Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods. Abstract

Similar documents
EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

ETNA (Efficiency Transfer for Nuclide Activity measurement)

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

Basic hands-on gamma calibration for low activity environmental levels

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

Inaccuracy Determination in Mathematical Model of Labsocs Efficiency Calibration Program

TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS

Gamma-Spectrum Generator

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron Activation System using 252 Cf Source

Evaluation of Nondestructive Assay Characterization Methods for Pipe- Over-Pack Containers

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES

Indian J.Sci.Res.2(3) :25-32, 2011

ETNA (Efficiency Transfer for Nuclide Activity measurement)

Digital simulation of neutron and gamma measurement devices

Determination of the activity of radionuclides

arxiv:nucl-ex/ v2 21 Jul 2005

EFFICIENCY CALIBRATION STUDIES FOR GAMMA SPECTROMETRIC SYSTEMS: THE INFLUENCE OF DIFFERENT PARAMETERS

Researchers at the University of Missouri-Columbia have designed a triple crystal

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

240NU215 - Monte Carlo Simulation of Radiation Transport

Calibration of A Small Anode Germanium Well Detector

Airo International Research Journal October, 2015 Volume VI, ISSN:

The Use of Self-Induced XRF to Quantify the Pu Content in PWR Spent Nuclear Fuel

MCNP6 simulation validation of fast neutron coincidence detection system for nuclear security and safeguards applications

NGN PhD Studentship Proposal

Library of uranium and plutonium reference spectra Introduction

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Compton suppression spectrometry

Simulation of 4π HPGe Compton-Suppression spectrometer

MCNP analysis and optimization of a double crystal phoswich detector

ITU-Nuclear Safeguards and Security Unit

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

Slides by: Prof. Abeer Alharbi

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

1 Introduction. 2 Method. Robert Metzger 1,*, Kenneth Van Riper 2, and George Lasche 3

Nucleonica Lab : An Overview of the Nuclear Science Portal with hands-on exercises

Response Function of the BGO and NaI(Tl) Detectors Using Monte Carlo Simulations

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

Characterization of Large Structures & Components

Comparison of Several Detector Technologies for Measurement of Special Nuclear Materials i

Characterization of a Portable Neutron Coincidence Counter Angela Thornton and W. Charlton Texas A&M University College Station, TX

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

THE RESPONSE FUNCTION OF ISOCART (ORTEC) GAMMA-RAY SPECTROMETRY SYSTEM USING MONTE CARLO SIMULATIONS *

Radiation Detection for the Beta- Delayed Alpha and Gamma Decay of 20 Na. Ellen Simmons

Journal of Radiation Protection and Research

in Cross-Section Data

University of Montenegro Centre for Nuclear Competence and Knowledge Management Podgorica,, Montenegro

Radiation Detection and Measurement

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

General Overview of Radiation Detection and Equipment

Application of Spectral Summing to Indeterminate Suspect Low-Level Drums at Los Alamos National Laboratory

Monte Carlo modelling of a NaI(Tl) scintillator detectors using MCNP simulation code

Preparation of Standard Source as a Petri Dish for Plant by Using

VERIFICATION OF MONTE CARLO CALCULATIONS OF THE NEUTRON FLUX IN THE CAROUSEL CHANNELS OF THE TRIGA MARK II REACTOR, LJUBLJANA

EXPERIMENTAL DETERMINATION OF THE URANIUM ENRICHMENT RATIO

A New MCNPX PTRAC Coincidence Capture File Capability: A Tool for Neutron Detector Design

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

PRACTICAL GAMMA SPECTROSCOPY ASSAY TECHNIQUES FOR LARGE VOLUME LOW-LEVEL WASTE BOXES

The IAEA-CU world wide open proficiency test on the determination of radionuclides in soil, spinach and water

Upcoming features in Serpent photon transport mode

AN ALTERNATIVE PERSPECTIVE ON THE WEIGHTED POINT MODEL FOR PASSIVE NEUTRON MULTIPLICITY COUNTING

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

Sample Spectroscopy System Hardware

Chapiter VII: Ionization chamber

MC simulation of a PGNAA system for on-line cement analysis

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

A Two-Dimensional Point-Kernel Model for Dose Calculations in a Glovebox Array. D. E. Kornreich and D. E. Dooley

Shielding of Ionising Radiation with the Dosimetry & Shielding Module

Detection and measurement of gamma-radiation by gammaspectroscopy

Training Course / Nucleonica Overview

Characterization of semiconductor detectors for γ-ray and x-ray spectrometry

Gamma ray coincidence and angular correlation

Quality Assurance. Purity control. Polycrystalline Ingots

Acoustics and Ionising Radiation Formulation and Strategy. 13 November 2008 Alan DuSautoy

High Precision Nondestructive Assay to Complement DA. H.O. Menlove, M.T. Swinhoe, and J.B. Marlow Los Alamos National Laboratory

4- Locate the channel number of the peak centroid with the software cursor and note the corresponding energy. Record these values.

III. Energy Deposition in the Detector and Spectrum Formation

"Neutron Flux Distribution"

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

K 40 activity and Detector Efficiency

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Semiconductor Detector

PhD Qualifying Exam Nuclear Engineering Program. Part 1 Core Courses

Detection of 106 Ru, via the decay of its daughter 106 Rh, in gammaray

Experimental Determination of Cross Sections for (n,x) Nuclear Reactions

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Compton Camera. Compton Camera

7 th International Summer School 2015, JRC Ispra: Operational Issues in Radioactive Waste Management and Nuclear Decommissioning

ANGLE 4. Advanced Gamma Spectroscopy Efficiency Calibration Software

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

A Germanium Detector with Optimized Compton Veto for High Sensitivity at Low Energy

The Efficiency Characterization of Germanium Detectors at Energies less than 45 kev

Simulating Gamma-Ray Telescopes in Space Radiation Environments with Geant4: Detector Activation

Transcription:

Detection efficiency of a BEGe detector using the Monte Carlo method and a comparison to other calibration methods N. Stefanakis 1 1 GMA Gamma measurements and analyses e.k. PO Box 1611, 72706 Reutlingen, Germany arxiv:1411.4225v1 [physics.ins-det] 16 Nov 2014 (Dated: November 18, 2014) Abstract In this paper we model by using the Monte Carlo simulation code PENELOPE [1, 2] a Broad Energy Germanium (BEGe) detector and determine its efficiency. The simulated geometry consists of a point source located close to the detector as well as volume sources with cylindrical geometry. A comparison of the simulation is made to experimental results as well as to analytical calculations. PACS numbers: Keywords: Typeset by REVTEX 1

I. INTRODUCTION Gamma-ray spectroscopy is one of the most used methods for the characterisation of nuclear waste, because of its non destructive character without the need of sample preparation. Usually the necessary efficiency calibration is done experimentally using standard radioactive sources of the same geometry, density and chemical composition as the sample to be measured. This procedure is not very effective. One effective procedure to overcome these difficulties is the use of Monte Carlo simulation. It can be defined as a methodology for obtaining estimates of the solution of mathematical problems by means of random numbers. In the efficiency calculation the history of the primary particle namely the photon and of the secondary particles resulting from the interaction in the detector is tracked until all its energy is dissipated. Monte Carlo method allows the peak efficiencies to be calculated taking into account the characteristics of the detector and the measurement geometry. For example, one generalist Monte Carlo simulation code is the MCNP[3] used to simulate neutrons, electrons, photons interactions. The user creates an input file that is subsequently read by MCNP. This file contains informationabout the problem in areas such as: the geometry specification, the description of materials and selection of cross-sections evaluations, the location and characteristics of the neutron, photon, or electron source, the type of answers or tallies desired, and any variance reduction techniques used to improve efficiency. One user friendly approach based on Monte Carlo analysis is the In Situ Object Counting System (ISOCS) from the company CANBERRA [4]. In that case the measurement parameters are entered in a geometry editor called Geometry Composer using mathematical templates for most of the commonly encountered geometries. Then the efficiency of the measurement is calculated. This approach supposes that the detector is characterised; A procedure which is time- an cost-expensive. The user is also limited to the available geometrical templates. In this paper we apply the Monte Carlo code PENELOPE, developed to simulate the interaction of photons, electrons and positrons with matter, in order to calculate the efficiency calibration of a BEGe detector. On PENELOPE the existing geometry package permits the generation of random electron-photon showers in material systems consisting of homogeneous bodies limited by quadric surfaces, i.e., planes, spheres, cylinders, etc. [1]. 2

Detector Parameters Dimension (mm) Ge crystal diameter 70 Ge crystal height 25 Al holder 1 Vacuum layer 3 Al case 2 Al Filter 4 Cd Filter 0.5 TABLE I: Detector parameters The simulated geometry considered in this paper is a Plutonium point source located close to the detector. Other more realistic geometries e.g. the cylindrical geometry, as well as comparison to analytical calculations are presented. II. RESULTS AND DISCUSSION A. Point source 1. Measurement geometry The detector was simulated as a simple cylinder made of Germanium(Ge) of radius 35mm and height 25mm hold by an aluminium (Al) cylinder of width 1mm. The detector core is wrapped by an aluminium case of width 2mm. The space between the aluminium holder and the aluminium case is vacuum (2 3mm). The high quality vacuum is necessary to minimize the collection of contaminations on the detector surfaces which cause degradation of the performance. It also deans as a thermal shield to prevent the transfer of heat from the outer surfaces to the detector. A material filter made of Aluminium 4mm and Cadmium 0.5mm is placed in frond of the detector. The filter is used in order to attenuate the low energy gamma radiation. In Fig. 1 we plot a two dimensional xz cross section of the detector geometry. The detector parameters have been chosen to simulate a BEGe detector and are seen in Table I. Several types of electrical contacts are used in order to apply the electric field such as Li diffused layer for one contact and an evaporated or ion-implanted metal film for the other. They produce a layer of attenuating material, from which the charge carriers are not 3

collected and therefore called dead layers. For the sake of simplicity the Germanium dead layer was not simulated. The existence of a Germanium dead layer would have decreased the active detector volume and also attenuated the low energy gamma rays. The Pu-239 point source is located 66cm from the detector window. The incident beam photons move along the z-axis (i.e. upwards) and impinge normally on the surface of the detector. 2. simulated spectra Using Monte Carlo code a narrow beam of energy 129 KeV [5 7] corresponding to a gamma ray from Pu-239 enters the detector. The measurement time is 300 sec. 10 6 particles were simulated. We present in Fig. 2 the distribution of the deposited energy in the Germanium crystal P(E) (ev 1 ) as a function of the energy. This plot can be considered equivalent to the simulated spectra. The photopeak at the energy 129KeV is seen as well as the Compton edge. 3. measurement efficiency The count rates or areas of individual peaks in the spectrum are related to the amount of radioactive material deposited in the body by a factor of efficiency. This efficiency depends on several factors including the intrinsic efficiency of the detector, the geometry of the measurement (solid angle, position, shape of the individual to be measured and the radionuclide distribution in the body) and the properties of the nuclide. Experimentally the detection efficiency Eff Eγ at energy E γ is determined as follows: Eff Eγ = N E i Amtf (1) where N Ei is the net area under the full energy peak corresponding to energy photons E γ emitted by a radionuclide with known activity A, f is the emission probability, m is the sample mass and t is the counting time. Usually calibration methods of multienergy gamma ray emmiters like Eu-152 are used. This nuclide has half life of 13.5 years and it covers the energy range from 122 KeV to 1408 KeV. The disadvantages are the Compton edges and the coincident summing corrections associated with each photopeak. As a next step an interpolation is performed in order to obtain the efficiency for every energy value. One of 4

the output of the Monte Carlo code is the probability distribution of the energy deposited in the Germanium crystal. From this distribution the peak efficiency values (total absorption detection efficiency) can be directly calculated. However the number of counts defined as belonging to the measured full-energy peak need not correspond to the number of full energy absorption events calculated by the monte carlo code. In particular the low energy tail of a measured peak may be only partly included in the experimental peak area, but all of these events are included in the calculated peak area. Therefore the experimental results are only qualitatively compared to the simulated ones. The full energy peak efficiency of the detector as a function of the energy is presented in Fig. 3. The calculated values for the efficiency are compared to the values of the efficiency of a BEGe detector using the ISOCS calibration software from Camberra, as well as to the detection efficiency calculated using the internet portal nucleonica for the same measurement conditions (see [8]). The deviation may be attributed to the fact that the detector parameters are not accurately known and therefore a direct comparison is not possible. B. Volume sources 1. Semi empirical method Having calculated or measured the efficiencies for simple geometries e.g. point sources one can use semi-empirical methods in order to derive the efficiency for other complicated geometries e.g. disc or cylinder. Let us consider a cylindrical homogeneous source with its symmetry axis coinciding with the detector axis. The detection efficiency of this structure can easily be calculated numerically. We can consider this source as a multilayer disk source with each layer at a distance x from the detector. The efficiency is calculated as [7]: ǫ(d) = 2 1 d+h R ǫ R 2 P (r,x)rdrdx (2) h d 0 where h,r is the source height and radius respectively and d denotes the distance to the source end nearest to the detector. ǫ P (r,x) is the radial dependence of the point source efficiency (area efficiency) given for example by x 2 ǫ P (r,x) = ǫ P (0,x) (3) x 2 +r 2 5

ǫ P (0,x) is the point efficiency at the detector axis at distance x from the detector. ǫ P (0,x) = 1 2 [1 (1+ R2 D x 2 ) 1/2 3 8 R2R2 D x 4 (1+ R2 D x 2 ) 5/2...] (4) where R D is the detector radius. 2. Numerical method We are going to study the effect of the volume of the source to the simulated spectra as well as in the efficiency. We consider the measuring geometry of Fig. 4. The detector parameters are the same as in section II where a point source was studied. Here the source is cylindrical of height 1cm and radius 1cm, positioned 10cm apart from the frond edge of the detector. A uniform activity concentration was assumed. We consider the Nuclide Plutonium 239. The simulated spectra is presented in Fig. 5. For simplicity only two gamma energies of the isotope having large emission probability were studied. It is seen that the gamma peak at 129KeV is affected from the Compton background created from the peak at 413KeV. We plot in Fig. 6 the efficiency of the peak measurement as a function of the energy. III. CONCLUSIONS We studied numerically the efficiency of a BEGe detector in the point source as well as in the volume source geometry. The results are well compared to experimental measurements. The efficiency calculated with the monte carlo method is comparable to the efficiency obtained by other methods e.g. use of calibration standards and can be directly integrated to the measurement procedure. [1] Salvat, F., Fernández-Varea, J.M., Sempau, J., PENELOPEA code System for Monte Carlo Simulation of Electron and Photon Transport (OECD Nuclear Energy Agency, Issy-les- Moulineaux, France 2003). [2] Lépy, M.C., et al., Intercomparison of efficiency transfer software for gamma-ray spectrometry. Appl. Radiat. Isot. 55, 493 (2001). 6

[3] Briesmeister, J.F., MCNP-A General Monte Carlo N-Particle Transport Code. Report LA- 12625-M Version 4B. Los Alamos National Laboratory, Los Alamos, NM (1997). [4] ISOCS Calibration software, www.canberra.com. [5] D. Reilly, N. Ensslin, and H. Smith, Jr., Passive Nondestructive Assay of Nuclear Materials (PANDA), LA-UR-90-732, 153-157 (1991). [6] Mini table de Radionucléides, Laboratoire National Henri Becquerel, EDP Sciences (2007). [7] Gamma- and x-ray spectrometry with semiconductor detectors, K. Debertin and R.G. Helmer, Elsevier 2001. [8] N. Stefanakis A comparison of a measurement using the Canberra BEGe Detector and a simulation using nucleonica. Poster presentation at Nucleonica Training Course 18-19 April 2013 FTU Karlsruhe. http://www.nucleonica.com/wiki/images/e/ee/stefanakis FTU 2013.pdf 7

FIG. 1: Measurement geometry 0,001 0,0001 P(E) (1/eV) 1e-05 1e-06 1e-07 1e-08 0 50000 1e+05 Energy (ev) FIG. 2: Distribution of energy deposited in the Ge crystal 8

0,001 Efficiency 0,0001 1e-05 1e+05 Energy (ev) FIG. 3: Detection efficiency in the range 0 bis 645KeV FIG. 4: Measurement geometry for the volume source 9

3e-07 2,5e-07 2e-07 P(E) (1/eV) 1,5e-07 1e-07 5e-08 0 0 1e+05 2e+05 3e+05 4e+05 Energy (ev) FIG. 5: Distribution of energy deposited in the Ge crystal 10

2e-05 1,5e-05 Efficiency 1e-05 5e-06 0 1e+05 2e+05 3e+05 4e+05 5e+05 Energy (ev) FIG. 6: Detection efficiency for the volume source geometry 11