Method for the determination of dimethyl sulfate

Similar documents
Method for the determination of 1,3-butadiene

Methods for the determination of vinyl chloride

Hexane isomers (except n-hexane) and methylcyclopentane

Methods for the determination of ethylene oxide

HYDROCARBONS, AROMATIC 1501

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Determination of Volatile Substances Proof of Food Adulteration

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

STANDARD OPERATING PROCEDURES

FORMALDEHYDE by GC 2541

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

METHYLTIN CHLORIDES 5526

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Alkali metal hydroxides and alkaline earth hydroxides

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Analytical Trap Comparison for USEPA Method 8260C

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

N-Methylformamide and N-methylacetamide in urine

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

VOC TEST REPORT M1. 6 April Regulation or protocol Conclusion Version of regulation or protocol

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Application Note 032

637. Thiamethoxam. HPLC method

Test Report. Tremco-Illbruck GmbH & Co KG. Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer. June 2011

A biphasic oxidation of alcohols to aldehydes and ketones using a simplified packed-bed microreactor

Tar analysis by Solid Phase Adsorption (SPA) associated with Thermal Desorption (TD) and Gas Chromatography (GC) analysis

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

VOC TEST REPORT CDPH

Chemistry Gas Chromatography: Separation of Volatile Organics

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Method for Characterization of Gum Rosin by Capillary Gas Chromatography

Gas Chromatography (GC)

Sulfotepp impurities in Chlorpyrifos EC formulations

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

VOC TEST REPORT AgBB

METHOTREXATE mg/m 3 (arbitrary). There is no OSHA PEL or AGGIH TLV for

Headspace Technology for GC and GC/MS: Features, benefits & applications

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations

VOC EMISSION TEST REPORT BREEAM NOR

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

2401 Gas (liquid) Chromatography

TDTS. Keywords. Introduction. TD Operation; Calibration; Impedance; Discrimination

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Supporting Information

NCASI METHOD CI/SG/PULP-94.02

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

VOC TEST REPORT CDPH

Abstract. Keywords. Author Information. Air Monitoring Methods. The MAK Collection for Occupational Health and Safety 2018, Vol 3, No 2

Selection of a Capillary GC Column

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

HPLC Praktikum Skript

VOC TEST REPORT CDPH

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

VOC TEST REPORT M1. 16 August Regulation or protocol Conclusion Version of regulation or protocol

Test Report in accordance CDPH-IAQ

Introduction to Capillary GC

VOC TEST REPORT M1. 23 October Regulation or protocol Conclusion Version of regulation or protocol

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

ENVIRONMENTAL analysis

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Peroxides (peracetic acid and hydrogen peroxide)

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Method for Nitrite determination on Low-Range Samples

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Report AFK0242/18 TABLE OF CONTENTS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Transcription:

German Social Accident Insurance Deutsche Gesetzliche Unfallversicherung Analytical Subcommittee of the Chemistry Board of Experts* Carcinogenic substances Order number: BGI 505-7-05 Established methods: Issued: February 2007 Method for the determination of dimethyl sulfate This method has been tested and recommended by the German Social Accident Insurance for the determination of dimethyl sulfate at workplaces. Both personal and stationary sampling can be performed for the assessment of workplaces. Sampling is carried out with a pump and adsorption on Tenax ¾. Analysis is performed by gas chromatography with mass selective detection (GC/MS) after thermal desorption. Chemical names: Dimethyl sulfate, sulfuric acid dimethyl ester CAS number: 77-78-1 Molecular formula: C 2 H 6 O 4 S Molar mass: 126.13 g/mol * Please direct correspondence to Berufsgenossenschaft RCI, Bereich Pråvention, Box 101480, 69004 Heidelberg, Germany; analytik@bgrci.de The MAK-Collection Part III: Air Monitoring Methods, Vol. 12. DFG, Deutsche Forschungsgemeinschaft Copyright 2011 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim ISBN: 978-3-527-32818-5

Analytical Methods 72 Summary This method can be used with personal or stationary sampling to determine the mean concentration of dimethyl sulfate at workplaces over the sampling time. Principle: Limit of quantification: Selectivity: Advantages: Disadvantages: Apparatus: A pump draws a defined volume of air through a metal tube filled with Tenax. The adsorbed dimethyl sulfate is then thermally desorbed and determined by gas chromatography. Absolute: 23 ng of dimethyl sulfate Relative: 0.9 lg/m 3 of dimethyl sulfate for a 25 L air sample. The method is selective due to the combination of gas chromatographic separation and mass selective detection. Personal and selective measurements are possible. No indication of peak concentrations. Pump with gas meter or volumetric flow meter, adsorption tubes filled with Tenax, gas chromatograph with thermal desorber and mass selective detector.

73 Dimethyl sulfate Detailed description of the method Contents 1 Equipment, chemicals and solutions 1.1 Equipment 1.2 Chemicals and solutions 2 Sampling procedure 3 Analytical determination 3.1 Sample preparation and analysis 3.2 Instrumental operating conditions 4 Evaluation 4.1 Calibration 4.2 Calculation of the analytical result 5 Reliability of the method 5.1 Precision and recovery 5.2 Limit of quantification 5.3 Selectivity 6 Remarks 7 References 1 Equipment, chemicals and solutions 1.1 Equipment For sampling: ± Pump, suitable for a flow rate of 5 to 200 ml/min, e.g. PP1, Gilian, supplied by DEHA Haan & Wittmer, 71296 Heimsheim, Germany ± Gas meter or volumetric flow meter ± Adsorption tube filled with Tenax as the collection phase: Metal tubes serve as adsorption tubes. Their external diameter and length must be adapted to the thermal desorber used. The tubes are filled with 250 mg of Tenax, which is enclosed by two metal sieves. After being filled, the tubes must be previously purged several times in a stream of helium at 2508C in the thermal desorber. Then they are sealed with Swagelok ¾ screw caps with Teflon seals. For the sample preparation and analysis: ± Volumetric flasks, 10 ml ± Adjustable piston pipette, e.g. Multipette pro, Eppendorf, 22366 Hamburg, Germany ± Microliter syringe, 10 ll ± Spiking facility (consisting of a brass T-piece with septum) to apply calibration or internal standard solutions to adsorption tubes

Analytical Methods 74 ± Thermal desorber ± Gas chromatograph with mass selective detector (GC/MS) 1.2 Chemicals and solutions Dimethyl sulfate, 99%, e.g. from Aldrich, 82024 Taufkirchen, Germany Naphthalene, as the internal standard (IS), purity > 99%, e.g. from Aldrich Methyl-tert.-butyl ether (MTBE), purity 99.8%, anhydrous, e.g. from Aldrich Tenax ¾, 60±80 mesh, e.g. from Supelco, 82024 Taufkirchen, Germany Helium 5.0 (purity 99.999%), to operate the gas chromatograph and the thermal desorber Naphthalene stock solution: Internal standard solution (IS): Solution of approx. 5.32 mg of naphthalene per ml of methyl-tert.-butyl ether (MTBE). Approx. 26.6 mg of naphthalene are weighed to the nearest 0.1 mg in a 5 ml volumetric flask. The flask is then filled to the mark with MTBE and shaken. Solution of approx. 266 lg of naphthalene (IS) per ml of MTBE. 500 ll of the IS stock solution are pipetted into a 10 ml volumetric flask that already contains several milliliters of MTBE. The volumetric flask is then filled to the mark with MTBE and shaken. Calibration stock solution: Solution of approx. 20 mg of dimethyl sulfate per ml of MTBE. Approx. 150 ll of dimethyl sulfate are pipetted into a 10 ml volumetric flask that already contains several milliliters of MTBE and the dimethyl sulfate is weighed to the nearest 0.1 mg. The volumetric flask is then filled to the mark with MTBE and shaken. Calibration solutions: Solutions of approx. 50 to 1000 lg of dimethyl sulfate per ml of MTBE. 25 ll, 100 ll, 200 ll, 300 ll, 400 ll and 500 ll each of the calibration stock solution are added with the piston pipette to six 10 ml volumetric flasks containing a few milliliters of MTBE. The volumetric flasks are then filled to the mark with MTBE and shaken. 5 ll each of these solutions are transferred via the spiking facility to six separate Tenax tubes. These adsorption tubes cover a concentration range of dimethyl sulfate from 0.01 mg/m 3 to 0.198 mg/m 3 for an air sample volume of 25 L. The corresponding concentrations are shown in Table 1.

75 Dimethyl sulfate Table 1. Concentrations of the calibration solutions of dimethyl sulfate. Stock solution added [ll] Concentration of dimethyl sulfate [mg/l] Concentration of dimethyl sulfate [mg/m 3 ]* 25 50 0.010 100 200 0.040 200 400 0.079 300 600 0.119 400 800 0.158 500 1000 0.198 * based on an air sample volume of 25 L 2 Sampling procedure Shortly before a previously purged adsorption tube is used, it must be heated again for approx. 15 minutes in the thermal desorber at 250 8C in a stream of helium. Then 5 ll of the IS standard solution are applied via the spiking facility to the adsorption tube which is then sealed with a Swagelok ¾ screw cap with a Teflon seal. For sampling the adsorption tube is opened and connected to the pump. The pump and tube are worn by a person during sampling or they are used for stationary sampling. The flow rate is adjusted to 100 ml/min. At a sampling time of 4 hours this is equivalent to an air sample volume of approx. 25 L. After sampling the tube must be tightly sealed again. 3 Analytical determination 3.1 Sample preparation and analysis The loaded adsorption tube is opened and inserted into the thermal desorber. The adsorbed dimethyl sulfate is desorbed by heating and transferred to the cold trap of the device containing Tenax by means of helium and collected there. From there it is transferred to the separation column of the gas chromatograph by sudden heating (device-specific heating rate approx. 40 8C/s) and a chromatogram is recorded as described in Section 3.2. The quantitative evaluation is performed according to the internal standard method using the peak areas or heights of dimethyl sulfate and of naphthalene as the internal standard. 3.2 Instrumental operating conditions The characteristics of the method stated in Section 5 are determined under the following instrumental conditions:

Analytical Methods 76 Apparatus: ATD 400 thermal desorber from PerkinElmer LAS, 63110 Rodgau-Jçgesheim, Germany HP 6890 gas chromatograph with an MSD HP 5872 mass selective detector Operating conditions for thermal desorption Desorption oven: 250 8C Inlet split: Closed Cold trap: Filling Tenax ¾ Temperatures: ±30 8C (adsorption) 280 8C (injection) Desorption time: 5 minutes Desorption flow rate: 11.4 ml/min Outlet split: 20.0 ml/min Transfer line: 200 8C Operating conditions for gas chromatography Separating column: Fused silica capillary, stationary phase DB-1, cross-linked (dimethylpolysiloxane) internal diameter 0.25 mm, film thickness 1 lm, length 30 m Oven temperature: Initial temperature: 50 8C Heating rate 20 8C/min to the final temperature of 240 8C Detector temperature: 250 8C Carrier gas: Helium 5.0, 24 psi Transfer line MS: 280 8C Operating conditions for mass spectrometry Ionization type: Electron impact ionization (EI) (70 ev) MS quad: 150 8C MS source: 230 8C Solvent delay: 3 minutes Measurement mode: Selected ion monitoring (SIM) Registered masses Quantification/Qualification (m/z in amu): naphthalene 128 127 dimethyl sulfate 95 96 4 Evaluation 4.1 Calibration 5 ll each of the calibration solutions given as examples in Section 1.2 and 5 ll of the internal standard solution are applied via the spiking facility which is connected to an adsorption tube. Then air at a flow rate of approx. 50 ml/min is drawn for 10 minutes through the tube to transfer the respective calibration solution to the adsorp-

77 Dimethyl sulfate tion tube. The quantities of dimethyl sulfate used to spike the adsorption tubes and the air concentrations calculated for an air sample volume of 25 L are shown in Table 1. The tubes are then analyzed as described in Section 3.2. The calibration function is obtained by plotting the peak area ratios of dimethyl sulfate and the internal standard versus the weight ratios of dimethyl sulfate and the internal standard on the adsorption tubes. 4.2 Calculation of the analytical result The mass of dimethyl sulfate corresponding to the peak area in the sample (adsorbed mass on the Tenax tube) is calculated with the calibration function. The dimethyl sulfate concentration by weight in the air sample in mg/m 3 is calculated according to equation (1): c = m V g 1 where c is the concentration by weight of dimethyl sulfate in the air sample in mg/m 3 m is the mass of dimethyl sulfate adsorbed on the Tenax tube in lg V is the air sample volume in L g is the recovery (see Section 5.1) 5 Reliability of the method 5.1 Precision and recovery The precision in the minimum measurement range according to DIN EN 482 [1] and the recoveries were determined for three different concentrations (see Tables 2 and 3). The following validation solutions were prepared for this purpose: Validation stock solution: Solution of 22.5 mg of dimethyl sulfate per ml of MTBE. 170 ll of dimethyl sulfate were pipetted into a 10 ml volumetric flask that already contained several milliliters of MTBE and the dimethyl sulfate was weighed to the nearest 0.1 mg. Then the volumetric flask was filled to the mark with MTBE and shaken. Validation solutions: Solutions of 56, 562 and 1124 lg of dimethyl sulfate per ml of MTBE. 25 ll, 250 ll, and 500 ll each of the validation stock solution were added to three 10 ml volumetric flasks,

Analytical Methods 78 which contained several milliliters of MTBE. The volumetric flasks were then filled to the mark with MTBE and shaken. To determine the precision of the method for each of the investigated concentrations (Table 2) six adsorption tubes were connected to a multiple tube holder that enabled the flow rate of the individual pump to be adjusted for each tube separately. A glass tube filled with a plug of quartz wool was fitted in front of each adsorption tube by means of a short piece of Teflon tube. 5 ll each of the three validation solutions and 5 ll of the internal standard solution were applied to the quartz wool of the glass tubes. Then 25 L of ambient air were drawn through each tube at room temperature (23 to 25 8C) at a flow rate of 200 ml/min. Concentrations of 0.23 mg/m 3,0.11mg/ m 3 and 0.01 mg/m 3 at an air humidity of approx. 55% and 85% were thus obtained. The characteristics obtained for a relative air humidity of 55% are shown in Table 2. Table 2. Characteristics of the method at a relative air humidity of approx. 55%. Concentration of dimethyl sulfate [mg/m 3 ] Standard deviation (rel.) [%] Recovery 0.01 7.9 1.02 0.11 5.4 0.99 0.23 4.3 0.92 The mean standard deviation was 5.9% and the mean recovery 0.98. The characteristics obtained for a relative air humidity of 85% are shown in Table 3. Table 3. Characteristics of the method at a relative air humidity of approx. 85%. Concentration of dimethyl sulfate [mg/m 3 ] Standard deviation (rel.) [%] Recovery 0.01 5.3 0.84 0.11 4.4 0.89 0.23 3.1 0.92 The mean standard deviation was 4.3% and the mean recovery 0.88. The relative air humidity influenced the relative standard deviations only slightly. The recovery for the lower concentrations fell markedly at higher air humidity. The storage stability of dimethyl sulfate on the Tenax tubes was not notably changed due to the air humidity during sampling.

79 Dimethyl sulfate 5.2 Limit of quantification The limit of quantification was determined as described in DIN 32645 [2] by rapid estimation. The standard deviation required for the calculation was obtained by sixfold injection of a solution of dimethyl sulfate with a concentration of 0.025 mg/l (corresponding to 0.005 mg/m 3 for an air sample volume of 25 L). The absolute limit of quantification is 23 ng of dimethyl sulfate. The relative limit of quantification is 0.9 lg/m 3 for an air sample volume of 25 L. 5.3 Selectivity The method is selective due to the combination of gas chromatographic separation and mass selective detection. 6 Remarks Dimethyl sulfate can be stored in the adsorbed state at room temperature for at least 7 days. The calibration solutions as well as and the stock solution must be freshly prepared at the latest after 48 hours, as residual water in MTBE leads to hydrolysis of dimethyl sulfate. Further analytical methods for the determination of dimethyl sulfate in the air of workplaces are also published in this series of volumes. One method is entitled BGI 505-7-03 (Sampling with a pump, adsorption on Tenax TA, desorption, gas chromatography and sulfur-specific detection; issued in January 1987) [3] and another is entitled BGI 505-7-04 (Sampling with a pump, adsorption on Tenax TA, desorption, gas chromatography and mass-selective detection; issued in April 1997) [4]. Apart from the determination of dimethyl sulfate, these methods are also suitable for measuring other dialkyl sulfates in the air of workplaces. 7 References [1] DIN EN 482 (1994) Workplace atmospheres ± General requirements for the performance of procedures for the measurement of chemical agents. Beuth Verlag, Berlin. [2] DIN 32645 (1994) Chemical analysis; decision limit, detection limit and determination limit under repeatability conditions; terms, methods, evaluation. Beuth Verlag, Berlin. [3] BGI 505-7-03 (2002) Method for the determination of dimethyl sulfate (DMS). In: Deutsche Forschungsgemeinschaft (Kettrup A., Greim H. eds.) Analysis of Hazardous Substances in Air, Volume 5, 93±99. Wiley-VCH Weinheim. [4] BGI 505-7-04 (2002) Method for the determination of dimethyl sulfate (DMS). In: Deutsche Forschungsgemeinschaft (Kettrup A., Greim H. eds.) Analysis of Hazardous Substances in Air, Volume 5, 100±105. Wiley-VCH Weinheim. Author: W. Kråmer