DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Similar documents
Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Organochlorine Pesticides by GCxGC-ECD

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

ApplicationNOTE THE ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBS) BY GC-HIGH RESOLUTION MASS SPECTROMETRY USING THE MICROMASS AUTOSPEC ULTIMA NT

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Yokogawa Analytical Systems Inc., Nakacho, Musashino-shi, Tokyo , Japan.

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

MULTIRESIDUE ANALYSIS OF PRIORITY POLLUTANTS IN SURFACE WATERS USING EXACT MASS GC-TOF MS

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

U.S. EPA Method 8270 for multicomponent analyte determination

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

Highly Sensitive and Rugged GC/MS/MS Tool

Research of Polychlorinated Biphenyls (PCBs) in Vegetables by GC-MS/MS

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

Multi-residue analysis of pesticides by GC-HRMS

DETERMINATION OF PESTICIDES IN FOOD USING UPLC WITH POLARITY SWITCHING TANDEM QUADRUPOLE LC/MS/MS

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

Sensitive Detection of Pyrethroids in Surface Water and Sediment

Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis

GC/MS/MS Dioxin and Dioxin-like PCBs in Feed and Food. Jessica Westland, LSAG Applications Chemist

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

ApplicationNOTE Abstract

Rugged GC/MS/MS Pesticide Residue Analysis Fulfilling the USDA Pesticide Data Program (PDP) Requirements

Validation Report 18

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

Application of gas chromatography atmospheric pressure chemical ionization mass spectrometry for analysis of contaminants in environmental samples

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

Introduction to Gas Chromatography

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

ApplicationNOTE ULTRA TRACE ANALYSIS OF DIOXINS AND FURANS IN HUMAN ADIPOSE TISSUE USING SFE-LC EXTRACTION/CLEANUP AND THE WATERS AUTOSPEC ULTIMA NT

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Fast and Accurate Quantitation of Perfluorinated Sources from Textiles using Gas Chromatography-Triple Quadrupole Mass Spectrometry

Orbitrap GC-MS: An Opportunity to Help Address the Challenges of Chlorinated Paraffins Analysis

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

High Throughput Multi-Residue Screening of Drinking Water using the SPE-DEX~ and Pegasus GC-TOF-MS

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

METHOD 3600C CLEANUP

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Analysis of Pesticides and Contaminants using HS-SPME-GC/MS and DRS Detection. Blair Surridge, Tyler Dyer, Rene Bilideau, Paloma Bertolin Presoti

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

OF ANALYSIS FOR DETERMINATION OF PESTICIDES RESIDUES IN FOOD (CX/PR 15/47/10) European Union Competence European Union Vote

Three-Fold Increase in Productivity for Pesticide Residue Analysis in Baby Food Using Fast Triple Quadrupole GC-MS/MS

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Evaluating CLP and EPA Methods for Pesticides in Water Using Agilent J&W DB-CLP1/DB-CLP2 GC Columns

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis

Sulfotepp impurities in Chlorpyrifos EC formulations

Polychlorinated Biphenyl Compounds (PCBs) Determination from Water by Gas Chromatography Coupled with Mass Spectrometry (GC MS)

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Validation Report 8 Determination of pesticide residues in rice and wheat by GC-MS/MS and LC-MS/MS (QuEChERS method) Appendix 2

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Analysis, fate and risks of organic contaminants in river basins under water scarcity Valencia, 7-8 February POPs by GC-MS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Hints and Tips for APGC Coupled with Xevo TQ-S for Food and Environmental Analysis

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Signal, Noise, and Detection Limits in Mass Spectrometry

WADA Technical Document TD2003IDCR

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

PosterReprint INTRODUCTION

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Fast and Accurate Analysis of Fluorotelomer Alcohols and Acrylates using Triple Quadrupole GC-MS/MS

KSA Environmental Laboratory Inc.

GC Analysis of Polybrominated Flame Retardants Application

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

ENVIRONMENTAL analysis

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Analysis of Total PCBs and PCB Congeners and Trans-nonachlor in Fish by Gas Chromatography/ Negative Chemical Ionization Single Ion Mass Spectrometry

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Determination of urea in ultrapure water by IC-MS/MS

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Transcription:

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT Peter Hancock 1, Mike Hetmanski 2 and Richard J. Fussell 2 1 Waters Corporation, Atlas Park, Simonsway, Manchester M22 5PP, UK 2 Central Science Laboratory, Sand Hutton, York YO41 1LZ, UK INTRODUCTION The regulations governing permitted levels of persistent organic pollutants (POPs) in food products are becoming more stringent in response to increased awareness of the hazards they pose to humans. POPs, such as organochlorine pesticides (OCs) and polychlorinated biphenyls (PCBs), have been linked to adverse effects such as cancer, damage to the nervous system, reproductive disorders, and disruption of the immune system. They do not readily degrade in the environment, are lipophilic and tend to bioaccumulate as they move through the food chain. Therefore, these can potentially occur at high concentrations in fatty foods, such as meats and fish. Due to concerns about the effects of POPs, an international treaty restricting their use was adopted via the Stockholm Convention in 2004 1. Currently in the UK, maximum residue limits (MRLs) of OCs and synthetic pyrethroids (SPs) in meat products are between 0.02 1.0 mg/kg on a fat basis 2. To enforce these regulations, high quality analytical methods with adequate confirmation and limits of quantification must be used. OCs, PCBs and SPs have all been analysed by conventional gas chromatography (GC) with electron capture detection (ECD). This is a sensitive technique for compounds containing halogens but suffers extensively from matrix interferences and confirmatory evidence cannot be provided as it is only a two dimensional technique with retention time and intensity. To add a third dimension, mass spectrometry (MS) can be used with selected ion monitoring (SIM). Usually three or four selected ions are monitored for each compound but confirmatory evidence is reduced if one or more of the selected ions are affected by matrix interferences. The likelihood of interference will increase as the amount of sample preparation decreases. To minimize sample preparation and/or the likelihood of interference the selectivity of the detection technique should be increased. The approach described in this application note is the use of tandem quadrupole MS/MS for the quantification and unequivocal confirmation of OCs, International Council for the Exploration of the Seas (ICES) 7 PCBs and SPs at low μg/kg levels in animal fat without the need for an additional clean-up technique following gel-permeation chromatography (GPC). METHODS Extraction Method 1.25 g melted porcine fat was weighed into a 10 ml volumetric flask. For recovery, the samples were spiked at 0.00625 μg/ml or 0.05 mg/kg. The volume was adjusted with ethyl acetate / cyclohexane (1:1). A 1 ml extract was cleaned-up by GPC to give a final extract with a matrix equivalent of 0.125 g/ml in hexane. The internal standard, δ-hch, was added at a concentration of 0.050 μg/ml equivalent to 0.4 mg/kg. GC Method The samples were injected by splitless injection (4 μl, 250 ºC, purge at 30 ml/min after 2.1 min) into a carrier gas of helium at a constant flow rate of 1.0 ml/min delivered from an Agilent 6890 GC with a CTC CombiPal autosampler attached. The column employed was a J & W Scientific DB-17MS 30 m x 0.25 mm i.d., 0.25 μm. The following temperature ramp rate was used: 100 ºC (1 min) to 200 ºC (6 min) at 20 ºC/min, to 280 ºC/min (15 min) at 10 ºC/min. The total run time was 35 min. The temperature of the interface was held at 260 ºC during the chromatographic run.

MS Method The Waters Micromass Quattro micro GC tandem quadrupole mass spectrometer was used (Figure 1) in electron impact (EI+) mode. The ion source was operated at 180 ºC with an electron energy of 70 ev and a trap current of 200 μa. The mode of acquisition was multiple reaction monitoring (MRM) at an argon collision gas pressure of 3.0 x 10-3 mbar. The Quattro micro GC was tuned so that the precursor and product ions were resolved with a peak width at half height of less than 0.7 Da. The list of pesticide residues and the MRM transitions, along with the collision energies and retention times for the method are listed in Table 1. The data were acquired using Waters MassLynx software and processed using the Waters TargetLynx Application Manager. RESULTS AND DISCUSSION A typical reconstructed ion chromatogram (RIC) for the twenty OCs, seven PCBs and nine SPs is illustrated in Figure 2. The MRM transitions were arranged into thirteen function windows represented by the different colors in Figure 2. These functions can be overlapped slightly to allow for small changes in retention time. Use of function windows allows more time to be spent on each MRM transition, thereby, improving the signal to noise (S/N) ratio for the analytes that gave the lowest response. Dwell times were set so that approximately fifteen data points described each chromatographic peak. Figure 2. RIC for the OCs, PCBs and SPs. Comparison of SIM and MRM Two selected mass chromatograms are illustrated in Figure 3 for hexachlorobenzene (8.52 min) acquired in SIM mode at a concentration in matrix of 0.01 mg/kg. For complete confirmation, three or four selected masses would need to be monitored, so this could impact on the sensitivity shown here. In this example the S/N ratios of both chromatograms would allow for routine integration and quantification of hexachlorobenzene in SIM mode. Hexachlorobenzene 284 286 Figure 3. Hexachlorobenzene in SIM mode. Figure 1. Waters Micromass Quattro micro GC tandem quadrupole mass spectrometer.

Pesticide Retention Quantification Transition Confirmation Transition Time Precursor Ion Product Ion CE Precursor Ion Product Ion CE Hexachlorobenzene 8.38 284 249 15eV 286 251 15eV Alpha-HCH 8.78 219 183 5eV 181 145 15eV Gamma-HCH 10.00 219 183 5eV 181 145 15eV Beta-HCH 10.75 219 183 5eV 181 145 15eV Heptachlor 11.12 272 237 13eV 274 239 15eV PCB 28 11.37 256 186 15eV 258 186 15eV Delta-HCH, Internal Standard 12.20 219 183 5eV 181 145 15eV Aldrin 12.53 263 193 25eV 263 191 25eV PCB 52 12.74 290 220 23eV 292 220 23eV Oxychlordane 14.26 185 149 5eV 185 121 15eV Heptachlor epoxide 15.12 183 155 10eV 217 182 15eV trans-chlordane 15.42 373 266 15eV 375 266 15eV PCB 101 15.64 326 256 25eV 328 256 25eV cis-chlordane 15.84 373 266 15eV 375 266 15eV Alpha-Endosulfan 15.96 241 206 12eV 241 170 20eV pp-dde 16.68 246 176 21eV 318 248 18eV Dieldrin 16.87 241 206 12eV 237 143 20eV PCB 118 17.54 326 256 25eV 328 256 25eV Endrin 17.82 263 193 30eV 263 191 30eV PCB 153 17.86 360 290 25eV 362 290 25eV op-ddt 18.10 235 165 20eV 237 165 20eV pp-ddd 18.27 235 165 20eV 237 165 20eV Beta-Endosulfan 18.47 241 206 12eV 241 170 20eV PCB 138 18.89 360 290 25eV 362 290 25eV pp-ddt 18.95 235 165 20eV 237 165 20eV Resmethrin 19.12 171 143 5eV 171 128 15eV Bifenthrin 19.18 181 166 15eV 181 165 20eV Endosulfan sulfate 19.57 272 237 11eV 272 235 11eV PCB 180 20.16 394 324 22eV 396 324 22eV Lambda-Cyhalothrin 20.57 197 141 10eV 181 152 20eV Permethrin 22.70 163 127 5eV 163 91 10eV Cyfluthrin 23.70 163 127 5eV 163 91 10eV Cypermethrin 24.70 163 127 5eV 163 91 10eV Flucythrinate 25.00 199 157 9eV 157 107 10eV Fenvalerate 28.00 167 125 10eV 125 89 15eV Deltamethrin 31.20 253 93 15eV 181 152 20eV Table 1. Pesticide residues and MRM method parameters.

Two selected mass chromatograms are illustrated in Figure 4 for α-hch (8.92 min) acquired in SIM mode at a concentration in matrix of 0.01 mg/kg. In this example, there is matrix interference (8.83 min) immediately preceding the confirmation peak. For this compound in this matrix, monitoring by SIM is unlikely to allow routine integration and quantification. α-hch 181 219 Matrix matched standards were prepared in the concentration range 0.1 to 25 ng/ml. Equivalent mg/kg concentrations were at the 0.0008, 0.0016, 0.0040, 0.0060, 0.0100, 0.0200, 0.0400, 0.0800 and 0.2000 mg/kg. δ-hch was used as an internal standard to correct for any volumetric errors. The standards were injected in a typical batch analysis. The data was then processed using Waters TargetLynx Application Manager. Correlation coefficients of r 2 > 0.9970 were obtained for all compounds in matrix. A representative curve for α-endosulfan with a correlation coefficient of r 2 = 0.9998 is illustrated in Figure 6. Compound name: Alpha-Endosulfan Correlation coefficient: r = 0.999910, r^2 = 0.999821 Calibration curve: 0.440823 * x + -6.18479e-005 Response type: Internal Std ( Ref 6 ), Area * ( IS Conc. / IS Area ) Curve type: Linear, Origin: Exclude, Weighting: Null, Axis trans: None 0.080 Figure 4. α-hch in SIM mode. In contrast, changing to MRM mode the selectivity for α-hch (8.92 min) is significantly improved at the same concentration in matrix (Figure 5). In this example, routine integration and quantification by MRM could easily be achieved. Response 0.060 0.040 0.020 0.000 mg/kg 0.000 0.050 0.100 0.150 0.200 Figure 6. Representative calibration curve for α- endosulfan, 0.0008 0.2000 mg/kg. α-hch 219 183 181 145 Figure 5. α-hch in MRM mode. In summary, some compounds will have acceptable selectivity for quantification and confirmation via SIM mode. However, as the matrix complexity is increased and/or the sample clean-up is reduced, other compounds require the more selective MRM technique for routine quantification and confirmation. To test the extraction method described, four recovery experiments were performed in porcine fat, spiked at 0.05 mg/kg. Each analyte was compared to its calibration curve of matrix-matched standards. The mean recovery and relative standard deviation (RSDs) of each analyte is listed in Table 2. The mean recoveries were in the range 72 to 123%, with RSDs between 1 and 23%. Inclusion of SPs compromises the clean-up and results in higher concentrations of matrix co-extractives in the final extract. Further optimization of the extraction to improve precision is being carried out. The high response and selectivity provided by tandem quadrupole MS/MS ensured that additional clean-up of the GPC extracts was not necessary for the majority of pesticides.

% Mean Recovery % RSD % Mean Recovery % RSD Hexachlorobenzene 99 23 PCB 153 98 18 Alpha-HCH 110 7 op-ddt 85 10 Gamma-HCH 109 5 pp-ddd 116 14 Beta-HCH 109 16 Beta-Endosulfan 91 12 Heptachlor 105 12 PCB 138 97 17 Aldrin 110 5 pp-ddt 72 11 PCB 52 97 20 Resmethrin 109 18 Oxychlordane 108 14 Bifenthrin 112 16 Heptachlor epoxide 123 10 Endosulfan sulfate 113 15 cis-chlordane 114 17 PCB 180 93 21 PCB 101 95 21 Lambda-Cyhalothrin 111 13 trans-chlordane 109 14 Permethrin 102 10 Alpha-Endosulfan 107 10 Cyfluthrin 99 11 pp-dde 103 1 Cypermethrin 95 14 Dieldrin 107 9 Flucythrinate 108 19 PCB 118 94 20 Fenvalerate 102 12 Endrin 101 12 Deltamethrin 83 13 Table 2. Mean Recovery and % RSD for 0.05 mg/kg recovery samples (n = 4). Limits of detection (LOD) and confirmation (LOQ) were determined from the matrix matched standards, and are illustrated in Figure 7. The LOD was defined as the amount injected that gave a signal equivalent to three times the baseline noise. All the compounds have an LOD of 0.006 mg/kg or less. The LOQ was defined as the response for both the quantification and confirmation transitions sufficient to confirm the identity. As all the UK MRLs are greater than 0.02 mg/kg in animal products, this confirmatory method is able to meet the current legislation requirements. 14 12 LOD LOQ Two MRM transitions were chosen so that quantification and confirmation could be performed in a single run assuming that the ion ratio between the transitions was consistent between standards and samples. The confirmation criterion chosen was that the ion ratio of the sample was within ±10% of the standard. The ratio was then used to confirm or reject compounds in the extracts. To illustrate sensitivity, selectivity and confirmatory ability of tandem quadrupole MS/MS, heptachlor (Figure 8), PCB 52 (Figure 9) and bifenthrin (Figure 10) at the concentration level are compared to the matrix blank. Number of Residues 10 8 6 4 2 0 <0.0008 0.0008 0.0016 0.0040 0.0060 0.0100 0.0200 >0.0200 mg/kg For both heptachlor and PCB 52, the expected ion ratio from the solvent standards is 0.62. For the confirmation criteria to be passed any concentrations in the extracts must have a ratio between 0.56 and 0.68. At 0.01 mg/kg the ion ratios for heptachlor and PCB 52 are 0.60 and 0.63 respectively, so the presence of both can be confirmed at this level. Figure 7. Limits of detection and confirmation for all residues in mg/kg.

Peak Area Ratio 0.60 For bifenthrin, the expected ion ratio from the solvent standards is 0.86. For the confirmation criteria to be passed, any concentrations in the extracts must have a ratio between 0.78 and 0.95. At the ion ratio is 0.89 so the presence of bifenthrin can be confirmed at this level. Blank 272 237 Peak Area Ratio 0.89 274 239 Figure 8. Sensitivity, selectivity and confirmation of heptachlor,. Blank 181 166 181 165 Peak Area Ratio 0.63 Figure 10. Sensitivity, selectivity and confirmation of bifenthrin,. Blank 290 220 292 220 Figure 9. Sensitivity, selectivity and confirmation of PCB 52,.

CONCLUSIONS A method has been described for the determination and confirmation of OCs, PCBs and synthetic pyrethroids in animal fat using the Waters Micromass Quattro micro GC tandem quadrupole mass spectrometer. The Quattro micro GC provides high selectivity to reduce any matrix interferences, high sensitivity to reach the reporting levels required by European Union legislation and quantitative and confirmatory data in a single injection. The tandem quadrupole MS/MS technique should be applicable to a much larger range of analytes in a range of more complex food commodities. The TargetLynx Application Manager provides advanced quantification with a range of automatic quality control checks. REFERENCES 1. http://www.pops.int/ 2. Statutory Instrument No. 2591, The Pesticides (Maximum Levels in Crops, Food and Feeding Stuffs) Amendment No. 2 (2003). Waters, Micromass, Quattro micro, MassLynx and TargetLynx are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2005 Waters Corporation Produced in the U.S.A. August 2005 720001329EN SE-PDF