Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Similar documents
M [scale units/s] of the system

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Calibration of a pencil ionization chamber with and without preamplifier

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Retrieval Of GammaCell 220 Irradiator Isodose Curves With MCNP Simulations And Experimental Measurements

CALIBRATION OF IONIZATION CHAMBERS FOR COMPUTED TOMOGRAPHY BEAMS IN BRAZIL: THE PRESENT REALITY*

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Proposed model for PGAA at TRIGA IPR- R1 reactor of CDTN

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

Bureau International des Poids et Mesures

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

City University of Hong Kong

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Commissioning of the Beta Secondary Standard (BSS2)

Rad T 290 Worksheet 2

Intercomparison of Ionization Chambers in Standard X-Ray Beams, Radioprotection Level

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Derivation of factors for estimating the scatter of diagnostic x-rays from walls and ceiling slabs

X-ray Interaction with Matter

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Monte Carlo Simulations for a Preliminary Design of TRIGA IPR-R1 PGAA Facility

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y

Regulatory Considerations in the Licensing of a Mobile Backscatter X-ray Device used in Security Screening

On-Site Calibration Method of Dosimeter Based on X-Ray Source

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam

CALIBRATION OF DOSEMETERS USED IN MAMMOGRAPHY WITH DIFFERENT X RAY QUALITIES: EUROMET PROJECT NO. 526

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Use of a radioactive check device for redundancy check of ionization chambers

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

Training program for x-ray inspection film (RT)

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical

Bureau International des Poids et Mesures

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

Basic physics Questions

Atomic and nuclear physics

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Dose Rate Levels around Industrial Gamma Sources

EXPERIMENTAL DETERMINATION OF NEUTRONIC PARAMETERS IN THE IPR-R1 TRIGA REACTOR CORE

Personal Dose Monitoring System

We have seen how the Brems and Characteristic interactions work when electrons are accelerated by kilovolts and the electrons impact on the target

X-Ray Spectrum Generation for a Multiphase Flow Meter

Progress in Nuclear Science and Technology, Volume 6,

Atomic and nuclear physics

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Measurement Science and Standards. Ernesto Mainegra-Hing, Hong Shen, Malcolm McEwen Ionizing Radiation Standards Report PIRS-2350 June 2017

Atomic and nuclear physics

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Chapter 10 Acceptance Tests and Commissioning Measurements

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

RADIATION SAFETY TRAINING SEALED SOURCES

Physics of Radiography

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima

Reliability and Traceability in Radiation Calibration

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

Neutron collimator design of neutron radiography based on the BNCT facility arxiv: v1 [physics.ins-det] 3 May 2013

COMPARISON OF PERSONNEL RADIATION MONITORING DOSIMETERS DESIGNED FOR MEDICAL FIELD

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

Simulation Modeling in Dosimetry

Computation of energy imparted in diagnostic radiology

Physics of Radiography

Analyzing Radiation. Pre-Lab Exercise Type of Radiation Alpha Particle Beta Particle Gamma Ray. Mass (amu) 4 1/2000 0

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

DICOM Correction Item

Application of Optical Fiber in Radiation Detection Systems. Francisco Antônio Brandão Junior, Antonella Lombardi Costa* and Arno Heeren de Oliveira

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Radiation Protection & Radiation Therapy

Chapter V: Cavity theories

Health Physics Services Ltd

Accelerators and radiation spectra

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

Initial Certification

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

SECTION 8 Part I Typical Questions

PHYS 3650L - Modern Physics Laboratory

Non-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen

Shell Atomic Model and Energy Levels

Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes

Dosimetry: Electron Beams

ISO INTERNATIONAL STANDARD

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES

Chapter 30 X Rays GOALS. When you have mastered the material in this chapter, you will be able to:

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector

Radionuclide Imaging MII Detection of Nuclear Emission

Transcription:

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory Annibal T. Baptista Neto, Carlos M. A. Soares, Teógenes A. da Silva Nuclear Technology Development Center, Radiation Dosimetry Service, Av. Presidente Antonio Carlos, 6627, Campus da UFMG, Pampulha, Belo Horizonte, MG, Brazil Abstract. One important parameter related to the protection of people exposed to ionizing radiation is the accurate dose measurement. In order to guarantee metrological reliability, the International Organization for Standardization (ISO) specifies x and gamma reference radiations for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy. In this work the implementation of the ISO reference radiations (wide and narrow spectrum series) in the Agfa-Seifert x-ray machine was evaluated. Quality control tests were performed and beam parameters like half-value layer, homogeneity coefficient, and inherent filtration were simulated. Results showed that the control tests complied with the standard requirements. It was conclude that the implemented x-ray beams could be considered similar to the ISO reference radiations and they may be used for the purpose of calibrating dosemeters and doserate meters used in radiation protection field. KEYWORDS: reference radiation; x-ray calibration, ionizing radiation metrology 1. Introduction The exposure of workers and environment to radiation must be monitored in order to comply with regulatory requirements and to help the radiological protection of the people. Gamma and x-ray reference radiations for calibrating dosimeters and for determining their energy response as a function of photon energy were established by the International Organization for Standardization [1]. The x reference radiations are grouped in four series: the low air-kerma rates, the narrow spectrum, the wide spectrum and the high air-kerma rate series. Detailed information on the beam production and dosimetry are provided in the ISO standards [1,2]. The Nuclear Technology Development Center (CDTN), in Belo Horizonte, Brazil, has maintained a metrology laboratory to provide calibrations of dosemeters in gamma beams. After acquiring a 3 kv Agfa-Seifert x-ray machine, steps were done to offer calibration services of ionizing chambers and area radiation monitors as well as irradiation of personal dosemeters in x-ray ISO beams. 2. Methods and Results 2.1 X-ray equipment A Isovolt 3 HS Pantak Seifert x-ray machine that was installed at the CDTN Laboratory was used to have the reference radiations implemented. The characteristics of the x-ray machine are given in Tab. 1 [3]. The machine was covered with a 1 mm lead to shield scattered radiation; a 55 mm diameter hole in the exit of the x-ray allows collimating the beam. A shutter is positioned in the center of this hole; an opening and closure pneumatic command makes possible to control the time exposure without turning the machine off. 1

Table 1: Technical characteristics of the Agfa-Seifert x-ray machine. 2.2 Irradiation geometry Large focal spot Small focal spot Maximum tube voltage 3 kv 3 kv Maximum anode dissipation 4 W 16 W Tube current at maximum tube voltage 13 ma 5 ma Focal spot size 6.3 mm 3. mm Emergent beam angle º Target material tungsten Inherent filtration 7 mmbe Weight (tube only) 35 kg Geometry conditions and x-ray parameters were verified in order to establish the ISO narrow and wide spectrum reference radiations. Figure 1 shows the positioning set-up that was used for implementing the x-ray qualities. A conical lead collimator with 13 mm thickness and 36.5 mm entrance surface diameter and 42 mm exit surface diameter was placed after the x- ray tube. Two additional plan collimators with 5 and mm diameters were used to define the field size and to reduce the scattered radiation. Aluminum, cooper and tin absorbers with not less than 99,9% purity were used to reproduce the reference radiations and to measure the half-value layers (HVL). The radiation field size was enough to cover uniformly the sensitive volume of the reference ionization chamber that was positioned cm to the focal spot; absorbers were exactly positioned in the middle between the focal spot and the ionization chamber. Figure 1: Set-up for implementing the x-ray reference radiations. A PTW-Freiburg ionization chamber model TN314 with 155 mm useful diameter was connected to a Keithley electrometer model 6517A; it was positioned after the first lead collimator and used to monitor the fluctuations in the radiation field, assure the stability of the system and make possible the measurements corrections. An unsealed cylindrical 6 cm 3 Radcal Corporation model RC6 ionization chamber (Fig. 2) with small energy dependence was connected to a Keithley electrometer model 6517A for beam measurements. 2

Figure 2: The Radcal Corporation model RC6 ionization chamber Dimensions in milimeters Center of Sensitive Volume Active Volume 2.3 Radiation field size The radiation field size was visualized with an intensifying screen arranged behind the ionization chamber that was positioned at 1 m distance from the focal spot. The diameter of the field was estimated to be mm for the used geometry. Figure 3 shows a Radcal ionization chamber RC6 in the centre of the radiation field. Figure 3: Radiation field and the ionization chamber images in the intensifying screen 2.4 Radiation beam uniformity The uniformity is used for delimiting the beam area to be used for irradiation and calibration of dosemeters. The beam uniformity was verified for all radiation qualities in terms of air kerma under the same geometry mentioned before. The Radcal ionization chamber RC6 was moved in the horizontal and vertical direction, perpendicular to the radiation beam; the intersection of these lines was considered as the center of the radiation beam. The air kerma was verified at distances not larger than 5 mm along the horizontal and vertical lines. In all situations, results assured that mm and 9 mm diameter radiation fields had uniformity inside ± 3% and ± 5%, respectively. Figures 4 to 9 show the beam uniformity of the N to N15 narrow reference radiations series. 3

Figure 4: Beam uniformity of N beam quality Figure 5: Beam uniformity of N beam quality Beam Uniformity - N Beam Uniformity - N - - - - - - - - Figure 6: Beam uniformity of N beam quality Figure 7: Beam uniformity of N beam quality Beam Uniformity - N Beam Uniformity - N - - - - - - - - Figure 8: Beam uniformity of N1 beam quality Figure 9: Beam uniformity of N15 beam quality Beam Uniformity - N1 Beam Uniformity - N15 - - - - - - - - The tube potential and the inherent filtration at kv were previously measured [4]; the scattered radiation conditions were verified to all reference radiations and they showed to comply with the ISO requirements. 4

2.5 Implementation of the reference radiation Figure 1: Attenuation curve N3 beam quality 9 7 5 3 1 y = A1*exp(-x/t1) + y Chi^2/DoF =.18668 R^2 =.99976 y 5.31319 A1 94.44769 t1 1.57494,,5 1, 1,5 2, 2,5 3, Thickness (mmal) 1st HVL = 1.177 mmal 2nd HVL = 1.292 mmal Figure 11: Attenuation curve N beam quality The verification of the x-ray spectra was made by experimental measurements of the 1 st and 2 nd HVLs in the atenuation curve in term of aluminum only for the 3 kv beam quality. For all others series the cooper was used to establish the atenuation curve. It was posible to compare the beams implanted in CDTN and the reference beams specified by the ISO standard [1]. Figures 1 to 18 show the atenuation cuve for all series.table 2 presents the comparison of parameters of the reference radiations. In all cases, the 1 st and 2 nd HVLs agreed within ± 5%. Figure 12: Attenuation curve N beam quality 9 7 5 3 y = A1*exp(-x/t1) + y Chi^2/DoF =.26463 R^2 =.99966 y 8.8148 A1 91.553 t1.1348 9 7 5 3 y = A1*exp(-x/t1) + A2*exp(-x/t2) + y Chi^2/DoF =.713 R^2 =.99992 y -1.97367 A1 24.78 t1.13783 A2 77.63423 t2.46193 1 1,,2,4,6,8,1,12,14,16,18, 1st HVL =.8 2nd HVL =.939,,5,1,15,,25,3,35,,45,5,55 1st HVL =.2289 2nd HVL =.277 Figure 13: Attenuation curve N beam quality Figure 14: Attenuation curve N beam quality 9 7 5 3 y = A1*exp(-x/t1) + A2*exp(-x/t2) + y Chi^2/DoF =.3331 R^2 =.99962 y -4.89749 A1 98.54982 t1.9835 A2 6.36619 t2.9556 9 7 5 3 y = A1*exp(-x/t1) + A2*exp(-x/t2) + y Chi^2/DoF =.28453 R^2 =.99961 y 1.2761 A1 49.42568 t1 1.62794 A2 49.42568 t2 1.628 1 1,,2,4,6,8 1, 1,2 1st HVL =.576 2nd HVL =.597,,2,4,6,8 1, 1,2 1,4 1,6 1,8 2, 2,2 2,4 1st HVL = 1.152 2nd HVL = 1.171 5

Figure 15: Attenuation curve N1 beam quality Figure 16: Attenuation curve N15 beam quality y = A1*exp(-x/t1) + A2*exp(-x/t2) + y y = A1*exp(-x/t1) + A2*exp(-x/t2) + y 9 7 5 3 Chi^2/DoF =.18863 R^2 =.99971 y -5.58276 A1 44.98855 t1 1.888 A2.93 t2 3.87458 9 7 5 3 Chi^2/DoF =.7759 R^2 =.9999 y 3.897 A1 48.24735 t1 3.31748 A2 48.24735 t2 3.31748 1 1,,5 1, 1,5 2, 2,5 3, 3,5 1st HVL = 1.762 2nd HVL = 1.1 1 2 3 4 5 1st HVL = 2.43 2nd HVL = 2.568 Figure 17:Attenuation curve W beam quality Figure 18: Attenuation curve W25 beam quality y = A1*exp(-x/t1) + A2*exp(-x/t2) + y y = A1*exp(-x/t1) + A2*exp(-x/t2) + y 9 7 5 3 Chi^2/DoF =.4619 R^2 =.99994 y 5.2325 A1 47.45554 t1 4.15881 A2 47.45554 t2 4.15881 9 7 5 3 Chi^2/DoF =.2913 R^2 =.99996 y 2.9544 A1 45.719 t1 4.91417 A2 51.36861 t2 6.96567 1 1 1 2 3 4 5 6 7 1st HVL = 3.125 2nd HVL = 3.399 1 2 3 4 5 6 7 8 9 1st HVL = 4.258 2nd HVL = 4.555 Table 2: Parameters comparison of the reference radiations implemented in CDTN and established by ISO 1 st HVL (a) 2 nd HVL (a) Reference Radiation ISO 37-1 CDTN ISO 37-1 CDTN N3 1,15 1,18 1,3 1,29 N,84,81,91,94 N,24,23,26,27 N,58,58,62, N 1,11 1,15 1,17 1,17 N1 1,71 1,76 1,77 1, N15 2,36 2,43 2,47 2,57 W 3,8 3,13 3,31 3, W25 4,22 4,26 4, 4,56 (a) values in mmcu, except for N3 in mmal 3. Conclusion The results confirmed the similarity between the radiation beam qualities implanted in CDTN and the specified by ISO 37-1 standard; they certify their applicability for calibrating dosemeters and doserate meters, for determining their response as a function of photon energy, for carrying out 6

performance tests of instruments used in dosimetry for industrial, medical, environmental or any other radiation protection purposes. REFERENCES [1] INTERNATIONAL ORGANIZATION FOR STANDARDIZATION (ISO), X and gamma reference radiations for calibrating and determining their response as a function of photon energy. ISO 37/Part 1: Radiation characteristics and production methods, ISO, Geneva (1996). [2] NTERNATIONAL ATOMIC ENERGY AGENCY (IAEA), International Basic Safety Standards for Protection against Ionizing Radiation and for Safety of Radiation Sources, Safety Series 115, Safety Standards, IAEA, Vienna (1996). [3] AGFA NDTPANTAK SEIFERT GMBH & CO. KG, Operating Instructions Isovolt 1/225/3/45 HS, Arensburg, (3). [4] OLIVEIRA, PAULO M. C. ET AL,. Analysis of characteristic parameters of diagnostic x-ray beams for calibrating dosemeters [in Portuguese] Master Science degree dissertation, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, (8). 7