Performance Test of the Electronic Personal Neutron Dosemeter in Neutron Fields Simulating Workplaces of MOX Fuel Fabrication Facilities

Similar documents
Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Personal Dose Monitoring System

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

Measurements with the new PHE Neutron Survey Instrument

Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace

The development of new generation electronic personal dosimeters

Neutron and/or photon response of a TLD-albedo personal dosemeter on an ISO slab phantom

Development And Testing of a Thermoluminescent Dosemeter for Mixed Neutron-Photon-Beta Radiation Fields

Neutron Dosimetry with Ion Chamber-Based DIS System

Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

Calibration of the GNU and HSREM neutron survey instruments

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 11 May 2007

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

REALISTIC NEUTRON SPECTRA FOR RADIATION PROTECTION AND OTHER APPLICATIONS AT AERI, BUDAPEST

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

A new neutron monitor for pulsed fields at high-energy accelerators

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005

Neutron Metrology Activities at CIAE (2009~2010)

Recent activities in neutron standardization at NMIJ/AIST

Update on Calibration Studies of the Canadian High-Energy Neutron Spectrometry System (CHENSS)

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Neutron Users Club Meeting

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Neutron Dose near Spent Nuclear Fuel and HAW after the 2007 ICRP Recommendations

Neutron dose assessments for MATROSHKA using the HPA PADC dosemeter

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

He-3 Neutron Detectors

Radiation Protection Dosimetry (2007), Vol. 126, No. 1 4, pp Advance Access publication 23 May 2007

Altitude Variation of cosmic-ray neutron energy spectrum and ambient dose equivalent at Mt.Fuji in Japan

Introduction. Neutron Effects NSEU. Neutron Testing Basics User Requirements Conclusions

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng

Commissioning of the Beta Secondary Standard (BSS2)

Characterization of an 241 AmBe neutron irradiation facility by different spectrometric techniques

Elastic Recoil Detection Method using DT Neutrons for Hydrogen Isotope Analysis in Fusion Materials. Abstract

Recent developments in neutron metrology at the Institute for Radiological Protection and Nuclear Safety (IRSN)

Neutronics Experiments for ITER at JAERI/FNS

IRPA th International Congress of The International Radiation Protection Association

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Reactor radiation skyshine calculations with TRIPOLI-4 code for Baikal-1 experiments

Benchmark Test of JENDL High Energy File with MCNP

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Integral Benchmark Experiments of the Japanese Evaluated Nuclear Data Library (JENDL)-3.3 for the Fusion Reactor Design

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

CHEMISTRY 170. Radioisotopes

THE ACTIVE PERSONNEL DOSIMETER - APFEL ENTERPRISES SUPERHEATED DROP DETECTOR*

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Ion-Chamber Survey Meter OD-02

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

ISO INTERNATIONAL STANDARD. Reference radiation fields Simulated workplace neutron fields Part 1: Characteristics and methods of production

Researchers at the University of Missouri-Columbia have designed a triple crystal

Response characteristics of neutron survey instruments. Rick Tanner and David Bartlett, NRPB Hamid Tagziria and David Thomas, NPL

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Reliability and Traceability in Radiation Calibration

Calculation of Bubble Detector Response Using Data from the Matroshka-R Study

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

Neutron dosimetry and microdosimetry with track etch based LET spectrometer

ISO INTERNATIONAL STANDARD

CHEMISTRY 130 General Chemistry I. Radioisotopes

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Am-241 as a Metabolic Tracer for Inhaled Pu Nitrate in External Chest Counting

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Applications of MCBEND

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

A Monte Carlo Simulation for Estimating of the Flux in a Novel Neutron Activation System using 252 Cf Source

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Occupational Radiation Protection at Accelerator Facilities: Challenges

WM 07 Conference, February 25 March 01, 2007, Tucson, AZ

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

EXPERIENCE OF TEST OPERATION FOR REMOVAL OF FISSION PRODUCT NUCLIDES IN TRU-LIQUID WASTE AND CONCENTRATED NITRIC ACID USING INORGANIC ION EXCHANGERS

PERSONNEL DOSE EQUIVALENT MONITORING AT SLAC USING LITHIUM-FLUORIDE TLD S*

NEUTRON SPECTROMETRY AND DOSIMETRY WITH ANNs

Introduction. Principle of Operation

Survey Meter OD-01 Address: Phone: Fax: URL:

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Benchmark Tests of Gamma-Ray Production Data in JENDL-3 for Some Important Nuclides

Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience.

Chapter VIII: Photographic films

Atoms, Radiation, and Radiation Protection

Progress in Nuclear Science and Technology, Volume 6,

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Radiation Detection. 15 th Annual OSC Readiness Training Program.

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

Neutron Interactions with Matter

CVD Diamond History Introduction to DDL Properties of Diamond DDL Proprietary Contact Technology Detector Applications BDD Sensors

anti-compton BGO detector

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

INTRODUCTION TO MEDICAL PHYSICS 1 Quiz #1 Solutions October 6, 2017

Transcription:

Performance Test of the Electronic Personal Neutron Dosemeter in Neutron Fields Simulating Workplaces of MOX Fuel Fabrication Facilities N. Tsujimura a*, T. Yoshida a, C. Takada a, T. Nunomiya b and K. Aoyama b a Nuclear Fuel Cycle Engineering Laboratories, Japan Atomic Energy Agency, 4-33, Tokai-mura, Ibaraki-ken 319-1194, Japan. b Fuji Electric Systems Co., Ltd., Fuji-cho 1, Hino, Tokyo 191-8502, Japan. Abstract. Recently, several electronic personal dosemeters (EPDs) capable of measuring neutron dose equivalents were developed and have appeared on the market. The authors initiated the present study to assess performance of neutron EPDs. The EPD selected for this study was NRY21, manufactured by Fuji Electric Systems Co., Ltd. This dosemeter uses two silicon semiconductor detectors (a fast neutron sensor and a thermal neutron sensor) to provide wide measurable energy ranges from thermal to several tens of MeV. In the test, we focused on the energy and angular dependences of dose equivalent responses because they were of great importance in neutron dosimetry in work environments. Measurements of dose equivalent responses were performed in a moderated neutron field simulating neutron spectra likely to be encountered in workplaces at MOX fuel facilities. As a result, the dose equivalent responses of this EPD were evaluated to be 0.7 to 1.1 for a normal incidence of neutron irradiation. This is a satisfactory performance for neutron monitoring. KEYWORDS: electronic personal dosemeter (EPD); neutron; thermoluminescence albedo dosemeter; performance tests; moderated neutron field; MOX fuel facility 1. Introduction Recently, several electronic personal dosemeters (EPDs) capable of measuring neutrons dose equivalents were developed and have become available on the market. EPDs offer a number of advantages such as simultaneous measurements of photon, beta and neutron doses, the ability to set dose rate/dose alarms, computer interfacing with radiological control systems, etc. They appear to be promising devices which can replace or supplement the passive personal dosemeters; however, they also pose problems to some extent in terms of the accuracy of their measurements of neutron dose equivalents [1-2]. Although it is generally required for active personal neutron dosemeters to respond to dose equivalents to within a factor of two (although the latest IEC criteria [3] are actually more complicated than this) over a wide (or rated) energy and angle range, the overall performances observed in tests at monoenergetic neutron fields were less satisfactory than those for photon dosemeters. Among most of currently available EPDs for neutrons, the best dosemeter still depends on the neutron spectrum in which the measurement of the neutron dose equivalent is made. In other words, it is more practically important to establish the relationship between the neutron dose equivalent and the response of the dosemeter in the neutron field in which it is used. This idea is partly supported by the IEC s recommendation to test the responses of dosemeters to simulated workplace neutron fields as an optional test in addition to their responses to monoenergetic neutron or radioisotopic neutron irradiation [3], and also by the recent comprehensive studies performed under the framework of the EVIDOS project [4-5]. The Nuclear Fuel Cycle Engineering Laboratories (NCL) of the Japan Atomic Energy Agency (formerly, Japan Nuclear Cycle Development Institute) operates the pilot plants for spent fuel reprocessing and MOX fuel fabrication. In personal monitoring, thermoluminescent dosemeters (TLDs) have been utilized as the dose-of-record dosemeters since the 1970s. TLDs have been well characterized and understood, and hence have also functioned successfully in routine monitoring programs. Moreover, in the reprocessing plant, EPDs for photons have been used as supplemental dosimeters and have become firmly entrenched as the preferred method for achieving ALARA goals, * Presenting author, E-mail: tsujimura.norio@jaea.go.jp 1

while in the MOX fuel fabrication facilities no EPDs capable of measuring neutrons and photons are employed yet, mainly due to the inadequacies in their neutron dose measurement capability. For this reason, the authors initiated the present study to assess the performances of neutron EPDs to determine their capacity to be used as dosemeters supplemental to TLDs. From a practical point of view, we focused on their dose measurement accuracy and suitability for use in work environments in MOX fuel facilities. The neutron EPD assessed in this study is the NRY21 electronic neutron/gamma personal dosemeter manufactured by Fuji Electric Systems Co., Ltd. This dosemeter uses two silicon semiconductor detectors to cover a wide measurable neutron energy range from thermal to several dozen of MeV. This paper presents the test results for the NRY21 dosemeters in moderated neutron fields simulating the fields likely to be encountered at workplaces in MOX fuel facilities. 2. Dosemeter description 2.1 NRY21 Electronic Personal Neutron/Gamma Dosemeter The model NRY21 electronic personal neutron/gamma dosemeter is shown in Fig. 1. The history of this dosemeter can be traced back to the late 1980s; since then, a great deal of research and progress reports on its basic characteristics and field-test results at various nuclear installations have been published [6-11]. The dosemeter uses two silicon semiconductor detectors (a fast neutron sensor and a slow neutron sensor) for neutron measurements. Each sensor is a 10 10 mm 2 p-type silicon crystal on which an amorphous silicon is deposited. The fast neutron sensor detects recoil protons produced from a polyethylene radiator of ~0.08 mm (~80 µm) in thickness by neutron-hydrogen nuclei interactions. The slow neutron sensor detects alpha- and Li-particles produced from a natural boron layer formed on the crystal. The fast neutron sensor is sensitive to neutrons > ~1 MeV, which corresponds to neutrons with energies above the threshold level (set at ~0.9 MeV) for neutron- and gamma-pulses discrimination. In contrast, the slow neutron sensor, categorized as a typical albedo detector, is sensitive to neutrons < ~1 MeV, specifically from thermal to ~10 kev. The personal dose equivalent (H p (10)) is derived by adding the dose equivalents for two neutron energy intervals, namely a fast component H f (> 1 MeV) and a slow component H s (< 1 MeV), which are separately measured by the fast and slow neutron sensors, respectively. This relatively simple dose evaluation is advantageous for highly reliable operation and good maintainability. In the factory setting, the fast and slow neutron sensors are calibrated using a bare 252 Cf neutron source and a graphite pile-moderated thermal neutron field, respectively [11]. Figure 1: External view of the NRY21 dosemeter. (Size: 103 55 15 mm 3, weight: 110 g with battery) The energy dependence of the combined dose equivalent response (dosemeter-indicated H p (10) per reference H p (10)) of this dosemeter at normal incidence is shown in Fig. 2, which is reproduced from Fig. 4 of Nunomiya [11]. Although this dosemeter has two discrete neutron-measurable ranges, showing unity at two energy points calibrated by 252 Cf (~2 MeV) and thermal neutrons, it exhibits a significant response dip around several hundred kev. This is attributed to the low counting efficiencies for both recoil protons (compatible with good signal discrimination) and albedo neutrons in such 2

incident neutron energy regions. The energy dependence of the combination response over the energy region from thermal to MeV is similar to that of a nuclear track emulsion dosemeter (NTED) and thermoluminescent albedo dosemeter (TLAD). It is well known [12] that NTEDs are essentially insensitive to neutrons below ~800 kev, and therefore are suitable only for occupational environments in which the majority of dose equivalents comes from neutrons with energies above 1 MeV or in which the fraction of neutrons below 1 MeV is known. It is anticipated that the principal weakness of the NTEDs plus TLADs combination dosimetry system will consequently apply to that of the NRY21 dosimeters. d Errico [13] suggested that the NRY21 dosemeter can seriously underestimate the neutron dose equivalent in the energy region from 10 kev to 1 MeV, which is important for some nuclear power plants. Figure 2: Personal dose equivalent response of the NRY21 dosemeter as a function of neutron energy in angles of normal incidence. This graph is reproduced from Fig. 4 of Nunomiya [11]. 2.2 Panasonic UD-809 Thermoluminescence Albedo Dosemeter For comparison with the NRY21 dosemeters, we also tested the Panasonic UD-809 thermoluminescence albedo neutron dosemeter. This dosemeter has been used as the dose-of-record neutron dosemeter for a long time at nuclear facilities of the NCL. The UD-809 TLAD has four elements of Li 2 B 4 O 7 (Cu) phosphor. Element 1 is enriched in 7 Li and 11 B, and is insensitive to neutrons. Elements 2, 3, and 4 are enriched in 6 Li and 10 B and are very sensitive to neutrons. The reading to incident neutrons can be derived from readings of four elements with different filtration by cadmium and tin in front and behind the elements. 3. Test Methods 3.1 Test Facility Description The neutron irradiation tests were performed at the Instrument Calibration Facility (ICF) of the NCL. The ICF was specifically constructed to provide in-house calibration of the radiation-measuring devices being used for radiological monitoring in the reprocessing plant, MOX fuel fabrication facilities and other related research laboratories of NCL. The calibration facility is a one-story building made up of two irradiation rooms. The low-scatter neutron irradiation room is 13.0 m 12.5 m 5.3 m (height from the ground level) equipped with an open pit of 2.0 m in depth. The roof and walls which face the outside of the facility are made of an autoclaved aerated concrete with a thickness of 10 cm for reducing room-scattered neutrons. The floor is made of a steel grating which acts as a pseudo ceiling for the pit room (basement), which has an 8 m 12 m concrete floor and four concrete walls of 2-m in height. Although the radio-isotopic neutron sources of 252 Cf (nominal activity of ~1 GBq) and 241 Am-Be (~111 GBq) are generally used for routine calibration of the neutron-measuring devices, the moderated neutron calibration field what we call the simulated workplace neutron field consisting of a 252 Cf source plus moderators was also established for determining the calibration factors specific to workplaces and for the performance testing of neutron dosemeters [15]. Moderators 3

and a 252 Cf source were arranged to mimic the neutron spectral and dosimetric conditions of workplaces, based on the neutron spectral measurements at the representative workplaces of ~70 points. Figure 3 schematically shows a vertical cross-sectional view of the irradiation room and an experimental set-up. The 252 Cf source of known neutron emission rate is remotely moved to the irradiation positions of A and B on a guide tube from a storage vault using a pneumatic or mechanical source transfer system. Annular cylinders or slabs made of steel, graphite and acrylic plastic can be placed surrounding the source in order to alter the initial fission neutron spectrum. The source position A is at a height of 126 cm from the steel grating floor of a large low-scatter room and can produce neutron fields simulating the spectrum of neutrons transmitted through MOX-handling glove boxes and their shielding windows. Position B is at a height of 126 cm from the concrete floor of a concretewalled basement, and produces a neutron field simulating the high-scatter neutron spectrum typical of work environments where large glove boxes containing MOX fuels are installed in a concrete wallenclosed room. One major difference between positions A and B is the directionality of the fields. Due to the greater scattering in the small rooms, the fields of position B are more multidirectional than those of position A. Figure 4 shows reference neutron spectra measured with the Bonner multisphere spectrometer at the point of the test. The neutron emission rate is normalized to 1. Each spectrum is assigned by abbreviated moderator identification code. In the graphical presentation, ordinates are multiplied by the indicated factors to avoid overlaps. The effects of acrylic plastic shielding are apparent in these figures; the peaks in the 252 Cf spectrum shift to slightly lower energies as the amount of shielding increases, and the number of low-energy neutrons is significantly increased. The fission neutron spectrum is markedly degraded to the lower energies due to inelastic scattering with steel. Thermal and epithermal neutrons are fairly dominant components, especially at position B. Table 1 lists moderator arrangements and neutron field parameters, E (fluence average energy, MeV), h (spectrum average dose equivalent conversion coefficients, psv cm 2 ), and θ (average angle of neutron incidence, degree), showing neutron spectral and dosimetric characteristics. These parameters were determined by the angle-integrated neutron energy distribution measured with the Bonner multispheres spectrometer and the energy and angular distributions obtained from Monte-Carlo calculations. In column 7 of Table 1, the dosimetric parameters obtained from BS measurements at representative locations are also presented for comparison. The shapes of the neutron spectra and dosimetric parameters of the calibration field entirely reproduce and cover those likely to be found in workplaces. The field meets the ISO 12789 guidelines [16]. Figure 3: Vertical cross-sectional view of the irradiation room and experimental set-up. The 252 Cf neutron sources can be placed at the two positions of A and B. A ground level B 4

Figure 4: Neutron spectral fluence per unit lethargy interval for the 252 Cf source positions of A and B. The spectra were measured with a Bonner multi-sphere spectrometer at a distance of 100 cm. The neutron emission rate is normalized to 1. Each spectrum is assigned by abbreviated moderator identification code (see Table 1), and ordinates are multiplied by the indicated factors to avoid overlap. The vertical broken lines correspond to the peaks of the uncollided fission spectrum. Table 1 : Neutron dosimetric parameters showing characteristics of neutron spectra at a distance of 75 cm Moderator Source Moderator (a) E h (b) θ (c) Remarks (d) ID Code Position (MeV) (psv cm 2 ) (deg.) (Workplaces) Cf A None 2.0 (e) 380 (e) F A Steel 40 mm 1.5 362 ~6 P15 A PMMA 15mm 1.7 332 ~7 P35 A PMMA 35mm 1.4 271 ~8 P60 A PMMA 60mm 1.2 225 ~8 P100 A PMMA 100mm 0.97 183 ~8 FP A Steel 40 mm + PMMA 60mm 0.86 191 ~9 B-Cf B None 1.6 336 ~25 B-P B PMMA 100mm 0.80 157 ~30 B-G B Graphite 100mm 0.90 182 ~27 B-F B Steel 100mm 0.72 185 ~29 MOX powder can (source term) E = ~1.6 MeV h = 340 psv cm 2 Glove box area E = 0.45~0.75 MeV h = 100~170 psv cm 2 Graphite 100mm B-GP B 0.72 136 ~43 + PMMA 100mm Steel 100mm B-FP B 0.50 118 ~45 + PMMA 100mm (a) Moderator materials and thickness of moderators placed between a 252 Cf source and the dosemeters. (b) Spectrum-averaged conversion coefficients from neutron fluence to ambient dose equivalent. (c) Average angle of neutron incidence to the plane of which the normal vector points toward the source. These values were derived from the ratio of neutron spectral current to spectral fluence for neutrons incident from the frontal half space, based on the Monte Carlo computations. (d) Derived from the Bonner multisphere spectrometer measurements at the workplaces. (e) Minor differences of values from the ISO are ascribed to the neutron scattering with a source holder and source support structures. 5

3.2 Experimental set-up The experiments were performed using both the bare and moderated 252 Cf neutron source. An experimental set-up was already presented in Fig. 3. The response of the dosemeters was determined by exposing them at locations where the reference personal dose equivalent has been determined. All irradiations were made with EPDs and TLADs mounted on an ISO water-filled parallelepiped phantom with an outer size of 30 cm 30 cm 15 cm. The distance from the source center to the face of the phantom was 75 cm. Two EPDs, serial numbers 000001 and 000002, were simultaneously irradiated, while three TLADs were irradiated as a group. The effective center of the grouped dosemeters was positioned at a height of 126 cm above the floor, which is a typical height for a personal dosemeter worn at the chest. For irradiations at position A in the low-scatter condition, the phantom was set at 0º and 60º. Since experience showed that the room-scatter corrections are almost negligible for a distance of 75 cm [17], no corrections were made. For irradiations at position of B in the high-scatter condition, the phantom was fixed facing the source (0º), and room-scattered neutrons were included in the measurements. 4. Results and Discussion 4.1 Test Results We summarize the response data from the viewpoint of the variation in relation to spectral transition from a fission spectrum, i.e., an initial neutron spectrum. Figure 5 presents the neutron spectraldependence of dose equivalent responses of both NRY21 and TLADs, namely the relationship between the spectrum-weighted dose equivalent responses of both dosemeters (NRY21 and TLADs) and the spectrum-weighted neutron energy. Note that the figure is for broad neutron spectra, not for monoenergetic neutrons. In this figure, the responses of both dosemeters are normalized to unity at a 252 Cf (~2.0 MeV) since it corresponds to the starting point in the neutron spectral variation. A decrease in neutron energy represents a significant increase of the relative magnitude of thermal and epithermal neutron components. First, the test results for the TLADs are shown since they are easy to explain. The experimental data shown in Fig. 5 demonstrate the strong spectrum-dependent nature of the albedo-type dosemeter. As the neutron spectrum decreases in energy, TLADs tend to overly respond. The TLADs can overestimate the dose equivalent by a factor of up to 3 to 4 around the average neutron energy of ~0.5 MeV, which is typical around the MOX-handling glove boxes. In routine neutron monitoring with TLADs, a location-specific correction factor of 0.3 has been applied based on the field calibration method performed near the glove boxes in the 1990s. The value of this factor is reasonably consistent with a reciprocal of the experimental responses (3~4) shown in Fig. 5. In contrast to the results of TLADs, the test results on the NRY21 dosemeters show a nearly spectrumindependent response. The dose equivalent response of the NRY21 dosemeter was evaluated to be 0.7 to 1.1 at angles of normal incidence at irradiation position A and with the dosemeter s reference orientation facing the source at irradiation position B. A simple explanation of this neutron spectral dependence in response is that in the test neutron field, most neutrons that contribute significantly to the dose equivalent have energies above several hundred kev, and hence the fast neutron sensor can function well without missing large portions of neutrons with energy below the threshold, while the slow neutron sensor virtually does not work. The thickness of the moderators used in the test fields was less than 100 cm of acrylic plastic or steel; therefore, they did not alter the fission neutron spectrum sufficiently to introduce the large dose evaluation errors that are ascribed to the response inadequacies in Fig. 2. The energy response of the NRY21 dosemeter seems to be at least adequate for a bare and lightly-moderated fission sources. The angular response was also tested using a 252 Cf plus steel/pmma combination moderator source (moderator code = FP). The conclusion of Sasaki s work was that the NRY21 dosemeter has a cosinelike angular dependence for a bare 252 Cf neutron source. Similar results were obtained for a moderated 6

252 Cf source. The dose equivalent response at 60º relative to the 0º response was 0.6, slightly mitigated relative to the expected value (~0.5) from the cosine curve, because of the angular dependence of H p (10) itself. Figure 5: Personal dose equivalent response of the NRY21 dosemeter and TLAD as a function of neutron energy averaged over the incident neutron spectrum. The responses of both dosemeters are set to unity at uncollided 252 Cf neutrons (~2 MeV). 4.2 Discussion From the experimental results, we can draw the conclusion that for unmoderated or lightly moderated fission neutron fields likely to be found in the workplaces of the MOX fuel fabrication facilities, the NRY21 dosemeter functions satisfactorily in the factory calibration setting. Conversely, it is anticipated that this dosemeter may only measure a small fraction of the neutron dose equivalent in highly-moderated or degraded neutron spectra typical in power reactors because of the threshold at ~1 MeV for the fast neutron sensor. For instance, for a 1/E spectrum extending from 100 kev to thermal, the NRY21 dosemeter behaves just like an albedo neutron dosemeter, since the fast neutron sensor does not work. In such cases, instead of applying the factory calibration setting at thermal neutrons, the appropriate calibration factor specific to the workplace should be applied for the slow neutron sensor so as to account for the correct neutron dose equivalents. 5. Conclusion The authors investigated the neutron dose equivalent evaluation performance of the NRY21 electronic neutron personal dosemeter and discussed its suitability to work environments in MOX fuel fabrication facilities. All tests were conducted at the moderated 252 Cf neutron fields having neutron spectral and dosimetric characteristics similar to those found in the workplaces of MOX fuel facilities. Experiments showed that the response of this dosemeter was evaluated to be 0.7 to 1.1 in angles of normal incidence. This is satisfactory, and also provides the expected data on how the dosemeters will respond to different neutron spectra in actual workplaces. From the experiments, it is concluded that for lightly-moderated fission neutron fields like MOX fuel facilities, this dosemeter functions well in factory calibration setting. If the dosemeters were used in unknown neutron spectrum or highlymoderated spectra such as those encountered at commercial power reactors, new calibration factors, not factory calibration settings, should be used to improve their dosimetry performances. 7

REFERENCES [1] BARTLETT, D. T., et al., Active neutron personal dosemeters a review of current status, Radiat. Prot. Dosim. 86, 107-122 (1999). [2] BOLOGNESE-MILSZTAJN, T., et al., Active methods and instruments for personal dosimetry of external radiation: present situation in Europe and future needs, Proceedings of the 11 th International Congress of the International Radiation Protection Association (IRPA-11), Madrid, (2004) (CD-ROM). [3] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Radiation protection instrumentation Measurement of personal dose equivalents H p (10) and H p (0.07) for X, gamma, neutron and beta radiations Direct reading personal dose equivalent meters and monitors, IEC61526-2005 (2005). [4] D ERRICO, F., et al., Evaluation of individual dosimetry in mixed neutron and photon radiation fields (EVIDOS). Part I: scope and methods of the project, Radiat. Prot. Dosim. 125, 275-280 (2007). [5] SCHUHMACHER, H., et al., Evaluation of individual dosimetry in mixed neutron and photon radiation fields (EVIDOS). Part II: conclusions and recommendations, Radiat. Prot. Dosim. 125, 281-284 (2007). [6] NAKAMURA, T., et al., A real time wide energy range personnel neutron dosemeter with two silicon detectors, Radiat. Prot. Dosim. 27, 149-156 (1989). [7] NAKAMURA, T., et al., Development of real time personal neutron dosimeter with two silicon detectors, Proceedings of IRPA-8, Montreal, Vol. 1, 152-155 (1992). [8] SASAKI, M., et al., Development and characterization of real-time personal dosemeter with two silicon detectors, Nucl. Instrum. Meth. Res. A 418, 465-475 (1998). [9] AOYAMA, K., et al., The development of new generation electronic personal dosimeters, Proceedings of IRPA-10, Hiroshima, (2000) (CD-ROM). [10] NUNOMIYA, T., et al., Energy response for high-energy neutrons of multi-functional electronic personal dosemeter, Proceedings of IRPA-11, Madrid, (2004) (CD-ROM, file 3h48.pdf). [11] NUNOMIYA, T., et al., Development of advanced-type multi-functional electronic personal dosemeter, Radiat. Prot. Dosim., 126, 284-287 (2007). [12] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS, Determination of operational dose equivalent quantities for neutrons, ICRU Report 66, J. ICRU 1, (2001). [13] D ERRICO, et al., State of the art of electronic personal dosimeters for neutrons, Nucl. Instrum. Meth. Phys. Res. A 505, 411-414 (2003). [14] TSUJIMURA, N., et al., Characteristics of the simulated workplace neutron fields using a 252 Cf source surrounded with cylindrical moderators, Radiat. Prot. Dosim. 110, 117-121 (2004). [15] TSUJIMURA, N., et al., Development of moderated neutron calibration fields simulating workplaces of MOX fuel facilities, HOKEN BUTSURI (Jpn. J. Health Phys.) 40, 354-359 (2005), (in Japanese). [16] INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Reference Neutron Radiations Characteristics and methods of production of simulated workplace neutron fields, ISO 12789, Geneva (2000). [17] YOSHIDA, T., et al., Evaluation of influence by room-scattered neutrons and source-tophantom distance effect for calibration of personal neutron dosemeters, HOKEN BUTSURI (Jpn. J. Health Phys.) 40, 37-42 (2005), (in Japanese). 8