Shabir: Simultaneous HPLC Determination of Preservatives in Gels

Similar documents
VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Simultaneous Determination of Preservatives (Methyl Paraben and Propyl Paraben) in Sucralfate Suspension Using High Performance Liquid Chromatography

Journal of Pharmaceutical and Scientific Innovation

Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation)

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Development and Validation of RP-HPLC Method for the Determination of Methamphetamine and Propranolol in Tablet Dosage Form

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68

Available online at Scholars Research Library

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF GABAPENTIN IN PURE AND PHARMACEUTICAL FORMULATIONS

Dissolution study and method validation of alprazolam by high performance liquid chromatography method in pharmaceutical dosage form

Validated RP-HPLC Method for Estimation of Cefprozil in Tablet Dosage Form

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

KEYWORDS: Acetaminophen, Doxylamine succinate, Dextromethorphan hydrobromide.

Method Development and Validation for the Estimation of Darunavir in Rat Plasma by RP-HPLC

International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC

Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form

Impact factor: 3.958/ICV: 4.10 ISSN:

Mashhour Ghanem 1 and Saleh Abu-Lafi 2 * ABSTRACT ARTICLE INFO

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ].

STABILITY INDICATING RP HPLC METHOD FOR ANALYSIS OF DORZOLAMIDE HCl IN THE BULK DRUG AND IT S PHARMACEUTICAL DOSAGE FORM

Development and validation a RP-HPLC method: Application for the quantitative determination of quetiapine fumarate from marketed bulk tablets

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation

RP-HPLC Method Development and Validation of Dapagliflozin in Bulk and Tablet formulation

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Available online at Scholars Research Library

Development and Validation of Stability Indicating RP-HPLC Method for the Determination of Anagrelide HCl in Pharmaceutical Formulation

Appendix II- Bioanalytical Method Development and Validation

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

International Journal of Pharma and Bio Sciences V1(1)2010. HPLC method for analysis of Lercanidipine Hydrochloride in Tablets

Journal of Chemical and Pharmaceutical Research, 2017, 9(10): Research Article

INTERNATIONAL JOURNAL OF UNIVERSAL PHARMACY AND BIO SCIENCES

Volume 6, Issue 2, January February 2011; Article-015

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Scholars Research Library. Rapid and sensitive RP-HPLC analytical method development and validation of Pioglitazone hydrochloride

Department of Quality Assurance, Luqman College of Pharmacy, GULBARGA (K.S.) INDIA ABSTRACT

INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND LIFE SCIENCES

Research Article Available online at

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Rapid and simultaneous determination of paracetamol, ibuprofen and related impurity of ibuprofen by UPLC/DAD

Journal of Chemical and Pharmaceutical Research

Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form

Validated First Order Derivative Spectroscopic Method for the determination of Stavudine in Bulk and Pharmaceutical Dosage Forms

International Journal of Current Trends in Pharmaceutical Research. International Journal of Current Trends in Pharmaceutical Research

CHAPTER - 3 ANALYTICAL PROFILE. 3.1 Estimation of Drug in Pharmaceutical Formulation Estimation of Drugs

Validated HPLC Methods

Impact factor: 3.958/ICV: 4.10 ISSN:

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Journal of Drug Delivery and Therapeutics

Chapter 4: Verification of compendial methods

HPLC Method Development and Validation for the estimation of Esomeprazole in Bulk and Pharmaceutical Dosage Form

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

DEVELOPMENT AND VALIDATION OF NEW ANALYTICAL METHOD FOR BIOACTIVE COMPOUNDS

Development and Validation of Stability-Indicating RP-HPLC Method for Estimation of Atovaquone

METHOD DEVELOPMENT AND VALIDATION OF RALTEGRAVIR POTASSIUM AND RILPIVIRINE HCL BY HPLC AND HPTLC METHODS

Novus International Journal of Analytical Innovations 2012, Vol. 1, No. 3

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

International Journal of Pharmaceutical Research & Analysis

Simultaneous HPLC Determination of Methocarbamol, Paracetamol and Diclofenac Sodium

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION

Isocraticc Reverse Phase HPLC Method-Determination and Validation of Cilostazol

Department of Chemistry, JNTUACE, Kalikiri

RP-HPLC METHOD FOR THE ESTIMATION OF ZIPRASIDONE

Pelagia Research Library

Journal of Pharmaceutical and Biomedical Analysis Letters. Analysis Letters

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Chapter-4 EXPERIMENTAL WORK BY RP-HPLC

Development and Validation of RP-HPLC Method for the Estimation of Repaglinide in Bulk Drug and Pharmaceutical Formulation

Pelagia Research Library

CHAPTER - IV. Acharya Nagarjuna University, Guntur 105

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 9(1), 35-42

Fast methods for the determination of ibuprofen in drug products

Development and validation of RP-HPLC method for determination of marker in polyherbal marketed Kankasava formulations

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Development and validation of RP-LC method for lisinopril dihydrate in bulk and its pharmaceutical formulations

with diode array detection

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

INTERNATIONAL JOURNAL OF ADVANCED PHARMACEUTICAL SCIENCES

Development and validation of stability indicating reverse phase high performance liquid chromatography method for Timolol Maleate

Sumathi Gracy et al. Int. Res. J. Pharm. 2014, 5 (4) INTERNATIONAL RESEARCH JOURNAL OF PHARMACY

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Method development and validation of Tinidazole and Ciprofloxacin HCl in bulk and tablet dosage form by Rp-HPLC

Journal of Chemical and Pharmaceutical Research

SIMULTANEOUS ESTIMATION OF MEFENAMIC ACID AND PARACETAMOL IN SUSPENSION FORM BY USING UPLC

ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF IVABRADINE HCL IN BULK AND FORMULATION

Analytical Method Development and Validation of Lafutidine in Tablet dosage form by RP-HPLC

Dissolution Test 5 was validated using a Zodiac C18 brand of L1 column. The typical retention time for atorvastatin is about min.

Transcription:

Research Papers A New Validated HPLC Method for the Simultaneous Determination of 2-phenoxyethanol, Methylparaben, Ethylparaben and Propylparaben in a Pharmaceutical Gel G. A. SHABIR* xford Brookes University, School of Life Sciences, Headington Campus, xford, X3 0BP, UK Shabir: Simultaneous HPLC Determination of Preservatives in Gels A novel reversed-phase HPLC method has been developed and validated for the simultaneous determination of 2-phenoxyethanol, methylparaben, ethylparaben and propylparaben preservatives. The method uses a Lichrosorb C8 (150 4.6 mm, 5 m) column and isocratic elution. The mobile phase consisted of a mixture of acetonitrile, tetrahydrofuran and water (21:13:66, v/v/v), pumped at a flow rate of 1 ml/min. The UV detection was set at 258 nm. The method was validated with respect to accuracy, precision (repeatability and intermediate precision), specificity, linearity and range. All the parameters examined met the current recommendations for bioanalytical method validation. The developed method was successfully applied to the determination of commercially available pharmaceutical gel products for these preservatives. The procedure describes here is simple, selective and reliable for routine quality control analysis and stability tests. Key words: 2-phenoxyethanol, ethylparaben, HPLC, method validation, methylparaben, pharmaceutical gel, propylparaben Best practices in method development and validation are equally important in the analysis of both active components and preservatives (excipients, inactive components) used in manufacturing of drug products. 2-Phenoxyethanol (C 8 H 10 2, fig. 1a, PhE), methylparaben (C 8 H 8, fig.1b, MP), ethylparaben (C 9 H 10, fig. 1c, EP), and propylparaben (C 10 H 12, fig. 1d, PP) are used in single or in combinations in drug, cosmetic and food formulations as antimicrobial preservatives to prevent alteration of product preparations [1]. Formulators must be fully aware of the procedure for preservative systems in a product need to be analysed to establish their effectiveness throughout shelf life of the product [2]. Many existing analytical procedures are available in literature for the determination of present preservatives studied, either alone or in combination with other drugs by HPLC and other techniques [3-16]. Such a method is important as there seem to be an increasing trend in using combination of preservatives, not only in pharmaceutical formulations but also in food *Address for correspondence E-mail: gshabir.oxford@yahoo.com a c H industry and cosmetic products. Moreover, many of the reported methods use complicated and labour-intensive pre-treatment procedures such as steam distillation, multiple-steps and solid phase extractions. Therefore, the purpose of the present study was to develop and validate a new, simple, accurate, and robust reversed-phase HPLC method for the determination of PhE, MP, EP and PP preservatives in a single chromatographic run suitable for preservatives raw materials, bulk gel b d Fig. 1: Molecular structures of the separated preservatives The separated preservatives were (a) 2-phenoxyethanol (PhE), (b) methylparaben (MP), (c) ethylparaben (EP) and d. propylparaben (PP) Indian Journal of Pharmaceutical Sciences 421

samples and finished gel products release. Validating analytical method is a crucial part of successful product development, testing and quality. The determination of preservatives both alone or in formulated products is important and provides a difficult analytical challenge. As a best practice [17-19], in the subsequent investigation, the new and simple reversed-phase HPLC assay method was validated [20] for linearity, precision (repeatability and intermediate precision), accuracy, specificity and robustness. The developed and validated method was applied to the analysis of these preservatives in commercially available pharmaceutical gel products. MATERIALS AND METDS HPLC-grade acetonitrile and tetrahydrofuran (THF) were obtained from Sigma-Aldrich (Gillingham, UK). 2-Phenoxyethanol (pure> 99%), methylparaben (pure> 99%), ethylparaben (pure> 99%), propylparaben (pure> 99%) and formic acid were also purchased from Sigma-Aldrich (Gillingham, UK). Distilled water was de-ionised by using a Milli-Q system (Millipore, Bedford, MA). The Knauer HPLC system (Berlin, Germany), consisted of a Knauer pump model 1000, autosampler model 3950, photodiode-array (PDA) detector model 2600 and a vacuum degasser, all controlled by a ClarityChrom software, was used. RP-HPLC analysis was performed isocratically at ambient temperature using a Lichrosorb C8 (150 4.6 mm, 5 μm) column (Jones Chromatography, Hengoed, UK). The mobile phase consisted of a mixture of acetonitrile/ tetrahydrofuran/water (21:13:66, v/v/v) adjusted to ph 3.0±0.05 with formic acid was used. The flow rate was 1 ml/min and injection volume was 10 μl. The eluent was monitored with a UV detector set at 258 nm. All samples were diluted with mobile phase. Preparation of the standard and sample solutions: A combined standard stock solution of accurately weighted preservatives PhE (1.5 g), MP (0.290 g), EP (0.07 g) and PP (0.036 g) was prepared in 100 ml volumetric flask and dissolved in mobile phase (stock). Five millilitre aliquots of PhE, MP, EP and PP stock solution were added to a 100 ml volumetric flask, and diluted in mobile phase, yielding a final concentration of 750, 145, 35 and 18 μg/ml, respectively. An accurately weighed amount (1.0 g) of gel sample was placed in a 100 ml volumetric flask and dissolved in methanol. Ten millilitre aliquot solution was added to a 100 ml volumetric flask and diluted in 50 ml methanol and volume made up with mobile phase. Validation of the method: The linearity test was performed using five different amounts of PhE, MP, EP and PP in the range 650-850 μg/ml, 45-245 μg/ml, 20-50 μg/ml and 6-30 μg/ml, respectively. Solutions corresponding to each concentration level were injected in duplicate and linear regression analysis of the PhE, MP, EP and PP peak area (y) versus PhE, MP, EP and PP concentration (x) were calculated. Precision of the method was determined by repeatability (intra-day) and intermediate precision (inter-day variation). Repeatability was examined by analysing six determinations of the same batch of each preservative at 100% of the test concentration. The samples were stored at 30 for 15 days. The RSD of the areas of preservative peak were calculated. Intermediate precision (inter-day variation) was studied by assaying five samples containing the nominal amount of PhE, MP, EP and PP on different days. Solutions corresponding to each concentration level were injected in duplicate. The RSD values across the system were calculated. Recovery studies may be performed in a variety of ways depending on the composition and properties of the sample matrix. In the present study, three different solutions were prepared with a known added amount of pure PhE, MP, EP and PP compounds to give a concentration range of 50-150% of that in a test preparation. These solutions were injected in triplicate and percent recoveries of response factor (area/concentration) were calculated. The stability analytical solutions were also evaluated. Sample solutions chromatographed immediately after preparation and then re-assayed after storage at room temperature for 48 h. RESULTS AND DISCUSSIN The chromatographic separation of PhE, MP, EP and PP preservatives was carried out in the isocratic mode using a mixture of acetonitrile, tetrahydrofuran and water ph 3.0±0.5 (21:13:66, v/v/v) as mobile phase. The column was equilibrated with the mobile phase flowing at 1 ml/min for about 30 min prior to injection. The column temperature was ambient. Ten microlitres of standards solutions were injected automatically into the column. Subsequently, the liquid chromatographic behaviours of preservatives 422 Indian Journal of Pharmaceutical Sciences

were monitored with a PDA UV detector at 258 nm. Additionally, preliminary system suitability, precision, linearity, robustness and stability of solutions studies performed during the development of the method showed that the 10 μl injection volume was reproducible and the peak response was significant at the analytical concentration chosen. Chromatograms of the resulting solutions gave excellent separation and resolution (fig. 2). System suitability test was developed for the routine application of the assay method. Prior to each analysis, the chromatographic system must satisfy suitability test requirements (resolution and repeatability). Peak-to-peak resolution, between each peak measured on a reference solution must be above 1.5. System suitability test was performed from ten Fig. 2: Typical chromatogram of standard mixture of four preservatives PhE eluted at tr: 4.50 min; MP at tr: 6.00 min; EP at tr: 9.05 min and PP at tr: 15.03 min replicate injections of a solution containing 750, 145, 35 and 18 μg PhE, MP, EP and PP/ml, respectively. All peaks were well resolved and the precision of injections for all preservative peaks were acceptable. The percent relative standard deviation (RSD) of the peaks area responses were measured, giving an average between 0.13% and 0.34% (n=10). The tailing factor (T), capacity factor (K) and theoretical plate number (N) were also calculated. The results of system suitability in comparison with the required limits are shown in Table 1. The proposed method met these requirements within the accepted limits [21,22]. For the determination of method robustness within a laboratory a number of chromatographic parameters were evaluated during method development, such as flow rate, column temperature, mobile phase composition and ph, columns from different batches, and the quantitative influence of the variables were determined. For each parameter studied two injections of standard solutions were chromatographed. In all cases the influence of the parameters were found within a previously specified tolerance range. This shows that the method for determination of PhE, MP, EP and PP was reproducible and robust. Calibration curves were linear over the concentration range of 650-850 μg/ml for PhE, 45-245 μg/ml for MP, 20-50 μg/ml for EP and 6-30 μg/ml for PP. The results are presented in Table 2 and show a good correlation between the peak area of analytes and concentration with r > 0.9998. The % RSD were found to be less than 0.22 and 0.52 for intra-day and TABLE 1: SYSTEM SUITABILITY RESULS F THE PRPSED ANALYTICAL METD Parameters Recommended limits Results PhE MP EP PP Retention time (min) - 4.50 6.00 9.05 15.03 Injection repeatability* RSD 1 (%, n 5) 0.34 0.15 0.13 0.26 Resolution (Rs) R s 1.5-10.06 30.70 60.00 Capacity factor (K ) 2 4.54 5.34 6.45 9.67 Tailing factor (T) 2 1.125 1.100 1.133 1.217 Theoretical plate number (N) 2000 7245 8646 7654 8814 *Ten replicate injections. TABLE 2: LINEARITY ASSESSMENT F THE HPLC METD FR THE ASSAY F FUR PRESERVATIVES Components Concentration (μg/ml) Equation for regression line R 2 PhE 650-850* y = 12432x 7483.2 0.9999 MP 45-245 y = 53084x 179.38 0.9999 EP 20-50 y = 67400x 515 0.9998 PP 6-30 y = 69667x 361.2 0.9998 *K=5; n=2 Indian Journal of Pharmaceutical Sciences 423

TABLE 3: METD VALIDATIN RESULTS FR FUR PRESERVATIVES Validation steps Parameters Results Acceptance PhE MP EP PP criteria Repeatability RSD (%, n = 6) 0.307 0.097 0.317 0.283 X 2 Int. precision Day 1 RSD (%) 0.329 0.130 0.365 0.401 X 2 Day 2 RSD (%) 0.242 0.165 0.417 0.372 X 2 Standard stability (48 h data) Change in response factor (%) 0.13 0.13 0.13 0.15 X 2 System suitability RSD (%, n = 6) 0.11 0.15 0.09 0.26 X 2 RSD: Relative standard deviation. TABLE 4: RECVERY STUDIES F THE HPLC METD FR THE ASSAY F FUR PRESERVATIVES Applied concentration (% of target) (n = 3) Components 50 100 150 PhE 99.86±0.21* 100.00±0.12 99.88±0.14 MP 99.96±0.26 99.97±0.31 98.92±0.18 EP 100.00±0.45 100.00±0.19 99.77±0.36 PP 99.87±0.22 99.78±0.11 99.58±0.28 *The coefficient of variation inter-day precision respectively indicating that the method is reliable and reproducible (Table 3). For determining accuracy, three different solutions were prepared with a known added amount of pure PhE, MP, EP and PP compounds to give a concentration range of 50-150% of that in a test preparation. These solutions were injected in triplicate. The recovery was 100±2% for all samples with %RSD less than 3% (Table 4). The LC-PDA isoplot chromatogram was obtained which demonstrated a good separation of the PhE (RT= 4.50 min), MP (RT= 6.00 min), EP (RT= 9.05 min) and PP (RT= 15.03 min) from each other. A wavelength of 258 nm was found to be the most effective compromise to accomplish the detection and quantification of the four preservative components in a single run. The PhE, MP, EP and PP peaks are adequately resolved from each other, typical resolution values were >2. Therefore, this method demonstrates acceptable specificity. The stability of analytical solutions was evaluated for 48h. The results given in Table 3 showed there was no significant change (<0.15% response factor) in PhE, MP, EP and PP concentrations (750, 145, 35 and 18 μg/ml) over this period. Fig. 3: Chromatogram of four preservatives obtained from gel samples PhE eluted at tr: 4.55 min; MP at tr: 6.12 min; EP at tr: 9.32 min and PP at tr: 15.57 min. Analytical methods developed for use in quality control laboratories ideally are robust. Retention time for the analytes of interest will not change significantly from day-to-day or from laboratory-tolaboratory if the method is considered robust. To determine the robustness of the chromatographic methodology developed for PhE, MP, EP and PP, experimental conditions were purposely altered and chromatographic characteristics were evaluated. In particular the ph of the mobile phase was adjusted to 2.5 and 3.5, thus, the normal ph for the method was 3.0. The effected temperature was also studied. Standard solutions were prepared and injected at early 20 and again at 28. In all cases studied, the retention times of these preservatives (PhE, MP, EP and PP) were remains same 4.50, 6.00, 9.05 and 15.03 min, respectively. The coefficient of variation for retention time was lass then 1%. Excellent separation was always achieved, indicating that the analytical method remained selective for all components under the measured conditions. A system suitability test was performed to determine the accuracy and precision of the system by injecting six replicate injections of PhE, MP, EP and PP standard solutions. The RSD of the peak areas responses was measured. The RSD for PhE (0.11%), MP (0.15%), EP (0.09%) and PP (0.26%) as can be seen in Table 3. 424 Indian Journal of Pharmaceutical Sciences

The developed and validated method was applied to the determination of preservatives studied from pharmaceutical gel product. The chromatogram obtained from gel samples is shown in fig. 3. Peak identification of the preservatives in gel sample was based on the comparison between the retention times of standard compounds and was confirmed by spiking known standard compounds to the sample. A reversed-phase HPLC assay method with UV spectrophotometric detection on a C8 analytical column was developed successfully for the determination of 2-phenoxyethanol, methylparaben, ethylparaben and propylparaben preservatives. The method was validated and the results obtained were accurate and precise with RSD <1% in all cases and no significant interfering peaks were detected. The validated method was successfully applied to the determination of commercially available pharmaceutical gel products for these preservatives. The method can be used for the routine quality control analysis (batch analysis) of compounds in pharmaceutical gel products containing 2-phenoxyethanol, methylparaben, ethylparaben and propylparaben preservatives and the degradation products of the active compounds. ACKNWELDGEMENTS Author expresses his gratitude to Drs. Tony Bradshaw and Simon Giles (School of Life Sciences, xford Brookes University) for their support in buying materials for this research work. REFERENCES 1. Shabir GA. Determination of combined p-hydroxy benzoic acid preservatives in a liquid pharmaceutical formulation and assay by HPLC. J Pharm Biomed Anal 2004;34:207-13. 2. Denyer SP, Baird RM. Preservatives: Registrations and regulatory affairs-guide to Microbiological Control in Pharmaceuticals. West Sussex: Ellis Horwood; 1990. p. 314-40. 3. Perez- Lozano P, Garcia-Montoya E, rriols A, Minarro M, Tico JR, Sune-Negre JM. A new validated method for the simultaneous determination of benzocaine, propylparaben, and benzyl alcohol in a bioadhesive gel by HPLC. J Pharm Biomed Anal 2005;39:920-7. 4. Sottofatori E, Anzaldi M, Balbi A, Tonello G. Simultaneous HPLC determination of multiple components in a commercial cosmetic cream. J Pharm Biomed Anal 1998;18:213-7. 5. Dvorak J, Hajkova R, Matysova L, Novakova L, Koupparis MA, Solich P. Simultaneous HPLC determination of ketoprofen and its degradation products in the presence of preservatives in pharmaceuticals. J Pharm Biomed Anal 2004;36:625-9. 6. Popovic G, Cakar M, Agbaba D. Determination of bifonazole in creams containing methyl- and propyl p-hydroxybenzoate by derivative spectrophotometric method. J Pharm Biomed Anal 2003;33:131-6. 7. Hajkova R, Solich P, Dvorak J. Simultaneous determination of methylparaben, propylparaben, hydrocortisone acetate, and its degradation products in a topical cream by RP-HPLC. J Pharm Biomed Anal 2003;32:921-7. 8. Koundourellis JE, Malliou ET, Broussali TA. High performance liquid chromatographic determination of ambroxol in the presence of different preservatives in pharmaceutical formulations. J Pharm Biomed Anal 2000;23:469-75. 9. Shabir GA, Lough WJ, Shafique AA, Shar GQ. Method development and validation for the HPLC assay of phenylformic acid, 2,4-hexadienoic acid, methyl 4-hydroxybenzoate and propyl 4-hydroxybenzoate. J Liq Chromatogr Relat Technol 2006;29:1223-33. 10. Dalibor S, Jitka HR, Ferreira LC, Maria C, Montenegro BS. Determination of ambroxol hydrochloride, methylparaben and benzoic acid in pharmaceutical preparations based on sequential injection technique coupled with monolithic column. J Pharm Biomed Anal 2006;40:287-93. 11. Kokoletsi MX, Kafkala S, Tsiaganis M. A novel gradient HPLC method for simultaneous determination of ranitidine, methylparaben and propylparaben in oral liquid pharmaceutical formulation. J Pharm Biomed Anal 2005;38:763-7. 12. Belgaied JE, Trabelsi H. Determination of cisapride, its oxidation product, propyl and butyl parabens in pharmaceutical dosage form by reversed-phase liquid chromatography. J Pharm Biomed Anal 2003;33:991-8. 13. Mahuzier PE, Altria KD, Clark BJ. Selective and quantitative analysis of 4-hydroxybenzoate preservatives by microemulsion electrokinetic chromatography. J Chromatogr A 2001;924:465-70. 14. Lee MR, Lin CY, Li ZG, Tsai TF. Simultaneous analysis of antioxidants and preservatives in cosmetics by supercritical fluid extraction combined with liquid chromatography-mass spectrometry. J Chromatogr A 2006;1120:244-51. 15. Grosa G, Grosso ED, Russo R, Allegrone G. Simultaneous, stability indicating, HPLC-DAD determination of guaifenesin and methyl and propyl-parabens in cough syrup. J Pharm Biomed Anal 2006;41:798-803. 16. Kokoletsi MX, Kafkala S, Tsiaganis M. A novel gradient HPLC method for simultaneous determination of ranitidine, methylparaben and propylparaben in oral liquid pharmaceutical formulation. J Pharm Biomed Anal 2005;38:991-8. 17. Shabir GA, Lough WJ, Shafique AA, Bradshaw TK. Evaluation and application of best practice in analytical method validation. J Liq Chromatogr Relat Technol 2007;30:311-33. 18. Shabir GA. Validation of HPLC methods for pharmaceutical analysis: Understanding the differences and similarities between validation requirements of the U.S. Food and Drug Administration, the U.S. Pharmacopoeia and the International Conference on Harmonization. J Chromatogr A 2003;987:57-66. 19. Shabir GA. Step-by-step analytical methods and protocol in the quality system compliance industry. J Validation Technol 2004;10:314-24. 20. International Conference on Harmonization (ICH), Validation of analytical procedures: Text and Methodology, Q2 (R1), Geneva, Switzerland: 2005. 21. Reviewer Guidance: Validation of Chromatographic Methods, Food and Drug Administration (FDA), Centre for Drug Evaluation and Research (CDER), November 1994. 22. US Pharmacopoeia (USP) 32, Chromatography, Section (621), Rockville MD: United States Pharmacopeal Convention; 2009. p.1776. Accepted 15 July 2010 Revised 16 April 2010 Received 20 August 2009 Indian J. Pharm. Sci., 2010, 72 (4): 421-425 Indian Journal of Pharmaceutical Sciences 425