Cables and Components

J. T. Mihalczo. P. 0. Box 2008

A TER. Observation of Xe with the PNNL ARSA System PNNL UC-7 13

Transcription:

GA A23440 UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D by B. BRAY, C. HSIEH, T.N. CARLSTROM, and C.C. MAKARIOU AUGUST 2000

DISCLAIMER This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.

GA A23440 UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D by B. BRAY, C. HSIEH, T.N. CARLSTROM, and C.C. MAKARIOU This is a preprint of a paper presented at the 13th Topical Conference on High-Temperature Plasma Diagnostics, June 18 22, 2000 in Tucson, Arizona and to be published in Review of Scientific Instruments. Work supported by the U.S. Department of Energy under Contract No. DE-AC03-99ER54463 GA PROJECT 30033 AUGUST 2000

ABSTRACT The DIII D Thomson system measures electron density and temperature with eight pulsed ND:YAG lasers along three paths through the plasma vessel. The components of the Thomson system are absolutely calibrated so the measurements can be combined into a single profile from a normalized plasma radius (ρ) of about 0.1 to the edge of the plasma. A monochromator calibration and opto-electronic calibration measure the detectors' absolute sensitivity to background and pulsed light. A Rayleigh scattering calibration and transmission calibrations measure the transmission of light to the detectors. The calibration systems are being upgraded to reduce the effect of systematic errors on the temperature and density measurements. The systematic errors can be checked by a comparison of overlapping channels and estimated from fits to the profiles. The contributions of the systematic uncertainties relative to the statistical uncertainties of the measurement are discussed through simulations and experimental data. GENERAL ATOMICS REPORT GA-A23440 1

I. INTRODUCTION Thomson scattering has long been a primary diagnostic for measuring the electron temperature and density in large fusion experiments. The DIII D Thomson system 1 has reliably measured T e from tens of ev to 20 kev, and n e from 2 10 18 to 2 10 20 m 3 for the past decade with significant upgrades to extend the measurements to the lower divertor 2 and center of the plasma. Measurement precision is critical when trying to compare subtle changes in the profiles from one laser pulse to the next and between plasma shots taken on different days. The system has been designed for high transmission and low noise and as a result, the Thomson system has small statistical errors.(typically, δt e ~ 3% and δn e ~ 5%). Recently, extensive work has begun to understand better the systematic errors in the Thomson measurement and to reduce these errors to the level of the statistical errors. GENERAL ATOMICS REPORT GA-A23440 3

II. DESIGN The Thomson scattering system on the DIII D Tokamak uses eight 20 Hz ND:YAG lasers. Light from the lasers travel along three beam paths to the core, horizontal, and divertor Thomson entry ports. 3 The light is focussed in the vessel and wide angle lenses collect the scattered light from along the beam path and focus it onto an array of fiber optic cables which feed 44 polychromators. The six wavelength channel polychromators 4 were designed to maximize transmission, be compact, use 3.0 cm diameter optics, and be easy to align and maintain. Because 44 units are used, size and cost were factors in this design. The light is cascaded between high-performance interference filters inside the polychromators using relay lenses. The bandpass ranges of five filters were optimized to provide good temperature resolution over the range 10 ev to 20 kev. The stray light rejection of these filters at the laser wavelength is >10 5. A 3 nm wide filter at the YAG wavelength is used for Rayleigh calibration. An additional low temperature filter is used for divertor polychromators to extend the temperature range for cooler plasma in the divertor regions. Silicon avalanche photodiode detectors (APDs) are used to measure the Thomson scattered light. The APDs sensitivity to the near IR, fast response and stability make them very appropriate for detecting frequency shifted light pulses from an ND:YAG laser. The detector's gain varies about 3%/ C at 1060 nm and they are placed away from the circuit boards next to a water-cooled polychromator frame which minimizes the temperature variation (~0.2 C) of the system. Highpass and direct coupled outputs from the detector measure the scattered laser light and the background light respectively. 5 The highpass and dc signals are digitized and the data is analyzed immediately after the plasma discharge with a least-squares fit to the Thomson function derived by Selden. 6 The dc signals are used to estimate the contribution of the background to the variance of the high pass signal. Typically, the total uncertainty is dominated by the high pass signal but the background GENERAL ATOMICS REPORT GA-A23440 5

UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D B. BRAY, et al. can be significant for high power shots and channels which look toward the new upper divertor strike point. The fitter uses precalculated high pass signal levels as a function of T e and detector calibration to reduce the amount of computation after each plasma discharge. The profile results of T e and n e for an entire discharge are available for analysis within a minute after the end of a discharge. 6 GENERAL ATOMICS REPORT GA-A23440

III. CALIBRATION Many components of the Thomson system must be carefully calibrated to measure accurate electron temperature and density profiles. Three calibrations will be emphasized because they make significant contributions to the systematic errors of the measurements and upgrades have been made or are planned to reduce these errors. A. OPTO-ELECTRONIC CALIBRATION An absolutely calibrated photodiode (EG&G Model 690) is used to measure the detector sensitivity (counts/photon) in both the high-pass and direct coupled channels by calibrating a 1060 nm LED (RCA C30116) operated in both pulsed and dc modes. The ratio of the gains is a function of the electronics and is therefore independent of wavelength and only the relative gain ratios of the Thomson scattering detectors to the gain ratio of the Rayleigh YAG detector is needed for profile calculation. The calibration procedure was redesigned to eliminate the reference photodiode measurement from the final calibration used to calculate the profiles by simultaneous calibration of all detectors for an individual polychromator. The channel to channel variance of the digitizer (LeCroy FERA 4300B) for the short (65 ns) gates can be as much as a 4% correction to the observed gain ratio. The effect of the remaining 3% uncertainty in the gain ratio on the density and temperature systematic uncertainties has been modeled with a Monte Carlo simulation. B. WAVELENGTH CALIBRATION A monochromator (CVI Digikrom 240) is used to calibrate the wavelength response of the polychromators. This calibration is done using a dc light source and the direct coupled channel of the detectors. The ratio of high pass to dc gain corrects this calibration for the pulsed light from the lasers. Since the tranmission at the YAG wavelength is used to calibrate the measured GENERAL ATOMICS REPORT GA-A23440 7

UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D B. BRAY, et al. density, it is expecially important to accurately calibrate the YAG filter. This includes accurately mimicing the f/1.75 of the input optics because the transmission of the narrow filter (Fig. 1) is very sensitive to shifts due to the input angle of the light. The observed transmission of the YAG filter is strongly influenced by wavelength hysteresis, input angle and the f# of the input to the calibration fiber. This can introduce a 5% systematic variation in the density measurement. To reduce the effect of this calibration on the systematic errors, a new CW ND:YAG laser has been purchased to measure the YAG transmission. This will eliminate any uncertainty due to wavelength hysteresis. A new optics system is being designed as well to better mimic the f# of the collection optics and ensure the input light is normal to the surface of the input fiber. C. RAYLEIGH CALIBRATION The YAG light filters and detectors are used to measure the Rayleigh scattering sensitivity of the polychromators. During maintenence periods, we inject Ar gas into the vessel to calibrate the Rayleigh scattering sensitivity of the polychromators. Stray laser light from the exit window and baffles which scatters off the vessel wall can be detected at significant levels on the YAG filter 0.10 0.08 Counts/Photon 0.06 0.04 0.02 0.00 700 800 900 Wavelength (nm) 1000 1100 Fig. 1. Measured transmission for a six channel, Thomson scattering polychromator. 8 GENERAL ATOMICS REPORT GA-A23440

of some of the polychromators when there is no gas in the vessel. In order to perform an accurate Rayleigh calibration, extensive effort must be devoted to reducing the level of this stray light and especially any variation in the stray light due to laser beams wandering during calibrations. Typically, a series of measurements with various pressures from 0 to 3 Torr Ar can be performed in less than one hour. During this time the position of the alignment HeNe and ND:YAG lasers must be carefully monitored. The divertor region is expecially sensitive to the stray light because stray light from the exit window can shine directly on the divertor floor within the divertor channel viewing cones. During the 1999 2000 vent, the exit pipe was extended and exit baffles further restricted to reduce the level of stray light observed by the divertor channels by half. There is still an estimated 15% uncertainty in the divertor Rayleigh calibrations and electron density due to the variation in the stray light and alignment during calibration. D. LASER ALIGNMENT A new laser alignment feedback system for the Thomson system is being installed at DIII D. Currently, six CCD cameras (two for each of the core, horizontal and divertor systems) are used to monitor the drift of the HeNe alignment lasers during the day. The drift is monitored and corrected manually to keep the laser system aligned. The system must be carefully aligned since a small angular drift can shift the beam significantly along its 35 m path. New stepper motors have been installed in the Thomson system to allow a feedback system to correct the drift automatically. The system should be completed this year and will reduce the laser to laser variation in the Thomson density due to beam drift during plasma operations and the uncertainty in the Rayleigh calibration due to variation of the stray light level. GENERAL ATOMICS REPORT GA-A23440 9

IV. SYSTEMATIC ERROR COMPARISONS Figure 2 shows the viewing locations of the Thomson system and flux surfaces for a recent plasma discharge. The location of the spatial channels are defined by the intersection of the laser path and the viewing cone of the polychromators. Measurements are made from slightly below the midplane of the vessel to the top of the vessel and in the lower divertor. A set of closely spaced (1.3 cm) spatial channels view around the plasma edge region where the electron density and temperature can change rapidly. THOMSON R+1 R0 R-1 Fig. 2. Thomson viewing locations and flux surfaces for a recent lower divertor plasma discharge GENERAL ATOMICS REPORT GA-A23440 11

UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D B. BRAY, et al. The size of the observed systematic errors can be estimated for average profiles during quiescent periods of plasma discharges. Within one second eighty profile measurements are made in the core region and systematic errors due to the calibrations can be estimated from the deviations between the average data and a fit over most (ρ < 0.8) of the plasma. The standard deviation of this difference is about 3% for the temperature and 6% for the density (Fig. 3). This represents a significant improvement over previous systematic errors (4% and 10% respectively). Monte Carlo simulations which randomly shift the calibrations of the detectors to match the observed systematic errors suggest that the observed errors can account for most of the deviation in the average density and temperature profiles. The observed density variation is smaller than expected because recalibration of channels with large density variations has removed four points from the tails of this distribution. 3000 Shot 103029 Temperature 2000 1000 0 6 5 Density 4 3 2 1 0 0.0 0.2 0.4 0.6 ρ 0.8 1.0 1.2 Fig. 3. Averaged Thomson profile (crosses) for a recent plasma discharge compared to a fit (circles). 12 GENERAL ATOMICS REPORT GA-A23440

It is hoped that the proposed calibration system upgrades further diminish the systematic errors. The fitting of Thomson profiles is greatly simplified if the statistical errors are larger than the systematic errors and the improved understanding of the Thomson systematic errors increases our confidence in the measured profiles. GENERAL ATOMICS REPORT GA-A23440 13

REFERENCES 1 T.N. Carlstrom, et al., Rev. Sci. Instr. 63(10), 4901 (1992). 2 S.L. Allen, et al., J. Nucl. Mater. 241 243, 595 (1997). 3 P.K. Trost, et al., Rev. Sci. Instr. 61(10), 2864 (1990). 4 T.N. Carlstrom, et al., Rev. Sci. Instr. 61(10), 2858 (1990). 5 C.L. Hsieh, et al., Rev. Sci. Instr. 61(10), 2855 (1990). GENERAL ATOMICS REPORT GA-A23440 15

ACKNOWLEDGMENT Work supported by U.S. Department of Energy under Contract No. DE-AC03-99ER54463. GENERAL ATOMICS REPORT GA-A23440 17