Desalting and Buffer Exchange with Corning Spin-X UFCentrifugal Concentrators

Similar documents
Technical data and operating instructions. Vivaspin 4. Vivaspin 4 10K device for in vitro diagnostic use

Amicon Ultra-15 10K Centrifugal Filter Devices

Tube O Reactor. (Micro: Cat. # K, K, K) (Medi: Cat. # K, K, K)

Olympic B3 Summer Science Camp 2015 Lab 0

Preparing Extension Products for Electrophoresis

RayBio Nickel Magnetic Particles

Micro Float-A-Lyzer READY-TO-USE MICRODIALYSIS DEVICES. Product Information & Operating Instructions

Mag-Bind Soil DNA Kit. M preps M preps M preps

Spectra/Por. Float-A-Lyzer READY-TO-USE DIALYSIS DEVICES. Product Information & Operating Instructions

Bis sulfone Reagents. Figure 1.

Non-Interfering Protein Assay Kit Cat. No

TrioMol Isolation Reagent

TrioMol Isolation Reagent

Phenol-Chloroform reagents. Selection guide. OH ; MW : High quality reagents for use in nucleic acid purification.

N6,6 Posidyne Membranes

E.Z.N.A. MicroElute Clean-up Kits Table of Contents

Inserting grids into capsule Removing capsules from grid box Immunolabeling using template guide Preparing for TEM insertion

Malachite Green Phosphate Detection Kit Catalog Number: DY996

TaKaRa BCA Protein Assay Kit

Dual-Range TM BCA Protein Assay Kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

RNA Transport. R preps R preps

Lowry Protein Assay Kit KB test (96 well plate)

[ Care and Use Manual ]

Tetracyclines (TCs) ELISA Test Kit

11,12-EET/DHET ELISA kit

This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power.

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

For purification/separation of organic and aqueous phases from a two phase mixture within a 96 or 24 array tube system

For the quantitative measurement of PEGylated molecules in plasma, serum and cell culture media

Aflatoxin M1 (AFM1) ELISA Kit

Component Product # Product # Cell Lysis Reagent 100 ml 500 ml Product Insert 1 1

Please use only the valid version of the package insert provided with the kit. This kit is intended for Research Use Only.

Mouse Glutathione S Transferase Alpha 1 (GSTa1)

Gluten (Gliadin) Lateral Flow IIR

Semicarbazide (SEM) ELISA Quantitation Kit. Manual

Modified Adams Assay for Phenolics in Wine

Canine brain natriuretic peptide,bnp ELISA Kit

ab Deproteinizing Sample Preparation Kit TCA

Nitrofuran (SEM) ELISA test Kit

ab Caspase 3, Caspase 8 and Caspase 9 Multiplex Activity Assay Kit (Fluorometric)

Data sheet. UV-Tracer TM Biotin-Maleimide. For Labeling of Thiol-groups with UV-detectable Biotin CLK-B Description

Human papillomavirus,hpv ELISA Kit

MQCA ELISA Test Kit. Catalog No. LSY-10046

Effective desalting and concentration of in-gel digest samples with Vivapure C18 Micro spin columns prior to MALDI-TOF analysis.

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

OMX-S basic Kit Instruction Manual March/09

Mag Sepharose is available in the following formats (Instructions for use are included): μl TiO 2

OESTREICH LAB CHIP PROTOCOL

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit

PCBs Lower Chlorinated ELISA Kit

Shodex TM ODP2 HP series columns

Polyethylene Glycol (PEG), High Sensitive ELISA

ab MDA Assay Kit (competitive ELISA)

High Sensitivity Polyethylene Glycol (PEG) ELISA Kit

Nitrofuran (AOZ) ELISA Quantitation Kit. Manual

MyBioSource.com. This package insert must be read in its entirety before using this product.

IRDye 800CW Protein Labeling Kit Low MW

Alkaline Phosphatase Labeling Kit-NH2

ab Carrez Clarification Reagent Kit

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA

Scientific Observations and Reaction Stoichiometry: The Qualitative Analysis and Chemical Reactivity of Five White Powders

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG

GL Chromatodisc. Usage Type Description. GL Chromatodisc Type A GL Chromatodisc Type P GL Chromatodisc Type N 13AI

TPH (Total Petroleum Hydrocarbons)

Spherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois PARTICLE COATING PROCEDURES

Sulfonamide Residues (Three) ELISA Quantitation Kit. Manual

Protocol for Coating QD-COOH on glass slides Chris Ochs 19/09/12 Modified by Kathy Lu 2/27/2013

Cyproheptadine ELISA Test Kit

Protein separation and characterization

Lab 2. Go Their Separate Ways: Separation of an Acid, Base, and Neutral Substance by Acid-Base Extraction

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

mag nanogram kit Description Kit uses

Bisphenol A ELISA Kit

THE UNIVERSITY OF NEWCASTLE- DISCIPLINE OF MEDICAL BIOCHEMISTRY. STANDARD OPERATING PROCEDURE PROCEDURE NO: GLP 021 MOD: 1st Issue Page: 1 of 6

DNA can be extracted from the following sample types using this procedure: Archived

Fast Protein Quantification Kit (Bradford) KB tests (96 well plate)

Detection limit: grain, feed 500 ppb; milk 50 ppb; cream, cheese 5 ppb


Monensin ELISA Kit. Catalog Number KA assays Version: 11. Intended for research use only.

GENERAL INSTRUCTIONS GENERAL PREPARATION

Zearalenone ELISA Kit

Ribo-Zero Magnetic Gold Kit*

Melatonin Research RIA

For the detection and quantitation of glyphosate in water (groundwater, surface water, well water).

Eppendorf UVette. Instructions for use

Zearalenone Plate Kit PN 53018B

Glove Selection Chart

RIDASCREEN Aflatoxin B 1 30/15

Canine Erythropoietin,EPO ELISA Kit

SDS-PAGE (Sodium Dodecyl Sulfate Polyacrylamide Gel Electrophoresis):

Sampling and DNA Extraction from Wastewater Activated Sludge Standard Protocol

Protocol for 2D-E. Protein Extraction

AFM1 (Aflatoxin M1) ELISA Kit

Crustacean kit Ⅱ Maruha Nichiro

PREVENTING HAZARDOUS WASTE FINES

Data Sheet. Azide Cy5 RNA T7 Transcription Kit

Human Anti-Ovary Antibody (IgG)ELISA Kit

Multi-Prep and Detectabuse Columns and 96 Well Deep-Well Plates for Sample Preparation

Introduction to Work in Laboratory

Transcription:

Desalting and Buffer Exchange with Corning Spin-X UFCentrifugal Concentrators Protocol Table 1. Maximum Recommended Maximum Volume and (Centrifugal Force) for Spin-X UF Concentrators Concentrator Spin-X UF 500 Spin-X UF 6 Spin-X UF 20 Maximum Volume and (Spin Force Swing Bucket) 5,000 to 50,000 MWCO PES Do not use 6 ml (4,000 x g) 20 ml (5,000 x g) 100,000 MWCO PES Do not use 6 ml (4,000 x g) 20 ml (3,000 x g) Maximum Volume and (Spin Force Fixed Angle) 5,000 to 50,000 MWCO PES 500 µl (15,000 xg) 6 ml (10,000 x g) 14 ml (8,000 x g) 100,000 MWCO PES 500 µl (15,000 xg) 6 ml (6,000 x g) 14 ml (6,000 x g) Spin-X UF 500 Centrifugal Concentrator Spin-X UF 6 Centrifugal Concentrator Spin-X UF 20 Centrifugal Concentrator Introduction Spin-X UF centrifugal concentrators, with vertical membrane technology, combine fast filtration with high recovery of target proteins. This makes Spin-X UF the technology of choice for desalting or buffer exchange, avoiding lengthy dialysis steps. While proteins are retained by an appropriate ultrafiltration membrane, salts can pass freely through, independent of protein concentration or membrane molecular weight cut-off (MWCO). In consequence, the composition of the buffer in the flow-through and retentate is unchanged after protein concentration. By diluting the concentrate back to the original volume, the salt concentration is lowered. The concentrate can be diluted with water or salt-free buffer if simple desalting is required; however, it is also possible to dilute the concentrate with a new buffer, thereby exchanging the buffering substance entirely. For example, a 10 ml protein sample containing 500 mm salt concentrated 100-fold still contains 500 mm salt. If this concentrate is then diluted 100-fold with water or salt-free buffer, the protein concentration returns to normal, while the salt concentration is reduced 100-fold to only 5 mm, (i.e., a 99% reduction in salt). The protein sample can then be concentrated again to the desired level or the buffer exchange can be repeated to reduce the salt concentration even further before a final concentration of the protein. This process is called diafiltration. For proteins with a tendency to precipitate at higher concentrations, it is possible to perform several diafiltration steps in sequence with the protein concentrated each time to only 5- or 10-fold. For example, if a precipitous protein sample is concentrated 5-fold then diluted back to the original volume and this process is repeated a further two times, this still results in a greater than 99% reduction in salt concentration without over-concentrating the protein. Desalting and Buffer Exchange Procedure 1. Select the most appropriate MWCO for your sample. For maximum recovery, select a MWCO 1/2 to 1/3 the molecular size of the species of interest. 2. Add the solution to be desalted to the upper chamber of the Spin-X UF concentrator. See Table 1 to determine maximum volume, check the instructions that came with the concentrator. 3. If the sample is smaller than the maximum concentrator volume, it can be diluted up to the maximum volume with the replacement buffer before the first centrifugation step. This will help increase the salt removal rate. 4. Centrifuge for the recommended amount of time at an appropriate spin speed for your Spin-X UF concentrator (Table 1). When fixed angle rotors are used, angle the concentrator so that the printed window faces upwards (outwards). Note: This step concentrates the proteins in the solution remaining in the upper chamber but does not change the salt concentration in either chamber. 5. Empty the lower filtrate container which should now contain the salt or buffer solution minus the proteins. Note: Retain filtrate until the concentrated sample has been analyzed. 6. Refill the upper concentrator chamber with an appropriate replacement solvent or buffer. This dilutes both the proteins and remaining salts. 7. Centrifuge again as before. Note: This step concentrates the proteins in the solution remaining in the upper chamber while removing almost all of the salt. 8. Empty filtrate container. Note: Retain filtrate until the concentrated sample has been analyzed. 9. Recover the concentrated, desalted sample from the bottom of the concentrate pocket in the upper chamber with a pipettor fitted with a fine tip. Optimizing Solute Recovery When highest solute recoveries are most important, in particular when working with solute quantities in the microgram range, Corning recommends considering the following key points: Select the smallest concentrator that suits the sample volume. Additionally, take advantage of the extra speed of Spin-X UF concentrators by refilling a smaller concentrator repeatedly. Select the lowest MWCO membrane that suits the application. When available, use swing bucket rotors rather than fixed angle rotors (except for the Spin-X UF 500 which must always be run in a fixed angle rotor). This reduces the surface area of the concentrator that will be exposed to the solution during centrifugation. Reduce centrifugal force to approximately half of the maximum recommended (Table 1). Avoid over concentration. The smaller the final concentrated solute volume in the upper concentrator chamber, the more difficult it is to achieve complete recovery. If feasible, after a first recovery, rinse the concentrator with one or more drops of buffer and then recover again. Test Results As the results below show (Tables 2 and 3), the efficient design of Spin-X UF concentrators allowed greater than 95% of the salt to be removed during the first centrifugation step. Only one subsequent centrifugation step was needed to increase the typical salt removal to 99% with greater than 92% recovery of the sample.

Spin-X UF20ConcentratorResults Table 2. Results of Desalting Procedure Using Spin-X UF 20 Concentrators MWCO 5 kda 30 kda 50 kda 100 kda Cytochrome C BSA BSA IgG 0.25 mg/ml 1 mg/ml 1 mg/ml 1 mg/ml Protein NaCl Protein NaCl Protein NaCl Protein NaCl Recovery Removal Recovery Removal Recovery Removal Recovery Removal Spin 1 100% 99% 97% 99% 97% 99% 90% 98% Spin 2 96% 100% 92% 100% 93% 100% 87% 100% TreatmentofCorning Spin-X UF Concentrators for Improved Recovery of Low-Concentrated Protein Samples Protocol Four of Spin-X UF 20 concentrators of each cut-off were tested with 20 ml of solution. Each of the solutions contained 500 mm NaCl. Each spin was performed at 4,000 xg. The devices with greater than 5kDa MWCO were spun for 30 minutes. Concentrators with 5 kda MWCO were spun 45 minutes. After the first and second spin, the retentate was brought up to 20 ml with ultra pure water from an Arium purification system (Sartorius Stedim Biotech). OD readings were taken at 410 nm for the Cytochrome C and 280 nm for the BSA and IgG samples. Salt concentration was measured with a Qcond 2200 conductivity instrument. Spin-X UF6 Concentrator Results Table 3. Results of Desalting procedure using Spin-X UF 6 concentrators MWCO 5 kda 30 kda 50 kda 100 kda Cytochrome C BSA BSA IgG 0.25 mg/ml 1 mg/ml 1 mg/ml 1 mg/ml Protein NaCl Protein NaCl Protein NaCl Protein NaCl Recovery Removal Recovery Removal Recovery Removal Recovery Removal Spin 1 98% 99% 92% 99% 93% 99% 92% 98% Spin 2 85% 100% 86% 100% 83% 100% 89% 100% Four Spin-X UF 6 concentrators of each cut-off were tested with 6 ml of solution. Each of the solutions contained 500 mm NaCl. Each spin was performed at 4,000 xg. The concentrators with MWCO greater than 5 kda were spun for 30 minutes. Concentrators with 5 kda cut-offs were spun 45 minutes. After the first and the second spin the retentate was brought up to 6 ml with ultra pure water from the Arium purification system (Sartorius Stedim Biotech). OD readings were taken at 410 nm for the Cytochrome C and 280 nm for the BSA and IgG samples. Salt concentration was measured with a Qcond 2200 conductivity instrument. Spin-X UF 500 Concentrator Spin-X UF 6 Concentrator Introduction With appropriate device size and membrane cut-off selected, Spin-X UF concentrators will typically yield recoveries for the concentrated sample greater than 90% when the starting sample contains over 0.1 mg/ml protein of interest. Although dependent on the sample characteristics, solute (protein) adsorption on the polyethersulfone (PES) membrane surface is typically very low (2 to 10 µg/cm 2 ) and in practice is not detectable. Typically, a higher molecular cut-off (MWCO) membrane will bind more solute than a low MWCO membrane. Sample losses through adsorption can increase to 20 to 100 µg/cm 2 when a solute in the filtrate is of interest and it must pass through the whole internal structure of the ultrafiltration (UF) concentrator membrane. While the relative adsorption to the plastic surface of the sample container will be proportionately less than on the membrane (due to the membrane s higher total internal surface area), binding to the plastic surface of the upper and lower chambers can still be a source of yield loss. Whenever possible, choose the lowest MWCO membrane in the smallest concentrator applicable. Swinging bucket rotors are preferred to fixed angle rotors because they reduce the surface area of the concentrator that will be exposed to the solution during centrifugation. An important factor not to be neglected is the thorough recovery of the retentate. Make sure to carefully remove all traces of solution from the upper sample container and, if feasible, rinse the device after recovering the sample with one or more drops of buffer and then recover again. The intention of the following passivation procedure is to improve recovery of protein samples in the nanogram to microgram concentration range by pretreating (passivating) the concentrator to block or coat its membrane and plastic surfaces to reduce their stickiness for the solute molecules of interest. For this purpose a range of solutions are suggested in Table 1. Passivation procedure for Spin-X UF concentrators 1. Wash the concentrators once by filling with high purity water and spin the liquid through according to the respective protocol. 2. Remove residual water thoroughly by pipetting. Caution: Take care not to damage the membrane with the pipette tip. 3. Fill concentrators with the blocking solution of choice as given in Table 1. Table 1. Passivation Solutions Spin-X UF 20 Concentrator Type Powdered milk BSA Tween 20 SDS Triton X-10 PEG 3000 Concentration 1% in high purity water 1% in PBS 5% in high purity water 5% in high purity water 0.5% in high purity water 5% in high purity water 4. Incubate the filled concentrators at room temperature for at least 2 hours (overnight is also possible except for Triton X-100 which is not recommended for overnight incubation).

5. Pour out the blocking solution. 6. Rinse the device 3 to 4 times very thoroughly with high purity water and finally spin through. 7. The concentrators are now passivated or blocked and ready for use. We recommend evaluating different passivation reagents by comparing passivated concentrators side by side against untreated concentrators. Note: It is necessary to rinse the concentrator thoroughly before each wash/spin to ensure that traces of passivation compound are entirely removed from the bottom of the upper chamber. Use the concentrator immediately for protein concentration or store it at 4 C filled with high purity water, to prevent the membrane from drying. Corning Spin-X UFConcentratorOrderingInformation Cat. No. Description Capacity Membrane Pack Size 431477 Spin-X UF 500 500 µl 5,000 MWCO 25 431478 Spin-X UF 500 500 µl 10,000 MWCO 25 431479 Spin-X UF 500 500 µl 30,000 MWCO 25 431480 Spin-X UF 500 500 µl 50,000 MWCO 25 431481 Spin-X UF 500 500 µl 100,000 MWCO 25 431482 Spin-X UF 6 6 ml 5,000 MWCO 25 431483 Spin-X UF 6 6 ml 10,000 MWCO 25 431484 Spin-X UF 6 6 ml 30,000 MWCO 25 431485 Spin-X UF 6 6 ml 50,000 MWCO 25 431486 Spin-X UF 6 6 ml 100,000 MWCO 25 431487 Spin-X UF 20 20 ml 5,000 MWCO 12 431488 Spin-X UF 20 20 ml 10,000 MWCO 12 431489 Spin-X UF 20 20 ml 30,000 MWCO 12 431490 Spin-X UF 20 20 ml 50,000 MWCO 12 431491 Spin-X UF 20 20 ml 100,000 MWCO 12 For additional product or technical information, please e-mail us at CLStechserv@corning.com, visit www.corning.com/lifesciences, or call 1.800.492.1110. Outside the United States, please call 1.978.442.2200. Corning Incorporated Life Sciences Tower2,4thFloor 900 Chelmsford St. Lowell, MA 01851 t 800.492.1110 t 978.442.2200 f 978.442.2476 www.corning.com/lifesciences CorningandSpin-XareregisteredtrademarksofCorningIncorporated,Corning,NY. CorningIncorporated,OneRiverfrontPlaza,Corning,NY14831-0001 Worldwide Support Offices ASIA/PACIFIC Australia/New Zealand t 65 6733-6511 f 65 6735-2913 China t 86-21-5467-4666 f 86-21-5407-5899 India t 91-124-235 7850 f 91-124-401 0207 Japan t 81(0) 3-3586 1996/1997 f 81(0) 3-3586 1291/1292 Korea t 82 2-796-9500 f 82 2-796-9300 Singapore t 65 6733-6511 f 65 6861-2913 Taiwan t 886 2-2716-0338 f 886 2-2716-0339 EUROPE France t0800916882 f0800918636 Germany t08001011153 f08001012427 The Netherlands t31206557928 f31206597673 United Kingdom t08003768660 f08002791117 All Other European Countries t31(0)206596051 f31(0)206597673 LATIN AMERICA Brasil t(55-11) 3089-7419 f(55-11) 3167-0700 Mexico t(52-81) 8158-8400 f(52-81) 8313-8589 2009 Corning Incorporated Printed in U.S.A. 4/09 POD CLS-AN-120

Corning Spin-X UF 6and20mL Concentrators Technical Data and Operating Instructions Forinvitrouseonly.

Introduction Corning Spin-X UF concentrators are disposable, single use only ultrafiltration devices with polyethersulfone membranes (PES) for the concentration and/or purification of biological samples. Spin-X UF 6 concentrators are suitable for sample volumes of 2 to 6 ml and the Spin-X UF 20 concentrators can handle samples up to 20 ml. Both products feature twin vertical membranes for unparalleled speed. The innovative design, ease of use, speed and exceptional concentrate recoveries are the main features of the concentrators. Storage Conditions and Shelf Life Spin-X UF concentrators should be stored at room temperature. The devices should be used before the expiration date printed on the box. Chemical Compatibility Spin-X UF concentrators are designed for use with biological fluids and aqueous solutions. For chemical compatibility details, refer to Table 4 (page 6). Centrifugal Operation Spin-X UF concentrators can be used in swing bucket or fixed angle rotors accepting standard conical bottom tubes. In a single spin, solutions can be concentrated in excess of 100-fold. Samples are typically concentrated in 10 to 30 minutes with macromolecular recoveries in excess of 95%. The longitudinal membrane orientation and thin channel concentration chamber provide optimum cross flow conditions even for particle laden solutions; the centrifugal force pulling particles and solids away from the membrane to the bottom of the device. Macromolecules collect in an impermeable concentrate pocket integrally molded below the membrane surface, thereby eliminating the risk of filtration to dryness. Required Equipment 1. Centrifuge with swing bucket or fixed angle rotor (minimum 25 ). Device Spin-X UF 6 Spin-X UF 20 Carrier Required To fit 15 ml (17 mm diameter) conical bottom tubes To fit 50 ml (30 mm diameter) conical bottom tubes 2. Pasteur or standard pipets for sample addition. Pipettors with gel loading tips are recommended for sample removal from the concentrate pocket. Rotor Compatibility Please note: Spin-X UF 20 (30 mm x 116 mm) is designed to fit into rotors that can accommodate Corning 50 ml conical bottom tubes, e.g., Beckman Allegra 25R with TS-5.1-500 swing-out rotor with BUC 5 buckets and 368327 adaptors; Beckman TA-10.250 25 fixed angle rotor with 356966 adaptors; Heraeus Multifuge 3 S-R with (Heraeus/Sorvall) 75006445 swing out rotor with 75006441 buckets and adaptors for These devices are not designed to fit into rotors that only accept round bottom 29 mm x 10 5 mm tubes, e.g., Sorvall SS34 or Beckman JA 20. Operation 1. Select the most appropriate membrane cut-off for your sample. For maximum recovery select a molecular weight cut off (MWCO) at least 50% smaller than the molecular size of the species of interest. 2. Fill concentrator with up to maximum volumes shown in Table 1 (page 4). Ensure screw closure is fully seated. 3. Insert assembled concentrator into centrifuge (when fixed angle rotors are used, angle concentrator so that the printed window faces upwards/outwards). 4. Centrifuge at speeds recommended in Table 2 (page 5), taking care not to exceed the maximum g force indicated by membrane type and MWCO. 5. Once the desired concentration is achieved, (see Tables 3a and 3b, page 5) for guide to concentration times), remove assembly and recover sample from the bottom of the concentrate pocket with a pipet. Desalting/Buffer Exchange 1. Concentrate sample to desired level. 2. Empty filtrate container. 3. Refill concentrator with an appropriate solvent. 4. Concentrate the sample again and repeat the process until the concentration of the contaminating microsolute is sufficiently reduced. Typically, 3 wash cycles will remove 99% of initial salt content. Removing the Spin-X UF Body from the Filtrate Tube The sleeve (seen from the end) is oval in cross section (Figure 1). The tube is round in cross section to give a tight fit to the sleeve. To release the tube from the sleeve, you must pinch the tube to press it into an oval shape before removing it with a twisting action. Helpful Hints 1.FlowRate Filtration rate is affected by several parameters, including MWCO, porosity, sample concentration, viscosity, centrifugal force and temperature. Expect significantly longer spin times for starting solutions with over 5% solids. When operating at 4 C, flow rates are approxi- Corning 50 ml conical bottom tubes. mately 1.5 times slower than at 25 C. Viscous solutions Figure 1. such as 50% glycerin will take up to 5 times longer to concentrate than samples in a predominantly buffer solution. 2 3

2. Prerinsing Membranes fitted to Spin-X UF concentrators contain trace amounts of glycerin and sodium azide. Should these interfere with analysis they can be removed by rinsing fill volume of buffer solution or deionized water through the concentrator. Decant filtrate and concentrate before processing sample solution. If you do not want to use the prerinsed device immediately, store it in the refrigerator with buffer or water covering the membrane surface. Please do not allow the membrane to dry out. 3. Sterilization of Polyethersulfone Membranes Polyethersulfone membranes should not be autoclaved as high temperatures will substantially increase membrane MWCO. To sterilize, use a 70% ethanol solution or sterilizing gas mixture. Technical Specifications Table 1. Technical Properties Concentrator Capacity Spin-X UF 6 Spin-X UF 20 Swing bucket rotor 6 ml 20 ml Fixed angle rotor 6 ml 14 ml Dimensions Total length 122 mm 116 mm Width 17 mm 30 mm Active membrane area 2.5 cm 2 6.0 cm 2 Hold up volume of membrane <10 µl <20 µl Dead stop volume* 30 µl 50 µl Materials of Construction Body Polycarbonate Polycarbonate Filtrate vessel Polycarbonate Polycarbonate Concentrator cap Polypropylene Polypropylene Membrane Polyethersulfone Polyethersulfone *Dead stop volume as designed in molding tool. This volume may vary depending on sample, sample concentration, operation temperature and centrifuge rotor. Table 2. Maximum Recommended Spin Force(xg) Spin-X UF 6 Swing Bucket Fixed Angle 5,000 to 50,000 MWCO PES 4,000 xg 10,000 xg 100,000 MWCO PES 4,000 xg 6,000 xg Spin-X UF 20 5,000 to 50,000 MWCO PES 5,000 xg 8,000 xg 100,000 MWCO PES 3,000 xg 6,000 xg Table 3a. Performance Characteristics Spin-X UF 6 (Timetoconcentrateup to30x[min.]at 20 Candsoluterecovery%) Rotor Swing Bucket 25 Fixed Angle Start Volume 6 ml 6 ml BSA 1.0 mg/ml (66,000 MW) Min. Rec. Min. Rec. 5,000 MWCO PES 20 98% 12 98% 10,000 MWCO PES 13 98% 10 98% 30,000 MWCO PES 12 98% 9 97% IgG 0.25 mg/ml (160,000 MW) 30,000 MWCO PES 18 96% 15 95% 50,000 MWCO PES 17 96% 14 95% 100,000 MWCO PES 15 91% 12 91% Table 3b. Performance Characteristics Spin-X UF 20 (Timetoconcentrateup to30x[min.]at 20 Candsoluterecovery%) Rotor Swing Bucket 25 Fixed Angle Start Volume 20 ml 14 ml BSA 1.0 mg/ml (66,000 MW) Min. Rec. Min. Rec. 5,000 MWCO PES 23 99% 29 99% 10,000 MWCO PES 16 98% 17 98% 30,000 MWCO PES 13 98% 15 98% IgG 0.25 mg/ml (160,000 MW) 30,000 MWCO PES 27 97% 20 95% 50,000 MWCO PES 27 96% 22 95% 100,000 MWCO PES 25 91% 20 90% 4 5

Table 4. Chemical Compatibility (2 hour contact time, compatible ph range ph 1 9) Acetic Acid (25.0%) 1 Acetone (10.0%) 3 Acetonitrile (10.0%) 3 Ammonium Hydroxide (5.0%) 2 Ammonium Sulphate (saturated) 1 Benzene (100%) 3 n-butanol (70%) 1 Chloroform (1.0%) 3 Dimethyl Formamide (10.0%) 2 Dimethyl Sulfoxide (5.0%) 1 Ethanol (70.0%) 1 Ethyl Acetate (100%) 3 Formaldehyde (30%) 1 Formic Acid (5.0%) 1 Glycerine (70%) 1 Guanidine HCI (6M) 1 Hydrocarbons, aromatic 3 Hydrocarbons, chlorinated 3 Hydrochloric Acid (1M) 1 Imidazole (500 mm) 1 Isopropanol (70%) 1 Lactic Acid (5.0%) 1 Mercaptoethanol (10 mm) 1 Methanol (60%) 2 Nitric Acid (10.0%) 1 Phenol (1.0%) 2 Phosphate Buffer (1.0 M) 1 Polyethylene Glycol (10%) 1 Pyridine (100%) 2 Sodium Carbonate (20%) 2 Sodium Deoxycholate (5.0%) 1 Sodium Dodecylsulfate (0.1M) 1 Sodium Hydroxide 3 Sodium Hypochlorite (200 ppm) 2 Sodium Nitrate (1.0%) 1 Sulfamic Acid (5.0%) 1 Tetrahydrofuran (5.0%) 3 Toluene (1.0%) 3 Trifluoroacetic Acid (10%) 1 Tween 20 (0.1%) 1 Triton X-100 (0.1%) 1 Urea (8 M) 1 Corning Spin-X UFConcentratorsOrderingInformation Cat. No. Description Capacity Membrane Pack Size 431482 Spin-X UF 6 6 ml 5,000 MWCO 25 431483 Spin-X UF 6 6 ml 10,000 MWCO 25 431484 Spin-X UF 6 6 ml 30,000 MWCO 25 431485 Spin-X UF 6 6 ml 50,000 MWCO 25 431486 Spin-X UF 6 6 ml 100,000 MWCO 25 431487 Spin-X UF 20 20 ml 5,000 MWCO 12 431488 Spin-X UF 20 20 ml 10,000 MWCO 12 431489 Spin-X UF 20 20 ml 30,000 MWCO 12 431490 Spin-X UF 20 20 ml 50,000 MWCO 12 431491 Spin-X UF 20 20 ml 100,000 MWCO 12 * 1 = acceptable, 2 = questionable, testing advised, 3 = not recommended. 6 7

Corning Spin-X UF 500 µl Concentrators Technical Data and Operating Instructions Forinvitrouseonly.

Introduction Spin-X UF concentrators are disposable, single use only ultrafiltration devices with polyethersulfone membranes (PES) for the concentration and/or purification of biological samples. Spin-X UF 500 concentrators are suitable for sample volumes of 100 to 500 µl. Thevertical membrane design and thin channel filtration chamber minimizes membrane fouling and provides high speed concentrations, even with particle laden solutions. Storage Conditions and Shelf Life Spin-X UF concentrators should be stored at room temperature. The devices should be used before the expiration date printed on the box. Chemical Compatibility Spin-X UF concentrators are designed for use with biological fluids and aqueous solutions. For chemical compatibility details, refer to Table 3 (page 5). Centrifugal Operation Spin-X UF 500 concentrators can be used in a benchtop fixed angle rotor, accepting 2.2 ml centrifuge tubes. In a single spin, solutions can be concentrated approximately 100-fold. Required Equipment 1. Centrifuge with fixed angle rotor (minimum 40 ) that can fit 2.2 ml (11 mm diameter) conical bottom tubes. 2. Fixed or variable volume pipettors with gel loading tips are recommended for sample addition and removal. Operation CorningSpin-XUF500 100 to 500 µl capacity 1. Select the most appropriate membrane cut-off for your sample. For maximum recovery select a molecular weight cut off (MWCO) at least 50% smaller than the molecular size of the species of interest. 2. Fill concentrator with up to a maximum volume of 500 µl as shown in Table 1 (Ensure lid is fully seated). 3. Insert assembled concentrator into a centrifuge with a fixed angle rotor. Minimum rotor angle is 40º. Angle concentrator so that the printed window faces upwards/ outwards. 4. Centrifuge at speeds up to 15,000 xg. 5. Once the desired concentration is achieved, remove assembly and recover sample from the bottom of the concentrate pocket with a pipettor and gel loading tip. See Table 2 (page 4) for a guide to concentration times. The filtrate tube can be sealed for storage. Desalting/Buffer Exchange 1. Concentrate sample to desired level. 2. Empty filtrate container. 3. Refill concentrator with an appropriate solvent. 4. Concentrate the sample again and repeat the process until the concentration of the contaminating microsolute is sufficiently reduced. Typically, 3 wash cycles will remove 99% of initial salt content. Helpful Hints 1.FlowRate Filtration rate is affected by several parameters, including MWCO, porosity, sample concentration, viscosity, centrifugal force and temperature. Expect significantly longer spin times for starting solutions with over 5% solids. When operating at 4 C, flow rates are approximately 1.5 times slower than at 25 C. Viscous solutions such as 50% glycerin will take up to 5 times longer to concentrate than samples in a predominantly buffer solution. 2. Prerinsing Membranes fitted to Spin-X UF concentrators contain trace amounts of glycerin and sodium azide. Should these interfere with analysis they can be removed by rinsing fill volume of buffer solution or deionized water through the concentrator. Decant filtrate and concentrate before processing sample solution. If you do not want to use the prerinsed device immediately, store it in the refrigerator with buffer or water covering the membrane surface. Please do not allow the membrane to dry out. 3. Sterilization of Polyethersulfone Membranes Polyethersulfone membranes should not be autoclaved as high temperatures will substantially increase membrane MWCO. To sterilize, use a 70% ethanol solution or sterilizing gas mixture. 2 3

Technical Specifications Table 1. Technical Properties Concentrator Capacity Swing bucket rotor Spin-X UF 500 Do not use Fixed angle rotor 500 µl Minimum rotor angle 40º Dimensions Total length Width 50 mm 11 mm Active membrane area 0.5 cm 2 Hold up volume of membrane <5 µl Dead stop volume* 5 µl Materials of Construction Body Filtrate vessel Concentrator cap Membrane Maximum Spin Force (Fixed angle only) Polycarbonate Polypropylene Polycarbonate Polyethersulfone 5,000 to 50,000 MWCO PES 15,000 xg >100,000 MWCO PES 15,000 xg *Dead stop volume as designed in molding tool. This volume may vary depending on sample, sample concentration, operation temperature and centrifuge rotor. Table 2. Performance Characteristics Spin-X UF 500 (Timetoconcentrateup to30x[min.]at 20 Candsoluterecovery%) Rotor 40 Fixed Angle Start Volume 500 µl BSA 1.0 mg/ml (66,000 MW) Min. Rec. 5,000 MWCO PES 15 96% 10,000 MWCO PES 5 96% 30,000 MWCO PES 5 95% Table 3. Chemical Compatibility (2 hour contact time; compatible ph range ph 1 9) Acetic Acid (25.0%) 1 Acetone (10.0%) 3 Acetonitrile (10.0%) 3 Ammonium Hydroxide (5.0%) 2 Ammonium Sulphate (saturated) 1 Benzene (100%) 3 n-butanol (70%) 1 Chloroform (1.0%) 3 Dimethyl Formamide (10.0%) 2 Dimethyl Sulfoxide (5.0%) 1 Ethanol (70.0%) 1 Ethyl Acetate (100%) 3 Formaldehyde (30%) 1 Formic Acid (5.0%) 1 Glycerine (70%) 1 Guanidine HCI (6M) 1 Hydrocarbons, aromatic 3 Hydrocarbons, chlorinated 3 Hydrochloric Acid (1M) 1 Imidazole (500 mm) 1 Isopropanol (70%) 1 * 1 = acceptable, 2 = questionable, testing advised, 3 = not recommended. Lactic Acid (5.0%) 1 Mercaptoethanol (10 mm) 1 Methanol (60%) 2 Nitric Acid (10.0%) 1 Phenol (1.0%) 2 Phosphate Buffer (1.0 M) 1 Polyethylene Glycol (10%) 1 Pyridine (100%) 2 Sodium Carbonate (20%) 2 Sodium Deoxycholate (5.0%) 1 Sodium Dodecylsulfate (0.1M) 1 Sodium Hydroxide 3 Sodium Hypochlorite (200 ppm) 2 Sodium Nitrate (1.0%) 1 Sulfamic Acid (5.0%) 1 Tetrahydrofuran (5.0%) 3 Toluene (1.0%) 3 Trifluoroacetic Acid (10%) 1 Tween 20 (0.1%) 1 Triton X-100 (0.1%) 1 Urea (8 M) 1 IgG 0.25 mg/ml (160,000 MW) 30,000 MWCO PES 10 96% 50,000 MWCO PES 10 96% 100,000 MWCO PES 10 96% 4 5

Corning Spin-X UF500ConcentratorOrderingInformation Cat. No. Description Capacity Membrane Pack Size 431477 Spin-X UF 500 500 µl 5,000 MWCO 25 431478 Spin-X UF 500 500 µl 10,000 MWCO 25 431479 Spin-X UF 500 500 µl 30,000 MWCO 25 431480 Spin-X UF 500 500 µl 50,000 MWCO 25 431481 Spin-X UF 500 500 µl 100,000 MWCO 25 6