Av. Professor Lineu Prestes 2242, São Paulo, SP Tel: +55 (11) Fax: + 55 (11)

Similar documents
The use of LA-SF-ICP-MS for nuclear forensics purposes: uranium isotope ratio analysis

Method development for analysis of single hot particles in Safeguards swipe samples

Evaluation of the n( 235 U)/n( 238 U) isotope ratio measurements in a set of uranium samples by thermal ionization mass spectrometry

The analysis of particles of nuclear material finding the proverbial needle in a hay stack

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

STRONTIUM ISOTOPES AS NATURAL TRACERS IN RESERVOIR OILFIELD AND IN GROUNDWATER SYSTEMS

Improvement of bulk analysis of environmental samples by using a multiple collector ICP-MS

Micrometrie particle's isotopics: An ultra-sensitive tool to detect nuclear plant discharge in the environment

Environmental Sample Analysis Advances and Future Trends

Particle Analysis of Environmental Swipe Samples

Autenticity control, provenance testing and fraud detection using mass spectrometry

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Keywords: Safeguards, Destructive Analysis, Environmental Sampling

NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON ELEMENTAL ASSAY U TITRATION

2. QUALITY CONTROL IN MASS SPECTROMETRY

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Error Analysis. The big questions are: - What value best represents a set of measurements and how reliable is it?

TECHNIQUE OF MEASURING THE FIRST IONIZATION COEFFICIENT IN GASES

INVESTIGATIONS OF THE FRICKE GEL (FXG) DOSIMETER DEVELOPED AT IPEN IRRADIATED WITH 60 Co GAMMA RAYS

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Th- 234 U model ages of some uranium standard reference materials

MASS SPECTROMETRIC TECHNIQUES FOR THE RAPID CHARACTERIZATION AND FINGERPRINTING OF NUCLEAR FUEL MATERIALS

DIFFERENCES IN TRACE ELEMENT CONCENTRATIONS IN WHOLE BLOOD OF SMALL SIZED ANIMALS USING ANAA

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE THERMAL IONISATION MASS SPECTROMETRY (TIMS)

Certification Report. REIMEP 18 Inter-Laboratory Comparison for the Measurement of Uranium Isotopic Ratios in Nitric Acid Solution

MECHANICAL AND THERMAL PROPERTIES OF COMMERCIAL MULTILAYER PET/PP FILM IRRADIATED WITH ELECTRON-BEAM

Alpha spectrometry enriched uranium urinalysis results from IPEN

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

PRELIMINARY CONCEPT OF A ZERO POWER NUCLEAR REACTOR CORE

arxiv:nucl-ex/ v2 21 Jul 2005

IAEA-CN-184/159. Environmental Sample Analysis Advances and Future Trends. D. Donohue

Effect of Sampling Depth on the Analyte Response in Laser Ablation Inductively Coupled Plasma Mass Spectrometry

PERFLUORINATED POLYMER GRAFTING: INFLUENCE OF PRE- IRRADIATION CONDITIONS

Defining quality standards for the analysis of solid samples

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 2) Module 4

ICP-MS. High Resolution ICP-MS.

Nu Plasma II - Collector Configuration

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Particle Analysis Finding the Needle in the Haystack

IAEA-SM-367/10/04/P SCREENING AND RADIOMETRIC MEASUREMENT OF ENVIRONMENTAL SWIPE SAMPLES

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Characterization of nuclear and radioactive materials seized from illicit trafficking in Ukraine

Reactor Physics Calculations for a Sub critical Core of the IPEN-MB-01 driven by an external neutron source

EUROPEAN COMMISSION Directorate General Joint Research Centre IRMM NUCLEAR CERTIFIED REFERENCE MATERIALS 2015

Standardization of JRC developed Nuclear Mass Spectrometry Methods with ASTM-International

Observations Regarding Automated SEM and SIMS Analysis of Minerals. Kristofor Ingeneri. April 22, 2009

AGE DETERMINATION OF HIGHLY ENRICHED URANIUM

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Joint Research Centre (JRC)

Isotope Dilution Mass Spectrometry

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS)

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Microparticle Reference Materials for Particle Analysis in Nuclear Safeguards Production & Characterisation

Introduction to Laser Ablation ICP-MS for the Analysis of Forensic Samples Application Note

Looking Forward through the Past: Status of the United States Nuclear Safeguards Reference Material Program at New Brunswick Laboratory

RADIOCHEMISTRY AND NUCLEAR CHEMISTRY Vol. II - Nuclear Forensics - Klaus Mayer and Maria Wallenius

Trace element quality control analysis of environmental samples at the Neutron Activation Analysis Laboratory, IPEN, São Paulo, Brazil

vacuum analysis plasma diagnostics surface science gas analysis

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

NUCLEAR CERTIFIED REFERENCE MATERIALS 2018

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Samples compliant with the WATER

Nuclear Forensics-A Review

Investigation of fundamental mechanisms related to ambient gas heating and hydrodynamics of laser-induced plasmas

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Journal of Analytical Atomic Spectrometry

Agilent Technologies at TIAFT 2013

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Original Paper Fission Gas Bubbles Characterisation in Irradiated UO 2 Fuel by SEM, EPMA and SIMS

THERMOLUMINESCENCE PROPERTIES OF OLIVINE

Detectors for the measurement of ionizing radiation

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Precise Pb isotope analysis of igneous rocks using fully-automated double spike thermal ionization mass spectrometry (FA -DS- TIMS)

Additional Analytical Methods. Detrital zircon samples were collected from nine fine-, medium-, and coarse-grained

FLUORINE IN WATER AND TEETH MEASURED WITH PIGE

Hiden SIMS Secondary Ion Mass Spectrometers. Analysers for surface, elemental and molecular analysis

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Preparation and Certification of IRMM-075. A. Verbruggen, A. Alonso, R. Eykens, F. Kehoe, H. Kühn, S. Richter, R. Wellum, Y.

Secondary ion mass spectrometry (SIMS)

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

Determination of Some Trace Elements in Marine Sediment using ICP-MS and XRF (A Comparative Study)

BLOOD BIOCHEMISTRY EVALUATION IN HORSES HYPERIMMUNE SERA PRODUCTION

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

High Precision Nondestructive Assay to Complement DA. H.O. Menlove, M.T. Swinhoe, and J.B. Marlow Los Alamos National Laboratory

TECHNETIUM-99 IN WATER

dating of geological samples by laser ablation ICPMS

Signal, Noise, and Detection Limits in Mass Spectrometry

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

POLAROGRAPHIC DETERMINATION OF Sn (II) AND TOTAL Sn IN PYRO AND MDP RADIOPHARMACEUTICAL KITS

ICP-OES Application Note Number 35

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

Library of uranium and plutonium reference spectra Introduction

Transcription:

2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 The use of Laser Ablation Sector Field Inductively Coupled Plasma Mass Spectrometry for swipe samples analysis: a viewpoint from Safeguard and Nuclear Forensics RAFAEL C. MARIN 1, JORGE E. S. SARKIS 2 and RAFAEL C. B. PESTANA 3 1 Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor Lineu Prestes 2242, 05508-000 São Paulo, SP Tel: +55 (11) 3133-9391 - Fax: + 55 (11)3133-9018 marin.rafael.c@gmail.com 2 jesarkis@ipen.br 3 rcbpestana@gmail.com ABSTRACT This work describes the utilization of laser ablation sector field inductively coupled plasma mass spectrometry (LA-SF-ICP-MS) for determining uranium isotopic composition in a simulated swipe sample by deposition of U 3 O 8 powder with natural enrichment level. This method is useful to measure 234 U/ 238 U and 235 U/ 238 U. The measurements were performed on a continuous ablation with low energy density and defocusing, which improved the signal stability, in a cluster of uranium particles. Optimization of measurements was achieved by adjusting the following parameters: RF power, laser beam diameter, defocusing of laser beam, laser energy, laser energy-density, auxiliary gas and sample gas. The 235 U/ 238 U isotope ratio was 0.00719 ± 0.00020 and its precision was 1.2 % RSD (relative standard deviation). Uncertainties were estimated following the International Organization for Standardization Guide to the Expression of Uncertainty in Measurement (ISO GUM), with a confidence level of 95.45% (k = 2.00). The results indicate that the Laser Ablation ICP-MS technique offers a rapid and accurate alternative for the measurement of uranium isotope ratios in uranium particle. The technique has the added advantage of allowing measurements straight on the sample (without further preparation), preserving the testimony which is very important for safeguards and nuclear forensics purposes. KEYWORDS: nuclear forensics; laser ablation; LA-SF-ICP-MS; uncertainty; uranium; isotope ratio; safeguards

Introduction Strengthened safeguards procedures arose to prevent clandestine nuclear programmes in States under safeguards agreements. These measures were demonstrated by the Programme 93+2 through a series of field trials. Based on this programme, the use of environmental samples for analyzing uranium enrichment and plutonium separation by means of swipe sampling was established. Swipe samples are taken by wiping various surfaces of nuclear facilities with cotton cloths measuring 10 cm x 10 cm (1; 2; 3). This methodology provides a reliable tool to verify the declarations of the operator, recreate the process history of the material and determine its use (4). Moreover, owing to the increasing number of illicit trafficking of nuclear materials, such procedures are an important element of a new field of science which is responsible to investigate (and determine the source) of seized nuclear materials nuclear forensics (5; 6). Globally, only a few laboratories have developed methodologies to analyze swipe samples. These belong to the Network of Analytical Laboratories (NWAL) (4), accredited by the International Atomic Energy Agency (IAEA). The techniques employed by these laboratories include thermal ionization mass spectrometry (TIMS) (7; 8; 9), secondary ion mass spectrometry (SIMS) (10; 11; 12), multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS), inductively coupled plasma mass spectrometry (ICP-MS) (13; 14; 15), among others. The most accurate and most commonly used technique for isotope characterization is TIMS, which is considered the primary technique for isotope ratio measurements by IAEA. However, it is relatively time consuming (several weeks) and requires extensive sample preparation. By contrast, SIMS takes a few days and can identify and perform isotope ratios measurements in individual particles previously deposited on a carbon disk. However, the less abundant isotopes of uranium ( 234 U and 236 U) can suffer poly-atomic interferences through the combination of lead with oxygen atoms and carbon atoms from the disk (4). ICP-MS has been increasingly used to perform isotope ratio measurements (13; 14; 15), owing to its ultra-trace levels capability (fgg -1 ) (4). It also reaches a precision of down to 0.026% relative standard deviation (RSD) (16; 17). Thus, this technique offers the advantages of ease of operation, multi-elemental analysis (17) and possibility to be coupled to several sample introduction methods, such as Laser Ablation. Recently, laser ablation has been employed in sector field inductively coupled plasma mass spectrometry (LA-SF-ICP-MS) for isotopic analysis of single uranium particles down to 10 μm, achieving precision between 0.9 5.1% RSD (relative standard deviation) (18) and in one case reaching a precision of 0.1% RSD (19). Moreover, even laser ablation inductively coupled plasma mass spectrometry (LA- ICP-MS), which is quadrupole based, has been used to perform uranium isotopic ratio measurements of single uranium particles previously located by fission track (FT), achieving a precision of 24.2 % RSD (20). These studies indicate that this technique offers a fast and accurate isotopic ratio analysis. The aim of this work is to determinate uranium isotopic composition of simulated swipe samples using the LA-SF-ICP-MS technique. The analyses were performed based on fundamental metrological concepts following recommendations in the Guide to the Expression of Uncertainty in Measurement (21).

Experimental Instrumentation The measurements of isotope ratio were performed using a SF-ICP-MS (Sector Field Inductively Coupled Plasma Mass Spectrometer) ELEMENT1, Thermo Electron Corp., Bremen, Germany. Sensitivity was approximately 1 x 10 6 cps for 10 ppb of 114 In. The Laser Ablation (LA) measurements were carried out using a Nd:YAG laser at a wavelength of 266 nm, LUV-266X (New Wave Research Merchantek, Carlsbad - CA, USA). Materials and Methods The main operational conditions were established using a Standard Reference Material (SRM) NIST 610 glass supplied by National Institute of Standards and Technology (NIST, USA), which has a concentration of (461.5 ± 1.1) ppm of uranium and was prepared in a rod form, then sliced into wafers. The method validation and the external correction factor were performed using a Certified Reference Material (CRM) 125-A, a UO 2 pellet certified by NBL (New Brunswick Laboratory, USA), whose the uranium isotopic composition for CRM 125-A is provided in Table 1. Table 1: Certified values for UO 2 pellet. All data are presented with their expanded uncertainty at 95% confidence level (k=2). Isotope Atom percent (%) 235 U 4.0574 ± 0.0028 238 U 95.9049 ± 0.0029 234 U 0.0374 ± 0.0003 236 U 0.0003 ± 0.0002 The simulated swipe sample was prepared with U 3 O 8 natural uranium (NU) particles deposited in an adhesive tape. The sample was then inserted into the laser ablation s chamber and several clusters of uranium material particle were observed. The material was then ablated and carried to the ICP-MS system by an argon flux.

Data evaluation Mass Discrimination Mass discrimination is an important effect due to the Coulomb repulsion force suffered by the positively charged ions formed in the plasma. It causes more deflection on light ions than heavy ions. For this reason the isotope ratios must be corrected for each isotope (13; 22). In LA-SF-ICP-MS applications, this effect is much more complex owing to the interrelationships between laser and target. To minimize these effects a power density of 10 9 Wcm -2 (19) and defocused beam (23) was used. The mass fractionation factor was determined by analyzing five independent replicates of UO 2 Certified Reference Material (CRM 125-A) as described by Marin, R.C. et al. (5). Estimates of Uncertainty Uncertainty estimates consider all sources of errors in a measurement, i.e. the uncertainty consists of the uncertainty from a standard reference materials certificates, the uncertainty from the repeatability of the measurements and systematic errors. Such estimates allow a better understanding of the results and facilitate comparison with different methodologies. The process of estimating uncertainty adopted in the present study is based on the ISO GUM (21) and has been described in detail in previous studies (5; 24) Results and discussion The main operational parameters of the LA-SF-ICP-MS system were optimized using a GLASS standard reference material NIST 610 supplied by National Institute of Standard and Technology (NIST, USA, 1992). The analyses were performed in low resolution mode (R = 350) and the main parameters are given in table 2:

Table 2: Optimized parameters for LA-ICP-MS system. Parameters Liquid sample Laser ablation RF power (W) 1240 1000 Cooling gas (l/min) 16.00 16.00 Auxiliary gas (l/min) 0.91 0.96 Sample gas (l/min) 1.060 1.434 Resolution LOW LOW Run and passes 3 x 1 10 x 2 Mass windows (%) 150 % 5 % Samples per peak 20 800 Search window (%) 100 % 150 % Integration window 100 % 100 % (%) Integration type Average Average Scan type Escan Escan Repetition rate (Hz) - 10 Laser beam diameter - 200 (μm) Laser energy (mj) - 0.096 Laser energy density - 0.31 (J/cm 2 ) Defocusing (mm) - 1.125 After establishing the initial conditions, the swipe sample prepared for the 234 U/ 238 U and 235 U/ 238 U isotope ratios analyses. The signals of the isotopes measured ( 234 U, 235 U and 238 U) are very intense in the first few seconds, and then show a decreasing trend which begins to stabilize after seventy seconds. The signal acquisition has already begun by this time and is integrated during the following eighty five seconds. The isotopic ratios results represent the average of five independent measurements. Thus, they are followed by an inherent error which is represented by their associated uncertainty. It contains all sources of such errors which comprise several components (Table 3) that, when combined, provide the expanded uncertainty. The components of type A are from statistical analysis, whereas the type B components are based on other non-statistical sources.

Table 3 Uncertainty budget considered to uncertainty estimates. N normal distribution (divisor = 1 for type A uncertainty or divisor = 2 for type B uncertainty inherited from the coverage factor of the CRMs calibration certificate), ci coefficient of sensibility, ui standard uncertainties, vi degree of freedom for each uncertainty quote. Uncertainty budget Name Type Dist. Divisor ci ui 235/238 ui 234/238 vi Prob. CRM measurements A N 1 1 0.00009 0.0000049 4 Sample measurement A N 1 1 0.000037 0.000001 4 CRM certificate for B N 1 1 0.000028 0.000003 interest isotope CRM certificate for U B N 1 1 0.000015 0.000015 The results obtained for the isotopic ratios 234 U/ 238 U and 235 U/ 238 U (Table 4) gave RSDs of 3.0 % and 1.2 %, respectively. The reproducibility of the isotopic ratio 235 U/ 238 U, represented by the standard deviation, is compatible with the percent expanded uncertainty (u %) because they are in the same order of magnitude and thus represent a good estimate of uncertainty. There is good agreement between the precision obtained in this study and that reported by other authors for 235 U/ 238 U isotopic ratio analysis. Ponturier et al. reported RSD of 12.7 % when analyzing natural uranium single particles (20) and 9.6 % RSD when analyzing a sample provided during an interlaboratorial comparison programme (Nusimep-7) (4). Kappel et al. obtained RSD 0.65 % from analysis of glass particles doped with uranium of certified isotopic composition (25) and a range varying from 1.1 % to 3.0 % RSD when analyzing particles from CRM 9073-01-B (26). Lloyd et al. reported precision varying from 0.2 % to 1.8 % from an analysis of depleted uranium in soils (27). The isotopic ratio corrected by the mass discrimination factor (R corr ) for the isotope 235 U was 0.00719 ± 0.00020, which represents a percent atomic abundance of (0.71 ± 0.07) %. A relative difference of -2.1 % was found between this value and the recently recommended atom ratio (0.7257 %) (27). The isotopic ratio 234 U/ 238 U was not concordant with the results of Kapple et al. (25) who also analyzed natural uranium. The relative difference between these values was about 37 % and may be explained by polyatomic interferences caused in minor isotopes at low resolution (18) and by low statistical counts.

Table 4 - Results obtained by LA-SF-ICP-MS analysis of swipe samples for the corrected uranium isotope ratio (R corr ), expanded uncertainty (U), percent expanded uncertainty (u %), effective freedom degree (v eff ) and coverage factor (k). All data are presented with their expanded uncertainty at 95% confidence level (k=2). Sample R corr U (k=2) u% (%) ν eff k (t-student) 235 U/ 238 U 0.00719 0.00020 2.8 7 2.51 234 U/ 238 U 0.000076 0.000031 41 7 2.51 Percent expanded uncertainties were calculated by combining the standard uncertainties (Table 3) and considering a Gaussian distribution (k=2). The result for the 235 U/ 238 U isotopic ratio was in good agreement with the percent expanded uncertainty reported by Kappel et al. which ranged from 1.8 % to 3.9 %. Nevertheless, the u% found for the 234 U/ 238 U isotopic ratios was very high, probably because of uneven signals and mixing effects generated during the simultaneous analysis of several particles. A t-student test was performed on the five replicates, because this number is considered low to ensure a Gaussian distribution. This test depends on the effective degree of freedom (ν eff ) and the distribution probability. The coverage factor calculated by the t-student distribution was 2.51, indicating that the distribution was slightly different from a normal distribution. Conclusions This study demonstrated that the use of LA-SF-ICP-MS to directly analyze swipe samples is reliable, robust, cost effective, and avoids time consuming preparation procedures when compared to wet-based ICP-MS methodologies. The precision and expanded uncertainty estimates were excellent for of the analysis of the 235 U/ 238 U isotopic ratio. They make this technique promising from the viewpoint of nuclear safeguards and nuclear forensics; in spite of its destructive analysis characteristics, it consumes only few grams of the sample thereby preserving the evidence. However, it is necessary to evaluate the results to ensure sufficient repetitions increase the degrees of freedom and produce a Gaussian distribution. Moreover, as previously concluded by other authors, the methodology needs to be improved in terms of locating and relocating the particles, thereby increasing the reliability of the technique when analyzing real-life swipe samples.

References 1. Donohue, D. L. Strenghtening IAEA safeguards through environmental sampling and analysis. Journal of Alloys and Compounds. 1998, Vols. 11-18, pp. 271-273. 2. Donohue, D. Environmental Sample Analysis Advances and Future Trends. IAEA. [Online] [Cited: april 10, 2013.] http://www.iaea.org/safeguards/symposium/2010/documents/papersrepository/159.pdf. 3. Fischer, D.M. The Evolution of Environmental Sampling for Safeguards. IAEA. [Online] [Cited: April 10, 2013.] 4. Pointurier, F. and Hubert, A. Performance of laser ablation: quadrupole-b ased ICP-MS coupling for the analysis of single micrometric uranium particles. Journal of Radioanalytical and Nuclear Chemistry. 2012, Vol. 296, pp. 609-616. 5. Marin, R. C., Sarkis, J. E. S. and Nascimento, M. R. The use of LA-SF-ICP-MS for nuclear forensics purposes: uranium isotope ratio analysis. Journal of Radioanalytical and Nuclear Chemistry. 2013, Vol. 295, pp. 99-104. 6. Mayer, K. and Wallenius, M. Nuclear Forensic Methods in Safeguards. ESARDA BULLETIN. 2008, Vol. 38, pp. 44-51. 7. Buchmann, J. H. et al. Environmental monitoring as an important tool for safeguards of nuclear material and nuclear forensics. journal of Radioanalytical and Nuclear Chemistry. 2006, Vol. 270, pp. 291-298. 8. Application of thermal ionisation mass spectrometry for oxygen isotope ratio measurements in uranium oxide. Pajo, L. et al. Helsink : s.n., 2000. 6th Finnish-Russian Radiochemistry Symposium. Finland. 9. Oliveira, OP jr et al. Demonstrating the metrological compatibility of uranium isotope amount ratio measurement results obtained by GSMS, TIMS and MC-ICPMS techniques. Int J Mass Spectrom. 2010, Vol. 291, pp. 48-54. 10. Esaka, F. et al. Particle isolation for analysis of uranium minor isotopes in individual particles by secondary ion mass spectrometry. Talanta. 2007, Vol. 71, pp. 1011-1015. 11. Esake, F. et al. Comparison of ICP-MS and SIMS techniques for determining uranium isotope ratios in individual particles. Talanta. 78, 2009, pp. 290-294. 12. Tamborine, G. et al. Application of secondary ion mass spectrometry to the identification of single particles of uranium and their isotopic measurement. Spectrochimica Acta Part B. 9, 1998, Vol. 53, pp. 1289-1302. 13. Oliveira, OP jr. et al. Isotope measurements in uranium samples using a quadrupole inductively coupled plasma mass spectrometer (ICPMS). J Radioanal Nucl Chem. 2002, Vol. 253, pp. 345-350. 14. Godoy, M. L. D. P. et. al. Application of ICP-QMS for the determination of plutonium in environmental sample for safeguards purposes. Journal of Environmental Radioactivity. 2007, Vol. 97, pp. 124-136. 15. Pestana, R.C.B. et al. New methodology for uranium analysis in swipe samples for nuclear safeguards purposes. Journal of Radioanalytical and Nuclear Chemistry. 10.1007/s10967-013-2467-1, 2013. 16. Boulyga, S.F. et al. Determination of uranium isotopic composition and 236U content of soil samples and hot particles using inductively coupled plasma mass spectrometry. Fresenius J Anal Chem. 370, 2001, pp. 612-617.

17. Becker, J.S. et al. Precise and accurate isotope ratio measurements by ICP-MS. Fresenius J Anal Chem. 2000, Vol. 368, pp. 23-30. 18. Varga, Z. Application of laser ablation inductively coupled plasma mass spectrometry for the isotopic analysis of single uranium particle. Analytica Chimica Acta. 2008, Vol. 625, pp. 1-7. 19. Pickhardt, C. et al. Laser ablation inductively coupled plasma mass spectrometry for direct isotope ratio measurements on solid samples. International Journal of Mass Spectrometry. 2005, Vol. 242, pp. 273-280. 20. Pointurier, F, et al. Application of Nanosecond-UV Laser Ablation Inductively Coupled Plasma Mass Spectrometry for the Isotopic Analysis of Single Submicrometer-Size Uranium Particles. Analytical Chemistry. 2011, Vol. 83, pp. 7841-7848. 21. Joint Comitee for Guides in Metrology. Evaluation of measurement data - guide to the expression of uncertainty in measurements. s.l. : JCGM 100, 2008. 22. Furusawa, H. A. et al. Uranium isotopic analysis by using Sector Field Inductively Mass Spectrometer. Journal of Radioanalytical and Nuclear Chemistry. 1999, Vol. 242, pp. 647-651. 23. Isotope ratio measurements of uranium by LA-HR-ICP-MS. Marin, R. C. and Sarkis, J. E. S. Belo Horizonte : s.n., 2011. INAC. 24. Bürger, S. F. et al. Implementation of Guide to the Espression of uncertainty in Measurement. International Journal of Mass Spectrometry. 2010, Vol. 294, pp. 65-76. 25. Direct uranium isotopic analysis of single micrometer-sized glass particles. Kappel, S. et al. s.l. : Journal of Environmental Radioactivity, 2012, Vol. 113, pp. 8-15. 26. Evaluation strategies for isotope ratio measurements of single particles by LA-MC-ICP- MS. Kappel, S. et al. s.l. : Anal. Bioanal. Chem, 2013, Vol. 405, pp. 2943-2955. 27. Precise and accurate isotopic analysis of microscopic uranium-oxide grains using LA- MC-ICP-MS. Lloyd, N.S. et al. s.l. : JAAS, 2009, Vol. 24, pp. 752-758. DOI: 10.1039/b819373h.