Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Similar documents
Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Maximizing Sample Throughput In Purge And Trap Analysis

Solid Phase Micro Extraction of Flavor Compounds in Beer

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Optimizing. Abstract: is standardd. procedures. altered to

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Determination of Volatile Organic Compounds in Air

Methanol Extraction of high level soil samples by USEPA Method 8260C

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Maximizing Production While Minimizing Costs

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Optimizing Standard Preparation

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analytical Trap Comparison for USEPA Method 8260C

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

2017 EPA Method Update Rule and EPA Method 624.1

Ed George and Anaïs Viven Varian, Inc.

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Volatile Organic Compounds in Every Day Food

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

STANDARD OPERATING PROCEDURES

EPA TO-17 Volatile Organic Compounds

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

Headspace Technology for GC and GC/MS: Features, benefits & applications

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Author. Abstract. Introduction

Application. Gas Chromatography February Introduction

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Thermal Desorption Technical Support

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Meeting NJ Low Level TO-15 Air Testing Method Requirements

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Study of Residual Solvents in Various Matrices by Static Headspace

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Analysis of BTEX in Natural Water with SPME

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Target Compound Results Summary

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

STANDARD OPERATING PROCEDURES

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Residual solvents analysis using an Agilent Intuvo 9000 GC system

CALA Directory of Laboratories

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Application Note 032

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

U.S. EPA Method 8270 for multicomponent analyte determination

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Transcription:

Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note Environmental Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract The United States Environmental Protection Agency (USEPA) Method 8260 has an extensive list of analytes that can be analyzed by purge and trap sampling. Two of the more troublesome compounds on this list are Ethanol and 1, 4-Dioxane. Both of these compounds are water miscible and Selective Ion Monitoring (SIM) is required in order to detect these compounds at lower levels. The advent of SIM/Scan monitoring has made it easier to analyze for these compounds by traditional purge and trap sampling. However, due to the miscibility of the compounds and their propensity to stick to the sparge vessel of the purge and trap, purge and trap sampling needs to be optimized. This application will compare linearity, method detection limits, precision and accuracy and carryover of several purge and trap sampling parameters. Introduction: During the manufacture of pharmaceuticals, 1, 4-Dioxane is used as a cleansing agent. It is also a byproduct of plastic manufacturing. The most likely exposure to 1, 4-Dioxane is at an industrial site. However, if any 1, 4-Dioxane is released into the environment during manufacturing there is a potential for it to migrate into ground water. The fact that 1, 4-Dioxane is so miscible in water makes degradation of the chemical challenging. Ethanol, on the other hand, has been a popular gasoline additive. Since it burns more quickly and completely than gasoline, emissions from car exhaust are decreased. The downside of this is underground storage tank leakage and fuel spills cause ground water and drinking water to be contaminated with both the fuel and the ethanol additive. Since ethanol is also very miscible in water, the detection of ethanol contamination in water can be difficult. There have been several innovations that help overcome the obstacles of detecting 1, 4-Dioxane and Ethanol. The first one is the SIM mode of the mass spectrometer. This mode allows better detection of hard to extract compounds. Furthermore, mass spectrometers are now equipped to run SIM/Scan, this not only helps detect more difficult compounds, but also enables the detection of an extensive list of USEPA Method 8260 compounds without having to run the samples twice. Advancements in purge and trap sampling have also facilitated better detection of these

Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK compounds. Most effective in this has been the ability to heat the samples. However, there are drawbacks to better detection. The most problematic of these is the tendency of 1, 4-Dioxane and Ethanol to stick to the glass ware. This susceptibility has caused many headaches in environmental labs. This application will investigate seven variations of purge and trap sampling. The experimental modifications were chosen to optimize the 1, 4-Dioxane and Ethanol responses and minimize the carryover issues found with these compounds. Thus, purge temperature, sparge vessel variations and glassware bake temperatures will be examined. The results of these experimental distinctions will be evaluated in order to recommend the optimum purge and trap sampling parameters for your lab. Experimental: The sampling system used for this study was the EST Analytical Evolution concentrator affixed with a Vocarb 3000 trap. The Centurion WS autosampler equipped with the syringe option was employed as the autosampler. The separation and analysis were performed by an Agilent 7890A Gas Chromatograph (GC) and 5975C inert XL Mass Spectrometer (MS). The GC was configured with a Restek Rxi-624 Sil MS 30m x 0.25mm x 1.4µm column. Table 1 shows the GC/MS parameters while the purge and trap parameters used for this study are listed in Tables 2 and 3. The experimental variations listed in Table 3 take advantage of two patented innovations from EST Analytical. The first patent used is the sample heater, US Patent Number 8075842 B1. This invention allows the sparge vessel to be heated during bake thus limiting carryover of the heavier volatile compounds. The second patent is the ability to automate sampling waters and transferring them to the soil station for purging, US Patent Number 8850905 B2. This advancement was utilized for iteration E in order to provide a new sparge vessel for each sample. GC/MS Agilent 7890A/5975C inert XL Inlet Split/Splitless Inlet Temp. 220ºC Inlet Head Pressure 12.153 psi Mode Split Split Ratio 40:1 Column Rxi-624Sil MS 30m x 0.25mm I.D. 1.4µm film thickness 45ºC hold for 1 min, ramp 15ºC/min to Oven Temp. Program 220ºC, hold for 1.33 min, 14 min run time Column Flow Rate 1mL/min Gas Helium Total Flow 44mL/min Source Temp. 230ºC Quad Temp. 150ºC MS Transfer Line Temp. 180ºC Scan Range m/z 35-300 Scans 5.2 scans/sec SIM Ions (0.7min to 3.49min) 45, 46 SIM Ions (3.5min to 14min) 58, 88 Solvent Delay 0.7 min Table 1: GC/MS Experimental Parameters

Purge and Trap Concentrator EST Evolution Trap Type Vocarb 3000 Valve Oven Temp. 150ºC Transfer Line Temp. 150ºC Trap Temp. 35ºC Moisture Reduction Trap (MoRT) Temp. 39ºC Purge Time 11 min Purge Flow 40mL/min Dry Purge Temp. ambient Dry Purge Flow 40mL/min Dry Purge Time 1.0 min Desorb Pressure Control On Desorb Pressure 6psi Desorb Time 0.5 min Desorb Preheat Delay 15 sec Desorb Temp. 260ºC Moisture Reduction Trap (MoRT) Bake Temp. 210ºC Bake Temp 270ºC Sparge Vessel Bake Temp. 120ºC Bake Time 8 min Bake Flow 85mL/min Purge and Trap Auto-Sampler EST Centurion WS Sample Type Water Water Volume or 10ml Internal Standard Vol. 5µl Table 2: Purge and Trap Parameters Altered Parameters Baseline Parameter s A Iteration B Iteration C Iteration D Iteration E Iteration F Iteration G Sparge Vessel Type Fritted Fritted Fritted Fritted 40ml Vial, 10ml purge volume Fritless Bulbless Fritless Bulbless Purge Temperature (ºC) Room Temp. 40 60 40 40 40 40 Sparge Bake Temperature (ºC) 110 110 110 Not Applied Not Applied* 110 Not Applied *For Iteration E, the patented Water/Soil prep mode was used. Table 2: Purge and Trap Sampling Test Parameters For each parameter iteration, see Table 2, a calibration curve was established with a linear range of 0.5 to 200ppb using standards from Restek. After each of the prescribed curves was determined; Method Detection Limits (MDLs), precision and accuracy and carryover studies were performed. A series of seven low level standards were run in order to establish MDLs per 40CFR Part 136, Appendix B. Next, seven replicate samples of the 50ppb standard were run in order to establish the precision and the accuracy of the experimental parameters. Finally, a series of four 50ppb standards were run with each standard followed by three blanks. The amount of carryover was then analyzed in the first blank. The data from each parameter variation was compiled and

compared in order to determine the optimum conditions for this analysis. An average of the results for each analysis condition is listed in Table 4. Averaged Results Results Summary Parameter Iteration A B C D E F G Curve Relative Standard Deviation 7.75 8.09 8.30 6.73 7.84 7.60 7.58 Curve Response Factor 1.10 1.06 1.09 1.02 1.15 1.03 1.11 Method Detection Limit 0.500 0.316 0.973 0.327 0.394 0.326 0.288 Precision at 50ppb 4.45 3.96 3.89 2.11 2.90 3.89 4.46 Percent Recovery at 50ppb 107.01 95.29 97.15 96.28 97.12 97.59 99.72 Table 4: Averaged Results Summary The primary compounds of interest for the study were Ethanol and 1, 4-Dioxane. For this reason, the data for these compounds was compiled separately in order to better distinguish the benefits or detriment of each purge and trap parameter set. The experimental results of the precision and percent recovery studies are displayed in Tables 5 and 6. Precision at 50ppb Compound Parameter Iteration A B C D E F G Ethanol 13.43 7.29 12.48 2.47 2.54 6.18 3.68 1,4-Dioxane 9.44 7.64 14.86 2.10 4.34 7.20 2.51 Table 5: Ethanol and 1, 4-Dioxane Precision at 50ppb Percent Recovery at 50ppb Compound Parameter Iteration A B C D E F G Ethanol 170.24 108.45 125.90 110.68 93.71 80.05 96.53 1,4-Dioxane 148.44 128.01 122.49 110.95 101.55 80.27 97.45 Table 6: Ethanol and 1, 4-Dioxane Percent Recovery at 50ppb For the carryover studies, Ethanol and 1, 4-Dioxane were examined along with the carryover of 1, 2, 4-Trichlorobenzene, Naphthalene and 1, 2, 3-Trichlorobenzene. The reason the heavier compounds were included in this study was a concern that changing the sampling parameters may help the Ethanol and 1, 4-Dioxane carryover, but harm the carryover of the heavier compounds of interest. Table 7 shows a listing of the carryover results, in parts per billion, while Figure 1 displays the percent carryover of the different purge and trap experimental parameters for just the Ethanol and 1, 4-Dioxane.

Average Carryover after 50ppb Compound Iteration A B C D E F G Ethanol 63ppb 76ppb 38ppb 58ppb ND ND ND 1,4-Dioxane 53ppb 60ppb 52ppb 44ppb ND ND ND 1,2,4-Trichlorobenzene ND ND ND 0.50ppb ND ND 0.53ppb Naphthalene ND ND ND 0.59ppb ND ND 0.61ppb 1,2,3-Trichlorobenzene ND ND ND 0.53ppb ND ND 0.57ppb ND signifies Non Detect Table 7: Carryover in First Blank after a 50ppb Standard Ethanol 1,4 Dioxane 160.00 140.00 120.00 100.00 80.00 60.00 40.00 20.00 0.00 Baseline RT purge, 110ºC Sparge Bake 40ºC Purge, 110ºC Sparge Bake 60ºC Purge, 110ºC Sparge Bake 40ºC Purge, No Sparge Bake Water Extraction 40ºC Purge 40ºC Purge, 110ºC Sparge Bake, Fritless Bulbless Sparger 40ºC Purge, NoSparge Bake, Fritless Bulbless Sparger Figure 1: Ethanol and 1, 4-Dioxane Percent Carryover after a 50ppb Standard Graphic After all the studies were completed, two purge and trap parameter variations showed the best precision, percent recovery, and carryover while meeting the USEPA Method 8260 linearity and response factor requirements. An abbreviated listing of the results for these two iterations are shown in Table 8.

Compound Curve Linearity Water Extraction 40ºC Purge (E) Response Factor Precision (%RSD) % Recovery Fritless Bulbless Sparge Vessel, 40ºC purge, Sparge Bake (F) Curve Linearity Response Factor Precision (%RSD) % Recovery Dichlorodifluoromethane 3.3 0.543 5.03 96.45 7.73 0.42 6.62 100.91 Chloromethane 11.39 1.016 3.71 90.88 12.68 0.963 4.44 87.32 Vinyl Chloride 2.08 0.927 4.72 97.24 5.31 0.889 5.13 95.08 Bromomethane 11.04 0.547 3.2 89.98 14.45 0.568 5.25 90.29 Chloroethane 11.48 0.613 4.18 90.66 12.83 0.623 4.76 87.49 Ethanol 0.999* 0.012 2.54 93.71 14.71 0.009 6.18 80.05 Trichlorofluoromethane 3.86 0.614 4.8 98.25 7.91 0.54 6.45 98.41 1,1-Dichloroethene 4.26 0.611 3.99 98.96 6.46 0.517 5.73 101.07 Acetone 13.93 0.346 3.3 93.93 13.09 0.392 3.06 89.45 Carbon Disulfide 5.64 2.112 4.27 98.19 7.28 1.544 5.49 100.13 Methylene Chloride 13.51 0.786 2.12 91.85 9.99 0.728 2.31 92.19 MTBE 3.38 2.113 1.55 100.14 2.39 2.174 1.72 104.56 1,1-Dichloroethane 5.99 1.419 2.97 100.83 8.15 1.245 3.67 106.21 2-Butanone 5.86 1.574 1.21 93.14 9.97 1.901 4.7 95.12 Chloroform 10.6 1.33 2.6 92.31 5.51 1.211 3.06 99.47 Benzene 2.89 3.079 2.92 97.77 2.23 2.735 3.8 100.65 1,2-Dichloropropane 5.64 0.458 2.46 99.63 4.85 0.41 3.17 105.24 1,4-Dioxane 7.47 0.01 4.34 101.55 13.72 0.009 7.2 80.27 Toluene 12.28 1.136 3.5 92.8 4.15 0.958 4.79 102.29 2-Hexanone 6.69 0.367 1.64 98.85 5.14 0.448 1.81 106.37 Chlorobenzene 6.08 1.34 2.19 97.14 2.93 1.156 3.1 97.02 Ethylbenzene 7.62 2.321 3.45 98.97 6.16 1.994 4.27 97.77 Xylene (m+p) 7.87 1.804 3.18 97.64 5.89 1.527 4.06 98.29 Xylene (o) 7.56 1.819 2.45 99.5 5.6 1.614 3.62 98.74 Bromoform 14.45 0.259 1.52 116.43 13.74 0.286 1.42 109.97 1,1,2,2-Tetrachloroethane 6.15 1.266 1.09 94.79 4.33 1.213 2.06 100.72 1,2-Dibromo-3-chloropropane 7.97 0.245 0.91 104.86 4.89 0.271 1.19 104.47 1,2,4-Trichlorobenzene 7.68 1.291 2.6 98.69 5.61 1.02 2.71 98.64 Naphthalene 12.46 3.787 1.52 95.1 5.1 3.365 1.63 101.97 Hexachlorobutadiene 9.3 0.556 4.49 97.58 7.86 0.376 5.07 96.15 1,2,3-Trichlorobenzene 12.74 1.253 1.52 93.05 5.08 0.955 2.12 99.03 Table 8: Results of the Water Extraction and the Fritless/Bulbless Sparge Vessel With Sparge Bake Studies

Figure 2: 50ppb Chromatogram of Water Extraction Standard Conclusions: All seven purge and trap parameter iterations passed the USEPA Method 8260 requirements for linearity and method detection limits. The problem with some of the experimental parameters was found in the carryover and precision and accuracy studies. When using the traditional fritted sparge vessel there was a large amount of carryover for the Ethanol and 1, 4-Dioxane. Since the carryover was so high, the precision and accuracy data suffered. The Fritless/Bulbless sparge vessel, on the other hand, displayed much lower carryover for the Ethanol and 1, 4-Dioxane especially when the patented sparge bake was not used. However, the 1, 2, 4-Trichlorobenzene, Naphthalene and 1, 2, 3-Trichlorobenzene carryover after the 50ppb standard was above the lower limit of the curve without the sparge bake. Thus, the sparge bake would be recommended in order to limit the carryover of the heavier compounds. The optimum purge and trap parameters proved to be the patented water extraction technique. This technique provided linearity and method detection limits that met the USEPA Method 8260 requirements, while providing excellent precision and accuracy data. During water extraction, the sample is transferred to an empty 40ml vial and then purged in the soil station of the Centurion. The fresh vial provides a clean sparge vessel for every sample thus limiting the problems of Ethanol and 1, 4-Dioxane adhering to the glass while reducing carryover of the heavy compounds. This study demonstrated the EST Analytical Evolution and Centurion, in conjunction with the patented water extraction sampling technique, provided the best results for Ethanol, 1, 4-Dioxane and the extensive USEPA Method 8260 compound list. References: 1. Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS); United States Environmental Protection Agency Method 8260C, Revision 3, August, 2006. 2. Jurek, Anne, A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260, https://www.estanalytical.com/a-single-calibration-method-for-water-andsoil-samples/, EST Analytical, published 3, 2015, Web 4, 2016. For More Information

For more information on our products and services, visit our website www.estanalytical.com/products. EST analytical shall not be liable for errors contained herein or for incidental or consequential damages in connection with this publication. Information, descriptions, and specifications in this publication are subject to change without notice