Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Similar documents
Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

PHONE: (303) FAX: (303) WEBSITE:

LM-79 Report TEST REPORT NO.M037D

LM-79 Report TEST REPORT NO. L523A

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

Checked X.CAO

Checked X.CAO

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl5s50u1

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl4s50u1

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA.

Cat# LED High Bay

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A)

Morris Products 53 Carey Road, Queensbury, NY 12804

March 26, Title: TEMPO 21 Report. Prepared for: Sviazinvest, OJSC. Prepared by: Cree Durham Technology Center (DTC) Ticket Number: T

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

Page 2 of 14 Report No.: Test sample... : Type of test object... : LED Downlight LED Downlight

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es)

NVLAP LAB CODE LM Test Report. For. Shenzhen Guanke Technologies Co. Ltd. (Brand Name: )

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

LED Entryway Light RoHS

Cat# LED Linear High Bay 100W

NVLAP LAB CODE LM Test Report. For SIMKAR CORPORATION. (Brand Name: ) LED Lamp. Model name(s): SALXS350U1. Model differences: N/A

NVLAP LAB CODE LM Test Report. For SIMKAR CORPORATION. (Brand Name: ) LED Lamp. Model name(s): SALXS450U1. Model differences: N/A

Model No Stock No Calibration Due Date Chroma Programmable AC Source PS-AC02

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

Test Report. Report No.: EED35I Page 1 of 12. Customer : Revolution Lighting Technologies : 4139 Guardian Street, Simi Valley, CA93063

GM LIGHTING 9830 W 190th St, Torrance, CA 90503

Photometric Test Report

Photometric Test Report

Maxlite Jersey Blvd Rancho Cucamonga, CA Electrical and Photometric tests as required by the IESNA test standards.

Transcription:

Report Number: Model: RESTL-2013-0012 E-HL6S11CSZ Date: 11/08/2013 Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report Prepared For: Joshua Krohn Cree, Inc 9201 Washington Avenue Racine, WI 53406 Prepared By: Approved By: Linjie Li, Photometric Test Technician Tim Henning, Specialized Testing Manager C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 1 of 10 This report contains data that are not covered by the NVLAP accreditation.

Product Information Manufacturer Model Number (SKU) Serial Number LED Type e-conolight E-HL6S11CSZ PL02124-001 CXA2530 5000K Product Description Cast black painted finned metal housing, four circuit boards each with one led array, two-piece molded plastic reflector with specular finish and one conical aperture per led array and textured bottom mounting surface, clear flat glass lens with black painted interior and one aperture per led in cast black painted metal lens frame. Driver Information (Where Applicable) BLL040FMA-042-0850C Aperture Size Length Width 6 3/4 15 7/8 10 1/2 Sample The following sample was submitted for evaluation C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 2 of 10

Key Photometric Data Sphere Output Goniophotometer Luminous Flux 11743 11382.1 lm Efficacy 81.38 79.21 lm/w Correlated Color Temperature (CCT) 5178 K Color Rendering Index (CRI) 72.6 R 9-21.0 Duv 0.001223 S/P Ratio* 1.80 Sphere Goniophotometer Electrical Measurements 120V 277V 120V Input Wattage 144.3 143.94 143.7 W Input Current 1.211 0.556 1.215 A Input Voltage 120.06 276.85 119.94 V Power Factor 0.992 0.935 0.987 Off-State Power 0 0 0 W Total Harmonic Distortion (Voltage) 0.16 0.29 0.20 % Total Harmonic Distortion (Amperage) 12.49 19.75 12.45 % Luminous Intensity Distribution Goniophotometer Center Beam Candlepower (Max Candela) 40162.4 Cd Beam Angle 19.9 Field of View 65.8 Zonal Lumens (0 30 ) 8392.7 (73.7) lm (%) Zonal Lumens (0 40 ) 10652.0 (93.6) lm (%) Zonal Lumens (0 60 ) 11365.0 (99.8) lm (%) Zonal Lumens (60 90 ) 17.1 (0.2) lm (%) Key Test Parameters Sphere Output Goniophotometer Stabilization Time 85 81 min Total Operating Time (Stabilization + Test) 86 101 min Ambient Temperature 25.7 25.2 C C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 3 of 10

CCy Chromaticity Coordinates NVLAP Lab Code 500089-0 x y u v u v Duv 0.3403 0.3501 0.2088 0.3222 0.2088 0.4833 0.0012 Color Rendering Index Details Ra R1 R2 R3 R4 R5 R6 R7 R8 R9 R10 R11 R12 R13 R14 72.6 70.8 77.0 79.6 73.3 71.1 67.4 81.3 60.0-21.0 43.5 69.4 42.3 71.4 88.2 Chromaticity Diagram 0.43 3000 K 2700 K 0.41 4000 K 3500 K 0.39 4500 K 0.37 5700 K 5000 K 0.35 6500 K 0.33 5178 0.31 0.29 0.28 0.33 0.38 0.43 0.48 CCx C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 4 of 10

Radiant Flux (W) Spectral Distribution NVLAP Lab Code 500089-0 λ(nm) W/nm λ(nm) W/nm λ(nm) W/nm 360 0.000490 530 0.172151 700 0.022181 370 0.000342 540 0.186283 710 0.017029 380 0.000104 550 0.192363 720 0.013076 390 0.000162 560 0.195535 730 0.009741 400 0.000708 570 0.195852 740 0.007172 410 0.005676 580 0.191848 750 0.005430 420 0.026918 590 0.182704 760 0.004067 430 0.083244 600 0.168542 770 0.003023 440 0.199605 610 0.151221 780 0.002245 450 0.306788 620 0.131804 790 0.001667 460 0.169065 630 0.112270 800 0.001235 470 0.078572 640 0.093215 810 0.000918 480 0.041142 650 0.075985 820 0.000642 490 0.037246 660 0.061022 830 0.000351 500 0.060729 670 0.047984 510 0.103309 680 0.037349 520 0.144937 690 0.028946 0.35 0.3 0.25 0.2 0.15 0.1 0.05 0 Dominant Wavelength 568 nm Peak Wavelength: 449 nm Spectral Power Distribution (W/nm) Wavelength (nm) C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 5 of 10

Zonal Lumen Summary NVLAP Lab Code 500089-0 Zone Lumens % of Total Zone Lumens % of Total 0-5 825.1 7.2% 90-95 0 0% 5-10 1,785.7 15.7% 95-100 0 0% 10-15 1,754.2 15.4% 100-105 0 0% 15-20 1,501.1 13.2% 105-110 0 0% 20-25 1,322.4 11.6% 110-115 0 0% 25-30 1,204.1 10.6% 115-120 0 0% 30-35 1,194.8 10.5% 120-125 0 0% 35-40 1,064.5 9.4% 125-130 0 0% 40-45 554.6 4.9% 130-135 0 0% 45-50 130.1 1.1% 135-140 0 0% 50-55 20.3 0.2% 140-145 0 0% 55-60 8.1 0.1% 145-150 0 0% 60-65 5.8 0.1% 150-155 0 0% 65-70 4.0 0.0% 155-160 0 0% 70-75 3.1 0.0% 160-165 0 0% 75-80 2.5 0.0% 165-170 0 0% 80-85 1.5 0.0% 170-175 0 0% 85-90 0.2 0.0% 175-180 0 0% Total 11382.1 lm 100% Spacing Criteria Spacing Criterion (0-180) 0.34 Spacing Criterion (90-270) 0.34 Spacing Criterion (Diagonal) 0.40 C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 6 of 10

Candela Plots NVLAP Lab Code 500089-0 Isofootcandle Plot C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 7 of 10

Candela Tabulations 0 0 0 0 0 0 40162 37 3403 73 6 109 0 145 0 1 39445 38 3110 74 6 110 0 146 0 2 37762 39 2767 75 5 111 0 147 0 3 35466 40 2402 76 5 112 0 148 0 4 33018 41 2030 77 5 113 0 149 0 5 30700 42 1660 78 5 114 0 150 0 6 28591 43 1308 79 4 115 0 151 0 7 26672 44 993 80 4 116 0 152 0 8 24720 45 731 81 4 117 0 153 0 9 22433 46 526 82 3 118 0 154 0 10 19958 47 389 83 3 119 0 155 0 11 17578 48 217 84 2 120 0 156 0 12 15622 49 99 85 1 121 0 157 0 13 13973 50 74 86 1 122 0 158 0 14 12538 51 65 87 0 123 0 159 0 15 11329 52 62 88 0 124 0 160 0 16 10283 53 36 89 0 125 0 161 0 17 9442 54 25 90 0 126 0 162 0 18 8720 55 21 91 0 127 0 163 0 19 8100 56 20 92 0 128 0 164 0 20 7526 57 18 93 0 129 0 165 0 21 6983 58 16 94 0 130 0 166 0 22 6506 59 15 95 0 131 0 167 0 23 6077 60 15 96 0 132 0 168 0 24 5700 61 13 97 0 133 0 169 0 25 5370 62 12 98 0 134 0 170 0 26 5082 63 11 99 0 135 0 171 0 27 4831 64 11 100 0 136 0 172 0 28 4625 65 10 101 0 137 0 173 0 29 4467 66 9 102 0 138 0 174 0 30 4337 67 8 103 0 139 0 175 0 31 4211 68 8 104 0 140 0 176 0 32 4107 69 7 105 0 141 0 177 0 33 4013 70 7 106 0 142 0 178 0 34 3916 71 6 107 0 143 0 179 0 35 3803 72 6 108 0 144 0 180 0 36 3639 C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 8 of 10

Integrating Sphere Equipment List Description Manufacturer Model Serial Number 2M Sphere Everfine 2M 1004156T CCD Array Spectrometer Gooch & Housego OL 770VIS 10413180 Programmable AC Source Adaptive FC200 2280220 Power Analyzer Chroma 66202 66202 0005018 Goniophotometer Equipment List Description Manufacturer Model Serial Number AC Power Source Elgar CW1251P 1248A02602 Type C Goniophotometer LSI / UL 6440T 6440PN2028 Spectroradiometer Gooch & Housego 770VIS/NIR 12415189 Power Meter Yokogawa WT210 91M945458 Test Methods Used: Title ANSI C82.77:2002 CIE Pub. 13.3:1995 CIE Pub. 15:2004 IES LM-58:1994 IES LM-65:2001 IES LM-79:2008 Description Harmonic Emission Limits- Related Power Quality Reqt s for Lighting Equipment Method of Measuring and Specifying Color Rendering of Light Sources Colorimetry Spectroradiometric Measurements Single-Ended Compact Fluorescent Lamps Life Test Performance Electrical and Photometric Measurements of Solid-State Lighting Products Reference Standard Used: Equipment 2m Sphere Type C Goniophotometer and Spectrometer Description Tungsten Halogen Omni-Directional 75W Calibration Lamp, Serial Number F119 Tungsten Halogen Omni-Directional 500W Calibration Lamp, Serial Numbers 13C069. 13C070, 13C071. For color calibration of spectrometer, 13C074. C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 9 of 10

Disclaimers: This report must not be used by the customer to claim product certification, approval or endorsement by NVLAP, NIST or any agency of the federal government. The results contained in this report pertain only to the tested sample. This report shall not be reproduced, except in full, without written approval of the RESTL. Items marked with an asterisk are not covered by the NVLAP accreditation. In the event that the recorded temperature is outside of 25 + 1 C, this is considered a non-standard condition. Some areas of this report may not explicitly call out the equipment used to produce the data. The following list identifies the equipment and associated data tables or figures that may appear in this report but are not clearly identified. Integrating Sphere: Chromaticity Coordinates, Color Rendering Index Details, Chromaticity Diagram, Spectral Power Distribution, Dominant Wavelength and Peak Wavelength. Goniophotometer: Zonal Lumen Summary, Spacing Criteria, Candela Plot, Isofootcandle Plot and Candela Tabulations. In the event that testing is subcontracted out to an external laboratory, the associated information that appears in the equipment list and report items associated with that equipment are the actual equipment used, and test results produced, by that laboratory. Additional Comments: C r e e, I n c 9 2 0 1 W a s h i n g t o n A v e n u e R a c i n e, W I 5 3 4 0 6 Page 10 of 10