Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Similar documents
Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Samples compliant with the WATER

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Analysis of domestic sludge using the Agilent 4200 MP-AES

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Analysis Repeatability of Trace and Major Elements in a Water Sample

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

OES - Optical Emission Spectrometer 2000

Trace Analyses in Metal Matrices Using the ELAN DRC II

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Trace Metal Dynamic Extraction from Five Pumps in Hydrochloric Acid. Submitted to: Reto Schoeb Levitronix GmbH. Submitted by:

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

CT Associates, Inc. Trace Metal Dynamic Extraction from Three Centrifugal Pumps in Hydrochloric Acid. Submitted to: Juergen Hahn Levitronix GmbH

ICP-OES Application Note Number 35

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Contamination Monitoring of Semiconductor Processes by VPD HR-ICPMS

Single particle analysis using the Agilent 7700x ICP-MS

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

ICP/MS Multi-Element Standards

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Defining quality standards for the analysis of solid samples

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

OPTIMIZING METALS SAMPLE PREP

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

ICP-MS Analysis of Bulk & Process Chemicals for Semiconductor Processes. Ann O Connell Chemical Analysis Eng Intel Corporation

Agilent 8900 Triple Quadrupole ICP-MS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Optimizing Analytical Performance in ICP-OES Applications

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Current State of Extraction Don t Be Deceived! Sharon F. Webb, Ph.D. Director of Quality Program

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Agilent Technologies at TIAFT 2013

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom.

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom.

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Sample Analysis Design PART II

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

Microsoft Excel Directions

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Element Cube Project (x2)

materials and their properties

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Transcription:

Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid Application Semiconductor Authors Abe G. Gutiérrez Elemental Scientific 2440 Cuming St Omaha, NE 68131 USA abe@icpms.com Junichi Takahashi Koichi Yono Agilent Technologies Musashino Center Building 1-19-18 Naka-cho Musashino-shi Tokyo 180-0006 Japan Tom Gluodenis Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract This application note describes the analysis of metallic impurities in semiconductor-grade hydrofluoric acid by inductively coupled plasma mass spectrometry (ICP-MS). The Agilent 7500s ICP-MS instrument with its ShieldTorch technology was used to analyze 38% (w/w) hydrofluoric acid directly for all of the elements commonly required by the semiconductor industry. Combined with the high-sensitivity characteristic of the ShieldTorch System, the off-axis ion lens assembly in the Agilent 7500s provided excellent signal/background resulting in exceptional detection limits for all elements (0.02- to 5-ppt range). Excellent spike recovery data at the 5-ppt level further highlights the suitability and accuracy of the Agilent 7500s for the determination of all Semiconductor Equipment and Materials International (SEMI) required elements in hydrofluoric acid at ultra-trace levels. Introduction The semiconductor industry uses solutions of hydrofluoric (HF) acid for wet chemical cleaning or etching of silicon dioxide on silicon wafers and integrated circuits (ICs) during the manufacturing process. These cleaning processes usually involve immersion into etching baths containing dilute hydrofluoric acid. It is critical during the cleaning process that these chemical etching baths do not contaminate the wafer surface and ICs. Metallic contaminants in semiconductor devices can diffuse on the surface, diffuse into the device itself, and accumulate at interfaces. This results in surface conduction that decreases minority carrier lifetime, degrades electrical device performance, and lowers product yield. In fact, over 50 percent of the yield losses in IC manufacturing are caused by micro-contamination 1. To minimize this risk, many incoming chemicals are analyzed for their chemical purity and the etching baths are sampled regularly during their useful lifetimes. This application note describes the use of an Agilent 7500 Inductively Coupled Plasma Mass Spectrometry (ICP-MS) system, fitted with an HFresistant sample introduction system, to analyze semiconductor-grade HF (38% w/w) with excellent recoveries (75 to 125%) for a 5-ppt spike added directly to the concentrated acid. Recent advances in ICP-MS and sample introduction technology allow the routine, direct assay of

38% (w/w) HF for all Semiconductor Equipment and Materials International (SEMI) required elements, including K, Ca, Fe and Zn, at single ppt levels. The chemically inert, O-ring-free perfluoroalkoxyalkane (PFA) sample introduction system allows direct, contamination-free introduction of the concentrated acid into the ICP-MS. The combination of a PFA microflow nebulizer, Peltier-cooled spray chamber, and robust 27-MHz plasma ensure complete matrix decomposition and ionization of the sample including difficult elements such as boron and zinc, which have high ionization potentials. This eliminates the need for the use of internal standards, thereby eliminating a potential source of contamination. Matrix-based interferences on key elements, such as Ca, K, and Fe, were virtually eliminated using cool plasma conditions and Agilent Technologies exclusive ShieldTorch Technology. In addition, the unique Omega II ion lens system in the Agilent 7500s ICP-MS results in a 10- to 100-fold enhancement in the signal-to-background (S/B) ratio for elements across the entire mass range. Switching between operating modes is automated using Agilent's Multi-Tune software functionality and requires that each autosampler vial be sampled only once, thereby minimizing any potential for sample contamination. Analytical results are collated and presented in a single report. Normal/Cool plasma switching has proven to be rapid and stable over extended periods of operation resulting in the excellent S/B, signal stability, and detection capability required for this analysis 2. Experimental Instrumentation The instrument used in this application was an Agilent 7500s ICP-MS equipped with the ShieldTorch System (STS) and an HF-resistant, PFA sample introduction system (Elemental Scientific, Omaha NE, USA). The Elemental Scientific PFA sample introduction system (part number ES-2161-7500) consists of a PFA-100 high efficiency, self-aspirating micro flow nebulizer, 35 mm PFA PureCap end cap, 35 mm PFA spray chamber, 2-mm O-ring free Pt injector and precision quartz torch. The Agilent 7500s was operated in soft extraction mode 3,4. In this mode of operation, a slight positive voltage is applied to the first lens of the 7500s twin extraction lenses. This automated opti- mization procedure dramatically reduces the background across the entire mass range without any sacrifice in sensitivity. The result is a 10- to 100- fold increase in the S/B ratio for most elements. Potentially troublesome argon polyatomic ions on key elements, like K, Ca, Fe, Ge, V, and Zn, were virtually eliminated using Agilent's STS combined with cool plasma operating conditions. Unlike traditional cool plasma conditions, which require plasma powers of 500 to 550 W, the Agilent 7500s is able to effectively eliminate argon based polyatomic interferences while operating at higher, more robust plasma conditions. In addition, the STS is highly efficient at reducing the plasma potential to approximately 1 ev, significantly lower than with non-shield plasma locking devices. This leads to better ion focusing and more efficient interference removal, thereby alleviating the need for sample pretreatment or complex scanning reaction/collision cell technologies. Sample Preparation All sample analysis was performed directly without any preparation or pretreatment, which avoids contamination and leads to improved detection limits. Analyses were performed using the method of standard additions (MSA) on an ultrapure-grade HF sample. Calibration curves were prepared at concentrations ranging from 0.5 to 10 ng/l (ppt). A single MSA curve was established and converted into an external calibration curve, thereby eliminating the need to spike each subsequent sample to be analyzed. Once an external calibration curve has been established from the MSA curve in this manner, all other sample concentrations can be determined against it. No internal standards were used in this study to minimize the risk of sample contamination. Following quantitation, HF samples were spiked at a level of 5 ppt (ng/l) to assess the accuracy of the measurement. All sample handling and data analysis was performed under Class 100 cleanroom conditions. Table 1. ICP-MS Operating Conditions Cool plasma Normal plasma Parameter conditions conditions RF power 630 W 1300 W Sampling depth 18.5 mm 7 mm Carrier gas flow 0.9 L/min 0.94 L/min Makeup gas flow 1.0 L/min 0.1 L/min Extraction lens 1-80 V + 4.7 V Extraction lens 2-20 V - 5 V 2

ICP-MS Analysis The sample was self-aspirated at an uptake rate of 100 µl/min. Instrument operating conditions are given in Table 1. Switching between normal and cool plasma conditions is fully automated using Agilent s Multi-tune software. Multi-tune allows complete data collection in a single acquisition, with a single visit to the sample vial, thereby saving time and reducing the potential for sample contamination. In addition, all resulting data is compiled in a single report. Stabilization time in switching between plasma operating modes is minimal at 20 seconds. The stability of switching between operating modes over prolonged periods was previously demonstrated to be 3% RSD over an 8-hour period 2. Results and Discussion The results in Table 2 show detection limits (ppt) and spike recovery data for a standard suite of semiconductor elements in undiluted HF. Note the exceptional performance even for more difficult elements such as Ca, K, and Fe. Table 2. Detection Limits (Parts-Per-Trillion) and Spike Recoveries (%) in Undiluted HF (38%) 7500 DL Blank conc Blank plus Percent Plasma Element Mass (ppt) (ppt) 5 ppt recovery mode Li 7 0.05 0.0 5.0 99 Cool Be 9 0.2 0.1 5.0 100 Normal B 11 3 19.2 24.4 103 Normal Na 23 0.09 2.5 7.6 102 Cool Mg 24 0.1 0.6 5.8 105 Cool Al 27 0.2 2.0 7.8 117 Cool K 39 0.1 4.4 9.8 108 Cool Ca 40 2 21.7 26.6 98 Cool Ti 48 5 76.8 81.7 98 Normal V 51 3 42.6 48.1 112 Normal Cr 52 1 13.0 18.2 104 Cool Mn 55 0.2 1.4 6.3 99 Cool Fe 56 0.3 3.8 9.3 111 Cool Ni 58 0.1 0.1 5.1 102 Cool Co 59 0.2 0.5 5.7 104 Cool Cu 63 0.3 0.6 5.6 100 Cool Zn 68 0.8 2.2 6.4 83 Cool Ga 69 0.05 0.1 4.9 97 Cool Ge 74 5 52.2 56.2 79 Normal Rb 85 0.02 0.0 5.1 102 Cool Sr 88 0.1 0.1 5.3 103 Cool Zr 91 4 26.2 31.4 103 Normal Nb 93 0.5 2.3 7.2 96 Normal Mo 98 2 37.4 41.7 85 Normal Ag 107 0.08 0.2 4.9 95 Cool Cd 114 0.5 6.8 12.0 104 Normal Sn 118 0.2 1.2 5.8 93 Normal Sb 121 1 7.0 12.3 107 Normal Ba 138 0.06 0.2 4.8 93 Normal Ta 181 0.1 1.8 6.6 97 Normal W 182 0.9 4.4 9.8 107 Normal Au 197 0.2 0.2 5.1 96 Normal Tl 205 0.08 0.1 5.1 100 Normal Pb 208 0.08 0.2 5.1 98 Normal Bi 209 0.05 0.1 4.9 96 Normal U 238 0.02 0.0 4.7 93 Normal 3

Representative calibration curves are given in Figure 1. Excellent correlation coefficients were obtained considering the low concentration levels used. The good fit at the single ppt level on each calibration curve shows this level of analyte can be reliably quantified in undiluted HF. Detection limits were calculated using three times the standard deviation (n = 3) of the raw counts of the HF blank divided by the slope of the calibration curve. The reported detection limits are generally excellent and are dependent upon the trace metal content in the blank. This provides opportunity for even further improvement upon the reported values. All detection limits are in the range 0.02 to 5 ppt (ng/l) in the 38% HF. Figure 1. Representative calibration curves for B and Zn obtained using the method of standard addition (Blank, 2, 4 and 10 ppt) and K and Fe (Blank, 0.5, 2 and 10 ppt). 4

A 2-hour stability study was performed by adding a 10-ppt standard to the 38% HF and analyzing the spiked sample over a 2-hour period. Instrument stability over this period was excellent with %RSD values typically less than 3% over the period. Figure 2 is a stability plot of representative elements. 1.4 1.2 1.0 Normarized counts 0.8 0.6 Li-7 (3.82%) Na-23 (4.71%) Mg-24 (3.75%) Fe-56 (2.39%) Ni-58 (3.25%) Co-59 (2.70%) 0.4 Al-27 (4.16%) Cu-63 (2.88%) K-39 (2.92%) Ga-69 (3.68%) 0.2 Cr-52 (3.30%) Mn-55 (3.54%) Rb-85 (3.71%) Ag-107 (4.68%) 0.0 0 30 60 90 120 Time (minutes) Figure 2. Analysis of 10 ppt multi-element standard spiked into 38% HF measured repeatedly over a 2-hour period. 5

Conclusions Direct analysis of concentrated hydrofluoric acid (38% w/w) at the 5-ppt level is possible using the Agilent 7500s with ShieldTorch System and a trace metal-free, high-efficiency, inert sample introduction kit (Elemental Scientific, Omaha, NE). As is often the case when making ultra-trace measurements, analytical results are highly dependent upon laboratory environment and analytical methodology. In this study, all data was obtained under cleanroom conditions with minimal sample handling. The excellent detection limits and spike recovery data highlight the suitability and accuracy of the Agilent 7500s ICP-MS to determine low ppt level impurities in the concentrated acid even for difficult elements like B, Zn, K, Ca and Fe. The robust 27-MHz plasma ensures complete sample decomposition providing plasma stability when operating in cool plasma mode as well as ionization of elements with a high ionization potential, such as boron and zinc. The patented ShieldTorch System enhances system sensitivity while reducing the spectral interferences on key elements while the dual extraction lenses and off-axis Omega lens system reduce the S/B by a factor of 100. The result is uncompromised quantification at the low ppt level without the need for sample pretreatment or complex scanning reaction/collision cell technologies. Samples were analyzed using the MSA. Only the first sample is spiked with the calibration standards. The MSA regression generated from the first sample is then converted to an external calibration curve, eliminating the need to spike all subsequent samples. This protocol enhances sample throughput while eliminating potential sources of contamination. Finally, Agilent's Multi-mode software ensures walk-away automation even when changing instrument operating modes. All data is collected in a single acquisition, with a single visit to the sample vial, and presented to the user in a single report. www.agilent.com/chem References 1. Handbook of Semiconductor Wafer Cleaning Technology by Werner Kern. Chapter 2, section 2.3, page 8 2. Tye, C.T., Gutierrez, A., Agilent Technologies ICP-MS Journal, 2000, 7, 2. 3. Sakata, K., Yamada, N., Sugiyama, N., Spectrochim. Acta B, 2001, 56, 1249. 4. Tye, C.T., Sakata, K., Agilent Technologies ICP-MS Journal, 2000, 8, 7. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. For more information about semiconductor measurement capabilities, go to www.agilent.com/chem/semicon. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2002 Printed in the USA March 4, 2002 5988-5421EN