Analysis and Quantification of Mycotoxins in Cereals Using MRM HR on the SCIEX X500R QTOF System with SCIEX OS

Similar documents
Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

Chemical Components Identification of Cistanche Deserticola Using the X500R QTOF System

Rapid and Sensitive Analysis of Antibiotics in Children s Urine Using the X500R QTOF System

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Powerful Scan Modes of QTRAP System Technology

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Simultaneous Determination of 26 Mycotoxins in Sesame Butter Using Modified QuEChERS Coupled with UHPLC-ESI/Triple Quadrupole Mass Spectrometry

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Multi-residue analysis of pesticides by GC-HRMS

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Analyst Software. Automatic Optimization Tutorial

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

Untargeted Screening on the X500R QTOF

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

WADA Technical Document TD2003IDCR

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Overview. Introduction

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

4-Methyldiethcathinone (4-MDEC)

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

1-(1,3-benzodioxol-5-yl)-2-(ethylamino)hexan-1-one

Analyst Software. Peptide and Protein Quantitation Tutorial

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

2-[[2-(4-ethyl-2,5-dimethoxy-phenyl)ethylamino]methyl]phenol

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Determination of Pharmaceuticals in Environmental Samples

Key Features of Targeted Metabolite Screening. Using the TripleTOF System, an Accurate Mass Metabolite Spectral Library and MasterView Software

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

4-methyl-1-phenyl-2-pyrrolidin-1-yl-pentan-1-one

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

PHL-IL and Mycotoxins. Andreia Bianchini, PhD University of Nebraska - Lincoln

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Speaker/Author: W.P. Nxumalo*, ** Co-authors: Y. Naudé**, and M. Fernandes-Whaley*

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

5Cl-AKB48 and 5Br-AKB48

N-(1-carbamoyl-2-methyl-propyl)-1-(5-chloropentyl)indazole-3- carboxamide

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Definitive EtG/EtS LC-MS/MS Analysis:

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Automated and accurate component detection using reference mass spectra

Multi-Mycotoxin testing A routine approach

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Methyl 1-[(4-fluorophenyl)methyl]indazole-3-carboxylate

How to screen and identify unexpected and unwanted compounds in food

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Sampling, sample preparation and analyte isolation

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

This document is a preview generated by EVS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Forensic and toxicological drug screening using liquid chromatography-tandem mass spectrometry with MS/MS library based identification

Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

Theory and Practical Consequences.

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Transcription:

Analysis and Quantification of Mycotoxins in Cereals Using MRM HR on the SCIEX X500R QTOF System with SCIEX OS Yang Zong 1, Cheng Haiyan 1, Zhong Hongjian 2, Li Lijun 1, Jin Wenhai 1 Phil Taylor 3 1. SCIEX, Asia Pacific Application Support Center (Shanghai); 2. Henan Academy of Agricultural Sciences; 3 SCIEX, Warrington UK. Introduction Mycotoxins are toxic secondary metabolites produced by fungi under favorable conditions which can cause acute and chronic illness in humans and animals [1]. The most common mycotoxins that threaten human health include aflatoxin (AF), deoxynivalenol (DON), ochratoxin A (OTA), fumonisin (FB), zearalenone(zen), and T-2 toxin(t2) [2], which are widely found in many grains and oils, and their products. To date, many countries and regions have set strict mycotoxin limits [3], and China has revised its Food Safety Law in 2015 to explicitly include biotoxins in contaminants of major concern for the first time. China has also set limits for mycotoxins such as deoxynivalenol (DON), zearalenone (ZEN), aflatoxin B1(AFB1), and ochratoxin (OTA); for instance, the state food safety standard for cereals set the limit for ZEN at 60µg/kg, aflatoxin B1(AFB1) at 5-20µg/kg, and OTA at 5µg/kg[4]. Abnormal climate conditions and insect damage have caused extensive mycotoxin contamination of grain in recent years, leading to severe damage and huge losses. Reports show that in China, 31 million tons of grain are contaminated with mycotoxins annually during production, storage, and transport; this accounts for approximately 6.2% of total grain production. Rapid, highthroughput, and highly accurate mycotoxin detection methods with simple sample pretreatment are critical to ensuring the quality and safety of China s grain supplies, strengthening the monitoring for mycotoxins in grain, as well as protecting human and animal health. The X500R QTOF system combines high mass resolution and mass accuracy with quantitative capabilities. As with triple quadrupole- mass spectrometry, it is possible to collect multiple reaction monitoring (MRM) ion pair data; however, the QTOF can obtain high-resolution MSMS ions with ultra-rapid scanning speeds (100Hz) in a mode called MRM HR scanning mode [Fig 2]. This scanning mode offers high selectivity in the same manner as quadrupole devices and provides quantitative analysis through high-resolution secondary product ion peak areas and qualitative analysis through the calculation of ion abundance ratios. This capability helps to avoid false positives or other matrix effects due to interferences. Figure 1. Scheduled MRM HR selectivity compared to TOFMS for FB1. Increased selectivity is demonstrated by monitoring the ion transition using MRM HR compared to monitoring solely the target precursor ion and reduces the potential for false negatives or incorrect integration of matrix peaks. Key Advantages: MRM HR for Mycotoxins Sample pretreatment is simple and fast, only 10 minutes from sample preparation to testing Fifteen common mycotoxins are included in the panel; because mycotoxin reference products are expensive and difficult to obtain, this study produced high-resolution MS/MS spectral data allowing identification without reference products Scheduled MRM HR scanning mode, retention time locking, and established MRM HR ion parameters greatly reduce matrix effects and increase method reproducibility and accuracy. The method is easily applied, saving on development time and costs. p 1

Sample Preparation Method Figure 2. High Resolution MRM (MRM HR ) ion path. Experimental methods Table 1. Chromatography Conditions LC Column Mobile Phase A Mobile Phase B Flow Rate Column Temperature Shim-pack XR ODS, 2.0x 75mm 1.6µm Water with 2mM ammonium acetate and 0.1% Formic Acid Pure acetonitrile 0.3mL/min 40 C Injection Volume 2µL This sample preparation method is simple, fast, and is suitable for wheat, corn, rice, and sesame; a total of 10 grain samples were collected from different production regions. Compound information Compound CAS no. Formula Aflatoxin B1 AFB1 001162-65-8 C17H12O6 Aflatoxin B2 AFB2 007220-91-7 C17H14O6 Aflatoxin G1 AFG1 001165-39-5 C17H12O7 Aflatoxin G2 AFG2 007241-98-7 C17H14O7 Ochratoxin A OTA 000303-47-9 C20H18ClNO6 Table 2. Mass Spectrometry Conditions Scanning Method Scanning range 100 1000 Da Scheduled MRM HR CUR 30 psi CAD 7 psi IS Voltage 5500V / -4500V TEM 550 C GAS 1 55 psi GAS 2 50 psi T-2 toxin T-2 021259-20-1 C24H34O9 Fumonisin B1 FB1 116355-83-0 C34H59NO15 Fumonisin B2 FB2 116355-84-1 C34H59NO14 Fumonisin B3 FB3 136379-59-4 C34H59NO14 Aflatoxin M1 AFM1 6795-23-9 C17H12O7 Nivalenol NIV 023282-20-4 C15H20O7 Deoxynivalenol DON 051481-10-8 C15H20O6 3-Acetyl-deoxynivalenol 3-AcDON 876926-22-6 C17H22O7 15-Acetyl-deoxynivalenol 15-AcDON 088337-96-6 C17H22O7 Zearalenone ZEN 017924-92-4 C18H22O5 X500R SCIEX OS MRM HR Work Flow The quantification capabilities of the X500R QTOF highresolution mass spectrometer are comparable to those of the triple quadrupole MRM device; however, the SCIEX OS platform design includes a unique MRM HR work flow. User defined MRM ion pair information can be entered similarly to quadrupole devices in the SCIEX OS software [Fig 3]. The precursor ion p 2

window is set to unit resolution, as in quadrupole mode, and the product ions are separated using a TOF mass filter; thus highresolution product ions are produced [Fig 1]. MRM HR ion pair data can be imported directly from high-resolution secondary libraries into MRM HR method lists [Fig. 3]. Data for 5 ion pairs from each compound can be imported based on the sensitivity ranking of the secondary product ion in the library. MRM HR lists with imported data include retention time, as well as unique voltage parameters for each ion pair such as declustering potential (DP) and collision energy (CE). ammonium acetate causes a more obvious signal increase than ammonium formate, while a lower salt concentration (2mmol/L ammonium acetate) has a stronger effect than higher concentrations (5mmol/L and 10mmol/L). FB1 and FB2 show a strong effect on signal production when acid (0.1%) is added. Ultimately, a weak aqueous eluent containing 0.1% formic acid and 2mmol/L ammonium acetate was selected as the mobile phase A. 2. Target ion selection This study optimizes the effects of adduct ions of various toxins in a detailed manner. For example, T-2 toxin has peaks for hydrogen, ammonium, and sodium adducts; given that the sodium peak is not fragile, this study compared hydrogen and ammonium adduct peaks and found that the ammonium peak signal was 30 times that of the protonated peak [Fig 5]. In addition, the effects of formic acid adducts in NIV and DON were greater than those of dehydrogenation peaks [Fig 6]. Fig 3. SCIEX OS Software Scheduled MRMHR Method Editor, Applying Fragment Ion Mass ±10ppm m/z Fig 5. Chromatogram of T-2: ammonium adduct (top) produced a greater absolute signal versus protonated precursor (bottom). Fig 4. Import and Autofill MSMS Scan Information Experimental results 1. Mobile phase optimization Comparison of the influence of mobile phases with varying ratios of modifier on ionization of various mycotoxins shows that Fig 6. Chromatogram of DON: formate adduct (top) produced greater absolute signal than dehydrogenated precursor (bottom). p 3

3. Strong resistance to matrix interference Sample preparation involved a simple liquid-liquid extraction and dilution, which does not remove much background sample matrix and can therefore leave the sample subject to interference effects. Four grain matrices were assessed for matrix effects based on the ratio of the areas of the target in the blank solvent and the matrix solvent. As the ratio approached 100%, matrix effects became insignificant; above 100% matrix effects were considered enhancement ; below 100% there was matrix suppression. Matrix effects of high-resolution primary scanning and MRM HR scanning were also compared. When TOFMS scanning of 15 mycotoxins was performed, matrix effects ranged between 43.1% and 125.6%, and when MRM HR scanning was performed, matrix effects ranged between 88.5% and 109.2%. The results showed that MRM HR scanning is much more selective and resistant to matrix interference than primary TOFMS scanning. Figures 1 and 6 show that after the addition of corn matrix to fumonisin B1, there was strong interference with the TOF-MS primary scan, while the MRM HR scan showed less noise and greater selectivity. Aflatoxin AFB1 is highly sensitive, and matrix inhibition effects are minimal when TOF-MS scanning is performed. Thus, this study shows that MRM HR scanning is more selective and resistant to matrix inhibition, increasing method accuracy and reproducibility, which can in turn greatly decrease positive and false negative results. Fig 6. Scheduled MRMHR selectivity for AFB2 compared to TOFMS; increased selectivity with MRMHR reduces the potential for false negatives or incorrect integration of different peaks. 4. Simultaneous quantification and qualification At concentrations between 0.05 and 50ng/mL, there were strong linear responses for all mycotoxins [Fig 7]; the correlation coefficients were greater than 0.99, fully meeting quantitative analysis requirements. Fig 7. The Regression Calibrations of Mycotoxins SCIEX OS software calculates ion abundance ratios in a similar manner to quadrupole devices, and different compounds have qualitative and quantitative differences in terms of ion abundance ratios. International and EU regulations [Table 3] can be used to determine the range of ratios. p 4

When data of actual samples were analyzed, as shown in Fig 9, they were processed quickly and easily, and the interface was intuitive. The result showed a standard curve, a sample and matrix spiked ion abundance ratio, accuracy, sample concentration, and retention time. One can intuitively determine that wheat produced in a certain area contains aflatoxin B1 (AFB1) based on a sample ion abundance ratio of 0.357, a matrix spiked ion abundance ratio of 0.343, and concentration of 2.5µg/kg. This study examined 10 samples including wheat, corn, etc., from various provinces. See Table 3 for the detection results. Table 4. Mycotoxin content of wheat (n=6) and corn (n=4) from different regions. A blank space in the table indicates not detected. Fig 8. The MRM HR Ratio tolerance limits can be user-defined in the SCIEX OS software Wheat Corn µg/kg 1 2 3 4 5 6 1 2 3 4 AFB1 1.6 1.1 0.6 2.5 2.1 1.9 2.2 6.2 2.1 0.8 AFB2 0.2 0.5 0.9 Table 3. EU regulations from the SANTE/11945/2015 document AFG1 AFG2 OTA T-2 DON 331.1 498.8 210.2 112.2 99.9 120.2 98.9 76.9 88.1 120.3 FB1 FB2 FB3 AFM1 NIV 150.1 122.3 144.5 112.6 90.3 160.4 3-AcDON 13.5 10.6 9.9 13.2 40.2 26.0 16.8 19.2 10.3 11.5 15- AcDON 19.9 11.0 13.4 16.6 54.2 31.9 15.9 18.3 21.9 22.1 ZEN 20.1 15.6 4.8 8.9 9.9 2.7 2.8 4.5 0.9 6.5 Summary Fig 9. Quantification results for AFB1 in wheat as they appear in the SCIEX OS software user interface This report describes the use of the novel SCIEX X500R QTOF system to establish techniques for the detection of common mycotoxins in grain samples. The MRM HR scanning capability offers the high sensitivity of existing quadrupole devices as well as high-resolution quantitative product ion detection (peak area) and qualitative confirmation (ion abundance ratios) at ultra-rapid scanning speeds. Because this capability is selective and resistant to matrix interference, sample preparation techniques were able to be further simplified, decreasing labor-intensive steps and improving work efficiency. Since mycotoxin reference samples are expensive and difficult to obtain, a high-resolution secondary library of mycotoxins was created. This removes the need for reference samples and permits easy identification of mycotoxin species in a sample. In summary, this method is p 5

sensitive, meets international and EU requirements for mycotoxin limits, and fully complies with customer analytical requirements. It can serve as a valuable reference in the application of high-resolution instrumentation to the detection of mycotoxins in grain samples. This study included a total of 10 samples of wheat and corn from different production regions. Testing showed AFB1, DON, NIV, ZEN, 3-AcDON and 15-AcDON were widely distributed throughout the wheat and corn sample, illustrating the prevalence of mycotoxin contamination of grain samples and highlighting the need for greater monitoring and oversight. References 1. Hussein H S, Brasel J M. Toxicity, metabolism, and impact of mycotoxins on humans and animals [J]. Toxicology, 2001, 167(2): 101-134 2. Sulyok M, Berthiller F, Kraska R, et al. Development and validation of a liquid chromatography/tandem mass spectrometric method for the determination of 39 mycotoxins in wheat and maize [J] Rapid Commun Mass Spectrom, 2006, 20(18): 2649-2659 3. Commission Regulation (EC) No 1881/2006, Setting Maximum Level for Certain Contaminants in Food Stuffs 4. GB2761-2011, People s Republic of China State Standard for Mycotoxin Limits in Food Products AB Sciex is doing business as SCIEX. 2018 AB Sciex. For Research Use Only. Not for use in diagnostic procedures. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Document number: RUO-MKT-02-8863-A p 6