Vela User s Guide
The Vela User s Guide is not intended as a replacement for the Operator s Manual. You must become completely familiar with the Vela Operator s Manual before using the Vela ventilator. Table of Contents Vela Ventilator Front Panel...1 Vela Ventilator Rear Panel...3 Circuit Assembly...5 Touch-Turn-Touch /Touch-Turn-Accept...9 Extended Function Testing...11 Fi02 Monitor Calibration...13 New Patient Set-up...15 Setting the Ventilation Breath Type and Mode...17 Monitored Parameters...19 Graphics...23 Synchronized Nebulizer...29 Oxygen Computer Chart...31 Contact Information...33 Notes...34
This page intentionally left blank.
Vela Ventilator Front Panel EFFORT INDICATOR MODE INDICATOR SCREEN INDICATOR ALARM STATUS INDICATOR MONITORED PARAMETERS PRIMARY CONTROLS MESSAGE WINDOW PRESSURE A/C 391 ml Vti 380 ml Vte 28 cmh2o Ppeak 5 cmh2o PEEP 21 % FiO2 20 BPM Rate 40 20 0-20 80 40 0-40 -80 250 500 750 0-250 22 cmh2o Insp Press Paw (cmh O) 2. V (lpm) V t (ml) 1.0 Sec Insp Time MAIN 2 4 6 8 10 12 2 4 6 8 10 12 2 4 6 8 10 12 5 cmh2o LPM % PEEP Flow trig %O2 Man Brth Silence ADV LIMITS SETUP PRINT Lk Comp Vol Min SETTINGS 5 21 ALARM ALARM SILENCE RESET INSP FREEZE HOLD MANUAL EXP BREATH HOLD 100% NEBULIZER O2 PANEL LOCK ACCEPT POWER INDICATORS DC AC DC STATUS CANCEL NEBULIZER OUTPUT FROM PATIENT TO PATIENT DATA DIAL PRESSURE RELIEF EXPIRATORY FLOW SENSOR 1 Vela User s Guide L1536 Rev. C Vela Front Panel
This page intentionally left blank. 2 Vela User s Guide L1536 Rev. C
Vela Ventilator Rear Panel A - Power Switch B - Fan and Fan Filter C - Printer Port D- VGA Port E - Non-Operational Port A B C D E F G F - High Pressure Oxygen Fitting G - Low Pressure Oxygen Fitting H H - Nurse Call System Connection I - Alarm Speaker I J - Power Cord K - Fuses L - Ground M - Alarm Volume Adjustment N - Fiber Optic Output Port K J M L N 3 Vela User s Guide L1536 Rev. C Vela Rear Panel
This page intentionally left blank. 4 Vela User s Guide L1536 Rev. C
Circuit Assembly 1. Install the diaphragm on the exhalation valve. Carefully seat the rim of the diaphragm on the exhalation valve and gently press around the rim to ensure it is seated evenly as shown. 2. Line up the fins on the exhalation valve body with the exhalation valve housing. 5 Vela User s Guide L1536 Rev. C Circuit Assembly
Circuit Assembly 3. Rotate clockwise until it clicks into place. 4. Attach the flow sensor with the flow tubes in an upright position. 6 Vela User s Guide L1536 Rev. C
Circuit Assembly 5. To connect the smart connector for the variable orifice flow sensor, pull back the locking sleeve. Push into the receptacle and slide the locking sleeve forward to secure. To remove, repeat these steps in reverse order. 7 Vela User s Guide L1536 Rev. C
Circuit Assembly FROM PATIENT TO PATIENT PRESSURE RELIEF NEBULIZER OUTPUT EXPIRATORY FLOW SENSOR EXPIRATORY LIMB OF PATIENT CIRCUIT INSPIRATORY LIMB OF PATIENT CIRCUIT Complete Patient Circuit Assembly 8 Vela User s Guide L1536 Rev. C
Touch-Turn-Touch / Touch-Turn-Accept Changes to controls and displays are accomplished with a three-step technique. 1. Touch the control or display to be changed. 2. Turn the data dial to reach the desired setting. 15 BPM Rate 500 ml Volume 3. Touch the control again OR press "ACCEPT" to confirm the change. Changes not accepted by either method within 15 seconds revert to previous settings. Press "CANCEL" prior to accepting proposed changes to return to previous settings. 9 Vela User s Guide L1536 Rev. C Touch-Turn-Touch / Touch-Turn-Accept
This page intentionally left blank. 10 Vela User s Guide L1536 Rev. C
Extended Functions Testing The extended functions menu allows access to stored data, performance calibrations and customization of the front panel. Touch the screen indicator to access the extended functions menu. When the screen select menu appears, press the Extended Functions button. Extended Functions Menu 11 Vela User s Guide L1536 Rev. C Extended Functions Testing
Extended Functions Testing Events Stores data events for Service evaluation & troubleshooting. * Transducer Data Allows setting of transducer analog outputs for Service. * Transducer Test Allows Service testing of transducer function. * Version Info Displays software version information and turbine serial number. Date/Time Displays total hours of ventilator operation and date/time configuration. Fi02 Monitor Calibration Allows calibration of internal oxygen sensor. (Comprehensive & Standard models only). Vent Set-up: Allows setting of these functions: Low Min Vol OFF Enable or disable an "OFF" setting Panel Lock Enable or disable the front panel lock switch. Fi02 Monitor Turns the Fi02 monitor on or off. (Comprehensive & Standard models only). Altitude units of measure Toggles between feet and meters for the altitude setting. Altitude setting Allows setting of altitude for accurate volume measurement. Language buttons Select the desired language for the front panel. Video Normal/Inverse Reverses the color configuration of the graphic interface. For a complete explanation of each function, refer to the Vela Service Manual (P/N L1534). * Denotes function for use by trained Service technician. 12 Vela User s Guide L1536 Rev. C
Fi02 Monitor Calibration When the Fi02 Mon Calibration button is touched, the calibration menu appears. You have a choice of doing an ambient air calibration or a 100% oxygen calibration. Either can be done on or off a patient; however, be certain that the patient can tolerate room air or 100% oxygen for the calibration procedure, which takes approximately 4 minutes to complete. To exit the Fi02 Calibration screen, touch the EXIT button. 13 Vela User s Guide L1536 Rev. C Fi02 Monitor Calibration
This page intentionally left blank. 14 Vela User s Guide L1536 Rev. C
New Patient Set-Up 1. Power on the ventilator and perform operational verification testing as described in Chapter 2 of your Operator's Manual. 2. Select "NEW PATIENT". Touch "PATIENT ACCEPT" Selecting "NEW PATIENT" clears all trends and saved loops, and sets all controls to default values. Selecting "RESUME CURRENT" resumes ventilation at previous control settings. PATIENT SELECT RESUME CURRENT NEW PATIENT PATIENT ACCEPT 3. Select Active or Passive Humidification. 4. Enter patient ID if desired. Touch the Patient ID box to open a virtual keyboard and use Touch-Turn- Touch TM /Touch-Turn-Accept TM to enter alphanumeric characters. 5. Select LEAK COMP ON/OFF. Note: Default for NPPV A/C, NPPV SIMV; NPPV CPAP PSV; LEAK COMP ON 6. Touch "SETUP ACCEPT" The Set-up screen can be accessed at any time by pressing the "Setup" button on the bottom right corner of the touch screen. LEAK COMP ON PATIENT IDENTIFICATION HUMIDIFIER ACTIVE ON SETUP ACCEPT 15 Vela User s Guide L1536 Rev. C New Patient Set-up
This page intentionally left blank. 16 Vela User s Guide L1536 Rev. C
Setting the Ventilation Breath Type and Mode 1. Touch the Mode Indicator to open the mode window. 2. Select the desired mode and set the primary controls. 3. Touch "MODE ACCEPT" to accept the new mode and control settings. 4. Advanced settings and alarm limits may also be adjusted at this time. The buttons that open the advanced setting and alarm limits windows are located in the lower right hand corner of the touch screen. MODE SELECT Volume A/C Pressure A/C PRVC A/C NPPV A/C Volume SIMV Pressure SIMV PRVC SIMV NPPV SIMV NPPV CPAP PSV APNEA MODE Volume Pressure APNEA SETTINGS APRV BiPhasic CPAP PSV MODE ACCEPT 17 Vela User s Guide L1536 Rev. C Setting the Ventilation Breath Type and Mode
This page intentionally left blank. 18 Vela User s Guide L1536 Rev. C
Monitored Parameters Set Main Screen Monitored Parameters by using the Touch-Turn-Touch TM /Touch-Turn-Accept TM technique. Main Screen Monitors Five monitored parameters are continuously displayed to the left of the graphics display. Selections are made from a scrolling menu. Monitored parameters appearing on the main screen may be different from those that appear on the optional loops screen. 19 Vela User s Guide L1536 Rev. C Monitored Parameters
Monitored Parameters To configure an extended display for easy charting, 1. Touch the Screen Indicator in the top center of the Main Screen display. 2. The Screen Select box will display. 3. Select MONITOR from the selection box that appears. SCREEN SELECT MAIN LOOP MONITOR TRENDS MANEUVER EXTENDED FUNCTIONS 20 Vela User s Guide L1536 Rev. C
Monitored Parameters 18.5 0 26.7 0.0 0.0 ml Vte ml Vti ml Spon Vt ml Mand Vt L Ve 0.0 0.0 0.0 0 0.00 L Spon Ve bmp Rate bmp Spon Rate sec Ti sec Te 0 0.00 0 8 0 cmh20 Ppeak cmh20 Pmean cmh20 PEEP PSIG 02 Regulated % Fi02 A 15-window display is configurable from a scrolling menu. Use the Touch-Turn-Touch TM /Touch-Turn-Accept TM technique to configure this display. 21 Vela User s Guide L1536 Rev. C
This page intentionally left blank. 22 Vela User s Guide L1536 Rev. C
Graphics Configure main screen graphics for scale and sweep speed. 1. Touch the vertical or horizontal axis to highlight. 2. Turn the data dial to adjust scale or sweep speed. 3. Touch the axis again or press accept to confirm the change. 23 Vela User s Guide L1536 Rev. C Graphics
Graphics Configure the optional Loops Screen 250 500 750 1. Touch the vertical or horizontal axis to highlight. 2. Turn the data dial to adjust the scale. 3. Touch the axis again or press accept to confirm the change. 24 Vela User s Guide L1536 Rev. C
Graphics The FREEZE button freezes the current screen until pressed again. When the screen is frozen you can scroll through displayed waveforms or loops using the Data Dial to move the cursor. A flow/volume loop in "freeze" mode is depicted below. As the dotted line cursor traces the "frozen" loop curve, flags display the values along the curve of the loop. Dashed Cursor Line 270 0.0 Flag showing X and Y values at various points along the loop tracing 250 500 750 Flow/Volume Loop Tracing Cursor, currently overlays the "X" axis at Zero 25 Vela User s Guide L1536 Rev. C
Graphics Saving loops 1. Press the "Freeze" button to freeze the graphics display. 2. Press "Save Loop" 270 0.0 250 500 750 3. Saved loops will appear with a time reference. A total of 4 loops can be saved at any time. The oldest loop is discarded when the 5th loop is saved. 26 Vela User s Guide L1536 Rev. C
Graphics Create Reference Loops 1. Press the "Freeze" button to freeze the graphics display. 3. Press the "Ref Loop Off" button to toggle the reference loop on. 4. Press the "Freeze" button to return to live loops display over the reference loop. 2. Touch the saved loop you want as a reference. Discontinue Reference Loops To discontinue the reference, press the "Freeze" button to freeze the graphics display and press the "Ref Loop On" button to turn off the reference loop. Press "Freeze" again to return to a live loop screen. 27 Vela User s Guide L1536 Rev. C
This page intentionally left blank. 28 Vela User s Guide L1536 Rev. C
Synchronized Nebulizer A high pressure oxygen source gas connection is required to power the in-line nebulizer. Nebulizer flow delivery is synchronized with inspiration. Nebulization lasts for up to 60 minutes and is user adjustable in increments of one minute. Pressing the nebulizer button again ends nebulization. Caution: Nebulizer output is not volume compensated during volume control breaths. Delivered volumes and peak inspiratory pressures will increase during nebulization. To ensure consistent tidal volume delivery, tidal volume should be reduced for the duration of the treatment. Nebulizer output is approximately 50 ml per half second of inspiratory time. Tidal volume is unaffected during pressure control breaths. Fi02 will increase slightly during nebulization. Delivered Fi02 during nebulization can be estimated using the following calculation; 2 2 ALARM SILENCE FREEZE MANUAL BREATH ALARM RESET INSP HOLD EXP HOLD New Fi02 = [(Vt x Fi02) + (0.1 x Ti)] [Vt + (0.1 x Ti)] NEBULIZER 100% O2 29 Vela User s Guide L1536 Rev. C Synchronized Nebulizer
This page intentionally left blank. 30 Vela User s Guide L1536 Rev. C
Oxygen Input flow (L/min) Oxygen Computer Chart 30 To Find O2 Input Flow: 1. Select desired FiO2 on horizontal axis. 2. Project up to current minute volume (Ve from monitor display). 3. Project horizontally to left vertical axis and read oxygen flow. To Find O2 Concentration: 1. Select current O2 input flow on vertical axis. 2. Project horizontally right to current minute volume (Ve from monitor display). 25 20 15 10 Ve = 20 10 6 4 3 2 1 18 16 14 12 8 3. Project vertically down to horizontal axis and read oxygen concentration. 5 0.21.30.40.50.60.70.80.90 1.0 FiO2 31 Vela User s Guide L1536 Rev. C Oxygen Computer Chart
This page intentionally left blank. 32 Vela User s Guide L1536 Rev. C
Contact Information VIASYS Healthcare Customer Care-Technical Service Hours: 7:30 AM to 4:30 PM (PST) M-F Phone: (760) 778-7200 Fax: (760) 778-7377 VIASYS Healthcare Customer Care Helpline Hours: 24 Hours, 7 days a week Phone: (800) 934-2473 Fax: (760) 778-7377 VIASYS Healthcare- Critical Care Division 1100 Bird Center Drive Palm Springs, CA 92262-8099 U.S.A. Phone: (760) 778-7200, (800) 328-4139 Fax: (760) 778-7274 www.viasyscriticalcare.com 33 Vela User s Guide L1536 Rev. C Contact Information
Notes 34 Vela User s Guide L1536 Rev. C