Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

Similar documents
TOMAHAQ Method Construction

Self-assembling covalent organic frameworks functionalized. magnetic graphene hydrophilic biocomposite as an ultrasensitive

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Methods for proteome analysis of obesity (Adipose tissue)

MS-based proteomics to investigate proteins and their modifications

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Tandem mass spectra were extracted from the Xcalibur data system format. (.RAW) and charge state assignment was performed using in house software

Increasing the Multiplexing of Protein Quantitation from 6- to 10-Plex with Reporter Ion Isotopologues

Optimization and Use of Peptide Mass Measurement Accuracy in Shotgun Proteomics* S

Relative Quantitation of TMT-Labeled Proteomes Focus on Sensitivity and Precision

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

Figure S1. Interaction of PcTS with αsyn. (a) 1 H- 15 N HSQC NMR spectra of 100 µm αsyn in the absence (0:1, black) and increasing equivalent

Workflow concept. Data goes through the workflow. A Node contains an operation An edge represents data flow The results are brought together in tables

Tutorial 1: Setting up your Skyline document

Supporting Information

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

DIA-Umpire: comprehensive computational framework for data independent acquisition proteomics

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

Quantitation of TMT-Labeled Peptides Using Higher-Energy Collisional Dissociation on the Velos Pro Ion Trap Mass Spectrometer

Workshop: SILAC and Alternative Labeling Strategies in Quantitative Proteomics

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

HOWTO, example workflow and data files. (Version )

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

SRM assay generation and data analysis in Skyline

Spectrum-to-Spectrum Searching Using a. Proteome-wide Spectral Library

TANDEM MASS SPECTROSCOPY

SILAC and TMT. IDeA National Resource for Proteomics Workshop for Graduate Students and Post-docs Renny Lan 5/18/2017

Purdue-UAB Botanicals Center for Age- Related Disease

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

Analyst Software. Peptide and Protein Quantitation Tutorial

Q Exactive TM : A True Qual-Quan HR/AM Mass Spectrometer for Routine Proteomics Applications. Yi Zhang, Ph.D. ThermoFisher Scientific

Advanced Fragmentation Techniques for BioPharma Characterization

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

LogViewer: A Software Tool to Visualize Quality Control Parameters to Optimize Proteomics Experiments using Orbitrap and LTQ-FT Mass Spectrometers

A Quadrupole-Orbitrap Hybrid Mass Spectrometer Offers Highest Benchtop Performance for In-Depth Analysis of Complex Proteomes

The Agilent 6495 Triple Quadrupole LC/MS: Peptide Quantitation Performance

Application Note. Edgar Naegele. Abstract

A Description of the CPTAC Common Data Analysis Pipeline (CDAP)

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Breaking the speed limit: Fast, high-resolution peptide and tryptic digest separations using fused-core particles

Integrated enzyme reactor and high resolving chromatography in sub-chip dimensions for. sensitive protein mass spectrometry

Overview - MS Proteomics in One Slide. MS masses of peptides. MS/MS fragments of a peptide. Results! Match to sequence database

HR/AM Targeted Peptide Quantitation on a Q Exactive MS: A Unique Combination of High Selectivity, Sensitivity and Throughput

Effective Strategies for Improving Peptide Identification with Tandem Mass Spectrometry

Monitoring Protein PEGylation with Ion Exchange Chromatography

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

Effective desalting and concentration of in-gel digest samples with Vivapure C18 Micro spin columns prior to MALDI-TOF analysis.

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

A TMT-labeled Spectral Library for Peptide Sequencing

SUPPORTING INFORMATION Antimicrobial activity of chlorinated amino acids and peptides

Supplemental Information

Optimization and Use of Peptide Mass Measurement Accuracy in Shotgun Proteomics

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Isotopic-Labeling and Mass Spectrometry-Based Quantitative Proteomics

LECTURE-13. Peptide Mass Fingerprinting HANDOUT. Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of

The Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Protein Reduction, Alkylation, Digestion

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Translational Biomarker Core

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

Metal-catalyzed oxidation of Aβ and the resulting reorganization of the Cu binding sites promote ROS production. Supporting Information

Backbone modification of a parathyroid hormone receptor-1 antagonist/inverse agonist

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Computational Methods for Mass Spectrometry Proteomics

PeptideProphet: Validation of Peptide Assignments to MS/MS Spectra. Andrew Keller

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Thermo Finnigan LTQ. Specifications

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Structure of the α-helix

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Protein analysis using mass spectrometry

Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry

Protocol. Product Use & Liability. Contact us: InfoLine: Order per fax: www:

At-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + =

Choosing the metabolomics platform

Yifei Bao. Beatrix. Manor Askenazi

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Tandem MS = MS / MS. ESI-MS give information on the mass of a molecule but none on the structure

Information Dependent Acquisition (IDA) 1

Supporting Information

Protocol. Product Use & Liability. Contact us: InfoLine: Order per fax: www:

TUTORIAL EXERCISES WITH ANSWERS

High-Field Orbitrap Creating new possibilities

Supporting Information

Finnigan LCQ Advantage MAX

OMX-S basic Kit Instruction Manual March/09

OMX-S pro Instruction Manual* Oct/07

Transcription:

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 * 1 Department of Chemistry, Pomona College, Claremont, California 91711, United States 2 Proteome Exploration Laboratory, Division of Biology and Biological Engineering, Beckman Institute, California Institute of Technology, Pasadena, California 91125, United States *Corresponding Author Tel: (626) 395-2339, shess@caltech.edu *Corresponding Author Tel: (909) 607-8832, jane.liu@pomona.edu Supporting Information Extended Methods Figure S1 Figure S2 Figure S3 Figure S4 Table S1 Table S2 Table S3 S- 1

Extended Methods Sample Preparation. We used six- plex TMT reagents to build a model peptide mixture sample from protein digests of Saccharomyces cerevisiae. The six- plex TMT system allows for quantification in up to six reporter channels through reporter ions at m/z 126, 127, 128, 129, 130 and 131. For each of the six reporter channels, 4 µg of tryptic- digested yeast peptides were resuspended in 24 µl of 100 mm triethyl ammonium bicarbonate (TEAB); the ph of the solution was confirmed to be between 8.5-9.0. To this mixture was added 6.6 µl of the TMT reagent, dissolved in acetonitrile (41 µl for 0.8 mg reagent). The reaction was incubated at room temperature for 1 h, protected from light. The reaction was quenched with 1.3 µl 5% hydroxylamine (in 100 mm TEAB), for 15 min, at room temperature, protected from light. The total volume of each reaction was brought to 40 µl with 100 mm TEAB, such that each sample had a peptide concentration of 100 ng/µl. The reactions were mixed to give a 10:2:1:1:2:10 mass ratio: 25 µl (2.5 µg) 126- labeled peptides, 5 µl (0.5 µg) 127- labeled peptides, 2.5 µl (0.25 µg) 128- labeled peptides, 2.5 µl (0.25 µg) 129- labeled peptides, 5 µl (0.5 µg) 130- labeled peptides, and 25 µl (2.5 µg) 131- labeled peptides. The combined mixture of TMT- labeled peptides was concentrated in vacuo and then resuspended in 0.2% formic acid. The peptide mixture was then desalted by HPLC using a C18 Macrotrap (Michrom Bioresources) (buffer A: 0.2% formic acid in H2O; buffer B: 0.2% formic acid in acetonitrile), followed by concentration in vacuo. Based on the UV chromatogram, >99% of starting peptide was retained through the labeling and chromatography process. Full description of the LC- MS methods All LC- MS experiments were performed on an Orbitrap Fusion mass spectrometer (Thermo Fisher Scientific) coupled to an ancillary EASY- nlc II nano- HPLC (Thermo Fisher Scientific) equipped with a Nanospray Flex ion source (Thermo Fisher Scientific). Samples were directly loaded onto a 20- cm, 50 µm I.D. Picrofrit microcapillary column (New Objective), packed in- house with ReproSil- Pur C18AQ 1.9 µm resin (120 Å, Dr. Maisch GmbH, Ammerbuch, Germany). Separation was achieved by applying a 1.6-32% ACN gradient in 0.2% formic acid over 120 minutes at 200 nl/min. Specifically, the gradient was as follows: 2-6% solvent B (7.5 min), 6-25% solvent B (82.5 min), 25-40% solvent B (30 min), 40-100% solvent B (1 min) and 100% B (9 min). Solvent A was 97.8% H2O, 2% acetonitrile, 0.2% formic acid; solvent B was 19.8% H2O, 80% acetonitrile, and 0.2% formic acid. The column was enclosed in a column heater operating at 75 C. Electrospray ionization was enabled by applying a voltage of 2.5 kv through the coated tip. S- 2

The Orbitrap Fusion was operated in data- dependent acquisition mode for both the MS 3 - IT and MS 3 - OT methods. The FTMS 1 and ITMS 2 settings were the same for both methods. For the FTMS 1 spectra, a survey scan of 400-1500 m/z was collected in the Orbitrap at a resolution of 120,000, an AGC target of 2x10 5, and a max injection time of 50 ms. The 10 most intense ions were selected for MS analysis. Precursor ions were filtered according to charge state (2-7), dynamic exclusion (70 s with a ± 10 ppm window), and monoisotopic precursor selection. MS 1 precursors were first interrogated by ITMS 2 using CID (35% collision energy, rapid scan rate, scan range of 400-2000 m/z). Precursors were isolated using a 0.5 m/z isolation window. They were accumulated to an AGC target of 4,000 or a max injection time of 150 ms. The intensity threshold was 1x10 3. For the SPS- MS 3 scan in the Orbitrap, parameters included an AGC target of 5x10 4, max injection time of 250 ms, HCD normalized collision energy of 55, quadrupole isolation width of 2 m/z, SPS selection of the 10 most intense ITMS 2 fragment ions, an Orbitrap resolution of 60,000, and a scan range of 100-500 m/z. For the SPS- MS 3 scan in the Ion Trap, parameters included an AGC target of 1x10 3-1x10 4, max injection time of 250 ms, HCD normalized collision energy of 55, quadrupole isolation width of 2 m/z, SPS selection of the 10 most intense ITMS 2 fragment ions, rapid scan mode, and a scan range of 100-500 m/z. Data Analysis. Thermo raw files were converted to MGF files using ReAdW4Mascot2 (v. 20090305a). MS 2 spectra were searched against the SGD yeast database (5,911 sequences) and a contaminant database (259 sequences) using MSGF+ (v. 20140630). Precursor tolerance was 8 ppm. CID was specified as the fragmentation method and low- res LCQ/LTQ was specified as the MS 2 mass analyzer. Trypsin was specified as the digestion enzyme. Number of tolerable tryptic termini was 2. Fixed modifications included carboxyamidomethyl of cysteine and TMT6plex modification of peptide N- termini and lysine. Variable modifications included oxidation of methionine and protein N- terminal acetylation. A decoy search was included and PSM and peptide false discovery rates were both less than 1%. MS 3 reporter ion mass tolerances were 0.5 Da for MS 3 - IonTrap and 0.02 Da for MS 3 - Orbitrap. Reporter ion intensities were corrected for isotope impurities. In- house scripts were used to pair the MS 2 identifications with the MS 3 spectra and to perform protein inference from the peptides identified. S- 3

Figure S1. Summary of data for each of the three AGC targets tested for the MS 3 - HCD scan when using the linear ion trap for the MS 3 scan. The number of peptide- spectrum matches (PSM) and the number of unique, identified peptides (Peptides) that were quantified in all six channels are shown. S- 4

Figure S2. Accuracy and precision of PSM quantification when collecting the MS 3 - HCD scans using the linear ion trap with three different AGC targets. Boxplots indicate the median (middle line), the interquartile range (box) and the 5 th and 95 th percentile (whiskers) for each experiment. Dashed lines indicate the values for the idealized fractional intensities. S- 5

Figure S3. Number of peptide- to- spectrum matches (PSM) that had reporter ions quantified in the indicated amount of reporter channels. For the MS 3 scans collected in the linear ion trap, 95% of the PSM were quantified in all six channels; for the MS 3 - Orbitrap scans, 70% of the PSM were quantified in all six channels. For normalization and quantitative downstream data analysis for all samples, it is necessary to have reporter ion intensities observed in all six reporter channels. Thus, we focused on only those PSM that were quantified in all six channels. Error bars represent the standard deviation of five technical replicates. S- 6

Figure S4. Accuracy and precision of protein quantification when collecting the MS 3 - HCD scan using the linear ion trap (dark blue, left) or the Orbitrap (light blue, right). The proteins analyzed here were all identified by only 1 peptide that was quantified in all six reporter channels; see the main text for a similar analysis for proteins identified by >1 peptide. Box plots indicate the median (middle line), the interquartile range (box) and the 5 th and 95 th percentile for each experiment (which includes five technical replicates). The dashed lines indicate the values for the idealized fractionalized intensities. S- 7

Table S1. Quantification Data when Varying AGC Target Values in the Linear Ion Trap (MS 3 - HCD Scan). AGC Target Median % Error (PSM) Median % Error (Proteins) Proteins a 1.0e3 6.89 5.15 1,632 5.0e3 5.05 4.46 1,395 1.0e4 4.83 4.43 1,335 a Number of proteins that were identified by at least 1 peptide that was quantified in all reporter channels. The data presented here is from one analysis, each, for the three different AGC targets tested. S- 8

Table S2. Number of PSM as a Function of Reporter Channels. Number of Channels PSM a (Mean ± SD) MS 3 - Ion Trap Percent of total PSM PSM (Mean ± SD) MS 3 - Orbitrap Percent of total PSM 0 b 7 ± 4 0.048 143 ± 4 1.4 1 8 ± 3 0.053 143 ± 22 1.4 2 15 ± 3 0.10 246 ± 18 2.3 3 39 ± 5 0.27 288 ± 19 2.7 4 118 ±15 0.81 582 ± 36 5.5 5 521 ±46 3.6 1,790 ± 26 17 6 13,916 ± 226 95 7,299 ± 226 70 Total 14,624 ± 243 10,490 ± 186 a Number of peptide- to- spectrum matches (PSM) that had reporter ion intensities quantified in the corresponding number of channels. The average and standard deviation of five technical replicates are provided. b MS 2 scans that either did not trigger an MS 3 scan or resulted in an MS 3 scan with no quantified reporter ions result in a PSM that was quantified in 0 channels. S- 9

Table S3. Protein Identification Rates. a MS 3 Scan Proteins Identified b Proteins Quantified c Proteins Quantified >1 peptide d Ion trap 2,295 ± 38 2,229 ± 34 1,602 ± 32 Orbitrap 1,726 ± 18 1,342 ± 25 933 ± 17 a The average and standard deviation of five technical replicates are provided. b Number of proteins that were identified by at least 1 peptide. c Number of proteins that were identified by at least 1 peptide that was quantified in all reporter channels. d Number of proteins that were identified by at least 2 peptides that were quantified in all reporter channels. S- 10