Time-Resolved μ-xrf and Elemental Mapping of Biological Materials

Similar documents
PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

ELECTRIC FIELD INFLUENCE ON EMISSION OF CHARACTERISTIC X-RAY FROM Al 2 O 3 TARGETS BOMBARDED BY SLOW Xe + IONS

IMPROVEMENT OF DETECTION LIMITS OF A PORTABLE TXRF BY REDUCING ELECTRICAL NOISE

ABNORMAL X-RAY EMISSION FROM INSULATORS BOMBARDED WITH LOW ENERGY IONS

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

AEROSOL FILTER ANALYSIS USING POLARIZED OPTICS EDXRF WITH THIN FILM FP METHOD

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

DEVELOPMENT OF A NEW POSITRON LIFETIME SPECTROSCOPY TECHNIQUE FOR DEFECT CHARACTERIZATION IN THICK MATERIALS

USABILITY OF PORTABLE X-RAY SPECTROMETER FOR DISCRIMINATION OF VALENCE STATES

MEASUREMENT CAPABILITIES OF X-RAY FLUORESCENCE FOR BPSG FILMS

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

ANALYSIS OF LOW MASS ABSORPTION MATERIALS USING GLANCING INCIDENCE X-RAY DIFFRACTION

A COMPACT X-RAY SPECTROMETER WITH MULTI-CAPILLARY X-RAY LENS AND FLAT CRYSTALS

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

INTERNATIONAL JOURNAL OF SCIENTIFIC & TECHNOLOGY RESEARCH VOLUME 3, ISSUE 2, FEBRUARY 2014 ISSN

INFLUENCE OF GROWTH INTERRUPTION ON THE FORMATION OF SOLID-STATE INTERFACES

CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE

Horst Ebel, Robert Svagera, Christian Hager, Maria F.Ebel, Christian Eisenmenger-Sittner, Johann Wernisch, and Michael Mantler

GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

POLYCAPILLARY OPTICS BASED NEUTRON FOCUSING FOR SMALL SAMPLE NEUTRON CRYSTALLOGRAPHY

SYNCHROTRON X-RAY MICROBEAM CHARACTERIZATION OF SMECTIC A LIQUID CRYSTALS UNDER ELECTRIC FIELD

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES

IMPROVING THE ACCURACY OF RIETVELD-DERIVED LATTICE PARAMETERS BY AN ORDER OF MAGNITUDE

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

ULTRATHIN LAYER DEPOSITIONS A NEW TYPE OF REFERENCE SAMPLES FOR HIGH PERFORMANCE XRF ANALYSIS

ION-EXCHANGE FILMS FOR ELEMENT CONCENTRATION IN X-RAY FLUORESCENCE ANALYSIS WITH TOTAL REFLECTION OF THE PRIMARY BEAM.

VORTEX A NEW HIGH PERFORMANCE SILICON DRIFT DETECTOR FOR XRD AND XRF APPLICATIONS

XUV 773: X-Ray Fluorescence Analysis of Gemstones

QUANTIFICATION OF MERCURY, LEAD, AND CADMIUM IN AQUEOUS SOLUTIONS BY ENERGY K X-RAY FLUORESCENCE SPECTROSCOPY

NEW X-RAY DETECTORS FOR XRF ANALYSIS. Jan S. Iwanczyk & Bradley E. Patt Photon Imaging, Inc., Northridge, CA 91324

Electron-Induced X-Ray Intensity Ratios of Pb Lα/Lβ and As Kα/Kβ by kev Applied Voltages

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

STRESS ANALYSIS USING BREMSSTRAHLUNG RADIATION

Application of X-ray Spectrometry at X-ray Absorption Edges for Investigation of Human Albumin

SILICON DRIFT DETECTORS FOR HIGH RESOLUTION, HIGH COUNT RATE X-RAY SPECTROSCOPY AT ROOM TEMPERATURE

Micro-XRF excitation in an SEM

Micro-XRF: Principles, Methodology and Applications in CH Studies. Birgit Kanngießer

COMPARISON OF THREE UNIVERSAL CURVES FOR THE ESCAPE PROBABILITY OF X-RAY EXCITED ELECTRONS - I. THEORY

MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT

QUANTITATIVE ENERGY-DISPERSIVE ELECTRON PROBE X-RAY MICROANALYSIS OF INDIVIDUAL PARTICLES

THE IMPORTANCE OF THE SPECIMEN DISPLACEMENT CORRECTION IN RIETVELD PATTERN FITTING WITH SYMMETRIC REFLECTION-OPTICS DIFFRACTION DATA

RADIOISOTOPE RH-101 AS X-RAY SOURCE FOR INSTRUMENTS ON SPACE MISSIONS

Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol

A MODIFIED APPROACH TO HOMOGENEITY TESTING AT MICROSCALE

Available online at ScienceDirect. Energy Procedia 71 (2015 ) Yokohama, Kanagawa, Japan

BORON IN GLASS DETERMINATION USING WDXRF

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

OBSERVATION OF SURFACE DISTRIBUTION OF PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY DURING D 2 GAS PERMEATION THROUGH PD COMPLEXES

print first name print last name print student id grade

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

In Situ High-Temperature Study Of Silver Behenate Reduction To Silver Metal Using Synchrotron Radiation

Mining Analyzer XRF-3360

APPLICATION OF MICRO X-RAY FLUORESCENCE SPECTROMETRY FOR LOCALIZED AREA ANALYSIS OF BIOLOGICAL AND ENVIRONMENTAL MATERIALS

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

Student Projects for

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

ACCURATE QUANTIFICATION OF RADIOACTIVE MATERIALS BY X-RAY FLUORESCENCE: GALLIUM IN PLUTONIUM METAL

Depth Distribution Functions of Secondary Electron Production and Emission

AP5301/ Name the major parts of an optical microscope and state their functions.

Quantum e$ciency of the CCD camera (XIS) for the ASTRO-E mission

FINDING DESCRIPTORS USEFUL FOR DATA MINING IN THE CHARACTERIZATION DATA OF CATALYSTS

X-RAY MICRODIFFRACTION STUDY OF THE HALF-V SHAPED SWITCHING LIQUID CRYSTAL

MONTE-CARLO MODELING OF SILICON X-RAY DETECTORS

Portable type TXRF analyzer: Ourstex 200TX

ED(P)XRF: SCREENING ANALYSIS AND QUANTITATIVE ANALYSIS with POLARIZED

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

Latest advances in identifying mineral composition variation by the M4 TORNADO AMICS

PERFORMANCE OF A BOREHOLE XRF SPECTROMETER FOR PLANETARY EXPLORATION

Overview of X-Ray Fluorescence Analysis

Detectors in Nuclear and High Energy Physics. RHIG summer student meeting June 2014

Issues With TXRF Angle Scans and Calibration

Repeatability of Spectral Intensity Using an Auger Electron Spectroscopy Instrument Equipped with a Cylindrical Mirror Analyzer

An Analysis of Secondary Enhancement Effects in Quantitative XRFA

object objective lens eyepiece lens

X-ray CCD calibration system using #uorescent lines

SYSTEM FOR PIXE TRACE ELEMENT ANALYSIS AT MTF. M. Beňo, J. Dobrovodský, D. Vaňa, S. Minárik

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

Generation of surface electrons in femtosecond laser-solid interactions

AN EXAFS STUDY OF PHOTOGRAPHIC DEVELOPMENT IN THERMOGRAPHIC FILMS

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Laser heating of noble gas droplet sprays: EUV source efficiency considerations

X Rays & Crystals. Characterizing Mineral Chemistry & Structure. J.D. Price

RIETVELD REFINEMENT WITH XRD AND ND: ANALYSIS OF METASTABLE QANDILITE-LIKE STRUCTURES

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

Spin-resolved photoelectron spectroscopy

HOW TO APPROACH SCANNING ELECTRON MICROSCOPY AND ENERGY DISPERSIVE SPECTROSCOPY ANALYSIS. SCSAM Short Course Amir Avishai

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

CHEM-E5225 :Electron Microscopy X-Ray Spectrometry

MC simulation of a PGNAA system for on-line cement analysis

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

Transcription:

296 Time-Resolved μ-xrf and Elemental Mapping of Biological Materials K. Tsuji 1,2), K. Tsutsumimoto 1), K. Nakano 1,2), K. Tanaka 1), A. Okhrimovskyy 1), Y. Konishi 1), and X. Ding 3) 1) Department of Applied Chemistry; Graduate School of Engineering; Osaka City University; Abstract 3-3-138 Sugimoto; Sumiyoshi-ku, Osaka 558-8585; Japan Telephone/Fax: +81-(0)6-6605-3080; E-mail: tsuji@a-chem.eng.osaka-cu.ac.jp 2) JST-PREST; 4-1-8 Honcho Kawaguchi, Saitama; Japan 3) Institute of Low Energy Nuclear Physics; Beijing Normal University; No. 19, Xin Jie Kou Wai Da Jie; Beijing 100875; China Telephone: 62207171; E-mail: xld@ht.rol.cn.net We have developed a micro-xrf instrument using a polycapillary x-ray lens. In addition, a time-resolved MCA (multi channel analyzer) was installed for x-ray energy analysis. This micro-xrf system with the time-resolved MCA enables continuous x-ray analysis as a function of time at localized region on the sample. After the micro x-ray beam was fixed on the quinoa seed, x-ray fluorescence intensities of several elements emitted from the quinoa were monitored. Furthermore, elemental distributions in biological materials, such as wheat and rice grains were investigated by the micro-xrf method. INTRODUCTION In x-ray fluorescence analysis, a non-destructive elemental analysis is possible at ambient air pressure. This is a unique advantage because other instrumental analytical methods such as EPMA require special sample preparation and high vacuum condition. In addition, XRF analysis at small region, that is, micro-xrf (μ-xrf), has been possible by applying recent technological advances. For example, a polycapillary x-ray lens [1-6] is useful for obtaining micro x-ray beam in laboratory. We have studied applications of μ-xrf with polycapillary x-ray lens [7,8]. One of interesting samples is in vivo biological material such as seed and plants. It is difficult to measure them by other micro analytical methods, because they can not be usually put into a vacuum chamber, and the damage due to electron beam is a serious problem for the sample. XRF at ambient air pressure will be the best method for them. Rice, wheat and quinoa seeds are important grains to as nutrient in the human body. Elemental distributions of minerals in grains give us useful information for processing of food materials besides basic research on nutrition and botany. The minerals in grains have been analyzed by atomic spectrometry such as ICP-AES (-MS) and AAS (atomic absorption spectroscopy) [9]. However, only averaged composition is obtained by these methods. In addition, sample preparation for these atomic spectroscopic methods is a time-consuming process and requires a skill for sample preparation. On the other hand, μ-xrf directly gives us elemental distributions without any special sample preparation.

This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website www.dxcicdd.com ICDD Website - www.icdd.com

297 EXPERIMENTAL DETAILS Figure 1 shows our experimental setup for μ-xrf [7,8]. A polycapillary x-ray lens (made by Beijing Normal University) was attached to an x-ray tube (Mo target). The focus size of x-ray emission in the x-ray tube was 1 x 10 mm. Finally, a spot size of about 80 μm was obtained at a focal output distance of 15 mm from the end of the lens. A whole grain sample was put on an X-Y-Z sample stage, which was controlled by a motor controller and a computer. A CCD camera was used for observation and confirmation of the position of the sample. A silicon drift detector (SDD; X-Flash Detector Type 1201, RÖNTEC, Germany; Sensitive area: 10 mm 2, Energy resolution <150 ev, FWHM at 5.9 kev) was used for detection of x-ray fluorescence. The distance between the sample and the SDD was less than 30 mm. A time resolved multi-channel analyzer (time-resolved MCA; NT-2400M, Laboratory Equipment Co., Japan) was used to obtain the time-dependent XRF intensity. By using this equipment, measuring life time, interval time and total time are controlled and x-ray fluorescence area-intensities (ROI: range of interest) are automatically saved by computer. The cotton and water shown in Fig.1 were used for observation of growth process of the quinoa seed. All measurements were performed at ambient air pressure. CCD X-ray tube Polycapillary X-ray lens Quinoa Cotton Water Si Drift Detector Sample stage Time resolved MCA Motor controller Computer Fig. 1 Experimental setup for μ-xrf.

298 RESULTS AND DISCUSSIONS First, a black wheat grain was measured to know the distribution of minerals in the grain. Figure 2 shows element maps for K,,, and. X-ray generator was operated at a tube voltage of 30 kv and a tube current of 20 ma. One pixel was 100 x 100 μm, and preset time for each pixel was 10 s. Each element showed different distributions. For example, K and distributed on the grain coat, while, and dominantly existed at embryo. On the edge surface near the SDD (right side of each element map in Fig. 2), excited x-ray fluorescence was not absorbed by sample itself; therefore, high x-ray fluorescence intensity is usually observed compared with XRF intensity at other positions. 3mm K Fig. 2 Element maps obtained for black wheat. 2 mm K

299 Fig. 3 Element maps obtained for buckwheat (soba wheat). Figure 3 shows the elemental distributions obtained for buckwheat (soba wheat). Similar result was obtained, although K, and distributed on the coat while dominantly distributed on the embryo. Figs. 2 and 3 indicate that μ-xrf is useful for elemental mapping of biological samples such as grains. Next, a quinoa sample was measured by using a time-resolved MCA. As shown in Fig.1, the quinoa seed was put on the cotton using double sides tape. The end of the cotton was put into the water (pure water; Milli-Q); therefore, the seed could be germinated. We measured the x-ray fluorescence from the seed in the growth process.x-rays 2 mm X-rays Fig. 4 Growth of quinoa seed and micro x-ray beam was fixed at germinated region. Figure 4 shows the growth of the quinoa seed. After germination of the seed, micro x-ray beam was fixed at the germinated region (stem). The time-dependence of XRF intensities was monitored. 52500 50000 Intensity /counts 47500 45000 2500 Cu 0 S -2 0 2 4 6 8 10 12 14 16 18 Time / hour

Fig. 5 Time dependence of x-ray intensities of each element, measured at stem of the quinoa seed without nutrition solution. 300 Figure 5 shows the time dependence of x-ray intensities of K,,,, P, S and. A preset time of 100 s and an interval time of 200 s were applied. During the growth in 16 hours, no change in the time dependent intensities was observed in the stem region. Intensity /counts 20000 15000 10000 5000 K K 0 0 1 2 3 4 5 6 Time / hour Fig. 6 Time dependence of x-ray intensities of each element at root of the quinoa seed. A nutrition solution was put into the water at 2 hours after starting the measurement. Figure 6 shows the time dependent XRF intensities (K lines) during the growth at the root of the quinoa. A nutrition solution is used to fasten the growth of the plant. The nutrition solution, including N, P, Mg,, Cu, and Mo, was added to the water at 2 hours after starting the measurement. Before adding the nutrition solution, K signal decreased. However, after adding the nutrition solution, the decease of K signal stopped and slightly increased. In contrast, signal increase until 2 hours; however, it deceased after the nutrition solution was added. The reason of these trends of x-ray intensities is not clear; however, it was demonstrated that the time-resolved MCA is useful to acquire the time dependence of x-ray fluorescence. CONCLUSIONS A micro XRF instrument was developed using a polycapillary x-ray lens. In addition, a time-resolved MCA, which enables continuous x-ray analysis as a function of time, was developed. Using this analytical system, we studied the growth of a quinoa seed. After the micro x-ray beam was fixed on the Quinoa sample, x-ray fluorescence intensities of several elements emitted from the quinoa were monitored. It was demonstrated that the time-resolved MCA is useful to acquire the time dependence of x-ray fluorescence. It is expected that the

time-resolved MCA is useful for studying and monitoring the compositional changes during growth processes as well as environmental monitoring of harmful elements. 301 REFERENCES [1] Kumakhov, M.A., Nucl. Instrum. Meth., 1990, B48, 283. [2] Yiming, Y.; Xunliang, D., Nucl. Instrum. Meth. Phys. Res., 1993, B52, 126. [3] Wang, L.; Rath, B. K.; Gibson, W. M.; Kimball, J. C.; MacDonald, C. A., J. Appl. Phys., 1996, 80, 3628. [4] Gao, N; Ponomarev, I. Yu.; Xiao, Q. F.; Gibson, W. M.; rpenter, D.A., Appl. Phys. Lett., 1996, 69, 1529. [5] Adams, F.; Janssens, K.; Snigirev, A., J. Anal. At. Spectrom., 1998, 13, 319. [6] Gao, N; Ponomarev, I.Y., X-Ray Spectrom., 2003, 32, 186-194. 132-135. [7] Emoto, T.; Sato, Y.; Konishi, Y.; Ding, X.; Tsuji, K., Spectrochim. Acta B, 2004, 59, 1291. [8] Tsuji, K.; Emoto, T.; Matsuoka, M.; Miyatake, Y.; Nagamura, T.; Ding, X., Powder Diffraction, 2005, 20, 1. [9] Borjesson, J.; Mattsson, S.; X-ray Fluorescence Analysis in Medical Sciences, in X-Ray Spectrometry: Recent Technological Advances edited by Tsuji, K.; Injuk, J.; Van Grieken, R.; John Wiley & Sons (2004).