CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS

Similar documents
Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Gy can be used for any type of radiation. Gy does not describe the biological effects of the different radiations.

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

Quantities and Units for Use in Radiation Protection

11/23/2014 RADIATION AND DOSE MEASUREMENTS. Units of Radioactivity

Commissioning of the Beta Secondary Standard (BSS2)

Chapter 2 Dose Quantities and Units for Radiation Protection

Nuclear Physics and Astrophysics

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Survey Meter OD-01 Address: Phone: Fax: URL:

1. RADIOACTIVITY AND RADIATION PROTECTION

05/11/2013. Nuclear Fuel Cycle Ionizing radiation. Typical decay energies. Radiation with energy > 100 ev. Ionize an atom < 15eV

Akira Endo ICRP Committee 2 & ICRU Report Committee 26 Japan Atomic Energy Agency

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

2 Units of Radiation Protection

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

Nuclear Fusion and Radiation

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Ion-Chamber Survey Meter OD-02

Image quality assessment. Question: which is a better image? Answer: what are you trying to do?

Radioactivity. Lecture 7 Dosimetry and Exposure Limits

Ionization Chamber. Pocket dosimeter

Personal Dose Monitoring System

ICRP Symposium on the International System of Radiological Protection

A new neutron monitor for pulsed fields at high-energy accelerators

WHAT IS IONIZING RADIATION

Lab #13: MEASUREMENT OF RADIATION DOSES

Extensions to the Beta Secondary Standard BSS 2 Consolidated version from

11 Gamma Ray Energy and Absorption

Radiation Quantities and Units

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Secondary Particles Produced by Hadron Therapy

General Physics (PHY 2140)

General Physics (PHY 2140)

Neutron Dosimetry with Ion Chamber-Based DIS System

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

Radiation Protection & Radiation Therapy

Atoms, Radiation, and Radiation Protection

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Higher -o-o-o- Past Paper questions o-o-o- 3.6 Radiation

Chapter 2. Atomic Structure and Nuclear Chemistry. Atomic Structure & Nuclear Chemistry page 1

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

LET! (de / dx) 1 Gy= 1 J/kG 1Gy=100 rad. m(kg) dose rate

CHARGED PARTICLE INTERACTIONS

Nuclear Spectroscopy: Radioactivity and Half Life

Uncertainty and detection limit in decontamination measurements of gamma emitting radio nuclides in unshielded environments Tom Koivuhuhta

Calibration of a pencil ionization chamber with and without preamplifier

INTERNAL RADIATION DOSIMETRY

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Working Correctly & Safely with Radiation. Talia Tzahor Radiation safety officer Tel:

Nuclear Radiation. Natural Radioactivity. A person working with radioisotopes wears protective clothing and gloves and stands behind a shield.

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Mitigation of External Radiation Exposures

Dosimetry. Sanja Dolanski Babić May, 2018.

Chapter V: Cavity theories

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

DOSE CONVERSION FACTORS. David C. Kocher and Keith F. Eckerman Health and Safety Research Division Oak Ridge National Laboratory

Technical basis of occupational dosimetrygeneral

Chapter 29. Nuclear Physics

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

Interactions with Matter Photons, Electrons and Neutrons

Industrial Hygiene: Assessment and Control of the Occupational Environment

ICRP Symposium on the International System of Radiological Protection

Neutron Users Club Meeting

Radiation Safety. PIXE PAN 2008 Ed Stech University of Notre Dame

Occupational Radiation Protection at Accelerator Facilities: Challenges

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Monte Carlo Simulation concerning Particle Therapy

Differentiating Chemical Reactions from Nuclear Reactions

Physics of Radiography

Calculation of Effective Dose Conversion Coefficients for Electrons

Radiation Safety Training Session 1: Radiation Protection Fundamentals and Biological Effects

Development and Test of a GEM-Based TEPC for Neutron Protection Dosimetry

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

M [scale units/s] of the system

College Physics B - PHY2054C

2015 Ph.D. Comprehensive Examination III. Radiological Sciences - Medical Physics

Physics 219 Help Session. Date: Wed 12/07, Time: 6:00-8:00 pm. Location: Physics 331

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Quantities, Units and Definitions

Activities of the International Commission on Radiological Protection

ICTP-IAEA Joint Workshop on Nuclear Data for Science and Technology: Medical Applications. 30 September - 4 October, 2013

Reliability and Traceability in Radiation Calibration

Neutron interactions and dosimetry. Eirik Malinen Einar Waldeland

Basic physics Questions

Fundamentals of radiation protection

Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes

Organ doses determined using a RANDO phantom for different radionuclide depositions and photon energies

Organ and effective dose rate coefficients for submersion exposure in occupational settings

Chapter. Nuclear Chemistry

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Study of the Different Modes of the External Exposure to the Terrestrial Gamma Rays at Seila Area, Southeastern Desert, Egypt

ISO INTERNATIONAL STANDARD

SECTION 8 Part I Typical Questions

Ba (Z = 56) W (Z = 74) preferred target Mo (Z = 42) Pb (Z = 82) Pd (Z = 64)

Chapter 16 Basic Precautions

Transcription:

CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS F. SCARLAT 1, A. SCARISOREANU 1, M. OANE 1, E. BADITA 1, E. MITRU 1 1 National Institute for Laser, Plasma and Radiation Physics, Bucharest-Magurele, Romania Abstract. This work presents the operational parameters used for personal and radiation field monitoring: personal dose equivalent H p (10, R, α) and ambient dose equivalent H*(10, R, α) for powerful penetration radiation, the measurement method using both secondary standard ionization chambers from STARDOOR laboratory and the results of measurements made in the room adjacent to the irradiation hall where the 7 MeV liniar accelerator is located, also including the results of the measurements conducted at the Sr-90-20 MBq radioactive source for te sting the ionizing chambers. 1. Introduction As regards the external exposure to radiations, there are four operational parameters which show remarkable interest for radiation field characterization in radioprotection purposes. These ICRU parameters are: the personal dose equivalent, penetration, H p(d), personal dose equivalent, superficial, H S (d), ambient dose equivalent, H*(d) and directional dose equivalent, H (d), where d is the depth under a specific point on the body, which is proper f or the penetration powerful radiation. All these parameters are based on the absorbed dose equivalent concept in one point rather than on the equivalent dose concept. In this paper, the descriptions of the concepts for the four operational parameters used for the radiation field characterization is followed by the presentation of some details on the standard chambers, on the measurement method and the dosimetric measured results made for the room adjacent to the irradiation hall where the 7 MeV linear accelerator is located and at the radioactive check device (Sr-90-20 MBq) used for ionization chamber testing. 2. Concepts of equivalent / effective dose The equivalent dose proper for an organ or tissue T is given by relation H T = Σ R w R x D T,R, [Sv] (1) where w R is the radiation weighted factor for R radiation and D T,R is the absorbed dose for a tissue or organ T, due to R radiation. Table 1 shows the values for different kinds of radiations. Table 1.Values of w R for some radiations Radiation w R w R = A + B exp (-C/6) ± ± γ, e, μ, for all energies 1 1 Neutrons, E < 1MeV 1-10 A = 2.5; B = 18.20; C = (ln(e)) 2 Neutrons, E = 1MeV - 50 MeV 5-20 A = 5.0; B = 17.00; C = (ln(2e)) 2 Neutrons, E > 50 MeV 5 A = 2.5; B = 3.50; C = Protons, other than recoil one, E > 2 MeV 5 (ln(0.04e)) 2 Alpha particles, fission fragments, heavy 20 5 nucleus 20 317

The effective dose, E, recommended by ICRP 60, is defined like E = Σ T w T x H T, [Sv] (2) where w T is the weighted factor for the tissue or the organ T, its values being shown in Table 2. Table 2. Weighted factors for organs / tissue (w T ) Organ or tissue wt Σ T w T Bone surface, skin 0.01 0.02 Bladder, breast, liver, esophagus, thyroid 0.05 0.25 Bone marrow(red), colon, lung, stomach 0.12 0.48 Gonads 0.20 0.20 Remainder Tissues (10 in total: Adrenals, Brain, trachea small intestine, muscle, Pancreas, Kidneys, spleen, thymus & uterus) 0.05 0.05 Total Body 1.00 1.00 The total effective dose collected by a person in a g group age is determined with the relation: E = E external + j e(g) j,inh x A j,inh + j e(g) j,ing x A j,ing + k D k (g) x Q k, [Sv] (3) where E external represent the interest effective dose from the external exposure, [Sv], e(g) j,inh and e(g) j,ing are the effective doses engaged o n the incorporation unit of one radionuclide j, [Sv/Bq], by inhalation, ingestion respectively, by one person from a g group age, A j,ing and A j,inh represent the radionuclide incorporation j, [Bq], by ingestion, inhalation respectively, D k [Gy] is calculated with relation (4): D k = 1/ρ Σ j Ek E k Eγ N j σ kj (E γ ) f j (E k,e γ ) φ ( E γ ) de γ E k, [Gy] (4) where: D k represents the absorbed dose from the charged particle k resulted from nuclear reaction (γ,p), (γ,d), (γ,t), (γ, He 3 ), (γ,α) and (γ,n), in any organ or tissue T, with Q k radiation quality factor; E k the charged particle energy k; N j - atomic number density of the nucleus j in the phantom; σ ij (E γ ) the efficiency section of the nucleus j to produce particle k in reaction to the energy of the phantom E γ, f j (E k,e γ ) the energetic spectrum of particle k produced in the reaction between one photon with E γ energy and j nucleus; φ E ) - the mean value of the photon spectrum in each organ or tissue, ρ phantom tissue density. 3. Concepts of personal/ambient dose equivalent For to define the parameters related to the effective dose equivalent and the skin dose equivalent the stipulation of some radiation fields derived from the radiation field is useful. In ICRU (ICRU, 1985) the expanded and aligned terms are given to characterize this derived radiation field. In the expanded field, the fluency and its angular and energy distribution have the same values everywhere in the interest volume like the real field in the reference point. In the aligned and expanded field, the fluency and its angular distribution are the sam e like in the expanded field, but the fluency is unidirectional. ( γ 318

For personal monitoring two concepts are introduced. The first one, the personal dose equivalent- penetration, Hp(d), is proper for an organ or tissue placed deep into the body and that can be irradiated with powerful penetration radiation. The second concept, the personal dose equivalent- superficial, Hs(d), is proper for superficial organs and tissue that are to be irradiat ed with powerful and soft penetration radiation. The personal dose equivalent penetration, Hp(d) is the dose equivalent in soft tissue, defined under a specific point on the body on d depth, which is adequate for powerful penetration radiation. As regards the monitoring of the area and the medium, two concepts which connect the external radiation field to the effective dose and equivalent dose in the skin, are introduced. The first one, the ambient dose equivalent, H*(d), in one point in the radiation field, is the dose equivalent that can be produced by a properly aligned and expanded field, in ICRU sphere, to a d depth, on the opposite direction ray of the aligned field. The measuring unit is [Sv]. The directional dose equivalent, H (d), in one point of the radiation field is the dose e quivalent that can be produced by the proper expanded field in ICRU sphere to a d depth on a ray in one specified direction. Its measuring unit is [Sv]. Reference axis Measuring point 300 40 200 300 Fig. 1. The device model 4. Secondary standard chambers Hp(10) secondary standard chamber 34035 type, was developed by Ankerhold et all at PTB. The device consists of one (30 x 30 x 15) cm ionization chamber made in a PMMA phantom used for personal dose equivalent measurement, Hp(10), namely, the radiation quantity which includes the radiation field and the phantom scattered radiation. The Hp(10) chamber response is presented in Figure 2. The chamber is connected to a re ference dosimeter T10005-50406 model. Introducing the ionization chamber parameters in UNIDOS dosimeter memory and also its calibration factor, the absorbed dose equivalent, measured in Sv, at d = 10 mm depth under the marked point on the irradiated body, is obtain. The secondary standard chamber Hp(10) was calibrated at PTW regarding photon with 15 kev 1400 kev energy measurement for quality radiation of narrow spectrum (N), corresponding to ISO 4037-1 with gamma radiation emitted by radionuclide S-Cs 137 (661.6 kev; T m 2 1/2 =11.500 days, 0.079 μsv /h MBq) with 15,2 ± 3 % msv/h rate. The 6 calibration factor value for this secondary standard chamber is N H = 3.17 10 Sv/C, and its relative expanded uncertainty is 3 %. 319

k(r, α) 2.0 1.8 0 60 1.6 45 75 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0 E ph (kev ) Fig. 2. Correction factors k(r, α) for radiation quality N-10 si N-300 and S-Cs ISO to incidence angles α, of 0 o, 45 o, 60 o, 75 o and for S-Co to angles α = 0 o and 60 o. E is the ph photon mean energy given to N-10 and to N-120 by Ankerhold (Since Ankerhold (2000)) and for N-150 to N-300, S-Cs and S-Co with ISO/FDIS 4037-3. The spherical ionization chamber H*(10), TN 32002 type, was calibrated at PTW for to be used like a secondary standard chamber for ambient dose equivalent measurement, for photons between 25 kev 50 MeV energy range. The radiation quality used for calibration was chosen in conformity with IEC60731 / ISO 4037-1 and the radiation was emitted by a Co 60 source with ambient dose equivalent rate ranging 50-1000 msv/h. The calibration factor value given in the calibration certificate delivered by PTW, is N H = 2,863 10 4 Sv/C, and its measurement uncertainty is equal with 2,5 %. eading / R Dose 1,1 1,0 0,9 0,96 0,96 0,97 0,98 0,99 Photon equivalent dose 50 100 150 200 Photon Energy [kev] 60 Co Fig. 3. Energy dependence for spherical ionization chamber type 32002. The H*(10) chamber response given in Fig. 3, shows the dependence of the photon energy in the domain between 50 kev and Co-60 energy and the photon dose equivalent. 5. Measurement and calculation method The personal dose equivalent determination, Hp(10, R, α), for X and γ rays, of R quality, at α incidence angle between the beam and the normal to the phantom surface, is given by the relation: H p (10, r, α)= N H k(r, α)q, (5) 320

were N H is the calibration factor for the reference radiation quality N-60 and the reference angle of radiation incidence α = 0º; k(r, α) correction factor for the radiation quality R and the angle of radiation incidence α and Q is charge measured by the chamber. The calibration factor N H with respect to the personal dose equivalent Hp(10), is 137 determined with the radiation quality S-Cs and the incidence angle α = 0º. It is given by the expression: o K N S Cs, 0 a H = hpk (10; ), Q (6) where K a is conventional true value of the air kerma fr ee in air; h pk (10) is conversion coefficient from K a to Hp(10) and S-Cs shows that the measurements were made for a PMMA phantom to a reference radiation S-Cs and incidence angle α = 0º. The correction factor k(r,α) for the radiation quality R = S-Cs 137 ( E pk = 662keV ) at α incidence angle is given by formula: K a 1 k ( R, α ) = ( hpk (10; R, α ) ) Q N H, (7) where h pk (10, R, α), Q, K a and N H as given above. The ambient dose equi valent determination in radiation f ield of R quality is made with the relation H*(10) = N 60 k(r ) Q, (8) where N 60 is the cal ibration factor at the reference radiation qu ality N-60; k(r )- is the correction factor at the radiation quality R and Q is the charge measured by the chamber. * K a N 60 = k K (10, N 60), (9) Q where K a - the conventionally true value of the air kerma free-in-air for the reference quality N-60; k R *(10, N-60) is the conversion coefficient from K a to H*(10) for the quality N-60. For the determination of the calibration factor N 60 and the correction factors k(r), the measurements were carried out for energies ranging 9-1250 kev using the radiation qualities of the ISO, narrow-spectrum series (N 10 N 300 ) and the ISO gamma qualities S-Cs and S-Co. 6. Results and discussions All measurements were made by the ionization chamber H p(10) vertically positioned to 140 cm distance from the wall of the hall in which the accelerator is located. In sets 1 and 2, the accelerator was turned-off, and in other sets (3 and 4) the accelerator was operational in the electron and bremstrahlung modes (5 and 6). Tables show the preliminary results for personal dose equivalent, measured in the room adja cent to the ha ll the accelerator is located. The results led to the conclusion that, when the accelerator is turned-off, Hp(10) of 74.16 va lue and 73.11 respectively, the equivalen t dose o f 303 μsv/h is not exceeded, and when the accelerator is operational, Hp(10) of 91.98 value and 85.92 respectively, the equivalent dose of 303 μsv/h is not exceeded. Analyzing the values in the table, it results that only the population allowable equivalent dose of 5.7 μsv/h is exceeded in both accelerator conditions. For checking, a Sr-90 radiation source with 20 MBq activity (Figure 4) delivered by PTW, with test certificate, was used and a < 1 μsv/h dose at 10 cm distance was predicted. Table 4 includes the dosimetric measurement results made with the source placed to 10 cm distance from the secondary standard chambers oriented on a measurement direction of the secondary standard chamber detector. 321

Table 3. The results of measurements Set 1 (Without radiation beam) Set 3 (Electron Set 5 (Bremsstrahlung operation) operation ) Hp(10) [μsv/min] 1.23 1.53 1.637 Hp(10) [μsv/min] Hp(10) [μsv/min] 6 3 Hp(10) [μsv/10 min] 12.3 Hp(10) [μsv/10 15.3 Hp(10) [μsv/10 16.34 6 min] 3 min] Personal Dose 74.1 91.9 98.24 Equivalent,Hp(10) 6 8 Tissue Equivalent Dose for public / training / occupational exposure, H T = 500 /150 /50 Σ R w R x D T,R msv/an Effective Dose for public / training / occupational exposure, E = Σ T w T x H, (ICRP 60) 20 / 6 / 1 msv/an T Set 2 (Without radiation beam) Set 4 (Electron operation) Set 6 (Bremsstrahlung operation ) Hp(10) [μsv/min] 1.21 Hp(10) [μsv/min] 1.43 Hp(10) [μsv/min] 1.660 Hp(10) [μsv/10 min] 12.1 Hp(10) [μsv/10 14.3 Hp(10) [μsv/10 16.60 8 min] 2 min] Personal Dose 73.1 85.9 99.65 Equivalent,Hp(10) 1 2 Tissue Equivalent Dose for public/ training / occupational exposure, H T = 500/150/50 Σ R w R x D T,R msv/an Effective Dose for public / training / occupational exposure, E = Σ T w T x H T, (ICRP 60) 20 / 6 / 1 msv/an The measurement was made with the housing the source, open, closed and the last measurement was recorded in absence of the source. It was noticed that when the source was absent, the obtained value was smaller than the case in which the was closed but both values were smaller than in case in which the was open. Table 4. Obtained values for personal / ambient dose equivalent, using Sr-90 of 20 MBq. Measure mode Personal dose equivalent, Measure Ambient dose equivalent Hp(10) mode H*(10) Without 1.05 [μsv/min] = 63.0 Without 0.70 [μsv/h] source [μsv/h] source Close 1.41 [μsv/min] = 84,6 Open 3.2 [μsv/h] ; SSD = 0 cm [μsv/h] Open 1.44 [μsv/min] = 86,4 Open 1.7/ 0.5 [μsv/h] ; SSD = [μsv/h] 10/100 cm 7. Conclusions The paper presents the method to measure the ambient/personal dose equivalent and the dosimetric results, the measurements being conducted by the use of a secondary standard chamber Hp(10) 34035 type. After conducting the measurements in the room of the hall in which the 7 MeV linear electron accelerators is located, in INFLPR, it was found that the obtained values corresponding to the equivalent dose for occupational exposed individuals, 322

sh ow small values which fall-in CNCAN allowable limits. As an average, the obtained values are 3.76 times lower than the allowable equivalent dose limit for occupational exposure and 14.26 times higher than the allowable equivalent dose limit for population exposure. The obtained values, after the ambient dose equivalent measure, decrease with the distance between source and H*(10) secondary standa rd ionization chamber detector. For the measurements made with closed, is realized the CNCAN condition about the ambient dose equivalent that must be smaller than 1 μsv/h at 0,1 m distance. References [1] Ankerhold, U, Radiat. Prot. Dosim. 123 (2), 137 142, 2007. [2] Ankerhold, U., Behrens, R., Ambrosi, P., Radiat. Prot. Dosim. 86 (3), 167-173, 1999. [3] Ankerhold, U., Behrens, R. and Ambrosi, P., PTB Bericht, PTB-Dos-34, ISBN 3-89701- 513-7, 2000. [4] International Standard ISO 4037-3, X and Gamma Reference Radiations for Calibrating Dosemeters and Dose rate Meters and for Determining their Response as a Function of Photon Energy Part 3: Calibration of Area and Personal Dosemeters and the Measurement of their Response as a Function of Energy and Angle of Incidence, 1997. [5] Calibration Certificate, no. 0801543 for Ionization Chamber type TN32002-0278, PTW- Freiburg, Physikalisch-Technische Werkstatten, Germany, 16.05.2008. [6] Calibration Certificate, no. PTB-6.3-4033197 for Ionization Chamber type T34035 Hp(10), Physikalisch-Technische Bundesanstalt Braunschweig und Berlin, Germany, 04.12.2007. 323