Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Similar documents
Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis

Certificate of Analysis

Certificate of Analysis

Certificate of Analysis

Certificate of Analysis

Certificate of Analysis

Quality? Who Decides? Mitzi M. Rettinger Vice President, Sales & Marketing Cerilliant Corporation Round Rock, TX

Certificate of Analysis

Certificate of analysis Certified reference material code: CRM-00-GTX1&4

Certificate of Analysis

Certificate of Analysis

The Uncertainty of Reference Standards

Certificate of analysis Certified reference material code: CRM-00-NEO

ASPIRIN (ACETYLSALICYLIC ACID) CERTIFIED REFERENCE MATERIAL

Certificate of Analysis

Certificate of Analysis

PHENYLEPHRINE HYDROCHLORIDE CERTIFIED REFERENCE MATERIAL

USP Class 1 Residual Solvent Mixture

Tri-n-propyltin Chloride Mixture

Metolachlor ESA Sodium Salt Standard

HPLC Reverse Phase Test Mix #1

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications

Delta-9-THC Standard. Catalog # Lot # A082745

Certified BTEX in Unleaded Gas Composite

(RS)-2-(4-isobutyrylphenyl)propanoic acid Ibuprofen Impurity J

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Technical Procedure for General Laboratory Equipment

ANALYSIS CERTIFICATE OF

Aldehydes/Ketones DNPH Standard

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

A Fast and Sensitive Method for Residual Hydrazine Analysis in Pharmaceutical Samples

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

637. Thiamethoxam. HPLC method

Annex to the Accreditation Certificate D-RM according to ISO Guide 34:2009 and DIN EN ISO/IEC 17025:2005

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4.

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

Period of validity: to date of issue:

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

CERTIFICATE OF. Single Analyte Custom Grade Solution. ICP Assay NIST SRM 3120a Lot Number: Calculated NIST SRM Lot Number: See Sec. 4.

PHENACETIN MELTING POINT STANDARD CERTIFIED REFERENCE MATERIAL

Certificate of Analysis

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Sulfotepp impurities in Chlorpyrifos EC formulations

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Additionally, minor editorial changes have been made to update the monograph to current USP style.

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

CORESTA Recommended Method No. 86

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# Calculated See Sec. 4.

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) EDTA NIST SRM 928 Lot Number: 928

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Appendix II- Bioanalytical Method Development and Validation

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3162a Lot Number:

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

1.11 Measurement Uncertainty

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3122 Lot Number:

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3112a Lot Number: Cr3

H 3 CO H 3 CO S CH 3

TECHNICAL BRODIFACOUM

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Single Analyte Mass Spec Solution ± 0.60 µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT# Calculated See Sec. 4.

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Certificate of Analysis

Technical Procedure for Measurement Assurance

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

การใช REFERENCE STANDARDS ในการควบค มค ณภาพยา

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Certificate of Analysis

Method - Determination of aromatic amines in mainstream cigarette smoke

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3103a Lot Number:

BUTYL ALCOHOL in urine by GC-FID Code GC05510

The analysis of organic acid content of additives, premix, feed, and water.

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

GC-CI-MS analysis of TMS derivatives

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

Transcription:

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard Page 1 of 9 Catalog Number: O Lot: Expiration Date: Description: Packaging: Storage: Shipping: Intended Use: Instructions for Use: July 2018 Methylone HCl in Methanol. Nominal concentration is adjusted for HCl content. Solution in 2 ml amber USP Type I glass ampoule containing not less than 1 ml of certified solution. O NH O Store unopened in freezer (-10 ⁰C to -25 ⁰C). Sonication required prior to use. Ambient. See Stability Section. This Certified Reference Material is suitable for the in vitro identification, calibration, and quantification of the analyte(s) in analytical and R&D applications. Not suitable for human or animal consumption. Users should quantitatively transfer desired volume using established good laboratory practices to spike into matrix or to dilute to the desired concentration. Each ampoule is intended for one-time use. Regulatory: USDEA Exempt Canadian TK # 61-602 Safety: Danger. See Safety Data Sheet HCl Expiration Date has been established through real time stability studies. Ampoules are overfilled to ensure a minimum 1 ml volume can be transferred when using a 1 ml Class A volumetric pipette. For quantitative applications, the minimum sample size for intended use is 1 µl. Analyte Certified Concentration Value Methylone 1.000 ± 0.005005 mg/ml Uncertainty of the concentration is expressed as an expanded uncertainty in accordance with ISO 17025 and Guide 34 at the approximate 95% confidence interval using a coverage factor of k = 2 and has been calculated by statistical analysis of our production system and incorporates uncertainty of the mass balance purity factor, material density, balance, and weighing technique. This standard is prepared gravimetrically and mass results are reported on the conventional basis for weighing in air. Nominal concentration is calculated based on: the actual measured mass; Mass Balance Purity Factor of the analyte(s); measured mass of the solution; and the density of the pure diluent at 20 C. Concentration is corrected for chromatographic purity, residual water, residual solvents and residual inorganics. Nominal concentration is adjusted for HCl content. No adjustment required before use. Additional certification information available upon request. Metrological Traceability This standard has been prepared and certified under the ISO Guide 34, ISO/IEC 17025, ISO 9001 and ISO 13485 standards. This standard meets the requirements of a Certified Reference Material and a Primary Standard as defined by ISO and is traceable to the SI and higher order standards through an unbroken chain of comparisons. This standard has been gravimetrically prepared using balances that have been fully qualified and calibrated to ISO 17025 requirements. All calibrations utilize NIST traceable weights which are calibrated externally by a qualified ISO 17025 accredited calibration laboratory to NIST standards. Qualification of each balance includes the assignment of a minimum weighing by a qualified and ISO 17025 accredited calibration vendor taking into consideration the balance and installed environmental conditions to ensure compliance with USP tolerances of NMT 0.1% relative uncertainty. Balance calibration adjustments are performed weekly utilizing the balance's internal adjustment mechanism. Calibration verifications are performed pre-use. Weigh tapes from the calibration verification are included in the production batch record for this standard. Production data package available upon request. Fill volume is gravimetrically verified throughout the dispensing process using qualified and calibrated balances. Concentration is verified against an independently prepared calibration solution gravimetrically prepared. Each raw material utilized has been identified and thoroughly characterized through the use of multiple analytical techniques. Spectral data is provided on subsequent pages of this COA. The density and material Mass Balance Purity Factor is traceable to the SI and higher order reference standards through mass measurement and instrument qualification and calibrations. Cerilliant certifies that this standard meets the specifications stated in this certificate and warrants this product to meet the stated acceptance criteria through the expiration/retest date when stored unopened as recommended. Product should be used shortly after opening to avoid concentration changes due to evaporation. Warranty does not apply to ampoules stored after opening. Darron Ellsworth, Quality Assurance Manager October 09, 2015 Date

Solution Standard Verification Standard Solution Assay Parameters Calibration Curve Analysis Method: HPLC/UV Calibration Curve: Linear Regression Column: Ascentis Express Phenyl Hexyl, 2.7 µm, 50 x 3.0 mm Number of Points: 4 Mobile Phase: A:: Acetonitrile Linearity (r) : 1.000 B:: 0.1% Trifluoroacetic acid in Water Gradient: Time (min) 0.0 0.5 3.0 5.0 5.1 6.0 %B 98 98 50 50 98 98 Page 2 of 9 Concentration accuracy and within- and between-bottle homogeneity are analytically verified against an independently prepared calibration solution and to the prior lot. Solution standard verification demonstrates confirmation that the specified requirements for the Primary Standard have been fulfilled and validated under ISO 13485. %A 2 2 50 50 2 2 Standard Solution New Lot Existing Lot Lot Number FN093012-10 Verified Concentration (mg/ml) %RSD - Homogeneity Actual Results Acceptance Criteria Actual Results Acceptance Criteria 1.024 ± 3% 1.0 3% 1.014 ± 3% 0.6 3% Concentration is verified through multiple analyses and is calculated as the average of multiple analyses compared to an independently prepared calibration solution. Within-sample and between-sample homogeneity of the New Lot is ensured through rigorous production process controls statistically analyzed to evaluate risk and verified by analysis. Multiple samples pulled from across the lot using a random stratified sampling plan were analyzed to verify homogeneity. % RSD results shown above for the New Lot demonstrate ampoule-to-ampoule homogeneity.

Analyte Certification - Mass Balance Purity Factor Page 3 of 9 Each analyte is thoroughly identified and characterized using an orthogonal approach. A mass balance purity factor is assigned incorporating chromatographic purity and residual impurities. The mass balance purity factor and salt adjustment are utilized to calculate the weighing adjustment necessary to ensure accuracy of the solution standard concentration. Material Lot Number CAS Number Chemical Formula Molecular Weight (salt) Molecular Weight (base) Salt Adjustment Methylone HCl PC11041301 186028-80-8 C 11 H 13 NO 3 HCl 243.69 207.23 1.176 Material Characterization Summary Analytical Test Method Results Primary Chromatographic Purity by HPLC/UV Analysis SP10-0102 99.5% Secondary Chromatographic Purity by GC/FID Analysis SP10-0101 99.1% Identity by GC/MS Analysis SP10-0105 Consistent with Structure Identity by LC/MS Analysis SP10-0107 Consistent with Structure Identity by 1 H-NMR Analysis USP <761>, SP10-0116 Consistent with Structure Residual Solvent Analysis by GC/FID Headspace AM1087 0.24% Residual Water Analysis by Karl Fischer Coulometry AM1346 0.53% Inorganic Content by Microash Analysis SP10-0135 < 0.2% Purity Factor 98.70% The primary chromatographic purity is calculated as the average of two independently performed analyses utilizing two different methods. Acceptance criteria requires the purity values to be within 0.5% of each other. The primary chromatographic purity value is used to calculate the Mass Balance Purity Factor. A secondary chromatographic purity method is utilized as a control. Mass Balance Purity Factor = [(100 - wt% residual solvent - wt% residual water - wt% residual inorganics) x Chromatographic Purity/100]. Mass Balance Purity Factor does not include adjustment for chiral and/or isotopic purity. 1 Validated analytical method

Spectral and Physical Data Page 4 of 9 HPLC/UV Column: Mobile Phase: Gradient Program: Flow Rate: Wavelength: Ascentis Express C18, 2.7 µm, 2.1 x 50 mm A:: Acetonitrile B::0.1% Trifluoroacetic acid Time (min) %A %B 0.0 2 98 0.5 2 98 3.0 50 50 5.0 50 50 5.1 2 98 6.0 2 98 0.6 ml/min 282 nm Data File Name: Instrument: Sample Name: Acquired: PRM-435 P 2013-11-28 01-27-00\1BB-0301.D LC#13 PC11041301 November 28, 2013 Peak # Ret Time Area Height Area % 1 2.47 0.42 0.26 0.05 2 2.54 833.52 499.81 99.54 3 2.70 0.13 0.13 0.02 4 2.80 0.62 0.57 0.07 5 2.95 0.27 0.24 0.03 6 3.03 0.11 0.07 0.01 7 3.06 0.31 0.29 0.04 8 3.13 0.12 0.10 0.01 9 3.24 0.16 0.14 0.02 10 3.60 1.00 0.77 0.12 11 3.69 0.33 0.28 0.04 12 3.77 0.17 0.14 0.02 13 3.86 0.17 0.09 0.02

Spectral and Physical Data (cont.) Page 5 of 9 GC/FID Column: DB-5ms, 30 m x 0.53 mm ID, 1.5 µm film thickness Temp Program: 35 C (hold 5 min) to 280 C at 10 C/min hold 9 min Injector Temp: Cool-on-Column Detector Temp: 325 C Data File Name: Instrument: Sample Name: Acquired: PRM-435 P GC 10 2013-11-20 20-33-31\010B0201.D GC#10 PC11041301 November 20, 2013 Peak # Ret Time Area Height Area % 1 20.48 3.28 1.25 0.03 2 21.63 7.37 2.71 0.06 3 22.46 11911.30 4362.73 99.11 4 22.52 84.40 19.32 0.70 5 22.84 5.57 2.58 0.05 6 25.13 2.72 0.68 0.02 7 26.82 1.51 0.37 0.01 8 28.47 2.58 0.58 0.02 Residual Solvent Analysis by GC/FID Headspace Column: Temp Program: Carrier Gas: DB-ALC1 30 m x 0.53 mm, 3 µm film thickness 40 C (12 min) to 220 C at 40 C/min (5.5 min) Helium Flow Rate: 2.0 ml/min Detector Heater Temp: 250 C Injector: Headspace Sampler HS Oven Temp: 60 C Vial Equilibration: 10 minutes Data File Name: Instrument: Sample Name: Acquired: GC 9 2013-11-25 17-55-01\012F0301.D GC#9 PC11041301 November 25, 2013 Peak Compound Area Weight % 1 Unknown # 1 115.04 0.24 2 NMP NA NA Total 0.24

Spectral and Physical Data (cont.) Page 6 of 9 1 H NMR Instrument: JEOL ECS 400 Solvent: DMSO-D 6

Spectral and Physical Data (cont.) Page 7 of 9 LC/MS Column: Ascentis Express C18, 2.7 µm, 3.0 x 50 mm Flow Rate: 0.4 ml/min Mobile Phase: A:: 0.1% Formic acid in Water Scan Range: 100-1200 amu B:: Acetonitrile Ionization: Electrospray, Positive Ion Gradient: Time (min) %A %B Data File Name: W11261317 0.0 95 5 Instrument: Waters XEVO G2 QTOF 1.0 95 5 Sample Name: PC11041301 2.5 80 20 Acquired: November 26, 2013 4.0 80 20 5.0 95 5 8.0 95 5 PRM-435 PC11041301 W11261317 559 (2.033) Cm (559:579) 208.0970 100 Methylone hydrochloride Cone Voltage 15 CE 6 1: TOF MS ES+ 1.25e7 Theoretical [M + H] + : 208.0974 Found [M + H] + : 208.0970 % 209.1002 125.9861 230.0786 437.1671 477.1073 517.1043 731.4724 814.4664 0 100 200 300 400 500 600 700 800 900 1000 1100 m/z

Spectral and Physical Data (cont.) Page 8 of 9 GC/MS

Page 9 of 9 Stability Short term stability studies have been performed under accelerated conditions for periods up to four weeks. Short term data is utilized to predict long term stability and to support transport conditions and normal laboratory use. Real-time stability studies are performed at the recommended storage conditions over the life of the product. Short Term Stability : A summary of accelerated stability findings for a related product (M-141, Methylone-D 3 HCl) is listed below. Storage Condition Freezer Mean Kinetic Temperature (MKT) -15 C Refrigerator 4 C Room Temperature 21 C 40 C 40 C Transport/Shipping : Stability data supports transport of this product at ambient conditions. Time Period/Result No decrease in purity was noted after four weeks. Short Term Storage: Stability data supports short term storage up to 12 months at refrigerate conditions. Long Term Stability: Long term stability has been assessed for freezer storage (-10 C to -25 C) conditions. Stability of a minimum of 51 months has been established through real-time stability studies. COA Revision History Revision No. Date 00 May 09, 2014 01 October 09, 2015 Reason for Revision Initial version Revised Retest Date of July 2018 to Expiration Date of July 2018. Updated Long Term Stability from 44 months to 51 months.