PC Control / Touch Control

Similar documents
Mobile coulometer for water determination anywhere you want

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable

TitriSoft 2.5. Content

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Operating Instructions. METTLER TOLEDO DL37 KF Coulometer

Karl-Fischer Titration the method for determining water

Compact Titration Compact Titrator G20. One Click Titration Simple & Dependable

Easy and reliable KF titrations From 1 ppm to 100%

From 1 ppm to 100% Karl Fischer water content determination

Software BioScout-Calibrator June 2013

Karl Fischer Titration Of Food : Towards Complete Automation

ON SITE SYSTEMS Chemical Safety Assistant

Automatic determination of the bromine number and the bromine index in petroleum products

St art. rp m. Km /h 1: : : : : : : : : : : : :5 2.5.

Karl Fischer Volumetric Titration Theory and Practice

ISSP User Guide CY3207ISSP. Revision C

User's Manual altimeter V1.1

[Agilent Technologies] [Electrochemistry Products] Quick Reference

Automatic titrators, TitroMatic

Titration of a strong acid with a strong base with Cobra4

Lightcloud Application

NH 3 -selective electrodes

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

ph Probe ReallyEasyData com

Non-Acoustical Inputs

User's Guide. DISTO online. Leica Geosystems

O P E R A T I N G M A N U A L

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides.

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Using Conductivity to Find an Equivalence Point

Application Determination of water in petrochemistry products using Titrator TitroLine 7500 KF trace

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

Circular Motion and Centripetal Force

J.T.Baker Brand HYDRA-POINT KARL FISCHER REAGENTS. Europe Middle East Africa

Sample Alignment (2D detector) Part

Chemistry MultiMeasure Sensor PS-2170

TitraLab 580 and 585. KF Titration Workstations. Reference Manual D21T083

Experimental Procedure

MULTI_T. Digital Titrator for Multi-tasking Titrations

Experiment C-10 Titration of a Strong Acid and a Strong Base

Work and Energy. This sum can be determined graphically as the area under the plot of force vs. distance. 1

[RNH]SO3CH3 [RNH]SO 4 CH [RNH]I

pka AND MOLAR MASS OF A WEAK ACID

Acid-Base Titration. Evaluation copy

Ka of Unknown Acid In this experiment you will determine the Ka of an unknown acid by titration with the sodium hydroxide.

Jasco V-670 absorption spectrometer

Acid-Base Titration. Sample

Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry.

Table of content. Understanding workflow automation - Making the right choice Creating a workflow...05

MULTIFUNCTION COMPUTER METER CX-742 USERS MANUAL

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

15. Studio ScaleChem Getting Started

Appendix A Prototypes Models

mylab: Chemical Safety Module Last Updated: January 19, 2018

TA Instruments ITCRun TM Software Getting Started Guide

Example Application 12. Embankment Loading on a Cam-Clay Foundation

The Quantizing functions

ION570. Reference Manual. Ion Analyser D21M073

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Product manual. Mylos KNX Shutter actuator 2CSY1104C/S

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Graphical User Interfaces for Emittance and Correlation Plot. Henrik Loos

XR Analog Clock - Manual Setting Model Troubleshooting Guide

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Application Bulletin

Virtual Cell Membrane Potential Tutorial IV

Sample Alignment Part

0. Table of contents. Author: Jaap Snijder

ABB i-bus KNX Analogue Input AE/A 2.1

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.

PAA-ANALYZERS PRINCIPLE OF OPERATION

756 KF Coulometer. Water determination right down to trace levels with full automation capabilities. Karl Fischer titrators

Geavanceerde bepaling van water in poeders en vaste stoffen via geautomatiseerde Karl Fischer titratie. Peter van de Spreng 8 November 2011

Lab 5 Enthalpy of Solution Formation

really easy with Metrohm Metrohm AG is the only company whose range of products covers all important methods of ion analysis:

Determination of Total Base Number (TBN)

KEM SET. for FOOD. Salinity. Brix Density Concentration. Acid Value POV. Water Content

Automatic Watershed Delineation using ArcSWAT/Arc GIS

Conductometric Titration & Gravimetric Determination of a Precipitate

Acid/Base Interactions

Validation report and uncertainty budget of Coulometric Karl Fischer titrator with an oven system

PiCUS Calliper Version 3

Virtual Cell Version 4.0 Membrane Potential

KEM SET. for Beverage. Gas volume Internal gas pressure Air content Oxygen concentration Inside pressure Density Brix

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

User Manuel. EurotaxForecast. Version Latest changes ( )

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

Minimum (-31) every 10 s -35 (-31)

Information Dependent Acquisition (IDA) 1

How to Create a Substance Answer Set

VELA. Getting started with the VELA Versatile Laboratory Aid. Paul Vernon

What is a WQSensor? Software Installation. Uninstalling WQSensors Software. NexSens Technology, Inc. TABLE OF CONTENTS

Determining the K sp of Calcium Hydroxide

Investigation reply tool operation manual

MOHID Land Basics Walkthrough Walkthrough for MOHID Land Basic Samples using MOHID Studio

Operating Instructions for Digital Manometer. Model: MAN-SD

University of Colorado Denver Anschutz Medical Campus Online Chemical Inventory System User s Manual

Citric Acid Analysis L-6-1. MOISTURE (Karl Fischer)

Spectrophotometric Determination of pka of Phenol Red

Titration Workstations

Transcription:

PC Control / Touch Control Version 6.0 / 5.840.0150 New Features Manual 8.840.8007EN

Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com PC Control / Touch Control Version 6.0 / 5.840.0150 New Features Manual 8.840.8007EN 07.2009 jb

Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.

Table of contents Table of contents 1 General 1 1.1 Reagents... 1 1.1.1 Editing reagent data... 2 1.1.2 Reagent monitoring... 2 1.2 Favorites... 5 1.2.1 Creating favorites... 6 2 Parameters 9 2.1 Coulometric Karl Fischer titrations (KFC)... 9 2.1.1 Control parameters... 9 2.1.2 Titration parameters... 11 2.1.3 Stop conditions... 12 2.1.4 Conditioning... 12 2.1.5 Cell... 15 2.1.6 Control device... 15 2.1.7 Sensor... 16 2.1.8 Stirrer... 17 2.2 Bromine index determination (BRC)... 18 2.2.1 Control parameters... 18 2.2.2 Titration parameters... 20 2.2.3 Stop conditions... 21 2.2.4 Conditioning... 21 2.2.5 Cell... 23 2.2.6 Control device... 24 2.2.7 Sensor... 24 2.2.8 Stirrer... 26 3 Appendix 27 3.1 Stirring rate... 27 Index 28 III

1 General 1 General 1.1 Reagents Main dialog: System Reagents This chapter describes how you can create a list of the reagents used in the system. Depending on the usage a distinction is made between two types of reagents: Reagent for volumetric determinations Reagents for coulometric determinations [New] [Delete] [Edit] The list of reagents gives the designation and type of each reagent configured. Add a new reagent to the list, see following chapter. Delete the selected reagent from the list. Edit the data of the selected reagent, see following chapter. 1

1.1 Reagents 1.1.1 Editing reagent data Reagent The designation of the reagent is used for unambiguous identification. Comment Input Input 24 characters maximum 24 characters maximum [Reagent monitor.] Set the parameters for the reagent monitoring, see following chapter. 1.1.2 Reagent monitoring In the dialog Edit reagent / Reagent monitoring the conditions for the monitoring of the reagent are defined. If one of the following values is reached, then the reagent has to be exchanged. The values are checked in the following cases: 2

1 General at the start of the determination. at the end of the determination. Number of determ. Working life Volume Reagent capacity Drift [Status] The number of determinations to be carried out with a certain amount of reagent depends on the type of sample and its amount. 1... 999 Working life of the reagent. 1... 999 days This parameter is only visible with volumetric reagents. Volume of the titrant dosed. 1.0... 999.9 ml This parameter is only visible with coulometric reagents. Water capacity of the reagent. 1... 9999 mg This parameter is only visible with coulometric reagents. During conditioning the measured drift has to be in the following range for 2 min: 'specified drift value + 50 µg/min'. 1... 999 µg/min Display the status overview of the current values of the reagent monitoring. 3

1.1 Reagents [Reagent exchange] [Reset] Edit the parameters for the reagent exchange. Dialog "Reagent monitoring / Status" In this dialog the current values of the reagent monitoring are displayed. Reset the values to zero. Dialog "Reagent monitoring / Reagent exchange" In this dialog the parameters for the reagent exchange are defined. Reagent exchange Memory Method The reagent can either be exchanged manually or automatically. manual auto manual manual If a monitored parameter has reached the limit set, a message is being displayed. Then the reagent has to be exchanged manually. auto If a monitored parameter has reached the limit set, the method defined below is started automatically. This parameter can only be edited when Reagent exchange = auto. Memory location the method is loaded from. All memory locations are selectable, even if they are currently not accessible. Internal memory Card 1 Card 2 Shared memory Internal memory Shared memory Memory released on an external system. For this an 847 USB Lab Link is necessary. This parameter can only be edited when Reagent exchange = auto. Method used for emptying the titration cell. Note Make sure that the memory is accessible. 4

1 General Input 32 characters maximum of stored methods 1.2 Favorites Main dialog: Control Favorites Main dialog: System System settings User admin. Edit Favorites This chapter describes how you can create so-called favorites. These favorites are comparable to the favorites/bookmarks of your Internet browser. On the main dialog, a button is created for each favorite. With pressing a single key or with a single mouse click you can trigger an action without having to click yourself through different dialogs. The following objects can be saved as a favorite: Methods Sample data silos Depending on whether you are working with or without login function, user-specific or common favorites can be created. Operation with login function: If you are working with the login function activated, user-specific favorites can be created. These can only be used by a certain user. User-specific favorites are created as follows: in the user administration of a user with administrator rights (System System settings User admin. Edit). in the dialog Control for the user logged in. Operation without login function: If you are working without login function, common favorites can be created. These favorites are available for all users. Common favorites are created in the dialog Control. 5

1.2 Favorites 1.2.1 Creating favorites [New] [Delete] [Edit] The list of favorites gives the designation and position of the button on the main dialog of each favorite configured. Add a new favorite to the list, see following chapter. Delete the selected favorite from the list. Edit the data of the selected favorite, see following chapter. 1.2.1.1 Editing favorites Position On the main dialog, a button is created for each favorite. These buttons are ordered in three columns at fixed positions. The position 1 can be found top left. 6

1 General 1... 14 1 Name Type Memory Method / Silo [More options] The designation of the favorite is used for unambiguous identification. Input 24 characters maximum Definition, whether the favorite is representing a single method or a complete sample data silo. Method Sample data silo Method Memory location the method or the sample data silo is loaded from. Only the currently accessible memory locations are selectable. Internal memory Card 1 Card 2 Shared memory Internal memory Shared memory Memory released on an external system. For this an 847 USB Lab Link is necessary. Method or sample data silo of the favorite. of stored methods or sample data silos Parameterize the autostart function, see following chapter. More options With [More options] the autostart function is parameterized. Autostart on (: ) If this parameter is activated, a new determination is started automatically at the end of a determination. This continues until the quantity specified has been reached (see Number of autostarts). Number of autostarts This parameter can only be edited when Autostart = on. Number of automatic starts. 7

1.2 Favorites 1... 9999 1 Silo Silo The number of automatic starts corresponds to the number of samples in the sample data silo. 8

2 Parameters 2 Parameters 2.1 Coulometric Karl Fischer titrations (KFC) 2.1.1 Control parameters Under [Control parameters], the control parameters for the endpoint are defined. Endpoint at Titration rate Measured value for the endpoint. 1250.0... 1250.0 mv 50.0 mv Three predefined sets of parameters can be selected for the titration rate. slow optimal fast user optimal slow For samples with a low water content or samples which release their moisture only slowly. optimal For all standard titrations. The parameters have been optimized for the most frequent applications. fast For uncritical samples with high water content. user The individual titration parameters can be modified. The settings of the individual titration rates are listed in table 1, page 10. User-defined parameters These parameters are only accessible when Titration rate = User. Dynamics This parameter defines the control range before the specified endpoint. Within the control range, the iodine is generated step by step, the generation is finely controlled. The closer the endpoint, the slower the iodine is generated until the rate defined under Min. rate is reached. The larger the control range, the slower the titration. Outside the control range, 9

2.1 Coulometric Karl Fischer titrations (KFC) iodine is being continuously generated, the rate is defined under Max. rate. 0.1... 1250.0 mv 70.0 mv Max. rate Min. rate Rate at which iodine is generated outside the control range. 1.5... 2241.0 µg/min maximum maximum Rate at which iodine is generated at the very beginning of the titration and in the control range at the end of the titration. This parameter has a decisive influence on the titration rate and thus also on the accuracy. The smaller the selected minimum rate, the slower the titration. 0.3... 999.9 µg/min 15.0 µg/min Table 1 s of the predefined titration rates for KFC Titration rate slow optimal fast Dynamics 120.0 mv 70.0 mv 30.0 mv Max. rate 1000.0 µg/min maximum maximum Min. rate 0.3 µg/min 15.0 µg/min 30.0 µg/min Stop criterion Stop criterion The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled. If no stop criterion has been selected then the titration will not be stopped. The stop conditions (see Chapter 2.1.3, page 12) always lead to a stop, even if the stop criterion has not been reached. drift time rel. drift rel. drift drift The titration is stopped when the stop drift has been reached. time The titration is stopped when the endpoint has been exceeded during a certain period of time (Delay time). 10

2 Parameters rel. drift The titration is stopped when the sum of the drift at the start of the titration and the relative stop drift has been reached. The titration will not be stopped until the stop conditions have been fulfilled. Stop drift Delay time Relative stop drift This parameter can only be edited when Stop criterion = drift. The titration is stopped when the endpoint and the stop drift have been reached. 1... 999 µg/min 5 µg/min This parameter can only be edited when Stop criterion = time. When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped. 0... 999 s 10 s This parameter is can only be edited when Stop criterion = rel. drift. The titration is stopped when the endpoint and the sum of the drift at the start of the titration and the relative stop drift have been reached. 1... 999 µg/min 5 µg/min 2.1.2 Titration parameters Under [Titration parameters], the parameters influencing the sequence of the entire titration are defined. Pause Extraction time Waiting time, for example for dissolving the sample. During this time no iodine is generated. 0... 999999 s 0 s Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time 11

2.1 Coulometric Karl Fischer titrations (KFC) (see Chapter 2.1.3, page 12). Entering an extraction time may e.g. be advisable if a Karl Fischer oven is used. 0... 999999 s 0 s Temperature Time interval MP Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp. measurement = automatic or continuous is defined (see sensor dialog of the command). 20.0... 150.0 C 25.0 C Time interval for entering a measuring point in the measuring point list. The measuring point list is limited to 1000 measuring points. 0.1... 999999.0 s 2.0 s 2.1.3 Stop conditions Under [Stop conditions], the conditions for stopping a titration are defined, if this does not occur automatically. This could be the case when the endpoint set is not reached or if the stop criterion (see "Stop criterion", page 10) is not fulfilled. Stop time The titration is stopped when the specified time has elapsed since the start of the titration. 1... 999999 s 2.1.4 Conditioning Under [Conditioning], the conditions required for conditioning are defined. Conditioning on (: on) If this parameter is activated, then the first time the titration is started the working medium will be titrated to the endpoint with the specified control parameters. The status is kept stable. The actual method run does not begin until [START] has been pressed once more. 12

2 Parameters Start drift Drift correction Drift value If the measured drift is smaller than this value for a certain time (so-called stabilizing time), Conditioning OK is displayed and the titration can be started. The stabilizing time is defined in the dialog Conditioning / Conditioning options. 1... 999 µg/min 20 µg/min The endpoint amount can be corrected taking the drift value into account. For this, the drift is multiplied with the drift correction time and this value is then subtracted from the endpoint amount. The drift correction time is the time interval between the end of conditioning and the end of the determination. auto manual auto auto The value of the current drift is automatically applied at the start of the titration. manual If the drift is known throughout a longer period of time, this can be entered manually. No drift correction takes place. This parameter can only be edited when Drift correction = manual. Drift for manual drift correction. 0.0... 99.9 µg/min 0.0 µg/min Measured value display on (: ) If this parameter is activated, the currently measured value is displayed during the conditioning. 13

2.1 Coulometric Karl Fischer titrations (KFC) Automatic start [Autom. start] Automatic start Threshold value on (: ) If this parameter is activated, the determination is automatically started when the measured voltage suddenly changes. This setting will be ignored as long as the working medium has not been conditioned. If the change in voltage is higher than this value, the determination will automatically be started. 0... 999 mv 50 mv Conditioning options [Cond. options] Stabilizing time Cond. stop time Delay reconditioning Waiting time, during which the measured drift has to be smaller than the start drift defined until Conditioning OK is displayed. The start drift is defined in the dialog Edit command / Conditioning. 0... 999999 s 0 s Maximum permissible time over which conditioning may take place. Conditioning is stopped when the specified time has elapsed. on (: ) 1... 999999 s If this parameter is activated, the conditioning will not start before all the method commands have been processed. Otherwise, the conditioning will immediately start after the titration command. Note This parameter must be activated if commands follow which must not run simultaneously with the conditioning. Example: Emptying the titration cell followed by adding new working medium. 14

2 Parameters 2.1.5 Cell Generator electrode Generator current Reagent monitoring Under [Cell], the parameters for the titration cell are defined. Type of generator electrode. without diaphragm with diaphragm without diaphragm without diaphragm Generator electrode without diaphragm. with diaphragm Generator electrode with diaphragm. Polarization current at the generator electrode. 100 ma 200 ma 400 ma auto 400 ma, if Generator electrode = without diaphragm. auto The current is adapted to the conductivity of the reagent and automatically reduced near the endpoint., if Generator electrode = with diaphragm. of the reagent from the list of reagents. The selection depends on the titration mode. Reagents are defined under System Reagents. The reagent monitoring is not active. of configured reagents 2.1.6 Control device Under [Control device], the control device is selected the determination is carried out with. Control devices are defined under System Device manager. Note This button is only displayed when several control devices have been configured. 15

2.1 Coulometric Karl Fischer titrations (KFC) Control device of the control device from the list of devices. Only those devices are displayed which are able to carry out the command. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another. of configured control devices 2.1.7 Sensor Sensor I(pol) Electrode test Temp. measurement Under [Sensor], the parameters for the sensor are edited. of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System Sensors. You can also enter a sensor name which is not contained in the sensor list. When a determination is started there is a check whether the sensor is contained in the sensor list. of configured sensors The polarization current is the current that is applied to a polarizable electrode during the voltametric measurement. 5 µa 10 µa 20 µa 30 µa 10 µa on (: ) For polarizable electrodes and the generator electrode, an electrode test can be carried out. The following will be tested: Polarizable electrodes Is the electrode connected? Is there a short circuit? The electrode test is carried out when this command is started. Generator electrode: Is the electrode connected? Is the generator system defective? Is the resistance of the sample solution very high? The electrode test is carried out continuously during the execution of this command. Type of temperature measurement. 16

2 Parameters continuous automatic automatic continuous A temperature sensor must be connected. The temperature is measured continuously. automatic If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters). The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters). 2.1.8 Stirrer Stirrer Stirring rate Under [Stirrer], the parameters for the stirrer are edited. of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed. 1 2 3 4 1 No stirrer will be used. Setting the stirring rate. It can be set in steps of 15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in chapter 3.1, page 27. The optimum stirring rate can be tested in the manual control. The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means: "+": counterclockwise rotation " ": clockwise rotation 15... 15 8 Switch automatically on (: on) If this parameter is activated, the stirrer will be switched automatically at the end of the titration, measurement, etc. 17

2.2 Bromine index determination (BRC) 2.2 Bromine index determination (BRC) 2.2.1 Control parameters Under [Control parameters], the control parameters for the endpoint are defined. Endpoint at Titration rate Measured value for the endpoint. 1250.0... 1250.0 mv 200.0 mv Three predefined sets of parameters can be selected for the titration rate. slow optimal fast user optimal slow For samples with slow bromine addition. optimal For all standard titrations. The parameters have been optimized for the most frequent applications. fast For uncritical samples with fast bromine addition. user The individual titration parameters can be modified. The settings of the individual titration rates are listed in table 2, page 19. User-defined parameters These parameters are only accessible when Titration rate = User. Dynamics This parameter defines the control range before the specified endpoint. Within the control range, the bromine is generated step by step, the generation is finely controlled. The closer the endpoint, the slower the bromine is generated until the rate defined under Min. rate is reached. The larger the control range, the slower the titration. Outside the control range, bromine is continuously generated, the rate is defined under Max. rate. 0.1... 1250.0 mv 400.0 mv 18

2 Parameters Max. rate Min. rate Rate at which bromine is generated outside the control range. 10.0... 19876.0 µg/min 1000.0 µg/min maximum Rate at which bromine is generated at the very beginning of the titration and in the control range at the end of the titration. This parameter has a decisive influence on the titration rate and thus also on the accuracy. The smaller the selected minimum rate, the slower the titration. 1.0... 999.9 µg/min 25.0 µg/min Table 2 s for the predefined titration rates for BRC Titration rate slow optimal fast Dynamics 400.0 mv 400.0 mv 400.0 mv Max. rate 500.0 µg/min 1000.0 µg/min 2000.0 µg/min Min. rate 15.0 µg/min 25.0 µg/min 50.0 µg/min Stop criterion Stop criterion The titration is stopped when the endpoint has been reached and this stop criterion has been fulfilled. If no stop criterion has been selected then the titration will not be stopped. The stop conditions (see Chapter 2.2.3, page 21) always lead to a stop, even if the stop criterion has not been reached. drift time rel. drift drift & time drift & time drift The titration is stopped when the stop drift has been reached. time The titration is stopped when the endpoint has been exceeded during a certain period of time (Delay time). rel. drift The titration is stopped when the sum of the drift at the start of the titration and the relative stop drift has been reached. 19

2.2 Bromine index determination (BRC) The titration will not be stopped until the stop conditions have been fulfilled. drift & time The titration is stopped when the stop drift has been exceeded during the time set and the endpoint has been reached. Stop drift Delay time Relative stop drift This parameter can only be edited when Stop criterion = drift or drift & time. The titration is stopped when the endpoint and the stop drift have been reached. 1... 999 µg/min 15 µg/min This parameter can only be edited when Stop criterion = time or drift & time. When the endpoint has been reached, the specified time is allowed to elapse after the last dosing and the titration is then stopped. 0... 999 s 40 s This parameter is can only be edited when Stop criterion = rel. drift. The titration is stopped when the endpoint and the sum of the drift at the start of the titration and the relative stop drift have been reached. 1... 999 µg/min 5 µg/min 2.2.2 Titration parameters Under [Titration parameters], the parameters influencing the sequence of the entire titration are defined. Pause Waiting time, for example for dissolving a sample. During this time no bromine is generated. 0... 999999 s 0 s 20

2 Parameters Extraction time Temperature Time interval MP Minimum duration of the titration. The titration will not be stopped during the extraction time, even if the endpoint has already been reached. The titration is however stopped if a stop condition is fulfilled during this time (see Chapter 2.2.3, page 21). 0... 999999 s 0 s Temperature entered manually. The temperature is being continuously measured when a temperature sensor is connected and when Temp. measurement = automatic or continuous is defined (see sensor dialog of the command). 20.0... 150.0 C 25.0 C Time interval for entering a measuring point in the measuring point list. The measuring point list is limited to 1000 measuring points. 0.1... 999999.0 s 2.0 s 2.2.3 Stop conditions Under [Stop conditions], the conditions for stopping a titration are defined, if this does not occur automatically. This could be the case when the endpoint set is not reached or if the stop criterion (see "Stop criterion", page 19) is not fulfilled. Stop time The titration is stopped when the specified time has elapsed since the start of the titration. 1... 999999 s 2.2.4 Conditioning Under [Conditioning], the conditions required for conditioning are defined. 21

2.2 Bromine index determination (BRC) Conditioning Start drift Drift correction Drift value on (: on) If this parameter is activated, then the first time the titration is started the working medium will be titrated to the endpoint with the specified control parameters. The status is kept stable. The actual method run does not begin until [START] has been pressed once more. If the measured drift is smaller than this value for a certain time (so-called stabilizing time), Conditioning OK is displayed and the titration can be started. The stabilizing time is defined in the dialog Conditioning / Conditioning options. 1... 999 µg/min 20 µg/min The endpoint amount can be corrected taking the drift value into account. For this, the drift is multiplied with the drift correction time and this value is then subtracted from the endpoint amount. The drift correction time is the time interval between the end of conditioning and the end of the determination. auto manual auto The value of the current drift is applied automatically at the start of the titration. manual If the drift is known throughout a longer period of time, this can be entered manually. No drift correction takes place. This parameter can only be edited when Drift correction = manual. Drift for manual drift correction. 0.0... 999.9 µg/min 0.0 µg/min Measured value display on (: ) If this parameter is activated, the currently measured value is displayed during the conditioning. 22

2 Parameters Conditioning options [Cond. options] Stabilizing time Cond. stop time Delay reconditioning Waiting time, during which the measured drift has to be smaller than the start drift defined until Conditioning OK is displayed. The start drift is defined in the dialog Edit command / Conditioning. 0... 999999 s 0 s Maximum permissible time over which conditioning may take place. Conditioning is stopped when the specified time has elapsed. on (: ) 1... 999999 s If this parameter is activated, the conditioning will not start before all the method commands have been processed. Otherwise, the conditioning will immediately start after the titration command. Note This parameter must be activated if commands follow which must not run simultaneously with the conditioning. Example: Emptying the titration cell followed by adding new working medium. 2.2.5 Cell Generator electrode Under [Cell], the parameters for the titration cell are defined. Type of generator electrode. without diaphragm with diaphragm with diaphragm without diaphragm Generator electrode without diaphragm. with diaphragm Generator electrode with diaphragm. 23

2.2 Bromine index determination (BRC) Generator current Reagent monitoring Polarization current at the generator electrode. 100 ma 200 ma 400 ma auto auto auto The current is adapted to the conductivity of the reagent and automatically reduced near the endpoint. of the reagent from the list of reagents. The selection depends on the titration mode. Reagents are defined under System Reagents. The reagent monitoring is not active. of configured reagents 2.2.6 Control device Under [Control device], the control device is selected the determination is carried out with. Control devices are defined under System Device manager. Note This button is only displayed when several control devices have been configured. Control device of the control device from the list of devices. Only those devices are displayed which are able to carry out the command. The control devices do not have to be connected. This allows to transfer methods more easily from one system to another. of configured control devices 2.2.7 Sensor Sensor Under [Sensor], the parameters for the sensor are edited. of the sensor from the sensor list. The selection depends on the measuring mode. Sensors are defined under System Sensors. You can also enter a sensor name which is not contained in the sensor list. When a 24

2 Parameters determination is started there is a check whether the sensor is contained in the sensor list. of configured sensors I(pol) Electrode test Temp. measurement The polarization current is the current that is applied at a polarizable electrode during the voltametric measurement. 125.0... 125.0 µa (Increment: 2.5) 1.0 µa 1.0 µa 1.0 µa on (: ) For polarizable electrodes and the generator electrode, an electrode test can be carried out. The following will be tested: Polarizable electrodes Is the electrode connected? Is there a short circuit? The electrode test is carried out when this command is started. Generator electrode: Is the electrode connected? Is the generator system defective? Is the resistance of the sample solution very high? The electrode test is carried out continuously during the execution of this command. Type of temperature measurement. continuous automatic automatic continuous A temperature sensor must be connected. The temperature is measured continuously. automatic If a temperature sensor is connected then the temperature will be measured continuously. Otherwise, the temperature entered manually will be used (see dialog of the titration and measuring parameters). The temperature will not be measured. The temperature entered manually is used (see dialog of the titration and measuring parameters). 25

2.2 Bromine index determination (BRC) 2.2.8 Stirrer Stirrer Stirring rate Under [Stirrer], the parameters for the stirrer are edited. of the MSB connector the stirrer is connected to. Every MSB connector will always be displayed. 1 2 3 4 1 No stirrer will be used. Setting the stirring rate. It can be set in steps of 15 to +15. The default setting 8 corresponds to 1000 rpm. The formula for calculating the rotational speed is specified in chapter 3.1, page 27. The optimum stirring rate can be tested in the manual control. The algebraic sign of the stirring rate changes the direction in which the stirring is done. When the stirrer is viewed from above, this means: "+": counterclockwise rotation " ": clockwise rotation 15... 15 8 Switch automatically on (: on) If this parameter is activated, the stirrer will be switched automatically at the end of the titration, measurement, etc. 26

3 Appendix 3 Appendix 3.1 Stirring rate The stirring rate can be adjusted in steps of 15 to +15. The approximate rotational speed can be calculated with the following formula: Example: Stirring rate set: 8 Rotational speed/min (r/min) = 125 stirring rate Rotational speed in rpm = 125 8 = 1000 r/min 2000 1500 1000 500-15 -10-5 0 5 10 15 Figure 1 Rotational speed depending on stirring rate 27

Index Index B BRC Parameters... 18 C Cell BRC... 23 KFC... 15 Command BRC... 18 KFC... 9 Conditioning BRC... 21 KFC... 12 Control device Parameters... 15, 24 Control parameters BRC... 18 KFC... 9 E Electrode Parameters... 16, 24 F Favorite... 5 Add... 6 Delete... 6 Edit... 6 K KFC Parameters... 9 M Methods Save as favorite... 5 P Parameters BRC... 18 KFC... 9 R Reagent... 1 Add... 1 Delete... 1 Edit... 1 Reagent exchange... 4 Reagent monitoring... 2 S Sample data silo Save as favorite... 5 Sensor Parameters... 16, 24 Stirrer Parameters... 17, 26 Stirring rate... 27 Stop conditions BRC... 21 KFC... 12 Stop criterion BRC... 19 KFC... 10 T Titration parameters BRC... 20 KFC... 11 28