Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Similar documents
hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Gamma Spectroscopy. References: Objectives:

GAMMA RAY SPECTROSCOPY

Scintillation Detector

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Gamma Ray Spectroscopy

Analysis of γ spectrum

Experiment 6 1. The Compton Effect Physics 2150 Experiment No. 6 University of Colorado

MEASURING THE LIFETIME OF THE MUON

EEE4106Z Radiation Interactions & Detection

DETECTORS. I. Charged Particle Detectors

THE COMPTON EFFECT Last Revised: January 5, 2007

Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Nuclear Physics and Astrophysics

The Mössbauer Effect

Radionuclide Imaging MII Detection of Nuclear Emission

Detection and measurement of gamma-radiation by gammaspectroscopy

Chapter 4 Scintillation Detectors

Gamma Ray Spectroscopy, Using NaI(Tl) Detectors. Department of Physics The University of Hong Kong

Mass of the electron m 0

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

SCINTILLATION DETECTORS AND PM TUBES

Radiation Detection and Measurement

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

AN34 Experiments in Nuclear Science Laboratory Manual Fourth Edition

PHY 192 Compton Effect Spring

X-RAY SPECTRA. Theory:

Radiation Detection. Laboratory & Computational Physics 2. Last compiled March 2, Constructed from legacy notes with amendments by Innes Bigaran

Radiation (Particle) Detection and Measurement

Radiation Detection. Laboratory & Comp. Physics 2

Lecture 3 - Compton Scattering

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

Rice University Physics 332 LIFETIME OF THE MUON I. INTRODUCTION...2! II. MEASUREMENT PROCEDURES...3! III. ANALYSIS PROCEDURES...7!

Chapter NP-4. Nuclear Physics. Particle Behavior/ Gamma Interactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 IONIZATION

Radioactivity and Ionizing Radiation

Overview: In this experiment we will study the decay of a radioactive nucleus, Cesium. Figure 1: The Decay Modes of Cesium 137

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

III. Energy Deposition in the Detector and Spectrum Formation

Nuclear Spectroscopy (1) Precautions

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Quality Assurance. Purity control. Polycrystalline Ingots

Absolute activity measurement

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

"Neutron Flux Distribution"

Overview: In this experiment we study the decay of a radioactive nucleus, Cesium 137. Figure 1: The Decay Modes of Cesium 137

Experiments 30a and 30b INTERACTION OF GAMMA RAYS WITH MATTER

Total probability for reaction Yield

Chapter Seven (Nuclear Detectors)

David B. Cassidy. Department of Physics and Astronomy, University of California, Riverside, USA. Varenna, July 09

dn(t) dt where λ is the constant decay probability per unit time. The solution is N(t) = N 0 exp( λt)

Copyright 2008, University of Chicago, Department of Physics. Gamma Cross-sections. NaI crystal (~2" dia) mounted on photo-multiplier tube

Chapter 2 Problem Solutions

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

PHYS 3446 Lecture #12

Particle Detectors. How to See the Invisible

Radiation Signals and Signatures in a Detector (Gamma spectroscopy) Sangkyu Lee

1.4 The Tools of the Trade!

Advanced lab course for Bachelor s students

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

1.5. The Tools of the Trade!

CHAPTER 3 Prelude to Quantum Theory. Observation of X Rays. Thomson s Cathode-Ray Experiment. Röntgen s X-Ray Tube

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000

ORTEC AN34 Experiment 10 Compton Scattering

Physics 736. Experimental Methods in Nuclear-, Particle-, and Astrophysics. Lecture 3

THE MÖSSBAUER EFFECT

Modern Physics Laboratory Beta Spectroscopy Experiment

EQUIPMENT Beta spectrometer, vacuum pump, Cs-137 source, Geiger-Muller (G-M) tube, scalar

Physics 126 Practice Exam #4 Professor Siegel

Chapter Four (Interaction of Radiation with Matter)

Particle Energy Loss in Matter

Rb, which had been compressed to a density of 1013

λ φ φ = hc λ ev stop φ = λ φ and now ev stop λ ' = Physics 220 Homework #2 Spring 2016 Due Monday 4/11/16

Particle Energy Loss in Matter

Planck s Quantum Hypothesis Blackbody Radiation

Physics 201: Experiment #1 Franck-Hertz Experiment

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)

Foundations of Modern Physics by Tipler, Theory: The dierential equation which describes the population N(t) is. dn(t) dt.

RANGE OF ALPHAS. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

Radioactivity. is related to de/dx. The range, R, is defined by the integral of de/dx:

CHAPTER 3 The Experimental Basis of Quantum Theory

ATOMIC WORLD P.1. ejected photoelectrons. current amplifier. photomultiplier tube (PMT)

Experiment objectives: measure the ratio of Planck s constant to the electron charge h/e using the photoelectric effect.

INTRODUCTION TO MEDICAL PHYSICS 1 Quiz #1 Solutions October 6, 2017

Trace Element Analysis of Geological, Biological & Environmental Materials By Neutron Activation Analysis: An Exposure

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

Applied Nuclear Physics (Fall 2006) Lecture 19 (11/22/06) Gamma Interactions: Compton Scattering

Energy loss of alpha particles - Prelab questions

University of Michigan Physics : Advanced Laboratory Notes on RADIOACTIVITY January 2007

Particle Interactions in Detectors

Physics 111 Homework Solutions Week #9 - Friday

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

BETA-RAY SPECTROMETER

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Computer 3. Lifetime Measurement

Transcription:

Experiment VI Gamma Ray Spectroscopy 1. GAMMA RAY INTERACTIONS WITH MATTER In order for gammas to be detected, they must lose energy in the detector. Since gammas are electromagnetic radiation, we must understand how such radiation interacts with matter. There are three distinct modes of interaction: photo-electric effect, Compton scattering, and pairproduction. 1.1 Photo-electric effect A gamma (or photon) strikes an electron, releasing the electron from its surroundings and giving it kinetic energy. The photon gives up its entire energy in this process. 1.2 Compton scattering A gamma scatters from a free electron. The process produces a photon of reduced energy and an electron with kinetic energy. Note that Rayleigh scattering, or elastic scattering, may be considered a special case of Compton scattering with the electron replaced by a more massive object, e.g., a nucleus. Compton scattering is shown in Fig. 1. 77

The relation between the incoming and outgoing gamma energies is given by the Compton scattering formula: E = 1 + E mc 2 E ( 1 cosθ ) (1) where mc 2 = 0.511 MeV, is the rest-energy of the electron and the other variables are defined in Fig. 1. 1.3 Pair production A gamma passes near a massive nucleus. The energy of the gamma is converted to mass (an electron-positron pair) with kinetic energy. For this process to occur, the gamma must have at least twice the rest-energy of the electron, or 2 x 0.511 MeV. 78

The cross-sections for each of these interactions in sodium iodide (NaI) as a function of gamma energy is shown in Fig. 2. It is important to note that in each of the modes of interaction, the gamma s energy is converted to an electron having kinetic energy. The electron, and charged particles in general, traveling through matter, lose energy in a manner different from gammas. Fig. 2 Electromagnetic interaction cross-sections for NaI. 79

2. Electron interaction with matter A vast majority of the volume of bulk matter contains electrons. Those electrons feel a Coulomb force as another electron moves through the bulk matter. The cumulative forces acting on the moving electron bring it to rest in a short distance. For example, if a 1 MeV gamma enters a NaI crystal, the most likely mode of interaction is Compton scattering (see Fig. 2). If the gamma undergoes a head-on collision with an electron in the crystal, the gamma will be scattered by180 degres. Applying eq.(1), show that this electron will have kinetic energy of about 0.8 MeV. Is this sufficient energy to kick the electron loose from the bound state in the crystal? Is this electron relativistic? Such an electron will move only about 1 mm in NaI before losing all of its kinetic energy. Approximate experimental values of the range of electrons in NaI as a function of electron energy are shown below: 80

3. THE DETECTOR 3.1 Sodium Iodide (NaI) scintillators In this experiment the gammas are detected by a sodium iodide crystal doped with a small amount of thallium, NaI(Tl). The gammas entering the NaI crystal may undergo photoelectric effect, but more probably will Compton scatter from the electrons in the crystal (see Fig. 2). There are two products of this scattering: a scattered electron and a gamma of reduced energy. The electron loses energy very quickly in the crystal. The rate of energy loss, -de/dx, is governed by the Bethe formula (see Knoll for details). In the process of moving through the crystal, the scattered electron excites electrons in the crystal's valence band into the conduction band, leaving a trail of electron-hole pairs. Holes migrate to the thallium impurity sites and ionize them. The ionized thallium atoms capture electrons, ultimately return to the ground state and give off visible light. Meanwhile, the remaining gamma moves through the crystal losing energy only when it scatters from another electron. With each successive scattering, the remaining gamma has less energy. At some point the gamma energy is small enough that the photoelectric cross section is dominant and the remaining gamma gives up its energy entirely to an ejected electron, which loses energy in the same manner as the electrons mentioned above. 3.2 Photo-multiplier (PM) tube The light produced in the scintillator passes through a glass window and into the photomultiplier tube where it liberates an electron by the photoelectric effect in the photo-cathode of the tube. The photo-electrons are accelerated through a series of metal plates called dynodes. When an energetic electron strikes a dynode, several more electrons are ejected and are, in turn, accelerated to the next dynode. After typically 10 to 14 dynodes, the electron gain may be about 81

10 6. The electrons are collected at the anode of the PM tube. What polarity (+ or -) would you expect at the dynode output? At the anode output? 4. NUCLEAR DECAY SCHEMES At the end of this write-up you will find the nuclear decay schemes for the isotopes you will use in this experiment. Each column of horizontal lines represents the nuclear energy-levels of a specific isotope of some element. Diagonal arrows indicate the process involved in changing from one element to another. Vertical arrows indicate changes in nuclear energy-level within one element. Gammas are emitted when a nucleus changes from a higher to a lower energy nuclear state within the same isotope. 82

5. PROCEDURE 5.1 PM pulses on the scope a. Place a Cs-137 source near the scintillator. The PM tubes we will be using may be operated safely at (negative) 2000 volts, although we will generally use them at 1500 volts or less. b. Turn on the high voltage power supply. Note that there are two switches on the supply. Starting with both switches off, turn on the power switch. Wait a few seconds for a click and then turn on the high voltage switch. Adjust the high voltage to about 1500 volts. c. Attach the 50 ohm terminator to an input of the scope, and attach a coaxial cable from the PM anode output to the terminated scope input. d. Adjust the scope s gain, trigger slope and trigger level until the PM pulses are clearly visible. e. Sketch the pulses in your lab notebook. What is the pulse polarity? Does this polarity make sense, given how the PM works? What is the duration of the pulses? What is the range of pulse heights? What are the rise and fall times of the pulses (see Fig.5)? 83

Fig. 5 Defining rise and fall for a positive and negative pulse. 5.2 Termination a. Remove the terminator and re-connect the cable directly to the scope. Look at the pulses again and try to measure their duration and height with no terminator at the scope input. b. The RG58-U coaxial cable you are using has a characteristic impedance of 50 ohms, (independent of its length!). On the other hand, the input impedance of the scope is 1 MΩ. This change in impedance from low to high at the cable-scope interface will cause a reflection of the pulse without a change in polarity. The pulses observed on the scope are the sum of the incident and reflected pulses. Using the terminator places 50Ω in parallel with the scope s 1MΩ input impedance. What is the value of this parallel combination? How will the reflection be affected by the use of the 50Ω terminator? 84

5.3 Pulse heights Let us now try to make sense of the observed pulse heights. Make use of the following estimates to calculate the expected pulse heights from a Cs-137 source: a. The Cs-137 gamma energy is 0.662 MeV. b. The scintillation efficiency of NaI is about 10%, i.e., 10% of the gamma energy goes into visible scintillation light. c. The energy of each scintillation photon is about 4 ev. d. The light collection efficiency of the NaI-PM combination is about 70%. e. The photon to electron conversion (photo-electric) efficiency is about 10%. f. Each photo-electron at the entrance of the PM yields about 10 6 electrons collected at the anode. Using these estimates and the pulse duration you measured earlier, calculate the expected pulse height for Cs-137. Hint: a-f above should give the charge in each pulse. Additionally, the duration should give the current. This current through the terminator s resistance should give the voltage. Compare the estimated pulse heights with what you measured earlier. 5.4 Vary high voltage and gamma energy Still using the Cs-137 source, vary the high voltage and record its effect on the pulse height. Return to 1500 volts and change the source. Use sodium-22 (Na-22), cobalt-57 (Co-57), Co-60, In each case sketch the pulses observed and record their pulse heights in your notebook. 85

5.5 Pulse height analysis We would like to perform a quantitative analysis of the pulse heights. To do so, a histogram of heights may be created by using a pulse height analyzer (PHA). On the PHA the horizontal axis, channel number is proportional to pulse height. The vertical axis gives number of pulses having each height. Such a histogram is usually called a spectrum. The UCS-20 spectrometer may be used as a PHA. 5.6 PHA set-up a. Plug the coaxial cable from the photo-multiplier s (PM) anode into the pre-amp input on the rear panel of the UCS-20. This input is designed to take negative pulses from a PM tube without a 50 ohm termination. b. Turn on the UCS-20. Pull down the mode menu and select pulse height analysis. Quit the program and re-start it. (This action works past a minor bug in the software.) c. Start the UCS-20 software. d. Under the Settings menu, select Conversion Gain, 2048. This setting puts the pulse height range of 0 to 10 volts into channels 0 to 2047. e. Drag the LLD tab (lower level discriminator) to approximately the 5% position. This setting serves to eliminate low-level electronic noise pulses. f. Place a Co-60 source near the scintillator. Click the start box to start taking data. You may have to reduce the high voltage to get the entire spectrum on the screen. Note that for each source, there is a maximum energy gamma and therefore, an upper limit to the pulseheight. Thus, if the spectrum shows activity all the way to the upper (right) end of the 86

display, some of the spectrum must be off-scale to the right. In this case reduce the pulse heights by reducing the high voltage. g. Adjust the high voltage and/or the PHA gain so that both full-energy peaks appear near the upper end of the display. Refer to the nuclear decay schemes at the end of the manual to help identify the origin of the peaks. Leave the gain and HV set for the duration of the experiment. 5.7 Energy calibration We wish to re-scale the horizontal axis of the PHA from channel number to energy in MeV (million electron-volts). To do so we will use two known gamma energies. Be sure to leave the HV and preamp gain fixed for the rest of the experiment in order to preserve your calibration. a. Look up the energies of the two gammas of Co-60. Using your spectrum of Co-60, determine which peak is due to the 1.33 MeV gamma. b. Use the mouse to move the PHA cursor to the 1.33 MeV peak position. Record the channel number for this peak. c. Erase the data and acquire a spectrum for Co-57. Identify its peaks and their energies from the decay schemes. d. Move the cursor to the 0.122 MeV peak and record this channel number. e. Under the Calculations menu, select Energy Calibrate. f. Place the cursor in the channel corresponding to the 0.122 MeV peak. g. In the calibration window, click the Energy button and enter 0.122, and click OK. h. Click the Units button and enter MeV. i. Repeat the above steps to enter 1.33 MeV at its appropriate channel. 87

j. Click the Calibrate button. The horizontal axis should now be re-scaled and labeled in units of MeV. 5.8 Measure other gamma energies a. For the Na-22, Co-57 and Co-60 sources, acquire spectra and sketch them into your notebook. Measure the energies of the gamma peaks you observe and compare them to the decay schemes for each source. b. For each of the gammas, plot the energy vs. PHA channel number. How are they related? Does this relation make sense in light of the physical processes involved in the detection? Please explain. 5.9 Cs-137 spectrum a. Acquire a spectrum of Cs-137 with sufficient data to give a fairly smooth spectrum. Sketch the spectrum. Measure the gamma energy. b. Using the Compton formula, calculate the predicted energy of the Compton edge..note that this feature represents the maximum energy given by the gamma to an electron due to a single, head-on collision. Move the cursor to the calculated energy of the Compton edge. Identify this feature on your sketch. c. Calculate the energy of the backscatter peak. This feature represents gammas which have undergone a large angle scatter outside the scintillator and then enter the scintillator, losing all of their remaining energy in the scintillator. Move the cursor to the calculated energy. Identify this feature on your sketch of the spectrum. d. and the full-energy peak? 88

5.10 Natural Line Width of the Cs-137 Transition (calculation) From the Cs-137 decay scheme, determine the lifetime of the state which gives rise to the 0.662 MeV gamma. Use the Uncertainty Principle, ΔE Δt h, to estimate its natural line-width. 5.11 Detector energy-resolution The sources emit gammas with very well-defined energies. By contrast, the full-energy peaks displayed on the PHA have substantial width. The ability of the detector to resolve two closely spaced peaks is related to the peak widths. The widths are dominated by statistical fluctuations in the number of photo-electrons leaving the photo-cathode of the PM. Poisson statistics tells us that if N photo-electrons are produced at the photo-cathode, the statistical fluctuation in that number will be in the range N ± N From the estimates given in a- e above, calculate the expected number of electrons produced, N for each Cs-137 gamma. The fractional fluctuation is thus given by N / N = 1/ N. Using this expression, calculate the expected spread in energy (peak width). Measure the Cesium peak width and compare it to your calculation. Compare the measured line width to the natural line width of the Cs-137 gamma, calculated above. Can you measure the natural line width with your detector? 89