Data report for elemental analysis of IMPROVE samples collected during April, May, June 2006 UC Davis - Submitted May 14, 2008 SUMMARY

Similar documents
Data report for elemental analysis of IMPROVE samples collected during OCTOBER, NOVEMBER, DECEMBER 2005 UC Davis SUMMARY

Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 18, 2010 SUMMARY

Quantitative XRF Analysis. algorithms and their practical use

Overview of X-Ray Fluorescence Analysis

FINDING A NEEDLE IN A HAYSTACK: PERFORMANCE EVALUATION OF PORTABLE XRF INSTRUMENTS FROM THREE MANUFACTURERS

ARTICLE IN PRESS. Journal of Aerosol Science

LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong

DOCUMENT HISTORY. Initials Section/s Modified Brief Description of Modifications

X-Ray Fluorescence and Natural History

REGIONAL TRAINING COURSE ON METHODS AND TOOLS TO IDENTIFY SOURCES OF AIR POLLUTION

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Unit 1 Part 2 Atomic Structure and The Periodic Table Introduction to the Periodic Table UNIT 1 ATOMIC STRUCTURE AND THE PERIODIC TABLE

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

Name: SCH3U Worksheet-Trends

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Operation of the Coupled Cyclotron Facility at Michigan State University

XUV 773: X-Ray Fluorescence Analysis of Gemstones

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

AMMS-100 Atmospheric Metal Monitoring System

Inspection Documents for Type 2.1 Coverage:

Inspection Documents for Type 2.1 Coverage:

Technical note on seasonal adjustment for M0

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

Microsoft Excel Directions

Standardless Analysis by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited

Reporting Category 1: Matter and Energy

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

Page 1 of 9. Website: Mobile:

Calibration of the IXPE Instrument

Spectroscopy on Mars!

Chiang Rai Province CC Threat overview AAS1109 Mekong ARCC

Give the number of protons, neutrons and electrons in this atom of aluminium. Why is aluminium positioned in Group 3 of the periodic table? ...

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important ones are...

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Reporting Category 1: Matter and Energy

CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE

Chemistry 31A Autumn 2004 Professors Chidsey & Zare Exam 2 Name:

OXEA - Online Elemental Analyzer

Horizontal rows are called periods. There are seven periods on the periodic chart.

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

ATLAS Tile Calorimeter Calibration and Monitoring Systems

Enhancements to the CTBTO IDC radionuclide processing pipeline for particulate samples achieving significant improvement of automatic products

Cross-section Measurements of Relativistic Deuteron Reactions on Copper by Activation Method

Introduction to LIBS COMMUNITY USER WORKSHOP ON PLANETARY LIBS (CHEMCAM) DATA. Sam Clegg and the ChemCam team

1 Arranging the Elements

Inspection Documents for Type 2.1 Coverage:

Inspection Documents for Type 2.1 Coverage:

SFT3000S Measurement of Sn-Cu Coating

Three hour lab. Chem : Sept Experiment 2 Session 2. Preparation Pre-lab prep and reading for E2, Parts 3-5

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

GAMINGRE 8/1/ of 7

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

CASSY Lab. Manual ( )

Inspection Documents for Type 2.1 Coverage:

OES - Optical Emission Spectrometer 2000

Latest advances in identifying mineral composition variation by the M4 TORNADO AMICS

Portable type TXRF analyzer: Ourstex 200TX

PIXE Protocol Setup for Cluster Analysis of Archeological Samples

Mendeleev arranged the elements in order of their atomic mass (atomic weight).

Course theme. Three hours of lab Complete E1 (Parts 1, 2, 3, 4, and 5B) Prepare discussion presentation Prepare team report.

Three hours of lab Complete E1 (Parts 1, 2, 3, 4, and 5B) Prepare discussion presentation Prepare team report. Course theme

S8 TIGER Series 2 for ASTM D 6443

The ATLAS Liquid Argon Calorimeter: Construction, Integration, Commissioning Ph. Schwemling on behalf of the ATLAS LAr Group

Inspection Documents for Type 2.1 Coverage:

This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract

X-ray Absorption Spectroscopy

Pulse Height Analysis System (PHA) designed for W7-X Presented by Monika KUBKOWSKA

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

TRACE ELEMENT ANALYSIS OF DIAMOND BY LAM ICPMS: STANDARDISATION, RESULTS AND DIRECTIONS

Honors Chemistry - Unit 4 Bonding Part I

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

SUPPORTING INFORMATION

Determination of the activity of radionuclides

How many grams of sodium metal is required to completely react with 2545 grams of chlorine gas?

ENGINE SERIAL NUMBERS

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Wind Resource Data Summary Cotal Area, Guam Data Summary and Transmittal for December 2011

Angelika Dehn Rob Koopman 10 Years GOME on ERS-2 Workshop

DRAFT BRUKER XRF SPECTROSCOPY USER GUIDE: SPECTRAL INTERPRETATION AND SOURCES OF INTERFERENCE

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals.

MODERN ATOMIC THEORY AND THE PERIODIC TABLE

Chemistry 1 First Lecture Exam Fall Abbasi Khajo Levine Mathias Mathias/Ortiz Metlitsky Rahi Sanchez-Delgado Vasserman

Chapter 1. I- Fill the following table. Element symbol and the mass no. n p n n n e. number. II - Choose the correct answer for the following: Ca-40

Three hour lab. Chem : Feb Experiment 2 Session 2. Experiment 2 Session 2 Electrons and Solution Color

ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom.

Nuclear Magnetic Resonance (NMR)

4.1 Atomic structure and the periodic table. GCSE Chemistry

Creating Empirical Calibrations

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

Transcription:

Data report for elemental analysis of IMPROVE samples collected during April, May, June 2006 UC Davis - Submitted May 4, 2008 SUMMARY This report summarizes the quality assurance performed during elemental analysis of the IMPROVE samples collected in April, May and June of 2006. The elemental analyses include the determination of most elements with atomic numbers from to 26 (Na-Fe) by energy dispersive X-ray fluorescence (XRF) with a Cu-anode system, most elements from 27 to 40 (Ni-Zr) and 82 (Pb) by XRF with a Moanode system, and hydrogen by Proton Elastic Scattering Analysis (PESA) with the Crocker cyclotron. The following data assessments and quality controls are obtained for all analyses: Concentration calibration and verification (calibration check) Energy calibration Laboratory replicates (reanalysis) Systems comparison Field blanks Analysis details and key events are summarized below. Apr and Jun 2006 samples were analyzed on the second Cu-anode system Cu-Vac2, and May 2006 samples were analyzed on the first Cu-anode system Cu-Vac. A new calibration was performed on Cu-Vac2 on 4/20/07 (June 2006 samples) to reflect work done on the system. No new calibrations were performed on the Cu-Vac system. The Mo system was recalibrated to reflect maintenance (alignment and securing of the detector collimator, and window cleaning) on 0/5/06 (April 2006 samples). During the analysis of Apr-Jun 2006 samples all routine calibration checks on the Mo system were within criteria. A new set of standards (Mylar foils) was introduced for PESA calibration and calibration checks on April 26, 2007. Section. Overview of Elemental Analysis Systems The elements Na and Mg (considered qualitative only) and Al to Fe are reported from two XRF systems with a Cu-anode grounded X-ray tube, Cu-Vac and Cu-Vac2. Both systems operate under vacuum. Default settings for (20 kv, 0 ma for 000sec/sample) were used for each system. The elements Ni to Zr and Pb are reported from a similar system with a Mo-anode grounded X-ray tube operating in air. Samples were analyzed for 000 seconds at 23 ma and 35 kv (default settings for ). The PESA system operates under vacuum and uses a proton beam (4.5 MeV H+) from the Crocker cyclotron to quantify the concentration of hydrogen (H). Samples were analyzed for 5 seconds, with an average current value of approximately 50 na collected on a Faraday cup.

Section 2. General Statistics of April, May and June 2006 data XRF and PESA analyses were carried out on 649 samples collected in April 2006, 685 samples collected in May 2006, and 609 samples collected in June 2006. All samples were analyzed between 4 September 2006 and 4 December 2006 on the Mo-anode XRF system, between 8 March 2007 and 7 April 2007 on Cu-Vac, between 8 March 2007 and 3 May 2007 on Cu-Vac2, and on 3/23/07, 4/26-27/07, 5/-2/07, and 5/2-22/07 on the PESA system. Table summarizes the second quarter 2006 detection rates on the three systems, with rates for March 2006 included for comparison. PESA Z element 4-2006 5-2006 6-2006 3-2006 H 00% 97% 00% 00% Cu-anode XRF Z element 4-2006 5-2006 6-2006 3-2006 Na 60% 38% 63% 4% 2 Mg 77% 23% 7% 9% 3 Al 97% 82% 93% 72% 4 Si 00% 96% 00% 99% 5 P 2% 0% 3% 0% 6 S 00% 98% 00% 00% 7 Cl 20% 9% 5% 5% 9 K 00% 98% 00% 00% 20 Ca 00% 98% 00% 00% 22 Ti 00% 97% 99% 98% 23 V 98% 82% 97% 84% 24 Cr 90% 32% 82% 37% 25 Mn 00% 98% 00% 98% 26 Fe 00% 98% 00% 00% Mo-anode XRF Z element 4-2006 5-2006 6-2006 3-2006 28 Ni 57% 53% 58% 43% 29 Cu 89% 95% 96% 93% 30 Zn 00% 00% 00% 00% 33 As 70% 74% 58% 50% 34 Se 8% 9% 9% 82% 35 Br 00% 99% 00% 00% 37 Rb 86% 82% 80% 82% 38 Sr 96% 97% 97% 95% 40 Zr 30% 6% 20% 25% 82 Pb 94% 97% 00% 97% Table. Percentage of cases in which the element was detected on each system. March 2006 data included for reference.

Section 3. Quality Control 3. Concentration calibration and verification (calibration checks) Both XRF systems are calibrated with thin film foil standards produced by Micromatter. The standards used for samples from the second quarter of 2006 are listed below in Table 2. Because their concentrations are relatively high, standards are analyzed at reduced X-ray tube current (2.6 ma on XRF-Cu systems and 0mA on XRF-Mo system) to maintain counting live times comparable with those of actual IMPROVE samples. Standard Certified Elemental Concentrations Serial # +/- 5% (µg/cm 2 ) NaCl Na: 9., Cl: 29.4 658 MgF 2 Mg: 20.6 659 Al Al: 40.7 6520 SiO Si: 23.9 652 GaP* P: 4.5 6500 CuS x S: 2.9 Cu: 37.6 6523 KCl Cl: 22.5 K: 24.9 6296 CaF 2 Ca: 24.9 6525 Ti Ti: 3.7 6504 V V: 2.2 6505 Cr Cr: 5.8 6507 Mn Mn: 4.6 6506 Fe Fe: 4.7 6508 Ni Ni: 0.5 6509 Cu Cu: 2.4 650 ZnTe* Zn: 5.2 65 GaAs* Ga: 8 As: 8.7 652 Se Se: 2.9 653 CsBr Br: 5. 654 RbI Rb: 5.7 655 SrF 2 Sr: 0.9 656 Pb Pb La: 6 Pb Lb: 6 657 Table 2. Micromatter standard foils used for all analyses..some standards (*) have variable stoichiometry; they are not use directly in calibration of the systems but serve only as indicators. Spectra from the foil standards are processed and analyzed by the same software used for samples. Calibration factors relating spectral counts to elemental concentrations are determined from the ratio of an element s observed peak area to the concentration quoted by Micromatter. During the analyses of Apr-Jun 2006 samples, the performance of all systems was monitored weekly. Fifteen standards are analyzed on the Cu system and 9 on the Mo system, and the ratios of reported to quoted values are calculated. If the ratios lie within the acceptance limits 0.9. for all quantitative elements, then the system is considered stable and the existing calibration factors continue to be used. Deviations beyond 0% trigger an investigation of the problem and possible system recalibration. After a recalibration, all samples analyzed since the last successful calibration verification are reanalyzed with the new calibration factors. Figures to 3 below show the calibration checks and system recalibrations performed during the period in which samples of interest were analyzed on each system. The analysis dates for each sample month are listed in the legends. March 2006 dates are included for reference. The y-axes indicate the ratio of the values reported for each standard to the value quoted by Micromatter.

Ga Cu S K Cl Ca Ti V Cr Mn Fe Fe Ni Cu Zn Ga As Se Br Rb Sr Pb La Pb Lb.5 Ratio Found/Quoted Standard..05 0.95 Mar 2006 8/9/06-9/3/06 Apr 2006 9/4/06-0/6/06 May 2006 0/7/06- /0/06 Jun 2006 /0/06-2/4/06 0.9 0.85.5 Na Cl Mg Al Si P S Cl K Ca Ti V Cr Mn Fe Ni new cal new cal new cal new cal new cal new cal new cal new cal..05 0.95 0.9 0.85 2//07 3//07 4//07 5//07 8//2006 9//2006 0//2006 //2006 2//2006 //2007 Figure. Mo XRF system performance chart based on standards Ratio Found/Quoted Standard Mar 2006 2/2/07 3/8/07 May 2006 3/8/07-4/7/07 Figure 2. XRF Cu-Vac system performance chart based on standards.

.5 Na Cl Mg Al Si P S Cl K Ca Ti V Cr Mn Fe Ni new calib new calib new calib Ratio Fo un d/quoted Standard..05 0.95 Jan 2006 2//07-3/2/07 Apr 2006 3/8/07-4/3/07 Jun 2006 4/20/07-5/3/07 0.9 0.85 3//07 4//07 5//07 6//07 Figure 3. XRF Cu-Vac2 system performance chart based on standards. A descending trend of standards ratios was observed on the XRF-Mo system between 8/6/06 and 9/26/06. Even though the standards ratios did not exceed acceptance limits, the system was recalibrated on October 5, 2006 to reflect maintenance done on the system, including cleaning of the detector window and collimator adjustment. All calibration checks on the Mo system during the analysis of Apr-Jun 2006 samples were within our criteria. May 2006 samples were analyzed on the XRF Cu-Vac system. A failing socket adaptor for the X- ray tube was replaced toward the end of this analysis run. A descending trend of standards ratios was observed during the time of analysis, but no recalibration was performed as all ratios met the acceptance criteria. Unusually low detection rates were obtained for several elements (e.g. Cr) as was also the case with March 2006 samples, the previous group analyzed on Cu-Vac. The entire March and May 2006 sample collection was reanalyzed on the CuVac2 system to assure the quality of these data. The reanalyses yielded no grounds to invalidate the original results. April and June 2006 samples were analyzed on XRF Cu-Vac2 system. New calibration factors on April 24 th, 2007 were the result of the repairs on the system. Descending trends were noted in ratios, especially for Fe. During the analysis, all calibration checks for quantitative elements met criteria. On April 26, 2007, six new Mylar foils (thicknesses 2.5, 3.5 and 9.3 microns) were combined with two previously used foils (/8 mil thick) to form a new set of references for PESA system calibrations and calibration checks. The hydrogen concentration for each of these standards was determined based on their weights and chemical composition of the Mylar. The calculated values for each PESA standard are listed below: PESA Standard 2 3 4 5 6 7 8 Calculated H amount (ug/cm2) 9.80 9.80 36.73 36.73 23.78 23.78 7.59 7.59 As with XRF, the calibration factor is based on the average ratio of observed counts for the eight PESA standards to their calculated H concentration. The PESA system is recalibrated at the beginning of every analytical session, because of variations in the ion source production, amplitude harmonics, and optics. The eight Mylar blanks used as

calibration standards are reanalyzed approximately every 00-200 samples to verify the calibration throughout the session. If the ratio of reported to calculated concentrations for these standards drifts outside the 0.95-.05 range during an analysis run., the cyclotron is re-tuned, the system is recalibrated, and the samples are reanalyzed. Figure 4 shows calibration verifications during the analysis of Apr-Jun 2006 samples. During analysis of Apr 2006 and May 2006 samples (circled points) the average ratio of all standards had fallen below acceptable criteria twice. The analyses were continued and the affected samples were reanalyzed during a subsequent PESA analysis run on July 25 th, 2007. New Mylar PESA standards April, May,June 2006 samples.2.5 Ratio Found/Standard..05 0.95 0.9 Apr 06 May 06 Jun 06 cyclotron retuned 0.85 0.8 36000 37000 38000 39000 40000 4000 42000 43000 44000 45000 Run Number Figure 4. PESA standards for Apr-Jun 2006 samples 3.2 X-ray energy calibration In addition to the peak counts associated with a known concentration (concentration calibration), the energy channel associated with a known fluorescence line must also be determined; this is the energy calibration. Energy calibrations were performed for the analyses of each sample month on each system to establish relationships of the form energy = intercept + slope * channel The following energy calibration equations (in energy units of KeV) were used for the analysis: XRF-Cu change XRF-Cu2 change XRF-Mo change intercept slope full scale from Jun05 intercept slope full scale from Oct05 intercept slope full scale from Jun05 Apr-06-0.0358 0.076 8.750 0.00% -0.08626 0.036033 8.362-0.03% May-06-0.0377 0.0667 8.497 0.08% -0.08789 0.036027 8.358-0.05% Jun-06-0.0358 0.076 8.750 0.00% -0.07943 0.036008 8.357-0.06% 3.3 Reanalysis The reproducibility of XRF and PESA data is tracked over time by reanalyzing selected sample filters. Different reanalysis protocols are used for the XRF and PESA reanalyses, reflecting the different impacts of their exciting beams on the Teflon filter substrate, as explained in previous reports.

Filters to be reanalyzed by PESA are selected from the previous quarter s X-module (collocated A- module) samples. During the analysis of Apr 2006 samples, 24 MEVEX, OLYMX, PMRFX filters from Sep 2005 were reanalyzed four times. For May 2006 samples, 29 MEVEX, OLYMX, PMRFX filters from Nov 2005 were reanalyzed three times. Similarly, reanalyses for June 2006 samples were performed on 28 SAFOX, SAMAX AND TRCRX filters from Dec 2005 three times. Figure 5 compares the original and repeat analyses. 0000 0000 reanalyzed with Apr 2006 samples reanalyzed with May 2006 samples 000 000 000 0000 original MEVEX, OLYMX, PMRFX Sep 2005 (ng/cm²) 00 00 000 0000 original MEVEX, OLYMX, PMRFX Nov 2005 (ng/cm²) reanalyzed with Jun 2006 samples 0000 000 Figure 5. PESA reanalysis of selected Sep, Nov and Dec 2005 samples during analyses of Apr-Jun 2006 network samples. Reported uncertainties are indicated by error bars; agreement is indicated by sloping lines. 00 00 000 0000 original SAFOX, SAMAX, TRCRX Dec 2005 (ng/cm²)

XRF reanalyses were performed on the same fixed collection of samples employed in previous reports. The collection, denoted trays REANAL and REANAL2, was reanalyzed with the XRF-Mo and Cu- Vac2 systems approximately monthly during analyses of Apr-Jun 2006 samples. The trays were not reanalyzed on the Cu-Vac system during analysis of May 2006 samples, so the figures 8 and 9 show the available reanalysis results before and after period of interest. All results are summarized in the figures 6- below. For the Mo and Cu-Vac systems the mean loadings calculated based on 2 consecutive runs (about a year of data) are used as a benchmark for comparison. As in previous reports, the original run (Nov06) on Cu-Vac2 serves as a baseline for the newer system. The average ratio of observed deviations (from mean or original) to reported measured uncertainties for each element is calculated and shown on vertical axis. This format highlights any systematic trend in the measurements and provides a test, at actual sample loadings, of the stability of calibrations based on the heavily-loaded foil standards. 5 Reanalysis R on Mo Average Relative Deviation (Observed/Expected) 4 3 2 0 - -2-3 -4 S K CA TI V CR MN FE NI CU ZN AS PB SE BR RB SR Mar 2006 8/9/06-9/3/06 Apr 2006 9/4/06-0/6/06 May 2006 0/7/06- /0/06 Jun 2006 /0/06-2/4/06-5 aug6a aug6b oct6a oct6b nov6a dec6a Reanalysis Date Figure 6. Reanalyses of REANAL samples on XRF-Mo system. Horizontal arrow indicates when Apr-Jun 2006 network samples were analyzed. 5 Reanalysis R2 on Mo Average Relative Deviation (Observed/Expected) 4 3 2 0 - -2-3 -4 S K CA TI V CR MN FE NI CU ZN AS PB SE BR RB SR Mar 2006 8/9/06-9/3/06 Apr 2006 9/4/06-0/6/06 May 2006 0/7/06- /0/06 Jun 2006 /0/06-2/4/06-5 aug6a aug6b oct6a oct6b nov6a dec6a Reanalysis Date Figure 7. Reanalyses of REANAL2 samples on XRF-Mo system. Horizontal arrow indicates when Apr-Jun 2006 network samples were analyzed.

5 4 REANAL on XRF-CuVac YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Average Relative Deviation (Observed/Expected) 3 2 0 - -2-3 -4 Mar 2006 2/2/07 3/8/07 May 2006 3/8/07-4/7/07-5 Jan7a Feb7a May7a Reanalysis Date Figure 8. Reanalyses of REANAL samples on XRF-Cu system. Average Relative Deviation (Observed/Expected) 5 4 3 2 0 - -2-3 -4-5 REANAL2 on XRF-CuVac YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Jan7a Feb7a May7a Reanalysis Date Mar 2006 2/2/07 3/8/07 May 2006 3/8/07-4/7/07 Figure 9. Reanalyses of REANAL2 samples on XRF-Cu system.

Average Relative Deviation (Observed/Expected) 5 4 3 2 0 - -2-3 -4-5 REANAL on XRF-CuVac2 YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Mar7a Mar7b Apr7a Apr7b May7b Reanalysis Date Jan 2006 2//07-3/2/07 Apr 2006 3/8/07-4/3/07 Jun 2006 4/20/07-5/3/07 Figure 0. Reanalyses of REANAL samples on XRF-Cu2 system. Horizontal arrow indicates when Apr and Jun 2006 network samples were analyzed. Average Relative Deviation (Observed/Expected) 5 4 3 2 0 - -2-3 -4-5 REANAL2 on XRF-CuVac2 YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Mar7a Mar7b Apr7a May7b Reanalysis Date Jan 2006 2//07-3/2/07 Apr 2006 3/8/07-4/3/07 Jun 2006 4/20/07-5/3/07 Figure. Reanalyses of REANAL2 samples on XRF-Cu2 system. Horizontal arrow indicates when Apr and Jun 2006 network samples were analyzed. 3.4 Systems comparison Additional comparison between selected elements measured independently by the Cu and Mo systems is performed for each data set. The elements Calcium and Iron are reported from the Cu system (Cu- Vac or Cu-Vac2) but are also quantified by the Mo system. Figures 2 and 3 compare the two measurements of these two elements for the samples from Apr-Jun 2006. Reported uncertainties are

shown as bars for each sample, and reported MDL s are indicated by green and pink points for both systems. The increase in analytical uncertainty closer to the MDL s can be observed for all cases. April 2006 May 2006 00000 00000 0000 0000 FE from Mo (ng/cm²) 000 00 0 FE from Mo (ng/cm²) 000 00 0 0. 0. 0 00 000 0000 00000 FE from Cu2 (ng/cm²) 0. 0. 0 00 000 0000 00000 FE from Cu (ng/cm²) June 2006 00000 0000 Figure 2. Comparison of Iron data obtained independently from Cu (x-axis) and Mo (y-axis) systems. Cu refers to Cu-Vac and Cu2 to Cu-Vac2. FE from Mo (ng/cm²) 000 00 0 0. 0. 0 00 000 0000 00000 FE from Cu2 (ng/cm²)

April 2006 May 2006 00000 00000 0000 0000 CA from Mo (ng/cm²) 000 00 0 CA from Mo (ng/cm²) 000 00 0 0. 0. 0 00 000 0000 00000 CA from Cu2 (ng/cm²) 0. 0. 0 00 000 0000 00000 CA from Cu (ng/cm²) 00000 0000 June 2006 Figure 3. Comparison of Calcium data obtained independently from Cu (x-axis) and Mo (y-axis) systems. CA from Mo (ng/cm²) 000 00 0 0. 0. 0 00 000 0000 00000 CA from Cu2 (ng/cm²) Calcium and iron determinations by the Mo system contain more uncertainty than those from the Cu systems, and are accordingly not used to report concentrations. Their value in these system comparisons is the additional qualitative check they provide on both systems performance. 3.5 Field blanks Twenty field blanks for April 2006, 2 field blanks for May 2006 and 9 field blanks for June 2006 samples were exposed at selected sites on selected sampling events. The field blanks were analyzed on both, XRF-Mo and Cu-Vac or Cu-Vac2 systems. The Cu-anode system used for analysis of the field blanks was the same system used for the corresponding samples. As in previous reports, 95 th, 90 th and 75 th percentile field blank loadings are shown for each system in the tables below. They are given as percentiles of well measured network sample loadings during April, May and June 2006. Loadings are considered well measured when their uncertainties are less than 0%. Thus, the 95 th percentile field blank loading for Fe (Apr 2006 samples) was at or above about 0.5% of all well measured sample loadings on the Cu system (0.2% on the Mo system).

Cu anode Apr 2006 95 %ile 90 %ile 75 %ile Na Mg Al Si P S Cl K Ca Ti V Cr Mn Fe 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.% 0.3% 0.2% 0.0% 0.0% 0.0% 0.5% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% The field blanks data from Cu system for May and June 2006 samples are not yet available. Mo anode Apr 2006 Fe Ni Cu Zn As Pb Se Br Rb Sr Zr 95 %ile 0.2% 0.0% 0.0% 28.8% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 90 %ile 0.2% 0.0% 0.0% 28.% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 75 %ile 0.0% 0.0% 0.0% 5.2% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% May 2006 Fe Ni Cu Zn As Pb Se Br Rb Sr Zr 95 %ile.4% 0.0% 0.0% 48.5% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 90 %ile.0% 0.0% 0.0% 29.3% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 75 %ile 0.0% 0.0% 0.0%.5% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% Jun 2006 Fe Ni Cu Zn As Pb Se Br Rb Sr Zr 95 %ile.4% 0.0% 0.0% 43.5% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 90 %ile 0.7% 0.0% 0.0% 33.7% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 75 %ile 0.2% 0.0% 0.0%.7% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% 0.0% Field blank loadings are negligible compared to samples for all elements except Zn, as noted in previous reports.