Journal of Chromatography A, 1114 (2006)

Similar documents
anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

GC/MS Application Note

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Method for the determination of dimethyl sulfate

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

ChromTech options for PAL systems

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Yokogawa Analytical Systems Inc., Nakacho, Musashino-shi, Tokyo , Japan.

CALIBRATION OF GC/MS METHOD AND VALIDATION OF THE MODIFIED SAMPLE PREPARATION FOR DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS

Determination of 24 PAHs in Drinking Water

ENVIRONMENTAL analysis

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry

Sulfotepp impurities in Chlorpyrifos EC formulations

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Selection of a Capillary

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Determination of Volatile Substances Proof of Food Adulteration

Volatile organic compounds (VOCs):

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Tar measurement by the Solid Phase Adsorption (SPA) method

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Selection of a Capillary GC Column

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

SPE AND GC MS INVESTIGATION OF ORGANIC CONTAMINANTS IN ATMOSPHERIC PRECIPITATION

Introduction. Abstract

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Gas Chromatography (GC)

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

STANDARD OPERATING PROCEDURES

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

U.S. EPA Method 8270 for multicomponent analyte determination

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

Journal of Chemical and Pharmaceutical Research

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Determination of Elemental Sulfur in Explosives and. Explosive Residues by Gas Chromatography-Mass Spectrometry. Kimberly S.

Application. Gas Chromatography February Introduction

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography

Introduction to Capillary GC

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

GAS CHROMATOGRAPHY (GC)

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Quantification of Rice Aroma, 2-Acetyl-1-Pyrroline (2-AP), Using TurboMatrix Headspace Trap Coupled with GC/NPD and GC/MS

METHOD 3600B CLEANUP

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Application. Gas Chromatography March 1998

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Understanding Gas Chromatography

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Supporting Information

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Gas chromatography. Advantages of GC. Disadvantages of GC

HYDROCARBONS, AROMATIC 1501

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

Transcription:

Journal of Chromatography A, 1114 (2006) 234 238 Fast and sensitive environmental analysis utilizing microextraction in packed syringe online with gas chromatography mass spectrometry Determination of polycyclic aromatic hydrocarbons in water Aziza El-Beqqali a, Anders Kussak a, Mohamed Abdel-Rehim b,c, a University College of South Stockholm, SE-141 89 Stockholm, Sweden b AstraZeneca R&D Södertälje, DMPK & BAC, SE-151 85 Södertälje, Sweden c Karlstad University, Department of Chemistry, SE-651 88 Karlstad, Sweden Received 1 December 2005; received in revised form 29 January 2006; accepted 10 February 2006 Available online 3 March 2006 Abstract A new sensitive, selective, fast and accurate technique for online sample preparation was developed. Microextraction in a packed syringe (MEPS) is a new miniaturised, solid-phase extraction (SPE) technique that can be connected online to GC or LC without any modifications. In MEPS approximately 1 mg of the solid packing material is inserted into a syringe (100 250 ml) as a plug. Sample preparation takes place on the packed bed. The bed can be coated to provide selective and suitable sampling conditions. The new method is very promising. It is very easy to use, fully automated, of low cost and rapid in comparison with previously used methods. The determination of polycyclic hydrocarbons (PAHs) in water was performed using MEPS as sample preparation method online with gas chromatography and mass spectrometry (MEPS GC MS). The results from MEPS as sample preparation were compared with other techniques such as stir bar sorptive extraction (SBSE) and solid-phase microextraction (SPME). The method was validated and the standard curves were evaluated by the means of quadratic regression and weighted by inverse of the concentration: 1/x for the calibration range 5 1000 ng/l. The MEPS applied polymer (silica-c8) could be used more than 400 times before the syringe was discarded. The extraction recovery was about 70%. The results showed close correlation coefficients (R > 0.998) for all analytes in the calibration range studied. The accuracy of MEPS GC MS was between 90 and 113% and the inter-day precision (n = 3 days), expressed as the relative standard deviation (RSD%), was 8 16%. MEPS reduced the handling time by 30 and 100 times compared to SPME and SBSE, respectively. 2006 Elsevier B.V. All rights reserved. Keywords: Online sample preparation; Microextraction in packed syringe; Gas chromatography/mass spectrometry; Polycyclic aromatic hydrocarbons; Validation 1. Introduction Polycyclic aromatic hydrocarbons (PAHs) are a large group of organic compounds that are formed by the incomplete combustion of organic matter by natural (e.g. carbonisation) and anthropogenic processes. The carcinogenic and genotoxic potential of these compounds has attracted most attention. A number of PAHs as well as coal-tar and some occupational exposures to combustion emissions containing these compounds have shown carcinogenicity in experimental animals and genotoxicity and mutagenicity in vitro and in vivo [1]. Several analytical pro- Corresponding author at: AstraZeneca R&D Södertälje, DMPK & BAC, SE-151 85 Södertä lje, Sweden. Tel.: +46 8 55325604; fax: +46 8 55329026. E-mail address: Mohamed.abdel-rehim@astrazeneca.com (M. Abdel-Rehim). cedures have been developed for PAHs analysis in water, soil, air and food matrices. Common extraction methods of PAHs from wastewater samples involve liquid liquid extraction (LLE) [2] using non-polar solvents or solid-phase extraction (SPE) [3]. Solid-phase microextraction (SPME) [4] and stir bar sorptive extraction (SBSE) were also used to extract PAHs from aqueous samples [5,6]. In addition, there is a need for development of more selective sorbents for sample clean-up and enrichment [7 10]. Microextraction in packed syringe (MEPS) is a new technique for miniaturized solid-phase extraction that can be connected online to gas chromatography (GC) or liquid chromatography (LC) without any modifications [11 14]. In MEPS, approximately 1 mg of the solid packing material is inserted into a syringe (100 250 L) as a plug. The sample (10 250 L) is withdrawn through the syringe by an autosampler. When the 0021-9673/$ see front matter 2006 Elsevier B.V. All rights reserved. doi:10.1016/j.chroma.2006.02.024

A. El-Beqqali et al. / J. Chromatogr. A 1114 (2006) 234 238 235 2.2. Instrumentation The GC MS system consisted of an HP 6890-Plus gas chromatograph and a mass selective detector model 5973 (Agilent, Palo Alto, CA, USA) equipped with a programmed temperature vaporiser (PTV) and a Combi Pal autosampler (Crelab, Uppsala, Sweden). The PTV system was an OPTIC 2 (ATAS International, Veldhoven, The Netherlands). The PTV conditions were: vent flow 100 ml/min, vent time 3 min (evaporation time), purge flow 2 ml min 1 (purge pressure 5 p.s.i.), split flow 50 ml/min and split open time 2 min. The injector temperature was set at 45 C and after the evaporation period the temperature was raised by 5 C/s to 300 C. Helium was used as carrier gas and was obtained from ScanGas (Stockholm, Sweden). 2.3. GC conditions Fig. 1. (A) Packed syringe and (B) a schematic drawing of MEPS (the process is fully automated). sample has passed through the solid support, the analytes are adsorbed to the solid phase. The solid phase is then washed once by water to remove the interfering material. The analytes are then eluted with an organic solvent such as methanol or the LC mobile phase (20 50 L) directly into the instrument s injector. The process is fully automated. Any adsorption material such silica based (C2, C8, C18), restricted access material (RAM) or molecular imprinted polymers (MIPs) can be used (Fig. 1). In this study, we describe a new automated method utilizing MEPS as sample preparation online with GC MS for the analysis of PAHs in water. Compared to liquid liquid extraction and commercial solid phase extraction, it saves time and reduces solvent waste significantly. 2. Experimental 2.1. Reagents and materials Fluorene, anthracene, fluoranthene, pyrene and chrysene were delivered by Sigma Aldrich (Stockholm, Sweden). [ 2 H 7 ]Ropivacaine was used as internal standard (I.S.) supplied by the Department of Medicinal Chemistry, AstraZeneca (Södertälje, Sweden) and methanol LiChrosolv grade by Merck (Darmstadt, Germany). Milli-Q water (Millipore, Billerica, MA, USA) was used to prepare the samples. The column used was an HP50 (50% phenyl dimethylpolysiloxane) fused-silica capillary column (25 m 0.25 mm I.D., 0.31 m film thickness) obtained from Agilent (Palo Alto, CA, USA). The gas flow rates were 2 ml/min. The GC oven temperature was programmed for an initial hold of 3 min at 90 C; the temperature was increased at 50 C/min to 280 C. Conditions for MS measurements were: MS transfer line at 280 C, ion source at 230 C, electron impact ionisation (EI) at 70 ev, SIM mode with dwell time 50 ms, solvent delay: 4.5 min. The ions corresponding to anthracene, chrysene, fluoranthene, fluorine and pyrene m/z: 178, 228, 202, 166 and 202, respectively. A MSD ChemStation data system (version B.01.00) was used for data processing. 2.4. Preparation of samples Stock solutions of the analytes in methanol were prepared. From the stock solution a stepwise dilution series were made in water. Spiked water samples were prepared by adding few microliters of stock solution to 1.0 ml water, after which 50 L of the internal standard (1000 ng/l) was added. The concentration range of standard curves was between 5 and 1000 ng/l. 2.5. MEPS conditions MEPS was performed using 1 mg of solid-phase material (silica-c8) inserted into a 250 L gas-tight syringe. The solidphase material is fixed in the syringe as a plug with filters (polyethylene, 0.2 m) on both sides. The packed syringe was conditioned first with methanol and then with water (50 L) before being used for the first time. The water sample (50 L each) is drawn through the syringe 60 times by the autosampler. The analytes are then eluted with 30 L methanol directly into the GC injector. The multiple pulling/pushing of the sample by the syringe increase the extraction recovery. Also, using a small amount of the adsorbing (1 mg) make it easy to wash it and use the same syringe many times. In MEPS standard syringe with removable needle was used and no modifications were needed either for autosampler or GC.

236 A. El-Beqqali et al. / J. Chromatogr. A 1114 (2006) 234 238 Table 1 Regression parameters, limit of detection (LOD) and limit of quantification (LOQ) Anthracene Chrysene Fluoranthene Fluorene Pyrene R 2 (n = 6) 0.9990 0.9981 0.9992 0.9982 0.9989 LOD (ng/l) 5 5 5 1 1 LOQ (ng/l) 5 5 5 5 5 3. Result and discussion 3.1. MEPS method development The method using microextraction in packed syringe was used with silica-c8 (1 mg) as sorbent material. The recovery was measured as the response of a processed spiked sample expressed as peak area and compared to that of pure standard solution (1000 ng/l). To increase the sensitivity of the method, the sample was pumped up and down through the syringe 60 times by the autosampler with a speed of 20 L/s and the syringe was warmed up to 40 C under the elution step using methanol as elution solution. The extraction time was only about 2 min and the recovery was 70%. To avoid carry over the syringe was washed five times by methanol after each run. 3.2. Calibrations Calibration curves of water samples spiked with PAH standards were performed in the range 10 1000 ng/l, with [ 2 H 7 ]ropivacaine as internal standard. Table 1 shows the correlation coefficients of between 0.9981 and 0.9992 for the different analytes. The results are shown in Table 1. In addition, data from calibration curves of spiked water samples with back calculated values, accuracy and precision data are shown in Table 2. 3.3. Selectivity accuracy and precision The developed method using MEPS with silica-c8 material followed by GC MS showed good selectivity for the studied compound. Typical chromatograms presented in Fig. 2, from a water sample spiked with PAHs (100 ng/l) and from a blank water sample. The intra-day precisions (relative standard deviation, RSD) at three different concentrations for quality control (QC) samples were about 5.6 14% (n = 6) and the accuracy 11 to 16% (n = 6). The inter-day precisions (RSD) were 8.3 16% (n = 18). The accuracy varied from 10 to 13% (n = 18). The accuracy and precision data are summarized in Table 3. 3.4. Limit of quantification (LOQ) and carry-over The limit of detection (LOD), defined as the lowest detectable concentration (S/N 3), was in the range:1 5 ng/l and the limit of quantification (LOQ) defined as the lowest measurable concentration to 5 ng/l for the different PAH analytes. The carryover was tested by injecting heptane after the highest standard concentration. To eliminate the memory effect, the MEPS is Table 2 Back-calculated values of the calibration samples Compound Concentration (ng/l) Mean calculated concentration (n =6) Accuracy (%) RSD (%) Anthracene 10 11 110 5.3 50 42 84 12 100 115 115 9.8 200 173 86 10 400 403 101 10 500 524 105 8.8 800 777 97 7.0 1000 1019 102 3.7 Chrysene 10 12 121 1.4 50 53 107 1.2 100 87 87 16 200 196 98 5.2 400 448 112 13 500 526 105 13 800 789 99 5.3 1000 968 97 4.0 Fluoranthene 10 11 107 8.9 50 50 100 21 100 90 91 13 200 185 93 13 400 415 104 12 500 517 103 13 800 755 94 4.8 1000 975 98 3.7 Fluorene 10 9.0 86 7.6 50 51 103 4.5 100 105 105 2.7 200 225 112 4.8 400 373 93 7.6 500 480 96 11 800 765 96 5.9 1000 952 95 7.8 Pyrene 10 9.0 85 8.3 50 57 115 7.3 100 102 102 13 200 192 96 7.0 400 408 102 7.4 500 532 106 6.0 800 769 96 8.6 1000 1004 100 1.9 washed four times by methanol and five times by water after every injection. The carry-over was about 0.2 1%. 3.5. Comparison of extraction methods The results of the present study were compared with the results from other sample preparation techniques such as SBSE and SPME (Table 4). The present method (MEPS) showed comparable values of the accuracy and precision with the other methods for the analysis of PAH in water samples. In addition MEPS reduced the extraction time for the analytes studied by 30 100 times compared to earlier studies using SPME and SBSE. Compared with SPME, MEPS is more sensitive and more robust. Furthermore, the present method used a low sample volume of

Table 3 The accuracy and precision at various concentrations in water A. El-Beqqali et al. / J. Chromatogr. A 1114 (2006) 234 238 237 Compound Concentration (ng/l) Intra-assay accuracy (%, n = 6) RSD (%) Inter-assay accuracy (%, n = 18) RSD (%) Anthracene 150 116 14 90 16 300 97 11 103 11 600 89 11 98 12 Chrysene 150 105 8.5 101 13 300 98 9.1 106 11 600 98 12 95 13 Fluoranthene 150 95 14 99 11 300 98 7.1 104 10 600 91 5.6 99 9.1 Fluorene 150 108 8.2 113 8.3 300 110 12 107 11 600 96 9.9 92 10 Pyrene 150 112 12 107 15 300 101 6.6 105 8.3 600 94 13 100 9.6 Table 4 Comparison of accuracy and precision using MEPS, SPME and SBSE Compound (50ng/L) Accuracy (%) Precision RSD (%) Limit of detection (ng/l) Extraction time (min) MEPS SPME [4] SBSE [6] MEPS SPME [4] SBSE [6] MEPS SPME [4] SBSE [6] MEPS SPME [4] SBSE [6] Anthracene 84 81 99 12 3 6 5 100 1.2 2 30 200 Chrysene 107 81 100 1 4 5 5 90 0.2 2 30 200 Fluoranthene 100 84 100 9 4 4 5 100 1.2 2 30 200 Fluorene 103 96 97 5 5 4 1 40 0.7 2 30 200 Pyrene 115 86 100 7 3 3 1 40 0.7 2 30 200 1.0 ml, compared to 10 and 30 ml using SBSE and SPME, respectively. 4. Conclusions Fig. 2. GC MS chromatograms obtained from (A) spiked water sample with 1: anthracene; 2 chrysene; 3: fluoranthene; 4: fluorine; 5: pyrene (50 100 ng/l) and [ 2 H 7 ]ropivacaine (I.S.) and (B) blank water. A new sensitive, selective and accurate online sample preparation technique was developed and validated for the determination of PAH traces in aqueous samples. The reproducibility of the applied method for PAH analysis is below 10% at 50 ng/l. Detection limits are between 1 and 5 ng/l. Compared with SBSE, MEPS reduced sample preparation time by about 100 times and it is fully automated. Compared with SPME, the new technique is more robust. In SPME the sampling fibre of SPME is quite sensitive to the nature of the sample matrix. Also, much higher extraction recovery (higher sensitivity) can be obtained using MEPS compared to SPME. The packed syringe can also be used several times, more than 400 times for water samples. This study demonstrated that the present method could be attractive for the analysis of other groups of compounds, e.g. pesticides, herbicides, and phenols. This approach to sample analysis is very promising for many reasons: (1) it is easy to use; (2) fully automated online procedure; (3) rapid; and (4) the cost of analysis is minimal when compared to conventional solid-phase extraction. The method can be used for small sample volumes (50 L) as well as large volumes (>1 ml).

238 A. El-Beqqali et al. / J. Chromatogr. A 1114 (2006) 234 238 References [1] European Commission, SCF/CS/CNTM/PAH/29 Final, Opinion of the Scientific Committee on Food on the risks to human health of polycyclic aromatic hydrocarbons in food, 2002. [2] D. Lin, Y. Tu, L. Zhu, Food Chem. Toxicol. 43 (2005) 41. [3] M.S. García-Falcón, B. Cancho-Grande, J. Simal-Gándara, Food Chem. 90 (2005) 643. [4] E. Cortazar, O. Zuloaga, J. Sanz, J.C. Raposo, N. Etxebarria, L.A. Fernandez, J. Chromatogr. A 978 (2002) 165. [5] V.G. Zuin, L. Montero, C. Bauer, P. Popp, J. Chromatogr. A 1091 (2005) 2. [6] B. Kolahgar, A. Hoffmann, A.C. Heiden, J. Chromatogr. A 963 (2002) 225. [7] J. O Reilly, Q. Wang, L. Setkova, J.P. Hutchinson, Y. Chen, H.L. Lord, C.M. Linton, J. Pawliszyn, J. Sep. Sci. 28 (2005) 2010. [8] H. Kataoka, Curr. Pharm. Anal. 1 (2005) 65. [9] H. Kataoka, Trends Anal. Chem. 22 (2003) 232. [10] E.M. Thurman, K. Snavely, Trends Anal. Chem. 19 (2000) 18. [11] M. Abdel-Rehim, J. Chromatogr. B 801 (2004) 317. [12] M. Abdel-Rehim, Z. Altun, L. Blomberg, J. Mass Spectrom. 39 (2004) 1488. [13] Z. Altun, M. Abdel-Rehim, L. Blomberg, J. Chromatogr. B 813 (2004) 129. [14] M. Vita, P. Skansen, M. Hassan, M. Abdel-Rehim, J. Chromatogr. B 817 (2005) 303.