Infrared / K Thermometer Remote Temperature Measurement ( Non-Contact ) 1327K INSTRUCTION MANUAL

Similar documents
HL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual

INFRARED THERMOMETER TM0819

IR F ( C) ENGLISH INSTRUCTION MANUAL. Thermometer

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration

Stick-Style Water-Resistant Infrared Thermometer for Foodservice Applications with 4:1 Distance-to-Sight Ratio Model

Model ST-616CT. Instruction Manual. Infrared Thermo-Hygrometer. reedinstruments. www. com

InfraRed Thermometer with Laser Pointer

INSTRUCTION MANUAL INF Fax: (503)

Introduction Cautions

INFRARED THERMOMETER Model : TM-959

INSTRUCTION MANUAL DTH Fax: (503)

Instruction Manual SEFRAM Thermal Camera

Gun-Type Professional Infrared Thermometer MODEL: ST 653. Instruction Manual

Dual laser + type K thermometer INFRARED THERMOMETER

EM90 INFRA-RED THERMOMETER & THERMOCOUPLE METER

IR 80-H. Non-Contact Infrared Temperature Sensor/Transmitter. Temperature Range: 500~1700

566/568. Infrared Thermometers. Getting Started

DIGITAL THERMOMETER WITH TRICOLOR LCD INSTRUCTION MANUAL

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000

Infrared Thermometer Model: and

GS-IR 8816 Series Infrared Thermometer

Non - Contact Forehead Infrared Thermometer. Care Q7 USER MANUAL - 1 -

63/66/68. Infrared Thermometers. Users Manual

IR 40. Non-Contact Infrared Temperature Sensor/Transmitter. Temperature Range: 0~1000

INSTRUCTION MANUAL Fax: (503)

DIAGNOSTICS NO TOUCH FOREHEAD THERMOMETER

New! 2010 Model. Model No

FLT5A FLT5A. abc. abc. Thermometer User Guide. Thermometer User Guide. Internet:

Infrared Thermometer Model CA879

DIGITAL THERMOMETER OPERATOR S MANUAL

User s Guide. Shown with Built-in Laser Sighting. OS531, OS532, OS533, OS534 OMEGASCOPE Handheld Infrared Thermometer

DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual

MN-ISO-2000FN-E. ISO-2000FN COATING THICKNESS GAGE OPERATION MANUAL

Noncontact Thermometer

RADIATION-METER TM-91/TM-92

Operating instructions Infrared temperature sensor TW20xx TW21xx

INSTRUCTION MANUAL RS 51 & RS 52 DIGITAL THERMOMETER

IR-712 IR-712-EUR 12:1 Infrared Thermometer. IR-720 IR-720-EUR 20:1 Infrared Thermometer. IR-730 IR-730-EUR 30:1 Infrared Thermometer.

VIBRATION TACHOMETER OPERATION MANUAL. 3 in 1. Model : VT Your purchase of this

Manual Dew point scanner PCE-DPT 1

D501 and D501-RS. Microscanner D-Series IR Thermometers. no effect. no effect. no effect. no effect no effect

R I T. Title: GCA Stepper Operations. Semiconductor & Microsystems Fabrication Laboratory Revision: F Rev Date: 08/09/ SCOPE

Infrared Hand-Held Thermometers

IT Series Infrared thermometer Using manual

Bright Advance Corporation

IR/Thermistor Pen. Instruction Manual

Hot solutions for temperature measurement. Contact and non-contact thermometers from Fluke

ABB temperature measurement Radiation thermometry. Measurement made easy. Process temperature measurement practice--non-contacting

Coating Thickness Gauge TW-8. art. nr Thoma Tel./Fax: Operation Manual

Data Sheet / Instructions. Dewpoint MeterO

INF195 Infrared Thermometer

Every object with a temperature above 0 Kelvin= -273,15 C emits

Dewpoint Meter Data Sheet / Instructions

Mini Environmental Quality Meter

Infrared Hand-Held Thermometers

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

MODE button MAX/MIN button CHANNEL button / C/ F / RCC button SNOOZE / LIGHT button Battery compartment A/C in-jack 3. 4.

Model: 53005/ Printed in JAPAN 2012/05

1. Electrostatic Lab [1]

MODEL: MO Digital DC Milliohm Meter OPERATING INSTRUCTION

Non contact temperature measurement

INSTRUCTION MANUAL FOR PORTABLE DIGITAL INFRARED RADIATION THERMOMETER IRT-200-HS

IR-712 IR-720 IR-730

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

IR-SA Series Infrared radiation thermometer for Processing Model : IR-SAI IR-SAS IR-SAH

Traceable Big-Digit Type-K Thermometer Big digits are readable from 15 feet away. For additional Type-K probes, see page 41.

Model DIT-516/517: OPERATING MANUAL

TIR100-2 Measurement of Thermal Emissivity

Acceleration/Velocity/Displacement VIBRATION METER

VIBRATION METER Acceleration & Velocity Model : VB-8201HA

VIBRATION METER. Instruction Manual.

Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202

The Fluke 66 and 68 Infrared Thermometers

SUNLIKE DISPLAY. Model No: SC2004C GENERAL SPECIFICATION

Surface-mounting Thermostats ATH Type Series

Infrared Thermosensor ES1C

Surface Mount UV LED. NUVA33 Series PART NUMBERING SYSTEM. WAVELENGTH CODES Code Nominal Wavelength

optris LaserSight Operators manual

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

PORTABLE DIGITAL RADIATION THERMOMETER Model : PSC-90

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

Operator's Manual. optris BR 400. Calibration source

MANUAL TT-220 TT-220

Automatic titrators, TitroMatic

NX6314EH. Preliminary. Data Sheet. LASER DIODE nm InGaAsP MQW-DFB LASER DIODE FOR 1.25 Gb/s FTTH P2P AND 3 Gb/s BTS DESCRIPTION APPLICATIONS

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

DM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability

Ground Penetrating Radar & By Mike Kelty Western Technologies Inc.

Today s Weather 24 Hour Forecaster

Thermal Imaging Best Practices: The Measuring Object. Page 2

High Resolution High Sensitivity SDR Super Wide Dynamic Camera

Designed to go where you work

INFRARED TEMPERATURE SENSORS OPERATION AND SELECTION

THERMYS 150. Advanced field reference thermometer / temperature calibrator

SMBBIR45A Broad band High Power Emitter. AnodeMark 5.2. heatsink

2. OPERATIONAL CONDITIONS

Dr. Michael Müller. Thermal Management for LED applications

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

Transcription:

Infrared / K Thermometer Remote Temperature Measurement ( Non-Contact ) 1327K INSTRUCTION MANUAL

TABLE OF CONTENTS TITLE PAGE INTRODUCTION...1. SAFETY INFORMATION...2. FEATURES...4. SPECIFICATIONS...4. GENERAL LAYOUT AND OPERATING CONTROLS...7. LCD DISPLAY DESCRIPTION...9. EMISSIVITY ADJUSTMENT...9. TEMPERATURE MEASUREMENT...13. SETTING MODE...15. BATTERY REPLACEMENT...17

INTRODUCTION The RS 1327K hand-held thermometer is a convenient and easy to use instrument, allowing temperature measurement either by an Infrared non-contact method, or by using a plug-in type-k thermocouple probe. The thermometer has an integral laser to assist sighting the Infrared sensor for making non-contact measurements of items that are inaccessible or hazardous to touch such as rotating machinery, high-voltage bus-bars, or inaccessible objects. The Infrared mode does not rely on physical contact with the object whose temperature is to be taken, but measures the radiated infrared energy from the object. The thermocouple connector will accept a standard liquid, air or surface thermocouple probe as required to suit the particular application. The Liquid Crystal Display has a backlight function which may be enabled to assist measurements made in a dark place. Other features include reading auto-hold, auto power-off, maximum and minimum readings, alarm levels and adjustable emissivity, all of which can be selected for maximum operating convenience. E-1

. SAFETY INFORMATION 1. Read the following safety information carefully before attempting to operate or service the meter. Laser Safety Warning LASER RADIATION DO EN 60825-1:1994/A11:1996 If the MEAS button is pressed while the symbol is shown on the LCD, the meter will radiate the laser. To prevent injury, do not look or stare into the beam. Do not point the laser directly at eyes or indirectly off reflective surfaces. If the target object has a smooth surface, the laser may be reflected and create a hazard. Do not radiate the laser at flammable gasses or other possible sources of ignition. 2. Precautions: a). Do not operate the thermometer in the presence of strong electrical or magnetic fields, as this may prevent correct operation. b). Use and store the thermometer away from direct sunlight or other strong source of light, high temperatures, high humidity or dust. c). If the thermometer has been kept in area of high or low temperature, wait until it has reached a stable temperature before use. E-2

d). Condensation may form on the focal lens if the thermometer is moved quickly from a cold to a hot environment. Before taking measurements, wait for the condensation to dissipate. e). Do not touch the focal lens. 3. Environmental conditions: a). Altitude up to 2000 meters. b). Relatively humidity: 80% max. c). Operating ambient temperature: 0 to 50 C 4. Maintenance & Clearing: a). Repairs or servicing should only be carried out by competent persons using the correct parts, test equipment and procedures. If repair or servicing is required, contact the supplier or RS Components. The address is given at the end of these instructions. b). Periodically clean the case with a dry cloth. Do not use abrasives or solvents. 5. Safety symbols: Complies with the EMC Directive 89/336/EEC E-3

. FEATURES C/ F Selectable. Backlit LCD display. Sighting laser. Auto-hold function. Maximum/minimum reading recorder function. Auto power-off. Type K thermocouple thermometer.. SPECIFICATIONS 3-1 General Information: Display: Backlit LCD display. Auto power-off time: Approx.15 sec. Data memory capacity: 99 set. (Direct reading on LCD display) Over-range indication: OL or -OL. Low battery indication: The symbol is displayed on LCD when the battery voltage drops below the operating voltage. Power supply: Single 9V battery 006P 9V, IEC6F22, or NEDA1604. Battery life: Approx. 100 hours (with laser pointer and backlight off) Operating temperature: 0 C to 50 C (32 F to 122 F), Operating humidity: Below 80%RH. E-4 +

Storage temperature Humidity: Dimensions: Weight: Accessories: 10 C to 60 C (14 F to 140 F), Below 70%RH. 172(L) 118(W) 46(H) mm 6.8(L) 4.6(W) 1.8(H) inches. Approx. 220g with battery. Instruction manual, carrying case, battery. 3-2 Electrical specifications: Measuring range: -35 C to 500 C (-31 F to 932 F) Resolution: 0.1 C, 0.2 F Accuracy: ± 2% of reading or ± 2 C or 4 F (whichever is greater). Temperature coefficient: 0.1 times the applicable accuracy specification per C from 0 C to 18 C and 28 C to 50 C (32 F to 64 F and 82 F to 122 F). Measurement rate: 2 times per second. Spectral Response: 6 to 14um. Field of view: 12:1; optics ratio with a 1 min target. Emissivity: 0.17 ~ 1.00 (adjustable, preset at 0.95) Sighting: Laser marker < 1mw (class 2). Sensor: Thermopile. E-5

Type K Measuring range : -150 C to 1350 C (-238 F to 1999 F). Resolution : 0.1 C, 1 C, 0.1 F, 1 F Measurement rate : Once per second. Basic accuracy : (@ 23±5 C calibration) accuracy are ± ( % of reading + degree) at 18 C to 28 C with relative humidity up 80%. Function Resolution Range Accuracy C F 0.1 C -150 C~0 C ( 0.2%rdg + 1.0 C) 0 C~200 C ( 0.1%rdg + 1.0 C) 1 C 200 C~1350 C ( 0.2%rdg + 2 C) 0.1 F -238 F~32 F ( 0.2%rdg + 2 F) 32 F~200 F ( 0.1%rdg + 2 F) 1 F 200 F~1999 F ( 0.2%rdg + 4 F) Temperature Coefficient: 0.1 times the applicable accuracy specification per C from 0 C to 18 C and 28 C to 50 C ( 32 F to 64 F and 82 F to 122 F ). Note : The basic accuracy specification does not include the error of the probe. Please refer to the probe accuracy specification for additional details. E-6

. GENERAL LAYOUT AND OPERATING CONTROLS CAUTION AVOID EXPOSURE - LASER RADIATION IS EMITTED FROM THIS APERTURE LASER RADIATION - DO NOT STARE INTO BEAM OUTPUT < 1mW WAVELENGTH 630-670nm CLASS (II) LASER PRODUCT Complies with IEC 60825-1:(1998-01) Edition 1.1, EN 60825-1:1994/All:1996 K C F Type K E-7

1. Display. Button 2. MODE 3. 4. M R Press once Enter MAX/MIN recording mode. MAX MIN Current reading (MAX/MIN) From normal mode will turn the backlight on or off From Read mode, decrements the memory location number. From SET mode decreases the value of the selected parameter. 8. Focal lens. 9. Type K thermocouple input socket. E-8 Press and hold for 3 seconds From MAX/MIN mode will exit this mode and return to normal operation. From normal mode will enter setting mode. Store the displayed reading. From normal mode will enable or disable the laser. From Read mode increments the Read the stored data memory location number. from memory. From SET mode increases the value of the selected parameter. 5. Measuring trigger : Trigger for turning the thermometer on. Pull the trigger to turn on the thermometer and make a measurement. In SET mode, pull the trigger to store the required value and exit SET mode. Press and hold the MODE button, then pull the trigger to enter continuous measurement mode. In continuous measurement mode, pull the trigger to turn the meter off. 6. Battery cover. 7. Laser aperture.

. LCD DISPLAY DESCRIPTION K C F 1. Hold function 11. Lo alarm 2. Maximum reading 12. Alarm function 3. Minimum reading 13. Hi alarm 4. Low battery indicator 14. Memory and Read address number 5. Maximum/minimum value 15. Read memory data 6. Emissivity value setting 16. Type K measuring indicator 7. Measured value 17. Memory full indication 8. Measure indication 18. Store data to memory 9. Laser enabled indicator 19. Units C, F 10. SET symbol. Emissivity Adjustment 1. Emissivity: Emissivity is a value between 0.17 and 1 which indicates an object s ability to emit infrared energy. Emissivity is determined primarily by the objects composition and surface finish. The thermometers sensitivity to emissivity ε is preset at 0.95, because this setting will be applicable to 90% of surfaces encountered, but can be adjusted as required. E-9

2. Emissivity adjustment Apply black tape, matt-black paint or black magic marker to the object if it is safe to do so. ε is set at 0.95 to indicate correctly from a dark surface. Aim the sensor at the black surface, pull the trigger and note the reading as T1. Remove the black tape or matt-black paint and aim the sensor at the same area again. Pull the trigger and note the reading as (T2). Change the emissivity value ε until T2 and T1 are equal as follows: a). Enter setting mode (refer to MODE button above). b). Press MODE button to select theε value setting. c). Press or buttons to set desired value. d). Pull the trigger to store the setting value and exit set mode. The following table gives typical emissivity values for common materials Emissivity VALUES Typical Emissivity Values; Metals SURFACE EMISSIVITY Zinc (oxidized) 0.1* Galvanized iron 0.3 Tin-plated steel 0.1* Gold (polished) 0.1* Silver (polished) 0.1* Chromium (polished) 0.1* Iron and Steel Cast iron (polished) 0.2 Cast iron (turned at 100 C) 0.45 Cast iron (turned at 1000 C) 0.6 to 0.7 Steel (ground sheet) 0.6 Mild steel 0.3 to 0.5 E-10

Steel plate (oxidized) 0.9 Iron plate (rusted) 0.7 to 0.85 Cast iron (rough) rusted 0.95 Rough ingot iron 0.9 EMISSIVITY VALUES Typical Emissivity Values; Metals SURFACE EMISSIVITY Molten cast iron 0.3 Molten mild steel 0.3 to 0.4 Stainless steel (polished) 0.1 Stainless steel (various) 0.2 to 0.6 Aluminum Polished aluminum 0.1* Aluminum (heavily oxidized) 0.25 Aluminum oxide at 260 C 0.6 Aluminum oxide at 800 C 0.3 Aluminum Alloys various 0.1 to 0.25 Brass Brass (polished) 0.1* Brass (roughened surface) 0.2 Brass (oxidized) 0.6 Copper Copper (polished) 0.05* Copper plate (oxidized) 0.8 Molten copper 0.15 Lead Lead (pure) 0.1* Lead (oxidized at 25 C) 0.3 Lead (oxidized, reated to 200 C) 0.6 Nickel and its alloys Nickel (pure) 0.1* Nickel plate (oxidized) 0.4 to 0.5 Nichrome 0.7 Nichrome (oxidized) 0.95 E-11

Emissivity Values; Non-metals - Refractory & Building materials Red brick (rough) 0.75 to 0.9 Fire clay 0.75 Asbestos 0.95 Concrete 0.7 Marble 0.9 Carborundum 0.85 Plaster 0.9 Alumina (fine grain) 0.25 Alumina (coarse grain) 0.45 Silica (fine grain) 0.4 Silica (coarse grain) 0.55 Zirconium silicate up to 500 C 0.85 Zirconium silicate at 850 C 0.6 Quartz (rough) 0.9 Carbon (graphite) 0.75 Carbon (soot) 0.95 Timber (various) 0.8 to 0.9 Miscellaneous Enamel (any color) 0.9 Oil paint (any color) 0.95 Lacquer 0.9 Matte black paint 0.95 to 0.98 Aluminum lacquer 0.5 Water 0.98 Rubber (smooth) 0.9 Rubber (rough) 0.98 Plastics (various, solid) 0.8 to 0.95 Plastic films (05 mm thick) 0.5 to 0.95 Polythene film (03 mm thick) 0.2 to 0.3 Paper and cardboard 0.9 Silicone polish (03 mm thick) 0.7 * Emissivity value varies with purity E-12

. TEMPERATURE MEASUREMENT 1. Non-contact measurement : Point the instrument at the target and pull the trigger to turn it on and start measuring (use the laser to select the target if required). Release the trigger to stop measuring and auto-hold the reading. The meter will turn off automatically after 15 seconds. Auto detection type K thermocouple plug-in or pull-out. If a type K thermocouple is connected to the instrument, the emissivity symbol ε will automatically change to K and the temperature measured by the thermocouple will be displayed. The Infrared sensor is disabled when a thermocouple is detected by the instrument. 2. MAX/MIN hold function : Pull & hold the trigger, then press MODE once to cycle through maximum (MAX), minimum (MIN) and current reading (MAX/MIN) modes. Press and hold the MODE button for 3 seconds to exit this mode. 3. Continuous measurement : Start with the instrument turned off. Press and hold the MODE button and then pull the trigger to turn the meter on. Release the MODE button and the instrument will enter continuous measurement mode. The H symbol will not be visible in the LCD, indicating the Hold function is disabled To exit continuous measurement mode: Pull the trigger to stop continuous measurement and turn the meter off. MAX/MIN reading: When in continuous measurement mode, press the MODE button once to cycle through maximum (MAX), minimum (MIN) and current reading (MAX/MIN) modes. Press and hold the MODE button for 3 seconds to exit this mode. E-13

Note: Laser offset: The laser beam is offset 18mm(0.71in) from the focal lens. Choose a sampling spot that is large enough to include the laser offset. Surface Temperature: The thermometer will measure the temperature of the first surface it detects, e.g. a glass cover, dust or fog. Ensure the object whose temperature is to be measured is not obstructed. 4. Selecting C/ F units: When the instrument is turned on, the units that appear in the LCD are the last units measured. To change the temperature units, pull the trigger to turn the meter on, then press and hold MODE for 3 seconds until the SET mark appears on the LCD. Press or to select the desired C or F units and then pull trigger to store the setting. 5. Data memory store and read procedure : To clear the data from the memory: With the instrument turned off, press and hold the M button, then pull trigger until CLR appears on the LCD. Data memory: Press and hold the M button for 3 seconds to store the current set of readings into memory. E-14

Reading the memory data: Press and hold the R button for 3 seconds to enter READ mode. Press the or buttons to select to desired memory data location. Press the MODE button to exit READ mode.. SETTING MODE Pull trigger to turn the meter on. Press & hold the MODE button for 3 seconds until SET appears on the LCD. 1. Parameter settings for measuring: Under the infrared parameter setting mode, the LCD will show SET. Press the MODE button to select setting of C / F, ALM ON/OFF, ALM ALM and emissivity ε. Pull the trigger to exit setting mode and return to normal measurement mode. 2. Parameter settings: C / F : Temperature unit / F, press or to select units C or F. ALM (Alarm function ON/OFF): Press or to select alarm function on or off. ALM (Lo Alarm Function): Press or to set a value as an alarm value. When this Lo alarm value is exceeded, the beeper will sound and the ALM symbol will appear on the LCD. ALM (Hi Alarm Function): Press or to set up a value as an alarm value. When this Hi alarm value is exceeded, the beeper will sound and the ALM symbol will appear on the LCD. Example: To sound the alarm if the temperature exceeds 100 C or falls below 0 C, set the Hi Alarm point to100 C and the Lo Alarm point to 0 C. E-15

Setting procedure: a). Press and hold the MODE key for 3 seconds until the LCD displays the SET mark. b). Press button to select C unit. c). Press the MODE button once to enter ALARM ON/OFF setting mode, then press the or key to select ON mode. d). Press the MODE button once to enter ALARM (Lo Alarm point) setting mode. Press the or button until the display reads 0.0 C. e). Press the MODE button once to enter ALARM (Hi Alarm point) setting mode. Press the or button until the display reads100.0 C f). Pull the trigger to store the setting and exit the setting mode. ε (Emissivity): Press or to adjust the value. : Button for increasing the value. Hold down to increase the parameter value rapidly. : Button for decreasing the value. Hold down to decrease the parameter value rapidly. Note: After the setting procedure is terminated, the parameter will be stored until changed. Under setting mode, the backlight, laser, memory store and read functions are disabled. Pull the trigger to exit setting mode. E-16

BATTERY REPLACEMENT When the symbol + appears on the LCD screen, there is insufficient power in the battery to provide an accurate measurement. Replace the battery with a fresh one as follows: a) Ensure the instrument is turned off b) Open battery compartment cover on the rear of the grip. c) Remove the exhausted battery from the compartment. d) Fit a new battery of the type given in the specifications. e) Close the battery cover. f) Turn the instrument on and check for correct operation. OPEN E-17