PHITS calculation of the radiation field in HIMAC BIO

Similar documents
Development of Active Space Radiation Detector, A-DREAMS-2 at NIRS

InterComparison for Cosmic-ray with Heavy Ion Beams At NIRS

THE mono-energetic hadron beam such as heavy-ions or

Republic. Sept. 3rd th WRMISS in Budapest 1

Contribution of the Nuclear Physics Institute of the AS CR to the BEOBAL Project

NSRL-ICCHIBAN Brief Report, ICCHIBAN-7&8 Announcement and Future ICCHIBAN Experiments

Simulation of light ion transport in a water phantom using Geant4.

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

František SPURNÝ, Iva JADRNÍCKOVÁ. Nuclear Physics Institute Department of Radiation Dosimetry, Academy of Sciences of the Czech Republic, Prague

Measurement of induced radioactivity in air and water for medical accelerators

Outline. Physics of Charge Particle Motion. Physics of Charge Particle Motion 7/31/2014. Proton Therapy I: Basic Proton Therapy

DOSIMETRY ON THE FOTON M2/BIOPAN-5 SATELLITE

Calculation of Bubble Detector Response Using Data from the Matroshka-R Study

The heavy ion irradiation facility at KVI-CART

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Portable, Low-cost Proportional Counters for Space, Atmospheric and Ground based Applications

Physics with Exotic Nuclei

Physics of particles. H. Paganetti PhD Massachusetts General Hospital & Harvard Medical School

CRaTER Science Requirements

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Detection and measurement of gamma-radiation by gammaspectroscopy

LONG-TERM MONITORING OF THE ONBOARD AIRCRAFT EXPOSURE LEVEL WITH Si-DIODE BASED SPECTROMETER ) F. Spurný (1), Ts. Dachev (2)

External MC code : PHITS

Characterization of Heavy Ion Beams at the Heavy Ion Medical Accelerator in Chiba Using a Li-drifted 5-mm Silicon Detector

Emphasis on what happens to emitted particle (if no nuclear reaction and MEDIUM (i.e., atomic effects)

III. Energy Deposition in the Detector and Spectrum Formation

Simulations of MATROSHKA-R experiment 2006 at the ISS using PHITS

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in.

Energy Dependence of Biological Systems Under Radiation Exposure

The interaction of radiation with matter

R&D of emulsion technology to study fragment interaction to improve ion therapy

Advacam, Prague, Czech Republic. Nuclear Physics Institute, Czech Academy of Sciences, Prague, Czech Republic

Radiation (Particle) Detection and Measurement

Simulation Studies for a Polarimeter at the International Linear Collider (ILC)

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Monte Carlo Simulation concerning Particle Therapy

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

China high-intensity accelerator technology developments for Neutron Sources & ADS

Yukio Uchihori, Nakahiro Yasuda, Hisashi Kitamura, Iva Jadrnickova, Eric R. Benton, A. Shurshakov on behalf of ICCHIBAN Working Group and

EEE4106Z Radiation Interactions & Detection

Fast detectors for Mössbauer spectroscopy )

Characterization of heavy charged particle fields using fluorescent nuclear track detectors

Detectors in Nuclear Physics (48 hours)

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Gy can be used for any type of radiation. Gy does not describe the biological effects of the different radiations.

Elastic Recoil Detection Method using DT Neutrons for Hydrogen Isotope Analysis in Fusion Materials. Abstract

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Chapter V: Cavity theories

Safety Co-ordinator : Patrick Walden, TRIUMF, ext : 7340

Development of Ring-Imaging Cherenkov Counter for Heavy Ions

Synchrotron Radiation a Tool for Precise Beam Energy Measurements at the ILC

Development of beam delivery systems for proton (ion) therapy

MONTE CARLO SIMULATION FOR EVALUATION OF DOSE DISTRIBUTION IN PROTON THERAPY *

Introduction to Accelerator Physics Part 1

Progress in measuring GMR in unstable nuclei: Decay detector calibration and inverse reaction experiment. J. Button, Y.-W. Lui, and D.H.

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY

pp physics, RWTH, WS 2003/04, T.Hebbeker

Práctica de laboratorio número 6: Non-Rutherford scattering near the MeV 12 C(p,p) 12 C resonance

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

REALISTIC NEUTRON SPECTRA FOR RADIATION PROTECTION AND OTHER APPLICATIONS AT AERI, BUDAPEST

Radiation Protection At Synchrotron Radiation Facilities

Improvements and developments of physics models in PHITS for radiotherapy and space applications

Neutron Imaging at Spallation Neutron Sources

Radiation hardness studies of the Shashlyk DCS parts:

JRPR. Measurement of Neutron Production Doubledifferential Cross-sections on Carbon Bombarded with 430 MeV/Nucleon Carbon Ions.

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

AIRFLY: Measurement of the Air Fluorescence induced by electrons

EXPERIMENTAL STUDY OF NEUTRON FIELDS PRODUCED IN PROTON REACTIONS WITH HEAVY TARGETS. Nuclear Physics Institute AS CR, Rez Czech Republic

Characterization of 3D thermal neutron semiconductor detectors

RITU and the GREAT Spectrometer

Planning and preparation approaches for non-nuclear waste disposal

Radiological Issues at JLab

Lomonosov Moscow State University. NUCLEON Chemical Composition and Energy Spectra of Cosmic Rays at TeV

The scanning microbeam PIXE analysis facility at NIRS

1.4 The Tools of the Trade!

Bubble Detector Characterization for Space Radiation

CHARGED PARTICLE INTERACTIONS

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Calibration and applications of modern track detectors CR-39/PM-355 in nuclear physics and high temperature plasma experiments

Interaction of Ionizing Radiation with Matter

arxiv: v2 [physics.med-ph] 29 May 2015

Contents. Charged Particles. Coulomb Interactions Elastic Scattering. Coulomb Interactions - Inelastic Scattering. Bremsstrahlung

Overview and Status of the Austrian Particle Therapy Facility MedAustron. Peter Urschütz

Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI)

Calorimetry I Electromagnetic Calorimeters

Towards Proton Computed Tomography

Radiation damage calculation in PHITS

The photoneutron yield predictions by PICA and comparison with the measurements

S364 Experiment: Resonance in Isobaric Charge-Exchange reactions

Radiation Detection and Measurement

A Comparison between Channel Selections in Heavy Ion Reactions

PREX Simulation Update

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Transcription:

PHITS calculation of the radiation field in HIMAC BIO Ondřej Ploc, Yukio Uchihori, Hisashi Kitamura, Lembit Sihver National Institute of Radiological Sciences, Chiba, Japan Nuclear Physics Institute, Prague, Czech Republic Chalmers University of Technology, Gothenburg, Sweden

Outline Introduction Beam-line description Depth-dose curves Beam profile Example of application Silicon detector calibration using only one heavy ion beam Fragments contribution LET spectra Conclusions 2

Introduction HIMAC - Heavy Ion Medical Accelerator in Chiba Three irradiation rooms for experiments: (i) General Physics, (ii) Biology (BIO), (iii) Medium energy http://www.nirs.go.jp/eng/research/charged_particle/index.shtml 3

Introduction HIMAC BIO is an irradiation room used for heavy ion experiments related to biology and physics Goals to provide detailed information about the components of the beam line at HIMAC BIO to express the exact beam composition via fragment contribution to absorbed dose calculated by PHITS at the location of irradiated samples To compare PHITS simulations with measurements using a semiconductor detector Liulin 4

HIMAC BIO Advantage: Bragg curve is well defined Broad parallel beam (φ 10cm) Flat beam profile Disadvantage: Fragments and secondary particles because beam goes through scatterer (different for different ions) and 6.7m of air Ion+nominal energy in MeV/u: He 150, C 135, C 290, C 400, Ne 400, Si 490, Fe 500 Measurements performed behind different thicknesses of PMMA filters 9 filters are available (in mm): 0.5, 1, 2, 4, 8, 16, 32, 64, 128 5

HIMAC BIO geometry used for PHITS calculations Synchrotron room 6

HIMAC BIO geometry used for PHITS calculations 7

HIMAC BIO geometry used for PHITS calculations Ring collimator Four leaf collimator (FLC) Al Range shifter Brass collimators Beam dump samples Wobbler magnets Beam profile monitor Beam exit Scatter filters 8

Beam exit HIMAC BIO, beam line Synchrotron room F-collimator Vacuum tube Ring collimator Vacuum tube 2 Main monitor + SEM Beam profile monitor Experimental room Four leaf collimator (FLC) Al Range shifter Brass collimators samples Wobbler magnets 1055 892 1172 Materials on the beam line Scatter filters 540 281 30.5 Dump Thickness / cm Water equivalent thickness / cm Air 668.8 0.8 Aluminum windows (11) 0.011 0.03 Scatter filters (Ta or Pb) Variable, depends on ion and energy Range shifter (PMMA) Variable 9

Scatter filters Scatter filters Ta Pb 1 2 3 4 5 6 7 8 Brass F-collimator φ out 16cm φ in 10cm Ion and energy / MeV/u Ta / mm Pb / mm Water eq. / mm He 150 mono 0.215 0 3.57 C 135 mono 0.215 0 3.57 C 290 mono 0.434 0 7.20 C 290 SOBP 0.649 0 10.77 C 400 mono 0.754 0 12.52 Ne 230 mono 0.539 0 8.95 Ne 400 mono 1.02 0 16.93 Si 490 mono 0.434 1.6 25.35 Fe 500 mono 0.215 1.6 21.72 Al holders 10

Scatter filters Configuration for Ne 400 MeV/u, all charged particles The density of energy distribution is visibly higher behind the scatter filters 11

Scatter filters Configuration for Ne 400 MeV/u, Oxygen ions only 12

Scatter filters Configuration for Ne 400 MeV/u, He only 13

Measured and calculated Bragg curves Measurement with Markus ionization chamber behind different thicknesses of PMMA filters absorbed dose / Gy 3.0E-04 2.5E-04 2.0E-04 1.5E-04 1.0E-04 5.0E-05 C290 mono PHITS calculation IC measurement Real energy(phits): 278.7 MeV/u 0.0E+00 0 20 40 60 80 100 120 140 Depth in PMMA [mm] Calculated Bragg curves fitted well to the measured ones => well designed geometry was confirmed 14

Range and beam energies in HIMAC BIO Ion Nominal energy / (MeV/u) Range in PMMA / cm Measured with IC Calculated with PHITS Beam energy in front of PMMA filters / (MeV/u) SRIM PHITS He 150 12.63 12.58 145.35 145.5 C 135 2.73 2.91 112.54 114.0 C 290 12.83 12.61 274.43 278.7 C 400 22.23 22.01 382.85 389.7 Ne 400 12.48 12.48 370.05 373.8 Si 490 11.88 11.77 445.68 447.0 Differences in range < 2 mm and in energies 2% 15

Beam profile Experiments: by Hisashi Kitamura 2003-2007 Markus ionization chamber He, C, Ne, Si, Ar, Fe, Kr, Xe Several energies, diameters, binary filters for each ion PHITS calculations: Scatterers as described above Wobbler magnets 2 dipole electromagnets to change the beam diameter in x and y axis Magnetic field intensity: 0.5 kg Radius of gap: 10cm 16

Beam profile 105% C 290 MeV/u 100% 95% 90% 85% IC PHITS 80% 75% -70-50 -30-10 10 30 50 70 x / mm PHITS simulation: transverse dose distribution at the front surface of water column 17

Liulin Energy deposition spectrometer Active volume: silicon diode (2 1 0.03 cm 3 ) Size: 98 65 25 mm 3 (MDU01&02) 103 103 47 mm 3 (MDU07) Absorbed dose calculation D 1 = m 256 Si N i i= 1 ε ε i is energy deposition i 18

Energy deposition calibration of Liulin Original method: Energy deposition ε i = i 81.3keV i ADC channel number 19

Comparison of two identical Liulins MDU 1 Front aluminum cover removed from both Liulins PE desk removed from MDU 1 MDU 2 180 160 140 C 400 MeV/u Beam: C 400 MeV/u PMMA: 0.0 mm 86.0 mm 178.5 mm 208.5 mm D / µgy 120 100 80 60 40 20 0 0 50 100 150 200 250 Thickness of PMMA filters / mm 20

Identical Liulins in C 400MeV/u Relative number of events 1.0E+00 1.0E-01 1.0E-02 1.0E-03 1.0E-04 1.0E-05 1.0E-06 Liulins in C 400MeV/u, 0mm of PMMA 79.5 84.0 4.5 MDU1 MDU2 0 50 100 150 200 250 ADC Channel number Relative number of events 1.0E+00 1.0E-01 1.0E-02 1.0E-03 1.0E-04 1.0E-05 1.0E-06 Liulins in C 400MeV/u, 86mm of PMMA 95.3 100.5 5.2 MDU1 MDU2 0 50 100 150 200 250 ADC Channel number 1.0E+00 Liulins in C 400MeV/u, 178.5mm of PMMA 1.0E+00 Liulins in C 400MeV/u, 208.5mm of PMMA Relative number of events 1.0E-01 1.0E-02 1.0E-03 1.0E-04 1.0E-05 146.5 7.8 154.3 MDU1 MDU2 Relative number of events 1.0E-01 1.0E-02 1.0E-03 1.0E-04 12.9 235.1 248.0 MDU1 MDU2 1.0E-06 0 50 100 150 200 250 ADC Channel number 1.0E-05 0 50 100 150 200 250 ADC Channel number 21

Recalibration of Liulin using PHITS 22

Recalibration of Liulin using PHITS Gaussian fits of all peaks (primary C, fragments) 0.0 mm 86.0 mm 175.5 mm 208.5 mm 23

Recalibration of Liulin using PHITS 300 MDU 1 (no Al, no PE) Energy deposition / MeV 250 MDU 2 (no Al) Energy deposition / MeV ADC channel number 250 200 150 100 50 linear fit y = 13.465x - 0.4254 ADC channel number 200 150 100 50 linear fit y = 12.713x + 0.2618 0 0 0 5 10 15 20 Energy deposition / MeV 0 5 10 15 20 Energy deposition / MeV 24

Comparison of two identical Liulins 1.0E+00 Liulins in C 400MeV/u, 0mm of PMMA 1.0E+00 Liulins in C 400MeV/u, 86mm of PMMA Relative number of events 1.0E-01 1.0E-02 1.0E-03 1.0E-04 1.0E-05 Primary C Pile-up MDU1 MDU2 Relative number of events 1.0E-01 1.0E-02 1.0E-03 1.0E-04 1.0E-05 MDU1 MDU2 1.0E-06 0 5 10 15 20 25 Energy deposition / MeV 1.0E-06 0 5 10 15 20 25 Energy deposition / MeV 1.0E+00 Liulins in C 400MeV/u, 178.5mm of PMMA 1.0E+00 Liulins in C 400MeV/u, 208.5mm of PMMA 1.0E-01 MDU1 1.0E-01 MDU1 Relative number of events 1.0E-02 1.0E-03 1.0E-04 1.0E-05 MDU2 Relative number of events 1.0E-02 1.0E-03 1.0E-04 MDU2 1.0E-06 0 5 10 15 20 25 Energy deposition / MeV 1.0E-05 0 5 10 15 20 25 Energy deposition / MeV 25

Comparison of Liulin spectra with PHITS 1.0E+00 Liulins in C 400MeV/u, no BF 1.0E+00 Liulins in C 400MeV/u, 86.0 mm 1.0E-01 1.0E-01 Relative number of events 1.0E-02 1.0E-03 1.0E-04 PHITS MDU2 Relative number of events 1.0E-02 1.0E-03 1.0E-04 PHITS MDU2 1.0E-05 1.0E+00 0 5 10 15 20 25 Energy deposition / MeV Liulins in C 400MeV/u, 178.5 mm 1.0E-05 1.0E+00 0 5 10 15 20 25 Energy deposition / MeV Liulins in C 400MeV/u, 208.5 mm 1.0E-01 1.0E-01 Relative number of events 1.0E-02 1.0E-03 1.0E-04 PHITS MDU2 Relative number of events 1.0E-02 1.0E-03 1.0E-04 PHITS MDU2 1.0E-05 0 5 10 15 20 25 Energy deposition / MeV 1.0E-05 0 5 10 15 20 25 Energy deposition / MeV 26

Liulin in Ne 400 MeV/u 180 160 Ne 400 MeV/u 1.0E+00 Liulin MDU2 in Ne 400MeV/u D / µgy 140 120 100 80 60 40 20 0 0 20 40 60 80 100 120 140 Thickness of PMMA filters / mm Relative number of events 1.0E-01 He 1.0E-02 Be 1.0E-03 1.0E-04 B C 0mm 1mm 51.5mm N O 0 5 10 15 20 25 Energy deposition / MeV F Ne 1.0E+00 Liulin in Ne 400MeV/u, 51.5mm of PMMA Relative number of events 1.0E-01 1.0E-02 1.0E-03 1.0E-04 1.0E-05 MDU2 PHITS 0 10 20 30 40 27 Energy deposition / MeV

Contribution of fragments to total dose at the sample location Absorbed dose contribution 100 10 1 0.1 1.68 1.28 0.93 0.0mm 26.5mm 90.5mm C 290MeV/u 4.61 2.67 2.08 1.95 1.51 1.05 0.88 1.02 0.55 0.55 0.37 0.28 0.26 0.28 0.17 0.10 0.07 0.07 97.9494.85 88.90 0.01 p d H He Li Be B C 28

Contribution of fragments to total dose at the sample location Absorbed dose contribution 100 10 1 0.1 0.0mm 51.5mm 94.5mm 1.12 1.31 1.32 1.57 1.56 2.33 0.56 0.62 0.27 0.20 0.15 0.05 0.05 0.03 Ne 400MeV/u 0.50 0.40 0.32 0.23 0.22 0.26 0.22 0.25 0.09 97.78 85.79 77.49 7.21 3.90 2.80 3.81 3.72 1.87 2.24 2.42 0.39 0.30 0.01 p d H He Li Be B C N O F Ne 29

Evaluation of LET spectra with Liulin L w = ksi / w L ε LSi = d d = 300µ m Si ρw ρ Si Ion Energy / (MeV/u) Water was used instead of Si-diode! k Si/w (Si to water conv. Coefficient) He 150 1.36 C 135 1.23 C 290 1.24 C 400 1.26 Ne 400 1.24 Liulin E dep and LET spectra calculated with PHITS, C 135 MeV/u 30

Conclusions Detail simulation of geometry of HIMAC BIO was developed Calibration of Liulin can be done using only one heavy ion beam Liulin is capable to detect fragments; the difference between measured and calculated energy deposition spectra in peaks can be important, more research on this topic is in process The contribution of fragments to total absorbed dose is high, from 2% to 23% depending on primary heavy ion, energy and thickness of PMMA filters Si to water conversion coefficient estimated using PHITS code differ depending on ion and energy (1.24 1.36) Further research on calculation of neutron contribution in the radiation field is needed 31

THANK YOU FOR YOUR ATTENTION! 32