Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Similar documents
ICP-OES Application Note Number 35

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Elemental analysis of river sediment using the Agilent 4200 MP-AES

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

OES - Optical Emission Spectrometer 2000

Analysis of domestic sludge using the Agilent 4200 MP-AES

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Analysis Repeatability of Trace and Major Elements in a Water Sample

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Optimizing Analytical Performance in ICP-OES Applications

Samples compliant with the WATER

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

a. An emission line as close as possible to the analyte resonance line

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Unité d Agronomie, Centre de recherches INRA de Bordeaux, B.P. 81, Villenave d Ornon Cedex, France

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

Rapid Sample Assessment and Simplified Method Development with IntelliQuant

Effect of Sampling Depth on the Analyte Response in Laser Ablation Inductively Coupled Plasma Mass Spectrometry

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Single particle analysis using the Agilent 7700x ICP-MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Procedures for the determination of stable elements in construction materials from the nuclear reactors at Risø National Laboratory

Analysis of mineral elements in wines using ICP-AES (inductively coupled plasma / atomic emission spectrometry) (Resolution Oeno 478/2013)

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Test Method: CPSC-CH-E

INVESTIGATION OF ICP-OES ANALYSIS FOR DETERMINATION OF TRACE LEAD IN LEAD-FREE ALLOY

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Trace Analyses in Metal Matrices Using the ELAN DRC II

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Automated determination of the uniformity of dosage in quinine sulfate tablets using a fiber optics autosampler

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

The Measurement of Lead in Food Products by Graphite Furnace AA and SpectraAA-40

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

110 Lowell Rd Hudson, NH USA Tel:

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Performance Evaluation of an Axially Viewed Horizontal Inductively-Coupled Plasma for Optical-Emission Spectrometry

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES)

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

NATURE AN POLUUTION DEGREE OF WATER ASSETS STUDIED BY INDUCTIVELY COUPLED PLASMA-OPTICAL EMISSION SPECTROSCOPY

Evaluation, Control and Diagnosis of an ICP Through Simple Experiments

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

OPTIMIZING METALS SAMPLE PREP

Elemental Scientific. Spike Recovery. Elemental Scientific

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Spectroscopy Division, and * Central Workshop, Bhabha Atomic Research Centre, Trombay, Mumbai , India

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Agilent G2350A Atomic Emission Detector (AED)

Product Descriptions and Offerings

Transcription:

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Tran T. Nham Introduction With the benefits of typically 5 10 fold improvement in detection limits of the axially viewed ICP compared to the conventional radially viewed ICP system [1 5], the applications of axially viewed ICP-OES have been growing rapidly, especially in the area of environmental pollution control. Fast sample throughput is essential for routine water analysis, this study highlights a newly developed axially viewed CCD simultaneous ICP-OES as a high speed analysis tool for the water laboratory.

Experimental Instrumental In this work, a Vista AX (axially viewed plasma) CCD simultaneous ICP-OES and a Vista RL (radially viewed plasma) CCD simultaneous ICP-OES were used (Agilent Technologies, Inc.). The Vista features an echelle polychromator with a cross dispersion prism to form a two dimensional diffraction pattern across its focal plane. It is equipped with a custom designed CCD detector -the patented Vista Chip [6]. The polychromator can be purged with a low flow of either argon or nitrogen gas for the detection of low UV wavelengths. With more than 70,000 pixels, the VistaChip covers wavelengths continuously from 167 to 784 nm. The detector has one million pixels per second read out speed, and the read-out circuitry is duplicated on each side of the chip. The signal processing speed of the VistaChip is 80 times faster than a conventional system [7]. The sample introduction system consists of a glass concentric nebulizer and a glass cyclonic chamber. The nebulizer flow is controlled by a mass flow controller. The operating parameters are shown in Table 1. Table 1. Instrument Operating Conditions Power 1.2 kw Plasma gas flow 15 L/min Auxiliary gas flow 1.5 L/min Nebuliser flow 0.75 L/min Pump rate 15 rpm Integration time 10 s Replicates 3 Background correction Fitted Reagents and Standards All chemicals and reagents used were of high purity analytical reagent grade. Standards and blank for detection limit measurements were prepared in 1% v/v HNO 3. Standards and blank for the water analysis were prepared in 3.5 % v/v HNO 3 to matrix match with the NIST water samples. Results Detection Limits The detection limits of the system were measured as the concentration giving signals equal to three times the standard deviation of a blank solution. They were measured at an integration time of 10 s and 10 replicate readings over three con- secutive days to meet the requirements of the US EPA definition of instrument detection limit (IDL) for example, to test compliance against the Contract Required Detection Limit (CRDL). Table 2 illustrates the detection limit of 27 elements with the Vista axially viewed and radially viewed ICP systems. The current CRDL of 22 elements specified by the US EPA is also given in Table 2. The CRDL numbers in brackets indicate proposed changes to the CRDLs under ILMO 5.0 [9]. Table 2. Detection Limits, Current US EPA CRDLs [8] and Proposed CRDLs in Brackets [9] Wavelength IDL axial IDL radial CRDL [8,9] Element (nm) (µg/l) (µg/l) (µg/l) Ag 328.068 0.5 2 10 (5) Al 396.152 1.5 6 200 As 188.980 3.5 22 10 (5) B 249.678 0.2 0.5 Ba 455.403 0.04 0.3 200 (20) Be 234.861 0.05 0.3 5 (1) Bi 223.061 3 18 Ca 315.887 1.5 8 5000 Cd 214.439 0.3 1.5 5 (2) Co 238.892 0.5 4.5 50 (5) Cr 267.716 0.5 2.2 10 (5) Cu 324.754 0.3 1.8 25 Fe 259.940 0.4 5 100 K 766.491 0.5 6.5 5000 Mg 279.800 2.2 20 5000 Mn 257.610 0.05 0.3 15 Mo 202.032 0.8 5 Na 589.592 0.2 1 5000 Ni 231.604 1.3 10 40 (20) Pb 220.353 3 25 3 Sb 206.834 2.9 28 60 (5) Se 196.026 5 35 5 Si 251.430 3.5 32 Sr 421.552 0.01 0.1 Te 214.282 6 35 Tl 190.794 3 24 10 (5) V 292.401 0.4 3.5 50 (10) Zn 213.857 0.3 1.0 20 (10) The detection limits with the axially viewed ICP are generally an order of magnitude better than those from the radially viewed ICP. All detection limits with the axially viewed ICP are lower than the US EPA CRDL. 2

The detection limit of the radially viewed ICP also meets the majority of the current and proposed CRDL with the exception of As, Tl and Sb. Of course, ICP-OES detection limits can be further enhanced by using alternate sample introduction systems such as vapor generation or ultrasonic nebulization. Warm Up Time One of the factors in improving productivity is to shorten the instrument warm up time. The instrument warm up time is defined as the time taken from ignition of the plasma until the system is stabilised and ready for analysis. Figure 1 demonstrates the warm up time from cold start. As shown in Figure 1, the instrument takes about 35 minutes to stabilize from plasma ignition. While you are waiting for the instrument to warm up, the software allows you to set up a new worksheet, or manipulate the data from the previous runs, or generate a report. The time could also be used for loading samples onto the autosampler. Table 3. Sample Throughput Rate Integration time(s) Total analysis time(s) Samples per hour 1 80 45 3 87 41 5 93 38 10 108 33 20 140 26 30 172 21 With an integration time of 10 s, the sample throughput is about 220 samples per 8 hour day which includes calibration of a blank and two standards and recalibration after every 20 samples. With its automatic data Quality Control Protocols (QCP) software, the Vista can operate unattended 24 hours per day with results tested against acceptance criteria. The instrument need only be stopped for periodic maintenance of the sample introduction system. Figure 1. Instrument warm up time. Short Term Precision and Long Term Stability Short term precision of the system was measured with a integration time of 10s and 30 replicates, and is typically 0.5 %RSD or less. Long term stability was measured by performing a calibration on a multielement standard and a reagent blank, then continuously measuring the standard solution as a sample every 15 minutes without recalibration. Figure 2 shows that the long term stability over 4 hours ranged from 0.21 0.98 %RSD. Ar 404.442 nm, Ba 455.403 nm and Zn 206.200 nm lines were selected together with Mg II 280.270 nm and Mg I 285.213 nm line intensity ratio. These test elements were recommended for drift diagnostics for the ICP. Sample Throughput Rate To demonstrate the speed of analysis of the Vista systems, the sample analysis rate was measured under a range of conditions. The sample analysis rate was measured using 25 s sample delay time, 10 s stabilisation time, 30 s rinsing time between samples and fitted background correction. The speed of analysis for 35 elements with 3 replicate readings is shown in Table 3. Figure 2. Four hour stability. Water Analysis Because of the excellent detection limit capability of the axial system, the water analysis was conducted on this unit. The results of the analysis of NIST SRM 1643C and 1643D water samples for both major and trace elements are listed in Table 4. The measured values compare well with the certified values. 3

Table 4. Results of Water Analysis by Vista Axial ICP NIST 1643 C (µg/l) NIST 1643 D (µg/l) Element Measured Certified Measured Certified Ag 2.37 ± 0.22 2.21 ± 0.30 1.31 ± 0.11 1.27 ± 0.06 Al 121.0 ± 2.4 114.6 ± 5.1 127.2 ± 3.6 127.6 ± 3.5 As 79.2 ± 3.9 82.1 ± 1.2 54.5 ± 1.5 56.02 ± 0.73 B 121.9 ± 1.8 119 ± 1.4 138.8 ± 3.6 144.8 ± 5.2 Ba 50.9 ± 0.1 49.6 ± 3.1 495.0 ± 6.0 506.5 ± 8.9 Be 24.2 ± 0.2 23.2 ± 2.2 13.01± 0.1 12.53 ± 0.28 Bi 11.3 ± 5.0 (12) 13.5 ± 3.0 (13) Cd 12.8 ± 0.2 12.2 ± 1.0 6.38 ± 0.38 6.47 ± 0.37 Co 24.7 ± 2.0 23.5 ± 0.8 25.3 ± 0.5 25 ± 0.59 Cu 22.4 ± 2.6 22.3 ± 2.8 21.0 ± 3.0 20.5 ± 3.8 Fe 101.6 ± 3.5 106.9 ± 3.0 80.2 ± 2.0 91.2 ± 3.9 Mn 37.6 ± 0.04 35.1 ± 2.2 40.1 ± 0.68 37.66 ± 0.83 Mo 105.1 ± 2.1 104.3 ± 1.9 110.1 ± 2.0 112.9 ± 1.7 Ni 61.6 ± 1.1 60.6 ± 7.3 60.9 ± 3.0 58.1 ± 2.7 Pb 38.4 ± 0.5 35.3 ± 0.9 19.7 ± 0.8 18.15 ± 0.64 Sb 56.5 ± 0.9 54.1 ± 1.1 Se 11.8 ± 2 12.7 ± 0.7 14.4 ± 0.6 11.43 ± 0.17 Sr 278.0 ± 1.5 263.6 ± 2.6 315.9 ± 2.4 294.8 ± 3.4 Te 3.6 ± 0.8 (2.7) 1.3 ± 0.4 (1) Tl 8.5 ± 2.4 (7.9) 7.6 ± 0.5 7.28 ± 0.25 V 29.5 ± 0.9 31.4 ± 2.8 34.3 ± 0.6 35.1 ± 1.4 Zn 89.2 ± 2.6 73.9 ± 0.9 79.1 ± 1.3 72.48 ± 0.65 mg/l mg/l mg/l mg/l Ca 37.4 ± 0.3 36.8 ± 1.4 31.21 ± 0.46 31.04 ± 0.50 K 2.27 ± 0.03 (2.3) 2.31 ± 0.02 2.356 ± 0.035 Mg 9.88 ± 0.41 9.45 ± 0.27 8.31 ± 0.17 7.989 ± 0.035 Na 11.62 ± 1.25 12.19 ± 0.36 22.14 ± 0.43 22.07 ± 0.64 Si 2.84 ± 0.56 (2.7) Summary The Vista CCD simultaneous ICP-OES has been shown to be an excellent productivity tool for the water analysis laboratory. The Vista axially-viewed ICP system provides 5-10 times better detection limits than the radially-viewed ICP system, and easily meets the current and proposed CRDLs for the US EPA. The short and long term stability of the Vista are excellent. The measured values of the NIST SRM water samples are in good agreement with the certified values. The Vista instrument provides quick warm up time and fast sample throughput. Data in parentheses are not certified 4

References 1. D. R. Demers, Evaluation of the axially viewed (end-on) inductively coupled argon plasma source for atomic emission spectroscopy, Appl. Spectrosc., 33, 1979, 584. 2. L. M. Faires, T. M. Bieniewski, C. T. Apel, T. M. Niemczyk, Top-down versus side-on viewing of the inductively coupled plasma, Appl. Spectrosc., 39, 1985, 5. 3. Y. Nakamura, K. Takahashi, O. Kujirai, H. Okochi, C. W. McLeod, Evaluation of an axially and radially viewed inductively coupled plasma using an Echelle spectrometer with wavelength modulation and second-derivative detection, J. Anal. At. Spectrosm., 9, 1994, 751. 4. J. C. Ivaldi, J. F. Tyson, Performance evaluation of an axially viewed horizontal inductively coupled plasma for optical emission spectrometry, Spectrochim Acta, 50l, 1995, 1207. 5. T. T. Nham, Performance evaluation and applications of an axially-viewed ICP-AES, paper No. 223, presented at the 13th Analytical Chemistry Conference in Darwin Australia, 1995. 6. A. T. Zander, C. B. Cooper III, R. L. Chien, Optical detector for Echelle spectrometer, U.S. Patent number 5,596,407, Jan 21 1997. 7. T. W. Barnard, M. I. Crockett, J. C. Ivaldi, P. L. Lundberg, D. A. Yates, P. A. Levine, D. J. Sauer, Solid-state detector for ICP-OES, Anal. Chem., 65, 1993, 1231. 8. Contract Laboratory Program Statement of Work for Inorganics Analysis Multi-media Multi-concentration SOW No. 788, U.S. EPA, Washington D. C., 1988. 9. USEPA Contract Laboratory Program Statement of Work for Inorganics Analysis and Classical Chemistry Parameters. Multi-media Multi-concentration, Draft ILMO 5.0C, March 1999. 10. M. Carre, E. Poussel, J. M. Mermet, Drift diagnostics in inductively coupled plasma atomic emission spectrometry, J. Anal. At. Spectrosm., 7, 1992, 791 For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 1999 Printed in the USA November 1, 2010 ICPES-28