HIGH SPEED GC-TOFMS FOR ESSENTIAL OILS PROFILING IN ROUTINE QUALITY ASSESSMENT

Similar documents
UNIQUE SOLUTION FOR FLAVOUR AND FRAGRANCE IDENTIFICATION BY MEANS OF GC-MS (TOF) / CONDENSED PHASE FTIR

Benzene Contamination in BabyFood and Beverages by New Generation of Static Headspace Autosampler coupled to Fast GC-TOFMS

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

TIME OF FLIGHT GC/MSTOF

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

VGA-100 Instrument Detection Limits and Linearity

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Determination of Hydrocarbon Components in Petroleum Naphthas

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Determination of Volatile Substances Proof of Food Adulteration

GC/MS BATCH NUMBER: T50100

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

Multi-residue analysis of pesticides by GC-HRMS

Author. Abstract. Introduction

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Separation and Characterization of Indian and Australian Sandalwood Oils

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

Pittcon P

Benefits of Equipping your Lab with a Time-of-Flight Mass Spectrometer

U.S. EPA Method 8270 for multicomponent analyte determination

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

A New Web-Based Application for Modeling Gas Chromatographic Separations

Case 1:13-cv WBH Document Filed 12/30/16 Page 406 of 519. Exhibit 53

GAS CHROMATOGRAPHY (GC)

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Highly Sensitive and Rugged GC/MS/MS Tool

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative.

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Headspace Technology for GC and GC/MS: Features, benefits & applications

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Expectations for GC-MS Lab

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

IJPRD, 2011; Vol 3(7): October 2011 (51-55) International Standard Serial Number

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Introduction to GC/MS

GC/MS BATCH NUMBER: CB7102

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

GC Application Note. Dilution/standard addition: Efficient and traceable.

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

GC-FTIR Analysis of Lemon Extract: Gram-Schmidt Chromatogram. Solvent Peak(2.20) H 3 C CH Air Peak(2.08)

TOF. Plus. Time of Flight GC/MS. The New Concept of High Speed TOF-MS. Extremely Compact Design. High Sample Throughput

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

The leaf essential oil of Abies grandis (Doug. ex D. Don) Lindl. (Pinaceae): revisited 38 years later

Application. Gas Chromatography February Introduction

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

Phytochemical Analysis of the Essential Oil from Aerial parts of Pulicaria undulata (L.) Kostel from Sudan

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

This document is a preview generated by EVS

Chapter 27: Gas Chromatography

CERTIFICATE OF ANALYSIS - GC PROFILING

WADA Technical Document TD2003IDCR

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Supplementary Materials: Model-based peak alignment of metabolomic profiling from comprehensive two-dimensional gas chromatography mass spectrometry

Analysis of Fast Moving Consumer Goods (FMCG) using GCMS with static and dynamic headspace (HS)

Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

GC-CI-MS analysis of TMS derivatives

Jack Cochran, Lindsey Shear-Laude, Alex Hodgson VUV Analytics Emerald Conference. #emeraldconference

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

CHAPTER 6 GAS CHROMATOGRAPHY

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Chapter 31 Gas Chromatography. Carrier Gas System

Transcription:

HIGH SPEED GC-TOFMS FOR ESSENTIAL OILS PROFILING IN ROUTINE QUALITY ASSESSMENT APPLICATION NOTE Authors: Daniela Cavagnino DANI Instruments SpA viale Brianza, 87 Cologno Monzese Milano Italy Fast GC-TOFMS for faster Essential Oil Characterization Dual Detector Configuration MS/FID for qualitative and quantitative data Dedicated F&F library for Retention Index and Spectra matching Suitable for QA/QC laboratory in F&F Industry Introduction Natural flavoring materials, such as essential oils, are complex hydrophobic mixtures of volatiles components from natural origin, usually steam distilled or solvent extracted from plants. These oils are widely used as odorants and flavoring ingredients in a variety of consumer products such as perfumes, cosmetics, cleaning products, and also as food and beverage taste enhancers. The composition of essential oils can vary significantly with place of origin, harvest season and weather. Oils may be blended, cut with other materials, or otherwise adulterated. The quality assessment of the essential oils is therefore of primary importance to assure consistency in the production since even minor differences in the oil composition can lead to significant alteration of the flavor. An increasing number of accurate quantitative analysis in routine control are required due to stringent regulatory constraints and the possibility to speed up the separation and increase the number of samples per day is extremely appealing. Due to its robustness and easy applicability, gas chromatography (GC) with FID detection is a common analytical technique used to ascertain EO quality and purity. However, without the support of an MS spectrometer, the full characterization and identification of the compounds is very hard and time-consuming. Besides, the significant cost per analysis and the large number of samples-per-day to be analyzed, are requiring faster response. This paper aims to highlight the analytical performance of a dedicated Fast GC system hyphenated to a new bench top High Speed TOF-MS and set up with a Dual-Detector MS-FID configuration for fast qualitative and quantitative characterization of different essential oils. The analysis time is extremely reduced by using short and narrow bore capillary columns and the high acquisition speed of the DANI MasterTOF allows for mass spectra information and suitable deconvolution of co-eluted components.

Experimental The analytical requirements for a comprehensive Essential Oil characterization includes reliable identification based on deconvolution and Spectra/ Retention Index matching against a dedicated F&F library to assure highly reliable identification of the components. Comprehensive qualitative and quantitative information in a single run can be achieved with a Dual-Detector configuration and a postcolumn splitting between a FID and a MS. For this purpose the DANI MasterGC-TOFMS has been set up accordingly, as shown in Figure 1. In this configuration, two pieces of deactivated empty fused silica capillaries are used as transfer lines to connect the analytical column to the two detectors. Because of the different outlet pressure and flows to the MS and the FID detector, the dimensions of the two capillaries must be carefully calculated and balanced, in order to assure the proper restriction for the desired split of sample between the two detectors. In this case the capillary transfer line dimensions have been calculated to have equal splitting of the eluent. The MasterAS as liquid autosampler has been used for the analyses, using 10m, 0.1mm id narrow bore columns to achieve fast GC separations. Highest temperature ramps for the GC oven are also required to properly exploit the efficiency of the short and narrow bore columns. Different essential oils have been tested and injected as neat samples without previous dilution. Analytical conditions for both conventional and Fast GC-TOFMS analyses are summarized in Table 1. The MS data have been acquired and processed with DANI MasterLAB software which allows proper deconvolution and identification by matching the reconstructed spectra with a dedicated DANI F&F Library, for the complete characterization of the samples. The FID data have been acquired with Clarity Data System (Data Apex). The data acquisition from the two detectors is achieved simultaneously by running the two software at the same time. Figure 1. Master GC-TOFMS/FID Dual Detector configuration

Master GC & Master AS Conventional GC Fast GC Column StabilWax 30m, 0.25mm id, 0.25µm th. SolgelWax 10m x 0.1mm x 0.1 µm Ovent Temp. Pr. 40 C, 5 C/min to 240 C 40 C (1min), 15 C/min to 100 C, 35 C/min to 240 C Injector Temp SSL 250 C SSL 250 C Split Ratio 1:500 1:500 Carrier Gas He Const Flow 0.8 ml/min He Const Flow 1mL/min Injection Vol. 0.2 µl 0.2 µl FID 280 C Master TOF MS EI Source 70 ev 70 ev Source Temp 200 C 200 C Transfer Line 250 C 250 C Temp. Mass Range 50-600 amu 50-600 amu Sampling Rate 10 Hz 25 Hz Dual-Detector Configuration Capillary Restrictors Split 1:1 (applied only to Fast GC) To the FID 0.3m, 0.1mm id To the TOFMS 1m, 0.1mm id a) 1. Pinene 2. Paracymene 3. Camphor Results and discussion The use of short and narrow bore columns combined with faster GC oven temperature programs, allows to significantly reduce the analysis time and match the increasing demand of high sample throughput in quality laboratories. Conventional analytical methods for essential oils profiling, generally produce analyses time of about 30-50min, especially when a complete characterization of the samples is required. Exploiting the fast GC approach, it is possible to gain up to 6-fold reduction of the run time, maintaining the required resolution, precision and accuracy of the data. Besides, the resulting compression of the chromatogram leads to compression of each single peak significantly enhancing the signal-to-noise ratio. b) 1. Pinene 2. Paracymene 3. Camphor 1 2 3 30 min 1 2 3 4.5 min Peak # Compound Conventional GC Fast GC RT (min) PW (min) SNR RT (min) PW (min) SNR 1 Pinene 9.08 0.14 1067 1.37 0.05 3996 2 Paracymene 13.11 0.16 4899 2.13 0.07 6591 3 Camphor 19.35 0.17 1636 3.34 0.04 3372 Figure 2. Rosemary Oil Separation: a) Conventional GC-MS; b) Fast GC-MS; Example of peaks compression and increased signal-to-noise ratio

Figure 2 shows the characterization of a rosemary oil in conventional and fast GC-TOFMS conditions, according to the method parameters reported in Table 1. Using a 10 m, 0.1 mm id column and a faster column heating rate, the reduction in analysis time is only advantageous in terms of separation, sensitivity and productivity, as shown by the preserved resolution and gain in signal-tonoise ratio of the highlighted peaks. The capability of the MS detector to correctly describe the narrow peaks with a suitable acquisition speed is of primary importance to maintain peak resolution and to correctly exploit the narrow bore columns efficiency. In the landscape of the MS spectrometers available on the market, the DANI High Speed Time of Flight is the most affordable solution capable of the fastest acquisition speed up to 1000 spectra/s saved to disk. It permits to acquire the signal at a suitable acquisition speed without sacrificing sensitivity and matching the typical limits of detection of quadrupole in SIM mode, but over the entire mass range of interest. The intrinsic benefit of the Time of Flight technology compared to scanning systems as quadrupole analyzers, is in fact the capability to acquire simultaneously over the entire mass range of interest without need of scanning. This feature guarantee much faster data acquisition and higher ion transmission through the analyzer since no filtering is applied. The fast acquisition speed and the suitable number of data points across the chromatographic peaks make the deconvolution algorithm to be much more performing, allowing a reliable identification of co-eluted compounds or traces behind the background signal. Figure 3. Nutmeg Oil separation. Comparison between conventional GC-MS separation in 38 min and Fast GC-MS in 8 min, with maintened resolution

In Figure 3 an example of fast GC-TOF separation of a Nutmeg Oil sample is reported, compared to a conventional analysis. The fast separation is accomplished in the same time a conventional approach is delivering the first peak, while the chromatographic profile is evidently maintained. The identification of the compounds has been achieved after deconvolution and spectra / Retention Index matching against a dedicated MS library. The deconvolution algorithm recognizes small differences in the ions profile of the spectrum of each data point, discovering the presence of more compounds having only a slightly different retention time, but appearing fully co-eluted in the TIC profile, as shown in Figure 4. The matching of the deconvoluted spectra is very good, assuring reliable compounds identification. In Table 2 the list of identified compounds in the Nutmeg Oil sample is reported. A filter has been applied in the processing method settings, with spectra matching higher than 90% and a RI window of ±30 unit. Figure 4. Deconvolution of co-eluted compounds and reconstructed MS spectra with library matching

RT COMPOUND Δ RI MATCH 1.3800 Alpha-Pinene -17.7 95 1.3933 Alpha-Thujene -10.0 95 1.5599 Camphene -23.2 100 1.7899 Beta-Pinene -16.7 97 1.8590 Sabinene -13.5 95 1.9891 Delta-3-Carene -24.6 99 2.0780 Alpha-phelandrene -17.2 91 2.0827 Myrcene -10.4 97 2.0967 Pseudolimonene -14.7 93 2.1598 Alpha-Terpinene -17.1 99 2.2629 Limonene -24.1 96 2.3106 Beta-Phellandrene -24.9 97 2.5159 Gamma-Terpinene -25.7 98 2.5493 Cis-Beta-Ocimene 6.5 96 2.6449 Paracymene -27.0 98 2.7123 Terpinolene -21.6 98 3.5141 Dehydroparacymene -12 96 3.6513 Trans-Sabinene hydrate 1.8 99 3.6822 Alpha-Cubebene -10.2 95 3.8529 Alpha-Copaene -12.8 99 4.0641 Linalol -0.5 97 RT COMPOUND Δ RI MATCH 4.2562 Isobornyl Acetate -16.3 96 4.2594 Bornyl Acetate -11.6 98 4.3081 Trans-Alpha-Bergamotene 1.3 93 4.3453 Terpinen-4-ol -1.1 94 4.3666 Caryophyllene -7.8 95 4.5825 Pinocaveol 0.7 92 4.6447 Citronellyle Acetate -3.0 99 4.7745 Alpha-Terpineol -24.2 98 4.8000 Alpha-Terpinyl Acetate -6.8 99 4.8858 Germacrene D -32.5 95 4.9567 Beta-Bisabolene -12.8 98 5.0675 Acetate Geranyle -7.1 97 5.0923 Delta-Cadinene 12.5 98 5.4179 Cis-Calamenene -10.2 93 5.5643 Safrol -10.2 100 6.1249 Methyl Eugenol 9.5 100 6.7169 Eugenol -0.2 96 6.9540 Elemicine na 98 7.0988 Myristicine -18.7 98 7.3706 IsoEugenol E -19.9 96 8.0973 Vanilline na 95 Table 2. Nutmeg Oil Compounds Identification In a conventional dual channel configuration, a GC-FID and a GC-MS chromatograms are obtained separately and compared. The post-column splitting in a dual-detector configuration enables to obtain both TIC and FID data at the same time, permitting a correct and easier comparison of the data. Two examples are shown in Figure 5 for Mint Oil and Lavander Oil, where for peaks A, B and C the difference in retention time are less than one second. The proper low dead volume capillary connections to both the detectors with appropriate dimensions allow to equally split the carrier gas flow, minimizing possible different hold up time and maintaining an excellent peak alignment across the chromatogram. The constant flow control for the carrier gas is also fundamental to keep the post column splitting during the GC oven heating.

Figure 5. Peak Alignment for Mint Oil with Dual Detector Configuration

FID Area % has been calculated for quantitation, correcting the areas with appropriate response factors, as reported in literature [1]. Different types of essential oils with different complexity were characterized in fast GC in less than 8 minutes, as reported in Figure 6. Peppermint Oil Bergamot Oil Mandarin Oil Sweet Lemon Oil Figure 6. Fast GC-TOFMS separation of essential oil in less than 8 min

- Linalol + Linalol Narrow bore column with chiral phase has been also used to separate enantiomer pairs, useful to help in authentication of natural products. The presence of racemic pairs most often indicates adulteration or unnatural origin. In Figure 7, the separation of nutmeg oil on chiral phase is illustrated, with the detail of the chiral separation of optical isomers of Linalol. Figure 7. Characterization of nutmeg oil on chiral stationary phase The essential oil composition is strictly correlated to the sample origin and direct comparison of the chromatograms can reveal major differences due to the provenience. However, in the case of three lemon oils with different origin analyzed and compared (Figure 8), the highest content of Geraniol found in one sample could be correlated to an adulteration process, since this compounds typically occurs in small quantity in lemon oil. Figure 8. Characterization of different types of Lemon Oil

Conclusions For EO profiling, the combination FID / TOFMS is advisable to satisfy both qualitative and quantitative aspects for a fast and comprehensive characterization. Shorter analysis time and combined FID/TOFMS data provide maximum analytical information with ID confirmation and quantitation in a single run, for highest productivity. The post-column splitting in constant flow mode is the recommended configuration assuring consistent retention times and constant splitting between the detectors. Proper transfer line capillary dimensions, taking account of the different outlet pressure, allows to modulate the required split ratio between the detectors while assuring the required peak RT alignment. FID area counts with proper response factors provide percentage abundance information, while high speed TOF MS data allows for a fast and reliable compounds identification by spectra and retention index matching with dedicated F&F library. The quality of the data and the fast and automatic reprocessing procedure supported by efficient deconvolution algorithms, are attractive features targeting the daily routine analysis of QA/QC laboratories. Reference [1] C. Bicchi et al., Flavour Fragr. J., 2008; 23: 382-391 www.dani-instruments.com The contents of this publication are for reference and illustrative purposes only. Information, descriptions and specification are subject to change without notice. DANI Instruments assumes no responsibility and will not be liable for any errors or omissions contained herein or for incidental, consequential damages or losses in connection with the furnishing, performance, or use of this material. AN_151