Ultrasonic Measurement of Minute Displacement of Object Cyclically Actuated by Acoustic Radiation Force

Similar documents
Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

Modified Phased Tracking Method for Measurement of Change in Thickness of Arterial Wall

D. BARD DIVISION OF ENGINEERING ACOUSTICS, LUND UNIVERSITY

Vibro-Acoustography and Vibrometry for Imaging and Measurement of Biological Tissues

Workshop 2: Acoustic Output Measurements

Noninvasive Measurement of Stiffness and Density of Bone for Its Diagnosis Using Ultrasound

Shear waves in solid-state materials

The Physics of Doppler Ultrasound. HET408 Medical Imaging

Structure of Biological Materials

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Today s menu. Last lecture. Ultrasonic measurement systems. What is Ultrasound (cont d...)? What is ultrasound?

Acoustic wave reflection from the transition layer of surficial marine sediment

Chapter 2 Surface Acoustic Wave Motor Modeling and Motion Control

Microwave-induced thermoacoustic tomography using multi-sector scanning

Today s menu. Last lecture. Measurement of volume flow rate. Measurement of volume flow rate (cont d...) Differential pressure flow meters

ESTIMATION OF LAYER THICKNESS BY THE COST FUNCTION OPTIMIZATION: PHANTOM STUDY

Classification of services in Acoustics, Ultrasound and Vibration

Measurement of shear viscoelasticity using dual acoustic radiation pressure induced by continuous-wave ultrasounds

Lesson 05: Intensity Measurements and Bioeffects. This lesson contains 18 slides plus 6 multiple-choice questions.

1. Background and Principle of Elastography

Technical University of Denmark

Vibro-acoustography: An imaging modality based on ultrasound-stimulated acoustic emission

Research on sound absorbing mechanism and the preparation of new backing material for ultrasound transducers

Echocardiography is a widely used modality for. Effect of Subaperture Beamforming on Phase Coherence Imaging. Hideyuki Hasegawa and Hiroshi Kanai

Sound wave bends as it hits an interface at an oblique angle. 4. Reflection. Sound wave bounces back to probe

Friction Drive Simulation of a SAW Motor with Slider Surface Texture Variation

PROPERTY STUDY ON EMATS WITH VISUALIZATION OF ULTRASONIC PROPAGATION

SCATTERING OF ULTRASONIC WAVE ON A MODEL OF THE ARTERY J. WÓJCIK, T. POWAŁOWSKI, R. TYMKIEWICZ A. LAMERS, Z. TRAWIŃSKI

Evaluation of a surface acoustic wave motor with a multi-contact-point slider

On the beam deflection method applied to ultrasound absorption measurements

Pixel-based Beamforming for Ultrasound Imaging

Non-Destructive Testing of Concrete Based on Analysis of Velocity Dispersion of Laser Ultrasonics

19 th INTERNATIONAL CONGRESS ON ACOUSTICS MADRID, 2-7 SEPTEMBER 2007

Shear Wave Propagation in Soft Tissue with Ultrasound Vibrometry

DAMPING CONTROL OF A PZT MULTILAYER VIBRATION USING NEGATIVE IMPEDANCE CIRCUIT

MULTIFREQUENCY ULTRASOUND RADIATION FORCE EXCITATION AND MOTION DETECTION OF HARMONICALLY VIBRATING SCATTERERS

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

DETERMINATION OF THE B/A OF BIOLOGICAL MEDIA BY MEASURING AND MODELING NONLINEAR DISTORTION OF PULSED ACOUSTIC WAVE IN TWO-LAYER MEDIA TAMARA KUJAWSKA

UNIVERSITY OF SOUTHAMPTON

Nondestructive Determination of Elastic Constants of Thin Plates Based on PVDF Focusing Ultrasound Transducers and Lamb Wave Measurements

ISUOG Basic Training The Principles of Doppler Ultrasound. Juriy Wladimiroff

Physics and Knobology

An Energy Circulation Driving Surface Acoustic Wave Motor

Physical and Biological Properties of Agricultural Products Acoustic, Electrical and Optical Properties and Biochemical Property

Lamb Wave Behavior in Bridge Girder Geometries

Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2006 Ultrasound Lecture 1

N db compared with that in the single pulse harmonic imaging mode, whereas

6th NDT in Progress Lamb waves in an anisotropic plate of a single crystal silicon wafer

Basic System. Basic System. Sonosite 180. Acuson Sequoia. Echo occurs at t=2z/c where c is approximately 1500 m/s or 1.5 mm/µs

Caution! Pay close attention to the special operation and safety instructions in the manual of the ultrasonic echoscope.

Acoustic Scattered Field Computation

ANALYSIS OF FREQUENCY CHARACTERISTICS ON NON-INVASIVE ULTRASONIC-DOPPLER FLOW MEASUREMENT FOR METAL PIPES

Fast simulation of nonlinear radio frequency ultrasound images in inhomogeneous nonlinear media: CREANUIS

Measurements of acoustic properties of aqueous dextran solutions in the VHF/UHF range

Impedance of a sphere oscillating in an elastic medium with and without slip

Mandatory Assignment 2013 INF-GEO4310

Noncontact Human Force Capturing based on Surface Hardness Measurement

Angular Spectrum Decomposition Analysis of Second Harmonic Ultrasound Propagation and its Relation to Tissue Harmonic Imaging

Lamb Waves in Plate Girder Geometries

13th International Symposium on Nondestructive Characterization of Materials (NDCM-XIII), May 2013, Le Mans, France

Numerical analysis of sound insulation performance of double-layer wall with vibration absorbers using FDTD method

Magnetoacoustic tomography with current injection

Chapter 1. The Nature of Sound. The Nature of Sound

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm.

Nanoacoustics II Lecture #2 More on generation and pick-up of phonons

IMAGING THE MECHANICAL PROPERTIES OF TISSUE WITH ULTRASOUND: AN INVESTIGATION OF THE RESPONSE OF SOFT TISSUE TO ACOUSTIC RADIATION FORCE

High Intensity Focused Ultrasound Test & Measurement Techniques

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

EE 5345 Biomedical Instrumentation Lecture 6: slides

For more information, please contact

Measurement of liquid surface properties by laser-induced surface deformation spectroscopy

ACOUSTIC EFFECTS OCCURRING IN OPTICAL BREAKDOWN WITH A LIQUID BY LASER RADIATION

Chapter 2. Interaction with Soft Tissue

Modeling shear waves through a viscoelastic medium induced by acoustic radiation force

MODELLING AND SIMULATION OF ULTRASOUND NON LINEARITIES MEASUREMENT FOR BIOLOGICAL MEDIUMS

31545 Medical Imaging systems

On the study of elastic wave scattering and Rayleigh wave velocity measurement of concrete with steel bar

Phase-Sensitive Lateral Motion Estimator for Measurement of Artery-Wall Displacement Phantom Study

ACOUSTIC TRANSMISSION WITH MODE CONVERSION PHENOMENON

Experimental approach on natural frequency of window vibration induced by low frequency sounds

Donald School Journal of Ultrasound Physical Bases in Obstetrics of Medical and Ultrasound Gynecology, April-June 2009;3(2):1-9

Development of water immerse type ultrasound probe with PZT film deposited by hydrothermal method

Validation of FEM simulation of EMATs for versatile EMAT configurations

Noncontact measurement technique for wide range of viscosity of µl-order liquid sample

Change in Ultrasonic Backscattered Energy for Temperature Imaging: Factors Affecting Temperature Accuracy and Spatial Resolution in 3D

Medical Imaging Physics Spring Quarter Week 3-2

UNCERTAINTIES PROPAGATION STUDY OF THE TRANSIT TIME METHODS IN ULTRASONIC ANEMOMETERS.

Ultrasonic Liear Motor using Traveling Surface Acoustic Wave

Research of a novel fiber Bragg grating underwater acoustic sensor

Simulation of Contrast Agent Enhanced Ultrasound Imaging based on Field II

CROSS-TALK RESPONSE ANALYSIS ON A PIEZOELECTRIC ULTRASONIC MATRIX ARRAY

Surface Acoustic Wave Atomizer with Pumping Effect

31545 Medical Imaging systems

Electrical Engineering 3BA3: Structure of Biological Materials

ACOUSTIC EMISSION SOURCE DETECTION USING THE TIME REVERSAL PRINCIPLE ON DISPERSIVE WAVES IN BEAMS

Outline of today Medical Imaging systems. Wave types. 1. Discussion assignment on B-mode imaging

A method of the forward problem for magneto-acousto-electrical tomography

Supplement (videos)

Absolute Measurement and Relative Measurement of Ultrasonic Nonlinear Parameters

Transcription:

Jpn. J. Appl. Phys. Vol. 42 (2003) pp. 4608 4612 Part 1, No. 7A, July 2003 #2003 The Japan Society of Applied Physics Ultrasonic Measurement of Minute Displacement of Object Cyclically Actuated by Acoustic Radiation Force Kazuaki MICHISHITA, Hideyuki HASEGAWA and Hiroshi KANAI Graduate School of Engineering, Tohoku University, Sendai 980-8579, Japan (Received November 18, 2002; accepted for publication March 12, 2003) Recently, several researchers have investigated the potential of the use of acoustic radiation force for imaging the mechanical properties of tissue. However, these methods have some problems as to safety and the spatial resolution. Therefore, an alternative ultrasonic remote actuation and measurement method is proposed in this paper, in which the intensity of the applied continuous ultrasonic wave is suppressed to be lower than the safety guideline (1 W/cm 2 ) recommended by the Japan Society of Ultrasonics in Medicine (JSUM). A minute displacement with amplitude of less than 1 mm is cyclically generated by the radiation force with a low frequency f of several hertz. For simultaneous measurement of the minute displacement, cyclic radiation force is applied intermittently by maintaining its envelope waveform of the low-frequency component of f Hz. At the same time, an ultrasonic correlation-based method, namely, the phased tracking method, is employed to measure the minute displacement. In basic experiments, the minute displacement of several micrometers was generated in silicone rubber by applying radiation force, and it was successfully measured by the phased tracking method. [DOI: 10.1143/JJAP.42.4608] KEYWORDS: acoustic radiation force, noninvasive measurement, measurement of minute displacement, viscoelasticity measurement, tissue characterization 1. Introduction In recent years, some remote actuation methods based on acoustic radiation force have been reported. Fatemi and coworkers proposed an imaging method called ultrasoundstimulated acoustic emission (USAE). 1,2) Their system consists of two confocal ultrasonic transducers, and two ultrasound beams with two slightly different frequencies f and (f þ f) are transmitted. Acoustic radiation pressure P R ðtþ exerted on the interface between different media is a function of the energy density, eðtþ, and the specific acoustic impedances, Z 1 and Z 2, of the medium. 3) The energy density, eðtþ, is proportional to the square of the sum of the sound pressures, p 1 ðtþ and p 2 ðtþ, generated by the two transducers. In the intersectional area of the two beams, therefore, an oscillatory radiation pressure P R ðtþ with the frequency difference, f, is applied to the interface. The radiation force produces acoustic emission which is closely related to the mechanical frequency response of the medium. By measuring the acoustic emission with a hydrophone, hard inclusions in a soft material were experimentally detected. However, the spatial resolution in the depth direction corresponds to the size of the intersectional area. Nightingale et al. proposed an alternative imaging method in which the focused ultrasound is employed for applying the radiation force to a soft tissue during short durations (less than 1 ms). The viscoelastic properties of the tissue are investigated from the magnitude and the transient response of the displacement, dðtþ, of the tissue. 4) In order to generate measurable displacement, dðtþ, using several ultrasonic pulses, high-intensity pulsed ultrasound of 1,000 W/cm 2 is employed. According to the safety guideline shown by JSUM, however, the intensity of ultrasound is recommended to be less than 240 mw/cm 2 and 1 W/cm 2, for the pulsed wave and the continuous wave, respectively. 5) Therefore, the intensity of the pulsed ultrasound employed by Nightingale et al. is far greater than that indicated in the safety guideline. In this study, in order to apply ultrasound within the range of the safety guideline and to detect the minute displacement, dðtþ, generated by the resultant acoustic radiation force, the two continuous-wave (CW) ultrasound beams of f and f þ f are employed to cyclically actuate the acoustic radiation pressure at low frequency (f) to increase the signal-to-noise ratio. For the measurement of the resultant minute displacements, an ultrasound correlation-based method, the phased tracking method, 6) is employed. By sweeping the low frequency f, the frequency characteristics of the object are determined. Moreover, from the amplitude characteristics of the displacement, dðtþ, continuously generated by the acoustic radiation pressure, P R ðtþ, the viscoelasticity of the object can be determined. We constructed an experimental setup and confirmed the measurement of the minute cyclic displacement by the phased tracking method using several micrometers. 2. Principle When continuous plane-wave ultrasound is incident on an interface between two different media, as illustrated in Fig. 1, a constant force, which is called an acoustic radiation force, is exerted on the interface. Acoustic radiation pressure, P R ðtþ, is defined as the radiation force per unit area of the object, as follows: P R ðtþ ¼ð1 þ R 2 ÞheðtÞi; ð2:1þ where R and heðtþi are the reflection coefficient of sound pressure and the temporally averaged energy density at the E-mail address: hkanai@ecei.tohoku.ac.jp Fig. 1. The acoustic pressure pðtþ of the incident wave exerts the acoustic radiation pressure, P R ðtþ, on an object. 4608

Jpn. J. Appl. Phys. Vol. 42 (2003) Pt. 1, No. 7A K. MICHISHITA et al. 4609 interface, respectively. In eq. (2.1), the incident wave through the interface is assumed to be perfectly absorbed into the object. Using the density, 1 and 2, and sound speed, c 1 and c 2, of each medium, the reflection coefficient, R, and temporally averaged energy density, heðtþi, are defined by R ¼ Z 2 Z 1 Z 2 þ Z 1 ¼ 2c 2 1 c 1 2 c 2 þ 1 c 1 ; Z T ð2:2þ heðtþi ¼ 1 1 1 c 2 fpðtþg 2 dt; ð2:3þ 1 T 0 where pðtþ is the sound pressure at the interface and T is the averaging period. JSUM shows the safety guideline, in which the intensity of the CW ultrasound should be less than 1 W/cm 2. By assuming that density 1 and sound speed c 1 of water are 10 3 kg/m 3 and 1,500 m/s, respectively, the acoustic radiation pressure, P R ðtþ, exerted on the interface of the totally absorbing object is calculated as 6.67 Pa when the ultrasound intensity is 1 W/cm 2. The energy density, eðtþ, of the incident wave is proportional to the square of the sound pressure pðtþ of the ultrasound beam. When two ultrasonic beams with slightly different frequencies, f and f þ f, cross each other, the sound pressure, p sum ðtþ, at the interface is expressed by the sum of sound pressure of two ultrasonic beams as follows: p sum ðtþ ¼p 0 cos!t þ p 0 cosð! þ!þt; ð2:4þ where p 0,!, and!, are the amplitude of the sound pressure pðtþ of each ultrasound beam, angular frequency of the incident wave (! ¼ 2f), and difference in angular frequency (! ¼ 2f), respectively. For this case, the energy density eðtþ is expressed by eðtþ ¼ 1 2 1 c 2 p sum ðtþ 1 ¼ 1 1 c 2 1 ¼ p2 0 1 c 2 1 2 p 0 cos!t þ p 0 cosð! þ!þt 1 þ cos!t þ cosð2! þ!þt displacements dðtþ generated by the radiation force, the viscoelasticity of the object can be characterized. 3. Experimental Setup An experimental setup is shown in Fig. 2. In order to measure the displacement, dðtþ, we employed ultrasonic diagnostic equipment (Toshiba SSH-160A) with a sector array probe (center frequency: 5 MHz). The equipment was modified to detect the minute displacement dðtþ of an object by the phased tracking method. The object (45 40 10 mm silicone rubber; static elastic modulus E S was measured as 20 kpa) was placed at the bottom of a water tank, and the ultrasonic probe was fixed 5 cm away from the object. For the application of radiation pressure, P R ðtþ, an ultrasonic transducer (Tokimec 5Z10I, center frequency: 5 MHz) was driven by a sum of two CWs with two slightly different frequencies (5 MHz and 5 MHz + f) generated by a function generator (Sony Tektronix AFG2020). When CW ultrasound is employed for actuation, interference occurs between the CW ultrasound for actuation and the pulsed ultrasound for displacement measurement. In order to avoid this interference, it is necessary to stop actuation during transmission and reception of this ultrasonic pulse for measurement. Therefore, we made an electrical switch to control the cessation of the CW ultrasound for actuation. Figure 3 shows a time chart for actuation and measurement. Only for actuation period (A) when the control signal in Fig. 3(b) was turned on, did the electrical switch pass the signals for actuation. Then, the signal was amplified and the ultrasound for actuation was transmitted from the transducer during period (A). It is necessary to set the length of period (A) so that the successively transmitted þ 1 2 cos 2!t þ 1 2 cos 2ð! þ!þt : ð2:5þ From the second term of the right-hand side of eq. (2.5), it is found that the energy density eðtþ of the incident field has a component with the frequency difference f. Therefore, the cyclically oscillatory radiation pressure, P R ðtþ, at the frequency difference f of eq. (2.1) is given by P R ðtþ ¼ð1þR 2 Þ p2 0 1 c 2 cos!t: 1 ð2:6þ Thus, using the two crossed ultrasound beams at the slightly different frequencies, f and f þ f, the oscillatory radiation force can be cyclically applied to the intersectional area of the beams. The amplitude and phase characteristics of the displacement, dðtþ, which is generated as a response to the radiation force, relate to the viscoelasticity of the target. By sweeping the frequency difference f and detecting the Fig. 2. Experimental setup for measurement of the displacement of an object cyclically actuated by ultrasound.

4610 Jpn. J. Appl. Phys. Vol. 42 (2003) Pt. 1, No. 7A K. MICHISHITA et al. Fig. 3. Time chart for actuation and measurement. (a) Timing clock from the diagnostic scanner. (b) Period (A) of actuation controlled by the electrical switch. (c) The sound pressure, p sum ðtþ, consisting of the sum of two ultrasound beams with slightly different frequencies, f and f þ f, applied intermittently only during actuation period (A). (d) The acoustic radiation pressure, P R ðtþ, generated by the sound pressure in Fig. 3(c). (e) Timing of the transmission of the ultrasonic pulse for measurement from the probe. (f) The sampling gate of the RF echo data in probe period (P). measurement pulse is prevented from interfering with reverberations of the ultrasound for actuation. Therefore, as shown in Fig. 3(c), the ultrasound for actuation was transmitted intermittently, and then the radiation pressure P R ðtþ became a series of pulses whose envelope consisted of the sinusoidal wave with the frequency difference f, as shown in Fig. 3(d). At the rising edge of the trigger pulse in Fig. 3(e), on the other hand, an ultrasonic pulse was transmitted from the ultrasonic probe in Fig. 2. In order to track the displacement dðtþ of the object, the quadrature-demodulated signals of the RF echo data were A/D converted during the gates shown in Fig. 3(f). In order to avoid interference between the ultrasound for actuation and that for measurement, the data which were stored in the periods other than probe period (P) were not used for the measurement of the displacement dðtþ of the object. 4. Experimental Results Figure 4(a) shows the M-mode image of the object. In Fig. 4, the difference frequency f ¼ 7 Hz was employed. Acoustic radiation pressure, P R ðtþ, shown in Fig. 4(b) was calculated based on eq. (2.6) as follows: The density, 2, and the sound speed, c 2, of the object were measured as 0:98 10 3 kg/m 3 and 1:0 10 3 m/s, respectively, in a separate experiment. By assuming the density, 1, and the sound speed, c 1, of water to be 1:0 10 3 kg/m 3 and 1:5 10 3 m/s, respectively, the pressure reflection coefficient, R, and the energy reflection coefficient, R 2, were respectively calculated as 0:21 and 0.044 using eq. (2.2). Fig. 4. The results of the experiment with the frequency difference f ¼ 7 Hz. (a) M-mode image of the object. The black line at the depth of x ¼ 7:725 mm shows the tracked displacement. (b) The estimated acoustic radiation pressure, P R ðtþ, given by eq. (2.6). (c) The vibration velocity of the object, v A ðtþ, obtained by the phased tracking method. (d) The displacement of the object, dðtþ. By measuring the acoustic field of the ultrasound for actuation with a hydrophone (Force Institute, MH-28-10), the amplitude of the spatially averaged sound pressure, p 0, was obtained beforehand and plotted as a function of the voltage applied to the transducer, as shown in Fig. 5. In this study, the amplitude of the applied voltage, V 0, was 32 V. Therefore, by neglecting the attenuation of the medium surrounding the object, the component of the sound pressure, p 0, which is perpendicular to the interface of the object, was calculated as 61 kpa. The absorption coefficient of the object was also measured beforehand. Figure 6(a) shows the measurement system used. The ultrasonic transducer was driven by a sum of two CWs with slightly different frequencies (5 MHz and 5 MHz + 10 Hz, the amplitude of the applied voltage, V 0, for each CW: 30 V). The sound pressure, p 0, of the ultrasound was measured with the hydrophone which was fixed at 5 cm away from the transducer. As shown in Fig. 6(b), in the case of interposing the object between the transducer and the hydrophone, the measured sound pressure, p 1 ðtþ, was Fig. 5. Spatially averaged sound pressure p 0 transmitted from the transducer as a function of the voltage applied to the transducer.

Jpn. J. Appl. Phys. Vol. 42 (2003) Pt. 1, No. 7A K. MICHISHITA et al. 4611 Fig. 6. (a) The experimental setup for measurement of the absorption coefficient of the object. (b) The sound pressure, p 1 ðtþ, was measured using a hydrophone with the object interposed between the transducer and the hydrophone. The sound pressure, p 2 ðtþ, was measured without the object. (c) Ultrasonic intensities, I 1 ðtþ and I 2 ðtþ, calculated from p 1 ðtþ and p 2 ðtþ, respectively. smaller than that, p 2 ðtþ, without the object due to attenuation by the object. Figure 6(c) shows the ultrasonic intensities, I 1 ðtþ and I 2 ðtþ, calculated from sound pressures, p 1 ðtþ and p 2 ðtþ, respectively. Temporally averaged intensities of I 1 ðtþ and I 2 ðtþ were calculated as 0.23 W/cm 2 and 8:8 10 3 W/ cm 2, respectively. The absorption coefficient,, is defined by the ratio of the transparent intensity, I 2 ðtþ, to the incident intensity, I 1 ðtþ. Thus, the absorption coefficient,, was calculated as 0.038. Since the acoustic intensity was attenuated by about 96% by the object, the object can be assumed to be a totally absorbing material. From these results, by assuming the object to be a totally absorbing material (R ¼ 0), the acoustic radiation pressure, P R ðtþ, exerted on the interface of the object was calculated as 1.7 Pa, using eq. (2.6) as shown in Fig. 4(b). The vibration velocity, vðtþ, of the object actuated by the acoustic radiation force was measured by the phased tracking method, as shown in Fig. 4(c). By integrating the velocity vðtþ, the displacement of the object, dðtþ, was Fig. 7. For each frequency difference f, three experimental results are shown: the estimated acoustic radiation pressure P R ðtþ given by eq. (2.6), the vibration velocity vðtþ of the object measured by the phased tracking method, and the displacement dðtþ of the object. (a) f ¼ 7 Hz. (b) f ¼ 9 Hz. (c) f ¼ 11 Hz. (d) f ¼ 13 Hz. obtained as shown in Fig. 4(d). In Fig. 4(d), it is seen that the object was cyclically actuated at the frequency difference, f, of 7 Hz with an amplitude of the order of a few micrometers. For the frequency difference, f, swept from 7 Hz to 13 Hz, the results are shown in Fig. 7. For each f, an almost cyclic component with f can be found in the measured displacement. By assuming the Voigt model to be a viscoelastic model of the object, the absolute value of the complex elastic modulus, je j, of the object is given by je j¼je S þ j!j ¼ P Rmax d max d 0 ; ð4:1þ where P Rmax and d max are the amplitude of the radiation pressure P R ðtþ and that of the change in thickness from the the initial thickness, d 0, of the object (d 0 ¼ 10 mm), respectively. E S and are the static elastic modulus and the viscosity coefficient, respectively. In Fig. 8, the absolute value of the complex elastic modulus, je j, of the object is plotted as a function of the frequency difference, f. Circles and diamonds show the results measured by the proposed method and using the laser

4612 Jpn. J. Appl. Phys. Vol. 42 (2003) Pt. 1, No. 7A K. MICHISHITA et al. Fig. 8. The absolute value of the complex elastic modulus, je j, of the object as a function of the frequency difference f, obtained by eq. (4.1). Circles and diamonds show the results measured by the proposed method and using the laser Doppler velocimeter, respectively. Doppler velocimeter, respectively. In eq. (4.1), the absolute value of the complex elastic modulus, je j, gradually increases with the frequency difference, f, which reflects the viscoelastic properties of the object. 5. Conclusion In this study, we constructed an experimental setup for cyclic remote actuation of the object and showed the measurement results of minute displacements, dðtþ, in the micrometer order actuated by the acoustic radiation pressure, P R ðtþ, by using the phased tracking method. A single transducer was driven by the sum of the two continuous waves having slightly different frequencies, f and f þ f. The duration of the ultrasound for remote actuation was controlled to prevent ultrasonic displacement measurement from the interference. The frequency characteristics of the absolute value of the complex elastic modulus je j calculated from the resultant displacement, dðtþ, generated by the radiation pressure, P R ðtþ, showed the viscoelastic properties of the object. From these results, it was shown that the proposed method has the potential to accurately measure the viscoelasticity of a regional area of the object. 1) M. Fatemi, L. E. Wold, A. Alizad and J. F. Greenleaf: IEEE Trans. Med. Imag. 21 (2002) 1. 2) M. Fatemi and J. F. Greenleaf: Proc. Natl. Acad. Sci. USA 96 (1999) 6603. 3) G. R. Torr: Am. J. Phys. 52 (1984) 402. 4) K. Nightingale, M. S. Soo, R. Nightingale and G. Trahey: Ultrasound Med. Biol. 28 (2002) 227. 5) JSUM: Jpn. J. Med. Ultrason. 11 (1984) 41 [in Japanese]. 6) H. Kanai, M. Sato, Y Koiwa and N. Chubachi: IEEE Trans. UFFC 43 (1996) 791.