SOP-C-120 Determination of ph

Similar documents
Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

QAM-I-116 Preparation of Labware

QAM-I-111 Operation and Calibration of the Conductivity Meter

-fr-[*-] ^'A: Approval: /": f. z /zs- /, e. Determination of Nitrate/Nitrite as Nitrogen. EffectiveDate: 3 it

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

ALLOWAY METHOD OUTLINE

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Laboratory Certification Workshop

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Guide to ph Analysis.

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Document Control No.: QM-AD002 SOP Code: AP09 Issue Date: Revision No.: 04 Page 1 of 6 Title: ph MEASUREMENT

Approximate Volatile Acids by Titration

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Trihalomethane Formation Potential (THMFP)

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

aa + bb cc + dd Equation 1

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Prince George s Community College PL 2: CHARACTERIZATION OF A MONOPROTIC WEAK ACID BY POTENTIOMETRIC TITRATION

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

symphony Multisensor Probes

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Titration of a strong acid with a strong base with Cobra4

Determination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements

STANDARD OPERATING PROCEDURE

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Chemistry Determination of Mixed Acids

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Working Copy. tlrq /rb I 'B-16. Determination of Gross Alpha/Beta Activity. Revision 0. Laboratory Manag erllqao/rso. Effective Date. lnitials.

1. 4 2y 1 2 = x = x 1 2 x + 1 = x x + 1 = x = 6. w = 2. 5 x

Experiment 7 Buffer Capacity & Buffer Preparation

Determining the K sp of Calcium Hydroxide

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Measure - Conductivity STANDARD Last std: Sep 10:17 am. Touch meas to measure sample. Touch std to access standardize mode ATC. 25.

Method Abstract. Flow Solution Fluoride, USGS by ISE and FIA

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

GETTING THE END POINT TO APPROXIMATE. Two hours

SECTION D.2 AMMONIA NITROGEN

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

DR/4000 PROCEDURE. IRON, Total

DR/4000 PROCEDURE NITRATE. 2. The display will show:

By D.B. Radtke, J.K. Kurklin, and F.D. Wilde. Temperature... T Equipment and supplies A Maintenance, cleaning, and storage...

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

Technical Procedure for General Laboratory Equipment

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Pharmaceutical Active Ingredients Group Standard 2017 HSR100425

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Total Sulfide, Methylene Blue Method

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

GENERAL INSTRUCTIONS GENERAL PREPARATION

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Pharmaceutics IV PHR4308 Practicals Handbook

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Acid Deposition and Oxidant Research Center (ADORC)

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

GENERAL INSTRUCTIONS GENERAL PREPARATION

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Chloramine (Mono) and Nitrogen, Free Ammonia

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Preparation of Viscous Solutions. Quantitative wet lab; students work in pairs and individually.

How Do Certain Factors Affect the Rate of a Chemical Reaction?

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Method NumberUSTUR 150

# 12 ph-titration of Strong Acids with Strong Bases

Method to 0.50 mg/l NH 3 N Powder Pillows

STANDARD OPERATING PROCEDURES

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

6. Multiple Reactions

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

To measure ph s in a variety of solutions and mixtures and to account for the results obtained.

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

National standard of People s Republic of China

OPERATING INSTRUCTIONS

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

STANDARD OPERATING PROCEDURES

symphony Reference Electrodes

Method to 0.50 mg/l NH 3 N Powder Pillows

Chlorine, Free and Total, High Range

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

Experiment 3: Acids, Bases, and Buffers

Transcription:

i. Identification of the method a. SM 4500-H + (approved 2011) and EPA 9045C (approved 1997) ii. Applicable matrix or matrices a. Waters or aqueous solutions and suspensions iii. Limits of detection and quantitation a. Normal limits are from 1 to 10 ph units without specialized probes. b. Report to the nearest 0.1 ph Standard Units. iv. Scope and application, including parameters to be analyzed a. Calibration and verification are done near the ph of the solution or sample being measured. v. Summary of the method a. Electrometric ph measurement is determination of the activity of the hydrogen ions by potentiometric measurement using a standard hydrogen electrode and a reference electrode. vi. Definitions a. ph- indicator of acidity or basicity, defined as the negative logarithm of the hydrogen ion concentration in an aqueous solution b. Standard Units (S.U.)- an assigned numeric value of ph c. ATC (Automatic Temperature Compensation)- a feature found on most ph meters which corrects for temperature differences between calibration standards and samples measured. vii. Interferences a. The glass electrode is relatively free from interference from color, turbidity, colloidal matter, oxidants, reductants, or high salinity, except for a sodium error at ph 10. Reduce this error by using special low sodium error electrodes. b. Extreme temperatures will interfere with ph measurement, even using ATC. c. Oils and greases can foul the probe. d. ph measurements are affected by temperature in two ways: mechanical effects that are caused by changes in the properties of the electrodes and chemical effects caused by equilibrium changes. In the first instance, the slope increases with increasing temperature and electrodes take time to achieve thermal equilibrium. This can cause long-term drift in ph. Because chemical equilibrium affects ph, standard ph Rev. 9 Page 2 of 6 TIAER

buffers have a specified ph at indicated temperatures. Always report temperature at which ph is measured. e. Combination electrodes incorporate the glass and reference electrodes into a single probe. Use a low sodium error electrode that can operate at high temperatures for measuring ph over 10 because standard glass electrodes yield erroneously low values. For measuring ph below 1 standard glass electrodes yield erroneously high values; use liquid membrane electrodes instead. viii. Safety a. Refer to QAM-S-101 for general laboratory safety. ix. Equipment and supplies a. Accumet AB 15 Plus ph/millivolt Meter, or equivalent. b. Accumet ph Electrode with Ag/AgCl reference system, combination probe, or other equivalent electrode. c. Magnetic stirrer and Teflon-coated stirring bar. d. Thermometer or temperature sensor for automatic compensation e. Beakers: Preferably use polyethylene or TFE beakers. f. Flow chamber: Use for continuous flow measurements or for poorly buffered solutions. x. Reagents and standards a. Reagents i. Deionized water (DI) for rinsing b. Standards (refresh daily from the stock) i. ph 7 buffer solution, commercially prepared ii. ph 10 buffer solution, commercially prepared iii. ph 4 buffer solution, commercially prepared iv. other levels of buffer solutions as appropriate for the samples tested xi. Sample collection, preservation, shipment and storage a. Holding Time: as soon as possible; ph is best measured in the field. If not possible, ph is measured as soon as practicable in the laboratory and data are designated as Laboratory ph with temperature, time and date labels. Laboratory ph is measured within 15 minutes of sampling, or data is flagged as questionable. b. Refrigerate sample to >0-6 C if not measured immediately. c. Samples are not collected or delivered by the laboratory. Rev. 9 Page 3 of 6 TIAER

xii. Quality control a. Refer to QAM-Q-101 for quality control requirements. xiii. Calibration and standardization a. Calibration for this instrument consists of two-buffer set points with one other buffer calibration check in accordance with QAM-I-105. The range of buffer standards generally bracket the ph of the samples analyzed. b. Before initial analysis of the day, perform a two-buffer set point calibration using two of ph 4.0, 7.0, or 10.0 buffer solutions in accordance with QAM-I-105 Operation and Calibration of the ph Meter. The ph 10.0 & 7.0 buffers are for routine waters, solutions and suspensions. When analyzing wastes or samples with extreme higher or lower ph, use a more appropriate buffer closer to the expected ph range. Other buffers may be used to achieve desired range of measurement. c. Perform a calibration verification on at least one other ph buffer, usually 7.0 ph. A ICV/CCV is of a different source than the calibration buffers, but the LCS may be one of the calibration buffers. If the observed readings are within ± 0.1 ph units, acceptance criteria have been met. xiv. Procedure a. Prior to analysis, refrigerated samples are warmed to about the same temperature as the standard buffer solutions. If ATC is not used, bring samples to within ±1 C of calibration standards used. b. For soils, sediments and other solids, make an aqueous suspension by mixing the sample with DI water at a 1:1 ratio. The mixture is continuously stirred for 5 minutes, and then allowed to stand for about 1 hour prior to determination of ph in the aqueous phase as described below. c. Place sample in a clean glass beaker or other suitable container and mix the sample with stir bar or rod. d. Verify there is sufficient sample volume to cover sensing elements of the electrodes and to give adequate clearance for the stir bar. e. Rinse electrode with deionized water and ensure air port is open (if present), then rinse at least 3 times with the sample. f. While stirring, place electrode in sample. g. Press stdby button to put meter into active mode. Rev. 9 Page 4 of 6 TIAER

xv. h. After reading stabilizes, read the ph value on the display. i. Record ph to the nearest 0.1 unit in the log. j. Rinse electrodes with deionized water between measurements. For less clean samples, also rinse electrode with a small amount of the next sample to be measured before placing into fresh sample. k. Perform ICV/CCV/LCS/LCSD checks by measuring one standard at ph 7. ESDMS or current LIMS may calculate acceptability of the check. l. Press stdby button to return meter to Standby mode. m. Close air port on electrode (if present). n. Store probe in ph 4 buffer. Data analysis and calculations a. None. When using ATC, report ph adjusted to 25ºC, otherwise report the temperature at which the ph was measured. b. Report the time/date of both sample collection and ph measurement to show it was done within 15 minutes. xvi. Method performance a. Method performance, data assessment and acceptance, corrective action: refer to QAM-Q-101, "Laboratory Quality Control". Any ph data measured in the laboratory is designated as such. Per 40 CFR 136, ph is measured immediately on samples in the field. xvii. Pollution prevention a. Refer to QAM-W-101 for general waste management and pollution control requirements. xviii. Data assessment and acceptance criteria for quality control measures a. CCV, LCS, LCSD measurements are acceptable if they are within 0.1 S.U. of expected values. b. Refer to QAM-Q-101 or project plans for other general quality control requirements. xix. Corrective actions for out-of-control data a. Refer to QAM-Q-105 for corrective action implementation. xx. Contingencies for handling out-of-control or unacceptable data a. Refer to QAM-Q-105 and QAM-Q-101 for handling out-ofcontrol or unacceptable data. xxi. Waste management Rev. 9 Page 5 of 6 TIAER

a. Refer to QAM-W-101 for general waste management and pollution control requirements. No hazardous waste is generated by this procedure, unless the sample is hazardous initially. xxii. References a. Accumet AB15 Plus ph/millivolt Meter Instruction Manual, Accumet. b. Accumet ph/atc Combination Electrode Instructions, Accumet, 1994. c. National Environmental Laboratory Accreditation Conference (TNI Standard), The NELAC Institute, 2009. d. Standard Methods for the Examination of Water and Wastewater, latest online edition, ed. by Arnold E. Greenberg, et al., APHA, AWWA, Washington D.C., Method 4500 H+ (approved 2011). e. Test Methods for Evaluating Solid Waste, Physical/Chemical Methods, SW-846, Version 2, Method 9045C, EPA, Cincinnati, Ohio, 1997. f. Guidelines Establishing Test Procedures for the Analysis of Pollutants, 40 CFR 136.3, Title 40, Volume 20, Code of Federal Regulations, U.S. Government Printing Office, July 1, 2012. xxiii. Any tables, diagrams, flowcharts and validation data a. none Rev. 9 Page 6 of 6 TIAER