Uncertainty determination of correlated color temperature for high intensity discharge lamps.

Similar documents
Uncertainty determination of correlated color temperature for high intensity discharge lamps

Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors

Spectral and photopic studies for high intensity discharge (HID) lamps

Introduction to Colorimetry

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA.

Report No.: HZ p. (Lumens) Stabilization Time CRI (K)

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

Standard-Tech Co. Ltd Testing Center

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl5s50u1

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl4s50u1

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A)

Checked X.CAO

NIST Role in Supporting Solid State Lighting Initiative

LM Test Report. For. Beyond LED Technology. (Brand Name: Beyond LED) 2725 Mountain Industrial Blvd. Suite A-1 Tucker, GA 30084

Checked X.CAO

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es)

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

NVLAP LAB CODE LM Test Report. For. Shenzhen Guanke Technologies Co. Ltd. (Brand Name: )

Radiometry, photometry, measuring color

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT8N20-A

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-6

NVLAP LAB CODE LM Test Report. For CE INNOVATIONS LTD. (Brand Name: N/A) 911 Denison St Markham, ON L3R 3K4 Canada

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-9

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

LightLab I N T E R N A T I O N A L

Wet Locations Use 71400A 120 W PF V 2.9A 93% 144 pcs. Philips 130 LM/W 5000K Hrs. 5 Years. 24-1/4 Lbs 665*530*120 1 PCS

Measurement method for the proficiency testing program

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

LED Entryway Light RoHS

Experiment 4 Radiation in the Visible Spectrum

Cat# LED High Bay

LM Test Report. For MaxLite Inc. LED Commercial Down lights Model: RR82540W

Report of Test LLIA A

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

Morris Products 53 Carey Road, Queensbury, NY 12804

Report of Test LLIA A

LM Test Report. for. Forest Lighting Corporate. LED Tube. Model: FL-T8U BN

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

Fundamental Concepts of Radiometry p. 1 Electromagnetic Radiation p. 1 Terminology Conventions p. 3 Wavelength Notations and Solid Angle p.

LightLab I N T E R N A T I O N A L

Report No.: EASZF Page 2 of 14

TECHNICAL NOTE. Relating Photochemical and Photobiological Quantities to Photometric Quantities

Report of Test LLIA A

LightLab I N T E R N A T I O N A L

TRNSYS MODELING OF A HYBRID LIGHTING SYSTEM: BUILDING ENERGY LOADS AND CHROMATICITY ANALYSIS

LM Test Report. For Elec-Tech International Co., Ltd. T8 LED TUBE Model: XX(XX:41-50)

LM Test Report. For Elec-Tech International Co., Ltd. T8 LED TUBE Model: XX(XX:41-50)

PHONE: (303) FAX: (303) WEBSITE:

RoHS. Dimming: PWM/VR/0-10V QPL ID # PL21D5Q1B95E. Cat# W LED FlatPanel Slipfitter Mount. EPA Rating 2.

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

Lighting Sciences

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products

CERTIFIED TEST REPORT No

VI. Terminology for Display

PHOTOMETRIC TESTING & EVALUATION TO IES LM Sample Tested LN130-A30 LED HIGH BAY. Prepared for: Janie Chung

A*STAR Seminar on LED and Solid State Lighting. Standards and Technologies. Measurements. 05 August LIU Yuanjie National Metrology Centre

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

Lighting fundamentals

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

Handbook of LED and SSL Metrology

National Voluntary Laboratory Accreditation Program

Novel Flux Calibration Source. CORM 2007 David C. Gross May 10, 2007

TEST REPORT IEC Photobiological safety of lamps and lamp systems

Module 1. Illumination Engineering Basics. Version 2 EE IIT, Kharagpur 1

INFRAMET. 2.1 Basic laws

Design of visual environment

Stray Light Rejection in Array Spectrometers

LM-79 Report TEST REPORT NO. L523A

2. Lighting Terms. Contents

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

LightLab I N T E R N A T I O N A L

LightLab I N T E R N A T I O N A L

OPAC 101 Introduction to Optics

Fundamentals of light

Fundamentals of light

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Practical Optical Measurements of OLED Panels for Lighting Applications

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

application note LED measurement radiant power measurement

Photometric Test Report. Keystone Technologies

Photometric Test Report. Keystone Technologies

Photometric Test Report

Test Report. Project No: Report No: aR01 Report Issued Date: Customer Company & Address:

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

The official directions are written in Chinese, this English edition is for your reference only -1-

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Compliance with the National requirements of EUROPEAN GROUP DIFFERENCES AND NATIONAL DIFFERENCES for EN 62471:2008.

Module 1 LIGHT SOURCES Lecture 1. Introduction. Basic principles of Light and Vision

Transcription:

Uncertainty determination of correlated color temperature for high intensity discharge lamps A.B. El-Bialy 1, M.M. El-Ganainy 2 and E.M. El-Moghazy 3 1 University College for Woman for Art, science and education. Cairo, Egypt 2 National Institute for Standards (NIS), Giza, code 11211, Egypt 3 NIS and Ph.D. student in University College of Woman, Giza, code 11211, Egypt. emoghazy@yahoo.com Abstract: Color temperature is a description of the color of light sources. The chromaticity coordinates of the light source lying on the Planckian locus which is called (Commission Internationale de l'eclariage, referred to as CIE) CIE diagram and the source has color temperature (in Kelvin) equal to the blackbody temperature of the Planckian radiator. For light sources that don't have chromaticity coordinates that fall exactly on the Planckian locus but lie near it. In this case the chromaticity coordinates of such sources can be representing by correlated color temperature (CCT). Uncertainty of Correlated Color Temperature (CCT) or (T cp ) for high intensity discharge lamps (HID) is derived from (u, v) color coordinates. The method of the International organization for standardization (ISO) Guide is applied by Gardner to drive analytical expression for uncertainty in u and v chromaticity coordinates and an uncertainty in CCT for few Kelvins can be achieved. The color temperature standard achieved with the uncertainty is. ± 11.48 K for mercury lamp, ± 3.44 K for sodium lamp and ± 6.4 K for metal halide lamp). [A.B. El-Bialy, M.M. El-Ganainy and E.M. El-Moghazy, Uncertainty determination of correlated color temperature for high intensity discharge lamps. Journal of American Science 2011; 7(1):1092-1096]. (ISSN: 1545-1003). http://www.americanscience.org. Key words: lamp, correlated color temperature, Uncertainty and luminous flux. 1. Introduction: Color temperature is a characteristic of visible light that has important applications in photometry science (calibration and lighting), photography, videography, publishing, manufacturing, and other fields. The color temperature of a light source is the temperature of an ideal black-body radiator that radiates light of the same chromaticity as that light source. The temperature is usually stated in Kelvin (K). It is directly related to Planck's law and Wien's displacement law. The CIE color coordinates are derived by weighting the spectral power distribution (obtained by using a spectroradiometer). the chromaticity coordinates are usually given by normalized coordinates x and y. The (x, y) coordinates are called the chromaticity coordinates. (1) The CCT of a light source, also expressed in Kelvins, is defined as the temperature of the blackbody source that is closest to the chromaticity of the source in this case the CIE 1960 (Uniform Color Space) UCS (u, v) system is used.(2) A "modified uniform chromaticity scale diagram" suggested, based on certain simplifying geometrical considerations where (u, v) chromaticity coordinates was used instead of (x,y). This (u, v) chromaticity space became the CIE 1960 color space, which is still used to calculate the CCT. (3). Higher color temperatures (5,000 K or more) are cool (blueish white) colors, and lower color temperatures (3,000 K or lower) warm (yellowish white through red) colors. For incandescent lamp is called color temperature but for flourescent and high intensity discharge lamps is called Correlated color temperature. (4) In physics and color science, the Planckian locus is the path or locus that the color of an incandescent black body would take in a particular chromaticity space as the blackbody temperature changes. It goes from deep red at low temperatures through orange, yellowish white, white, and finally bluish white at very high temperatures. (5) In this work we have to calculate the uncertainty in u, v and CCT for one high pressure mercury lamp has symbol W1, one high pressure sodium lamp has the symbol W2, and one metal halide lamp has the symbol W3. 1092

Figure 1.The CIE 1931 x,y chromaticity space, also showing the chromaticities of black-body light sources of various temperatures (Planckian locus), and lines of constant correlated color temperature. 2. The experiment technique The measurements of CCT and u, v were done by HR 2000 spectroradimeter. The spectroradiometer system is made up of several elements: - input optics, (a source or sources, with power supplies and electrical measuring equipment). - Polychromator (monochromator)/array detector, - Data acquisition system (electronics for measuring detector output quantity combined with a data processing system). The spectroradimeter ocean optics HR 2000 Irradiance uncertainty: 4.7% 1093

Figure 2. Spectroradiometer measurements where illuminant A is used to take as reference spectrum borrowed from manual (6). In the present work we choose the one lamps of high pressure mercury 125 Watts, one lamp of high pressure sodium 150 watts and one lamp of metal halide 150 watts Such lamps have CCT from warm ( 2200 K) to cool light ( 6500 K). In the spectroradimeter measurements irradiance is total radiant flux incident on an element of surface divided by the surface area of elements in W/m 2. (7) Before any work the lamps should be seasoned until the photometric and electric characteristics remain constant. In the present work the HID lamps must be seasoned for 100 operating hours and should be cycled 11 hours on and one hour off. The metal halide and high pressure sodium lamps should be stored in the same position as seasoned. (8) 3.Theoretical background 3.1 The uncertainty of (u, v) The uncertainty in u is u c (u) = { ( u-4) 2 Σ u 2 c (E i ) x 2 i + u 2 [ 225 Σ u 2 c (E i ) y 2 i + 9 Σ u 2 c (E i ) z 2 i ] + 30u (u-4) Σ u 2 c (E i ) x i y i + 6u (u-4) Σ u 2 c (E i ) x i z i + 90 u 2 Σ u 2 c (E i ) y i z i } 1/ 2 / (Σ E i x i + 15 Σ E i y i + 3 Σ E i z i ). And similarly u c (v) = {9(5v -2) 2 Σ u 2 c (E i ) y 2 i + v 2 [Σ u 2 c (E i ) x 2 i + 9 Σ u 2 c (E i ) z 2 i ] + 6v (5v-2) Σ u 2 c (E i ) x i y i + 6 v 2 Σ u 2 c (E i ) x i z i + 18v (5v-2) Σ u 2 c (E i ) y i z i } 1/ 2 / (Σ E i x i + 15 Σ E i y i + 3 Σ E i z i ) (9) Correlated color temperature CCT: The CCT of a general source is defined the temperature of the nearest point on the Black-body locus. The standard uncertainty u c (T) in CCT is given by u c (T) = ( T/ u) 2 u 2 c (u) + ( T/ v) 2 u 2 c (v) + 2r uv ( T/ u) ( T/ v) u c (u) u c (v). (1) Where r uv is the correlation coefficient between u and v and T/ u = - 5918.47 + 9.69941 T - 0.00958899 T 2 + 1.88114x10-6 T 3-1.67343x10-10 T 4 + 5.42081x10-15. T/ v = - 385.70 +8.40689 T 0.00362952 T 2 + 3.71034 x10-8 T 3. The correlation coefficient between u and v is given by (1) is r uv = Σ ( u/ E i ) ( v/ E i ) u 2 c (E i ) / [ Σ ( u/ E i ) 2 u 2 c (E i ) Σ ( v/ E i ) 2 u 2 c (E i )] (9) x i, y i and z i are color matching functions (description of a color by the spectral concentration of a radiometric quantity such as radiance or radiant power as a function of wavelength) from 360 nm to 770 nm and obtain from standard table. Radiant power is total emitted by a light source per unit time. (7) Gardner obtains the uncertainty in CCT derived directly from systematic and random components of the spectral irradiadiance values. (10) 4. Results and discussions: By setting the lamps at their nominal voltage at distance one meter from input fiber (optics). The u,v and CCT data of each lamp obtained from the computerized spectroradimeter, tabulated in tables (1-4). In table 1 the values of u,v and CCT for each lamp. In table 2 the values of uncertainty of u,v and their squares. In table 3 we obtained the uncertainty of CCT. Finally in table 4 the values of operating voltage, current and watt for each lamp. We found that photometrically and electrically the W1 is the high uncertainty in CCT for high pressure mercury lamp, the high pressure sodium lamp has lower uncertainty and W3 for metal halide lamp is intermediate. Gardner (9) uses this method for calculating CCT for a high pressure sodium lamp reaching uncertainty of CCT for this lamp as 3.1 K assuming an uncertainty of spectral irradiance u c (E i ) is 0.01 but we measure experimentally the uncertainty of spectral irradiance of spectroradimeter is u c (E i ) = 4.7%. 5. Conclusion: For the first time in Egypt experimentally determination of the uncertainty of CCT for high intensity discharge lamps. The lamps under investigation may use as standard lamps for correlated color temperature in national institute for standards (NIS). By using the uncertainty for CCT we can obtain the uncertainty for mismatch factor, which is very important for calculation of luminous flux uncertainty. 1094

Table 1. The values of CCT and u and v were obtained by using the spectroradimeter. Lamps CCT u v W1 6036 0.236 0.327 W2 2200 0.32 0.361 W3 6306 0.235 0.333 Table 2. The values of uncertainty of u and v and their square Lamps u c (u) u c (v) u 2 c (u) u 2 c (v) W1 0.03 0.018 0.000888 0.000324 W2 0.046 0.018 0.002116 0.00032761 W3 0.015 0.009 0.000228 8.14506E-05 Table 3. The values of uncertainty of CCT for lamps Lamps T/ u T/ v ( T/ u) 2 ( T/ v) 2 Uncertainty of CCT (Kelvin) ±K W1-29488.2-17849.6 8.7E+08 3.2E+08 11.48 W2-8476.06-376.64 71843513 141858 3.44 W3-31058.5-19833 9.65E+08 3.9E+08 6.4 1095

Table 4. The values of Current, volt and power of the lamps Lamps Volt (V) Current (A) Power (W) W1 119 1.18 125 W2 91 1.71 133 W3 97 1.87 149 References 1. http://www.reefkeeping.com/issues/2006-05/sj/index.php 2. http://www.spie.org/x32415.xml 3. http: // www.answers.com/color_temperature 4. http://en.wikipedia.org/wiki/color_temperature 5. http://en.wikipedia.org/wiki/planckian_locus 6. CIE 1984 publication No. 63. International Commission on Illumination (CIE). Or the manual of instrument. 7. DeCusatis C. Handbook of applied photometry, optical society of America, Poughkeepsie, New York, 1994, 8. "IES Guide to Lamp Seasoning", Journal of illuminating engineering society, IES LM- 54-1984. 9. Gardner J.L., 2000, "Correlated color temperature- uncertainty and estimation", Metrologia, 37, 381-384 10. Gardner J.L., 2006, "uncertainties in source distribution temperature and correlated color temperature", Metrologia, 43, 403-407. 12/1/2010 1096