Recovery/bias evaluation

Similar documents
Ivo Leito University of Tartu

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

Validation of an Analytical Method

Uncertainty of Measurement (Analytical) Maré Linsky 14 October 2015

Measurement Uncertainty, March 2009, F. Cordeiro 1

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

Reproducibility within the Laboratory R w Control Sample Covering the Whole Analytical Process

NORDTEST NT TR 537 edition :11

The Role of Proficiency Tests in the Estimation of Measurement Uncertainty of PCDD/PCDF and PCB Determination by Isotope Dilution Methods

Validation of analytical methods. Adrian Covaci Toxicological Center, University of Antwerp

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

Method Validation and Accreditation

Appendix II- Bioanalytical Method Development and Validation

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Validation and Standardization of (Bio)Analytical Methods

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

The bias component in measurement uncertainty

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

What is measurement uncertainty?

Measurement uncertainty: Top down or Bottom up?

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Protocol for the design, conducts and interpretation of collaborative studies (Resolution Oeno 6/2000)

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

Estimating limit of detection for mass spectrometric analysis methods

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Calculation of Measurement Uncertainty

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Application of Detection and Quantification Concepts to Chlorine Residual Measurements

Measurement uncertainty revisited Alternative approaches to uncertainty evaluation

ACCURACY AND PRECISION

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

VAM Project Development and Harmonisation of Measurement Uncertainty Principles

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

LC-MS/MS in the Clinical Laboratory. Jo Adaway

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

Measurement Uncertainty Principles and Implementation in QC

Understanding the Uncertainty Associated with Analytical Results: Sources, Control and Interpretation of Results. Marc Paquet, Chemist M.Sc.

Figure 1. The supported liquid extraction process using the ISOLUTE SLE plate (single well shown)

BD Vacutainer Barricor Blood Collection Tube is the Tube of Choice for LC- MS/MS Analysis of Bioactive Cannabinoids in Plasma

Ch 3. EXPERIMENTAL ERROR

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

Chapter 4: Verification of compendial methods

Analytical Measurement Uncertainty

25-OH VITAMIN D 3 AND 25-OH VITAMIN D 2 IN PLASMA BY UV / VIS FAST CODE Z19710

Strategies for the Analysis of Therapeutic Peptides in Biofluids by LC-MS/MS. Lee Goodwin

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Pesticide Standards. Sadat Nawaz. Pesticides Residues Analyst CSL, York, UK

Validation Report 18

A basic introduction to reference materials. POPs Strategy

Really, A Revised MDL Procedure

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

25-OH VITAMIN D 3 IN PLASMA BY UV FAST CODE Z19610

Estimating MU for microbiological plate count using intermediate reproducibility duplicates method

The Theory of HPLC. Quantitative and Qualitative HPLC

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

Absolute Value Equations and Inequalities. Use the distance definition of absolute value.

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Performance characteristics of analytical tests

Definitive EtG/EtS LC-MS/MS Analysis:

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

ICP-OES DETERMINATION OF IRON. Introduction:

Proficiency testing: Aqueous ethanol. Test and measurement Workshop Marcellé Archer. 20 September 2011

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

A Pre-Clinical Pharmacokinetic-Pharmacodynamic Modelling and Biodistribution Studies of Donepezil Hydrochloride by Validated HPLC Method

Quantitative Analysis of β-lactam Antibiotics in Human Plasma by High Sensitivity LC/MS/MS Method

Practical and Valid Guidelines for Realistic Estimation of Measurement Uncertainty in Pesticide Multi-Residue Analysis*

Chemical derivatization

Method Development and Validation for the Estimation of Darunavir in Rat Plasma by RP-HPLC

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

Method Validation Characteristics through Statistical Analysis Approaches. Jane Weitzel

HPLC analysis of VFAs, Furfural and HMF

Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Examples of Method Validation Studies Conducted in Different Economies

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

The Analysis of Fluoroquinolones in Beef Kidney Using HPLC Electrospray Mass Spectrometry Application

Transcription:

Recovery/bias evaluation Ivo Leito University of Tartu ivo.leito@ut.ee 1

Overview What is bias? What are the main issues in bias evaluation? Time frame Random and systematic effects Approaches for bias evaluation Some examples 2

The measurement cycle Client Decision on result Client issue Data evaluation Define issue Report on measurement Evaluation Analysis Sampling Decision on measurement Client Interface Analytical chemist Analytical chemist 3

Benjamin Franklin 1706-1790 Tell me and I forget. Show me and I forget. Involve me and I remember. 4

Bias is What is bias? 1. difference between the measured value and the true value 2. Difference between the measured value and a reference value 3. Difference between the mean of a large number of replicate measured values and the true value 4. Difference between the mean of a large number of replicate measured values and a reference value For eliminating the random effects We never know the true value 26.05.2014 VIM: Bias is an estimate Analytical of a Measurements systematic measurement error 5

Bias and related concepts Repeatability/ Reproducibility Bias Uncertainty range of C i C i U C i + U C i -U C m C i C i +U Measurand, C C Ref Individual measured value Reference value Mean measured value 6

Which of these situations describe bias? 1. All the results of a specific day are systematically influenced by the calibration graph of that day 2. Delicate analyte partially decomposes during sample preparation leading to lowered results 3. The titrant concentration determined on a particular day is slightly lower or higher than the true concentration 4. Because of the specifics of the used sample preparation procedure the sample is digested incompletely, leading to lowered values 7

Does bias depend on the time frame? 1. Yes, bias determined within a single day is different from one determined on different days (and averaged) 2. No 8

Systematic and random effects Random and systematic effects can be grouped differently: Within-day bias Repeatability s r Long-term bias Intermediate precision s RW All effects causing error/uncertainty The longer is the time frame the more effects change their status : 26.05.2014 systematic Analytical Measurements random 9

Why is lab/method bias more useful than within-day bias? Within-day bias should be redetermined every day Long-term bias can be determined less frequently It is useful to work with the lowest possible bias s RW can be determined more reliably than bias It is good if most of the uncertainty sources are included into the random component s RW From now on in this session we only address 26.05.2014 the long-term Analytical bias Measurements (lab/method bias) 10

Example: LC-MS determination of a delicate bioactive compound in blood plasma Effect Systematic within day Systematic in long term Calibration graph of a specific day Y (N) Injection volume of autosampler is 5% higher than nominal N N Delicate analyte partially decomposes at room temperature before samples are loaded into cooled autosampler (N) (N) Repeatability of peak integration N N Ionization suppression in the ESI source by a co-eluting compound Y Y Baseline noise (N) (N) 11

Which are important issues in determining bias? Issue Bias s RW Sufficient number of replicates Y Y Sufficiently long timeframe Y Y Homogeneous sample Y Y Matrix match Y Y Concentration range match Y Y Reliable reference value Y N Determination of one can be hindered by the other Y N 12

Which are the best approaches for determining bias? Approach Analysing spiked blank matrix Replicate measurements of a routine sample Using a PT sample and consensus value as reference value Analysing a CRM Analysing a routine sample with a reference procedure How good? 13

How to calculate/express bias? Way of expressing Absolute bias Relative bias Recovery Recovery bias bias R R Formula C lab C C C _ mean lab _ mean uncorrected C C ref Ref C1 C C 0 C ref ref When to use? If bias is absolute or the C range is narrow If bias is proportional Sample with ref value, bias is proportional Spiking, bias is proportional 14

How to conduct a spiking experiment? Two analysts determined meropenem (an antibiotic) in blood plasma. Both needed to determine the recovery of the procedure. They obtained blank plasma samples and did the following: Analyst 1 took 500 μl of the blank plasma and added 400 μl of methanol for protein precipitation. He separated the precipitated proteins by centrifugation and transferred the supernatant into an HPLC vial. 100 μl of meropenem standard solution with suitable concentration was added to the supernatant and the resulting solution was injected into the HPLC system for analysis. Analyst 2 took 500 μl of the blank plasma, added 100 μl of meropenem standard solution and mixed well. She then added 500 μl of methanol for protein precipitation. She separated the precipitated proteins by centrifugation and injected the resulting supernatant into the HPLC system for analysis. Which analyst did it more correctly? Why? 26.05.2014 Analyte has to be added Analytical at Measurements as early stage as possible! 15

How to evaluate uncertainty due to bias? RMS bias ( bias n i 2 ) u( Cref ) u( Cref n i 2 ) 2 2 bias u( ) u( bias) RMS Cref This component accounts for the average bias of the laboratory results from the This component accounts for the average uncertainty of the reference values C 26.05.2014 ref Analytical Measurements C ref 16

http://sisu.ut.ee/measurement/ 17

Thank you for your participation! The slides are available from: http://www.ut.ee/ams/eurachem-2014/ More explanations and examples: http://sisu.ut.ee/measurement/ You are always welcome to contact me: ivo.leito@ut.ee 18