Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Similar documents
Application field / Industry branch:

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

DRAFT EAST AFRICAN STANDARD

Atomic Absorption Spectrometer ZEEnit P series

ZEEnit. Quality is the difference

ZEEnit. Quality is the difference

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

Eppendorf BioSpectrometer 6135/6136/6137

Sodium Chloride - Analytical Standard

novaa 800 D Atomic Absorption Spectrometer

Brooklyn College Department of Chemistry

On the use of Deming and quadratic calibrations in analytical chemistry

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Elemental Scientific. Spike Recovery. Elemental Scientific

Traceable UV-Vis Reference Materials

Analysis of domestic sludge using the Agilent 4200 MP-AES

Quantum Series Product Catalog

Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

110 Lowell Rd Hudson, NH USA Tel:

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

ISO INTERNATIONAL STANDARD. Water quality Determination of trace elements using atomic absorption spectrometry with graphite furnace

EVEN THE FINEST CANNOT ESCAPE. Wöhler SM 500 Suspended Particulate Analyser

CHEMISTRY WORKSTATION. The power of proven quality GIVING YOU PEACE OF MIND

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Electrical safety testing Checklist

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

The Plus for your lab!

Compact Titration Compact Titrator G20. One Click Titration Simple & Dependable

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

Easy and reliable KF titrations From 1 ppm to 100%

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

ZEEnit. Quality is the difference

Spectrometer User s Guide

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

Test Method: CPSC-CH-E

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

Fields of Application / Industry:

ISO INTERNATIONAL STANDARD. Fruit and vegetable products Determination of tin content Method using flame atomic absorption spectrometry

Product Descriptions and Offerings

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Elemental Scientific. Stability. Elemental Scientific

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

PURCHASE DIVISION DEPARTMENT QUALITY MANUAL. CPRI Section II -Technical Specification

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Approval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

TECHNICAL MANUAL 820 LX / 910 LX / 1300 LX

The Measurement of Lead in Food Products by Graphite Furnace AA and SpectraAA-40

Alpha spectrometry systems. A users perspective. George Ham. 26 th May Date Month Year

Determination of Trace Cations in Power Plant Waters Containing Morpholine

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Speed and flexibility

Atomic Absorption Spectrophotometry. Presentation by, Mrs. Sangita J. Chandratre Department of Microbiology M. J. college, Jalgaon

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Freezing point osmometry

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Ideal and powerful solution

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Application Note PlasmaQuant PQ 9000 Elite, PlasmaQuant MS Elite

Your partner in Science!

ISO INTERNATIONAL STANDARD

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

Test Procedure for ItalSens IS-1 Graphite Sensors

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

V. LAB REPORT. PART I. ICP-AES (section IVA)

Technology that counts

MN-ISO-2000FN-E. ISO-2000FN COATING THICKNESS GAGE OPERATION MANUAL

Technical Procedure for Glass Refractive Index Measurement System 3 (GRIM 3)

Kjeltec 2300 Analyzer Unit

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

R-Biopharm AG RIDA CUBE SCAN. The smart analyser. Product Information. R-Biopharm for the safety of your analysis.

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC)

Cadmium and lead content

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct

Avery Dennison LW Series Retroreflective Window Sticker Film Issued: July 2013

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Transcription:

Maintenance with OQ Operational Qualification OQ within the scope of a maintenance for the Atomic Absorption Spectrometer ZEEnit 700

Contents 1. GENERAL INFORMATION...4 2. DEFINITIONS...4 2.1 Validation...4 2.2 Verification...4 2.3 Qualification...4 3. ASCERTAINMENT OF THE TECHNIQUES AND ACCESSORIES TO BE TESTED... 5 4. FUNCTION TEST OF BASIC FUNCTIONS OF DEVICE AND OF DEVICE PERIPHERALS...5 5. OPERATIONAL QUALIFICATION (OQ)...7 5.1 Basic calibration of Graphite-furnace technique...8 5.1.1 Calibration procedure... 9 5.1.2 Results of calibration... 9 5.1.3 Recovery of a quality control sample...10 5.1.4 Determination of detection limit...11 5.2 Basic calibration of flame technique...12 5.2.1 Calibration procedure...13 5.2.2 Results of calibration...14 5.2.3 Recovery of a quality control sample...15 5.2.4 Determination of detection limit...16 5.3 Basic calibration of Flame technique in emission mode...17 5.3.1 Calibration procedure...18 5.3.2 Results of calibration...19 5.3.3 Recovery of a quality control sample...19 5.3.4 Determination of detection limit...20 6. FAULTS ASCERTAINED IN INSTRUMENT VALIDATION...21 7. CONCLUSION OF MAINTENANCE WITH OPERATIONAL QUALIFICATION...21 V: 1.0 2/ 22

Order no.: Software version: Customer no.: performed by: Serial number: Company User Department Street Zip Code, City Country Phone Fax E-mail V: 1.0 3/ 22

1. General information The ZEEnit 700 is a fully automatic, sequential multielement atomic absorption spectrometer of the latest generation. It can be used both in absorption mode and, for specific elements, also in emission mode. Being a compact AA spectrometer, it provides the performance of all AAS techniques. The combination of graphite furnace and flame AAS allows the entire dynamic measuring range of atomic absorption spectrometry from ultratrace to principal component analysis to be covered by a single system. The system with the requirements as laid down in EN ISO 9001. The used control and analytical software operates on the principles of Good Laboratory Practice (GLP). The ZEEnit 700 is particularly suitable for quality monitoring in pharmacy or, in general, for the control and monitoring of production processes, thus finding extensive applications in research and environmental care. 2. Definitions 2.1 Validation Described by FDA (Food and Drug Administration, USA) as follows: Document furnishing proof, that a defined process with a high degree of reliability will continuously yield a product that meets predefined specifications and quality features. 2.2 Verification Defined in EN 45020: Examination of generally accepted performance data of a device or a method that are valid for all applications that can be performed with the device or method. 2.3 Qualification Term used by Pharmaceutical Manufacturers Association (PMA, USA): The qualification deals with the testing of instruments and software products throughout the entire life cycle of the instrument system. It may be subdivided into different phases. Specification Qualification (SQ) Production Qualification (CQ) Design Qualification (DQ) Installation Qualification (IQ) Operational Qualification (OQ) Performance Qualification (PQ) V: 1.0 4/ 22

3. Ascertainment of the techniques and accessories to be tested In this section, based on the existent order, those techniques shall be defined that are to be tested on the AAS ZEEnit 700 within the scope of the Operational Qualification. Techniques and accessories that are not included in the supply shall not be subjected to OQ. Graphite furnace technique Flame technique 50 mm burner 100 mm burner Flame technique (emission, 50 mm burner) 4. Function test of basic functions of device and of device peripherals does not comply Functionality of hardware and software On pressing the green power key, he AAS ZEEnit 700 starts with the automatic initialization of the spectrometer. The autosampler (if installed) also starts with the automatic initialization (sample tray, dosing device and injector arm are moving to their initial positions). Counter-read by: V: 1.0 5/ 22

Communication of hardware and software The communication between the hardware and the control software of the device is checked by loading the control software (software for flame or graphite furnace technique). In the case of error-free communication, a method stored in the device memory will be loaded. In the case of a communication error, the message Time Out will appear. By pressing the Print key on the PC keyboard, the communication with the printer is tested. The printer will print the current screen contents. Discovery of faults occurring in the functional check Any faults occurring in the functional test are to be noted in writing on this page. If needed also dates for the rectification of faults that can not be realized within the maintenance shall be enter. If no faults occur in the functional check, this should be signed by the Analytik Jena AG specialist and countersigned from the competent head of laboratory or his/her representative. Faults occurring Comment Competent person Initials Date Repair of faults occurring Comment Additional assigned service date Competent person Initials Date does not comply The ZEEnit 700 and the included components did not show any apparent faults in the function tests performed. Counter-read by: V: 1.0 6/ 22

5. Operational Qualification (OQ) Prior to beginning with the OQ (and actual sample measurements), it is advisable to clean the atomizer systems provided and the sample-handling systems by clean-out (graphite furnace atomizer) or by thorough rinsing with nitric acid solution (approx. 1%). Start-up and successful performance of OQ of the system requires the warm-up of all device components (hollow-cathode lamps, spectrometer, etc.) of at least 20 minutes. Basic calibration of the AAS ZEEnit 600 / 650 is performed after instrument delivery at the customer's premises. Within the scope of OQ, basic calibration is performed by the analysis of elements that are characteristic for the respective technique or atomizer (graphite furnace. Basic calibration of the graphite furnace system is based on the analysis of vanadium. During the procedure, the analytical sensitivity of the spectrometer, expressed by the characteristic concentration C 0, and the instrument detection limit (IDL) are determined. After the calibration, additional tests are run for the precision of analytical determination (relative standard deviation in %) and the recovery rate of a quality control standard. V: 1.0 7/ 22

5.1 Basic calibration of Graphite-furnace technique Preparations Switch-on of ZEEnit 700 and loading of test utility V-val Checking the adjustment of MPE injector relative to graphite furnace and sample cups Start of formation program of graphite furnace Adjustment of hollow-cathode lamps Starting peak search (line maximum) Checking the current energy level of HCL/D2HCL Checking if atomizer system is free of any contamination Element vanadium does not comply MPE adjusted Formation factor of graphite furnace +10-10% Peak search successful (deviation < 0.5 nm) Energy level HCL/D2HCL between 60-80 % Atomizer being free of contamination Counter-read by: V: 1.0 8/ 22

5.1.1 Calibration procedure Calibration is performed by automatic, volume-graduated dilution by means of the MPE 60. Five reference standards are being prepared from a calibration blank solution (1% nitric acid (HNO 3 )) and a calibration stock standard (100 µg/l vanadium in 1% nitric acid (HNO 3 )). These reference standards have the following calibration concentrations: Vanadium (µg/l) 0.0 / 20.0 / 40.0 / 60.0 / 80.0 / 100.0 The chemicals used must have a purity appropriate for trace analysis (e.g. suprapure registered trademark of Merck). The calibration stock standards are being prepared from commercial element standard solutions. The used calibration standards must be freshly prepared every day. After successful calibration the following analytical performance data are being determined. Quality of determination R 2 Characteristic concentration c 0 Slope of calibration curve 5.1.2 Results of calibration Element Vanadium does not comply Quality of determination R 2 > 0.995 Characteristic concentration c 0 < 3.0 µg/l Slope of calibration curve > 0.001 A/µg/L If the required analytical performance data should not be met, the calibration procedure must be repeated. Counter-read by: V: 1.0 9/ 22

5.1.3 Recovery of a quality control sample The accuracy of the analytical measurement is to be checked by analysing a quality control sample that is independent of calibration. The result of the measured QC sample is stored on a quality control card. The accuracy test is performed with a test solution having a concentration of 50 µg/l vanadium in 1 % HNO 3. Element Vanadium Recovery rate 95 105 % does not comply Precision (RSD %) < 3 Counter-read by: V: 1.0 10/ 22

5.1.4 Determination of detection limit The detection limit is determined on the principle of the blank method by measuring the calibration blank eleven times with the instrument put into operation. The detection limit is calculated from the blank-value variance according to the 3-sigma criterion and output by software. The determination of the detection limit requires previous successful calibration of the spectrometer (see Basic Calibration). Element Vanadium does not comply Detection limit (DL) < 5 µg/l Counter-read by: V: 1.0 11/ 22

5.2 Basic calibration of flame technique Preparations Switch on ZEEnit 700 and loading of test routines Cu-val50 or Cu-val100. Igniting the flame, letting it burn in for approx. 20 minutes. Adjustment of hollow cathode lamp Starting peak search (line maximum) Checking the current energy level of HCL/D2HCL Results of preparations Element copper (50-mm burner / 100-mm burner) does not comply Burn-in period (approx. 20 min) Peak search successful (deviation < 0.5 nm) Energy level of HCL/D2HCL between 60-80 % Counter-read by: V: 1.0 12/ 22

5.2.1 Calibration procedure Calibration is performed by automatic, volume-graduated dilution by means of the AS52 Autosampler. If the scope of delivery does not include an autosampler, the calibration standards are to be prepared manually. Five reference standards are being prepared from a calibration blank solution (1% nitric acid (HNO 3) ; 0.1% potassium chloride (KCl)) and a calibration stock standard (10 mg/l copper in 1% nitric acid (HNO 3 ); 0.1% potassium chloride (KCl)). These reference standards have the following calibration concentrations: Copper 50-mm burner (mg/l) 0.0 / 0.5 / 1.0 / 1.5 / 2.0 / 2.5 Copper 100-mm burner (mg/l) 0.0 / 0.4 / 0.8 / 1.2 / 1.6 / 2.0 The chemicals used must have a purity appropriate for trace analysis (e.g. suprapure registered trademark of Merck). The calibration stock standards are being prepared from commercial element standard solutions. The used calibration standards must be freshly prepared every day. After successful calibration the following analytical performance data are being determined. Quality of determination R 2 Characteristic concentration c 0 Slope of calibration curve Counter-read by: V: 1.0 13/ 22

5.2.2 Results of calibration Element: Copper (50-mm burner) does not comply Quality of determination R 2 > 0.995 Characteristic concentration c 0 0.1 mg/l Slope of calibration curve > 0.07 A/mg/L Element: Copper (100-mm burner) does not comply Quality of determination R 2 > 0.995 Characteristic concentration c 0 0.06 mg/l Slope of calibration curve > 0.1 A/mg/L If the required analytical performance data should not be met, the calibration procedure must be repeated. Counter-read by: V: 1.0 14/ 22

5.2.3 Recovery of a quality control sample The accuracy of the analytical measurement is to be checked by analyzing a quality control sample that is independent of calibration. The result of the measured QC sample is stored on a quality control card. The accuracy test is performed with a test solution having a concentration of 1 mg/l copper in 1% HNO 3 and 0.1% KCl. Element: Copper (50- and 100-mm burner) does not comply Recovery rate 95 105 % Precision (RSD %) < 2 Counter-read by: V: 1.0 15/ 22

5.2.4 Determination of detection limit The detection limit is determined on the principle of the blank method by measuring the calibration blank 11 times with the instrument put into operation. The detection limit is calculated from the blank-value variance according to the 3-sigma criterion. The determination of the detection limit requires previous successful calibration of the spectrometer (see Basic Calibration). Element: Copper (50-mm burner) does not comply Detection Limit (DL) < 0.05 mg/l Element: Copper (100-mm burner) does not comply Detection Limit (DL) < 0.03 mg/l Counter-read by: V: 1.0 16/ 22

5.3 Basic calibration of Flame technique in emission mode Preparations Switch-on of ZEEnit 700 and loading of test routine Na-val Turning the burner into 90 position Igniting the flame, letting it burn in for approx. 20 minutes Performing energy balance (MIN/MAX) Results of preparations Element: Sodium does not comply Burn-in period (approx. 20 min) Energy balance successful (MIN/MAX, MAX 25 mg/l) Counter-read by: V: 1.0 17/ 22

5.3.1 Calibration procedure Calibration is performed by automatic, volume-graduated dilution by means of the AS52 Autosampler. If the scope of delivery does not include an autosampler, the calibration standards are to be prepared manually. 5 reference standards with the following concentrations were prepared using a blank calibration solution (aqueous 0,1% CsCl-solution) and a stock calibration standard (25 mg/l sodium (Na) in aqueous 0,1% CsCl-solution). Sodium 50-mm burner (mg/l) 0.0 / 2.0 / 4.0 / 6.0 / 8.0 / 10.0 The chemicals used must have a purity appropriate for trace analysis (e.g. suprapure registered trademark of Merck). The calibration stock standards are being prepared from commercial element standard solutions. The used calibration standards must be freshly prepared every day. After successful calibration the following analytical performance data are being determined. Quality of determination R 2 Slope of calibration curve Counter-read by: V: 1.0 18/ 22

5.3.2 Results of calibration Element: Sodium does not comply Quality of determination R 2 > 0.995 Slope of calibration curve > 0.055 Ems/mg/L If the required analytical performance data should not be met, the calibration procedure must be repeated. 5.3.3 Recovery of a quality control sample The accuracy of the analytical measurement is to be checked by analyzing a quality control sample that is independent of calibration. The result of the measured QC sample is stored on a quality control card. The accuracy test is performed with a test solution having a concentration of 5 mg/l sodium in distilled water. Element: Sodium does not comply Recovery rate 95 105 % Precision (RSD %) < 2 Counter-read by: V: 1.0 19/ 22

5.3.4 Determination of detection limit The detection limit is determined on the principle of the "blank method by measuring the calibration blank 11 times with the device put into operation. The detection limit is calculated from the blank-value variance according to the 3-sigma criterion. The determination of the detection limit requires previous successful calibration of the spectrometer (see Basic Calibration). Element: Sodium does not comply Detection limit (DL) < 0.010 mg/l Counter-read by: V: 1.0 20/ 22

6. Faults ascertained in instrument validation On this page, faults that occurred during instrument validation shall be put down in writing. If needed also dates for the rectification of faults that can not be realized within the maintenance shall be enter. If the tests did not show any faults, the qualified personnel of Analytik Jena AG as well as the responsible head of the laboratory or his deputy shall countersign the statement of this condition. Faults occurring Comment Competent person Initials Date Repair of faults occurring Comment Additional assigned service date Competent person Initials Date 7. Conclusion of maintenance with operational qualification The maintenance with operational qualification has been performed and concluded correctly. does not comply This provides verification that the ZEEnit 700 with the performance data guaranteed by Analytik Jena AG. The ZEEnit 700 is hereby released by the qualified signatories for Counter-read by: V: 1.0 21/ 22

Name technician (in block letters) Name customer (in block letters) Signature technician Signature customer Place, Date (DD/MM/YYYY) Place, Date (DD/MM/YYYY) Version 1.0 2007-10-11 2007 Analytik Jena AG Publisher: Analytik Jena AG Konrad-Zuse-Straße 1 07745 Jena Telefon +49 36 41 77-70 Telefax +49 36 41 77-92 79 www.analytik-jena.com info@analytik-jena.com V: 1.0 22/ 22