Screening for Hundreds of Pesticide Residues Using a GC/Q-TOF with an Exact Mass Pesticide Database in Food

Similar documents
All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Highly Sensitive and Rugged GC/MS/MS Tool

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Multi-residue analysis of pesticides by GC-HRMS

Agilent All Ions MS/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

MassHunter TOF/QTOF Users Meeting

Date ACCURATE-MASS DATABASE FOR SCREENING OF PESTICIDE RESIDUES IN FRUITS AND VEGETABLES BY GAS CHROMATOGRAPHY TIME-OF-FLIGHT MASS SPECTROMETRY

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Rapid Pesticide Screening and Identification Using the High Resolution All Ions MS/MS Technique

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ

Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Author. Abstract. Introduction

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

MassHunter Software Overview

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis

Application Note. Authors. Abstract. Introduction. Environmental

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

The Analysis of Organophosphate Pesticides by LC/MS Application

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Automated and accurate component detection using reference mass spectra

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Workflow Using Hyphenated Mass Spectrometry Techniques

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

GC/Q-TOF MS Surveillance of Pesticides in Food

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Screening and Quantitation of 240 Pesticides in Difficult Food Matrixes Using the Agilent 6545 Q-TOF LC/MS System

Making Sense of Differences in LCMS Data: Integrated Tools

ENVIRONMENTAL analysis

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

LC/Q-TOF Workflows for Comprehensive Micropollutant Analysis

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Rugged GC/MS/MS Pesticide Residue Analysis Fulfilling the USDA Pesticide Data Program (PDP) Requirements

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Agilent G3212 GC-APCI Source

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Multiresidue Analysis of 100 Pesticides in Food Samples by LC/Triple Quadrupole Mass Spectrometry Application

The new Water Screening PCDL

Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012

An Optimal Method for the Analysis of Pesticides in a Variety of Matrices

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Analysis of pesticide residues using various GC/MS systems. Radim Štěpán Czech Agriculture and Food Inspection Authority Prague, Czech Republic

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

U.S. EPA Method 8270 for multicomponent analyte determination

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Transcription:

Screening for Hundreds of Pesticide Residues Using a GC/Q-TOF with an Exact Mass Pesticide Database in Food Application Note Food Authors Amadeo R. Fernandez-Alba, Samanta Uclés, and Noelia Belmonte-Valles Pesticide Residue Research Group Department of Hydrogeology and Analytical Chemistry University of Almería Almería, Spain Abstract The Agilent 7200 GC/Q-TOF, in concert with the Agilent MassHunter Qualitative Analysis All Ions workflow, and the first commercially available exact mass pesticide library, was used to rapidly screen, identify, and generate quantitative information for pesticide residues in five different food matrices. This technique helps eliminate false positives, and has the speed and accuracy to significantly improve the productivity of pesticide screening and quantitative work. A screening workflow using the new GC/Q-TOF Pesticide Library and the unique All Ions software tools enables the detection of pesticide levels as low as 10 ppb in complex matrices. Joerg Riener Agilent Technologies GmbH Waldbronn, Germany

Introduction With increased international trade in food and food ingredients, there is even more emphasis on food safety. State-ofthe-art pesticide screening requires the consideration of more than 1,000 pesticides and their metabolites. Of these, as many as 600 to 700 compounds can be included in routine monitoring programs. Testing approaches must be able to handle many compounds at a time, while being able to avoid matrix interferences coming from many different food matrices. The increasing global emphasis on pesticide screening is reflected in the implementation of European Union (EU) guideline SANCO/12571/2013 [1]. The most recent revision specifies criteria for qualitative screening supported by databases or libraries. An accurate-mass approach for pesticide screening using quadrupole time-of-flight mass spectrometry (Q-TOF) ensures reliable pesticide identification, and enables a virtually unlimited number of compounds to be screened simultaneously. For many of the most important compounds, gas chromatography (GC) coupled to a Q-TOF mass spectrometer is the ideal analytical tool for screening, confirmation, and quantification of both target and unexpected compounds at trace levels, even in complex matrices. This application note introduces a workflow for the screening of pesticide residues in various foodstuffs using GC/Q-TOF and electron ionization (EI) in combination with a retention time locked GC method [2], midcolumn backflushing for increased method robustness [3], and a novel exact mass pesticide spectral library. Agilent MassHunter Software then automates the screening for more than 700 pesticides that are contained in a Personal Compound Database and Library (PCDL). The Agilent All Ions workflow chooses characteristic exact mass ions for each compound in the PCDL and extracts them from the chromatogram. To verify the hits, a coelution plot and coelution score are created to observe and express the covariance of the extracted accurate mass ions. The coelution score uses the retention time and the entire chromatographic peak information (including peak width and symmetry) to determine covariance of the characteristic ions. Experimental Reagents and standards All high-purity pesticide standards were obtained from Dr. Ehrenstorfer (Augsburg, Germany), Sigma-Aldrich (Steinheim, Germany) and Riedel-de Haën (Selze, Germany), and were stored at 30 C. Individual pesticide stock solutions (1,000 2,000 mg/l) were prepared in acetonitrile and stored in amber screw-capped glass vials in the dark at 20 C. Individual standard solutions, used for the optimization, and one 10 mg/l mix of all the standards in acetonitrile were prepared from the stock standards. The standard mix solution was used for the calibration by appropriate dilution in ethyl acetate. Ethyl acetate was obtained from Fluka Analytical Pestanal; acetonitrile was obtained from Sigma-Aldrich (Steinheim, Germany), and MgSO 4 was obtained from Panreac Quimica S.A. (Barcelona, Spain). Primary secondary amine (PSA) sorbent was obtained from Supelco (Bellefonte, Pennsylvania), and NaCl was from J.T. Baker (Deventer, The Netherlands). Instruments This study was performed using an Agilent 7890B GC system coupled to an Agilent 7200 Series GC/Q-TOF System. The instrument conditions are listed in Table 1, and the instrument system configuration is shown in Figure 1. Sample preparation Vegetable and fruit samples were obtained from local markets. Blank vegetable and fruit extracts were used to prepare the matrix-matched standards for validation purposes. In this way, five types of fruits and vegetables (apple, carrot, leek, tomato, and oranges) were extracted using the QuEChERS method, as previously described [4]. The vegetable extracts were spiked with the mix of standards at different concentrations (ranging from 10 to 200 µg/kg), and subsequently analyzed by GC/Q-TOF. This GC/Q-TOF screening approach complements GC/MS/MS target compound analysis. In addition, retrospective data analysis is also possible since chromatograms with full EI spectra are acquired. For any unexpected compounds, the user can quickly investigate the identities of such compounds with high resolution accurate mass data. If subsequent quantitative screening is considered important for future work, the critical ion information can easily be exported into a quantitative method. If necessary, hundreds of pesticides can be quantified in a single analysis. 2

Table 1. GC conditions Columns Gas Chromatograph and Mass Spectrometer Conditions Multimode injector Agilent HP-5MS Ultra Inert, 15.0 m 0.25 mm, 0.25 µm (p/n 19091S-431UI) Inlet multimode inlet, Outlet pressure controlled tee Agilent HP-5MSUI, 15.0 m 0.15 mm, 0.15 µm (p/n 19091S-431UI) Inlet pressure controlled tee Outlet vacuum Multimode inlet Splitless Injection port Injection mode Injection volume 1.0 µl Injection port liner Ultra inert liner split, straight, wool (5190-2293) Carrier gas Helium at 0.96 ml/min constant flow Oven program 60 C for 1 minute 40 C/min to 120 C for 0 minutes 5 C/ min to 310 C for 0 minutes Retention time locking Chlorpyrifos-methyl locked to 18.111 minutes Backflush Post run, 5 minutes, oven 300 C 40 psi at pressure controlled tee, inlet 1 psi Transfer line temperature 280 C Q-TOF MS conditions Instrument Agilent 7200 Q-TOF Ionization mode EI MS temperatures Source 280 C, Quadrupole 150 C Detection mode 45 550 m/z scan Spectra acquisition rate 5 spectra/s Aux EPC Purged ultimate union Agilent 7200 Q-TOF EI mode Data acquisition and analysis The data were acquired with the MassHunter Acquisition Software B.07.02. Data analysis for the pesticide screening was performed with the All Ions tool in MassHunter Qualitative Analysis Software (B.07.00) and the GC/Q-TOF Pesticide PCDL (p/n G3892A). Data analysis for pesticide quantitation was performed with the MassHunter Quantitative Analysis Software (B.07.01). Results and Discussion The All Ions tool Data analysis was performed using the MassHunter Qualitative Analysis Software (B.07.00). Users can set up parameters for the All Ions MS technique in a new tab in the Find by Formula (FbF) area of MassHunter Qualitative Analysis called Fragment Confirmation (Figure 2). The tab allows the user to specify how many of the most specific ions to extract. Limits can also be set for fragment ion Extracted Ion Chromatograms (EICs) based on retention time (RT) difference, minimum signal-to-noise (S/N) ratio, and coelution score. The EICs of the most specific EI fragments in each PCDL spectrum are extracted and evaluated using a unique coelution score parameter. The coelution score was derived from a technique similar to UV chromatography s Peak Purity [1], in which the software calculates a number that takes into account multiple factors, such as abundance, peak shape (symmetry), peak width, and RT. Figure 3A provides an example of overlaid EICs for the ions derived from bupirimate in carrot extract. All of the ions have the same chromatographic apex and shape, suggesting that they originated from the same compound. The normalized ratios of the fragment ions to the reference ion intensity are plotted across the RT and made available to the user for inspection in a coelution plot (Figure 3B). If all ions exhibit a ratio of approximately 1 across the middle of the reference ion peak, as in this example, there is strong confirmation that the fragments belong to the same compound. Agilent 15-m HP-5ms (0.25 mm id 0.25 µm) Agilent 15-m HP-5ms (0.25 mm id 0.25 µm) Figure 1. GC/Q-TOF configuration with midcolumn backflush. 3

Figure 2. Fragment Confirmation tab from the Find by Formula (FbF) tool in Agilent MassHunter Qualitative Analysis. Counts (%) 1.1 A 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 24.043 Fragment ion/precursor ion ratio 10 8 6 4 2 0.8 0 0.6 0.4 0.2 23.80 23.85 23.90 23.95 24.00 24.05 24.10 24.15 24.20 24.25 24.30 24.35 Acquisition time (min) B 0.1 0.08 Figure 3. 23.80 23.85 23.90 23.95 24.00 24.05 24.10 24.15 24.20 24.25 24.30 24.35 Acquisition time (min) Overlaid EICs for bupirimate in carrot with the reference ion in grey and qualifying ions in other colors. All of the ions have the same chromatographic apex (A) and calculated coelution plot (B). All fragment ions exhibit ratios of approximately 1 across the middle of the reference peak, indicating strong coelution. This provided confirmation for the identification of bupirimate in the sample. 4

time, as well as the target, or reference values, and the identification scores. The flexibility of the All Ions tool settings enables the user to fine-tune it for the specific application by selecting the desired number of qualifying ions, coelution score, mass extraction window, and other parameters (Figure 2). In addition, RT locking ensures precise identification of the pesticides. Viewing compound details Users can easily inspect the results quickly, scrolling through all compounds in the screen and efficiently viewing overlaid EICs, the coelution plot, spectrum results for each compound, and the identification parameters, in this case for bupirimate (Figure 4). The identification parameters panel is shown by itself in Figure 5, displaying the measured mass and retention Figure 4. Agilent All Ions tool results overview for pesticides spiked into carrot extract. Figure 5. Compound Identification Results pane from the Agilent All Ions tool. 5

As a validation study, 56 pesticides were spiked into five different matrices (apple, carrot, tomato, leek, and orange) at increasing concentration levels (Table 2). Most of the compounds were found at the lowest spiked level of 10 µg/ml (parts per billion, (ppb)) in all matrices, and their presence was verified by at least two additional fragment ions (as indicated by dark green cells) and their retention times. Quantitative analysis For unexpected compounds that are found, subsequent quantitative analysis may be considered necessary, and this can be set up simply by exporting the qualitative data to MassHunter Quantitative Analysis Software, using a Compound Exchange Format (CEF) file. Table 2. Compound Screening Results in Four Matrices (10 to 200 ppb) Tomato Carrot Apple Orange Leek Compound Dichlorvos Biphenyl Phenylphenol 2- Chlorpropham Trifluralin HCH alpha HCB HCH beta Propazine HCH gamma (lindane) Terbuthylazine Pyrimethanil Diazinon Pirimicarb Chlorpyrifos-methyl Parathion-methyl Vinclozolin Tolclofos-methyl Metalaxyl Fenpropidin Fenitrothion Chlorpyrifos Fenpropimorph Pendimethalin Fipronil Procymidone Endosulfan alpha Dieldrin DDE p,p - Myclobutanil Bupirimate Kresoxim-methyl Endosulfan beta Chlorobenzilate DDD p,p - DDT o,p - Oxadixyl 10 50 100 200 10 50 100 200 10 50 100 200 10 50 100 200 10 50 100 200 White = not found; dark green = Found; light green = Found, qualifier used for quantitation 6

Table 2. Compound Screening Results in Four Matrices (10 to 200 ppb)(continued) Tomato Carrot Apple Orange Leek Compound Endosulfan sulfate DDT p,p - TPP Iprodione Tetramethrin I Bromopropylate Tetramethrin II Bifenthrin g-cyhalothrin Acrinathrin Bitertanol Cypermethrin I Cypermethrin II Etofenprox Esfenvalerate (SS,RR) Azoxystrobin 10 50 100 200 10 50 100 200 10 50 100 200 10 50 100 200 10 50 100 200 White = not found; dark green = Found; light green = Found, qualifier used for quantitation The CEF file contains information necessary to set up a quantitative method: compound name, retention time, reference ion, fragment ions (to create qualifiers), and relative abundances. The MassHunter Quantitative Analysis Software automatically selects the reference and qualifier ions, saving tedious manual processing. After the method has been set up, suspect samples can be run to acquire quantitative results. Turnkey automation allows MassHunter to both acquire and quantitate data, and provides a report for the targeted compounds. MassHunter has long since provided a popular environment for reviewing quantitative results, and these software tools are available for GC/Q-TOF as well. In this case, Quantitative Analysis software allows viewing of quantifier and qualifier ions, but with an added level of confidence in the results provided by the scoring of the quality of identifications with accurate mass metrics. Figure 6 shows extracted ion chromatograms of bupirimate with two qualifier ions and their ratios plotted against the quantifier ion. During data processing, the MassHunter Quantitative Analysis software automatically flags qualifier ratios that are outside of user-specified limits. 10 4 A 2.4 2.2 2.0 1.8 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0 Counts 24.031 min. Relative abundance (%) 10 2 B 1.2 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 208.1458, 316.1564, 273.1016 Ratio = 30.6 (111.4 %) Ratio = 93.1 (100.3 %) Figure 6. 23.6 23.7 23.8 23.9 24.0 24.1 24.2 24.3 24.4 Acquisition time (min) 23.6 23.7 23.8 23.9 24.0 24.1 24.2 24.3 24.4 Acquisition time (min) EICs of the quantifier (target) ion (A), as well as qualifier ions and the expected qualifier to quantifier ion ratio (B) for bupirimate in apple at 10 µg/ml (ppb). 7

The necessary calibration range for most compounds is usually from 10 to 200 ng/ml. When quantitating many compounds in one run, it is likely that some compounds will produce much higher responses than others. For such compounds, this can lead to saturation of the highest calibration standard of 200 µg/ml. Unifying response factors is done easily when using GC/Q-TOF, because EI often offers a range of candidate ions from which to choose. The resolution of the mass spectrometer also offers the use of carbon 13 isotope ions. The user can simply choose the ion that is optimal for the calibration range. This can even be changed retrospectively due to the untargeted nature of acquisition with a GC/Q-TOF. Retrospective analysis is exactly what is required when curating a quantitative method for the first time using spikes and standards. This was also part of the objective for this study, and the results will be shown in a separate report. For now, those compounds whose quantitation benefitted from adjustment away from the dominant ions are shown in Table 2 (light green cells). Conclusions The Agilent 7200 Series GC/Q-TOF, in combination with Agilent MassHunter Qualitative Analysis Software and the GC/Q-TOF Pesticide PCDL, can be used effectively to screen for pesticide residues in a variety of matrices at concentrations as low as 10 ppb. Accurate identification is assured by use of the unique Agilent All Ions tool. The advantages of using GC/Q-TOF include increased confidence in compound confirmation provided by accurate mass-high resolution data, the ability to perform retrospective analysis (particularly for unexpected peaks), and the ability to seamlessly go from qualitative to quantitative analysis. These results are encouraging, because as new compounds appear on a laboratory s radar, not only can data collected in the past be re-interrogated, but a means is also available to create and expand optimized quantitative methods for the future. References 1. SANCO/12571/2013. Guidance document on analytical quality control and validation procedures for pesticide residues analysis in food and feed, 2013. 2. V. Giarrocco, B. Quimby, M. Klee, Retention Time Locking: Concepts and Applications, Agilent Technologies Application Note, publication number 5966-2469EN, December 1997. 3. H. Prest, Capillary Flow Technology for GC/MS: A Simple Tee Configuration for Analysis at Trace Concentrations with Rapid Backflushing for Matrix Elimination, Agilent Technologies Application Note, publication number 5979-8664EN, June 2008. 4. L. Rajski, A. Lozano, N. Belmonte-Valles, A. Uclés, S. Uclés, M. Mezcua, and A. R. Fernandez-Alba. Comparison of three multiresidue methods to analyse pesticides in green tea with liquid and gas chromatography/tandem mass spectrometry Analyst, 138, 921-931(2013). 5. H. P. Sievert, A. C. J. H. Drouen. Spectral matching and peak purity in Diode Array Detection in High Performance Liquid Chromatography, New York, Marcel Dekker, 51, (1993). For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2015 Printed in the USA May 8, 2015 5991-5894EN