Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Similar documents
Laboratory ID. Laboratory Name. Address. Phone. Lab Manager/Supervisor. QA Officer. Analyst(s) Auditor. Date(s) of Audit

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS

Working Copy. tlrq /rb I 'B-16. Determination of Gross Alpha/Beta Activity. Revision 0. Laboratory Manag erllqao/rso. Effective Date. lnitials.

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

ALLOWAY METHOD OUTLINE

CHAPTER VI ANALYTICAL METHODS & QUALITY CONTROL

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Liquid Scintillation Counter

ANALYTICAL METHODS & QUALITY CONTROL

Laboratory Certification Workshop

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

QAM-I-116 Preparation of Labware

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

Percent Solids Determination

Technical Notes for EPA Method Gross Alpha and Gross Beta Radioactivity in Drinking Water

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

SLAC Radioanalysis Laboratory

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

STANDARD OPERATING PROCEDURES

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

QSM 5.1 FAQs. For a test with a recommended maximum holding time measured in hours, the holding time shall be tracked by the hour.

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

QAM-Q-101 Laboratory Quality Control

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

A Brief Overview of Radiation and Analytical Water Testing for Radiological Contaminants.

ANALYTICAL REPORT. Attn: Richard Boelter

Qualification Code Reference Table Cont.

MANUAL FOR THE CERTIFICATION OF LABORATORIES ANALYZING ENVIRONMENTAL SAMPLES FOR THE IOWA DEPARTMENT OF NATURAL RESOURCES

ENVIRONMENTAL LABORATORY SECTOR MODIFIED WORKING DRAFT STANDARD (MWDS) VOLUME 1

Table of Contents I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:...

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

December 30, Laboratory Report This Page is to be Stamped Introduction: This report package contains total of 5 pages divided into 3 sections:

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350)

Standard Operating Procedure PCM SOP 1-1. Microscope Calibration, Preparation and Analysis of Air Samples by Phase Contrast Microscopy (PCM)

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

URANIUM IN SOIL. Analytical Procedure (2 GRAM SAMPLE) 1. SCOPE

CERTIFICATE OF ANALYSIS

SECTION D.2 AMMONIA NITROGEN

Radiochemistry Webinars Data Verification and Validation

DISTRIBUTION LIST To be filled out by Document Author or person requesting document deletion

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Certificate of Analysis FINAL REPORT

alscience nvironmental aboratories, Inc.

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Seawater Extraction Procedure for Trace Metals

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Shaker Table Extraction

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

ANALYSIS CERTIFICATE OF

Test Method: CPSC-CH-E

Technical Procedure for General Laboratory Equipment

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

Colorado Oil and Gas Conservation Commission

ANALYTICAL REPORT. Attn: Richard Boelter

Ohio EPA Laboratory Manual for Chemical Analyses of Public Drinking Water 2014

TNI Standard; EL-V1M4 Sections and (Detection and Quantitation) page1 of 8. TNI Standard VOLUME 1 MODULE 4

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

IN THE CIRCUIT COURT OF THE SEVENTEENTH JUDICIAL CIRCUIT IN AND FOR BROWARD COUNTY, FLORIDA. Div.: XXXXXXXX PRC MOTION TO COMPEL ADDITIONAL DISCOVERY

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

1.1. This is a method for the separation and measurement of 228 Ra in water via its beta emitting 228 Ac daughter.

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011

Certified Reference Material - Certificate of Analysis

VOTING DRAFT STANDARD

Radiological Traceability Program (RTP) Radiological and Environmental Science Laboratory (RESL) Svetlana Nour Kenneth Inn Jerry LaRosa Jackie Mann

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory

Certified Reference Material - Certificate of Analysis

MODULE 2B INDUSTRIAL HYGIENE LABORATORY ACCREDITATION PROGRAM (IHLAP) ADDITIONAL REQUIREMENTS

TNI V1M Standard Update Guidance on Detection and Quantitation

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

American Environmental Testing Laboratory Inc.

RADIOPHARMACY PURPOSE

Certificate of analysis Certified reference material code: CRM-00-GTX1&4

Center for Environmental Sciences and Engineering Laboratory Quality Assurance Plan

Characterization Survey Techniques and Some Practical Feedback

LOW-RANGE TOTAL CHLORINE

Maine DEP DRO. MY05TP220A(0-0.5) MY05TP222A(0-0.5) FD MY05TP223A(0-0.5) FD MY05TP220A (EB) EB-Equipment Rinsate Blank FD- Field Duplicate Samples

Method Validation and Accreditation

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

APPENDIX II QUALITY CONTROL SUMMARY TABLES

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

This procedure describes the monitoring activities in a laboratory quality control (QC) program to ensure the quality of test results.

Transcription:

NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION OFFICE OF QUALITY ASSURANCE ENVIRONMENTAL LABORATORY CERTIFICATION PROGRAM ON-SITE LABORATORY EVALUATION RADIOCHEMISTRY PROCEDURES Gross Alpha-Gross Beta Analysis -- Gas-Flow Proportional Counter Method(s): (DW) SM 7110 B (DW) 48-Hour Rapid Gross Alpha Test (DW) Other: Laboratory ID Laboratory Name Analyst(s) Auditor Date(s) of Audit Type of Audit Initial Biennial Special ELCP TNI/NELAP Notes:

Analytical Method & SOP Does the laboratory maintain an SOP for each accredited method? 4.2.8.5(e) Are the SOPs complete and acceptable? On each SOP, does it clearly indicate the effective date of the document, revision number, and signature(s) of approving authority? Are all relevant SOPs readily available to all personnel? Sample Control Is the laboratory involved in sampling? (If not, the skip next two questions) Does the laboratory provide necessary sampling equipment and/or supplies to its clients? Does the laboratory have a sampling plan and procedures for sampling? Does the laboratory have a procedure for chain of custody (COC) of samples? Does the laboratory maintain a COC form for each sample? Does the laboratory utilize a permanent chronological record to document receipt of all sample containers (e.g., LIMS database)? Does the laboratory assign a unique laboratory ID code to each sample, which is linked to its field ID code? Does the laboratory assign unique ID extension to subsamples? (applies primarily to bulk samples) Does the laboratory implement procedures for checking chemical preservation prior to or during sample preparation or analysis? Sample Collection and Preservation Is the representative sample collected from a free-flowing source of drinking water? Are samples stored in polyethylene, fluoropolymer, or glass containers? 6.6(b) 4.2.8.5 6.6(b)1 4.2.8.5(c) 6.6(b)4.i&ii 4.2.8.5(b) 6.6(b)2 N/A N/A 5.7.1 5.8.8 9.3.4 5.8.7.5 9.3.4.i. 5.8.7.3 9.3.3.iv 5.8.7.3(b)i 5.8.5.a) 1.7.4(b) 9.4(a) Section 3.1 7:18 Table 9.3 40 CFR Part 136 Table 2 7:18 Page 1 of 6

Table 9.3 Are samples preserved with HNO 3 to ph < 2? 40 CFR Part 136 Table 2 Are samples analyzed within 6 months, if preserved? 40 CFR Part 136 Table 2 Are samples analyzed within 48 hr for NJ Rapid Gross Alpha Test? Method 6.4(a)3.ii Is the method used applicable for the samples being tested? - : alpha emitters above 3.9 MeV, beta emitters above 0.1 MeV Is the detection limit for gross alpha particle activity and gross beta 3 pci/l and 4 pci/l, respectively? Is the sample counted with a precision of ± 100% at the 95% confidence level? Laboratory Facilities and Equipment Are bench surfaces of an impervious material covered with absorbent paper? Are trays constructed of stainless steel, plastic, or fiberglass and lined with absorbent paper? Gas-Flow Proportional Counter Make: Model: Year: Serial #: Is the gas-flow proportional counter in working order? Is the detector either a windowless (internal proportional counter) or a thin window type? Is each counting instrument properly grounded? Does each counting instrument have a regulated power supply, either internal or external? Are counting instruments not located in a room in which samples and standards are prepared or other types of chemical analyses are performed? Are counting pans, preferably stainless steel counting planchets, available? Is only Class A volumetric glassware used? Does the laboratory store radioactive Section 1.2 40 CFR 141.25 Table B and C 40 CFR 141.25 (c) 2i 2ii 6.3 (b) 2 6.3 (b) 2 1 1 1 SM 7110 B.2a 3.3 (a) 8i Page 2 of 6

standards, samples, and wastes in an 6.5(a)1 enclosed and properly labeled area? Does the laboratory store standards, samples, and radioactive wastes safely in containers that protect against flammability and against contamination of the laboratory? 6.5(a)1 Does the laboratory prepare standards and samples in an area specifically designated for and exclusively used for the preparation of radioactive standards and samples? Does the laboratory monitor the work area for radioactivity? Gas-Flow Proportional Counter Calibration Are the initial instrument calibration procedures included or referenced in the method SOP? Is sufficient raw data records retained to permit reconstruction of the initial instrument calibration? Are samples results quantitated from the initial instrument calibration, not from any continuing instrument calibration verification? Are performance checks for detection efficiency performed on a day-of-use basis (if a batch of samples lasts for more than one day then check may be performed at beginning and end as long as time interval is no greater than one week)? Are background measurements performed on at least a weekly basis? Does the laboratory maintain a written procedure for monitoring radiation measurement instrumentation for radioactive contamination? Does the procedure indicate the frequency of the monitoring and indicate criteria which initiate corrective action? Standards and Reagents Are all reagents used analytical reagent-grade or better? Does the source of water meet the required standard of quality for the method? Is the source of water monitored daily and a record kept? Does distilled or deionized water have a resistance value between 0.5 and 2.0 megohms/cm at 25 C? Are reference standards obtained from 6.5(a)2 6.5(a)2 1.7.1(a)i SM 7110B 1.7.1(a)ii 1.7.1(a)iii SM7110 B 1.7.1(b)iii 1.7.1(c)iii 1.7.1(d) 1.7.1(d) 1.7.2.5(a) 3.3 (a) 9 1.7.2.5(b) Section 6.1 Page 3 of 6

and/or traceable by NIST? 1.7.2.5(c) Is the laboratory s tap water used to prepare efficiency standards, or is a surrogate water matrix used? 7 Technical Notes Is 230 Th used to calibrate alpha activity? 6.4 Is 90 Sr/ 90 Y used to calibrate beta activity? Are reference standards accompanied by a certificate of calibration? Are all reagents and reagent solutions properly labeled? a) Identity b) Titer c) Strength or concentration d) Recommended storage requirements e) Preparation f) Expiration date g) Any other pertinent information Does the laboratory immediately discard any reagent or reagent solution that is past its expiration date? Demonstration of Capability (DOC) Has the laboratory conducted an initial DOC prior to using any method and at any time there is a change in instrument type, personnel, or method or any time it hasn t been performed in a 12 month period? Does the laboratory have a documented procedure for on-going DOC? Quality Control Is a method blank analyzed per preparation batch (max of 20 samples)? Is there no subtraction of the method blank from the sample results? Is a LCS analyzed at a minimum of one per preparation batch? Is the activity of the LCS at least 10 times the MDA and at a level comparable to that of routine samples? Is the LCS prepared with similar aliquot size to that of routine samples? How frequently are matrix spikes run? Is the activity of matrix spike analytes greater than 5 times the Minimum Detectable Activity (MDA)? Are duplicates performed on replicate aliquots of actual samples? Each day, are 10% duplicate analyses performed? (For Radon) (a)3i 7.2 TNI V1M3 1.7.6.1(c) 6.5(a)3 6.5(a)4 1.6.2 1.6.3.1 1.7.2.1(b) 1.7.2.1(c) 1.7.2.2(b) 1.7.2.2(e) 1.7.2.2(i) 1.7.2.3(a)v 1.7.2.3(b)iii 6.6(c)3 Page 4 of 6

Are the precision and accuracy assessed and expressed as percent 1.7.3.3(a)i recovery (%R) and relative percent difference (RPD)? Is corrective action taken when %R and RPD are outside established criteria? 1.7.3.3(a)ii Procedures Is a volume of sample containing no more than 100 mg residue used for alpha examination, or 200 mg for beta examination? If water samples are known or suspected to contain chloride salts, are those chloride salts converted to nitrate salts before the sample residue is transferred to a stainless steel planchet? Once the sample has been dried, is it cooled in a desiccator, weighed and kept dry until counted? What is the drying procedure/time for the samples/planchets? Is the initial counting of the plancheted sample initiated between 36 to 48 hr from the time of sample collection? Is the date and time of the initial and final counts recorded? If the calculated value from the initial gross alpha count is 5 pci/l, is that value reported (no further radiochemical analysis is required)? If the calculated value from the initial gross alpha count exceeds 5 pci/l, is that sample plancheted sample recounted between 20 to 28 hr after the initial count and this calculated value reported as the final result? Are the planchted samples stored in a desiccator prior to counting Proficiency Testing (PT) Does the laboratory obtain performance efficiency (PE) samples from USEPA s radiological PT program, or from the Department s designated PT program? Does the laboratory participate in the appropriate field of proficiency testing (FoPT) for initial and continued accreditation? Is PT analyses not contracted out to another laboratory? Does the laboratory report results for accreditation on or before the closing date of the study? SM 7110 B.4a.1 Section 8.1 SM 7110 B.4a.3 6.4(a)3ii 6.4(a)3iii 6.4(a)3iv 8.4 2.13(e) 4.0 5.1.2 5.2.2 Page 5 of 6

Does the laboratory have plans for resolving analytical problems when it receives a not acceptable performance score from a Proficiency Testing Provider (PTP)? 6.1 Does the laboratory have two consecutive, acceptable PT scores out of three within the last 18 months? Reports Are the reports complete and accurate? Are there any data qualifiers on reports? Does the report have authorized signatures? Is the following information included in the report: a) Identity of sample b) Sampling station c) Date of collection d) Volume of sample e) Type of test f) Time of counting g) Alpha activity in pci/l h) Beta activity in pci/l (if applicable) i) Counting error(for alpha and beta results) 4.1.3 5.10.2 5.10.2(g) 5.10.2(j) 6.7(f) Records Does the laboratory establish and maintain procedures for identification, collection, indexing, access, filing, storage, maintenance and disposal of quality and technical records? Do quality records include reports from internal audits and management reviews and records of corrective and preventive actions? Does the laboratory retain records concerning analyses including raw data records, quality control data records, chain-of-custody forms, and laboratory reports for at least five years? 4.13.1.1 4.13.1.1 6.7(a) Page 6 of 6