Measurements of secondary particles emitted by 12 C, 4 He and 16 O ion beams in view of innovative dose profiling technique in Particle Therapy

Similar documents
Measurement of secondary particle production induced by particle therapy ion beams impinging on a PMMA target

arxiv: v1 [physics.med-ph] 29 Aug 2016

arxiv: v1 [physics.ins-det] 9 Nov 2015

RDH activity at LNF in P. Ciambrone, E. De Lucia, E. Iarocci, I. Mattei, A. Paoloni, A. Sarti, E. Spiriti (Resp.)

Monte Carlo simulations of ripple filters designed for proton and carbon ion beams in hadrontherapy with active scanning technique

Secondary Particles Produced by Hadron Therapy

April 24, :10 MPLA S page 1. Dose monitoring in particle therapy

A study of monitoring performances with the INSIDE system. Francesco Pennazio on behalf of the INSIDE collaboration

Monitoring of Hadron Therapy by Means of Secondary Particles

Physics of Novel Radiation Modalities Particles and Isotopes. Todd Pawlicki, Ph.D. UC San Diego

Overview and Status of the Austrian Particle Therapy Facility MedAustron. Peter Urschütz

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

Results from Silicon Photo-Multiplier neutron irradiation test

M. Martišíková 1,3, C. Granja 2, J. Jakůbek 2, B. Hartmann 1,3, K. Gwosch 1, P. Soukup 2 and O. Jäkel 1,3,5

Sub-Nanosecond Timing for In-Beam PET in hadrontherapy

First Results and Realization Status of a Proton Computed Radiography Device

A Monte Carlo Study of the Relationship between the Time. Structures of Prompt Gammas and in vivo Radiation Dose in.

Range monitoring with DoseProfiler in Carbon ion therapeutic beam and future perspective with proton and neutron secondary products

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

A prompt gamma camera for real-time range control in Proton Therapy

Outline. Physics of Charge Particle Motion. Physics of Charge Particle Motion 7/31/2014. Proton Therapy I: Basic Proton Therapy

Radionuclide Imaging MII Positron Emission Tomography (PET)

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

8/3/2016. Chia-Ho, Hua St. Jude Children's Research Hospital. Kevin Teo The Hospital of the University of Pennsylvania

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

The reaction p(e,e'p)π 0 to calibrate the Forward and the Large Angle Electromagnetic Shower Calorimeters

Performance of upstream interaction region detectors for the FIRST experiment at GSI

THE mono-energetic hadron beam such as heavy-ions or

Ion- and proton-beams: Experience with Monte Carlo Simulation

Measurements of Carbon ion fragmentation on a thin gold target by the FIRST collaboration at GSI

Geant4 studies of the CNAO facility system for hadrontherapy treatment of uveal melanomas

TERA CONTRIBUTIONS TO PARTNER

A Novel Method for Fragmentation Studies in Particle Therapy: Principles of Ion Identification

Biological Dose Calculations for Particle Therapy in FLUKA

Examples for experiments that can be done at the T9 beam line

Accelerators for Hadrontherapy -- Present & Future --

Il picco di Bragg. G. Battistoni INFN Milano. 08/06/2015 G. Battistoni

Dipartimento di Fisica, Università degli Studi di Milano-Bicocca, Milano 2. Istituto di Fisica del Plasma, CNR, Milano

Gamma-ray emission by proton beam interaction with injected Boron atoms for medical imaging. Giada Petringa - Laboratori Nazionali del Sud -

Progress in using prompt gammas for ion range monitoring during hadrontherapy

Alfredo Mirandola a) * Centro Nazionale di Adroterapia Oncologica (CNAO Foundation), Pavia 27100, Italy

Geant4 simulations of the lead fluoride calorimeter

The heavy-ion magnetic spectrometer PRISMA

The EDM Polarimeter Development at COSY-Jülich

Development of beam delivery systems for proton (ion) therapy

The heavy ion irradiation facility at KVI-CART

EEE4106Z Radiation Interactions & Detection

Simulations in Radiation Therapy

a) Dipartimento di Fisiopatologia Clinica, Università degli Studi di Firenze, v.le Morgagni 85, I-

Fast and compact proton radiography imaging system for proton radiotherapy

Physics of Particle Beams. Hsiao-Ming Lu, Ph.D., Jay Flanz, Ph.D., Harald Paganetti, Ph.D. Massachusetts General Hospital Harvard Medical School

Equalisation of the PMT response to charge particles for the Lucid detector of the ATLAS experiment

Initial Studies in Proton Computed Tomography

S. Agosteo (1,2), A. Fazzi (1,2),, M.V. Introini (1,2), A. Pola (1,2), E. Sagia (1,3), V. Varoli (1,2)

The ALICE Inner Tracking System Off-line Software

The NArCos project (Neutron ARray for Correlation Studies)

Physics of particles. H. Paganetti PhD Massachusetts General Hospital & Harvard Medical School

A gas-filled calorimeter for high intensity beam environments

HEATHER. HElium ion Acceleration for radiotherapy. Jordan Taylor, Rob Edgecock University of Huddersfield Carol Johnstone, Fermilab

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

Towards Proton Computed Tomography

PANDA Muon System Prototype

Development of a High Precision Axial 3-D PET for Brain Imaging

A Complete Simulation of a Triple-GEM Detector

Today, I will present the first of two lectures on neutron interactions.

Compton Camera. Compton Camera

Isotope Production for Nuclear Medicine

Progress in measuring GMR in unstable nuclei: Decay detector calibration and inverse reaction experiment. J. Button, Y.-W. Lui, and D.H.

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

Compton Camera with PositionSensitive Silicon Detectors

Radiation protection issues in proton therapy

Dedicated Arrays: MEDEA GDR studies (E γ = MeV) Highly excited CN E*~ MeV, 4 T 8 MeV

Beam Diagnostics and Instrumentation JUAS, Archamps Peter Forck Gesellschaft für Schwerionenforschnung (GSI)

Use of fragmentation beams at LNS with CHIMERA detector

The Use of Neutron Generators for the Detection of Illicit Materials in the Sea Transportation System

Using Prompt gamma ray emission to address uncertainties in proton therapy

A fast triple GEM detector for high-rate charged-particle triggering

APPLIED RADIATION PHYSICS

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES *

A diamond detector for Space Research and Therapy. Radek Pleskac (GSI Darmstadt)

Validation of Geant4 Physics Models Using Collision Data from the LHC

Radioactivity and Ionizing Radiation

Analysis of Carbon Ion Fragmentation and Angular Fragment Distributions using a flexible set-up

Time-of-Flight PET using Cherenkov Photons Produced in PbF 2

Particle Energy Loss in Matter

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Prospects for achieving < 100 ps FWHM coincidence resolving time in time-of-flight PET

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

Particle Energy Loss in Matter

PoS(PhotoDet 2012)010

Measurement of the plasma astrophysical S factor for the 3 He(D, p) 4 He reaction in exploding molecular clusters

Information about the T9 beam line and experimental facilities

General Physics (PHY 2140)

General Physics (PHY 2140)

A dual scintillator - dual silicon photodiode detector module for intraoperative gamma\beta probe and portable anti-compton spectrometer

arxiv: v1 [physics.ins-det] 14 Jan 2019

Nucleus-Nucleus Interaction Modelling and Applications in Ion Therapy Treatment Planning

Precision neutron flux measurement with a neutron beam monitor

Transcription:

Measurements of secondary particles emitted by C, 4 He and 16 O ion beams in view of innovative dose profiling technique in Particle Therapy A. Rucinski b,d, G. Battistoni g, F. Collamati a,b, E. De Lucia c, R. Faccini a,b, M. Marafini a,b, I. Mattei g, S. Muraro g, R. Paramatti b, V. Patera b,d,e, D. Pinci b, A. Russomando a,b, f, A. Sarti c,d,e, A. Sciubba b,d,e, E. Solfaroli Camillocci a, M. Toppi c, G. Traini a,b, C. Voena a,b a Dipartimento di Fisica, La Sapienza Università di Roma, Roma, Italy b INFN Sezione di Roma, Roma, Italy c Laboratori Nazionali di Frascati dell INFN, Frascati, Italy d Dipartimento di Scienze di Base e Applicate per Ingegneria, La Sapienza Università di Roma, Roma, Italy e Museo Storico della Fisica e Centro Studi e Ricerche E. Fermi, Roma, Italy f Center for Life Nano Science@Sapienza, Istituto Italiano di Tecnologia, Roma, Italy g INFN Sezione di Milano, Milano, Italy Abstract Hadrontherapy is a technique that uses accelerated charged ions for cancer treatment. The high irradiation precision and conformity achievable during hadrontherapy treatments allows for local tumor control and sparing of the surrounding healthy tissues. Such a high spatial selectiveness requires the development of new dose monitoring techniques. It has been proved that the beam emits secondary particles in the path to the tumor, namely g from b + emitters, prompt g from nuclear de-excitation and charged particles, that can be used to monitor Bragg Peak (BP) position and the related dose release. In this contribution preliminary results obtained in the study on the neutral and charged secondary particles produced by C, 4 He and 16 O ion beams of therapeutical energy impinging on PMMA phantoms will be presented. The data acquisition have been performed at GSI (Darmstadt, Germany) and HIT (Heidelberg, Germany) facilities. A correlation between the secondary generation regions and BP position will be shown and the design of a monitoring device exploiting all the secondary information will be outlined. 1. Introduction Currently about 50 radiotherapy centers using proton and/or carbon ion beams for treatment of solid tumors are in operation around the word and several others are under construction or in the planning phase [1]. The benefit from application of ion pencil beams for radiotherapy is related to (i) their favorable, inversed in comparison with photons dose deposition profile (Bragg Peak - BP), (ii) reduced scattering due to the application of active scanning technique for beam delivery and (iii) increased in comparison to photons Radiobiological Effectiveness (RBE) resulting from high Linear Energy Transfer (LET) of those beams [2]. Recent research discussions consider the use of Helium and Oxygen beams for radiotherapy and few synchrotron based hospital centers offer those ions for research, mainly for the radio-biological studies [3]. Helium beams are considered as a compromise between the high LET of Carbon ions and low LET in plateau region of Protons, which could be an optimal solution for the treatment of radio-resistant tumors of pediatric patients [4]. The increased LET characteristic of Oxygen beams, exceeding the LET of Proton, Helium and Carbon beams, is expected to improve efficiency of radio-resistant tumor treatments. Independently of the ion beam used for treatment, major drawbacks of scanned pencil beam delivery (typical longitudinal profile up to teens of millimeters) in comparison with a passive, broad beam approach, are related to the high sensitivity of this technique to the patient mis-positioning and anatomy 355

356 A. Rucinski et al. variations. For this reason, one of the crucial research aspects of hadrontherapy is a development of on-line monitoring techniques allowing real time control of the beam range and dose released per raster point. The presence of such techniques in clinical routine would potentially allow for a reduction of the treatment planning margins and/or application of unconventional irradiation fields, which normally cannot be used due to the proximity of critical organs at risk located after the distal fall-off of the BP [5]. So far, the most established monitoring technique for ion therapy is based on the detection of the backto-back photons produced by the annihilation of positrons coming from b + emitters generated by the beam interaction with the patient, typically using Positron Emission Tomography (PET). However, the obtained information is indirect, and the signal level is lower in comparison with PET signals known from clinical diagnostics, which indicates a need for investigation on other methods [6]. Recent developments focus on the detection of particles produced by the beam interaction in the target, in particular charged particles, originating from the projectile and target fragmentation and prompt photons from nucleus de-excitation [7, 8]. The measurements presented here were performed with PMMA target and confirm substantial photon and charged secondary production from C, 4 He and 16 O beams at 90 and 60 with respect to the primary beam direction. This information is required to calibrate and operate the multimodal dose profile monitor which is currently assembled in our group and will be installed in the Centro Nazionale per l Adroterapia Oncologica (CNAO) in Pavia, Italy. The monitor system will be composed of a two PET heads module, for online monitoring of the b + emitters production, and of a range monitor detector able to detect, track and measure energy of both secondary charged particles and prompt gammas (Fig. 1 right). 2. Experimental setup The test beam was performed in the experimental cave of Heidelberg Ion Beam Therapy Center (HIT) and the obtained results were compared with the previously published results of the experiments performed with Carbon beams at 80 MeV/u in Catania, IT [9, 10] and 220 MeV/u in Darmstadt, DE [7, 8]. At HIT several millions of collisions of Helium (102, 5 and 145 MeV/u), Carbon (0, 160, 180, 220 MeV/u) and Oxygen (210, 260, 300 MeV/u) beams with PMMA targets were registered and secondary production was measured at large angles with respect to the primary beam direction (90 and 60 ). Those angular configurations are considered as optimal for dose monitoring as the spatial resolution in the fragment emission point improves with the emission angle, even if the emission statistics worsens and multiple scattering of particles inside the target increases at the large angles with respect to the primary beam direction [7]. The target box with a 4.8x4.8 cm 2 face orthogonal to the beam line was positioned at the beam isocenter 1 m away from the beam nozzle. The primary beam rate impinging the PMMA target was measured with a 0.2 cm thin plastic scintillator (Start Counter, SC) read out by two opposite Hamamatsu H10580 photomultiplier tubes. The angular distribution of the secondary particles produced in the target were studied by three isocentrically positioned detectors mounted on a movable support: (i) 0.1 cmthick plastic scintillatior (LTS), (ii) 21 cm-long drift chamber (DC) and (iii) an array of four LYSO 1.5x1.5x cm 3 crystals [9, 10, ]. The primary energy of the beam and PMMA target length (7.65, 10.0,.65 cm) were selected in such a way, that the BP position was in-line with the center of LTS, DC and LYSO detectors. For C beam PMMA target at one length (10 cm) was used. The DC [11], consisting of six alternated horizontal and six vertical wire layers, was used for three dimensional reconstruction of the charged secondaries tracks [7]. The readout and performances of the DC as well as the tracking algorithm and DC calibration have been reported elsewhere [10]. The sketch of the experimental setup is shown in Fig. 1 (left). Time and charge information from all described above detectors have been used to select prompt

Measurements of secondary particles emitted by C, 4 He and 16 O ion beams... 357 Beam SC 37cm 8cm PMMA LTS 50.5cm φ x 23cm DC y z LYSO Fig. 1: LEFT: Experimental setup at HIT. Beam - primary beam, SC - start counter, PMMA - target, LTS - plastic scintillator, DC - drift chamber, LYSO - matrix of four LYSO crystals, j - angular configuration of movable support with LTS, DC and LYSO detectors (90 and 60 with respect to the primary beam). The coordinate system is plotted in the box on the bottom left side of the figure. BP was expected within the PMMA target at the intersection of the dashed lines for C at 220 MeV/u, 4 He and 16 O beams. RIGHT: A sketch of the multi-modal in-beam dose monitor able to detect at the same time, back-to-back gammas from b + annihilation (PET HEADS, not covered in this contribution), charged secondary particles emerging from the patient and prompt gammas with energies higher than 1 MeV (DOSE PROFILER). gamma and charged secondaries. The lateral profile of the primary beams (FWHM) was 1 cm. The trigger was performed by requiring SC and LYSO coincidence within 100 ns. The maximal beam rate was of few MHz whereas the maximal trigger rate was 10 khz. 3. Preliminary results 3.1 Charged secondaries The charged secondary particles were selected by exploiting the DC information together with the energy released in the LYSO detector. In addition the kinetic energy of the particle was estimated from Time of Flight (TOF) calculated as the time difference between LTS and LYSO signals. Figure 2 (left) shows the number of DC cells (N hit ) fired by charged secondary particles produced in the 90 setup configuration by the 220 MeV/u Carbon beam. Most of the events cross all the DC planes, which appears as a peak for N hit =. The charged particles tracks were reconstructed requiring at least eight cells fired in the DC. The particle identification was performed by selecting the events on the plot of charge released in LYSO as a function of TOF of the particle. A non negligible production of charged fragments at large angles is observed for all beam types. The emission shape reconstructed from the tracked particles can be correlated to the beam entrance window and the Bragg Peak position as it was already discussed elsewhere [7]. In Fig. 2 (right) the reconstructed emission positions of the charged fragments projected along the beam direction (longitudinal profile; y-z plane) for the Carbon beams of all the exploited energies is shown for 90 setup configuration. The reconstructed emission position projected perpendicularly to the beam direction (lateral pro-

358 A. Rucinski et al. No. of cells hit in DC 500 400 300 C - 220 MeV/u counts / 0.8 cm [a.u.] 250 200 150 C - 220 MeV/u C - 180 MeV/u C - 160 MeV/u C - 0 MeV/u 200 100 100 50 0 2 4 6 8 10 14 16 18 20 N hits 0 15 10 5 0 5 10 z [cm] Fig. 2: LEFT: Distribution of the number of cells hit in the DC (N hit ) for secondary charged particles emitted from the interactions of C beams with the PMMA target. The peak at N hit = is an indication of a charged particle that went trough all the DC planes. RIGHT: Emission profiles of secondaries produced by Carbon beam at different energies. The PMMA target (10 cm long) ranged from z=-9 cm (beam entrance window) to z=+1 cm with BP position for C at 0 MeV, 160 MeV, 180 MeV, and 220 MeV approximately at z=-6 cm, z=-4 cm, z=-3 cm, and z=0 cm, respectively. file; x-y plane) is a convolution of the transverse profile of the primary beam (spot size), multiple scattering (MS) of the fragments in the PMMA target and DC tracking resolution, which has been studied in detail elsewhere [7]. The lateral profile can be used to evaluate the MS of the fragments inside the target knowing the remaining two contributions a priori. 3.2 Prompt-g Prompt-g events were selected requiring less than three fired DC cells. The bi-dimensional distribution of energy deposited in the LYSO crystal for these events as a function of their TOF was used to select the prompt photons. An example of such a distribution for C beam at 220 MeV/u is shown on Fig. 3 (left). The prompt photon signal corrected by the slewing effect induced by the front-end electronics fixed voltage threshold is a vertical band at 0 ns. The LYSO intrinsic noise is visible as the horizontal low energy band, while the diffused cloud is mainly due to neutrons at ToF values greater than of prompt photons. The number of prompt photons has been extracted from an unbinned maximum likelihood fit to the time pull distribution for each energy bin and energy spectra of the prompt gamma emitted from all the beams were obtained. In Fig. 3 (right) the energy spectrum for the prompt-g emitted in the interaction of the C beam with the PMMA target at 90 is shown. The observed spectra are given by the sum of the raw spectra corresponding to the nuclear de-excitation of both projectile and target fragments produced in the interaction of the ion beam with the PMMA target. While in Helium and Carbon beam interactions the target de-excitation is dominant, in the study of Oxygen spectral characteristics an non negligible contribution from projectile fragmentation was observed. Helium and Carbon energy spectra are in good agreement with the experimental results obtained with 80 MeV/u Carbon beam [9] and with 220 MeV/u Carbon beam [8].

Measurements of secondary particles emitted by C, 4 He and 16 O ion beams... 359 Energy [MeV] 14 10 8 6 4 2 10 10 1 Candidates / ( 0.5 MeV) 250 200 150 100 50 2 10 8 6 4 2 0 2 4 6 8 10 ToF [ns] 0 2 4 6 8 10 Energy (MeV) Fig. 3: LEFT Energy deposited in the LYSO crystal as a function of TOF. Vertical band at 0 ns corresponds to prompt-g events. RIGHT: Energy spectrum of prompt-g for C beam at 220 MeV. 4. Dose Profiler - charged secondary on-line monitor The non negligible production of charged secondary particles emitted by all the investigated primary beam interaction with a PMMA target allows to exploit the correlation of charged secondary protons emission profile with position of the primary beam for monitoring of the dose release in ion beam therapy. For this purpose a tracking device, "Dose Profiler" (Fig. 1 right), made out of a scintillating fibers system, coupled to a calorimeter and realized by means of LYSO crystals matrix for measurement of particles energy was designed and is currently under construction. The materials of the detector were selected minimizing the charged particles multiple scattering in the detector. The Dose Profiler dedicated to track protons generated by the primary beam inside the patient consists of six planes of squared (0.5x0.5 mm 2 ) scintillating fibers (2 orthogonal layers in each plane) connected to silicon photo multipliers (SiPM) and read out by Basic32 front-end boards. The secondary proton detection and tracking efficiency is expected to be above 90%. Depending on the position of the tumor in the patient, before reaching the detector the particles have to travel few centimeters inside the patient losing their energy. In order to reduce the impact of multiple scattering, the device design maximize the geometrical acceptance in order to improve statistics collected for a given dose to the patient. 5. Conclusions During the test beam at HIT several millions of collisions of C, 4 He and 16 O beams with a PMMA target were collected exploring the range of energies that is of interest for particle therapy. The experiment confirms an evidence of a non negligible secondary charged and prompt-g production at 60 and 90 with respect to the primary beam, supporting the concept of on-line dose monitoring with those secondaries in ion beam therapy. The correlation of charged fragments emission profile with the position of Bragg Peak is possible with Helium, Carbon and Oxygen beams at all investigated energies. The development of a prototype monitoring system "Dose Profiler" dedicated for tracking of charged secondary particles on-line with a purpose of emission profile reconstruction is on the advanced stage. First tests, calibration and installation of the device in the treatment room of Centro Nazionale di Adroterapia

360 A. Rucinski et al. Oncologica (CNAO) is foreseen in 2016. References [1] M. Jermann, Int J Particle Ther., available at the URL http://www.theijpt.org/doi/pdf/10.14338/ IJPT-15-00013. [2] J. S. Loeffler, M. Durante, Nat Rev Clin Oncol, http://dx.doi.org/10.1038/nrclinonc.2013.79. [3] C. Kurz et al., Phys. Med. Biol., http://iopscience.iop.org/article/10.1088/0031-9155/57/ 15/5017. [4] H. Fuchs et al. Radiotherapy and Oncology, http://dx.doi.org/10.1016/s0167-8140(15)30498-9. [5] A.C. Knopf and A. Lomax Phys. Med. Biol., http://iopscience.iop.org/article/10.1088/ 0031-9155/58/15/R131/pdf. [6] K Parodi et al., IJROBP, http://dx.doi.org/10.1016/j.ijrobp.2008.02.033. [7] L. Piersanti et al. Phys. Med. Biol., available at the URL http://iopscience.iop.org/article/10. 1088/0031-9155/59/7/1857/pdf. [8] I Mattei et al. Accepted for publication in JINST [9] C Agodi et al. http://iopscience.iop.org/article/10.1088/1748-0221/7/03/p03001/pdf. [10] C Agodi et al. http://iopscience.iop.org/article/10.1088/0031-9155/57/18/5667/pdf. [11] Z Abou-Haidar et al., http://iopscience.iop.org/article/10.1088/1748-0221/7/02/p02006/ pdf. [] C. Agodi et al., Nucl. Instrum. Methods Phys. Res. B, http://arxiv.org/pdf/02.1676v2.pdf.