Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Similar documents
SmartNotes. What is meant by the term interference?

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Automated and accurate component detection using reference mass spectra

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Advanced ICP-MS techniques for overcoming interferences in LA-ICP-MS bioimaging

Defining quality standards for the analysis of solid samples

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

of mass spectrometry

Determination of Carbon d13c isotopic composition from DIC samples

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Introduction to Fourier Transform Infrared Spectroscopy

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

Introduction to Fourier Transform Infrared Spectroscopy

Elemental Analysis: NCS characterization of paper

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Elemental Analysis: CHNS/O characterization of polymers and plastics

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

PRINCIPLE OF ICP- AES

Determination of trace anions in concentrated hydrofluoric acid

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Fast, Effective XPS Point Analysis of Metal Components

Atomic Absorption & Atomic Fluorescence Spectrometry

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Speed and flexibility

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

for XPS surface analysis

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

HELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer

Expanded capabilities in ammonium and amine detection

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Analysis of domestic sludge using the Agilent 4200 MP-AES

9/13/10. Each spectral line is characteristic of an individual energy transition

The key to your laboratory s success

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Trace Analyses in Metal Matrices Using the ELAN DRC II

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Analysis of noble metals at low levels in geological reference materials and ores

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Sample Analysis Design Polyatomic Interferences

δ 13 C of DIC and carbonate samples: comparison of traditional mass spectrometry methods with an infrared spectrometry method

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Ideal and powerful solution

Whole Tablet Measurements Using the Frontier Tablet Autosampler System

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Whole Tablet Measurements Using the Spectrum One NTS Tablet Autosampler System

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Sources of Errors in Trace Element and Speciation Analysis

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

all suspects Identify Thermo Scientific Extractables & Leachables Workflow Unknown impurity identification and quantification workflows for:

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Dionex IonPac AS28-Fast-4µm column

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

MetWorks Metabolite Identification Software

Yin and yang in chemistry education: the complementary nature of FTIR and NMR spectroscopies

Determination of Nanoparticle Sizeand Number Concentration using the npquant Evaluation Module for Qtegra ISDS. The world leader in serving science

Low level Os isotopic measurements using multiple ion counting

Lecture 7: Atomic Spectroscopy

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Pros and Cons of Water Analysis Methods

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

OES - Optical Emission Spectrometer 2000

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Transcription:

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES Technical Note 43332 Key Words Overcoming Interferences, Element Finder plug-in Introduction Interferences are common in Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) for many different reasons; matrix, sample transport, ionization, suppression or spectral overlaps. These are commonly encountered in environmental, metallurgy and geological applications where there is a mixture of very high and low concentrations of elements resulting in a spectrally rich matrix. Ideally, physical and chemical effects should be experimentally eliminated and the use of a solid state detector should allow the analyst to select an interference free wavelength for analysis, but if this is not possible then alternative techniques must be used. There are three main types of interferences that are encountered during ICP-OES analysis: physical, chemical and spectral. Physical interferences are characterized as a difference between the samples and calibration standards which affect the sample transport or nebulization these differences can be viscosity, density, matrix (from the sample itself or the digestion/preservation procedure). Chemical interferences occur when there is a difference in the way that the sample and the calibration standard react in the plasma such as ionization, molecular formation and plasma loading. Finally, spectral interferences are characterized by an overlap (either partial or direct) of the analyte of interest by an interfering element. These interferences are normally apparent during method development and can lead to suppression or enhancement of signals, and thus, false negative or positive results which ultimately degrade the accuracy and precision of the method. This technical note will focus on spectral interferences and how to overcome these using analytical approaches and software tools. Spectral Interferences Spectral interferences are sometimes severe in ICP-OES due to the effectiveness of the plasma as an excitation source and the complexity of the emission spectrum of the elements and their ions. These interferences can be a shoulder on a peak, a partial or direct spectral overlap. It is a challenge to soundly analytically correct for a direct spectral interference, so a wavelength experiencing this uncommon phenomenon would be eliminated at the early stages of method development. Partial and peak overlaps are dealt with by either off-peak background correction (as per Figures 1 and 2) or with the use of an experimentally derived mathematical correction. The correct use of background correction is paramount when developing a method and can be used to alleviate or negate the effect of a spectral interference as shown below in Figure 1. The sample contains 1 mg L -1 boron, which is being interfered with at the B 249.678 nm wavelength by the presence of 100 mg L -1 iron.

2 Figure 1. 1 mg L -1 Boron in the presence of 100 mg L -1 Iron. This background interference is visible on the smoothed subarray display and the interferent is identified by the Interferences list as Fe 249.653 nm. Due to this background interference (and minor wing overlap), the 1 mg L -1 boron is reading as 0.4804 mg L -1. By turning off the left background point, the method can be salvaged and the interference negated see Figure 2 below where the boron now reads at 1.004 mg L -1. If the analyst prefers to use two background points, it is possible to extend the width of the subarray in the Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) software to accommodate this. Figure 2. Use of flexible background correction. Figure 3 below displays the interferent - 100 mg L -1 iron, a 1 mg L -1 boron solution, and the sample, which contains both boron and iron to demonstrate the spectral overlap contributions. Figure 3. Spectral overlap. Qtegra ISDS can use an Inter-Element Correction (IEC) calculation to correct for specific interferences. This is a ratio of the intensity of the interfered analyte wavelength divided by the intensity of an undisturbed wavelength of the interfering element. Figure 4 below shows a sample of boron being interfered with by iron, which causes a false result for a 1 mg L -1 boron, solution of 1.208 mg L -1. This is due to the shoulder overlap from Fe 249.782 nm on B 249.773 nm.

To set up an IEC for a method, the analyst must first identify the contaminant this is done by clicking on the interferent peak and using the Interferences list (see below Figure 4) to correctly identify what element is causing the overlap. 3 Figure 4. Use of interferences list to identify the interfering element. When the interfering element has been identified, the analyst adds a primary wavelength (in this case Fe 259.940 nm) to the method and analyses a single element solution (to avoid further interferences, it must be a single element solution) as a sample, which allows the apparent intensity of iron to be read as boron. Right-clicking on the false positive intensity allows you to select the interfering element and IEC correction factor will automatically be calculated. Steps to determine and automatically calculate an IEC using Qtegra ISDS 1. Isolate the interferent 2. Add the primary wavelength for the interferent to the method 3. Analyze an appropriate concentration of the interferent (high purity, single element) as a sample 4. Right click on the false positive intensity of the interfered wavelength 5. Select the wavelength of the interferent and let Qtegra ISDS calculate the IEC correction factor Figure 5. Selection of interfering element in the Inter-Element Correction table. 6. The IEC correction factor will be applied to each sample that is analyzed in this LabBook and can easily be switched on and off via an icon in the headline of Qtegra ISDS. IECs should be re-assessed if there is a change in the sample type or matrix, which may affect the correction.

4 Element Finder plug-in for the Qtegra ISDS software The use of an IEC as described is a well-accepted correction technique and prescribed by many standard protocols such as Untied States EPA Method 200.7 Trace Elements in Water, Solids, and Biosolids by Inductivity Coupled Plasma-Atomic Emission Spectrometry. However, selection of wavelengths that are interference free would be an ideal solution in all cases this is especially true when carrying out IEC corrections close to the detection limit of the instrument (as the small uncertainty associated with any measurement can be critical). However, given the spectral complexity of many matrices wavelength selection is a challenge for many analyst as many components of the sample are unknown leading to unexpected interferences. The Element Finder plug-in for the Qtegra ISDS software is a method development tool that allows for the selection of analytical wavelengths that are free from spectral interferences taking into account the analytes present in the sample and the sample matrix. By utilizing Fullframes (an image of the enter sample spectrum, within the wavelength of the instrument) of the sample, the software intelligently selects the wavelengths that are interference free for the elements you have selected. In this way the Element Finder plug-in helps to save time and cost associated with long and cumbersome method development. Steps to utilize the Element Finder plug-in for Qtegra ISDS 1. Select the element you wish to analyze in the sample (in this case boron); we can see there are, potentially, seven wavelengths for analysis. Figure 6. Selection of analyte elements. 2. Optional step - Select the matrix elements in the sample (if they are known). Based on this the Element Finder plug-in will rationalize the wavelengths used for the analysis (if the matrix elements will interfere with the analyte elements).

3. Run a Fullframe using a typical sample containing your matrix and analyte elements (the matrix is iron and the analyte is boron). The Fullframe will be measured and then analyzed for all elements present in the sample. 5 Figure 7. The measurement of a Fullframe within the Element Finder plug-in. In this case we can see that the Fullframe analysis took place for 76 elements and two were identified in the sample. Elements that are not measureable by ICP-OES are excluded from the Fullframe analysis (elements with a short half-life etc). Figure 8. Results of the Fullframe analysis show there are two elements present in the sample (the analyte and the matrix). 4. Taking into account the matrix elements found during the Fullframe analysis the Element Finder plug-in will then suggest the wavelengths to be used for the analysis of the selected analyte. In this case we can see that B 208.959 nm has been selected for the analysis. The suggested wavelengths can then be imported into a LabBook and used for the analysis.

Conclusion Through careful method development and understanding of interferences, precise and accurate results can be achieved for almost any sample. This process is simplified with the Thermo Scientific Qtegra ISDS Element Finder plug-in which correctly identifies interfering elements, and enables the analyst to automate their method development. The icap 7000 Plus Series ICP-OES also offers the inherent benefits of CID detector technology, and with over 55,000 wavelengths to choose from, complementing the benefits of the Element Finder plug-in. Technical Note 43332 thermofisher.com/icp-oes 2016 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. TN43332-EN 0216C