Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Similar documents
Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

The new Agilent 1290 Infinity Quaternary LC

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

High Resolution Fast LC

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

The Agilent 1290 Infinity II Multi Method Solution

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Agilent 1200 Series Rapid Resolution LC and Rapid Resolution LC/MS Optimization Guide

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Method Development in Comprehensive 2D-LC Finding the Most Orthogonal Separation Systems for RPLC RPLC Using Column and Solvent Screening

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Agilent LC & LC-MS Cannabis Testing

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Fast methods for the determination of ibuprofen in drug products

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

HPLC Praktikum Skript

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Figure 1. Structures for Vitamin B2 and Vitamin B1.

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Routine MS Detection for USP Chromatographic Methods

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

A Look at Agilent HPLC Columns Choices

High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension

Modernizing the USP Monograph for Acetaminophen

Successfully Scaling and Transferring HPLC and UPLC Methods

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

LC Technical Information

Qualification Specifications for Agilent 1100/1200 HPLC Systems

Varian Galaxie Chromatography Data System for Preparative HPLC

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Getting More Resolution from Your LC Michael Woodman Columns and Consumables September 10, 2008

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Application Note. Edgar Naegele. Abstract

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think!

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

LC and LC/MS Column Selection Flow Chart

ION CHROMATOGRAPHY SYSTEM S 150

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Transcription:

Performance characteristics of the Agilent 129 Infinity Quaternary Pump Technical Overview Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Abstract This Technical Overview presents Proof of Performance for the new Agilent 129 Infinity Quaternary Pump. The 129 Infinity Quaternary Pump combines the performance of a binary high-pressure mixing UHPLC pump with the flexibility of a low-pressure quaternary pump for conventional or fast runs on 3- and 4.6-mm id columns in addition to applications that need 2.1-mm id columns. Precision of retention times as well as precision and accuracy of step gradients were evaluated and compared to results obtained on the Agilent 129 Infinity Binary and 126 Infinity Quaternary Pumps.

Introduction The Agilent 129 Infinity Quaternary Pump provides excellent performance through innovative design elements such as: Integrated, high efficiency degassing unit with low internal volume facilitates fast change-over of solvents when purging and priming the pump Highest solvent mixing efficiency before solvent reaches the pump head through the new Agilent Inlet Weaver, based on established multilayer mixing technology Active damping with independently controllable high resolution pump drives and firmware-embedded tuning algorithms significantly reduce ripples and associated UV noise Optional 38-µL Agilent Jet Weaver for extra mixing volume Inline filter in front of the injector protects the column from contamination Multipurpose valve enables automatic software-embedded functionality : Switching the mixer in and out of fl o wp a t h Backflushing the inline filter Automatic purging Figure 1 shows the design of the 129 Infinity Quaternary Pump. This Technical Overview evaluated the following important pump parameters: The system delay volume and the mixing noise were determined based on step gradient tracer experiments. The composition accuracy and precision were determined based on step gradient tracer experiments. The precision of retention times was evaluated under different chromatographic conditions. 4-Channel degasser MCGV (Multi-Channel- Gradient-Valve) Figure 1 Agilent 129 Infinity Quaternary Pump. Table 1 Instruments used for the performance experiments. Inlet-weaver Results obtained on the Agilent 129 Infinity Binary Quaternary Pump and the 126 Infinity Quaternary Pump were compared. Experimental The instruments used are listed in Table 1. Pump head Switchable jet weaver mixer Multipurpose valve In-line filter Module Agilent 129 Infinity LC System Product number Agilent 126 Infinity LC Product number Binary pump G422A Quaternary pump G424A G1311B Autosampler G4226A G1329B Autosampler thermostat G133B G133B Thermostatted column compartment G1316C G1316C Diode array detector G4212A G4212B 2

Chromatographic conditions for step gradients Column substitute: Restriction capillaries providing a backpressure of approximately 17 bar Mobile phases: A = water +2% isopropanol, B = water +2% isopropanol +1 mg/l caffeine Flow rate: 1 ml/min Restrictor temperature: 36 C Detection: Signal 273 nm/4 nm bandwidth, Ref 38/1 nm, Slit 4 nm, 2 Hz Step gradient 1: % B to 1% B in 1% steps, each step held for 5 minutes Step gradient 2: % B to 1% B in 1% steps, each step held for 5 minutes Chromatographic conditions for conventional gradient runs on a 4.6-mm id column Sample from Sigma Aldrich: Reversed Phase Test Mix, (Order No.: 47641-U), 1 1 ml (uracil, phenol, n,n-diethyl-m-toluamide, toluene) HPLC Gradient System Diagnostic Mix, (Order No.: 48271) 6 1 ml (phenol, methyl parabens, ethyl parabens, propyl parabens, butyl parabens, heptyl parabens, uracil) Sample preparation: Column: Mobile phase: Dilute each sample to 5 ml with water/acetonitrile 1:1. Mix the two diluted samples 1:1. Agilent ZORBAX SB-C18, 4.6 15 mm, 5 µm (p/n 7995218-595) A = Water, B = Acetonitrile Gradient: minutes 2% B, at 1 minutes 95% B Flow rate: 1 ml/min, maximum backpressure of 125 bar Stop time: 12 minutes Post time: 5 minutes Injection volume: 5 µl, exterior needle wash 3 seconds Column temperature: 3 C DAD: 254/4 nm, Ref 4/1 nm, Flow cell: 1 mm, Peak width: <.25 minutes (1 Hz) Chromatographic conditions for conventional isocratic runs on a 4.6-mm id column Sample: Isocratic sample (p/n 18-6877) Column: Agilent ZORBAX SB-C18, 4.6 15 mm, 1.8 µm (p/n 829975-92) Isocratic: 3/7 = A/B Flow rate: 1.2 ml/min Stop time: 8 minutes Injection volume: 1 µl, exterior needle wash 3 seconds Column temperature: 4 C DAD: 254/4 nm, Ref 38/8 nm, Peak width: <.25 minutes, (1 Hz) Chromatographic conditions for fast runs on a 3-mm id column Sample: RRLC Checkout sample (p/n 5188-6529) Column: Agilent Poroshell 12 EC-C18, 3 5 mm, 2.7 µm (p/n 699975-32) Gradient: minutes 3% B, 1 minute 95% B Flow rate: 3 ml/min, 55 bar max pressure Stop time: 1.5 minutes Post time: 1 minute Injection volume: 1 µl, exterior needle wash 3 seconds Column temperature: 6 C DAD: 245/1 nm, Ref 4/1 nm, Flow cell: 1 mm, Peak width: <.625 minutes (4 Hz) Chromatographic conditions for conventional runs on a 2.1-mm id column Column: Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 1 mm, 1.8 µm (p/n 959758-92) Sample: Sample from Sigma Aldrich: Reversed Phase Test Mix (Order No.: 47641-U), 1 1 ml (uracil, phenol, n,n-diethyl-m-toluamide, toluene) HPLC Gradient System Diagnostic Mix (Order No.: 48271) 6 1 ml (phenol, methyl parabens, ethyl parabens, propyl parabens, butyl parabens, heptyl parabens, uracil) Sample preparation: Dilute each sample to 5 ml with water/acetonitrile 1:1. Mix the two diluted samples 1:1. Gradient: At minutes 1% ACN, at 1 minutes 9% ACN Flow rate:.3 ml/min at 83 bar max pressure Oven temperature: 5 C DAD: 254/1 nm, REF 36/1 nm, 2 Hz Injection volume: 1 µl, exterior needle wash 3 seconds Chromatographic conditions for fast runs on a 2.1-mm id column Column: Agilent ZORBAX RRHD SB C18, 2.1 5 mm, 1.8 µm (p/n 8577-92) Sample: Set of 9 compounds, 1 ng/µl each, dissolved in water/acn (65/35), 1. acetanilide, 2. acetophenone, 3. propiophenone, 4. butyrophenone (2 ng/ml), 5. benzophenone, 6. valerophenone, 7. hexanophenone, 8. heptanophenone, 9. octanophenone Gradient: At minutes 1% ACN, at 1 minute 95% ACN Flow rate: 1 ml/min Oven temperature: 45 C DAD: 245/1 nm, REF 36/8 nm, 8 Hz Injection volume:.5 µl exterior needle wash 3 seconds 3

Software Agilent OpenLAB CDS ChemStation Edition C.1.4 Results and discussion The system delay volume, the mixing noise, the composition accuracy and the composition precision were measured using step gradients. For these experiments, the 129 Infinity Quaternary Pump was used without optional mixer. Norm. 5 4 3 2 1 1 2 3 4 Norm. 175 15 125 1 75 5 25 5 6 Red: Agilent 129 Infinity Binary Pump Blue: Agilent 129 Infinity Quaternary pump Green: Agilent 126 Infinity Quaternary Pump Black: programmed gradient min 1 12 14 16 18 2 22 24 min Experiment 1 Two step gradient experiments were performed: From to 1% tracer in 1% steps, each step held for 5 minutes From to 1% tracer in 1% steps, each step held for 5 minutes The same step gradient conditions were applied to the 129 Infinity Quaternary and Binary Pump and the 126 Infinity Quaternary Pump. Figure 2 shows the step gradient from % tracer up to 1% tracer in 1% steps for the three instruments. Table 2 summarizes the results for the three step gradients. The 129 Infinity Quaternary Pump provided comparable results with the 129 Infinity Binary Pump except for the delay volume and the composition accuracy. The 126 Infinity Quaternary Pump also showed excellent results, but the system delay was significantly larger compared to the other two LC systems. Figure 2 Step gradient from to 1% tracer in 1% steps on an Agilent 129 Infinity Binary and 129 Infinity Quaternary Pump and on an Agilent 126 Infinity Quaternary Pump. Table 2 Step gradient results. 126 Infinity Quaternary Pump 129 Infinity Quaternary Pump 129 Infinity Binary Pump Ripple at 1% step (related to 1% step).8%.1%.12% Ripple at 5% step.8%.1%.2% Ripple at 9% step.8%.6%.5% Precision of 1% step (for three cons. runs).18% RSD.1% RSD.14% RSD Precision of 5% step (for three cons. runs).6%rsd.2%.2% RSD System delay volume approx. 9 µl approx. 42 µl approx. 17 µl Composition accuray (specification) NA ±.4% ±.35% 4

The wide gradient range of the 129 Infinity Quaternary Pump from to 1% provided an excellent step gradient profile even from to 1% tracer in 1% steps. Steps 1, 2, and 3 were well defined, see Figure 3 for the 129 Infinity Binary Pump. The 126 Infinity Quaternary Pump provided optimum results from the 3% step onwards. This is reflected in the specified range from 5 to 95% organic for the 126 Infinity Quaternary Pump. The step gradient profiles proved that the 129 Infinity Quaternary Pump showed a similar profile as the 129 Infinity Binary Pump even for the demanding range from to 3% tracer. The 126 Infinity Quaternary Pump profile was different especially at the steps from to 3%. Precision of retention times was evaluated applying different chromatographic conditions on all three pumps using five additional experiments: Experiment 2 Conventional gradient runs were done on a ZORBAX SB-C18 4.6 x 15 mm, 5-µm column with 1 ml/min flow rate. Figure 4 shows an overlay of the chromatograms obtained using the 129 Infinity Quaternary Pump with Norm. 5 4 3 2 1 Norm. and without the optional mixer, 129 Infinity Binary Pump and the 126 Infinity Quaternary pump. The lowest delay volume was achieved by the 129 binary pump, as expected, followed by the 129 Infinity Quaternary Pump without the optional mixer. 25 2 15 1 5 Red: Agilent 129 Infinity Binary Pump Blue: Agilent 129 Infinity Quaternary Pump Green: Agilent 126 Infinity Quaternary Pump Black: programmed gradient 1 12.5 15 17.5 2 22.5 25 27.5 min 1 2 3 4 5 6 min Figure 3 Step gradient from to 1% tracer in 1% steps on an Agilent 129 Infinity Binary and Quaternary Pump and on an Agilent 126 Infinity Quaternary Pump. Experiment 2: Conventional gradient runs on 4.6 15-mm column applying 1 ml/min flow rate Experiment 3: Conventional isocratic runs on 4.6 15-mm id column applying 1.2 ml/min flow rate mau 1 8 6 4 Agilent 126 Infinity Quaternary Pump Agilent 129 Infinity Quaternary Pump with mixer Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Infinity Binary Pump 2 1 3 4 5 6 7 8 9 Experiment 4: Fast gradient runs on 3 5-mm column applying 3 ml/min flow rate 2 2 4 6 8 1 min Experiment 5: Conventional gradient runs on 2.1 1-mm column applying.3 ml/min flow rate Experiment 6: Fast gradient runs on 2.1 5-mm column applying 1.2 ml/min flow rate Column: Agilent ZORBAX SB C18, 4.6 15 mm, 5 µm Gradient: minutes 2% B, at 1 minutes 95% B Flow rate: 1 ml/min, maximum backpressure of 125 bar Figure 4 Overlay of the four chromatograms applying the conventional gradient, 1. uracil, 2. phenol, 3. methyl paraben, 4. ethylparaben, 5. propylparaben, 6. n,n-diethyl-m-toluamide, 7. butylparaben, 8. toluene, 9. heptylparaben. The precision data are based for all experiments on six consecutive runs. 5

Regarding retention time precision, the 129 Infinity Quaternary Pump with mixer and the 129 Infinity Binary Pump showed comparable performance, see Figure 5. The 129 Infinity Quaternary Pump without mixer also provided excellent precision data. The 126 Infinity Quaternary Pump showed very good results, which were far below the 129 Infinity Quaternary Pump specification. RSD RT (%).16.14.12.1.8.6 Precision of Retention time for conventional runs Specification Agilent 129 Infinity Binary Pump Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Infinity Quaternary Pump with mixer Agilent 126 Infinity Quaternary Pump The relative standard deviation range of the retention times for the 129 Infinity Quaternary Pump with optional mixer ranged from.22 down to.75% RSD and without mixer from.46 to.444% RSD. The range for the 129 Infinity Binary Pump was.53 to.17% RSD, and the range for the 126 Infinity Quaternary Pump was.74 to.59% RSD. Experiment 3 Conventional isocratic runs were done applying 1.2 ml/min on an Agilent ZORBAX SB-C18, 4.6 15 mm, 1.8 µm column. These conventional isocratic conditions were applied on the 126 Infinity Quaternary Pump, and on the 129 Infinity Quaternary Pump without and with the optional mixer, see Figure 6. As expected, after applying isocratic conditions, the retentions did not differ for the three instrument configurations. Figure 5 Precision of retention times for the Agilent 129 Infinity Binary and Quaternary Pump (with and w/o optional mixer) and for the Agilent 126 Infinity Quaternary Pump. mau 14 12 1 8 6 4 2.4.2 1 2 3 4 5 Peak number 6 7 8 9 Agilent 129 Infinity Quaternary Pump with mixer Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 126 Infinity Quaternary Pump 1 2 3 4 5 6 7 8 9 min Figure 6 Overlay of chromatograms using conventional isocratic conditions, 1. dimethylphthalate, 2. diethylphthalate, 3. biphenyl, 4. o-terphenyl. 6

The maximum standard deviation of retention times for the 129 Infinity Quaternary Pump was.1% RSD with a mixer and.51% RSD without a mixer for dynamically mixed mobile phases, see Figure 7. The 126 Infinity Quaternary Pump showed a maximum relative standard deviation for the retention times of.16% RSD. Experiment 4 Fast runs below 1.2 minutes run time were done on an Agilent Poroshell 12 EC C18, 3 5 mm, 2.7 µm column applying a flow rate of 3 ml/min. All four instrument configurations were tested with these demanding chromatographic conditions. The configuration with the lowest delay volume, the 129 Infinity Binary LC, provided the shortest run times, see Figure 8. Whether the first peaks eluted isocratic or were affected by the gradient was determined by the delay volume of the used LC instrument. RSD RT (%).14.12.1.8.6.4.2 Precision of retention times for isocratic runs (dynamically mixed) Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Infinity Quaternary Pump with mixer Agilent 126 Quaternary Pump Dimethylphthalate Diethylphthalate Biphenyl o-terphenyl Column: Agilent ZORBAX SB C18, 4.6 mm 15 mm, 1.8 µm Isocratic: 3/7 = A/B Flow rate: 1.2 ml/min Figure 7 Precision of retention times for isocratic conditions comparing an Agilent 129 Infinity Quaternary and an Agilent 126 Infinity Quaternary pump. mau 6 5 1 2 3 4 5 6 7 8 9 4 3 2 1.2.4.6.8 1 1.2 min Agilent 126 Infinity Column: Agilent Poroshell 12 EC C18, 3 5 mm, 2.7 µm Quaternary Pump Agilent 129 Infinity Gradient: min 3% B, 1 min 95% B Quaternary pump with mixer Flow rate: 3 ml /min, 55 bar max pressure Agilent 129 Infinity Quaternary pump w/o mixer Agilent 129 Infinity Binary Pump Figure 8 Overlay of chromatograms obtained using high flow rate, 1. Acetanilide, 2. Acetophenone, 3. Propiophenone, 4. Butyrophenone, 5. Benzophenone, 6. Valerophenone, 7. Hexanophenone, 8. Heptanophenone, 9. Octanophenone. 7

The precision for all instrument configurations was excellent for these demanding chromatographic conditions. Nevertheless, the 126 Infinity Quaternary LC configuration did not show the same extremely good overall precision of retention times as the 129 Infinity Binary and Quaternary LC configurations. The 129 Infinity with mixer and the binary pump showed comparable results, see Figure 9. The relative standard deviation range of the retention times for the 129 Infinity Quaternary Pump was.27 to.57% RSD with mixer and.132 to.22% RSD without mixer. The range for the 129 Infinity Binary Pump was.214 to.76% RSD and the range for the 126 Infinity Quaternary Pump was.29 to.28% RSD. SD (sec) 1.2 1..8.6.4.2.132.177.2136.2124 1 2 3 4 Figure 9 Precision of retention times for high flow rates. Standard Deviation for an Agilent 129 Quaternary Pump in seconds.1644.1482.1386.1422 SD specification SD (sec).1218 5 6 7 8 9 Peak number The first peak had a retention time as short as approximately.13 minutes and the standard deviation was as low as.13 seconds, which is far below the specification, see Figure 1. Considering these extremely low standard deviation values, the relative standard deviation results were emphasized impressively. RSD RT (%).25.2.15.1 Precision of RT for fast gradients runs at 3 ml/min flow rate Agilent 129 Infinity Binary Pump Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Infinity Quarternary Pump with mixer Agilent 126 Infinity Quaternary Pump.5 1 2 3 4 5 Peak number 6 7 8 9 Figure 1 Standard deviation of retention times. 8

Experiment 5 A conventional application on a narrow bore column was done using the Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 1 mm, 1.8 µm applying a flow rate of.3 ml/min. The 129 Infinity Quaternary Pump is well suited for applications on long narrow bore columns at moderate flow rates. This type of application is frequently used in combination with mass spectrometry. Figure 11 shows an overlay of the obtained chromatograms. The delay volume was influencing the peak elution time, but for applications, where resolution, not speed, is the first criteria, this is less important. The 129 Infinity Quaternary LC showed comparable results with the 129 Infinity Binary LC, for the applied chromatographic conditions, see Figure 12. mau 7 6 5 4 3 2 1 Column: Agilent ZORBAX RRHD SB C18, 2.1 1 mm, 1.8 µm Mobile phase: water/acetonitrile Gradient: at minutes 1% ACN, at 1 minutes 9% ACN Flow rate:.3 ml/min at 83 bar max pressure Agilent 126 Infinity Quaternary Pump Agilent 129 Infinity Quaternary pump with mixer Agilent 129 Infinity Quaternary pump w/o mixer Agilent 129 Infinity Binary Pump 2 4 6 8 1 12 14 Figure 11 Overlay of chromatograms using a narrow bore column and.3 ml/min flow rate, compounds, see Figure 4. min The 126 Infinity Quaternary LC configuration showed results which were partially above the specification of the 129 Infinity Quaternary Pump. The relative standard deviation range of the retention times for the 129 Infinity Quaternary Pump was.116 to.14% RSD with mixer and without mixer.84 to.18% RSD. The range for the 129 Infinity Binary Pump was.91 to.11% RSD and the range for the 126 Infinity Quaternary Pump was.291 to.75% RSD. RSD RT (%).35.3.25.2.15.1.5 Precision of RT at low flow rates on 2.1 mm id column Agilent 129 Infinity Binary Pump Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Infinity Quaternary Pump with mixer Agilent 126 Infinity Quaternary Pump Specification 1 2 3 4 5 Peak number 6 7 8 9 Figure 12 Precision of retention times using a narrow bore column. 9

Experiment 6 Fast runs on 2.1 x 5-mm column applying 1.2 ml/min flow rate proved the usability of the 129 Infinity Quaternary Pump for fast measurements on narrow bore columns, which are frequently used for screening in connection with mass spectrometry. The delay volume influenced the elution times significantly, see Figure 13. The 129 Infinity Binary System with its delay volume of only 17 µl delivered all peaks within 1.25 minutes. The same conditions applied on the 129 Infinity Quaternary System caused an increase of the total run time of.35 minutes due to the delay volume of 42 µl. Typically, the 129 Infinity Binary Pump is best suited for all gradients or isocratic applications using narrow bore columns applying high flow rates with short run times. The precision of retention times was excellent for both systems, see Figure 14. The standard deviation for the first peak of the 129 Infinity Quaternary Pump, for example, was as low as.877 minutes (.53 seconds), which is far below the specification. This shows the excellent relative standard deviations values shown in Figure 14. mau 8 7 6 5 4 3 2 1 Column: Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 5 mm, 1.8 µm Mobile phase: water/acetonitrile Gradient: at minutes 1% ACN, at 1 minute 95% ACN Flow rate: 1 ml/min Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Infinity Binary Pump.25.5.75 1 1.25 1.5 1.75 2 Figure 13 Comparison of chromatograms obtained on the Agilent 129 Infinity Binary and Quaternary LC Systems applying fast chromatographic conditions using short 2.1-mm id columns, compounds, see Figure 8..2.18.16.14.12.1.8.6 Precision of RT for fast gradient runs on 2.1-mm id columns Agilent 129 Infinity Quaternary Pump with mixer Specification of the Agilent 129 Quaternary Pump Agilent 129 Infinity Binary Pump Agilent 129 Infinity Quaternary Pump w/o mixer Agilent 129 Quaternary Pump with mixer min The relative standard deviation range of the retention times for the 129 Infinity Quaternary Pump with mixer was.12 to.2% RSD, and without mixer.156 to.17% RSD. The range for the 129 Infinity Binary Pump was.24 to.1% RSD. This shows the outstanding performance of the 129 Infinity Binary Pump..4.2 1 2 3 4 5 Peak number Figure 14 Precision of retention times for fast runs using 2.1-mm id columns. 6 7 8 9 1

Conclusions The Agilent 129 Infinity Quaternary Pump provided excellent performance with respect to precision of retention times. Excellent results were obtained for conventional gradient and isocratic chromatographic conditions, for fast gradient conditions on 3-mm id columns and 3 ml/min flow rate and for applications using narrow bore columns using conventional conditions. For these applications, the 129 Infinity Quaternary Pump was comparable to the 129 Infinity Binary Pump. For fast chromatographic conditions, applying high flow rates on short 2.1-mm id columns the 129 Infinity Quaternary Pump showed excellent results with respect to precision of retention times, however, the 129 Infinity Binary Pump showed exceptional results with respect to elution times as well as precision of retention times. The applied step gradients showed excellent accuracy and precision over the complete gradient range from to 1% tracer. 11

www.agilent.com/chem/ 129quatpump Agilent Technologies, Inc., 212 Published in the USA, October 1, 212 5991-1193EN