Bureau International des Poids et Mesures

Similar documents
Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Bureau International des Poids et Mesures

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Bureau International des Poids et Mesures

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Abstract An indirect comparison of the standards for reference air kerma rate for

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Investigation of the standard temperature- pressure correction factor at low x-ray energies

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

First international comparison of primary absorbed dose to water standards in the medium-energy X-ray range

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017

M [scale units/s] of the system

Chapiter VII: Ionization chamber

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel

Updating reference dosimetry a decade after TG-51

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

BUREAU INTERNATIONAL DES POIDS ET MESURES

RECOMBINATION PARAMETERS OF SOME FABRICATED IONIZATION CHAMBERS

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

BUREAU INTERNATIONAL DES POIDS ET MESURES

Ionizing Radiation Dosimetry and Medical Physics

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

Limitations and benchmarks of EGSnrc

Measurement Science and Standards. Ernesto Mainegra-Hing, Hong Shen, Malcolm McEwen Ionizing Radiation Standards Report PIRS-2350 June 2017

General characteristics of radiation dosimeters

Comments on ICRU Report 64: Dosimetry of High-Energy Photon Beams based on Standards of Absorbed Dose to Water

Mass determination of 1 kg silicon spheres for Avogadro project

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

Determination of the Specific Heat Capacity of Graphite Using Absolute and Differential Methods Susanne Picard David Burns Philippe Roger BIPM

Monte Carlo simulations as a tool for the development of a new reference ionisation chamber

Progress Report on Radiation Dosimetry Standards at NMIJ/AIST and Photon Energy Spectra Measured in Fukushima

APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

NRC Activities and Publications, Report to CCRI(I) Meeting, BIPM May

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

The metrology of radioactivity. C. Michotte, BIPM

Accelerators and radiation spectra

Total radiation measurements of thermodynamic temperature

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Chapter 9: Calibration of Photon and Electron Beams

Basic physics Questions

CALIBRATION OF DOSEMETERS USED IN MAMMOGRAPHY WITH DIFFERENT X RAY QUALITIES: EUROMET PROJECT NO. 526

Bureau International des Poids et Mesures Comparison of ozone reference standards of the CHMI and the BIPM

The EPOM shift of cylindrical ionization chambers - a status report Hui Khee Looe 1, Ndimofor Chofor 1, Dietrich Harder 2, Björn Poppe 1

Ionization Chamber. Pocket dosimeter

Electron beam water calorimetry measurements to obtain beam quality conversion factors

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

Bureau International des Poids et Mesures

Final Report August 2010

Comparison of V and 10 V DC Voltage References

Gerhard Stucki, Sandor VörösV

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

1 Introduction. A Monte Carlo study

NPL s progress towards absorbed dose standards for proton beams

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the PTB, Germany and the BIPM in medium-energy x-rays

Approximation of ITS-90 with High Temperature Thermometers up to 1085 C

Electron Impact Ionization in EGSnrc

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Introduction to Blackbody Sources

Non-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen

COOMET Project 277/UA/03 COOMET L.S-1 Conducting of comparisons of comparators with photoelectric microscopes. Final report

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

A NEW REALIZATION OF TERRESTRIAL TIME

Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the PTB, Germany and the BIPM in low-energy x-rays

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

Transcription:

Bureau International des Poids et Mesures Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde, A.V. Oborin and E.N. Yurjatin September 2001 Pavillon de Breteuil, F-92312 Sèvres Cedex

Comparison of the air-kerma standards of the VNIIM and the BIPM in the medium-energy x-ray range D.T. Burns, N.D. Villevalde*, A.V. Oborin* and E.N. Yurjatin* Bureau International des Poids et Mesures, F-92312 Sèvres Cedex *D.I. Mendeleyev Institute for Metrology, Moskovski Prospekt 19, 198005 St. Petersburg, Russian Federation Abstract An indirect comparison has been made between the airkerma standards of the VNIIM and the BIPM in the medium-energy x-ray range. The results show the standards to be in agreement within the stated uncertainty. 1. Introduction An indirect comparison has been made between the air-kerma standards of the D.I. Mendeleyev Institute for Metrology (VNIIM), Russia, and the Bureau International des Poids et Mesures (BIPM) in the x-ray range from 100 kv to 250 kv. A cavity ionization chamber of type PTW M300001 was used as the transfer instrument. The measurements at the BIPM took place in December 1998 and those at the VNIIM in November 1998 and February 1999, using the reference conditions recommended by the CCRI [1]. 2. Determination of the air-kerma rate For a free-air ionization chamber standard with measuring volume V, the air-kerma rate is determined by the relation I W K& air 1 = k i (1) ρ airv e 1 g air i where ρ air is the density of air under reference conditions, I is the ionization current under the same conditions, W air is the mean energy expended by an electron of charge e to produce an ion pair in air, g air is the mean fraction of the initial electron energy lost by bremsstrahlung production in air, and Π k i is the product of the correction factors to be applied to the standard. The values used for the physical constants ρ air and W air /e are given in Table 1. For use with this dry-air value for ρ air, the ionization current I must be corrected for humidity and for the difference between the density of the air of the measuring volume at the time of measurement and the value given in the table 1. 3. Details of the standards Both free-air chamber standards are of the conventional parallel-plate design. The measuring volume V is defined by the diameter of the defining aperture and the length of the collecting region. The BIPM air-kerma standard is described in [2]. Details of the VNIIM standard, which has not previously been compared with the BIPM standard, are given in [3]. The main dimensions, the measuring volume and the polarizing voltage for each chamber are shown in Table 2. 1 For an air temperature T ~ 293 K, air pressure P and relative humidity ~50 % in the measuring volume, this involves a temperature correction T / T 0, a pressure correction P 0 / P, a humidity correction k h = 0.9980, and the factor 1.0002 to account for the change in the compressibility of dry air between T ~ 293 K and T 0 = 273.15 K. 1

Table 1. Physical constants used in the determination of the air-kerma rate Constant Value u i ρ air 1.293 0 kg m 3 0.000 1 W air / e 33.97 J C 1 0.001 5 u i is the relative standard uncertainty. Density of dry air at T 0 = 273.15 K and P 0 = 101 325 Pa. Table 2. Main characteristics of the standards Standard BIPM VNIIM Aperture diameter / mm 9.939 16.007 Air path length / mm 281.5 448.7 Collecting length / mm 60.004 100.233 Electrode separation / mm 180 300 Collector width / mm 200 300 Measuring volume / mm 3 4 655.4 20 171 Polarizing voltage / V 4 000 4 000 4. The transfer instruments 4.1 Determination of the calibration coefficient for a transfer instrument The air-kerma calibration coefficient N K for a transfer instrument is given by the relation N K K& = I tr (2) where K & is the air-kerma rate determined by the standard using (1) and I tr is the ionization current measured by the transfer instrument and the associated current-measuring system. The current I tr is corrected to the standard conditions of air temperature, pressure and relative humidity chosen for the comparison (T = 293.15 K, P = 101 325 Pa and h = 50 %). To derive a comparison result from the calibration coefficients N K,BIPM and N K,NMI measured, respectively, at the BIPM and at a national measurement institute (NMI), differences in the radiation qualities must be taken into account. Normally, each quality used for the comparison has the same generating potential at each institute, but the half-value layers (HVLs) may differ. A radiation quality correction factor k Q is derived for each comparison quality Q. This corrects the calibration coefficient N K,NMI determined at the NMI into one which applies at the equivalent BIPM quality and is derived by interpolation of the N K,NMI values in terms of log(hvl). The comparison result at each quality is then taken as kq N K,NMI R K,NMI = (3) N K,BIPM In practice, the half-value layers normally differ by only a small amount and k Q is close to unity. 2

4.2 Details of the transfer instrument A thimble-type cavity ionization chamber belonging to the VNIIM was used as the transfer instrument for the comparison. Its main characteristics are given in Table 3. The reference point for the chamber was taken to be on the cylindrical axis 13 mm from the chamber tip. The chamber was oriented so that the line marked on the chamber stem faced the source. Table 3. Main characteristics of the transfer chamber Chamber type PTW M300001 Serial number 0109 Geometry thimble External diameter / mm 7.00 Wall material Graphite-coated PMMA Wall thickness / mm 0.425 Nominal volume / cm 3 0.6 Polarizing voltage / V 300 Typical polarity correction k pol,tr 1.000 4 5. Calibration at the BIPM 5.1 BIPM irradiation facility and reference radiation qualities The BIPM medium-energy x-ray laboratory houses a constant-potential generator and a tungsten-anode x-ray tube with an inherent filtration of 2.3 mm aluminium. Both the generating potential and the tube current are stabilized using feedback systems constructed at the BIPM; this results in a very high stability and obviates the need for a transmission current monitor. The radiation qualities used in the range from 100 kv to 250 kv are those recommended by the CCRI [1] and are given in Table 4. The irradiation area is temperature controlled at around 20 C and is stable over the duration of a calibration to better than 0.1 C. Two thermistors, calibrated to a few mk, measure the temperature of the ambient air and the air inside the BIPM standard (which is controlled around 25 C). Air pressure is measured by means of a calibrated barometer positioned at the height of the beam axis. The relative humidity is controlled within the range 47 % to 53 % and consequently no humidity correction is applied to the current measured using transfer instruments. 3

5.2 BIPM standard and correction factors The defining plane of the aperture of the BIPM standard was positioned at 1 200 mm from the radiation source, with a reproducibility of 0.03 mm. The standard was aligned on the beam axis to an estimated uncertainty of 0.1 mm. The beam diameter in the reference plane is 105 mm for all radiation qualities; an off-axis displacement of 0.1 mm changes the measured current by no more than 3 10 4 in relative value at 100 kv. During the calibration of the transfer chamber, measurements using the BIPM standard were made at both polarities to correct for any polarity effect in the standard. The measured difference was 8 10 4 at 100 kv and around 3 10 4 at the other radiation qualities. The leakage current for the BIPM standard, relative to the ionization current, was measured to be around 2 10 4. The correction factors applied to the ionization current measured at each radiation quality using the BIPM standard, together with their associated uncertainties, are given in Table 5. Table 4. Characteristics of the BIPM reference radiation qualities Generating potential / kv 100 135 180 250 Additional Al filtration / mm 1.203 2 - - - Additional Cu filtration / mm - 0.232 1 0.484 7 1.570 1 Al HVL / mm 4.027 - - - Cu HVL / mm 0.148 0.494 0.990 2.500 µ air / m 1 0.035 5 0.023 5 0.019 8 0.017 2 K & BIPM / mgy s 1 0.21 0.20 0.29 0.38 Air-attenuation coefficient at 293.15 K and 100 kpa, measured at the BIPM for an air path length of 100 mm. Table 5. Correction factors for the BIPM standard Generating potential / kv 100 135 180 250 u ia u ib Air attenuation k a 1.010 0 1.006 6 1.005 6 1.004 9 0.000 3 0.000 1 Scattered radiation k sc 0.9948 0.9962 0.9967 0.9969-0.0007 Electron loss k e 1.0000 1.0023 1.0052 1.0078-0.0010 Ion recombination k s 1.0005 1.0005 1.0005 1.0005 0.0002 0.0001 Field distortion k d 1.0000 1.0000 1.0000 1.0000-0.0007 Aperture edge transmission k l 0.9999 0.9998 0.9997 0.9996-0.0001 Wall transmission k p 1.0000 1.0000 0.9999 0.9988 0.0001 - Humidity k h 0.9980 0.9980 0.9980 0.9980-0.0003 1 g air 0.9999 0.9999 0.9998 0.9997-0.0001 Nominal values for 293.15 K and 100 kpa; each measurement is corrected using the air density measured at the time. 4

The factors k a correct for the attenuation of the x-ray fluence along the air path between the reference plane and the centre of the collecting volume. They are evaluated using the measured air-attenuation coefficients µ air given in Table 4. In practice, the values used for k a take account of the temperature and pressure of the air in the standard at the time of the measurements. Ionization current measurements (both for the standard and for transfer chambers) are also corrected for changes in air attenuation arising from variations in the temperature and pressure of the ambient air between the radiation source and the reference plane. 5.3 Transfer chamber positioning and calibration at the BIPM The reference point for each chamber was positioned in the reference plane, with a reproducibility of 0.03 mm. The transfer chamber was aligned on the beam axis to an estimated uncertainty of 0.1 mm. The leakage current was measured before and after each series of ionization current measurements and a correction made using the mean value. The relative leakage current for the transfer chamber was typically 1 10 4. The relative standard uncertainty of the mean of each of two series of ten measurements at each radiation quality was typically 2 10 4 for the transfer chamber, although the repeatability was poorer, typically 4 10 4. 6. Calibration at the VNIIM 6.1 VNIIM irradiation facility and reference radiation qualities The medium-energy x-ray facility at the VNIIM is an ISOVOLT-400 industrial unit, with a tungsten-anode x-ray tube having an inherent filtration of around 4 mm aluminium. The stability of the generating potential is typically 3 10 3 in relative value. The x-ray output is monitored by means of a transmission ionization chamber whose aluminium mylar windows introduce a filtration of 2.136 mg cm 2. The typical relative standard uncertainty of short-term current measurents using the VNIIM standard, relative to the transmission monitor current, is typically 5 10 4 ; the longer-term reproducibility of the monitor is around 8 10 4. The characteristics of the VNIIM realization of the CCRI comparison qualities [1] are given in Table 6. Table 6. Characteristics of the VNIIM reference radiation qualities Generating potential / kv 100 135 180 250 Additional Al filtration / mm - 1.670 0.761 0.761 Additional Cu filtration / mm - 0.101 0.265 1.291 Al HVL / mm 4.120 - - - Cu HVL / mm - 0.490 0.982 2.496 µ air / m 1 0.031 0 0.022 2 0.019 8 0.017 1 K & VNIIM / mgy s 1 0.44 0.49 0.83 1.13 Air attenuation coefficient at 293.15 K and 100 kpa, measured at the VNIIM for an air path length of 150 mm. Two calibrated platinum resistance thermometers are used to measure the ambient air temperature, one positioned close to the transfer chamber and the second close to the 5

transmission monitor. Air pressure is recorded using a calibrated barometer positioned at the height of the beam axis. The relative humidity in the VNIIM measurement area is controlled within the range 42 % to 50 %. No humidity correction is applied to the transfer chamber current measurements. 6.2 VNIIM standard and correction factors The reference plane for the VNIIM standard was positioned at 1 000 mm from the radiation source, with a reproducibility of 0.05 mm. The standard was aligned on the beam axis to an estimated uncertainty of 0.2 mm. The beam diameter in the reference plane is 95 mm for all radiation qualities; an off-axis displacement of 0.2 mm changes the measured current in relative terms by no more than 5 10 4 at 100 kv. During the calibration of the transfer chambers, measurements using the VNIIM standard were made at both polarities to correct for any polarity effect in the standard. The measured difference was typically 4 10 4 in relative value. The relative leakage current was measured to be less than 2 10 4. The correction factors applied to the ionization current measured at each radiation quality using the VNIIM standard, together with their associated uncertainties, are given in Table 7. Table 7. Correction factors for the VNIIM standard Generating potential / kv 100 135 180 250 u ia u ib Air attenuation k a 1.014 0 1.010 0 1.008 9 1.007 7 0.000 3 0.000 1 Scattered radiation k sc 0.994 9 0.995 6 0.996 3 0.996 4-0.001 0 Electron loss k e 1.000 0 1.000 0 1.001 6 1.003 0-0.001 0 Ion recombination k s 1.000 9 1.001 0 1.003 5 1.004 4 0.000 2 0.000 2 Field distortion k d 1.000 0 1.000 0 1.000 0 1.000 0-0.000 5 Aperture edge transmission k l 0.999 9 0.999 7 0.999 7 0.999 6-0.000 1 Wall transmission k p 1.000 0 1.000 0 1.000 0 1.000 0 0.000 1 - Humidity k h 0.998 0 0.998 0 0.998 0 0.998 0-0.000 3 1 g air 0.999 9 0.999 9 0.999 8 0.999 7-0.000 1 Nominal values for 293.15 K and 100 kpa; each measurement is corrected using the air density measured at the time. The correction factor k a is evaluated using the measured air-attenuation coefficients µ air given in Table 6. In practice, the values used for k a take account of the temperature and pressure of the air in the standard at the time of the measurements. Ionization measurements (standard and transfer chambers) are also corrected for variations in the temperature and pressure of the ambient air between the radiation source and the reference plane. 6.3 Transfer chamber positioning and calibration at the VNIIM The reference point for the transfer chamber was positioned at the reference distance, with a reproducibility of 0.05 mm. Alignment on the beam axis was to an estimated uncertainty of 0.1 mm. 6

The leakage current of the transfer chamber was measured before and after each series of ionization current measurements and a correction made using the mean value. The relative leakage current was typically 2 10 4. The relative standard uncertainty of the mean of three series of thirteen measurements at each radiation quality was typically 3 10 4. 7. Additional corrections to transfer chamber measurements 7.1 Ion recombination, polarity and beam non-uniformity As can be seen from Tables 4 and 6, the air-kerma rates at the VNIIM are two to three times higher than those at the BIPM. Thus volume recombination effects will be greater for the transfer chamber calibrations at the VNIIM; recombination corrections k s,tr were determined experimentally at the VNIIM to be 1.000 0, 1.000 0, 1.000 4 and 1.000 8 for the four comparison qualities 100 kv, 135 kv, 180 kv and 250 kv, respectively. From these, the corresponding correction factors 1.000 0, 1.000 0, 1.000 1 and 1.000 3, respectively, were derived for the BIPM qualities by scaling according to the air-kerma rates given in Tables 4 and 6. Initial recombination effects were assumed to be negligible. The uncertainty arising from this treatment of ion recombination effects is estimated to be 1 10 4 in relative value. The transfer chamber was used with the same polarity at each laboratory and so no corrections k pol,tr are applied for polarity effects in the transfer chamber. From the typical value for the polarity effect in PTW M300001 chamber types given in Table 3, a relative uncertainty of 1 10 4 is estimated. No correction k rn,tr is applied at either laboratory for the radial non-uniformity of the radiation field. For small cylindrical transfer chambers the effect will be very small, less than 1 10 4 in relative value. 7.2 Radiation quality correction factors k Q As noted in Section 4.1, slight differences in radiation qualities may require a correction factor k Q. However, from Tables 4 and 6 it is evident that the radiation qualities at the BIPM and at the VNIIM are very closely matched in terms of HVL and so the correction factor k Q is taken to be unity for all qualities, with a negligible uncertainty. 8. Uncertainties The uncertainties associated with the primary standards are listed in Table 8, those for the transfer chamber calibrations in Table 9 and those for the comparison results in Table 10. The relative combined standard uncertainty u c of the comparison result R K,VNIIM takes into account correlations in the type B uncertainties associated with the physical constants and the humidity correction. Correlations between the BIPM and VNIIM values for k e and k sc are not taken into account, although these are derived from the same basic data. 9. Results and discussion The calibration coefficients determined at the BIPM and at the VNIIM are given in Table 11. For the 100 kv, 135 kv and 180 kv qualities, the pre- and post-comparison calibrations at the VNIIM agree at a level which is consistent with the uncertainties associated with chamber positioning, ionization current measurements and transmission monitor stability. However, the calibrations at 250 kv show a relative change of 2.5 10 3. 7

Table 8. Uncertainties associated with the standards Standard BIPM VNIIM Relative standard uncertainty u ia u ib u ia u ib Ionization current 0.000 3 0.000 2 0.000 5 0.000 2 Volume 0.000 1 0.000 5 0.000 2 0.000 5 Positioning 0.000 1 0.000 1 0.000 2 0.000 2 Correction factors (excl. k h ) 0.000 4 0.001 4 0.000 4 0.001 5 Humidity k h - 0.000 3-0.000 3 Physical constants - 0.001 5-0.001 5 K & std 0.0005 0.0021 0.0007 0.0022 Table 9. Uncertainties associated with the calibration of the transfer chambers Institute BIPM VNIIM Relative standard uncertainty u ia u ib u ia u ib K & std 0.0005 0.0021 0.0007 0.0022 Positioning of transfer chamber 0.000 1 0.000 1 0.000 2 0.000 2 I tr 0.000 4 0.000 2 0.000 3 0.000 2 Stability of transmission monitor - - 0.000 8 - Radial non-uniformity k rn,tr - 0.000 3-0.000 5 N K,std 0.000 7 0.002 1 0.001 1 0.002 3 Table 10. Uncertainties associated with the comparison results Relative standard uncertainty u ia u ib N N 0.0013 0.0022 K, VNIIM K,BIPM k s,tr - 0.0001 k pol,tr - 0.0001 0.0013 0.0022 R K,VNIIM u c = 0.002 6 Takes account of correlations in type B uncertainties. 8

The comparison results R K,VNIIM are also given in Table 11. General agreement between the standards is observed, the mean result for the four qualities being 0.998 2 (standard uncertainty of the distribution 8 10 4 ). The deviation from unity of this mean value is consistent with the stated comparison uncertainty of 2.6 10 3 (Table 10). No significant trend with radiation quality is observed, except perhaps for the result at 180 kv which is notably higher. Table 11. Calibration coefficients for the transfer chamber Generating potential / kv 100 135 180 250 N K,VNIIM (pre-comp) / Gy µc -1 48.19 48.09 48.28 48.32 N K,BIPM / Gy µc -1 48.27 48.19 48.30 48.47 N K,VNIIM (post-comp) / Gy µc -1 48.15 48.10 48.28 48.44 R K,VNIIM 0.997 9 0.998 0 0.999 6 0.998 1 In an attempt to explain the higher comparison result for the 180 kv quality, several possibilities were considered. Values for the correction factors k e and k sc for both standards have recently been calculated by Burns [4] using the Monte Calro Code EGSnrc [5]. Table 12 shows the values obtained and the effect of including these in the present comparison. No improvement is observed; the results are farther from unity and consistency with radiation quality is no better. Table 12. Analysis of comparison results using revised correction factors Generating potential / kv 100 135 180 250 Monte Carlo k e for VNIIM standard 1.000 0 1.000 0 1.000 8 1.002 5 Monte Carlo k e for BIPM standard 1.000 0 1.001 5 1.004 8 1.008 7 Monte Carlo k sc for VNIIM standard 0.992 7 0.993 8 0.994 5 0.995 8 Monte Carlo k sc for BIPM standard 0.995 2 0.995 9 0.996 4 0.997 4 Revised R K,VNIIM if these values adotped 0.995 3 0.997 3 0.997 7 0.995 6 A second quality-dependent effect was considered. The corrections for ion recombination determined at the VNIIM (using the 1 / I vs 1 / V method) are significantly different from those which one would calculate for the VNIIM standard following the work of Boutillon [6]. A reevaluation using [6] gives values for k s of 1.002 0, 1.002 2, 1.003 0, 1.003 8, respectively, the use of which leads to the revised comparison results 0.999 0, 0.999 2, 0.999 1 and 0.997 5, respectively. Bearing in mind the large change in the pre- and post-comparison calibrations for the 250 kv quality, this set of results may be considered to be more satisfactory. However, when combined with the new Monte Carlo values for correction factors, this improvement is lost and no strong conclusions can be drawn. It should be emphasized that Monte Carlo correction factors have not as yet been adopted by either institute and so the comparison results stand as those given in bold in Table 11. A summary of the results of BIPM comparisons of air-kerma standards for medium-energy x- rays, including the present comparison, is presented in Figure 1. 9

1.015 1.010 100 kv 135 kv 180 kv 250 kv 1.005 R K,NMI 1.000 0.995 0.990 0.985 ARPANSA BEV BNM-LCIE CSIR ENEA GUM NIST NMi NPL NRC OMH VNIIM Figure 1. Results of BIPM medium-energy x-ray comparisons, expressed as the ratio R K,NMI of the air-kerma rate detemined by the standard of the national metrology institute (NMI) to that determined by the BIPM standard. For NMIs that have compared more than once with the BIPM, only the results of the most recent comparison are included. 10

References [1] BIPM, Qualités de rayonnement, CCEMRI(I), 1972, R15. [2] BOUTILLON M., Mesure de l exposition au BIPM dans le domaine des rayons X de 100 à 250 kv, 1978, Rapport BIPM-78/3. [3] KOCHINA M.P., NOVOSELOVA T.N., Standard free-air plane-parallel ionization chamber type of IK 70-300, Investigation in the field of ionization radiation, Sc.Publ. VNIIM, 1985. [4] BURNS D.T., Free-air chamber correction factors for electron loss, photon scatter, fluorescence and bremsstrahlung, CCRI(I)/01-36 rev, 2001. [5] KAWRAKOW, I. and ROGERS D.W.O., The EGSnrc code system: Monte Carlo simulation of electron and photon transport, NRCC Report PIRS-701, 2000 (National Research Council of Canada). [6] BOUTILLON M., Volume recombination parameter in ionization chambers, Physics in Medicine and Biology, 1998, 43, 2061-2072. 11