FAST ANALYSIS OF GABA AND GLUTAMATE

Similar documents
Applied potential +600 mv vs. Ag/AgCl

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Appendix II- Bioanalytical Method Development and Validation

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

High Performance Liquid Chromatography

DEVELOPMENT AND VALIDATION OF NEW ANALYTICAL METHOD FOR BIOACTIVE COMPOUNDS

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Luminescence transitions. Fluorescence spectroscopy

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

Evolution of Solid Phase Homochirality for a. Proteinogenic Amino Acid

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

[ a ppl ic at ion no t e ]

Chapter content. Reference

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Quality control analytical methods- Switch from HPLC to UPLC

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Shodex TM ODP2 HP series columns

GLUTATHIONE IN WHOLE BLOOD BY FLUORIMETRY CODE Z30010

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

HPLC Praktikum Skript

UPLC Method Development and Validation

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

EFFECTS OF MOBILE PHASE COMPOSITION ON THE SEPARATION OF CATECHOLAMINES BY LIQUID CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

High Pressure/Performance Liquid Chromatography (HPLC)

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

Determination of Trace Cations in Power Plant Waters Containing Morpholine

LC and LC/MS Column Selection Flow Chart

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Biochemistry. Biochemical Techniques HPLC

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

CHAPTER - IV. Acharya Nagarjuna University, Guntur 105

High Performance Liquid Chromatography

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68

Gas Chromatography. Introduction

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

How to Troubleshoot a Failed System Suitability Test

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

LC Technical Information

LUMEFANTRINUM LUMEFANTRINE

3-METHYL INDOLE (Skatole) IN URINE BY FLUORIMETRY FAST - Code Z60010

Acetylcholinesterase Assay Kit

Determination of Caffeine by HPLC

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

NORBUPRENORPHINE (Buprenorphine s Metabolite ) BUPRENORPHINE in urine by GC/MS Code GC Method of Confirmation by GC-MS

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010

Definitive EtG/EtS LC-MS/MS Analysis:

Open Column Chromatography, GC, TLC, and HPLC

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

HYDROXYPROLINE IN URINE by UV/VIS Code Z01010

Enzymatic Assay of NITRIC OXIDE SYNTHETASE (EC ) ß-NADP = ß-Nicotinamide Adenine Dinucleotide Phosphate,

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

New approach to linear gradient elution used for optimisation in reversed-phase liquid chromatography

FORMALDEHYDE IN URINE by UV Code Z65010

Mechanisms of retention in HPLC

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Routine MS Detection for USP Chromatographic Methods

Dilution(*) Chromatography

Chromatography. Gas Chromatography

ACETONE IN URINE BY UV CODE Z42010

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

Pelagia Research Library

C18 Column. Care & Use Sheet

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Fast methods for the determination of ibuprofen in drug products

penta-hilic UHPLC COLUMNS

Use of Ultra-High Performance LC-MS for Challenging Assays. David Browne, Laboratory Manager (LC-MS-MS), Covance Laboratories, Harrogate

25-OH VITAMIN D 3 AND 25-OH VITAMIN D 2 IN PLASMA BY UV / VIS FAST CODE Z19710

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

CYCLOSERINE Final text for addition to The International Pharmacopoeia. (November 2008) CYCLOSERINUM CYCLOSERINE

Cation Exchange HPLC Columns

Volume 6, Issue 2, January February 2011; Article-015

Transcription:

13_ #1 NEUROSCIENCE APPLICATION NOTE Page : 1 FAST ANALYSIS OF AND GLUTAMATE THE SMARTEST LC-EC APPLICATIONS FOR NEUROSCIENCE ANALYSIS EVER MASTERMINDED Monoamines and the metabolites Noradrenalin Dopamine Serotonin 5-hydroxyindole acetic acid (5-HIAA) 3,4-dihydroxyphenylacetic acid (DOPAC) homovanillic acid (HVA) OPA derivatized amines and amino acids and Glutamate 4-aminobutyrate () Glutamate (Glu) Choline and Acetylcholine Choline (Ch) Acetylcholine (ACh) Markers for oxidative stress 3-nitro-L-Tyrosine 8-OH-DPAT Glutathione and other thiols INTRODUCTION Certain amino acids not only serve as building blocks for proteins, but can also serve as neurotransmitters in the brain. The amino acid derivative, - aminobutyrate, also called 4- aminobutyrate, () is a well-known inhibitor of presynaptic transmission in the Central Nervous System (CNS). The activity of is increased by Valium (Diazepam) and by anticonvulsant drugs. The concentration in the brain is -1 times higher than that of the monoamines or acetylcholine. Glutamate (Glu) is an excitatory neurotransmitter and a precursor for the synthesis of in ergic neurons. Glu activates the N-methyl-Daspartate (NMDA) receptors, which play a role in learning and memory and a number of other processes. Fast separation on a sub-µm UHPLC column Fully automated in-needle OPA derivatization Post separation step-gradient to eliminate late eluters Separation of Glu and within 1 minutes Summary ALEXYS and Glu Analyzer A fast and sensitive method is presented for the analysis of the amino acid neurotransmitters and Glu in micro dialysates based on the ALEXYS Neurotransmitter analyzer. Highly efficient separation is achieved using a sub- μm particle UHPLC column. An automated pre-column derivatization with o-phthaldialdehyde (OPA) is applied in the needle without the need of a mixing vial. A quick post separation step-gradient is used to eliminate late eluting substances. With this approach excellent detection sensitivity can be achieved with minimal sample consumption. Ultra-High Performance Liquid Chromatography (UHPLC) offers advantages in chromatographic resolution, analysis speed, and sensitivity over conventional HPLC systems. The combination of Electrochemical Detection (ECD) with UHPLC is a powerful solution to increase the sample throughput and sensitivity of neurotransmitter analysis in microdialysates and brain homogenates. Fig. 1. ALEXYS Neurotransmitter analyzer for and Glu.

Glu 14.63 17 13_ #1 Fast analysis of and Glutamate Page: Page : /9 Method [na] [na] LC-ECD conditions For the analysis of and Glu in brain dialysates a pre-column derivatization with OPA and sulphite has been used [1-]. 1 8 1.1 1..9.8.7 6.6 4 6 8 1 1 Fig.. Reaction scheme of the derivatization of primary alkyl amines with OPA and sulphite. 4 The N-alkyl-1-isoindole sulphonate derivatives that are formed are analyzed using UHPLC with ECD. The two amino acid derivatives of and Glu show a large difference in retention times. Therefore, a new method was developed for and Glu using the ALEXYS neurotransmitter analyzer. A 5 cm sub- micron C18 UHPLC column was used in combination with a post-separation step-gradient. The new ALEXYS system can operate at back pressures up to 65 bar (maximum rating 7 bar). Table 1 Conditions HPLC Flow cell Temperature ALEXYS Neurotransmitters analyzer with -Glu application kit & LC step-gradient upgrade µvt-3 with.7 mm GC WE and Ag/AgCl REF 4 C (separation and detection) Column Acquity UPLC HSS T3 1. x 5 mm column, 1.8 µm Flow rate Pressure µl/min about 4 bar This new approach has several advantages: 4 6 8 1 1 Fig. 3. Example chromatogram of the analysis of and Glu in pooled rat dialysate from the Nucleus Accumbens. OPA-sulphite derivatization Reagent - The derivatization procedure and composition of the OPA reagent was modified from Smith and Sharp []. The reagent is prepared in the following way: 5 mg OPA is dissolved in 5 µl methanol in a 1 ml glass autosampler vial. 5 µl 1M sodium sulphite solution is added. Followed by addition of 4.5 ml.1 M borate buffer (adjusted to ph 1.4 with 14 M sodium hydroxide). The OPA reagent including the 1M sodium sulphite solution should be prepared fresh each day for optimal performance. Automated in-needle OPA-sulphite derivatization. Fast and efficient separation using UHPLC Post separation step-gradient to eliminate late eluters In-needle derivatization procedure - The sample derivatization is completely automated by a user program in the autosampler. The procedure is significantly improved in comparison to the method described in application note 13-19 [3]. The time required to perform the derivatization is reduced by a factor of two (duration < 4 min instead of 8 min). Furthermore, sample and reagent are mixed in the auto sampler flow path (buffer tubing) and mixing vials are not required. This has as advantage that almost all vial positions (x 96 vial plates) can be used for samples, resulting in a total sample capacity in the auto sampler of 19 samples (4 vials are reserved for reagent). The user program comprises of the following steps: Aspiration of.5 µl reagent Aspiration of 9. µl sample Mixing of sample-reagent solution Injection of the derivatized sample on column.

5.9 8 13_ #1 Fast analysis of and Glutamate Page: 3 Page: 3 The procedure is depicted in Fig. 4 in more detail. [pa] 6 5 4 3 1 4 6 8 Fig. 5. Chromatogram of a rat dialysate showing several late eluters between 15 and 6 min (red arrow). Fig. 4. Automated in needle derivatization procedure (derivatization time < 4 min). Sample/reagent ratio - The sample/reagent ratio affects both the sample dilution factor and the reproducibility. A mixing ratio of,5:9 µl reagent:sample was found to give the most optimum result in combination with the above-mentioned derivatization procedure. The transfer of small volumes are handled reproducibly. In particular the final concentration in the derivatization mixture (OPA, sulphite as well as ph) must be sufficient and is primarily affecting reproducibility. It is evident from Fig. 5 that a series of components are eluting between t = 15 and 6 minutes. These peaks will interfere in consecutive runs if no precautions are taken. To remove the strong retained interferences from the column a post-separation step-gradient is applied using a mobile phase with 6% acetonitrile after elution of (from t = 1 14 min) as shown in Fig. 6. The rate of derivatization with OPA-sulphite reagent is strongly ph dependent. At high ph (> 9.5) the reaction occurs almost instantaneous [1]. Therefore, the OPA-sulphite reagent is buffered at ph 1.4 by means of a.1 M borate buffer to assure fast conversion of the amino acids. While microdialysates are often acidified with perchloric acid (PCA) immediately after sample collection to minimize sample degradation over time, it is suggested that microdialysates are not acidified prior to OPA derivation. The acidic sample ph will decrease the ph of the sample-reagent solution, which in turn will lead to a lower conversion of sample and subsequent loss of signal/sensitivity. For acidified samples, it might be necessary to adjust the composition/buffering capacity of the OPA/sulphite reagent to be able to reach a sufficiently high ph of the mixture to be able to derivatise the amino acids neurotransmitters in the sample. Step-gradient profile No late eluters 1 3 4 5 6 Fig. 6. Chromatogram of a rat dialysate with a post-separation stepgradient. No late eluting peaks are found. Post-separation step-gradient The analysis of microdialysate samples is hampered by late eluting substances. In Fig. 5 a typical chromatogram of a 9 minutes isocratic run of a pooled rat dialysate of the Nucleus Accumbens is shown.

Area (na.sec) Area (na.sec) Area (na.sec) Area (na.sec) Glutamate Glu 13_ #1 Fast analysis of and Glutamate Page: 4 Page : 4/9 Repeatability, linearity & LOD The biologically relevant concentration can typically range from 1nM - 5 nm for [5, 6] to several µm for Glu [7, 8], depending on the brain region under investigation. [pa] 4 Ringer(black), 1 nm (blue), 5 nm (red) and 1 nm (green) std in ringer [na] 1.5 1. - -4.5-6. 4 6 8 1 1 14 Fig. 7. Overlay of 8 chromatograms of 5 nm and.5 µm Glu in Ringer. The repeatability has been studied for 5 and 5 nm standards and 5 nm, 5 nm and.5 µm Glu standards in Ringer s solution using 1.5 µl flushed loop injections. See table for typical RSD values obtained. Table Repeatability of standards, peak Area (n=8) Glu, RSD, RSD 5 nm < 5 % < 3 % 5 nm < % < %.5 µm < % Calibration curve,.5-.5 µm Glu Calibration curve, 1-5 nm 4 6 8 1 1 Fig. 9. Overlay of chromatograms of a blank (ringer), 1, 5 and 1 nm and Glu standard in Ringer. Detection limits were calculated as the concentration resulting in a signal that is 3 times the peak-to-peak noise of the baseline. With the ALEXYS system, a Limit of Detection of 1 nm for both and Glu could be reached under the specified conditions (1.5 µl injections, flushed loop). This corresponds to an amount of 15 fmol on column ( 3 pg). Total sample consumption in derivatization/injection procedure is 9 µl. Note that under the Glu peak in the blank (ringer) an unidentified interference is present. In comparison to the expected biologically relevant concentration of Glu in dialysates (µm range) the intensity of the interference is relatively small. 1 1 1 8 6 4 1 8 6 4 1 3 Concentration Glu (nm) 1 3 Concentration Glu (nm) 1.5 1.5 1.5 4 6 Concentration (nm) 4 6 Fig. 8. Calibration plots (Peak Area) of Glu and in the concentration range.5.5 µm (Glu) and 1-5 nm ().. 1.5 Concentration (nm) The linearity of the method was determined in the concentration ranges of.5 -.5 µm and 1 5 nm for Glu and respectively. The method showed a good linear detector response with correlation coefficients of.998 or better for both and Glu.

Glu Glu 13_ #1 Fast analysis of and Glutamate Page: 5 Page: 5 1 Analytical parameters [pa] 6 4 C:\113\pH = 3. - Detector 1 C:\113\pH = 3.1 - Detector 1 C:\113\pH = 3. - Detector 1 C:\113\pH = 3.3 - Detector 1 C:\113\pH = 3.4 - Detector 1 C:\113\pH = 3.5 - Detector 1 C:\113\pH = 3.6 - Detector 1 C:\113\pH = 3.7 - Detector 1 C:\113\pH = 3.8 - Detector 1 C:\113\pH = 3.9 - Detector 1 C:\113\pH = 4. - Detector 1 ph 3.5, 4% ACN - -4 ph 3.5, % ACN -6 5 1 15 Fig. 1. Optimizing the separation using ph: overlay of chromatograms recorded in the ph range 3 4.of a & Glu standard mixture (T=35 C, separation & detection). Several analytical parameters have been investigated during method optimization. Both modifier concentration and ph of the mobile phase are the most important parameters for tuning of the chromatography in case baseline separation is not achieved for specific microdialysis samples. ph In Fig. 1 the influence of ph on the separation of and Glu is shown. The ph was varied in the range 3 to 4 with steps of.1 ph units. At higher ph is more strongly retained while Glu behaves opposite, it is eluting faster. It is evident that the ph is a powerful tool to tune the chromatography. With the ALEXYS -Glu analyzer, which can operate as a high-pressure gradient system, such optimization can be performed in an automated procedure, using two mobile phases (A and B) with identical contents but different ph. By changing the mixing ratio of the two mobile phases in the gradient method one can control the ph used for the separation. Modifier concentration The percentage modifier affects retention of all sample components, increasing the percentage modifier will decrease the retention times. Acetonitrile (ACN) is preferred as modifier above methanol as it will not increase the mobile phase viscosity [8] and system pressure. 5 1 15 Fig. 11. Overlay of two sets of chromatograms recorded with different modifier concentration ( and 4% ACN). Red: pooled rat dialysate from the Hippocampus. Blue: 5 µm & Glu standard mixture in Ringer. (T=35 C, separation & detection). Fig. 11 demonstrates that both the ph the modifier concentration are useful parameters for optimizing of the chromatography in case of co-elution of interferences in microdialysates. Also optimization of the modifier concentration can be performed in an automated way with the ALEXYS -Glu analyzer. A ph of 3.5 in combination with a modifier concentration of % ACN was found to be the optimum. Temperature Another parameter to take into consideration with respect to optimizing of the separation is the temperature. At higher temperatures components will elute faster, increasing the analysis speed. However it can also result in poorer separation. The oven temperature of the ALEXYS analyzer can be varied between 7 C above room temperature and 45 C and can be used as a parameter for optimisation of the separation if necessary. For this method a temperature of 4 C was chosen as the optimum with respect to separation versus analysis speed. Performance In Fig. 1 a chromatogram and peak table of a 5 nm and Glu standard in Ringer s solution is shown to illustrate the performance of the system with standards under optimized LC conditions.

8.9 3.69 Glu 1 13_ #1 Fast analysis of and Glutamate Page: 6 Page : 6/9 [pa] 1 C O N C L U S I O N 5 A fast and sensitive method is presented for the analysis of the amino acid neurotransmitters and Glu in micro dialysates based on the new ALEXYS neurotransmitter analyzer. The system combines good chromatographic performance with ease of use. The method includes an improved fully automated derivatization and a post-separation step-gradient to eliminate late eluters. -5 4 6 8 1 Fig. 1. Chromatogram of a.5 µm & Glu standard mixture in Ringer (1.5 µl injection). Separation at oven temperature of 4 C. Table 3 Peak table,.5 µm & Glu standard in Ringer Compound Name Glu Reten. [min] 3.69 8.9 Area [na.s] 1.59 1.41 Height [na].19.85 Capacity [-] 9.75 73.33 Asymmetry [-] 1.43 1.14 Eff/l [t.p./m] 9164 131 Typically for standards, efficiencies in the range of 7, 9, t.p/m for Glu and 1, 13, for are found. Analysis of microdialysates During method development several microdialysate samples were analysed to check the performance with real samples. Pooled basal-level rat microdialysates of different brain regions (Nucleus Accumbens and Hippocampus) were provided by Mr. Niels Leguit, Abbot Healthcare Products B.V., Weesp, the Netherlands. The samples were obtained by dialysis of 8 test animals for 16 hours at a flow rate of µl/min using perfusion fluid consisting of 147 mm NaCl, 4. mm KCl, 1. mm MgCl and.7 mm CaCl. After a sterility check, all samples (per brain region) were pooled and frozen at 8 C until analysis. References 1. W.A. Jacobs, o-phthalaldehyde-sulphite derivatization of primary amines for liquid chromatography-electro chemistry, J. Chromatography 39 (1987) 435-441. S. Smith, T. Sharp, Measurement of in Rat Brain Microdialysates Using o-phthaldialdehyde Sulphite Derivatization and High- Performance Liquid Chromatography with Electrochemical Detection, J. Chromatography B 65 () (1994) 8-33 3. Antec Leyden, Application note Gaba and Glutamate; Appendix: AS user program & settings, document 13-19 & 13-19A 4. Antec Leyden, Installation guide LC connection kit DCC I-I CS, document 18.718, (6) 3-9 5. H. L. Rowley, K. F. Martin, C. A. Marsden, Determination of in vivo amino acid neurotransmitters by high-performance liquid chromatography with o-phthalaldehyde-sulphite derivatization, J. of Neurosc. Meth. 57 (1) (1995) 93-99 6. S. Zhang, Y. Takeda, et. al., Measurement of and Glutamate in vivo levels with high sensitivity and frequency, Brain Research Protocols 14 (5) 61-66 7. J. Kehr, Determination of glutamate and aspartate in microdialysis samples by reversed-phase column LC with fluorescence and electrochemical detection, Journal of Chromatography B. 78 (1998) 7-38 8. L.R. Snyder, J.J. Kirkland, J.W. Dolan, Introduction to Modern Liquid Chromatography, Wiley, 3rd ed. (1) An example chromatogram of the analysis of and Glu in pooled rat dialysate from the Nucleus Accumbens is shown in Fig. 3. The insert in the the top-right corner is a zoom in on the peak. In Fig. 11 chromatograms are shown of pooled hippocampus rat dialysate (red curve). For the rat dialysate from the hippocampus a concentrations of 1.9 µm Glu and 1 nm was measured.

13_ #1 Fast analysis of and Glutamate Page: 7 Page: 7 Fig. 13. ALEXYS Neurotransmitter analyzer for & Glu based on the modular ALEXYS neurotransmitters base system combined with dedicated -Glu application kit and LC step-gradient upgrade. PART NUMBERS AND CONFIGURATIONS 18.91U ALEXYS Neurotransmitter Analyzer 18.5 ALEXYS -Glu kit 18.6 LC step-gradient upgrade 5.116 Acquity UPLC HSS T3 1. x 5 mm column, 1.8 µm 5.17 In-line filter (aqueous)

13_ #1 Fast analysis of and Glutamate Page: 8 Page : 8/9