Phase Separations Customer Education Program

Similar documents
LC Technical Information

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM

Comparison Guide to C18 Reversed Phase HPLC Columns

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

LC and LC/MS Column Selection Flow Chart

COMPARISON GUIDE. to C18 Reversed Phase HPLC Columns. Comparison Data on 60 Commonly Used C18 Phases. Stationary Phase Specifications

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Hypersil BDS Columns TG 01-05

Packings for HPLC. Packings for HPLC

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

OBSERVATIONS ON THE WETTING OF REVERSED-PHASE HPLC PACKINGS

High Performance Liquid Chromatography

HPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

New Product Update. HPLC Columns

for Acclaim Mixed-Mode HILIC-1 Column

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

High Performance Liquid Chromatography

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Kromasil. for your analytical HPLC. Kromasil. The way to peak performance in liquid chromatography

UPLC Method Development and Validation

InertSustainSwift C8

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

[ CARE AND USE MANUAL ]

State-of-the-art C18 HPLC Columns

Liquid Chromatography

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Dilution(*) Chromatography

Validation of ACQUITY UPLC Methods

ACE Ultra Inert Base Deactivated HPLC Columns

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

DATES: LAB: Liquid Chromatography Separation of Grape Kool-Aid

Alltech Alltima HP Introduction

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Fast Separation of Vastly Different Compounds by Isocratic HPLC

Successfully Scaling and Transferring HPLC and UPLC Methods

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Chromatography. Gas Chromatography

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

zorbax Packed by Phenomenex Material Characteristics High efficiency Guaranteed quality Phenomenex columns packed with Zorbax

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

C18 Column. Care & Use Sheet

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

[ Care and Use Manual ]

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Solid Phase Extraction Method Development Tips and Tricks

InertSustainSwift C8

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Comparison of different aqueous mobile phase HPLC techniques

Reversed Phase Solvents

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Validated HPLC Methods

Inertsil Technical Library

Determination of Polymer Modifier in Asphalt

Bringing uhplc Performance to the Separation of Peptides

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Gemini-NX Performance paired with Axia Technology

HICHROM. Chromatography Columns and Supplies. LC COLUMNS Zorbax. Catalogue 9. Hichrom Limited

CHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS

Thermo Scientific Syncronis HPLC Columns. Remarkable separations guaranteed time after time

Chemistry Instrumental Analysis Lecture 28. Chem 4631

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

It s hot! ZIC -chilic

8. Methods in Developing Mobile Phase Condition for C18 Column

Shodex TM ODP2 HP series columns

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Routine MS Detection for USP Chromatographic Methods

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

ACE Excel SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

Automatization for development of HPLC methods

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

HPLC Background Chem 250 F 2008 Page 1 of 24

Introduction to Chromatography

Analysis - HPLC A.136. Primesep 5 µm columns B.136

SPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Ensure reproducible Polymeric SPE workflow for high-throughput bioanalysis applications

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Peptide Isolation Using the Prep 150 LC System

HPLC Determination of Antidepressants and their Metabolites in Plasma

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Basic principles of HPLC

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Sample Preparation. Approaches to Automation for SPE

penta-hilic UHPLC COLUMNS

Method Development Kits

Acclaim Mixed-Mode WCX-1

Introduction to Chromatographic Separations

Transcription:

Phase Separations Customer Education Program Advanced HPLC Column Selection Using the Hydrophobicity Chart neutral neutral 0 base Time (min) base 50 Copyright 1998 Phase Separations

Course Library * Strategies and Techniques for Faster HPLC Methods Development * Methods Development Optimization Software --HIPAC * Advanced HPLC Column Selection Using The Hydrophobicity Chart * Solid Phase Extraction Technology * Selecting Optimum HPLC Column Dimensions and Stationary Phase Particle Size * Troubleshooting Common HPLC Problems * Purchasing High Performance Chromatographic Supplies * Understanding RP-HPLC Separations -- Effect of Silica Type/Activity * Reversed-Phase HPLC Methods Development Copyright 1998 Phase Separations

Course List - Order Number & Duration * Strategies and Techniques for Faster HPLC PSL908200 2.5 hours Methods Development *Methods Development Optization Software -- HIPAC PSL908201 45min * Advanced HPLC Column Characterization & Selection PSL908203 1.5 hours Using The Hydrophobicity Chart * Solid Phase Extraction Technology PSL908204 4 hours * Selecting Optimum HPLC Column Dimensions and PSL908205 1 hour Stationary Phase Particle Size * Troubleshooting Common HPLC Problems PSL908206 2 hours * Purchasing High Performance Chromatographic PSL908207 7 hours Supplies * Understanding RP-HPLC Separations PSL908208 1.5 hours -- Effect of Silica Type/Activity *Reversed-Phase HPLC Methods Development PSL908202 7 hours Copyright 1998 Phase Separations

...because Chromatography is still a Science SM Strategies & Techniques for Faster HPLC Methods Development Course Abstract: This course is designed for those individuals developing new reversed-phase HPLC methods or those planning to improve existing methods in order to increase the methods' sample throughput, robustness and cost effectiveness. A logical and efficient methods development process is described which covers column selection based on analyte type and sensitivity requirements utilizing different selectivities of chromatographic packings to solve difficult application problems. We will explore solvent system optimization strategies, as well as review the role that Method Development Optimization Software can play in this process. Key examples are shown, and each attendee will receive a complete set of the slides in hard- copy form to keep for future reference. A Certificate of Achievement will be sent to all attendees, which can be used for your files as documentation of this training program. Duration: 2.5 hours Order Course # PSL908200 1998 Phase Separations

presents: Strategies & Techniques for Faster HPLC Methods Development

Packings for Reversed-Phase Chromatography silica-based vs. polymeric chain length/type silica quality endcapping packings with embedded polar function use a well-bonded, endcapped C8 based on a high-purity silica from a reputable manufacturer!

How to achieve shorter analysis times -- without sacrificing resolution and efficiency Reduce particle size Use shorter columns Maintain constant ratio of column length to particle size

Faster Analysis for Stability Studies Chlordiazepoxide (x) Degradation Products x 2 10.00 20.00 30.00 a. 5 µm 0.7 ml/min Columns: Symmetry C 18 a. 5 µm 3.9 x 150 mm; b., c. 3.5 µm 4.6 x 100 mm Mobile Phase: acetonitrile/methanol/ TETA-MeCOOH ph 7.0 Detection: UV at 240 nm Sample: 100 µl of 40 µg/ml x 2 b. 3.5 µm 1.0 ml/min x 2 10.00 20.00 30.00 c. 3.5 µm 1.4 ml/min Peak 2 Analysis USP Times Plates a. 40 min 11,240 b. 30 min 13,400 c. 20 min 12,500 10.00 20.00 Minutes D. J. Phillips

Performance of Different Diameter Columns at Equal Linear Velocities a) 0.00 b) 0.00 4 4 4 4.6 mm x 150 mm 10.00 3.9 mm x 150 mm 10.00 3.0 mm x 150 mm Conditions: Column: Symmetry C 18 5 µm Mobile Phase: water/methanol/glacial acetic acid 79:20:1 Flow Rates: a) 1.4 ml/min b) 1.0 ml/min c) 0.6 ml/min d) 0.29 ml/min Sample: mixture of 6 sulfa drugs, 10 to 39 µg/ml Injection vol.: a) 14 µl, b) 10 µl, c) 6 µl and d) 3 µl c) 0.00 10.00 4 2.1 mm x 150 mm d) 0.00 Minutes 10.00 USP Plates USP Tailing Peak 4 a) 4.6 mm ID 5780 1.1 b) 3.9 mm ID 5710 1.1 c) 3.0 mm ID 4350 1.2 d) 2.1 mm ID 4140 1.2 D. J. Phillips

Method Development Strategy Gradient at low ph Starting Solvent Gradient at high ph Starting Solvent Gradient with Second Slope Calculate Isocratic Mobile Phase Isocratic Run in Starting Solvent Calculate Isoeluotropic Mobile Phases Isocratic Run in Second Solvent Isocratic Run in Third Solvent Mixture Optimization

Method Validation Criteria Precision Accuracy Method Validation Limit of Detection Limit of Quantitation Specificity Linearity and Range Ruggedness Robustness

Method Ruggedness Analyst to Analyst Instrument to Instrument Lab to Lab Column to Column Batch to Batch

Method Robustness Content of Organic Solvent Ionic Strength ph Temperature Other Additives

...because Chromatography is still a Science SM Methods Development Optimization Software HIPAC Course Abstract: One of the most significant trends today is the need for rapid HPLC Methods Development efforts. Recently, a tool to aid the analytical chemist in reducing the time necessary developing new applications has become available. This tool utilizes the chromatographic experience of analytical chemists combined with desktop computers and software, based on chromatographic behavior models, to rapidly optimize application conditions. In addition, this tool can be used to assess the robustness of potential applications, without the need for many multiple trial runs. A review of the HIPAC brand software designed for the optimization of binary, isocratic reversed- phase applications, and system optimization is given. Detailed use of the software is shown, as well as examples of the thorough, tutorial information included in the software. This tutorial information, and the use of the built- in chromatographic optimization examples, can also be effectively used in the training of lab personnel new to HPLC technology. Key examples are shown, and each attendee will receive a complete set of the slides in hard- copy form to keep for future reference. A Certificate of Achievement will be sent to all attendees, which can be used for your files as documentation of this training program. Duration: 45 minutes Order Course # PSL908201 1998 Phase Separations

HPLC Methods Development Optimization Software HIPAC TM B and S Binary-Isocratic/ inary-isocratic/system ystem Optimization

HIPACTM Chromatography Optimization Software Software Package Designed To Combine A Chromatographers' Experience, Experimental Data And HPLC Theory To Speed Methods Development Process Chromatographers' Experience Experimental Fast HPLC Data Theory Processing (Software)

HIPACTM Chromatography Optimization Software Complete Chromatography Methods Development Software Package Available In Individual Modules Targeted For The Type Of Applications In Your Lab * HIPAC B Isocratic Reversed-Phase And Binary Normal-Phase * HIPAC S Optimize System And Column Parameters * HIPAC G Reversed-Phase Binary Gradients * HIPAC TQ Ternary And Quarternary Isocratic

HIPACTM Chromatography Optimization Software Quickly Determines The Optimum Conditions For Your Application Based On Your Inputs

HIPACTM Chromatography Optimization Software

HIPACTM Chromatography Optimization Software Isoeluotropic Mobile Phase Selection

HIPACTM Chromatography Optimization Software

HIPACTM Chromatography Optimization Software

...because Chromatography is still a Science SM Advanced HPLC Column Characterization & Selection Using the Hydrophobicity Chart Course Abstract: Today, analytical chemists are frequently faced with the very challenging assignment of developing new HPLC methods, which are simple, rugged and robust, all in a short period of time. Traditional methods development techniques, coupled with the improper selection of HPLC column, could result in the spending of much unnecessary time to achieve the goals of the development effort, thus delaying the introduction of new products, which can be very costly. Phase Separations has developed a new, reversed- phase HPLC Column Selection chart, which scientifically differentiates many brands of HPLC columns. The resulting HYDROPHOBICITY Chart provides the analytical chemist with a performance ranking for each column. This allows the chemist to quickly select the best column for his/her application. With proper use, this selection can result in much faster methods development success, due to use This course provides the chemist with a complete understanding of chromatographic particle technology and performance. A wide range of HPLC columns from different vendors were evaluated. The development of the performance test, and how the chart is constructed is explained in detail. Examples of using the chart to solve application problems are given. A complete booklet of the course presentation materials will be furnished for future reference. A Certificate of Achievement will be given to all attendees, which documents your participation in this training program. Duration: 1.5 hours Order Course # PSL908203 1998 Phase Separations

...because Chromatography is still a Science SM Reversed-Phase HPLC Methods Development Course Abstract: This is a full day course is designed to provide critical information important to those developing new methods, as well as those redeveloping existing methods to improve performance in the areas of sensitivity, robustness and sample throughput. Examples are given, and a complete set of handouts is provided for future reference. A Certificate of Achievement will be given to all attendees, which documents your participation in this training program. Key topics include: Impact of Particle Technology on Chromatographic Performance Resolution Suggested Methods Development Strategy Benefits of Optimizing Column Dimensions and Particle Size Optimization Software Troubleshooting Tips Duration: 7 hours Order Course # PSL908202 1998 Phase Separations

Advanced HPLC Column Characterization & Selection Using the Hydrophobicity Chart neutral................................... base neutral base 0 Time (min) 50

Silica Pore Structure * Porous Silica > 99% of Surface Area in the Pores 10nm 100 Å * Pore Size Surface Area SA ~ 3 SV PD SA - Surface Area, m 2 /g SV - Specific Pore Volume, ml/g PD - Pore Diameter, Å or nm Typical Surface Area 100 -- 300 m 2 /g

Silanol Types H H Vicinal O O Si Si O O O Geminal OH OH Si O O H Si O O Lone Si Si Si O O O O O

Bonded Phase: Trifunctional Synthesis Cl Si OH + Cl Si Cl c c c c c c c c (Hydrolysis) OH Si Si o o Si o Si o c c c c c c c c c c c c c c c c + HCl Si Si o o Si c c c c c c c c

Bonded Pore C18 CH3 Si - O - Si -CH3 CH3 Å 50 CH3 Si - O - Si -CH3 CH3 Si-CCCCCCCCCCCCCCCCCC Si - OH Endcapped Silanol Si-CCCCCCCCCCCCCCCCCC Si-CCCCCCCCCCCCCCCCCC Si - OH CH3 Si - O - Si -CH3 CH3 CH3 Si - O - Si -CH3 CH3 CH3 Si - O - Si -CH3 CH3 Si - OH Si - OH Si-CCCCCCCCCCCCCCCCCC Si-CCCCCCCCCCCCCCCCCC Si-CCCCCCCCCCCCCCCCCC Si-CCCCCCCCCCCCCCCCCC Si-CCCCCCCCCCCCCCCCCC Si - OH Si-CCCCCCCCCCCCCCCCCC = C18 25 Å Si - OH = Silanol Group

Ligand Density (Surface Coverage) * Better measure of material's characteristics χ = %C [ %C 100 SA 1 - MW - 1 ] 100 nc 12 = µmoles/m2 SA - Specific Surface Area %C - % Carbon Load MW - Molecular Weight of Ligand nc - # of Carbon Atoms in Ligand Ligand Density Primary Ligand Density Silanols Ligand Density Surface Area Hydrolytic Stability Ligand Density

ION EXCHANGE (-) ph 14 Cation (+)pka 10.6 ph % Ionized 12.6 Neutral (+) ph 7 Anion 8.6 ~100% 6.7 ~100% (-) pka 4.7 ph 1 2.7 Neutral

Terminology Surface Silanol Groups * Changes surface charge as ph 2 7 OH O + H + Si Si Behaves as a Cation Exchanger (ph 2) (ph 7)

Tailing vs. Buffer ph Tailing Factor 4 3 2 1 Conventional C 18 Modern C 18 Ideal 2 3 4 5 6 7 8 Buffer ph B. A. Alden, T. H. Walter

Effect of Mobile Phase ph on Selectivity 12 base 8 Ret. Time 4 0 acid neutral 3 4 5 6 Buffer ph 7 8 9 Base ph 3 ph 8 Acid Neutral Acid Neutral Base 3 4 8 2 4 8

Reversed-Phase Selectivity Chart 20 Waters Spherisorb ODS 1 Resolve C 18 (Log Scale) 10 Platinum C 18 Zorbax Rx C 8 µbondapak C 18 Zorbax SB C 8 Alltima C 8 YMC J'Sphere L 80 Lichrosorb Select B Waters Spherisorb C 8 Lichrospher Select B Waters Spherisorb ODS 2 Purospher RP 18 Nucleosil C 18 YMC J'Sphere M 80 Alltima C 18 Silanol Activity 2 3 4 5 6 7 8 9 YMC-Basic Prodigy C 8 Nova-Pak C 8 Hypersil BDS C 8 Kromasil C 8 Zorbax XDB C 8 Inertsil C 8 Symmetry C 8 Hypersil ODS Zorbax SB C 18 Nova-Pak C 18 Hypersil BDS C 18 Hypersil Hypurity Elite Zorbax Rx C 18 Zorbax XDB C 18 Inertsil ODS-3 Inertsil ODS-2 Kromasil C 18 Prodigy C 18 Symmetry C 18 YMC J'Sphere H80 1 1 SymmetryShield RP 8 2 3 4 5 6 7 8 9 Hydrophobicity (Log Scale)

...because Chromatography is still a Science SM Solid Phase Extraction Technology Course Abstract: Solid Phase Extraction (SPE) is a powerful tool for sample preparation and has been growing in popularity because of the significant benefits in reduced sample preparation cost and increased sample throughput it can bring to your laboratory. In addition, utilizing a chromatographic bed in a sample preparation protocol can solve the difficult problems of increasing method sensitivity, accuracy and precision for your analytical applications. This seminar has been specifically designed for the chromatographer and will include: 1. Introduction and theory 2. Strategies for use cleanup, fractionation and trace concentration 3. Chemistries and device configurations 4. Experimental techniques and demonstrations 5. Method development and application assistance resources 6. Method troubleshooting tips 7. Applications using the latest in SPE technology A complete booklet of the course presentation materials will be furnished for future reference. A Certificate of Achievement will be issued to all attendees. Duration: 4 hours Order Course # PSL908204 1998 Phase Separations

Solid Phase Extraction Technology A Powerful Tool in Sample Preparation

Solid Phase Extraction (SPE) Technology Goals of Sample Preparation * Remove Interferences * Make Matrix Compatible with Analysis * Concentrate Sample to Increase Sensitivity * Derivatization * Protect Analytical Column + Simpler + Improve Accuracy + More Convenient + More Reproducible + More Cost Effective + Improve Safety

Solid Phase Extraction (SPE) Technology Precipitation Chemical Technique Utilizing the Addition of a Solvent to the Sample which causes One, or Some of the Compounds to Fall Out of Solution as Particulates

Solid Phase Extraction (SPE) Technology Liquid-Liquid Extraction (LLE) Chemical Technique Where an Immiscible Solvent is Added to the Sample which then Separates into 2 Distinct Liquid Phases. Some Sample Analytes will go into the Bottom Phase (Aqueous), Some will Separate into the Top Phase, (Organic)

Solid Phase Extraction (SPE) Technology How Does Chromatography Work? Stationary Phase Mobile Phase Analyte A Analyte B

Solid Phase Extraction (SPE) Technology Housing/Body Filters/Frits Chromatographic Bed (Sorbent) Luer Tip

Solid Phase Extraction (SPE) Technology Vacuum Manifolds

Solid Phase Extraction (SPE) Technology SPE Strategies Elute the product of interest, retain interferences want k 0 for analyte want k large for interferences Elute interferences, retain product want k 0 for interferences want k large for analyte * Concentrate product of interest ^ want k large for analyte / load large sample volume ^ elute concentrated analyte ^ enhanced sensitivity

Solid Phase Extraction (SPE) Technology Solid Phase Extraction (SPE) The Chromatographic Bed In The Cartridge Can Perform Three Critical Functions: 1 2 3 1) Chemical Clean-up (Pure Red) 2) Pre-Concentration (Dark Purple) 3) Fractionation (Pure Blue & Pure Red)

Solid Phase Extraction (SPE) Technology Breakthrough Study * Series of Experiments -- Passing Increasing Volumes of Sample Matrix Thru SPE Device * Follow Elution Protocol * Determine Analytical Results (% Recovery) * Plot Results vs Sample Volume * Determine Maximum Volume for Each of the Analytes -- Find Maximum Volume for Method

Solid Phase Extraction (SPE) Technology % Recovery as a Function of k 120 100 % Recovery 80 60 40 20 k = 10 k = 30 0 5 10 20 30 40 50 Volume Thru SPE Device (ml)

Solid Phase Extraction (SPE) Technology Mass Balance -- Measuring Analyte Concentration in all Fractions -- Will Show Breakthrough During Loading Proper Load Sample Analyte Fully Retained During Loading OverLoad Sample Analyte Breaks Through During Loading Poor Conditioning Drying Out Sample Analyte Breaks Through During Loading Poor Recovery

Solid Phase Extraction (SPE) Technology Fraction 1 (Clear) Water Polar Analytes (Clear) Fraction 2 8% IPA Pure Red Fractionation Fraction 3 35% IPA Pure Blue Fraction 4 70% IPA Non-Polar Analytes (Clear)

Solid Phase Extraction (SPE) Technology Low Concentration Analyte (Light Purple) Initially Retained and Concentrated as Large Sample Volume is Processed Interferences Discarded Then Concentrated Analyte is Eluted by a Different Elution Solvent Trace Concentration

Methods Development Approach Determine Nature of Analytes, and Sample Matrix Similar to Existing Method in Lab? Yes No Try Conditions - Evaluate for Capacity/ Breakthru, Recovery Reproducibility, Robustness and Ruggedness Meets Goals? Yes Validate Method No Yes Review SPE Bibliography, and Literature References for Exact or Similar Applications Any? No Determine Method Goals, and Strategy Call SPE Vendor Chromatography Mode Develop Method Conditions

Solid Phase Extraction (SPE) Technology Some Causes of Poor Recovery * Poor/No Condtioning * Drying Out Before Loading * Poor Chromatographic Conditions - k too Low (Breahthrough) - k too Large (Still Adsorbed) * Basic Compounds Strongly Retained By Deprotonated Silica Silanols of Sorbent * Metals in Silica Based Sorbents Can Interact With Metal Chelator Analytes

Room Temperature Stability of Sample in Solutions Betamethasone Valerate (20 µg/ml) a t=0 stock 3 Day 5 Day b b % Nominal Concentration 100 80 60 40 20 0 Saline Acetonitrile Methanol Spiked Solutions b b 0 5 10 15 20 25 Minutes a: betamethasone valerate b: breakdown products t=5 days saline t=5 days MeOH t=5 days MeCN

Results: Tetracyclines Compound Concentration % Recovery % RSD Minocycline 2.5 µg/ml 94.8 1.4 Tetracycline 2.5 µg/ml 104 0.55 1 AU 0.020 0.016 0.012 0.008 0.004 0.000 2 3 Column: SymmetryShield RP8, 5 µm, 3.0 x 150 mm Mobile Phase: 0.1% TFA in Water:Acetonitrile: Methanol (91:7:2) Detection: UV at 270 nm Flow Rate: 0.9 ml/min. Injection Volume: 20 µl Sample Identification: Peak 1: Minocycline Peak 2: Tetracycline Peak 3: Demeclocycline (I.S.) 10.0 20.0 30.0 Minutes Cheng

...because Chromatography is still a Science SM Selecting Optimum HPLC Column Dimensions and Stationary Phase Particle Size Course Abstract: In order to speed the introduction of new products to market, analytical chemists are now often faced with two additional challenges; analyzing compounds at lower concentration levels, as well as analyzing more samples per unit time. Developing new HPLC methods with high sensitivity and high sample throughput is becoming a significant priority. Phase Separations has created a new course that covers the key elements of optimizing HPLC column dimensions, and stationary phase particle size to meet these methods development goals. A discussion of the impact of the ratio of column length to particle size is given. This critical ratio determines the optimum performance of the column in the method, as it relates to efficiency, analysis time and back pressure. Examples are given, and a complete set of handouts is provided for future reference. A Certificate of Achievement will be given to all attendees, which documents your participation in this training program. Duration: 1 hour Order Course # PSL908205 1998 Phase Separations

Selecting Optimum HPLC Column Dimensions and Stationary Phase Particle Size

Benefits of Optimizing Column Dimensions Reduction in analysis time with decreasing column length Increase in mass sensitivity with decreasing column diameter Increase in resolution with increasing ratio of column length to particle size Increase in loadability with increasing column diameter

How to Achieve Shorter Analysis Times -- Without Sacrificing Resolution and Efficiency Reduce particle size Use shorter columns Maintain constant ratio of column length to particle size

Ratio of Column Length to Particle Size Determines 3 Key Performance Measures: * Maximum Column Efficiency * Shortest Analysis Time for Given Pressure * Pressure Drop at Given Analysis Time

Selection of Proper L/dp Ratio Type of Analysis L/dp Difficult > 50,000 Normal ~ 30,000 Fast < 15,000 Column 300mm -- 5µm 150mm -- 5µm 50mm -- 3µm

Constant Ratio of Column Length to Particle Size (30,000) 120 Resolution 100 80 60 40 20 10 µm x 30 cm 5 µm x 15 cm 3 µm x 9 cm 1 µm x 3 cm 0 1 10 100 1000 Analysis Time [min]

Performance Characteristics of a 5 µm and a 3.5 µm Symmetry Column 14000 100 Analysis Time [min] 10 12000 10000 Plate-Count 8000 6000 4000 5 µm, 3.9 mm x 150 mm 3.5 µm, 4.6 mm x 100 mm 2000 L/dp~30,000 B. A. Alden, U. D. Neue 0 0.1 1 Flow-Rate [ml/min]

Enhancing Sensitivity Sensitivity can be enhanced by: increasing efficiency decreasing asymmetry reducing column diameter using shorter columns decreasing detector noise using more sensitive detection modes decreasing capacity factor

Solvent Consumption -- Column Length Flow rate: 1 ml/min Column Length Elution Time Solvent % Savings 15 cm 6.5 min 6.5 ml -- 10 cm 4.5 min 4.5 ml 31% 5 cm 2.5 min 2.5 ml 62%

Solvent Consumption -- Column Diameter Linear Velocity: 0.118 cm/sec Column Length: 15 cm Elution Time: 8 min Column Diameter Flow Rate Solvent % Savings 4.6 mm 1.4 ml/min 11.2 ml -- 3.9 mm 1.0 ml/min 8.0mL 28% 3.0 mm 0.6 ml/min 4.8 ml 57% 2.1 mm 0.3 ml/min 2.4 ml 79%

Instrument Optimization Needed for Narrow Bore and Microbore Columns Instrument Bandspread: 25µl Areas to optimize: * Detector Flow Cell * Injector Sample Loop * 0.005" Tubing * Perfect Connections * Detector Time Constant < 0.2 (normally > 80µl)

Performance Monitoring To perform a measurement: - disconnect column from system - connect injector directly to detector Parameter Flow Rate Chart Speed Detector Sensitivity Time Constant Setting 1.0 ml/min 20 cm/min 0.5-1.0 AUFS 0.2 seconds or less dilute test mixture 1 to 10 in mobile phase inject 2 to 5 µl of this solution

Performance Monitoring Using 5 sigma efficiency method, measure the peak width at 4.4% of peak height Convert to microliters using the following equation: ( ) 2cm ( )( ) ( ) 1min 1 ml 1000µL = PW 20 cm min. ml 100 (µl) where: 1min/20cm = chart speed 1 ml/min = flow rate 1000 µl/ml = volume correction factor Typical LC System should be 100µL +/- 30µL Microbore System should be no greater than 20µL

...because Chromatography is still a Science SM Troubleshooting Common HPLC Problems Course Abstract: This course covers all commonly encountered HPLC problems and is designed for both the users and the developers of HPLC methods. It will assist the chromatographer in spotting potential problems during the operation of their HPLC and in eliminating instrument downtime. Covered subjects include: Key examples are shown, and each attendee will receive a complete set of the slides in hardcopy form to keep for future reference. A Certificate of Achievement will be given to all attendees, which documents your participation in this training program. problem prevention (column installation, equilibration, protection, storage) performance monitoring (instrument band spreading performance, column efficiency performance) Duration: 2 hours Order Course # PSL908206 examples and troubleshooting of common problems 1998 Phase Separations

SM Troubleshooting Common HPLC Problems

Course Outline * Column Maintenance Information * Problem Sources * Column Installation and Equilibration * Column Use * Performance Monitoring * Column Protection * Column Storage * Troubleshooting of Column Problems - Peak Shape Problems - Retention Time Problems ^ Hydrophobic Collapse - Miscellaneous Problems

Troubleshooting Problem CHEMISTRY COLUMN/GUARD SOLVENT SAMPLE HARDWARE PUMP INJECTOR DETECTOR INTEGRATOR Always do the easiest thing first: 1. Stop flow 2. Remove the column

Installation and Equilibration Waters Spherisorb 0.090 Parker Style other Waters Columns 0.130 Waters

Extra-Column Band Spreading Column Connection

Extra-Column Band Spreading.009".040".020" note the differences of the inner diameter of this tubing

Performance Monitoring Effect of Connecting Tubing on System Bandspreading.009".020".040" sample band dispersion inside tubing

Extra-Column Band Spreading The Observed Bandwidth (TOT) * Sum of the Bandspreading Contributions - Column (COL) - Extra-Column (EC) Instrument components TOT COL EC σ 2 = σ 2 + σ 2

Performance Monitoring To perform a Band Spread measurement: - disconnect column from system - connect injector directly to detector Parameter Flow Rate Chart Speed Detector Sensitivity Time Constant Setting 1.0 ml/min 20 cm/min 0.5-1.0 AUFS 0.2 seconds or less dilute test mixture 1 to 10 in mobile phase inject 2 to 5 µl of this solution

Performance Monitoring Using 5 sigma efficiency method, measure the peak width at 4.4% of peak height Convert to microliters using the following equation: ( )( 1min )( 1 ml ) ( 1000µL ) = 2cm PW 20 cm min. ml 100 (µl) where: 1min/20cm = chart speed 1 ml/min = flow rate 1000 µl/ml = volume correction factor Typical LC System should be 100µL +/- 30µL Microbore System should be no greater than 20µL

Column Protection Extension of column lifetime with Guard Column using a mixture of sulfa drugs as the sample A. Initial injection on Symmetry C 8 Sentry guard column B. After 550 injections on same Sentry guard column C. New Sentry Guard column for injection 551 on analytical column

Column Collapse voids - high back pressure, distorted and/or double peaks

Extra Column Effects Isocratic LC - Time Constant Differences left is 0.1 secs right is 10 secs note the noisy baseline on left chromatogram

Hydrophobic Collapse Chromatographers have observed complete loss of retention when working with low organic mobile phases. Vo? 40 min 40 min When we investigated this phenomenon we found that retention times were stable for over 20 hrs (77 injections) using 100% aqueous mobile phase. However, when flow was stopped, then restarted, retention was lost.. This observation suggested that the mobile phase is extruded from the pores when pressure is released from the column.

Retention Time Reproducibility Solvent Composition - Hydrophobic Collapse Temperature ph Ion Pairing Ionic Strength Extraneous Peaks Gradient Control

ph - Control

...because Chromatography is still a Science SM Purchasing High Performance Chromatographic Supplies Course Abstract: This course is designed specifically for Purchasing Professionals to enhance their understanding of the technology of Chromatography as it relates to improving their ability to effectively purchase these supplies while meeting corporate goals, such as improved customer satisfaction, and reduced purchasing costs through vendor consolidation. Key topics include: Chromatography technology / terminology (with a focus on HPLC) [includes live demonstration] End user performance needs Regulatory considerations Manufacturer / Supplier considerations Key purchasing specifications This full day course will provide you with the technical information you will need to intelligently discuss product specifications and performance, as well as outline critical issues in the development of an improved purchasing program for your chromatography supplies. Specific examples will be given showing how to achieve significant benefits by organizing your chromatography purchasing needs in product/ performance/ vendor priority. A complete booklet of the course presentation materials will be furnished for future reference. A Certificate of Achievement will be given to all attendees, which documents your participation in this training program. Duration: 7 hours Order Course # PSL908207 1998 Phase Separations

Purchasing High Performance Chromatographic Supplies A Course For Purchasing Professionals

Purchasing High Performance Chromatographic Supplies Topics * Chromatography Technology Backgrounder * Terminology * Hands-On Chromatographic Experiments * HPLC Supplies * GC Supplies * End-User Performance Needs * Manufacturer / Supplier Considerations * Regulatory Considerations * Key Purchasing Specifications

Purchasing High Performance Chromatographic Supplies Chromatography Technology Backgrounder Solvent HPLC Column Data Injector Pump Detector Waste

Purchasing High Performance Chromatographic Supplies Chromatography Terminology * Sample - The Original Representative Material Which Is To Be Analyzed -- Also Called The Sample Matrix * Analyte(s) - A Specific Compound(s) Contained In The Sample Which Is(Are) To Be Separated And Analyzed * Compound - Pure Chemical Component In A Sample, Also Called An Analyte Or Solute

Purchasing High Performance Chromatographic Supplies Chromatography Terminology * How Does A Chromatographic Column Work? Sample Band Analyte Bands

Purchasing High Performance Chromatographic Supplies Chromatography Terminology * How Does A Chromatographic Column Work? WHY Do They Separate? Yellow Likes The Mobile Phase And Goes As Fast As It Does - And Comes Out First Red Likes The Stationary Phase Somewhat, And Slows Down Blue Likes The Stationary Phase Best, And Slows Down The Most

Purchasing High Performance Chromatographic Supplies Chromatography Terminology * How Do You Get Peaks? Detector Sample Band Sample Peak Detector Cell

Purchasing High Performance Chromatographic Supplies Chromatography Terminology How Do You Identify Which Compounds Are Which When You Have A Series Of Peaks On You Chromatogram??? I Know From The Peaks' Retention Time Determined By Pure Standards Time

Purchasing High Performance Chromatographic Supplies Chromatography Terminology How Do You Know How Much Is Present In The Sample??? I Know From The Peaks' Area And Height Time

Purchasing High Performance Chromatographic Supplies Chromatography Terminology Look What Happens When It Rains Right After I WAX My Car! [Non-Polar Surface] Water Beads Up Polar Water DISLIKES Non-Polar Surface

Purchasing High Performance Chromatographic Supplies Chromatography Terminology * Solid Phase Extraction (SPE) Chemical Technique Using Column Chromatography In A Small, Single Use Disposable Cartridge Format. Can Perform Chemical Separation, Concentration And Fractionation Used With Vacuum Stations, Syringes And Automated Sample Processors

Purchasing High Performance Chromatographic Supplies Chromatography Terminology Big Ones Calibration Curve Molecular Weight Distribution "Finger Print" Mol. Wt. Retention Time

Purchasing High Performance Chromatography Supplies End User Performance Needs Stages Of A "Method" The Success Of The Method Is Critical: - Reputation Of The Researcher - Reduce Cost And Time To Market For A New Product If The Method Is Not Reproducible (Has To Be Redeveloped And Revalidated -- Can Be Incredibly Costly)

Purchasing High Performance Chromatography Supplies Manufacturer/Supplier Considerations Primary Manufacturer Column RePacker Specialized Distributor YOU 3rd Party Supplies Distributor Secondary Manufacturer/ Column Packer Column Packer Sourcing HPLC Columns

Purchasing High Performance Chromatography Supplies Manufacturer/Supplier Considerations * Manufacturers -- Some Names: Primary: Secondary: Alltech Bio-Rad GL Science Hamilton Hypersil Macherey-Nagel E. Merck Akso Nobel Polymer Labs Rockland HP Showa Denko Separations Group Toso Haas Waters Whatman YMC BrownLee Jones Keystone Phenomenex Supelco Packer: Bischoff Beckman Chrompack HPLC Technology HP Isolation Tech. Metachem Micra Resolution Systems Capital Hichrom Mac-Mod PE + Many Others

Purchasing High Performance Chromatography Supplies Manufacturer/Supplier Considerations Secondary Manufacturer/Packer Distributor Primary Packer/Technical Distributor End User Distributor Packer/ Repacker End User

Purchasing High Performance Chromatography Supplies Key Purchasing Considerations Brands Supplied By Different Vendors Brand # Vendors # Columns Hypersil 3 190 Inertsil 5 40 Kromasil 5 75 LiChrospher 3 8 Nucleosil 3 240 Waters Spherisorb 8 80

...because Chromatography is still a Science SM Understanding Reversed-Phase HPLC Separations Effect of Silica Type/Activity Course Abstract: This course is designed for those individuals developing new reversed-phase HPLC methods. Key topics include: Resolution Theory Controlling Chromatographic Parameters Silica Particle Technology Examples are given, and a complete set of handouts is provided for future reference. A Certificate of Achievement will be given to all attendees, which documents your participation in this training program. Duration: 1.5 hours Order Course # PSL908208 Bonding Process Packing Material Comparison Reproducibility Concerns 1998 Phase Separations

HPLC Method Development & Troubleshooting Understanding Reversed-Phase Separations 'The Effect of the Silica Chemistry' Phase Separations Customer Education Programme

Review of Chromatographic Terminology RETENTION, k' = (V 1 -V 0 )/V 0 SELECTIVITY, α = k' 2 /k' 1 PLATE COUNT, N = 16(V/W) 2

Resolution Resolution is a numerical measure of the separation of two compounds and is a function of N, k' and α. This equation is only valid for isocratic separations. Selectivity, retention and peak widths may be further enhanced by using gradients. R s =( N ). (α-1). k' 2 4 α (k' 2 +1)

k', N, α How They Control Resolution