Detection of X-ray spectra and images by Timepix

Similar documents
Hard X-ray Vela supernova observation on rocket experiment WRX-R

Kirkpatrick Baez X ray optics for astrophysics: Recent status

TIMEPIX3 First measurements and characterization of a hybrid pixel detector working in event driven mode

Characterization and Directional Visualization of Space Radiation Quanta in Low Earth Orbit with the compact Spacecraft Payload SATRAM

Biomimetics and astronomical X-ray optics

Scientific astrophysical payloads for pico and nano-satellites

Study of the radiation fields in LEO with the Timepix detector

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

Pole searching algorithm for Wide-field all-sky image analyzing monitoring system

X- & γ-ray Instrumentation

The X-Ray Universe. Potsdam University. Dr. Lidia Oskinova Wintersemester 2008/09

Measurement of the Ionizing Energy Depositions after Fast Neutron Interactions in Silicon

Abstract: J. Urbar [1], J. Scheirich [2], J. Jakubek [3] MEDIPIX CR tracking device flown on ESA BEXUS-7 stratospheric balloon flight

Energy calibration of the threshold of Medipix for ATLAS

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

MIRAM Miniaturized Radiation Monitor

Detector R&D in KAPAC

Mass Determination of Rn and Hg isotopes using MASHA

Status Report: Charge Cloud Explosion

Earth s thermal radiation sensors for attitude determination systems of small satellites

Detector R&D at KIPAC. Hiro Tajima Kavli InStitute of Particle Astrophysics and Cosmology

CCD OPERATION. The BBD was an analog delay line, made up of capacitors such that an analog signal was moving along one step at each clock cycle.

Peter Fischer, ziti, Universität Heidelberg. Silicon Detectors & Readout Electronics

Characterization of 3D thermal neutron semiconductor detectors

Special SLS Symposium on Detectors

René Hudec Czech Technical University in Prague, Faculty of Electrical Engineering, Czech Technical

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

arxiv: v2 [physics.ins-det] 8 Feb 2013

The Silicon-Tungsten Tracker of the DAMPE Mission

The distribution of electron energy is given by the Fermi-Dirac distribution.

Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up

Outline. Introduction, motivation Readout electronics, Peltier cooling Input J-FETsJ

ISPA-Tubes with YAP:Ce Active Windows for X and Gamma Ray Imaging.

Detector R&D at KIPAC

Advacam, Prague, Czech Republic. Nuclear Physics Institute, Czech Academy of Sciences, Prague, Czech Republic

How we wanted to revolutionize X-ray radiography, and how we then "accidentally" discovered single-photon CMOS imaging

SYNTHESIS OF SUPERHEAVY ELEMENTS USING THE MASS SPECTROMETER MASHA

Centre for Electronic Imaging

Performance of a Si PIN photodiode at low temperatures and in high magnetic fields

Light-pollution measurement with the Wide-field all-sky image analyzing monitoring system

Charge sharing in silicon pixel detectors

High-Sensitivity Gamma-Ray Imaging With Double Sided Strip Detectors

Study of Solder Ball Bump Bonded Hybrid Silicon Pixel Detectors at DESY

New trends in CdTe detectors for X and γ-ray applications

M. Martišíková 1,3, C. Granja 2, J. Jakůbek 2, B. Hartmann 1,3, K. Gwosch 1, P. Soukup 2 and O. Jäkel 1,3,5

Key words: avalanche photodiode, soft X-ray detector, scintillation γ-ray detector, imaging device PACS: 07.85;95.55.A;85.60.D

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

detector development Matthias Beilicke X ray Science Analysis Group meeting (12 April 2013, Monterey, CA) Collaborators: GSFC, BNL

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

High-performance direct conversion X-ray detectors based on sintered hybrid lead triiodide perovskite wafers

Development of a Silicon PIN Diode X-Ray Detector

Impact of high photon densities on AGIPD requirements

Enhanced lateral drift sensors: concept and development. TIPP2017, Beijing

A super-high angular resolution principle for coded-mask X-ray imaging beyond the diffraction limit of a single pinhole

Alpha-particle Stopping Powers in Air and Argon

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation.

History of Astronomical X-ray Telescopes in the Czech Republic

Studies of Hadron Calorimeter

He-3 Neutron Detectors

COMMISSIONING TESTS OF THE NEW NEUTRON RADIOGRAPHY FACILITY AT THE LVR-15 REACTOR ABSTRACT

Auxiliaire d enseignement Nicolas Ayotte

Fast Neutron and Gamma-Ray Detectors for the CSIRO Air Cargo Scanner

Development of the GridPix Detector for Dual Phase Noble Gas Time Projection Chambers

LAB 4: Gamma-ray coincidence spectrometry (2018)

The Mini-Calorimeter detector for the AGILE mission

1. Give short answers to the following questions. a. What limits the size of a corrected field of view in AO?

NUCL 3000/5030 Laboratory 2 Fall 2013

Development of a new MeV gamma-ray camera

1) CMOS-Sensors for vertex-detectors 2) CMOS-Sensors for X-ray imaging 3) Sensor with a 3-T-preamplifier 4) Sensor with nearly full depletion

SMILE Solar wind Magnetosphere Ionosphere Link Explorer Novel and global X-ray imaging of the Sun Earth connection

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Development of CZT strip detector modules for to 1-MeV gamma-ray imaging and spectroscopy

pickup from external sources unwanted feedback RF interference from system or elsewhere, power supply fluctuations ground currents

Study of Edgeless TimePix Pixel Devices. Dylan Hsu Syracuse University 4/30/2014

Lecture 8. Detectors for Ionizing Particles

EE 5344 Introduction to MEMS CHAPTER 5 Radiation Sensors

Muon Telescope at BEO Moussala *

ORIGAMIX, A CDTE-BASED SPECTRO-IMAGER DEVELOPMENT FOR NUCLEAR APPLICATIONS

A CdTe detector with a Gd converter for thermal neutron detection

RADIATION detectors based on crystalline silicon are well

Status Report: Multi-Channel TCT

Chandra was launched aboard Space Shuttle Columbia on July 23, 1999!!!

Your Partner in Environment Monitoring

Silicon Detectors. Particle Physics

DEPFET sensors development for the Pixel Detector of BELLE II

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves

Radiation Detector 2016/17 (SPA6309)

Position sensitive detection of thermal neutrons with solid state detectors (Gd Si planar detectors)

Geant4 simulation of SOI microdosimetry for radiation protection in space and aviation environments

Impact of high photon densities on AGIPD requirements

Development of New MicroStrip Gas Chambers for X-ray Applications

The Hermes Recoil Silicon Detector

Alpha-Energies of different sources with Multi Channel Analyzer

08 - Miscellaneous and historical detectors

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

A t XFEL experiment, sensors should have. Capacitance calculations in p + n silicon pixel sensors using three dimensional TCAD simulation approach

Advances in Compound Semiconductor Radiation Detectors. a review of recent progress

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici

Transcription:

Contrib. Astron. Obs. Skalnaté Pleso 47, 151 156, (2017) Detection of X-ray spectra and images by Timepix M. Urban, O. Nentvich, V. Stehlikova and L. Sieger Czech Technical University in Prague, Faculty of Electrical Engineering Technicka 2, Prague 166 27, Czech Republic (E-mail: urbanm24@fel.cvut.cz) Received: May 11, 2017; Accepted: June 28, 2017 Abstract. X-ray monitoring for astrophysical applications mainly consists of two parts - optics and detector. The article describes an approach based on a combination of Lobster Eye (LE) optics with Timepix detector. Timepix is a semiconductor detector with 256 256 pixels on one electrode and a second electrode is common. Usage of the back-side-pulse from an common electrode of pixelated detector brings the possibility of an additional spectroscopic or trigger signal. In this article are described effects of the thermal stabilisation, and the cooling effect of the detector working as single pixel. Key words: X-ray Spectrum Timepix detector Lobster Eye optics 1. Introduction Many tasks of recent astrophysics require to observe celestial X-ray sources over the wide field of view and wide spectral range. The possibility to detect the spectrum of source together with its image is a logical requirement of planned astronomical X-ray experiments. Due to the significant attenuation of X-ray radiation when passing through the atmosphere, most measurements are placed above it. Numerous satellites have already been launched for this reason, and more missions are planned. As examples of successfull missions can be mentioned Chandra (Weisskopf et al., 2000; Schwartz, 2014) and XMM Newton (Jansen et al., 2001) satellite or erosita (Merloni et al., 2012) as mission in preparation. All these missions use the Wolter-I type of X-ray optics (Wolter, 1952) and Charge-coupled device (CCD) X-ray detector. Thermal noise minimization is required for a CCD detector to maximise spectral resolution. It is achieved by cooling to temperatures approximately from -60 to -100 degrees Celsius. This requirement causes the disadvantage in high energy consumption or the need of a cooling medium. Another way to observe celestial X-ray sources with imaging device is a combination of Lobster Eye optics with Timepix detector. It can bring a wider field of view and lower requirements for thermal stabilisation of the detector, but only moderate angular resolution.

152 M. Urban, O. Nentvich, V. Stehlikova, and L. Sieger 2. Experiments There are many different ways to use Lobster Eye (LE), in astronomy (Hudec et al., 2015; Pína et al., 2016) and also in other fields of research, like colliders and medicine. Czech CubeSat VZLUSAT-1 can be mentioned as an example. This small satellite has an X-ray optics and several payloads for testing carbon fibre reinforced plastic onboard. It is prepared to launch in June 2017 (Urban et al., 2016; Pina et al., 2015). This satellite carries the wide-field monitor demonstrator based on Lobster Eye optics with Timepix detector (Daniel et al., 2016; Dániel et al., 2016; Baca et al., 2016; Blazek et al., 2017). Other example is in the proposed rocket experiment based on LE optics and the Timepix detector for wide-field sensing (Dániel et al., 2017). 3. X-ray detector - Timepix The Timepix detector is a pixel-particle-counting detector. It is a hybrid pixel detector with (256 x 256) pixels with a pitch of 55 µm, therefore the detection area has 196 square millimetres. This detector consists of a silicon semiconductor bump-bonded to Complementary Metal-Oxide-Semiconductor (CMOS) readout circuit. Each pixel has its own charge sensitive preamplifier, threshold discriminator and 14 bit counter. Timepix detector have approx. 3 kev threshold without any upgrades. (Esposito et al., 2011; Llopart et al., 2007) Figure 1. The hybrid semiconductor device Timepix (Platkevic, 2014; Urban et al., 2017)

Detection of X-ray spectra and images by Timepix 153 One of the improvements can be a usage of a back-side-pulse and thermal stabilisation. (Platkevic et al., 2011; Holik et al., 2016) The silicon chip is equipped with a single backside electrode (common electrode in Fig. 1) from the opposite side than the matrix, where the bias voltage is connected as well. The signal from this common electrode can be used for the X-ray detection as well. If we use a detector from this side, it behaves like a one-pixel detector instead of the matrix of (256 x 256) pixels on the other side. The advantage of this approach is that we can use the signal from a larger detection area without the noise generated by the readout electronics under the chip. The disadvantage is that the capacity of the reading cell is noticeably higher, making it significantly more demanding for the reading electronics, mainly for the first stage of the preamplifier. The Timepix detector is an uncooled and temperature unstabilised detector in standard usage. The signal is affected by two kinds of noises, thermal and electronic readout noise. Thermal noise is generated in the small pixels and can be reduced by acquisition time and cut by the threshold, which is set individually for each pixel. The major noise in the Timepix is generated by readout of pixels. This kind of noise is generated directly by bump-bonded applicationspecific integrated circuit (ASIC) chip on the silicon detection layer and readout interface (Platkevic et al., 2011). In case the ASIC electronics of the detector is turned off, and the common back-side electrode is used, the reading noise is minimised. Measurement by the common electrode is then affected by the thermal noise error caused only. This measurement can cause inaccuracies because the noise characteristics of the detector vary with its temperature. One of the possibilities for an improvement of the Full Width at Half Maximum (FWHM) of the signal, the noise reducing and thus shifting the threshold to lower energy values is thermal stabilisation or cooling. Low-energy photons incident can be detected by Timepix3 as well. This generation of the Timepix detector has the minimum threshold around 400-500 e, which corresponds with 1.5-2 kev with a silicon sensor. (Turecek et al., 2016; Frojdh et al., 2015; Poikela et al., 2014) 3.1. Thermal stabilisation The charge sensitive amplifier is one of the most important parts for signal evaluation and spectral resolution. The accuracy depends on the amplifier s FWHM respectively on detector capacitance and its thermal noise. The measurement setup in Fig. 2 was used for measurement. With respect to the limits of measuring facility, the energy corresponding to 109.7 kev was selected for principle verification. It is connected across the detector to the amplifier. Pulses are generated with frequency of 50 Hz. The amplified signal is shaped and connected to the Multichannel Analyzer (MCA) where is sorted into one of the 8 192 channels. The measurements were performed at the room temperature 25 C and at the temperature below the freezing point. The detector with the amplifier was

154 M. Urban, O. Nentvich, V. Stehlikova, and L. Sieger Figure 2. Principle of measurement placed in a controlled environment at temperature -12 C and protected against the water condensation during cooling. The detector s FWHM can be determined from the equation (1) N = NDet 2 + N Amp 2 (1) where N is total FWHM of the system, N Det is the FWHM of the detector and N Amp is the FWHM of the system with an amplifier without the detector. From the equation (1) we express N Det, which is our demanded value of FWHM. Values N and N Amp are substituted by the measured values for each temperature as well. The results of measurement are listed in Tab. 1. 1200 1000 T = -12 C T = +25 C Counts (-) 800 600 400 200 0 100 102 104 106 108 110 112 114 116 118 120 Energy (kev) Figure 3. Resulting sensed signal from the detector at the room temperature 25 C and below the freezing point -12 C

Detection of X-ray spectra and images by Timepix 155 Table 1. Resulting detector s FWHM at different temperatures. Temperature FWHM 25 C 4.88 kev -12 C 2.87 kev 4. Conclusion Timepix, the CMOS detector allows imaging and measuring of radiation spectra at the same time. The proposed modification of the detector connection and sense of the generated signal potentially enables to expand the detection spectrum to lower energies. The verification measurements were performed in this arrangement, where the detector works as the single pixel detector. Then we got additive improvement of sensed signal FWHM by another modification of this detector. The thermal stabilisation and even only slight cooling effect of the detector brings a reduction of FWHM. The resulting FWHM was decreased by 2 kev when the temperature was reduced by 37 C. That means the resulting spectral resolution is 2,9 kev for the detector with capacitance 100 pf for the whole (single) pixel. Acknowledgements. This work was performed in cooperation with the companies Rigaku Innovative Technologies Europe, s.r.o. and ADVACAM s.r.o.. It was supported by the Grant Agency of the Czech Technical University in Prague, grant No. SGS16/169/OHK3/2T/13 and by GA CR grant 13-33324S. References Baca, T., Platkevic, M., Jakubek, J., et al. 2016, Journal of Instrumentation, 11, C10007 Blazek, M., Pata, P., Inneman, A., & Skala, P. 2017, Advances in Astronomy, 2017, 316289 Dániel, V., Inneman, A., Pína, L., et al. 2017, in Proceedings of the SPIE, Vol. 10235, EUV and X-ray Optics: Synergy between Laboratory and Space Dániel, V., Pína, L., Inneman, A., et al. 2016, in Proceedings of the SPIE, Vol. 9978, CubeSats and NanoSats for Remote Sensing, 99780D Daniel, V., Urban, M., Nentvich, O., & Stehlikova, V. 2016, in Proceedings of the SPIE, Vol. 9978, CubeSats and NanoSats for Remote Sensing, 99780N Esposito, M., Jakubek, J., Mettivier, G., et al. 2011, Nuclear Instruments and Methods in Physics Research A, 652, 458 Frojdh, E., Campbell, M., De Gaspari, M., et al. 2015, Journal of Instrumentation, 10, C01039

156 M. Urban, O. Nentvich, V. Stehlikova, and L. Sieger Holik, M., Kraus, V., Georgiev, V., & Granja, C. 2016, Journal of Instrumentation, 11, C02032 Hudec, R., Pina, L., Inneman, A., & Tichy, V. 2015, in Proceedings of the SPIE, Vol. 9510, EUV and X-ray Optics: Synergy between Laboratory and Space IV, 95100A Jansen, F., Lumb, D., Altieri, B., et al. 2001, Astronomy and Astrophysics, 365, L1 Llopart, X., Ballabriga, R., Campbell, M., Tlustos, L., & Wong, W. 2007, Nuclear Instruments and Methods in Physics Research A, 581, 485 Merloni, A., Predehl, P., Becker, W., et al. 2012, ArXiv e-prints Pina, L., Hudec, R., Inneman, A., et al. 2015, in Proceedings of the SPIE, Vol. 9510, EUV and X-ray Optics: Synergy between Laboratory and Space IV, 951005 Pína, L., Hudec, R., Inneman, A. J., et al. 2016, in Proceedings of the SPIE, Vol. 9964, Advances in Laboratory-based X-Ray Sources, Optics, and Applications V, 99640B Platkevic, M. 2014, PhD thesis, Czech Technical University in Prague Platkevic, M., Jakubek, J., Vykydal, Z., & Granja, C. 2011, Journal of Instrumentation, 6, C11023 Poikela, T., Plosila, J., Westerlund, T., et al. 2014, Journal of Instrumentation, 9, C05013 Schwartz, D. A. 2014, Review of Scientific Instruments, 85, 061101 Turecek, D., Jakubek, J., & Soukup, P. 2016, Journal of Instrumentation, 11, C12065 Urban, M., Nentvich, O., Stehlikova, V., et al. 2016, submitted to Acta Astronautica Urban, M., Nentvich, O., Stehlíková, V., & Sieger, L. 2017, in EUV and X-ray Optics: Synergy between Laboratory and Space, Proceedings of the SPIE Weisskopf, M. C., Tananbaum, H. D., Van Speybroeck, L. P., & O Dell, S. L. 2000, in Proceedings of the SPIE, Vol. 4012, X-Ray Optics, Instruments, and Missions III, ed. J. E. Truemper & B. Aschenbach, 2 16 Wolter, H. 1952, Annalen der Physik, 445, 94