USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Similar documents
USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

425 Jordan Road SOP. No. 8 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 10

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Fusion UV/Persulfate TOC Analyzer

ALLOWAY METHOD OUTLINE

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Organic Carbon, Total (Low Level) (UV Promoted, Persulfate Oxidation) CAS # Total Organic Carbon TOC C

Fusion UV/Persulfate TOC Analyzer

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

United States EPA Method 415.3

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

DR/4000 PROCEDURE NITRATE. 2. The display will show:

SECTION D.2 AMMONIA NITROGEN

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

1) Alkaline Persulfate Digestion for TN, TDN and TDPTotal Dissolved Nitrogen. a) Scope and Application

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Hexavalent Chromium by Flow Injection Analysis (FIA)

Method for Nitrite determination on Low-Range Samples

Alkalinity, USEPA by Flow Injection Analysis (FIA)

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Oxygen Demand, Chemical

Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector

MEASUREMENT: PART II

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Test Method: CPSC-CH-E

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Chloride by Flow Injection Analysis (FIA)

Spectrophotometric Determination of Ferrocyanide in Effluents

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

DR/4000 PROCEDURE. IRON, Total

IV. EQUIPMENT a. Horiba LA 950 main unit b. Dry sample unit that has a silencer behind it c. Cell slide unit that holds

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Oxygen Demand, Chemical

Chemical Reactions: The Copper Cycle

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC)

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

Apply the ideal gas law (PV = nrt) to experimentally determine the number of moles of carbon dioxide gas generated

PAR Method Method to 2.0 mg/l Pb TNTplus 850

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Simplifying the Process: Automated USP 643 / EP

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Method to 0.50 mg/l NH 3 N Powder Pillows

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Sodium Chloride - Analytical Standard

NITROGEN, TOTAL, HR, Test N Tube

ICP-OES DETERMINATION OF IRON. Introduction:

Exercise 2-4. Titration of a Buffer Solution EXERCISE OBJECTIVES

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

Method to 0.50 mg/l NH 3 N Powder Pillows

Dr. OligoTM DNA / RNA / OLIGO SYNTHESIZERS

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Introduction to Spectroscopy: Analysis of Copper Ore

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

LOW-RANGE TOTAL CHLORINE

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

CHEMISTRY DEPARTMENT

Transcription:

Troy, NY 12180 Date: 12/12/2017 Page 1 of 7 USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure 1. Scope and Application 1.1 Analytes Dissolved inorganic carbon (DIC) 1.2 Reporting Limit 41.63 mol/l 0.5 mg/l 1.3 Applicable Matrices This method is used to determine the DIC concentrations in precipitation, dilute surface waters and soil waters. 1.4 Dynamic Range The analytical range for the determination of DIC is from 8.326 umoles C/L (0.1 mg C/L) to 832.6 umoles C/L (10.0 mg C/L). Sample concentrations that exceed this range must be diluted and reanalyzed. Sample concentrations below the reporting limit are flagged < in the LIMS. 1.5 Interferences Method interferences (positive bias) may be caused by contaminants in the carrier gas, dilution water, reagents, glassware, or other sample processing hardware (i.e. filtering apparatus). 2. Summary of Procedure The sample is acidified with phosphoric acid and sparged with ultra-high purity nitrogen. This process converts the DIC to carbon dioxide which is then measured by an infrared detector. 3. Safety Issues 3.1 Chemical Hazards A. Handle phosphoric acid (H3PO4) with care. B. All strong acids and bases should be mixed in a fume hood. C. Gloves, safety glasses, and lab coats should be worn when preparing and performing this analysis. D. For proper handling techniques for specific chemicals, consult the appropriate Safety Data Sheet (SDS). 4. Sample Preservation, Containers, Processing and Analysis Times 4.1 Sample Preservation Samples are unfiltered and stored at 4 C. 4.2 Containers

Troy, NY 12180 Date: 12/12/2017 Page 2 of 7 Samples are stored in 40-mL amber acid-washed glass vials with Teflon septa seals. 4.3 Processing and Analysis Times Sample processing: NA Lab analysis: one week LIMS entry: one week 5. Reagents and Standards 5.1 General Information All reagents are commercially purchased and should be stored in the original container. Date the reagent bottles when received and when opened. Note expiration date, if any. No verification of the reagents is necessary 5.2 Reagents A. Nitrogen Use commercially purchased ultra high purity (UHP) nitrogen. B. 21% Phosphoric Acid 1. Add 37 ml concentrated phosphoric acid (H3PO4) to a flask that contains 188 ml of Milli-Q water. 2. Mix thoroughly. 3. Store in a polyethylene bottle; label and date. 4. Prepare monthly. 5.3 Standards A. Inorganic Carbon Standard Stock Solution, 1000 mg C/L 1. Dry about 10 g anhydrous sodium carbonate (Na2CO3) at 105 C for 2 hours and allow to cool in dessicator. 2. Dry about 10 g of anhydrous sodium bicarbonate (NaHCO3) at 105 C for 2 hours and allow to cool in dessicator. 3. Dissolve 4.44122 g dried Na2CO3 in a 1,000-mL volumetric flask containing about 800 ml Milli-Q water, add 3.497 g of NaHCO3 and dissolve. 4. Fill close to final volume with Milli-Q water, mix, then fill to final 5. Store in a glass bottle at 4 C; label and date. 6. Prepare every six months. 7. To avoid contamination, aliquots of stock solution must not be withdrawn directly from the bottle. B. DIC Working Standard 1. Pipet 1 ml of 1000 mg C/L standard stock into a 100-mL volumetric flask. 2. Fill close to final volume with Milli-Q water, mix, then fill to final

Troy, NY 12180 Date: 12/12/2017 Page 3 of 7 3. Store in a brown glass bottle at 4 C; label and date. 4. Prepare every other day or as needed. Working Standard DOC Concentration Standard Stock Added (ml) Final Volume (ml) 1 0.0 mol/l (0.0 mg/l) 0.0 100 2 8.326 mol/l (0.10 mg/l) 0.05 100 3 20.81 mol/l (0.25 mg/l) 0.1 100 4 41.63 mol/l (0.5 mg/l) 0.2 100 5 83.26 mol/l (1.0 mg/l) 0.5 100 6 208.15 mol/l (2.50 mg/l) 1.0 100 7 832.6 mol/l (10.0 mg/l) 2.0 100 C. Inorganic Carbon Quality Control Stock Solution, 1000 mg C/L 1. Dry about 10 g anhydrous sodium carbonate (Na2CO3) at 105 C for 2 hours and allow to cool in dessicator. 2. Dry about 10 g of anhydrous sodium bicarbonate (NaHCO3) at 105 C for 2 hours and allow to cool in dessicator. 3. Dissolve 4.44122 g dried Na2CO3 in a 1,000-mL volumetric flask containing about 800 ml Milli-Q water, add 3.497 g of NaHCO3 and dissolve. 4. Fill close to final volume with Milli-Q water, mix, then fill to final 5. Store in a glass bottle at 4 C; label and date. 6. Prepare every six months. 7. To avoid contamination, aliquots of stock solution must not be withdrawn directly from the bottle. D. DIC QC Samples: 1. Pipet desired amount of QC stock into a 500-mL volumetric flask. 2. Fill close to final volume with Milli-Q water, mix, then fill to final 3. Store in a glass bottle at 4 C; label and date. 4. Prepare daily or as needed. QC Sample DOC Concentration QC Stock Added (ml) Final Volume (ml) Low 41.63 mol/l (0.50 mg/l) 0.125 250 High 249.78 mol/l (3.0 mg/l) 0.750 250

Troy, NY 12180 Date: 12/12/2017 Page 4 of 7 6. QC Procedure A. The standard curve is a linear plot of standard concentration vs. the peak area obtained from the working standards. The best-fit line is drawn, and the curve is accepted if the correlation coefficient is 0.99875 or greater. B. Quality-control samples are analyzed at the start of a run, after every 10 samples during the run, and at the end of the run. C. A quality control sample is acceptable if the analyzed value is within 10 percent of the QC-high range known value and 25 percent of the QClow range known value. D. If one of the QC samples fails the acceptance criteria, the run is stopped and the QC sample is re-run. If the QC sample fails again the run is stopped, the QC sample is remade from the substock, and/or the instrument is re-calibrated. Samples associated with the failed QC sample are re-analyzed. 7. Chemical Analysis Procedure 7.1 Instrumentation Tekmar-Dohrmann Fusion TOC Analyzer Teklink software v 1.1.5.2 7.2 Start-up A. Remove standards and QC samples from the refrigerator and allow to warm to room temperature before analysis. B. Check the waste container daily. If nearing capacity, empty the waste down the hood drain and flush the drain with water. C. Open the UHP nitrogen tank valve. The tank pressure of the gas supply should be above 200 psi. If the reading is 200 psi or lower, change the tank. The delivery pressure should be 80 psi for the proper functioning of the instrument. D. Double click on the Teklink icon. E. Connect to the Fusion using RS232COM1 connection F. Ready the instrument. G. The recommended flow rate for the Fusion is 200 cc/min ( + /- 10%). If necessary, adjust the pressure from the nitrogen tank regulator until the From Detector reading reaches 200 cc/min. H. Check supply of reagents and fill if necessary. I. If either the persulfate reagent or the acid has been replace, under Tools, Prime the Acid, Prime the Reagent in that order. J. Change supply of Milli-Q water in 2 L bottle, under Tools, Prime the Water. K. Allow the instrument to warm up until the UV lamp is at highest brilliance and the baseline has stabilized, approximately 5 minutes.

Troy, NY 12180 Date: 12/12/2017 Page 5 of 7 7.3 Blank and Calibration A. Under File, open Schedule from the previous run. Click Use for the Clean, Reagent/Acid Blank, and Calibration Standards. Save run as new run with MMDDYYYY format. B. Place the prepared 832.6 umoles C/L (10 mg/l) standard stock in Position A. From this stock the Fusion will automatically prepare all standards needed for the calibration. C. Click on Start. D. The Fusion will clean, perform daily blanks on reagents, and perform the calibration. Allow the Fusion to run through the clean, blank, and calibration procedure, approximately 45 minutes. E. Once the calibration has finished, select the daily calibration results to check the correlation coefficient. If the calibration is acceptable, print the calibration and proceed with samples. 7.4 Analysis A. Under File, open Schedule. Change the SSN s and set the status of the samples to Use. B. Save run with MMDDYYY format. C. Print out the schedule which serves as the bench sheet. D. Click on Save. E. Place the samples in the sample tray on the autosampler in the order corresponding to that on the bench sheet. F. Pour the QC samples into their respective bottles (QC-high in position B and QC-low in position C). G. At the end of the report, include a Clean procedure as the Sample Type. H. Click on Start. I. Any samples over-range will be automatically diluted using the Teklink s Intellidilute feature. The dilution factor will automatically used to determine the final results for the sample. J. The QC samples will be analyzed automatically during the run after every 10 samples. If all QC samples pass, the run will proceed until the tray is finished. K. If the QC samples fail at any time during the run, a message will pop up and analysis will cease. Re-pour the QC sample and reset the status to Use in the Report window. L. Click on Start. M. If QC samples fail again, remake the QC samples from the substock and/or recalibrate. N. Note any bad peaks or samples of questionable results.

Troy, NY 12180 Date: 12/12/2017 Page 6 of 7 7.5 Shutdown A. Click on Standby or the Fusion will automatically go to Standby Mode after 15 minutes of inactivity (no samples running). B. Log out of Teklink software. C. Turn off the UHP nitrogen gas. 7.6 Maintenance A. Check to ensure that the gas/liquid separator water level is filled to the waste outlet. B. Make sure that the carrier gas flow rate is 200 cc/min + /- 10%. C. The tin and copper should be replaced in the scrubber every 6 months or when the copper becomes discolored, whichever occurs first. 7.7 Data Processing and LIMS Entry A. Under File, Open Report, open the appropriate files for the run date. The files are in a date time format: MMDDTTTT. There may be multiple files for each day. B. Export the report as an HTML and save to \\igsahaewfs136/lab- DATA/DOC/DOC-2018/FusionData2018/Reports. C. Export the report as a.csv and save to \\igsahaewfs136/lab- DATA/DOC/DOC-2018/FusionData2018/Reports. D. A series of files will open with the results. Edit the files as needed and compile into a single Excel file with the data in chronological order. Do not include any data from the priming, cleaning, or calibration files. E. Save as an Excel worksheet (.xls file extension) in \\igsahaewfs136/lab- DATA/DOC/DOC-2018/FusionData2018/Reports F. Print the file. G. Double click on the Watershed LIMS icon. H. Click Custom, then Import Data. I. Under the Import drop down, choose Dissolved Inorganic Carbon. J. Choose and open the desired file. K. Choose Client, Units, type in the Test Date, and choose Analyst. L. Exclude and/or edit any data necessary. M. Click Client ID to Sample No. N. Click Set Data. O. Investigate problems for data that did not transfer or are duplicated. 8.0 Calculations and Data Reporting A. Data are output from the instrument in units of parts per milligram ppm). During the import to the LIMS, the LIMS automatically converts the ppm value from instrument to micromoles per liter by multiplying the raw data from Fusion by DIC conversion factor of 83.26

Troy, NY 12180 Date: 12/12/2017 Page 7 of 7 This method has been assigned USGS National Water Information System method code XXXXXX for filtered waters and X unfiltered waters. 9.0 Archiving Requirements 9.1 Data Data files are backed-up daily by an automated back-up program. Hard copies of the runs are filed and retained indefinitely. The laboratory LIMS system is backed up daily by an automated back-up program. 9.2 Samples Samples are stored at room temperature until the data can be verified. Sample bottles are cleaned and reused for new samples. 10.0 References Fusion User Manual; Tekmar Company. U.S. Environmental Protection Agency, 1987, Handbook of methods for acid deposition studies laboratory analysis for surface water chemistry: Washington, D.C., U.S. Environmental Protection Agency EPA 600/4-87/026 (variously paged). 11.0 Keywords Carbon, dissolved inorganic carbon, water samples, chemical analysis, water analysis.