Analysis of a national low dc resistance inter-laboratories comparison

Similar documents
1 Phone , fax , 2 Phone , fax ,

Determination of the Blockage Effect on a Thermal Anemometer using a Small Open Jet Wind Tunnel

Experimental Modal Analysis

792A AC/DC Transfer Standard

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

792A AC/DC Transfer Standard

INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A

Certificate of Accreditation to ISO/IEC 17025:2005

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

STATISTIQUE DE PERFORMANCE

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Document No: TR 12 Issue No: 1

Analysis of interlaboratory comparison when the measurements are not normally distributed

Calibration of Temperature Block Calibrators

Experiment 4: Resistances in Circuits

R SANAS Page 1 of 7

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Final Report August 2010

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

Calibration of Analog Scales Based on. a Metrological Comparison Between. the SIM Guide and OIML R 76-1

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

JAB NOTE4. ESTIMATION OF MEASUREMENT UNCERTAINTY (Electrical Testing / High Power Testing) Japan Accreditation Board (JAB)

Prepare for this experiment!

Calibration and verification: Two procedures having comparable objectives and results

Uncertainty Analysis in High-Speed Multifunction Data Acquisition Device. M.Catelani, L.Ciani, S.Giovannetti, A.Zanobini

Calibration of temperature sensors within Length Standard Section of NMIJ

The Fundamentals of Moisture Calibration

Vocabulary of Metrology

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

Uncertainty of Measurement A Concept

TEST REPORT ENVIRONMENTAL EN 60529:1991+A1:2000

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

IMPROVING SOIL MEASUREMENTS FOR BOUNDARY LAYER RESEARCH

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

A Case for Periodic Calibration or Verification of RTDs

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

Exercise 1: Thermocouple Characteristics

UNCERTAINTY ANALYSIS FOR LABORATORY ACCREDITATION. Peter B. Crisp. Fluke Precision Measurement Ltd, 52 Hurricane Way, Norwich, UK

Temperature Measurements

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Conformity assessment of a high-performance liquid chromatography calibration.

Interlaboratory Comparison & Proficiency Testing Program

World Meteorological Organization

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

Experimental Analysis of Random Errors for. Calibration of Scales Applying. Non-Parametric Statistics

BUREAU INTERNATIONAL DES POIDS ET MESURES

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

by S. Solve +, R. Chayramy +, M. Stock +, D. Vlad*

DUBLIN INSTITUTE OF TECHNOLOGY Kevin Street, Dublin 8.

For the National Voluntary Laboratory Accreditation Program

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

ISO/IEC requirements. General matters of quality system, requirements for documentation and the list of the documents required for a

Uncertainty, Measurement, and Models. Lecture 2 Physics 2CL Summer Session 2011

MECHANICAL ENGINEERING TECHNOLOGY ESSENTIALS FOR LABORATORY REPORTS

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

Lab E3: The Wheatstone Bridge

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

Euramet EM-S40. Bilateral Comparison KIM-LIPI / LNE. Final Report

For the National Voluntary Laboratory Accreditation Program

Practical Interpretation of Metrological Traceability on Gauge Block as to VIM 3

DEVELOPMENT OF A FLOW-THROUGH CELL FOR ACCURATE MEASUREMENTS OF LOW ELECTROLYTIC CONDUCTIVITY

Introduction of RIC Tsukuba (Japan, RAII)

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

BUREAU INTERNATIONAL DES POIDS ET MESURES

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

Experiment 5 Voltage Divider Rule for Series Circuits

Mason Technology Ltd

ISO INTERNATIONAL STANDARD. Rolling bearings Thermal speed rating Calculation and coefficients

Commissioning of the Beta Secondary Standard (BSS2)

A calibration facility for automatic weather stations

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

Glossary Innovative Measurement Solutions

PREDICTION OF THE OUTPUT VOLTAGE OF DC VOLTAGE STANDARDS

3B SCIENTIFIC PHYSICS

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Now let s look at some devices that don t have a constant resistance.

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

Interlaboratory Comparison & Proficiency Testing Program

IC Temperature Sensor Provides Thermocouple Cold-Junction Compensation

Design and Validation of an Automated Hydrometers Calibration System

The Uncertainty Principle

P1-APMP.EM-S9 on supplementary comparison. VNIIM/KRISS Bilateral Comparison of DC Magnetic Flux Density by Means of a Transfer Standard Coil

Interlaboratory Comparison & Proficiency Testing Program

Determination of Avogadro s Number via Electrolysis

MEASUREMENT OF WBGT INDEX IN AXIAL FLOW AS THE MAIN THERMODYNAMIC PARAMETERS VARY G.

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045

Study on a thermal diffusivity standard for the laser flash method measurements 1

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

CALIBRATION CERTIFICATE # 503

Transcription:

16 th International Congress of Metrology, 04008 (2013) DOI: 10.1051/ metrology/201304008 C Owned by the authors, published by EDP Sciences, 2013 Analysis of a national low dc resistance inter-laboratories comparison F. Galliana 1, E. Gasparotto 2, P. P. Capra 3 1,2,3 National Institute of Metrological Research, () str. delle Cacce, 91 10135 (TURIN Italy) Abstract. In this paper the description of an inter-laboratories comparison between the National Institute of Metrological Research () and nineteen Italian Secondary Calibration Electrical Laboratories, is given along with its main results in graphical way. The transportable standards were three Tinsley four terminal standard resistors to be calibrated in oil respectively at 5 A and 20 A (1 ), 1 A and 5 A (10 ), 0.1 A and 1 A (100 ). The comparison results were satisfactory, except only in a case of a Laboratory for which these measurements were outside its accreditation fields. 1 Introduction An inter-laboratory comparison aims to verify the compatibility of the measures of two or more Laboratories performed on the same measurand within their measurement uncertainties. Since its start the Italian Calibration Laboratories body (formerly Servizio di Taratura in Italia SIT) used to perform bilateral comparisons between the National Metrology Institutes and the Secondary calibration Laboratories to evaluate their technical competence in the framework of their first accreditation or in the periodical renewal of their accreditation. Normally, the results were elaborated according to the ISO Guide 43. Nowadays the number of Secondary Laboratories grew significantly, so the possibility to perform bilateral comparisons became quite difficult. A possible solution for this problem could be the use accredited of Inter-laboratory Comparisons providers (ILCs) [1, 2]. In this occasion the Italian Accreditation body (now ACCREDIA) gave to the National Institute of Metrological Research () the task to perform a an inter-laboratories comparison between and nineteen Italian Secondary calibration Electrical accredited Laboratories in the field of low dc resistance. The transportable standards for the comparison were three Tinsley mod 3504C 1, 10 and 100 four terminal standard resistors to be calibrated respectively at 5 A and 20 A (1 ), 1 A and 5 A (10 ), 0.1 A and 1 A. (100 ). The results of this comparison will be utilized by the accreditation body for the evaluation of the measurement competence of the participating Laboratories in the field of low dc resistance. The comparison took place from September 2012 to May 2013. During this period measured the travelling resistors before, after the circulation, and twice during the circulation. 2 Participating Laboratories The participating Laboratories were: - ASIT INSTRUMENTS, Orbassano (TO), accredited calibration laboratory (acl) 188; - ARO, Biassono (MI) acl 46; - AVIATRONIK, Samarate -(VA) acl 19; - CERMET, Bologna, acl 52; - DELO Services. s.r.l., Fizzonasco Pieve (MI), acl 121; - EMIT LAS, Milan, acl 24; - FASINTERNATIONAL, Milan, acl 128; - GAMMA MISURE, Calenzano - (FI), acl 56; - IMQ, Milan acl 21. - METRA, Augusta (SR), acl 188. - METRIX, S. Stefano Cisquina (AG), acl 171. - NEMKO, 4 Biassono (MI) acl 42; - QUALITY SERVICE, Venice, acl 112; - SIMAV S. Maurizio Can. - TO) acl 64. - SIMAV. Milan, acl 04; TECHNOSKY, Rome,acl 15; TELECOM, Turin, acl 72; TESI, Subbiano (AR), acl 101; TRESCAL, Zola Predosa (BO) acl 174. 3 Travelling Standards The transportable standards for the comparison were three Tinsley mod 3504C 1, 10 and 100 four terminal standard resistors to be calibrated respectively at 5 A and 20 A (1 ), 1 A and 5 A (10 ), 0.1 A and 1 A. (100 ). View of the model of these standards is given in Figure 1 and 2. The travelling package included Corresponding author: f.galliana@inrim.it This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2. 0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Article available at http://cfmetrologie.edpsciences.org or http://dx.doi.org/10.1051/metrology/201304008

Web of Conferences also a thermal collector for each resistor (Figure 1) to be inserted in the resistors to be in thermal contact Table 2. Power coefficients of the travelling resistors. Resistor 1 10 100 Power coeff 4.1.0 10 8 /mw 3.7 10 8 mw 8.1 10 8 /mw Figure. 1. Tinsley mod 3504C resistor with the thermal collector to put inside. with them and suitable to measure their temperature and a glass beaker to be filled with 500 ml of mineral oil. The measurements had to be carried out placing resistors inside it and acquiring the regime temperature of the thermal collector inside the resistor (Figure 2). 3.1 Time drift of the travelling resistors The 1 resistor showed a high stability following a 2 nd order polynomial drift of about 1.6 10 8 /day and 3.0 10 8 /day, respectively at 5 A and 20 A during the whole period in the measurements, with only a small rise in the middle period. 0.9994300 1 5 A 0.9994275 0.9994250 0.9994225 B1 B2 4.88112E-4-9.93588E-11 Figure 2. Tinsley mod 3504C resistor with the thermal collector inside the glass beaker. Before the start of the comparison the three resistors were characterized evaluating their power and temperature coefficients. These measurements were made by the Low Resistances Calibration Laboratory of and performed with a measurement method involving high precision standard resistors put in a thermo-regulated at 23±0.3) C oil-bath, a high performance current comparator bridge and a high stability power supply [3, 4]. The travelling resistors were placed in another thermo-regulated oil-bath with similar characteristics. The evaluated temperature and power coefficients of the travelling resistor are reported in Table 1 and 2. Table 1. Temperature coefficients of the travelling resistors. Resistor α β 1 1.0 10 5 / C 6.0 10 7 / C 2 10 9.3 10 6 / C 5.5 10 7 / C 2 100 8.9 10 6 / C 4.7 10 7 / C 2 0.9994200 0.9994175 20/08/2012 20/11/2012 20/02/2013 20/05/2013 Figure 3. measurements 2 nd order polynomial fit of the 1 resistor at 5 A. Also the 10 resistor showed a high stability with a 2 nd order polynomial drift of about 3.8 10 8 /day and 2.9 10 8 /day, respectively at 1 A and 5 A during the whole period of the comparison in the measurements, with also in this case a small rise in the middle period at 1 A, and smoothly decreasing at 5 A. 9.9991650 9.9991575 9.9991500 9.9991425 9.9991350 9.9991275 9.9991200 9.9991125 9.9991050 9.9990975 Equation y = Intercept + B1*x^1 + B2*x^2 Adj. R-Sq 0.95805 10 1 A B1 0.0013 3.3489 B2-2.755 6.8171 Figure 4. measurements 2 nd order polynomial fit of the 10 resistor at 1 A. 04008-p.2

16 th International Congress of Metrology Finally also the 100 resistor showed a high stability with linear decreasing drift of about 4.4 10 9 / day both at 0.1 A and 1 A during the whole period of the comparison in the measurements. 99.97435 99.97430 99.97425 99.97420 99.97415 99.97410 99.97405 99.97400 99.97395 99.97390 99.97385 100 0.1 A y = a + b*x Intercept 101.15937 0.4692 Slope -4.82545E-7 1.91016E-7 Figure 5. measurements linear fit on the 100 resistor at 0.1 A. 3.2 Calibration modalities The Laboratories had to perform the calibration of the resistors in four-terminal configuration at the current values and time applications shown in Table 3. If the adopted measurement method did not allow the choice of the measuring current, the applied current value and its application time had to be stated. Before starting the measurements it was suggested to wait two hours to stabilize the resistors in the oil. After performing the measurement at higher current, if was suggested to repeat the cycle after a waiting time of four hours. The reference ambient temperature of the laboratory had to be 23 C and the Laboratory had to indicate the ambient temperature at the time of calibration. The Laboratory had to acquire also the temperature at the thermal regime reached for each measurement current. This temperature had to be measured at the thermal shaft junction of the resistor with a thermometer or a temperature sensor; these values allowed to correct at 23 C the Laboratories measurements for a more correct evaluation of the compatibility with respect. Table 3: Resistors under calibration along with currents and their application times. Resistor 1 10 100 Meas. currents 5 A 20 A 1 A 5 A 0.1 A 1 A Application times 1 1 1 4 Measurement procedures The measurements were carried out with the previously mentioned measurement system involving high precision standard resistors and an high performance current comparator. The participating Secondary Laboratories measured the three standard resistors following their measurement procedures. In particular the involved measurement methods were: - Direct reading with a digital multimeter (DMM), in particular for the 100 standard at 0.1 A. For the other measurement points this method didn t allow to perform the measurements at upper current values, so those measurements were considered as not accredited ones. - Substitution method with a standard shunt or resistor connected in series with the resistor under calibration and supplied by a high stability current source (multifunction calibrator) and a DMM to measure the voltage drops on the standard and under calibration resistors. - Volt-ampere-metric method with a high precision current source (multifunction calibrator) and a DMM to measure the voltage drop on the resistor under calibration. 5 Analysis of the results To evaluate the measurement comparison the results of each measurement point were analyzed in the same way of [5, 6]. The measurement values obtained by the Laboratories and by were defined respectively as: m Lab ± u Lab and m I ± u I (1) where m Lab, u Lab, m I, u I are respectively the measurement values and the standard uncertainties of the Laboratories and of. Both and the Laboratories gave their results along with the reached regime temperature at each measurement current, so that all the measurement values were reported at the temperature of 23.0 C, starting from their temperature coefficients. For m I were taken into account the interpolated values of measurements at the date of the measurements of the Laboratories. With this interpolation, an uncertainty due to possible errors due to the fitting operation was added to the uncertainty. A new measurand: y Labr = (m Lab m I )/m I (2) was defined for each measurement point whose relative standard uncertainty is: 2 1 2 u y Lab, r I I 2 = [ u 2 (, )] 2 Lab + ui ulabui r mlab m (3) m where r(m Lab,m I ) is the correlation factor between the measurements of the Laboratories and those of, evaluated as the ratio between the type B component of the uncertainty with which calibrated the 04008-p.3

Web of Conferences reference standards of the Laboratories and the uncertainty declared by the same Laboratories for the comparison. Finally, the normalized error En with respect to for each measurement point and for each Laboratory was evaluated as: ylab, r En U 0.9995 = where U ylab,r = 2u ylab,r 2σ (4) (5) y Lab, r 6 Measurement results In Figure 6, the normalized errors En for each participant Laboratory and for each resistor and measurement current are reported as an estimate of the compatibility degree with measurements. 1.0010 1.0005 1.0000 0.9990 0.9985 1 20 A 0.9980 Figure 8. Results for the 1 at 20 A LAB 6 4 5 6 7 8 9 Normalized error En 2.5 2.0 1.5 1.0 0.5 0.0-0.5-1.0-1.5 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Laboratory Figure 6. Laboratories normalized errors. 1 mohm 5 A 1 mohm 20 A 10 mohm 1 A 10 mohm 5 A 100 mohm 0.1 A 100 mohm 1A. 10.020 10.015 10.010 10.005 10.000 9.995 9.990 9.985 10 1 A 9.980 LAB 5 4 5 6 7 8 9 In Figures from 7 to 12 and Laboratories results corrected at 23 C with their 2σ uncertainties, are shown. 1.0003 0.9996 0.9989 0.9982 0.9975 0.9968 0.9961 1 5 A 4 5 6 7 8 9 Figure 9. Results for the 10 at 1 A 10.015 10.010 10.005 10.000 9.995 9.990 10 5 A 9.985 date 4 5 6 7 8 9 Figure 10. Results for the 10 at 5 A. Lab 5 did not make this measurement. Figure 7. Results for the 1 at 5 A. 04008-p.4

16 th International Congress of Metrology 100.06 100.05 100.04 100.03 100.02 100.01 100.00 99.99 99.98 99.97 99.96 99.95 99.94 99.93 100 0.1 A 99.92 LAB5 4 5 6 7 8 9 3. P. Mac Martin, L. Kusters,, IEEE Trans. Instrum. Meas., Vol. IM 15, no. 5, pp:212 220, (1966). 4. P. Mac Martin, L. Kusters, IEEE Trans. Meas., Vol. IM 18, no. 4, pp.266 271, Dec. (1969). 5. F. Galliana, P.P. Capra, E. Gasparotto Measurement Vol. 42, pp. 1532 1540, (2009). 6. F. Galliana, P.P. Capra, E. Gasparotto, Measurement Vol. 45 no.3, pp. 615-621, (2012). Figure 11. Results for the 100 at 0.1 A. 100.04 100.02 100.00 99.98 99.96 99.94 100 1 A 4 5 6 7 8 9 99.92 Figure 12. Results for the 100 at 1 A 7 Discussion The condition that has to be satisfied for a correct traceability transfer from National Standards is that En < 1. In figure 6 it is shown that that the majority of the measurement points satisfied this condition. It was accurately verified that all the measurements performed under accreditation satisfied the previous condition. In fact the measurements with E n > 1 were made in particular by a Laboratory that, although accredited for dc resistance, these fields are out of its accreditation field. Other border line E n values were from Laboratories that were not able to perform the measurements at the correct currents (for example reading directly with a DMM) but issued calibration certificates according to their accredited procedures. Considering all these conditions it can be concluded that the traceability transfer to Secondary Electrical Laboratories the field of low dc resistance is correctly assured. References 1. ISO/IEC CD 17043 Conformity assessment General requirements for proficiency testing CASCO 05/2008, edition, 2008-03-14 previously ISO/IEC Guide 43-2 (1997). 2. EA-2/10. EA Policy for Participation in National and International Proficiency Testing Activities, rev.0, pp.20-20, (2001). 04008-p.5