arxiv: v2 [physics.ins-det] 23 May 2017

Similar documents
Development of a homogeneous, isotropic and high dynamic range calorimeter for the study of primary cosmic rays in space experiments

The TIC project: tracking gamma rays with a calorimeter

The LHCf experiment at LHC

Status of the PAMELA silicon tracker

Neutral particles energy spectra for 900 GeV and 7 TeV p-p collisions, measured by the LHCf experiment

PoS(Scineghe2014)007 THE GAMMA-400 SPACE OBSERVATORY: STATUS AND PERSPECTIVES. O. Adriani, on behalf of the GAMMA-400 collaboration

PAMELA satellite: fragmentation in the instrument

Istituto Nazionale di Fisica Nucleare, Sezione di Firenze and Physics Department of University of Florence Firenze, Italy

The High Energy cosmic-radiation Detection (HERD) Facility onboard China s Future Space Station

High Energy cosmic-radiation Detection (HERD) Facility onboard China s Space Station

arxiv: v1 [physics.ins-det] 20 Dec 2013

STATUS OF ATLAS TILE CALORIMETER AND STUDY OF MUON INTERACTIONS. 1 Brief Description of the ATLAS Tile Calorimeter

The LHCf experiment at LHC

Dual-Readout Calorimetry with a Mo-Doped PbWO4 Electromagnetic Section

CALICE Test Beam Data and Hadronic Shower Models

The GILDA mission: a new technique for a gamma-ray telescope in the energy range 20 MeV GeV

Measurement of Cosmic-Ray Electron and Positron Spectrum from CALET on the ISS

Study of muon bundles from extensive air showers with the ALICE detector at CERN LHC

The NUCLEON Space Experiment Preliminary Results. Skobeltsyn Institute of Nuclear Physics, Moscow State University, Moscow, , Russia

Correction for PMT temperature dependence of the LHCf calorimeters

Hands on LUNA: Detector Simulations with Geant4

PoS(ICRC2017)208. Particle tracking in the CALET experiment. Paolo Maestro (for the CALET Collaboration)

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

Measurements of Heavy Nuclei with the CALET Experiment

PoS(ICRC2017)1076. Studies of Cosmic-Ray Proton Flux with the DAMPE Experiment

PAMELA: a Satellite Experiment for Antiparticles Measurement in Cosmic Rays

Current and Future balloon and space experiments L. Derome (LPSC Grenoble) Tango, May 4-6th, 2009

Calibration and Performance of the ATLAS Tile Calorimeter During the LHC Run 2

Available online at ScienceDirect. Physics Procedia 74 (2015 )

The Large Area Telescope on-board of the Fermi Gamma-Ray Space Telescope Mission

CALOCUBE SVILUPPO DI CALORIMETRIA OMOGENEA AD ALTA ACCETTANZA PER ESPERIMENTI DI RAGGI COSMICI NELLO SPAZIO

Measurement of secondary particle production induced by particle therapy ion beams impinging on a PMMA target

Measurement of the neutron fraction event-byevent

Study of deuteron spectra under radiation belt with PAMELA instrument

Calorimetry I Electromagnetic Calorimeters

Hadronic Interaction Studies with ARGO-YBJ

Calorimetry in. in Nuclear and Particle Physics Experiments

Digital Hadron Calorimetry for the Linear Collider using GEM Technology

The space mission PAMELA

THE ELECTROMAGNETIC CALORIMETER OF THE AMS-02 EXPERIMENT

Temperature Dependence Calibration and Correction of the DAMPE BGO Electromagnetic Calorimeter

First indication of LPM effect in LHCf, an LHC experiment

Calibration of the CMS Electromagnetic Calorimeter with first LHC data

G. Gaudio, M. Livan The Art of Calorimetry Lecture V. The state of art Towards ILC calorimetry

Cosmic Ray Physics with the ARGO-YBJ experiment

Antimatter and DM search in space with AMS Introduction. 2 Cosmology with Cosmic Rays

CALICE scintillator HCAL

Geant4 simulations of the lead fluoride calorimeter

Digital Imaging Calorimetry for Precision Electromagnetic and Hadronic Interaction Measurements

Dark Matter Particle Explorer: The First Chinese Cosmic Ray and Hard γ-ray Detector in Space

Simulation and validation of the ATLAS Tile Calorimeter response

Zero degree neutron energy spectra measured by LHCf at s = 13 TeV proton-proton collision

The PAMELA Satellite Experiment: An Observatory in Space for Particles, Antiparticles and Nuclei in the Cosmic Rays

1 Introduction STATUS OF THE GLAST LARGE AREA TELESCOPE. SLAC-PUB December Richard Dubois. Abstract

Last Lecture 1) Silicon tracking detectors 2) Reconstructing track momenta

Cosmic Ray Composition

arxiv: v2 [physics.ins-det] 8 Feb 2013

Dual readout with tiles for calorimetry.

Recent results at the -meson region from the CMD-3 detector at the VEPP-2000 collider

CALET INSTRUMENT. Calorimeter CGBM

Thin Calorimetry for Cosmic-Ray Studies Outside the Earth s Atmosphere. 1 Introduction

Cosmic Ray Physics with the Alpha Magnetic Spectrometer

arxiv: v1 [physics.ins-det] 7 Dec 2018

Preliminary results from gamma-ray observations with the CALorimeteric Electron Telescope (CALET)

P A M E L A Payload for Antimatter / Matter Exploration and Light-nuclei Astrophysics

arxiv: v1 [physics.ins-det] 3 Feb 2014

Indirect Dark Matter search in cosmic rays. F.S. Cafagna, INFN Bari

arxiv: v4 [physics.ins-det] 15 Jan 2018

Validation of EM Part of Geant4

Ultra-High Energy Particles Astronomy with a space-based experiment

(a) Mono-absorber. (b) 4-segmented absorbers. (c) 64-segmented absorbers

Available online at ScienceDirect. Physics Procedia 74 (2015 )

High-energy Gamma Rays detection with the AMS-02 electromagnetic calorimeter. F. Pilo for the AMS-02 ECAL Group INFN Sezione di Pisa, Italy

CALICE Si-W EM Calorimeter: Preliminary Results of the Testbeams 2006

The Fermi Gamma-ray Space Telescope

arxiv: v1 [astro-ph.he] 2 Jul 2009

PERFORMANCE OF THE ATLAS LIQUID ARGON FORWARD CALORIMETER IN BEAM TESTS

Studies of Hadron Calorimeter

CERN Council - Open Session 18 December Lorenzo Bonechi - INFN Firenze, Italy On behalf of the LHCf Collaboration

PANDA Muon System Prototype

Highlights from the ARGO-YBJ Experiment

Conference Report Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

Interaction of Electron and Photons with Matter

A Comparison between Channel Selections in Heavy Ion Reactions

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

The Mini-Calorimeter detector for the AGILE mission

arxiv: v1 [physics.ins-det] 23 Oct 2017

Final project report

Very High Energy Gamma Ray Astronomy and Cosmic Ray Physics with ARGO-YBJ

REVIEW OF CRYSTAL CALORIMETERS

Fig. 11. Signal distributions for 20 GeV * particles. Shown are the measured Éerenkov (a) and scintillation (b) signal distributions as well as the

The Silicon-Tungsten Tracker of the DAMPE Mission

Future prospects for the measurement of direct photons at the LHC

IceCube. francis halzen. why would you want to build a a kilometer scale neutrino detector? IceCube: a cubic kilometer detector

Primary cosmic ray mass composition above 1 PeV as measured by the PRISMA-YBJ array

New Hadroproduction results from the HARP/PS214 experiment at CERN PS

arxiv: v1 [hep-ex] 18 Feb 2009

N U C L E O N. Satellite Mission. Present status.

Upgrade of the CMS Forward Calorimetry

LHCb Calorimetry Impact

Transcription:

Prepared for submission to JINST INSTR17: Instrumentation for Colliding Beam Physics 27 February 3 March, 2017 Novosibirsk, Russia CaloCube: a novel calorimeter for high-energy cosmic rays in space arxiv:1705.07027v2 [physics.ins-det] 23 May 2017 P.W. Cattaneo, a O. Adriani, b,c S. Albergo, d,e L. Auditore, f,g A. Basti, i E. Berti, b,c G. Bigongiari, h,i L. Bonechi, c S. Bonechi, h,i M. Bongi, b,c V. Bonvicini, j S. Bottai, c P. Brogi, h,i G. Carotenuto, k G. Castellini, c,l R. D Alessandro, b,c S. Detti, c M. Fasoli, m,n N. Finetti, c,o A. Italiano, e P. Lenzi, b,c P. Maestro, h,i P.S. Marrocchesi, h,i N. Mori, c M. Olmi, b,c G. Orzan, j L. Pacini, b,c,l P. Papini, c M.G. Pellegriti, e A. Rappoldi, a S. Ricciarini c,l M. Rossella, a A. Sciutto e,r P. Spillantini, b O. Starodubtsev, c F. Stolzi, h, j J.E. Suh, h,i A. Sulaj, h,i A. Tiberio, b,c A. Tricomi, d,e A. Trifirò, f,g M. Trimarchi, f,g E. Vannuccini, c A. Vedda, m,n G. Zampa, j N. Zampa, j a INFN Pavia, Via Bassi 6, I-27100 Pavia, Italy b University of Florence, Department of Physics and Astronomy, via G. Sansone 1, I-50019 Sesto Fiorentino (Firenze), Italy c INFN Firenze, via B. Rossi 1, I-50019 Sesto Fiorentino (Firenze), Italy d University of Catania, Department of Physics and Astronomy, via S. Sofia 64, I-95123 Catania, Italy e INFN Catania, via S. Sofia 64, I-95123 Catania, Italy f University of Messina, Dipartimento di Scienze MIFT, viale F. Stagno d Alcontres, I-98166 Messina, Italy g INFN Messina, c. da Papardo, sal. Sperone 31, I-98166 S. Agata, Messina, Italy h University of Siena, Department of Physical Sciences, Earth and Environment, I-53100 Siena, Italy i INFN Pisa, via F. Buonarroti 2, I-56127 Pisa, Italy j INFN Trieste, via Valerio 2, I-34127 Trieste, Italy k Institute of Polymer, Composites and Biomaterials (CNR), piazzale E. Fermi 1, I-80055 Portici (Napoli), Italy l IFAC (CNR), via Madonna del Piano 10, I-50019 Sesto Fiorentino (Firenze), Italy m University of Milano Bicocca, Department of Materials Science, via Cozzi 55, I-20125 Milano, Italy n INFN Milano Bicocca, piazza della Scienza 3, I-20126 Milano, Italy o University of L Aquila, Dipartimento di Scienze Fisiche e Chimiche, Coppito, I-67100 L Aquila, Italy p University of Trieste, Dipartimento di Fisica, via Valerio 2, I-34127 Trieste, Italy q MATIS (CNR), via S. Sofia 64, I-95123 Catania, Italy r IMM (CNR), Ottava strada 5, I-95121 Catania, Italy E-mail: paolo.cattaneo@pv.infn.it 1Corresponding author.

Abstract: In order to extend the direct observation of high-energy cosmic rays up to the PeV region, highly performing calorimeters with large geometrical acceptance and high energy resolution are required. Within the constraint of the total mass of the apparatus, crucial for a space mission, the calorimeters must be optimized with respect to their geometrical acceptance, granularity and absorption depth. CaloCube is a homogeneous calorimeter with cubic geometry, to maximise the acceptance being sensitive to particles from every direction in space; granularity is obtained by relying on small cubic scintillating crystals as active elements. Different scintillating materials have been studied. The crystal sizes and spacing among them have been optimized with respect to the energy resolution. A prototype, based on CsI(Tl) cubic crystals, has been constructed and tested with particle beams. Some results of tests with different beams at CERN are presented.

Contents 1 Introduction 1 2 The CaloCube concept 1 3 The Monte Carlo simulations and the expected performances 2 4 The prototype 3 5 Beam test and results 3 1 Introduction The scientific rational driving CaloCube1 [1] requirements is the direct measurement of the charges cosmic rays close to the knee, the PeV region where the inclusive spectrum of cosmic rays changes slope becoming steeper and the composition progressively heavier. So far this region has been accessible only with ground based experiments detecting induced air-showers. This indirect measurements have large acceptance but result in large systematic errors due to model dependence for what concerns energy resolution and composition studies. Direct CR detection overcomes those limitations, but suffers from limited exposure, due to constraint in the apparatus mass in space mission. This constraint have limited the measurement of H and He spectra to 100 TeV and heavier nuclei to no more than 1 TeV. The detection of the knees of the H and He spectra requires an acceptance of at least 2.5 m 2 sr 5 yr while the energy resolution is less constrained, being sufficient 40%. A possible detector design satisfying these requirements is a charge measuring device followed by a calorimeter. Additionally this calorimeter can measure the cosmic rays electrons and, if coupled to a trackerconverter system and an anticoincidence shield, the high-energy gamma radiation. This possibility requires an energy resolution better than 2% and a high h/e rejection factor above 1 TeV. 2 The CaloCube concept The goal of CaloCube is to achieve the above discussed performance with a space-borne calorimeter, within the constraint dictated by the limitation in weight ( 2 tons). The proposed solution consists in a 3D array of cubic scintillating crystals, optically isolated from each other, readout by one or more photodiodes (PDs), arranged to form a cube (see Fig. 1). The homogeneous detector geometry provides the possibility to collect particles from five faces out of six (bottom is excluded), so the geometrical acceptance for a fixed mass budget is maximised. The 1CaloCube is an R&D project financed by INFN for 3+1 years (end 2017). 1

cubic crystals, acting as active absorber, provides good energy resolution, while the high granularity provides 3D shower imaging, providing information for leakage correction and h/e separation. Such stringent requirements can be obtained using a calorimeter in conjunction with a de/dx detector, and for this purpose the CaloCube project was created, with the aim of designing and optimising of a calorimeter for measurements of high-energy cosmic rays in space [2, 3]. Figure 1. The CaloCube layout is based on a quasi-isotropic geometry with cubic symmetry, where the active volume is made of NxNxN small cubic scintillator crystals, optically isolated from each other, read from one or more solid state detectors (Si-Photodiode). 3 The Monte Carlo simulations and the expected performances The calorimeter response depends on the adopted geometry and on the material used for the single scintillating crystals. For this reason, an accurate Monte Carlo, based on the FLUKA package, has been developed to carefully study the calorimeter operation in different configurations. A NxNxN cubic geometry, with cubes of about 1 Molière radius size each, has been studied with different scintillating materials, sizes and gaps (distance between the adjacent cubes). Different configurations, comparable in terms of total weight (W tot 2 tons) and active volume fraction (about 78%) are shown in Table 1. Among the five materials, LYSO (Lu 1.8 Y 0.2 SiO 5 (Ce)) is the best CsI(Tl) BaF 2 YAP(Yb) BGO LYSO(Ce) l (cm) 3.60 3.20 2.40 2.30 2.10 Gap (cm) 0.30 0.27 0.20 0.19 0.18 N cubes 20 x 20 x 20 22 x 22 x 22 28 x 28 x 28 27 x 27 x 27 30 x 30 x 30 L (cm) 78.0 76.3 72.8 67.2 68.2 L int (λ int ) 1.80 2.31 3.09 2.72 3.01 L rad (X 0 ) 38.88 34.73 24.96 55.54 53.75 GF (m 2 sr) 9.56 9.15 8.32 7.10 7.35 Table 1. Main parameters of the simulated calorimeter geometries (see Sect. 3 for explanation). one for protons, due to the better shower containment, which compensates for the smaller volume (due to its high density of 7.1 g/cm 3 ). 2

The collection efficiency of the scintillation light depends on the type of coating used to reflect the light and isolate each crystal from the adjacent ones. Several different materials have been tested measuring the signal induced by a 5.5 MeV α emitted by a 241 Am source, and the results (in terms of signal amplitude) are shown in Fig. 2. 4 The prototype A small mechanical structure has been built, to allow the test of different geometric configurations on an accelerator beam. A prototype with 135 CsI(Tl) cubic crystals of 3.6 cm size, arranged in 15 planes of 3 x 3 cubes each, with a gap of 0.4 cm between them, has been built (Fig. 3) and tested on accelerator beams. This prototype has a shower containment of about 1.5 Molière radius (R M ) and an active depth of 28.4 X 0 and 1.35 λ int, and has been tested at CERN with different particle beams, summarised in Table 2. Each crystal is wrapped with few layers of Teflon tape and optically coupled to a single PD, VTH2090H from Excelitas, a large-area ( 100 mm 2 ) sensor that allows to clearly detect minimumionising protons with a signal-to-noise ratio of 15. One of the most challenging requirements is the very large dynamic range (10 7 ) ranging from 20 MeV for minimum ionising protons to 10% of the energy of a PeV proton. This will be accomplished by using also a second small area PD ( 1 mm 2 ). The front-end electronics is based on a custom designed high dynamic-range, low-noise ASIC, characterised by a dynamic range of 52.2 pc and an ENC 0.5 fc at 70 pf input capacitance. 5 Beam test and results Test Beam Energy Feb. 2013 ions from Pb + Be A/Z=2 13-30 GeV Mar. 2015 Ions from Ar + Poly 19-30 GeV Aug./Sep. 2015 µ, π, e 50, 75, 150, 180 GeV Table 2. Parameters of the beam tests. During the beam test at CERN in summer 2015, the prototype was exposed to µ ± beams, to obtain the response of each crystal to a minimum ionising particles (m.i.p.), thus determining the individual conversion factors between the deposited energy and the photodiode signal (Fig. 4). These factors were later used to equalise the response of all the cubes. Then an estimate of the energy resolution has been determined, exposing the calorimeter to e ± beams of different energy, and determining the total deposited energy, given by the sum of the equalised signals of all the cubes. A preliminary result is shown in Fig. 5, referring to a beam of 50 GeV electrons. In this case the measured total energy (expressed in m.i.p) is in good agreement with the expectation, and the corresponding resolution is at level of 1%. In the 2013 beam test the prototype was exposed to ion beams of 12.8 and 30 GeV/n, containing A/Z=2 fragments produced by a primary Pb beam colliding with a Be target. The single-crystal performances were studied, by selecting non-interacting ions. The observed dispersion among crystal responses was about 15%, with an average signal-to-noise ratio for deuterons of 10. The single-crystal responses were equalised by normalising to the energy 3

Figure 2. Signal amplitude due to 5.5 MeV α particles of a single CsI(Tl) scintillator crystal (36 mm) 2 with different coating materials. Figure 3. The prototype with 3 x 3 x 15 CsI(Tl) crystals tested under beam at CERN in 2015. deposit of non-interacting He nuclei (set by definition to 4 MIP), the most abundant fragments. Showers were classified on the basis of the starting point, which in the beam-test set-up univocally determines the shower containment. Fig. 6 shows the energy deposit (left) and its relative fluctuations (right) for He induced showers, versus the shower starting layer; in spite of the significant leakage for showers starting progressively deeper inside the calorimeter, the energy resolution is almost constant and better than 40% down to the fifth layer. A Monte Carlo model of the prototype has been developed in Fluka and its predicted response is shown in Fig. 7 in comparison with real data. A fine tuning of the Monte Carlo was necessary in order to reproduce the beam-test data. In particular, an additional spread of 4.5% on the single-crystal responses and an optical cross-talk of 14% were introduced. Acknowledgments This work was supported by the Istituto Nazionale di Fisica Nucleare (INFN), through the CaloCube project, and by the H2020 project AIDA-2020, GA no. 654168. The authors thank CERN for the allocation of beam time at the North Area test facility. References [1] R. D Alessandro et al., Calocube: A highly segmented calorimeter for a space based experiment, Nucl. Instrum. Meth. A824 (2016) 609 613. [2] M. Bongi et al., CALOCUBE: an approach to high-granularity and homogenous calorimetry for space based detectors, J. Phys. Conf. Ser. 587 (2015) 012029. 4

Counts sigma ~ 1% Total signal (MIPs) Figure 4. Response of a single crystal to a noninteracting track at 150 GeV (mainly µ). The distribution is well fit with a Landau distribution (red). Figure 5. Measured distribution of total energy (expressed in m.i.p.) released with a 50 GeV electrons beam, fit with with the expected distribution (red). The resulting energy resolution is about 1%. Figure 6. Energy deposit (left), in MIP, and energy resolution (right) measured for He nuclei, as a function of the layer where the shower starts. Figure 7. Distribution of the energy deposit (in MIP) of 30 GeV/n He ions for data (blue) and MC (red). [3] E. Vannuccini et al., CaloCube: A new-concept calorimeter for the detection of high-energy cosmic rays in space, Nucl. Instrum. Meth. A845 (2017) 421 424. 5