Evaluation of round robin test results for a size exclusion chromatography method for the analysis of nitrocellulose

Similar documents
Whole Life Assessment of Nitrocellulose in Double Base Propellants. Nathalie Mai Michael Isherwood Phil Gill

Progress toward reliable NC molecular mass distribution by GPC

Investigation of ageing behaviour of plasticised nitrocellulose accelerated ageing and real ageing

Gel Permeation Chromatography Basics and Beyond eseminar March 13, Jean Lane Technical and Applications Support LSCA, Columns and Supplies

Gel Permeation Chromatography

Differentiation of polymer branching and composition using the Mark Houwink plot

GPC - Gel Permeation Chromatography. aka Size Exclusion Chromatography- SEC

Comparison of Polymer Separation by Size Exclusion Chromatography and Asymmetric Flow Field Flow Fractionation

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

An Introductions to Advanced GPC Solutions

Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012

How to use GPC/SEC for compositional analysis

Gel Permeation Chromatography

Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Optimizing GPC Separations

Practical Steps in GPC Method Development

Tips & Tricks GPC/SEC: From a Chromatogram to the Molar Mass Distribution

Characterization of polyphenylene sulphide using the Agilent PL-GPC 220 High Temperature GPC System with triple detection

Determination of Polymer Modifier in Asphalt

How switching mobile phases can improve your OMNISEC results

GPC/SEC An essential tool for polymer analysis

Maximizing Performance Through GPC Column Selection

GPC/SEC standards. Product guide

Analysis of Star Polymers Using the Agilent 1260 Infinity Multi-Detector GPC/SEC System

Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How?

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection

GPC/SEC standards. Product guide

CALIBRATING GPC COLUMNS

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

Tips & Tricks GPC/SEC: Inter-Detector Delay

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

Amphiphilic diselenide-containing supramolecular polymers

Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System

GPC / SEC Theory and Understanding

Lecture 4 : Gel Permeation or Size Exclusion Chromatography

Latest Developments in GPC Analysis of Adhesive and Sealant Polymers Mark Pothecary PhD Americas Product Manager Malvern Instruments

Advanced GPC. GPC On Tour, Barcelona, 28 th February The use of Advanced Detectors in GPC

Supporting Information. Nitroxide Mediated Polymerization of methacrylates at moderate temperature

Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. Molecular weight of polymers. H i

One-pot polymer brush synthesis via simultaneous isocyanate coupling chemistry and grafting from RAFT polymerization

Supporting Information

MEASURING PROTEIN AGGREGATION WITH THE VISCOTEK SEC-MALS 20

IR-LC Deposition and Detection System. Polymer Deformulation and Additive Analysis by a Single GPC-IR Run

Macromolecular Chemistry

SUPPORTING INFORMATION

Xiangxiong Chen, Mohd Yusuf Khan and Seok Kyun Noh* School of Chemical Engineering, Yeungnam University, Dae-dong, Gyeongsan,

Controlling Multicompartment Morphologies Using Solvent Conditions and Chemical Modification

Supporting Information

High Performance Liquid Chromatography

The analysis of organic acid content of additives, premix, feed, and water.

Size Exclusion Chromatography: Method Development

The power of multi-detector SEC in the analysis of antibodies

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Molecular Weight Distribution of Living Chains in Polystyrene Pre-pared by Atom Transfer Radical Polymerization

Optimization of PLA/PLGA molecular weight and structure to suit final application requirements

PHASE 1: IUPAC SEC/GPC Round Robin Project Report. Title: Repeatability and Reproducibility of Sample Preparation and Analysis in High-Temperature SEC

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

EXCLUSION HIGH PERFORMANCE LIQUID CHROMATOGRAPHY ANALYSIS OF A CELLULOSE ACETATE-AMMONIUM NITRATE GAS GENERATOR PROPELLANT

Characterisation of Viscosity and Molecular Weight of Fractionated NR

Analysis of polyolefins by GPC/SEC

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

Benefit of light scattering technologies (RALS/LALS/MALS) and multidetection characterization in life science research?

Supporting Information

Novel Supercapacitor Materials Including OLED emitters

COURSE MATERIAL: Unit 3 (Part 1) Polymer Science LT8501 (Click the link Detail to download)

Supplementary Material (ESI) for Chemical Communications This journal is (c) The Royal Society of Chemistry 2008

Supplementary Information

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized by Conventional GPC/SEC Techniques

Supporting Information

Polymer analysis by GPC-SEC. Technical Note. Introduction

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

Supporting Information

GPC/SEC Analysis of Polymer Degradation

Phenogel. GPC/SEC Columns. Sample Elution. Technical Specifications 10 3 Å 10 6 Å

Using High Speed/High Resolution Size Exclusion Chromatography Separation of Polymeric Materials with Light Scattering Detection

Methods for the Determination of Nitrogen Content in Nitrocellulose. Beat Vogelsanger, Marc Müller, Beate Pausch, and Michael Ramin

Supporting Information

Supplementary Information

METHANOLYSIS OF ACETAL

Response to Questions Raised by Tom Phillips Concerning the Enforcement Analyitical Method for Avail and Nutrisphere Products. Verdesain Life Sciences

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Egualen Sodium Granules

Bulk ring-opening transesterification polymerization of the renewable δ-decalactone using

PHENOGEL GPC/SEC COLUMNS

(Published in Journal of Liquid Chromatography 2(3): (1979) GEL PERMEATION CHROMATOGRAPHY OF POLYMERS FROM WOOD

Supporting Information. segments

CHEMISTRY DEPARTMENT

GC-CI-MS analysis of TMS derivatives

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Synthesis of Peptide-Grafted Comb Polypeptides via Polymerisation of NCA-Peptides

Cationic Alkylaluminum-Complexed Zirconocene Hydrides as Participants in Olefin-Polymerization Catalysis. Supporting Information

Well-defined polyethylene-based random, block and bilayered molecular cobrushes

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Free radical and RAFT polymerization of vinyl

Efficient Magnesium Catalysts for the Copolymerization of Epoxides and CO 2 ; Using Water to Synthesize Polycarbonate Polyols

Ring-Opening Polymerization of N-Carboxyanhydrides Initiated by a Hydroxyl Group

Single-ion BAB triblock copolymers as highly efficient electrolytes for lithium-metal batteries

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Supplementary Information

Preparation of 1:1 alternating, nucleobase-containing copolymers for use in sequence-controlled polymerization

Transcription:

Evaluation of round robin test results for a size exclusion chromatography method for the analysis of nitrocellulose Alan Macdonald alan.f.macdonald@awe.co.uk www.awe.co.uk AWE/CRD11/AM10/0020

Background Propellants contain high nitrocellulose levels and life is traditionally assessed by stabiliser concentration. Mechanical integrity is important A correlation has been shown between mechanical properties of a nitrocellulose containing PBX and the molar mass distribution. Size Exclusion Chromatography (SEC) is a technique that can investigate the molar mass distribution of polymers but has been considered by some as a random number generator. A method for the molecular mass determination of nitrocellulose using SEC was proposed for inclusion in STANAG 4178. A round-robin inter-laboratory study was undertaken to assess variability of the method AWE/CRD11/AM10/0020 2

2008 / 2009 Method proposed for STANAG4178 and round-robin evaluation undertaken Method development and validation DCMT, Shrivenham Dec 2007 Method proposed for discussion Method development & data interpretation Fraunhofer ICT, Pfinztal Method development and validation for whole life assessment of conventional munitions (propellants) Nitrocellulose molecular mass determination by SEC has been refined for NC in PBX formulations at AWE AWE/CRD11/AM10/0020 3

Participants Organisation Nation Detectors AWE Plc UK RI ATK Radford USA RI Cranfield University UK RI Denel Land Systems South Africa RI Fraunhofer ICT Germany RI General Dynamics Canada UV/Vis Rheinmetall Nitrochemie Wimmis Switzerland UV/Vis PVTT Finland UV/Vis TNO Defence Netherlands UV/Vis AWE/CRD11/AM10/0020 4

Proposed Method M n (Dalton) (± 1 s.d.) M w (Dalton) (± 1 s.d.) Sample [BHT] drying by an existing STANAG Method Samples prepared at 1.5mgcm -3 in THF stabilised with <250ppm BHT [250ppm] 346200 ± 9050 (2.6%) 669000 ± 10500 (1.6%) Dissolution time: 7 days (Not ideal for production ± 0.048 labs) Polystyrene calibration standards across the whole nitrocellulose 335100 distribution. Broad 653500 polystyrene [100ppm] instrument performance ± 6900 check ± 11500 standard (2.1%) (1.8%) Mobile phase flow rate: 1cm 3 min -1 Triplicate sample injection per vial of 100µl [0ppm] 329400 ± 3900 639900 ± 8500 Polydispersity (± 1 s.d.) 1.933 1.950 ± 0.034 1.943 Columns: Two 10µ, 30 x 7.5cm mixed porosity ± 0.030 (1.2%) (1.3%) polystyrene divinyl benzene maintained at 35 C AWE/CRD11/AM10/0020 5

Round Robin Evaluation All Participants received: Two columns manufactured from the same batch of Polystyrene divinyl benzene Polystyrene narrow standards of the same batch Polystyrene NBS706 broad standard Four nitrocellulose samples (water wet) of differing nitrogen content (11.69% to 13.55%) All participants asked to report: Drying regime used and sample concentrations Calculated molecular weight distributions for the NBS706 and the nitrocellulose samples AWE/CRD11/AM10/0020 6

Size Exclusion chromatography M w = N M i i i N M i 2 Nitrocellulose - typical of naturally occurring polymers broad distribution tailing toward lower molecular weight M z N M 3 i i = 2 NiMi M n = N M i N i i t r AWE/CRD11/AM10/0020 7

Polystyrene Calibration 70 60 RI response [mv] 10 point calibration curve narrowly distributed polystyrene standards EasiCal TM PS-1, spatula A and B dissolved in two solutions lg(mp [g/mol]) 9.0 8.0 7.0 50 40 30 20 spatula B solution spatula A solution 3rd order polynomial Mp [g/mol] 7500000 2560000 841000 320000 148000 59500 28500 10850 2930 580 6.0 5.0 4.0 3.0 2.0 10 Ve [ml] 1.0 0 8 10 12 14 16 18 20 22 0.0 AWE/CRD11/AM10/0020 8

Data Interpretation 16 14 RI signal first step base line construction using elugram regions 1a and 1b 12 10 8 6 4 2 a 2 b second step elugram evaluation from 2a to 2b 2a: 1% from peak height off of baseline 2b: Set to a value where sample chains can still be considered as polymeric with NC this is about 3000 to 4000 g/mol 4000 g/mol correspond to 14-15 glucosid units for NC with N-content of 12.6 mass-% 2 0-2 1 a base line 1 b elution time or volume 5 8 11 14 17 20 23 26 29 Integration limits and base line setting [Ref Pontius H; Dörich M; Bohn M. P69, 39th International Conference of ICT 24-27 June 2008, Karlsruhe, Germany] AWE/CRD11/AM10/0020 9

Polystyrene Standard NBS706 Mn 180000 160000 140000 Mn [g/mol] NIST (NBS) certified value 1979 Mn = 130000 ± 4600 g/mol calculated from NIST(NBS) Mw by Mw/Mn = 2.1 120000 100000 NBS 1979 80000 60000 40000 20000 0 AWE/CRD11/AM10/0020 10

Polystyrene Standard NBS706 Mw 360000 330000 300000 Mw [g/mol] NIST (NBS) certified value for NBS SRM 706 1979 Mw = 273000 ± 9600 g/mol 2003 Mw = 287000 ± 6800 g/mol 270000 240000 210000 NBS 1979 NIST 2003 180000 150000 120000 90000 60000 30000 0 AWE/CRD11/AM10/0020 11

NBS706 Data Mn Mw Laboratory Mean Relative SD at 95% Conf Mean Relative SD at 95% Conf 1 98700 15.9 283070 7.7 2 92660 11.9 269380 1.0 3 95740 5.3 271500 10.5 4 95060 3.4 272660 0.9 5 92050 0.4 266993 0.8 6 123940 1.0 271500 1.5 7 94210 0.9 264850 1.5 8* 120020 48.1 285500 5.7 9 81580 0.6 271590 2.2 10* 83900 5.5 253640 5.4 11 97900 3.9 268910 1.6 * Average of two injections Mean 97500 (93490) 27.9 (13.0) 270730 (270960) 6.9 (6.0) AWE/CRD11/AM10/0020 12

NBS706 Data Evaluation 89.57 83.63 77.69 Data File: 2008-11-12 NBS706_02.vdt Method: 2910am-0003.vcm Outliers in Mn data identified by Grubbs Test Refractive Index Response (mv) 71.75 65.81 59.87 53.93 47.99 42.05 ANOVA analysis: Fcrit = 2.55 Mn Fcalc = 5.61 Not same family Mw Fcalc = 2.59 Same family? 36.11 30.17 7.17 8.42 9.66 10.91 12.16 13.40 14.65 15.89 17.14 18.39 19.63 Retention Volume (ml) Integration at PS 2930 daltons Consistent but includes large area of tail so influences Mn data Conclusion: confidence in Mw determinations for well characterised polymers AWE/CRD11/AM10/0020 13

Nitrocellulose Mn Data (12.15% N) 250000 Mn [g/mol] 200000 150000 100000 50000 0 AWE/CRD11/AM10/0020 14

Nitrocellulose Mw Data (12.15% N) 1200000 Mw [g/mol] 1000000 800000 600000 400000 200000 0 AWE/CRD11/AM10/0020 15

Nitrocellulose Data (12.15% N) Laboratory Mean Mn Relative SD at 95% Conf Mean Mw Relative SD at 95% Conf 1 158040 15.9 757637 10.6 2 151260 3.5 812680 11.7 3 176370 9.9 699720 11.5 4 222050 21.1 1068780 5.6 5 166720 11.0 684490 2.0 6 208310 0.9 427160 0.6 7 121070 2.0 540810 1.6 8 182350 10.3 597480 0.7 9 221920 30.6 511150 3.4 10 144140 4.3 523140 5.4 11 112840 2.5 442657 0.9 Mean 153990 32.0 621420 46.8 AWE/CRD11/AM10/0020 16

Nitrocellulose Data Evaluation ANOVA analysis of 11 data sets : Fcrit = 2.30 NC Fcalc M n M w 11.69% N 28.26 79.87 12.15% N 15.80 5.93 12.71% N 1.16 10.44 Only one case where the data (M n ) can be considered to be the same family at 95% confidence NBS706 data suggest variability is not related to the SEC systems High variability Natural variability of the sample Sample preparation 13.55%N 21.03 65.46 AWE/CRD11/AM10/0020 17

Nitrocellulose Data Evaluation Data File: 2008-11-12_NC_A2_STANAG_02.vdt Method: 2910am-0003.vcm Mn Mw 42.35 41.14 NC 11.69% N Mean Relative SD at 95% Conf Mean Relative SD at 95% Conf 126910 32.3 515570 47.7 121000 24.9 479090 31.7 Refractive Index Response (mv) 39.93 38.73 37.52 36.32 35.11 33.91 32.70 12.15% N 153990 32.0 621420 46.6 153990 32.0 582390 34.7 31.50 30.29 7.35 8.52 9.69 10.85 12.02 13.19 14.36 15.53 16.69 17.86 19.03 Retention Volume (ml) 12.71% N 13.55% N 166830 42.8 780800 89.3 164780 41.9 694880 47.3 121450 36.4 589300 74.4 121440 36.4 520230 41.0 Pre-peak affected two data sets high Mw Same SEC instrument sensitive RI detector AWE/CRD11/AM10/0020 18

Variability Organisation Drying method Sample preparation thought to be main area of variability Drying regime: Initially left to Lab preference Thermal load applied during drying variable. Some degradation possible. Consistency is now believed to be important Pontius et al; P80 40 th International Conference of ICT (2009) AWE Plc ATK Radford Cranfield Univ. Denel Fraunhofer ICT General Dyn. Nitrochemie PVTT TNO Defence samples on filter paper, 80 C for 2 h samples in cardboard tray, - 75 C for 3 h 65 C for 2 h under vacuum over silica gel left overnight under vacuum 65 C for 4 h under vacuum Stored in a vacuum desiccator 40 C for 24 h under vacuum over P 2 O 5 dried to constant mass 70 C for 6 h under vacuum dried to ~ 0.2 mass-% water content dried in hot air blower at 60 C for 1 h 100 C for 0.5 h dried in desiccator for 2 h at ambient temp. 60 C for 2 h 65 C for 24 h water content 0.2 0.4 mass-% AWE/CRD11/AM10/0020 19

Refined Method Sample drying at 60 C for 2 hours Samples prepared at 1.5mgcm -3 in THF stabilised with <250ppm BHT Dissolution time: 30 minutes at 37 C - 39 C with stirring Polystyrene calibration standards across the whole nitrocellulose distribution. Broad polystyrene instrument performance check standard Mobile phase flow rate: 1cm 3 min -1 Triplicate sample injection per vial of 100µl Columns: Two 10µ, 30 x 7.5cm mixed porosity polystyrene divinyl benzene maintained at 35 C AWE/CRD11/AM10/0020 20

Conclusions Good multinational support and commitment NBS706 Mw data demonstrates lab to lab reproducibility Nitrocellulose data has large variability Nitrocellulose sample prep regimes need to be standardised (drying in particular) 7 day dissolution not ideal for manufacturers An alternative fast dissolution method proposed with no impact on data Sample in - data out; 1 working day Refined method accepted for STANAG4178 AWE/CRD11/AM10/0020 21

Acknowledgements Paul Deacon, Ruth Knight AWE, Aldermaston, Reading, UK Phil Gill, Nathalie Mai Cranfield University, Shrivenham, UK Manfred A. Bohn, Heike Pontius Fraunhofer-ICT, Pfinztal, Germany Monique van Hulst, Wim de Klerk TNO Defence, Security & Safety, Rijswijk, The Netherlands Beat Vogelsanger, Ruth Sopranetti Rheinmetall Nitrochemie, Wimmis, Switzerland Carole Baker DOSG, MoD, Abbey Wood, Bristol, UK AWE/CRD11/AM10/0020 22