Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Similar documents
Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Study of Residual Solvents in Various Matrices by Static Headspace

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

USP<467> residual solvents

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Headspace Technology for GC and GC/MS: Features, benefits & applications

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

Fast methods for the determination of ibuprofen in drug products

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

C146-E209. Headspace Samplers. HS-20 Series

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals

Methanol Extraction of high level soil samples by USEPA Method 8260C

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Application. Gas Chromatography February Introduction

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Analysis of BTEX in Natural Water with SPME

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Automated and accurate component detection using reference mass spectra

GC METHOD FOR QUANTITATIVE DETERMINATION OF RESIDUAL 2-(2-CHLOROETHOXY)ETHANOL (CEE) AND N-METHYL-2-PYRROLIDINONE (NMP) IN QUETIAPINE

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

STANDARD OPERATING PROCEDURES

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

Maximizing Sample Throughput In Purge And Trap Analysis

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

DETERMINATION OF ORGANIC VOLATILE IMPURITIES IN NEPAFENAC BY GC METHOD

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Sulfotepp impurities in Chlorpyrifos EC formulations

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Determination of trace anions in concentrated hydrofluoric acid

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Introducing 7820A GC System. Reliability, Simplicity & Affordability!

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Ethylene Oxide and 1,4-Dioxane in Fatty Acids

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

GAS CHROMATOGRAPHY (GC)

Optimizing. Abstract: is standardd. procedures. altered to

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Transcription:

APPLICATION NOTE 0676 Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <67> method Authors Giulia Riccardino, Paolo Magni, Stefano Pelagatti, Manuela Bergna, Davide Bressanello, and Cristian Cojocariu Thermo Fisher Scientific, Runcorn, UK; Thermo Fisher Scientific, Milan, IT Keywords USP <67>, residual solvents, pharmaceuticals, valve and loop, static headspace, gas chromatography, GC, headspacegas chromatography, HS-GC, flame ionization detector, FID, TriPlus 00 Goal The aim of this work was to evaluate the performance of the new Thermo Scientific TriPlus 00 Gas Chromatography Headspace (HS) Autosampler for the determination of residual solvent content in water-soluble and waterinsoluble pharmaceuticals according to the United States Pharmacopeia <67> method (USP). Introduction Organic solvents are widely used in the synthesis of pharmaceutical products and cannot always be completely removed during the manufacturing processes. To ensure safety, final products are tested to assess whether the solvents used during the manufacturing processes have been efficiently removed or, if still present, their concentration is within the accepted limits.

According to the International Conference on Harmonization (ICH) guidelines, the USP <67> method describes the assay procedure and classifies residual solvents based upon their toxicities, setting the concentration limits according to their health hazard: Class : solvents with unacceptable toxicities Class : solvents with less severe toxicity Class : solvents with low toxicity These classes do not provide a full list of all the solvents that can be used in the manufacturing processes; therefore, the final products should be screened according to the solvents used during the production/ purification processes. As organic solvents have relatively low boiling points and are thermally stable, the analytical method of choice for Class and Class residual solvent determination is static headspace sampling coupled to gas chromatography (HS-GC), with either flame ionization detection (FID) or mass spectrometry (MS) as detectors of choice. Class solvents can also be determined with a nonspecific method such as weight loss on drying. Headspace sampling allows for the extraction of semivolatile and volatile compounds from complex liquid and solid matrices in a fast and simple way without the need for time-consuming sample preparation. The new Triplus 00 HS autosampler offers an innovative design of the pneumatic circuit with a direct connection between the heated valve and the GC column. This translates into highly precise sample introduction and excellent peak area repeatability. Additionally, continuous purging of the sample path ensures system robustness and reliability reducing the risk of contamination and carryover (important when high boiling residual solvents are analyzed). In this study, the results from the analysis of residual solvent according to the USP <67> criteria obtained with the TriPlus 00 HS autosampler and FID as detector of choice are reported. USP <67> system compliance, sensitivity, precision, robustness, and linearity were assessed according to the workflow described in the USP <67> method for both watersoluble and water insoluble pharmaceutical products. Experimental In all the experiments, a TriPlus 00 HS autosampler was coupled to a Thermo Scientific TRACE 0 GC equipped with a Thermo Scientific Instant Connect Split/ Splitless SSL Injector and a Thermo Scientific Instant Connect FID. Chromatographic separation of target chemicals was obtained on a Thermo Scientific TraceGOLD TG-6 GC column, 0 m 0. mm.8 µm (P/N 608-0) for procedure A and C, and on a Thermo Scientific TraceGOLD TG-WAXMS GC column, 0 m 0. mm 0. µm (P/N 6088-0) for procedure B. HS-GC-FID operative conditions, according to the USP <67> method, are given in Tables A and B. Table A. GC-FID analytical parameters for the TRACE 0 GC used for residual solvent content determination according to the USP <67> method, procedures: A, B, C TRACE 0 GC Parameters Procedure A/C Procedure B Inlet Module and Mode SSL, split Split Ratio 0: 0: Septum Purge Mode, Flow (ml/min) Constant, Carrier Gas, Carrier Mode, Flow (ml/min) He, constant flow,. Oven Temperature Program Procedure A/C Procedure B Temperature ( C) 0 0 Hold Time (min) 0 0 Temperature ( C) 0 6 Rate ( C/min) 0 6 Hold Time (min) 0 0 FID Temperature ( C) 0 Air Flow (ml/min) 0 H Flow (ml/min) N Flow (ml/min) 0 Acquisition Rate (Hz)

Table B. TriPlus 00 HS autosampler analytical parameters used for residual solvent content determination according to the USP <67> method, procedures: A, B, C TriPlus 00 HS Autosampler Parameters Incubation Temperature ( C) 80 Incubation Time (min) 60 Vial Shaking Fast Vial Pressurization Mode Pressure Vial Pressure (kpa) (Auxiliary Gas Nitrogen) 0 Vial Pressure Equilibration Time (min) Loop Size (ml) Loop/Sample Path Temperature ( C) 80 Loop Filling Pressure (kpa) 7. Loop Equilibration Time (min) Needle Purge Flow Level Injection Mode Standard Injection Time (min) Data acquisition, processing, and reporting Data was acquired, processed, and reported using the Thermo Scientific Chromeleon Chromatography Data System (CDS) software, version 7., a software platform compliant with Title of the Code of Federal Regulations, Part. Integrated instrument control ensures full automation from instrument set-up to raw data processing, reporting, and storage. Simplified e-workflows deliver effective data management ensuring ease of use, data integrity, and traceability. Chromeleon CDS also offers the option to scale up the entire analytical process in the laboratory from a single workstation to an enterprise environment. Sample preparation USP <67> Class, Class A, and Class B residual solvents solutions in dimethylsulfoxide (DMSO) were sourced from Restek (P/N 67, 60, 680, respectively). Stock and standard solutions for procedures A, B, C were diluted in water or DMSO as reported in the USP <67> method. HPLC-MS grade water and GC headspace grade dimethylsulfoxide (DMSO, purity.%) were used as diluents. Overthe-counter purchased aspirin (acetylsalicylic acid, 7 mg) and pain relief tablets (paracetamol, 00 mg, and caffeine, 6 mg) were used to prepare sample stock and test solutions as described in the regulation. To test the complete USP <67> workflow, a second stock of test solutions was prepared at a concentration level five times higher than the limits reported in Table, which represent the acceptable amount of residual solvents in the final product. Table. Concentration limits in ppm for Class, Class A, and Class B residual solvents Compound Name Concentration Limit (ppm) Class,-Dichloroethene 8,,-Trichloroethane 00 Benzene Carbon Tetrachloride,-Dichloroethane Class A Methanol 000 Acetonitrile 0 Dichloromethane 600 trans-,-dichloroethene 870 cis-,-dichloroethene 870 Tetrahydrofuran 70 Cyclohexane 880 Methycyclohexane 80,-Dioxane 80 Toluene 80 Chlorobenzene 60 Xylene* 70 Class B Hexane 0 Nitromethane 0 Chloroform 60,-Dimethoxyethane 00 Trichloroethylene 80 Pyridine 00 -Hexanone 0 Tetralin 00 * Usually 60% m-xylene, % p-xylene, % o-xylene, 7% ethylbenzene.

Results and discussion Procedure A: residual solvent screening and identification By using the Chromeleon CDS e-workflow a significant reduction in the number of steps needed to set up the analytical sequence was obtained, speeding up the analysis time and ultimately boosting laboratory productivity. Class, Class System Suitability, Class A standard solutions, and test solutions for water-soluble and water-insoluble pharmaceuticals were prepared using 0 ml crimped capped vials (P/N 0-CV). The following USP <67> method performance criteria were met for water-soluble and water-insoluble products: The peak-to-peak (PtP) signal-to-noise ratio (S/N) for,,-trichloroethane in Class standard solution was >: and all peaks in Class system suitability showed S/N >: (Figure ). PtP S/N calculations were made automatically in Chromeleon CDS according to the method described in the USP <6> system suitability section. Chromatographic resolution (Rs) between the critical pair, acetonitrile and dichloromethane, was automatically calculated using the chromatography data system applying the formula for electronic integrator reported in the USP <6> system suitability section, and confirmed to be >, meeting the acceptance criteria as required by the regulation (Figure ). a Peak ID:.,-Dichloroethene S/N 6.,,-Trichloroethane S/N 8. Carbon Tetrachloride S/N 8. Benzene = S/N 8.,-Dichloroethane S/N b Peak ID:.,-Dichloroethene S/N 76.,,-Trichloroethane S/N 0. Carbon Tetrachloride S/N. Benzene = S/N 7.,-Dichloroethane S/N Figure. Peak-to-peak signal-to-noise (S/N) ratios for Class system suitability solutions for water-soluble (a) and water-insoluble (b) products. Peaks with no annotation could not be identified from the FID data. Figure. Chromatographic resolution (Rs) between acetonitrile and dichloromethane for watersoluble (a) and water-insoluble products (b). Resolution met the regulation requirements (Rs.0) with calculated values of.8 and.7 for water-soluble and water-insoluble pharmaceuticals, respectively.

Testing a pharmaceutical product that passes procedure A requirements The pharmaceutical products (dispersive aspirin and paracetamol pain relief tablets) were analyzed unspiked for their residual solvent content. The results were compared with the standard solutions and confirmed that all solvents used during manufacturing process were efficiently removed as no residual solvents were detected (Figure ). As an example, the peak profile obtained for dispersive aspirin unspiked solution (green) compared to Class A standard solution (blue) is reported in Figure. Testing a pharmaceutical product failing procedure A The pharmaceutical products (dispersive aspirin and paracetamol pain relief tablets) were spiked with residual solvents and injected into the chromatographic system. The results were compared to the standard solutions. As peaks found in the spiked samples exceeded the limits reported in Table, a compound confirmation step was mandatory as described in the procedure B. As an example, the peak profile obtained for dispersive aspirin spiked solution (green) compared to Class A standard solution (blue) is reported in Figure. Class A residual solvent peaks detected in the spiked sample solution showed higher peak areas compared to the ones in the corresponding Class A standard solution. Peaks:. Methanol. Acetonitrile. Dichloromethane. trans-,-dichloroethene. cis-,-dichloroethene 6. Tetrahydrofuran 7. Cyclohexane 8. Methylcyclohexane.,-Dioxane 0. Toluene. Chlorobenzene. Ethylbenzene. m-xylene. p-xylene. o-xylene 6. Cumene 7 0, 8 6 6 Figure. Comparison between Class A standard solution (blue) and acetylsalicylic solution (green). Criteria are met as no residual solvent peaks could be detected in the test sample.

Peaks:. Methanol.,-Dichloroethene. Acetonitrile. Dichloromethane. trans-,-dichloroethene 6. Hexane 7. Nitromethane 8. cis-,-dichloerethene. Tetrahydrofuran 0. Chloroform. Cyclohexane/,,-Trichloroethane. Benzene.,-Dichloroethane/,-Dimetoxyethane. Trichloroethene. Methylcyclohexane 6.,-Dioxane 7. Toluene/Pyridine 8. -Hexanone. Chlorobenzene 0. Ethylbenzene. m-xylene/p-xylene. o-xylene. Cumene. Tetralin 7 6 7 0 6 8 0 7 6 8 Figure. Comparison between peak profiles obtained for water-soluble spiked test solution (green) and Class A standard solution (blue). Class A peaks in the spiked solution showed higher responses compared to Class A standard solution. Unmatched green peaks belonged to either Class or Class B residual solvents. Procedure B: peak identity confirmation According to the USP <67> for procedure B, the chromatographic TraceGOLD TG-6 GC column was replaced with a WAX column (TraceGOLD TG- WAXMS GC column, 0 m 0. mm 0. µm, P/N 6088-0). Class, Class System Suitability, Class A standard solutions and test solutions for watersoluble and insoluble pharmaceuticals were analyzed using the parameters reported in Table. System sensitivity and resolution requirements have been assessed for procedure B: PtP S/N for benzene in Class standard solution is >: and all peaks in Class system suitability showed S/N >: satisfying the regulation requirements (Figure ). Calculations for S/N were made automatically by the data system according to the method described in the USP <6> system suitability section. 6

a Peak ID:.,-Dichloroethene S/N.,,-Trichloroethane S/N 76. Benzene S/N.,-Dichloroethane S/N 0 b Peak ID:.,-Dichloroethene S/N 7.,,-Trichloroethane S/N. Benzene S/N.,-Dichloroethane S/N Figure. Peak-to-peak ratios for Class system suitability solutions for water soluble (a) and water-insoluble (b) products. Peaks with no annotation could not be confirmed from the FID data. The critical pair cis-,-dichloroethene and acetonitrile is baseline resolved with a chromatographic resolution of.8 and. for water-soluble and waterinsoluble Class A standard solutions, respectively (Figure 6), meeting the acceptance criteria required (Rs.0). Chromatographic resolution has been automatically determined in Chromeleon CDS by applying the formula for electronic integrator reported in the USP <6> system suitability section. Rs=.8 Rs=. Figure 6. Chromatographic resolution for critical pair cis-,-dichloroethene/acetonitrile for watersoluble (top) and water-insoluble products (bottom). Peaks are baseline resolved meeting the USP <67> chromatographic resolution requirements (Rs.0). 7

Testing a pharmaceutical product matching Procedure B confirmation Class, Class A, Class B standard solutions and spiked test solutions for water-soluble and waterinsoluble pharmaceuticals were injected into the chromatographic system and the peak profiles were compared. The peaks identified (procedure A) were confirmed (procedure B) as their responses were higher than the corresponding standards. Therefore, the levels of these residual solvents must be determined (procedure C). The chromatographic profile for water-soluble spiked solution and Class A standard solution is reported as an example in Figure 7. Procedure C: quantification Class, Class System Suitability, and Class A standard solutions were injected into the chromatographic system. HS-GC parameters applied for procedure C are reported in Table. Signal-to-noise (S/N) and chromatographic resolution (Rs) requirements for Class, Class System suitability solution, and Class A standard solution were the same as described and assessed in procedure A. Peaks:. Hexane. Cyclohexane/,-Dichloroethene. Methylcyclohexane. trans-,-dichloroethene.,,-tricholoethane 6. Methanol 7.,-Dimethoxyethane/Dichloromethane/Tetrahydrofuran 8. Benzene. cis-,-dichloroethene/trichloroethene 0. Acetonitrile. Chloroform. Toluene.,-Dioxane.,-Dichloroethane. -Hexanone 6. Ethylbenzene 7. p-xylene/nitromethane 8. m-xylene. Pyridine 0. o-xylene. Chlorobenzene. Tetralin 7 8 0 7 8 7 6 0 Figure 7. Comparison between peak profiles obtained for water-soluble spiked sample solution (green) and Class A standard solution (blue). Class A peaks in the spiked solution showed higher responses compared to Class A standard solution. Unmatched green peaks belonged to either Class or Class B residual solvents. 8

Quantification of the residual solvents in a pharmaceutical product Class, Class A, Class B standard and test solutions for quantification have been diluted as described by the USP <67> and injected into the chromatographic system. As an example, the peak profile for spiked aspirin compared to spiked standard test solution is reported in Figure 8. The calculated amount of each residual solvent (in ppm) identified with procedure A and confirmed in procedure B was derived by applying the formula reported in the USP <67> regulation for water-soluble and water-insoluble pharmaceuticals. Calculated concentrations were consistent with the levels used to fortify the samples. System repeatability System repeatability was assessed on n=8 consecutive injections for Class, Class A, and Class B standard solutions. The standard solutions were diluted in water or DMSO according to procedure A for water-soluble and water-insoluble products respectively. Sample preparation played a critical role for tested apolar solvents with high partition coefficients. As effect of the low affinity for water, %RSDs were higher when concentrated standard solutions were diluted in water with respect to DMSO. Peaks:. Methanol.,-Dichloroethene. Acetonitrile. Dichloromethane. trans-,-dichloroethene 6. Hexane 7. Nitromethane 8. cis-,-dichloerethene. Tetrahydrofuran 0. Chloroform. Cyclohexane/,,-Trichloroethane. Benzene.,-Dichloroethane/,-Dimetoxyethane. Trichloroethene. Methylcyclohexane 6.,-Dioxane 7. Toluene/Pyridine 8. -Hexanone. Chlorobenzene 0. Ethylbenzene. m-xylene/p-xylene. o-xylene. Cumene. Tetralin 7 0 6 8 0 7 6 8 Figure 8. Comparison between peak profiles obtained for spiked aspirin solution (green chromatogram) and standard test solution (blue chromatogram). Class, Class A, and Class B peaks in the spiked sample solution showed higher responses compared to standard test solution.

Peak area %RSDs obtained for Class, Class A, and Class B residual solvents are reported in Table with average values <% for all residual solvent classes when water and DMSO were used as diluent. Table. Peak area %RSDs obtained from n=8 consecutive injections using water and DMSO as diluents for the concentrated standard solutions. Average %RSDs for the assessments are <% for all residual solvent classes. Compound Name %RSD (n=8) Class Diluent: water Diluent: DMSO,-Dichloroethene. 0.7,,-Trichloroethane.0 0.8 Carbon Tetrachloride.. Benzene 0.8 0.,-Dichloroethane.6.0 Average %RSD.0. Class A Diluent: water Diluent: DMSO Methanol 0.7. Acetonitrile 0.8.6 Dichloromethane. 0.7 trans,-dichloroethene.0. cis,-dichloroethene. 0.8 Tetrahydrofuran 0.. Cyclohexane.6.8 Methycyclohexane.0.,-Dioxane.. Toluene.6 0.8 Chlorobenzene. 0.7 Ehylbenzene. 0. m-xylene/p-xylene. 0. o-xylene. 0.8 Average %RSD.7. Class B Diluent: water Diluent: DMSO Hexane. 0.8 Nitromethane.. Chloroform 0..0,-Dimethoxyethane. 0. Trichloroethylene. 0.7 Pyridine 0.8. -Hexanone 0.6 0. Tetralin 0. 0.6 Average %RSD. 0. 0

System linearity System linearity was assessed by serially diluting the stock solutions for Class, Class A, and Class B residual solvents as described in the USP <67> method (procedure C for water-insoluble pharmaceutical products). This way, four calibration levels were obtained: at.%, %, 0%, and 00% of the concentration limit. Prior to analysis, ml of each calibration solution was added to ml water corresponding to 0 mg real sample. Each calibration level was prepared and analyzed in triplicate. Residual solvents showed good linear responses with an average coefficient of determination R =0.8 as reported in Table. Moreover, the relative standard deviation (%RSD) of the residuals across each calibration level was <8% indicating good linearity. Table. Correlation coefficients (R ) and relative standard deviation of residuals (%RSD) obtained over four calibration levels at.,, 0, and 00%. Data analyzed in triplicate. Compound Name Concentration Range (μg/g) Class Correlation Coefficient (R ) Residuals Standard Deviation (% RSD),-Dichloroethene.0 8.0.00.0,,-Trichloroethane 87. 00 0.. Carbon Tetrachloride 0..0 0.7 6. Benzene 0..0 0..,-Dichloroethane 0.6.0 0.. Class A Methanol 7 000.00. Acetonitrile. 0.00.7 Dichloromethane 7 600 0.8. trans,-dichloroethene.8 0.. cis,-dichloroethene.8 0.8.0 Tetrahydrofuran 0 70.00. Cyclohexane. 880 0..0 Methycyclohexane 7. 80.00.,-Dioxane 7. 80.00. Toluene. 80 0.7.6 Chlorobenzene 60 0. 6. Ehylbenzene 6. 6 0.7. m-xylene 6.8 0 0.6 6.0 p-xylene 6.8 0 0.6 6.0 o-xylene. 0.7.6 Class B Hexane 6. 0 0.8.8 Nitromethane 6. 0 0.8.8 Chloroform 7. 60 0.7.6 Thrichloroethene 0 80 0.. -Hexanone 6. 0 0. 7.8 Tetralin.-00 0..0

Examples of calibration curves for benzene (Class ), methylcyclohexane (Class A), and trichloroethene (Classs B) are shown in Figure. Conclusions The results presented in this work demonstrate that the new TriPlus 00 HS autosampler in combination with the Trace 0 GC and FID detector delivers the outstanding performance for the analysis of residual solvents in pharmaceutical products meeting or exceeding all USP <67>method requirements. The innovative design of the pneumatic control and the flow path inertness ensure outstanding repeatability and precision in routine analysis. This was demonstrated by excellent peak area response obtained (average peak area %RSDs for n=8 consecutive injections was <%). Sensitive compound detection can be easily achieved with the Instant Connect FID. Moreover, the TraceGOLD TG-6 column allowed to easily meet and exceed USP <67> method resolutionrequirement (Rs.0), delivering expected chromatographic separation. Good linearity (as demonstrated by R and %RSD residual values) was obtained over the calibration range ensuring that the system can be used for routine quantitative assessment of residual solvents in pharmaceutical products. Figure. Examples of calibration curves for benzene, methylcyclohexane, and trichloroethene. Linearity is shown from 0.0 to 0. µg/ml (corresponding to 0..0 µg/g* in 0 mg pharmaceutical product) for benzene, at 7. to µg/ml corresponding to 7 80 µg/g*) for methylcyclohexane, at 0. to.0 µg/ml (corresponding to 0 80 µg/g*) for tricholoethene. For each calibration level n= replicates. Chromeleon CDS software (compliant with the Title CFR Part requirements) ensures sample integrity, traceability, and effective data management from instrument control to the final report. Overall these results demonstrate that the TriPlus 00 HS autosampler provides unparalleled levels of performance making it a reliable and robust analytical solution for routine laboratories.

References. General Chapter USP <67> Organic Volatile impurities, Chemical Tests, United States Pharmacopeia, 0.. Impurities: Guideline for Residual Solvents QC(R6), ICH Harmonised Guidelines, International Council for Harmonization of Technical Requirements for Pharmaceuticals for Human use, 06.. Thermo Fisher Scientific, Chromeleon CDS Enterprise - Compliance, Connectivity, Confidence, BR767-EN 078S.. General Chapter USP <6> Chromatography, Physical Tests, United States Pharmacopeia, First Supplement to USP 0-NF, 07. Learn More Today For more information about the TriPlus 00 Headspace Autosampler, to request a demo, or to discuss your GC/GC-MS needs, please contact your Thermo Scientific representative today. 08 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Restek is a registered trademark of Restek Corp. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN0676-EN 8S